Manual - Usm 100 H
Manual - Usm 100 H
CONTENTS
1.0 INTRODUCTION ..................................................................................................................... 3
1.1 PREFACE ........................................................................................................................... 3
1.2 FEATURES.......................................................................................................................... 3
1.3 FLOW MEASUREMENT PRINCIPLE ........................................................................................ 3
1.4 PART IDENTIFICATION ......................................................................................................... 5
1.5 APPLICATIONS .................................................................................................................... 7
1.6 DATA INTEGRITY AND BUILT IN TIME KEEPER ........................................................................ 7
1.7 PRODUCT IDENTIFICATION.................................................................................................. 7
1.8 SPECIFICATIONS ................................................................................................................. 7
2.0 MEASUREMENT...................................................................................................................... 8
2.1 BUILT IN BATTERY .............................................................................................................. 8
2.2 POWER ON ........................................................................................................................ 8
2.3 KEYPAD ............................................................................................................................. 9
2.4 MENU W INDOWS .............................................................................................................. 10
2.5 MENU W INDOW LIST ......................................................................................................... 11
2.6 STEPS TO CONFIGURE PARAMETERS ................................................................................. 11
2.7 TRANSDUCER MOUNTING ALLOCATION .............................................................................. 13
2.8 TRANSDUCER INSTALLATION ............................................................................................. 14
2.8.1 Transducer Spacing...................................................................................................... 15
2.8.2 V Method Installation..................................................................................................... 15
2.8.3 Z Method Installation..................................................................................................... 15
2.8.4 W Method Installation.................................................................................................... 15
2.9 INSTALLATION TESTING ..................................................................................................... 15
2.9.1 Signal Strength ............................................................................................................. 16
2.9.2 Signal Quality................................................................................................................ 16
2.9.3 Total Transit Time and Delta Time ................................................................................ 16
2.9.4 Transit Time Ratio......................................................................................................... 17
3.0 HOW TO CHECK AND SETUP ............................................................................................. 17
3.1 HOW TO CHECK THE INSTRUMENT WORKS PROPERLY.......................................................... 17
3.2 HOW TO CHECK THE LIQUID FLOW DIRECTION ..................................................................... 17
3.3 HOW TO CHANGE THE UNIT READINGS ................................................................................ 17
3.4 HOW TO SELECT A FLOW RATE .......................................................................................... 17
3.5 HOW TO USE THE TOTALISER MULTIPLIER ........................................................................... 18
3.6 HOW TO SET THE TOTALISER FUNCTIONS ........................................................................... 18
3.7 HOW TO RESET TOTALISERS.............................................................................................. 18
3.8 HOW TO RESTORE THE FACTORY DEFAULTS ....................................................................... 18
3.9 HOW TO USE THE DAMPER TO STABILISE THE FLOW RATE .................................................... 18
3.10 HOW USE THE ZERO CUT OFF FUNCTION ............................................................................ 18
3.11 HOW TO SET A ZERO POINT ............................................................................................... 19
3.12 HOW TO CHANGE THE FLOW RATE SCALE FACTOR .............................................................. 19
3.13 HOW TO SET AND LOCK THE PASSWORD............................................................................. 19
3.14 HOW TO USE THE INBUILT DATA LOGGER ............................................................................ 19
3.15 HOW TO USE THE FREQUENCY OUTPUT ............................................................................. 20
3.16 HOW TO USE THE TOTALISER PULSE OUTPUT ..................................................................... 20
3.17 HOW TO PRODUCE AN ALARM SIGNAL ................................................................................. 20
3.18 HOW TO USE THE BUILT IN BUZZER .................................................................................... 21
3.19 HOW TO USE THE OCT PULSE OUTPUT .............................................................................. 21
3.20 HOW TO SET THE BUILT IN CALENDER ................................................................................ 21
3.21 HOW TO ADJUST THE LCD CONTRAST ................................................................................ 21
Av da Invernada 12 - Vila Congonhas - CEP 04612-060 - São Paulo - SP
Fone / fax 011 5096 4654
Para honra e glória de ΙҲΘγΣ
Page 1 of 66
3.22 HOW TO USE THE RS232 SERIAL INTERFACE ...................................................................... 21
3.23 HOW TO VIEW THE TOTALISERS ......................................................................................... 21
3.24 HOW TO USE THE WORKING TIMER ..................................................................................... 22
3.25 HOW TO USE THE MANUAL TOTALISER ................................................................................ 22
3.26 HOW TO CHECK THE SERIAL NUMBER ................................................................................ 22
3.27 HOW TO CHECK THE BATTERY LIFE .................................................................................... 22
3.28 HOW TO CHARGE THE BATTERY ......................................................................................... 22
4.0 MENU WINDOW DETAILS.................................................................................................... 22
5.0 TROUBLE SHOOTING.......................................................................................................... 28
5.1 POWER-ON ERRORS ......................................................................................................... 28
5.2 W ORKING STATUS ERRORS ............................................................................................... 28
5.3 OTHER PROBLEMS AND SOLUTIONS .................................................................................. 29
6.0 COMMUNICATION PROTOCOL........................................................................................... 30
6.1 RS232 CONNECTOR PIN-OUT .......................................................................................... 30
6.2 COMMUNICATION PROTOCOL ............................................................................................ 30
6.2.1 Basic Commands......................................................................................................... 30
6.2.2 Protocol Prefix Usage .................................................................................................. 32
6.3 THE M COMMAND AND THE ASCII CODES.......................................................................... 33
7.0 WARRANTY AND SERVICE................................................................................................. 34
7.1 W ARRANTY ...................................................................................................................... 34
7.2 SERVICE .......................................................................................................................... 34
8.0 APPENDIX ............................................................................................................................ 35
8.1 BATTERY MAINTENANCE AND REPLACEMENT ..................................................................... 35
8.2 PIPE SIZE TABLES ............................................................................................................ 35
8.2.1 Standard Pipe size charts for Copper............................................................................ 35
8.2.2 Standard Pipe size charts for PVC ................................................................................ 37
8.2.3 Standard Pipe size charts for Steel pipe ....................................................................... 38
8.2.4 Standard Pipe Size Charts for Cast Iron Pipe............................................................... 48
8.2.5 Standard Pipe size charts for Ductile Iron Pipe ............................................................ 49
8.3 SOUND SPEED TABLES ..................................................................................................... 50
8.3.1 Sound Speed data of solids .......................................................................................... 50
8.3.2 Sound Speed in Water .................................................................................................. 52
8.3.3 Sound Speed in Liquids ............................................................................................... 53
1.0 INTRODUCTION
1.1 Preface
The hand held flow meter is a battery-powered ultrasonic flow meter with the capability of a
full-size flow meter. It is carefully designed for portability and ease of use.
