Iso 2360-2017
Iso 2360-2017
STANDARD 2360
Fourth edition
2017-07
Reference number
ISO 2360:2017(E)
© ISO 2017
ISO 2360:2017(E)
Contents Page
Foreword ........................................................................................................................................................................................................................................ iv
1 Scope ................................................................................................................................................................................................................................. 1
2 Normative references ...................................................................................................................................................................................... 1
3 Terms and definitions ..................................................................................................................................................................................... 1
4 Principle of measurement .......................................................................................................................................................................... 2
5 Factors affecting measurement uncertainty ........................................................................................................................... 3
5 .1 B as ic influence o f the co ating thicknes s ........................................................................................................................... 3
5.2 Electrical properties of the base metal .............................................................................................................................. 3
5 .3 Geo metry: B as e metal thicknes s .............................................................................................................................................. 4
5 .4 Geo metry: E dge e ff .................................................................................................................................................................... 4
ects
5 .5 f
Geo metry: S ur ace curvature ...................................................................................................................................................... 4
5.6 Surface roughness ................................................................................................................................................................................ 4
5.7 Cleanliness: Lift-off effect .............................................................................................................................................................. 5
5.8 Probe pressure ........................................................................................................................................................................................ 5
5.9 Probe tilt ....................................................................................................................................................................................................... 5
5.10 Temperature effects ............................................................................................................................................................................ 5
5.11 Intermediate coatings ....................................................................................................................................................................... 6
5 .1 2 E xternal electro magnetic fields ................................................................................................................................................ 6
6 Calibration and adjustment of the instrument ..................................................................................................................... 6
6.1 General ........................................................................................................................................................................................................... 6
6.2 f
Thicknes s re erence s tandards ................................................................................................................................................. 6
6.3 Methods of adjustment .................................................................................................................................................................... 7
7 Measurement procedure and evaluation .................................................................................................................................... 8
7.1 General ........................................................................................................................................................................................................... 8
7.2 Number of measurements and evaluation ...................................................................................................................... 8
8 Uncertainty of the results ............................................................................................................................................................................ 8
8.1 General remarks .................................................................................................................................................................................... 8
8.2 Uncertainty o f the calib ratio n o f the ins trument ...................................................................................................... 9
8.3 Stochastic errors ................................................................................................................................................................................. 10
8.4 Uncertainties caus ed by f acto rs s ummarized in Clause 5 .............................................................................. 10
8.5 C o mb ined uncertainty, exp anded uncertainty and final res ult .................................................................. 11
9 Precision .................................................................................................................................................................................................................... 11
9.1 General ........................................................................................................................................................................................................ 11
9.2 Rep eatab ility ( r) .................................................................................................................................................................................. 11
9.3 Rep ro ducib ility limit ( R) .............................................................................................................................................................. 12
10 Test report................................................................................................................................................................................................................ 12
Annex A (informative) Eddy-current generation in a metallic conductor .................................................................. 14
Annex B (informative) Basics of the determination of the uncertainty of a measurement of
the used measurement method corresponding to ISO/IEC Guide 98-3 ................................................... 18
Annex C (informative) Basic performance requirements for coating thickness gauges which
are based on the amplitude-sensitive eddy-current method described in this document .. 20
Annex D (informative) Examples for the experimental estimation of factors affecting the
measurement accuracy............................................................................................................................................................................... 22
Annex E (informative) Table of the student factor .............................................................................................................................. 27
Annex F (informative) Example of uncertainty estimation (see Clause 8) ................................................................ 28
Annex G (informative) Details on precision ............................................................................................................................................... 30
Bibliography ............................................................................................................................................................................................................................. 34
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work o f preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters o f
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular the different approval criteria needed for the
di fferent types o f ISO documents should be noted. This document was dra fted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www.iso .org/directives).
Attention is drawn to the possibility that some o f the elements o f this document may be the subject o f
patent rights. ISO shall not be held responsible for identi fying any or all such patent rights. Details o f
any patent rights identified during the development o f the document will be in the Introduction and/or
on the ISO list of patent declarations received (see www.iso .org/patents).
Any trade name used in this document is in formation given for the convenience o f users and does not
constitute an endorsement.
For an explanation on the voluntary nature o f standards, the meaning o f ISO specific terms and
expressions related to con formity assessment, as well as in formation about ISO’s adherence to the
World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT) see the following
URL: www.iso .org/iso/foreword .html.
This document was prepared by Technical Committee ISO/TC 107, Metallic and other inorganic coatings.
This fourth edition cancels and replaces the third edition (ISO 2360:2003), which has been technically
revised.
2 Normative references
The following documents are re ferred to in the text in such a way that some or all o f their content
constitutes requirements o f this document. For dated re ferences, only the edition cited applies. For
undated re ferences, the latest edition o f the re ferenced document (including any amendments) applies.
ISO 2064, Metallic and other inorganic coatings — Definitions and conventions concerning the measurement
of thickness
3.1
adjustment of a measuring system
set o f operations carried out on a measuring system so that it provides prescribed indications
corresponding to given values o f a quantity to be measured
Note 1 to entry: Adjustment o f a measuring system can include zero adjustment, o ffset adjustment, and span
adjustment (sometimes called gain adjustment).
Note 2 to entry: Adjustment o f a measuring system should not be con fused with calibration, which is a
prerequisite for adjustment.
Note 3 to entry: A fter an adjustment o f a measuring system, the measuring system must usually be recalibrated.
Note 4 to entry: Colloquially, the term “calibration” is frequently, but falsely, used instead o f the term “adjustment”.
In the same way, the terms “verification” and “checking” are o ften used instead o f the correct term “calibration”.
[SOURCE: ISO/IEC Guide 99:2007, 3.11 (also known as “VIM”)]
3.2
calibration
operation that, under specified conditions, in a first step, establishes a relation between the quantity
values with measurement uncertainties provided by measurement standards and corresponding
indications with associated measurement uncertainties and, in a second step, uses this information to
establish a relation for obtaining a measurement result from an indication
Note 1 to entry: A calibration may be expressed by a statement, calibration function, calibration diagram,
calibration curve, or calibration table. In some cases, it may consist o f an additive or multiplicative correction o f
the indication with associated measurement uncertainty.
Note 2 to entry: Calibration should not be con fused with adjustment o f a measuring system, o ften mistakenly
called “sel f-calibration”, nor with verification o f calibration.
Note 3 to entry: O ften, the first step alone in the above definition is perceived as being calibration.
[SOURCE: ISO/IEC Guide 99:2007, 2.39 (also known as “VIM”)]
4 Principle of measurement
Eddy-current instruments work on the principle that a high frequency electromagnetic field generated
by the probe system o f the instrument will produce eddy-currents in the base metal beneath the
coating on which the probe is placed (see Figure 1). These induced currents cause a change of the
electromagnetic field surrounding the probe coil and there fore result in a change o f the amplitude
o f the probe coil impedance. The induced eddy-current density is a function o f the distance between
the generating coil and the base metal sur face. Consequently, this impedance change can be used as
a measure o f the thickness o f the coating on the conductor by means o f a calibration with re ference
standards (see also Annex A).
In order to measure a change o f the coil impedance amplitude, the test coil is usually part o f an oscillator
circuit with a resonant frequency determined by the coil inductance and resistance. A change of the coil
impedance amplitude results in a shi ft o f the resonant frequency. Consequently, the measured resonant
frequency is a measure o f the coating thickness. The values are either pre-processed by digital means
or are directly displayed on a use fully scaled gauge.
The probe and measuring system/display may be integrated into a single instrument.
NOTE 1 Annex C describes the basic performance requirements of the equipment.
NOTE 2 Factors a ffecting measurement accuracy are discussed in Clause 5.
Key
1 ferrite core of the probe 5 induced eddy- current
5 . 1 B a s i c i n fl u e n c e o f th e c o a ti n g th i c kn e s s
T he s en s itivity o f a prob e, i . e . the me as u rement e ffe c t, de cre as e s with i nc re as i ng th ickne s s with i n the
me as u rement range o f the prob e . I n the lower me as u rement range, th i s me as u rement uncer ta i nty (i n
ab s olute term s) i s con s tant, i ndep endent o f the co ati ng th ickne s s . T he ab s olute va lue o f th i s u ncer tai nty
dep end s on the prop er tie s o f the prob e s ys tem and the s ample materi a l s , e . g. the homo geneity o f the
b a s e me ta l conduc tivity, the b as e me ta l rough ne s s and the s a mple s ur face rough nes s . I n the upp er
me as u rement range, the u ncer ta i nty b e come s approxi mately a con s tant frac tion of the co ati ng
th ickne s s .
s ys tem and fre quenc y. C on s e quently, the b as e me ta l conduc tivity c au s e s the me a s urement e ffe c t for
th i s me tho d . T he relation sh ip b e twe en co ati ng th ickne s s a nd the me a s u re d va lue dep end s s trongly
on the conduc tivity of the base me ta l . C on s e quently, ca l ibration pro ce du re s and me a s urements
shall be made on the same material. Different materials with different conductivities as well as local
fluc tuation s o f the conduc tivity or variation s b e twe en d i fferent s ample s c an c au s e (more or les s) errors
In order to reduce the influence o f roughness, a calibration should be carried out with an uncoated base
metal with a roughness equivalent to the coated sample base metal.
I f necessary, the definition o f the average coating thickness that is used should be stated between the
supplier and client.
NOTE When compared with the phase-sensitive method o f ISO 21968, the amplitude-sensitive eddy-current
measurement can be more a ffected by base metal roughness.
NOTE 2 An indentation o f the probe tip into so ft coatings can be reduced by placing a protective foil with
known thickness onto the coated sur face. In this case, the coating thickness is the measured thickness minus
the foil thickness. This procedure is not applicable when measuring non-magnetic metallic coatings on non-
conductive base materials.
5.9 Probe tilt
Unless otherwise instructed by the manu facturer, the probe shall be applied perpendicularly to the
coating sur face as tilting the probe away from the sur face normal can cause measurement errors.
The risk o f inadvertent tilt can be minimized by the probe design or by the use o f a probe-holding jig.
NOTE Most commercially available instruments are equipped with spring loaded probes, which ensure a
perpendicular placement on the sample surface.
5.10 Temperature effects
As temperature changes a ffect the characteristics o f the probe, it should be used under approximately
the same temperature conditions as when the instrument was calibrated.
NOTE 1 The influence o f temperature variations can be reduced by a temperature compensation o f the probe.
The manu facturer’s specification is taken into account.
NOTE 2 Temperature differences between the probe, electronics of the instrument, environment and sample
can cause large thickness errors. One example is the thickness measurement o f hot coatings.
Most metals change their electrical conductivity with temperature. Because the measured coating
thickness is influenced by changes in the electrical conductivity o f the base metal, large temperature
changes should be avoided (see 5.2).
electrical characteristics of that intermediate coating differ from those of the coating or base metal. If
a diff ff
erence do e s exi s t, then the me a s urements wi l l, i n add ition, b e a e c te d b y an i nterme d i ate co ati ng
f
th ickne s s o tmin f les s tha n tmin , then the intermediate coating, if non-
. I the th ickne s s i s gre ater than
T he me as u rement re s u lts c an be i n fluence d by s trong ele c tromagne tic i nter feri ng field s . In cases
showi ng u ne xp e c te d re s u lts or a s trong variation o f re s u lts , wh ich c an no t b e e xplai ne d b y o ther fac tors ,
th i s i n fluence shou ld b e ta ken i nto accou nt. I n th i s s ituation, a comp a ri s on me as u rement s hou ld b e
6.1 General
P rior to u s e, ever y i n s tru ment s ha l l b e c a l ibrate d or adj u s te d accord i ng to the i n s tr uc tion s o f the
ge ome tr y, a nd s u r face prop er tie s o f the b as e me ta l us e d for c a l ibration or adj u s tment shou ld b e s i m i lar
to tho s e for the te s t s p e ci men s i n order to avoid deviation s c au s e d by the fac tors de s c rib e d i n Clause 5.
O ther wi s e, the s e i n fluence s sha l l b e con s idere d i n the e s ti mation o f the me a s u rement u ncer ta i nty.
During calibration or adjustment, the instruments, standards and base metal should have the same
temperature as the test specimens to minimize temperature induced differences.
I n order to avoid the i n fluence o f i n s tr ument d ri ft, p erio d ic control me a s u rements with re ference
standards or control samples are recommended. If required, the instrument has to be re-adjusted.
NO TE M o s t i n s tr u ments auto m atic a l l y adj u s t them s el ve s du r i n g a fu nc tio n c a l le d “c a l ib ration”, c a r r ie d out
s ta nda rd s and thei r as s o ci ate d u ncer tai ntie s s ha l l b e known a nd u nambiguou sly do c umente d . T he
s u r face are a for wh ich the s e va lue s are va l id s ha l l b e marke d . T he th ickne s s va lue s s hou ld b e trace able
T he uncer ta i ntie s s ha l l b e do c u mente d with thei r con fidence level, e . g. U (9 5 %) , i . e . the prob abi l ity, th at
the “tr ue” va lue i s with i n the rep or te d u ncer tai nty i nter va l arou nd the do c u mente d th ickne s s va lue, i s
minimum 95 %.
P rior to u s e, foi l s a nd co ati ngs are to b e che cke d vi s ua l ly for damage or me ch an ic a l we ar as th i s wou ld
cau s e an i ncorre c t adj u s tment a nd thu s s ys tematic devi ation o f a l l me as u rement va lues .
In most cases, the foil material is plastic. In contrast to the magnetic method (see ISO 2178), conductive
materia l s , e . g. copp er a l loys , c an no t b e u s e d b e c au s e i n s uch foi l s , e ddy c u rrents c an b e i nduce d . T hey
NOTE When measuring non-magnetic metallic coatings on non-conductive base materials, the situation is
“inverted”.
The use o f foils as re ference standards, compared to selected coated base metals, benefits from the
possibility o f placing the foils directly on each base metal. The geometry influence and other factors are
already considered within the adjustment.
However, by placing the probe on foils, elastic or plastic de formation may occur, which can a ffect the
measurement result. Moreover, any gap between the pole o f the probe, foil and base metal has to be
avoided. Especially for concave specimens, or i f the foil is wrinkled or bent, the usually low pressure o f
the spring loaded guiding sleeve o f the probe may not be su fficient to ensure there is no gap.
Possible elastic or even plastic deformation of a reference foil depends on the applied force of the
probe and the probe tip diameter (see 5.9). Consequently, the calibration of such reference foils should
be carried out with comparable values of the applied force and tip diameter to avoid indentation
di fferences during the probe calibration. In this way, respective indentation errors are already taken
into account in the foil thickness value, i.e. this value can be smaller than the una ffected geometric
thickness. The values o f both the applied force and the tip diameter used at the foil calibration should
be known from the re ference foil manu facturer so that possible thickness errors can be estimated.
is determined by the ratio, E, of the resulting difference, t m − t r , and the combined uncertainty o f the
calibration measurement. This uncertainty (denominator o f E, k = 2) is considered to be caused by the
stochastic error of the measurement with n repeats (compare to 8.3) and the given reference standard
uncertainty, Ur. In case of E ≤ 1, the calibration is valid and cannot be further improved by means o f this
re ference standard, i.e. the di fference cannot be distinguished from the uncertainty. There fore, the
standard uncertainty o f the calibration, ucal (k = 1), is given by the combined uncertainty o f the
verification measurement but with respect to the 1 sigma level (k = 1).
However, in the case of E > 1, a significant deviation of the calibration within the uncertainty is detected
and an adjustment o f the instrument should be carried out in order to improve the calibration accuracy.
See Formulae (2) and (3):
t r − tm
E=
2 × u cal
(2)
s (t )
2
u cal = t ( 68 , 27 %, n − 1 ) × m + 0 , 5 × U r
2
(3)
n
NOTE 1 In case the tolerance, T, of the reference standard is given instead of Ur, i.e. (tr ± T), for example in a
certificate o f a certified re ference material, the respective standard uncertainty (for 68,3 % confidence level) can
be calculated as U r = T and the expanded uncertainty (for 95,4 % confidence level) as U r ( k = 2 ) = 1 , 653 × T
3 3
. The deviation from the usual factor 2 for normal distribution is due to the fact that tolerances follow rectangular
distributions.
The calibration uncertainty ucal is only valid in a small thickness range around tr. In the case of a larger
thickness range o f interest, the uncertainty ucal should be estimated on both sides o f the thickness
range. The linear interpolation between both values gives the uncertainty o f interest as a function o f
the thickness.
Very o ften, the accuracy o f the calibration is limited by the given uncertainty o f the re ference standard,
as the uncertainty o f the calibration cannot be smaller than the uncertainty o f the re ference standard
used. In order to improve the calibration, a re ference standard with a smaller uncertainty is necessary.
Usually, a normalization or zeroing on an uncoated base metal is recommended by the manu facturer
at the beginning o f a measurement. The resulting uncertainty o f this normalization is considered to be
already included in ucal .
NOTE 2 t(68,27 %, n – 1): student factor (degrees of freedom f = n - 1 and level o f confidence with P = 68,27 %).
Respective values are summarized in Annex E.
8.3 Stochastic errors
In general, repeated measurements are recommended in order to improve the accuracy o f the arithmetic
mean value, t , of the thickness values measured (see 7.2 ), i.e. to reduce the uncertainty of the thickness
result. In the case of n repeated measurements, the standard uncertainty, usto (k = 1), of the arithmetic
mean, t , can be estimated by using Formula (4) (Type A):
u sto = t ( 68 , 27 %, n − 1 ) ×
()
s t
(4)
n
The standard uncertainty, u sto , is a measure of all errors arising from unpredictable or stochastic
temporal and spatial variations o f influence quantities.
NOTE 1 The standard uncertainty, u sto , can be reduced by increasing the number o f repeated measurements.
This can be important, e.g. in case of rough sample surfaces.
NOTE 2 Not all contributions to the uncertainty, u sto , are o f random nature (Type A). This depends on the
design o f the experiment. For example, the measured thickness o f a larger sample with a thickness gradient
results in a high uncertainty, u sto , because o f the systematic thickness variation. In the case o f a reduced
measurement area, usto is reduced and the arithmetic mean value, t , gives a better description of the local
thickness.
Care should be taken to address the risk that Type B standard uncertainties (see 8.4), which might
contribute to Type A standard uncertainties, are not counted twice.
∆t bm = abs ( t min − t r ) or ∆t bm = abs ( t max − t r ) . This gives the standard uncertainty caused by the
variation of the base metal properties ubm(k = 1) as shown in Formula (5):
∆t bm
u bm = (5)
3
The same estimation o f the standard uncertainty shall be carried out for all relevant factors listed in
Clause 5. For example, in the case of an expected variation of the surface curvature resulting in ∆t cs
with respect to D.4, the standard uncertainty can be estimated as ucs(k = 1) as shown in Formula (6):
∆t cs
u cs = (6)
3
In case the influence o f a factor is minimized by means o f an adjustment, the remaining uncertainty
shall be considered.
Some o f these factors influencing the accuracy can be minimized by means o f flexible foils as re ference
standards, e.g. base metal properties (5.3) or surface curvature (5.5), if the calibration is carried out
with foils on the base metal with identical material and curvature properties as the sample of interest
shows. In this case, only expected variations o f the sample properties shall be considered.
8 . 5 C o m b i n e d u n c e r ta i n ty, e x p a n d e d u n c e r ta i n ty a n d fi n a l re s u l t
The combined uncertainty summarizes all standard uncertainty components (8.2, 8.3, 8.4 and any
potential others). In the simplified approach described, when estimating the uncertainties for a given
thickness, or a very small thickness range, the sensitivity coe fficients can be considered to be equal to 1
(see Annex B). This results in the combined uncertainty, uc , as shown in Formula (7):
2 2 2 2 (7)
u c= u cal + u sto + u bm + u cs + ...
As the final result, the expanded uncertainty U(k = 2) is calculated (2-sigma level, 95,45 %) as shown in
Formula (8):
U ( k = 2 ) = 2uc (8)
And the complete result o f the measurement with the thickness value, t , is calculated as shown in
Formula (9):
t = t ±U(k = 2) (9)
9 Precision
9.1 General
See Annex G for further information on determining precision.
9.2 Repeatability (r)
Repeatability, r, is the value less than or equal to which the absolute difference between two test results
obtained under repeatability conditions may be expected to be, with a probability o f 95 % (according to
ISO 5725-1:1994, 3.16). The repeatability limit, r, in accordance with this document and calculated with
a probability o f 95 %, is given in Table 1.
measurement)
rx
1 rx
approx. µm µm µm
12 1,0 1,0
25 1,7 2,3
125 2,7 12,5
9.3 Reproducibility limit (R)
Repro ducibi l ity l i m it, R, is the value less than or equal to which the absolute difference between two
te s t re s u lts ob tai ne d u nder repro ducibi l ity cond ition s may b e exp e c te d to b e, with a prob abi l ity o f 9 5 %
(triple measurement)
Rx
1 Rx
approx. µm µm µm
12 –a –a
25 5,0 5,3
125 6,0 13,0
a No calculation of R x1 and R x re p ro duc ib i l i ty co u ld b e do ne i n re s p e c t o f o n l y o ne s a mp le .
10 Test report
The test report shall include the following information:
a) a l l i n formation ne ce s s ar y for the identi fic ation o f the te s t s p e ci men;
c) the sizes of the test areas over which the measurements were made in square millimetres (mm 2 );
NOTE Other units of measurement can be used, with agreement between the supplier and client.
d) the location(s) of the test area(s) on each specimen;
e) the number of test specimens measured;
f) an identi fic ation o f the i n s tr ument, prob e and s ta nda rd s u s e d for the te s t, i nclud i ng re ference to
which the test was carried out, including the results of the individual determinations and their
arithmetic mean;
h) the name of the operator and testing organization;
i) a ny u nu s ua l fe atu res ob s er ve d and any ci rc u m s ta nce s or cond ition s that a re l i kely to a ffe c t the
k) d ate o f the te s t.
Annex A
(informative)
Eddy-current generation in a metallic conductor
A.1 General
E ddy- c u rrent i n s tr uments work on the pri nciple that a h igh fre quenc y ele c tromagne tic field generate d
by the prob e s ys tem o f the i n s trument pro duce e ddy c u rrents i n a n ele c tric a l conduc tor on wh ich the
prob e i s place d . T he s e i nduce d e ddy c u rrents c au s e a cha nge o f the ele c tromagne tic field s u rrou nd i ng
the prob e coi l s ys tem a nd there fore re s u lt i n a ch ange o f the ampl itude a nd/or the ph as e o f the prob e
coi l i mp e dance, wh ich c a n b e u s e d as a me a s u re o f the th ickne s s o f the co ati ng on the conduc tor (s e e
A.2 and A.5) or of the conductor itself (see A.3 and A.4).
T he e ddy- c u rrent generation i n a me ta l conduc tor i s shown i n Figure 1.
J(δ
T he e ddy- c u rrent den s ity, ) , cha nge s its magn itude with i ncre a s i ng d i s tance, δ , from the s u r face o f
the conductor (depth). At the depth, δ0 (s tandard p ene tration dep th) , the ele c tromagne tic field a nd
de term i ne d b y the s a mple conduc ti vity and the p erme abi l ity a nd the fre quenc y o f the prob e coi l
a) high frequency or/and high conductivity b) low frequency or/and low conductivity
Key
1 probe δ0 standard depth of penetration
2 eddy currents δ depth
J(δ) dens ity
F i g u r e A . 1 — S c h e m a t i c t o s h o w t h e i n f l u e n c e o ff r e q u e n c y a n d c o n d u c t i v i t y o n t h e s t a n d a r d
penetration depth
The standard penetration depth, δ0 , is a useful value for some important rough estimations. It may be
calculated, in mm, using Formula (A.1):
δ =
503 × F (A.1)
0 p
f ×σ × µ r
where
f is the probe operating frequency, in Hertz;
σ is the electrical conductivity o f the conductor, in megasiemens per metre;
μ r is the relative permeability o f the conductor (for non-magnetic materials μ r = 1);
Key
1 probe
2 non-conductive coating
3 conductive base metal
Figure A.2 — Schematic of the eddy-current density in the case of non-conductive coating on a
conductive base metal
To establish that the eddy-current density is a unique measure o f the coating thickness, this density
should not be a ffected or limited by the base metal thickness. In order to achieve this, the base metal
shall be thicker than the minimum base metal thickness. This minimum thickness, tmin , in mm, can be
estimated as shown in Formula (A.2) (see 5.3):
t min = 3 δ 0 (A.2)
NOTE tmin is very o ften called “saturation thickness”. I f the base metal thickness is lower than this minimum
thickness, tmin , the measured value o f the coating thickness will be a ffected and the value o f the thickness that is
obtained is too high.
However, in the special case o f very thin coatings with a very small conductivity, the amplitude-sensitive
eddy-current method can also be applied, because this coating is considered as non-conductive. A
typical example is thin chromium coating plated on copper with a coating thickness below 10 µm. In
this situation, the impact o f the eddy currents induced in the coating can be neglected. However, a
larger thickness results in an increasing eddy-current density in the coating resulting in an increasing
thickness error, although the conductivity o f the chromium coating is small. The possible thickness
error should be determined or estimated to decide whether this method is applicable or not.
A.3 Example 2: Conductive coating on a non-conductive base material
In this case, the eddy-current density is determined only by the thickness o f the conductive coating (see
Figure A.3 ). A larger coating thickness results in an increased interaction of the magnetic field of the
probe with the conductive coating and consequently in an increased eddy-current density. This e ffect
can be used as a measure o f the coating thickness.
Key
1 probe
2 conductive coating
3 non-conductive base metal
Figure A.3 — Schematic of the eddy-current density in the case of conductive coating on a non-
conductive material
The approximate maximum measurable thickness, tmax, in mm, may be calculated from Formula (A.3):
max = 0,8 δ0
t (A.3)
For example, the thickness range is limited by the penetration depth, δ0 . If the conductive coating
thickness is increased further, the resulting increase o f the generated eddy-current density starts to
become smaller, i.e. the measurement sensitivity will be reduced.
The amplitude-sensitive eddy-current method is only capable o f measuring a conductive coating on top
of a non-conductive material. In the case of a conductive coating on top of a conductive base metal,
the amplitude-sensitive method cannot distinguish between the coating and the base metal, i.e. the
entire eddy-current density generated in the coating and the base metal would be used to determine
the coating thickness. This results in incorrect thickness values.
and conduc tivity o f the co ati ng and the conduc tivity and p erme abi l ity o f the b a s e me ta l . T he co ati ng
Key
1 probe
2 conductive coating
3 conductive base metal
Figure A.4 — Schematic of the eddy-current density in the case of conductive coating on a
conductive and/or magnetic base metal
b a s e me ta l, i . e . the co ati ng th ickne s s (s e e Figure A.2). However, the probe impedance is also affected
by the magne tic prop er tie s o f the b a s e me ta l, re s u lti ng i n th ickne s s errors . T h i s add itiona l me a s u ri ng
e ffe c t i s opp o s ite to the e ffe c t o f the e ddy c urrents . Fu r thermore, the ampl itude - s en s itive e ddy- c u rrent
me tho d i s ver y s en s itive to cha nges or fluc tuation s o f the p erme abi l ity o f the b a s e me ta l, i . e . even when
the magne tic b a s e me ta l wa s ta ken i nto accou nt i n the c a l ibration, the th ickne s s res u lts a re exp e c te d
to show strong variations on different magnetic base metals and also at different locations on one
b a s e me ta l (e . g. s te el) . C on s e quently, the magne tic me tho d s p e c i fie d i n I S O 2 178 s hou ld b e u s e d i n th i s
situation.
O n ly i n the c a s e o f ver y th ick co ati ngs ab ove approxi mately 1 m m, the a mpl itude - s en s itive e ddy- c u rrent
method can also be used for this application. There are two reasons for this:
a) for s uch th ick co ati ngs , the relative i mp ac t o f the p erme abi l ity o f the b a s e me ta l i s s trongly re duce d;
b) i n order to me a s u re s uch th ick co ati ngs , the e ddy- c urrent prob e coi l shows a la rge d ia me ter and
con s e quently, the ac tive a re a o f the co ati ng i nclude d i n the me as u rement i s i ncre a s e d . I n th i s way,
the prob e i nte grate s va ri ation s o f the p erme abi l ity over the enti re ac tive are a, re s u lti ng i n more
stable results.
T he m i n i mu m th ickne s s o f the co ati ng ne ce s s ar y to u s e th i s me tho d for the s e appl ic ation s shou ld b e
de term i ne d with res p e c t to the exp e c te d accep table rep e atabi l ity a nd truene s s o f the me a s u rement.
Annex B
(informative)
Basics of the determination of the uncertainty of a measurement
of the used measurement method corresponding to ISO/
IEC Guide 98-3
B.1 General
Coating thicknesses are generally determined as the mean value o f several single measurements that
are carried out at a fixed section o f the layer’s sur face.
On the basis o f these measurements, a mean value is allocated to the measurand “coating thickness”. An
uncertainty value is assigned which provides in formation about the reliability o f the allocated value.
Analysis is carried out progressively and begins by drawing up a model equation that shows the
functional correlation between the indicated output value, t, and all the relevant influence quantities,
Hi , as shown in Formula (B.1):
t = F(H H H 0
H H )
,
1
,
2
, ...
i
...
n
(B.1)
To every influence quantity belongs a sensitivity coe fficient, ci , which indicates how strong a
modification Δ Hi affects the result t.
When the function, F, is given as analytic expression, the sensitivity coefficients may be calculated by
partial derivation; see Formula (B.2):
δt
c i = δH (B.2)
i
I f the kind o f the functional correlation in unknown, an approximation by means o f polynomial
functions is recommended.
In many practical cases, this formulation is expressed by a linear dependence, i.e. the sensitivity
coe fficients become constant. This situation arises, for example, in sections o f limited coating thickness.
In order to summarize the uncertainties o f various error influences appropriately, all single uncertainty
components may be re ferred to a level o f confidence o f 68,27 %, the so-called “standard uncertainty”.
There are two types o f uncertainties: Type A (see B.2 ) and Type B (see B.3).
B.2 Type A
The standard uncertainty o f Type A is a measure o f all random errors arising from unpredictable or
stochastic temporal and spatial variations o f influence quantities.
The standard uncertainty corresponds to the point o f confidence o f the mean value; see Formula (B.3)
and Formula (B.4):
u s to = t( 68 , 2 7 % , n − 1 ) ×
s( t )
(B.3)
n
∑( x − x )2 j
s=
j= 1 (B.4)
(n − 1 )
where
s is the empirical standard deviation of the repetition measurement n, and
t (68,27 %, n – 1) is the student factor (degrees of freedom f = n – 1 and level of confidence with
p = 68,27 %).
u ( 68 , 27 %) = U( 95 2, 45 %) (B.6)
In order to summarize all investigated uncertainties, the so-called “combined uncertainty” is calculated.
This is done by multiplying all standard uncertainties by their sensitivity coe fficients and adding them
up squared. In a simplified case, the sensitivity coe fficients are equally one; see Formula (B.7):
u= ∑ (ciui ) 2 (B.7)
i
Multiplying with an indicated coverage factor o f k ≥ 2 results in an expanded uncertainty, which should
be indicated in the actual result; see Formula (B.8):
U = k⋅u (B.8)
Annex C
(informative)
Basic performance requirements for coating thickness gauges
which are based on the amplitude-sensitive eddy-current method
described in this document
C . 1 Te c h n i c a l s p e c i fi c a ti o n
T he manu fac tu rer ’s te ch n ic a l s p e c i fic ation s hou ld at le as t provide the fol lowi ng te ch n ic a l i n formation
c arrie d out u nder cond ition s s p e ci fie d b y the manu fac turer;
l) measuring rate;
m) data memor y (de s ign, c ap acity, data com mu n ic ation) ;
C.2.1 Prior to the supply, after repair and at regular intervals after use
After the instruments and probes have been adjusted according to the manufacturer’s instructions,
the me as u ri ng acc u rac y shou ld b e che cke d and veri fie d b y u s i ng a plane and u nco ate d b a s e me ta l
and a representative number of coated calibration standards or calibration foils, whose coating or foil
th ickne s s e s shou ld b e e qua l ly d i s tribute d with i n the me a s uri ng ra nge o f the res p e c tive prob e .
T he a i m o f the che ck/veri fic ation o f the i n s tru ments i s to en s u re that the th ickne s s devi ation s a re
Annex D
(informative)
Examples for the experimental estimation of factors affecting the
measurement accuracy
D.1 General
Factors a ffecting the measurement accuracy are summarized and described in Clause 5. In practical
measurements, it is important to estimate the influence o f these factors or the resulting uncertainty.
Therefore, some examples of simple experiments are described in this annex in order to show how the
influence o f these factors can be estimated. These experiments also provide a basis for estimating the
respective uncertainty.
The factors described in D.2 to D.5 can cause di fferently pronounced influences for an instrument
working with combined measuring principles in one probe. Consequently, the factors should be
estimated separately for each measuring principle.
Key
d distance from the probe to the edge
Figure D.1 — Schematic representation of the test for edge effect
I f the instrument does not display negative values, it is recommended to use a thin foil (e.g. 10 µm)
between the probe and base metal to observe the decrease o f the thickness.
NOTE The procedure to stack several samples in order to simulate an increase o f the base metal thickness
allows a good estimation of t0 crit because the impact o f the air gap between the samples on the eddy-current
generation in the sample stack in comparison to the respective homogeneous material is almost negligible (eddy-
current flow is perpendicular to the probe axis). There fore, this simplified procedure can be carried out more
easily with good results instead o f producing base metals with variable thickness.
Figure D.2 — Schematic representation of the test for base metal thickness
In order to improve the accuracy o f the estimation o f the curvature influence, increase the number o f
samples with different diameters.
NOTE The same procedure can be used in cases where the samples show a concave curvature, however, this
concave curvature results in negative thickness readings. I f the instrument does not display negative values, it
is recommended to use a thin foil (e.g. 10 µm) between the probe and base metal to observe the decrease of the
thickness.
Figure D.4 — Schematic representation of the test for base metal conductivity test
Annex E
(informative)
Table of the student factor
Annex F
(informative)
Example of uncertainty estimation (see Clause 8)
range at t = 25 µm : ∆t bm = ±1 , 2 µm
F.2 Steps
F.2.1 The examp le s amp le is meas ured by fo llo wing thes e s tep s .
ur =
T 0 , 5 µm = 0 , 29 µm
=
3 3
i i) T he s tandard u ncer tai nty o f the veri fic ation me as u rement (on ly the s to chas tic comp onent
is considered) is:
u sto = t ( 68 , 27 %, n − 1 ) ×
s (t )
= 1 , 06 ×
0 , 11 µm = 0 , 04 µm
n 10
2 2
i i i) T he combi ne d uncer ta i nty i s u c = ( 0 , 04 µm ) + ( 0 , 29 µm ) = 0 , 29 µm
iv) T he e xp ande d u ncer ta i nty i s U cal ( k = 2 ) = 2 × u c = 0 , 58 µm
t − tr
v) The result is E = U =
1 , 14 µm
= 1 , 96
cal (k = 2) 0 , 58 µm
vi) Calibration is not correct. A significant deviation has been detected, because E = 1 , 96 > 1 ,
i.e. the difference between the measured value, t , and the given reference foil value,
t − t r , is larger than U cal ( k = 2 ) = 0 , 58 µm ; consequently the calibration accuracy can be
1) u c = ( 0 , 03 µm ) 2 + ( 0 , 29 µm ) 2 = 0 , 29 µm : u cal = 0 , 29 µm
e) Measure the sample.
1) Seven repeated measurements within the given measurement area of the sample.
2) Result ( n = 7 ): t = 22 , 8µm and s( t ) = 0 , 76µm
f ) Calculate all measurement uncertainty components and the combined uncertainty.
F.2 .3 Further conclusions: it is obvious that the resulting uncertainty is limited by the largest
uncertainty component, in this case, the possible base metal property variation (conductivity variation).
Therefore, an increase of the number of repeated measurements would reduce usto, however, the
combined uncertainty would not be strongly affected in this way.
Annex G
(informative)
Details on precision
The thicknesses o f the calibration foils were: 12 µm, 25 µm and 125 µm.
Coating thickness measurements were done directly a fter every calibration and adjustment.
G.6 Evaluation
G.6.1 General
The statistical evaluation was carried out following ISO 5725-2 and ISO/TR 22971.
Evaluation was carried out for each calibration method.
G.6.2 Evaluation o f first measuring point
The repeatability limit, rx1 , and the reproducibility limit, Rx
1 , were calculated from the first measuring
point measured in triplicate.
G.6.3 Evaluation o f all five measuring points
The repeatability limit, rx , and the reproducibility limit, R x , were calculated from all five measuring
points. For the first measuring point, the arithmetic mean from the triple measurements is used.
Table G.2 contains the results for repeatability limits and reproducibility limits calculated from the
first measuring point in comparison to the respective limits calculated from all five measuring points.
µm µm µm µm
12-R 1,0 _a 1,5 _a
12-S 1,0 _a 1,5 _a
25-R 1,6 4,1 2,2 3,7
25-S 1,7 5,0 2,3 5,3
125-R 2,7 5,6 12,3 12,3
125-S 2,3 6,0 12,5 13,0
rx
1 and Rx
1 Repeatability limit and reproducibility limit o f first measuring point (triple measurement).
rx and R x Repeatability limit and reproducibility limit o f all five measuring points.
a No calculation of R x1 and R x reproducibility could be done in respect o f only one sample.
Figure G.1 to Figure G.3 s how the res u lts o f th icknes s me a s urements b as e d on the th re e d i fferent
where
R is the reference method, and
S is the standard method (see also G.4).
Key
SD-12-R
SD-12-S
Figure G.1 — Comparison of reference and standard method calibration with 12 µm foil
Key
SD-25-R
SD-25-S
Figure G.2 — Comparison of reference and standard method calibration with 25 µm foil
Key
SD-125-R
SD-125-S
Figure G.3 — Comparison of reference and standard method calibration with 125 µm foil
Bibliography
Magnetic method
[4] I S O 57 2 5 -1 :19 9 4, Accuracy (trueness and precision) of measurement methods and results — Part 1:
[4] ISO 5725-2, Accuracy (trueness and precision) of measurement methods and results — Part 2: Basic
method for the determination ofrepeatability and reproducibility ofa standard measurement method
[6] I S O/I E C Guide 9 9 : 2 0 0 7, International vocabulary of metrology — Basic and general concepts and
associated terms (VIM)
guidance for the use of ISO 5725-2:1994 in designing, implementing and statistically analysing
interlaboratory repeatability and reproducibility results
ICS 25.220.20
Price based on 34 pages
© ISO 2017 – All rights reserved