The hand held flow meter is based on clamp-on transit-time flow measurement principle. It
measures the flow rate of liquid in a pipe from outside of the pipe by using a pair of ultrasonic
transducers. In general, the liquid should be full in the pipe, and should contain very little
particles or bubbles. Examples of applicable liquids are: water (hot water, chill water, city
water, sea water, etc.), sewage, oil (crude oil, lubricating oil, diesel oil, fuel oil, etc.), chemicals
(alcohol, acids, etc.), waste, beverage and liquid food, solvents and other liquids.
Due to the nature of clamp-on technique, the transducer installation is simple and no special
skills or tools are required. Besides, there is no pressure drop, no moving parts, no leaks and
no contamination.
The hand held flow meter utilizes our proprietary technologies such as advanced signal
processing, low-voltage transmitting, small signal receiving with self-adapting, and etc. It also
incorporates the latest surface-mounting semiconductors and mini PCB design techniques.
The built-in rechargeable Ni-H battery can work continuously for more than 10 hours without
recharge.
The hand held flow meter has also a built-in data-logger, which allows storage of 2,000 lines of
data. The stored information can be downloaded to a PC through its RS232 connection port.
The hand held flow meter also provides digital output such as frequency output or pulsed
totaliser output.
1.2 Features
The hand held flow meter is designed to measure the velocity of liquid within a closed conduit.
It uses the well-know transit-time technology. The transducers are a non-contacting, clamp-on
type. They do not block the flow, thus no pressure drop. They are easy to install and remove.
The hand held flow meter utilizes a pair of transducers that function as both ultrasonic
transmitter and receiver. The transducers are clamped on the outside of a closed pipe at a
specific distance from each other. The transducers can be mounted in V-method where the
Page 3 of 63
sound transverses the pipe twice, or W-method where the sound transverses the pipe four
times, or in Z-method where the transducers are mounted on opposite sides of the pipe and
the sound crosses the pipe once. The selection of the mounting methods depends on pipe and
liquid characteristics.
The hand held flow meter operates by alternately transmitting and receiving a frequency-
modulated burst of sound energy between the two transducers and measuring the transit time
that it takes for sound to travel between the two transducers. The difference in the transit time
measured is directly and exactly related to the velocity of the liquid in the pipe, as shown in the
following figure.
Where
θ is the angle between the sound path and the flow direction
Tup is the time for the beam travelling from upstream the transducer to the downstream
transducer
Tdown is the time for the beam travelling from the downstream transducer to the
upstream transducer
∆T = Tup – Tdown
Transducers:
The hand held flow meter flow meter can be applied to a wide range of pipe flow
measurements. The pipe size ranges 0.5”-240” (15mm-6000mm). A variety of liquid
applications can be accommodated: ultra-pure liquids, potable water, oil, chemicals, raw
sewage, reclaimed water, cooling water, river water, sea water, plant effluent, etc. Because
the transducers are non-contacting and have no moving parts, the flow meter will not be
affected by flow pressure or liquid properties. Standard transducers are rated to 100ºC.
Higher temperatures can be accommodated. For further information, please consult the
manufacturer for assistance.
All user-entered configuration values are stored in the built-in non-volatile flash memory that
can retain the data for over 100 years, even when the power is lost or turned off. Password
protection is provided to avoid inadvertent configuration changes or totalizer resets.
A time-keeper is integrated in the flow meter. It works as the time base for flow totalizing. The
time-keeper remains operating as long as the battery’s terminal voltage is over 1.5V. In case
of battery failure, the time-keeper will not keep running and the time data will lost. The user
must re-enter proper time values after the battery failure is recovered. Improper time values
will affect the totalizers as well as many other functions.
Each set of the hand held flow meter series flow meter has a unique product identification
number or ESN written into the software that can only be modified with a special tool by the
manufacturer. In case of any hardware failure, please provide this number which is located on
menu window M61 when contacting the manufacturer.
1.8 Specifications
Linearity 0.5%
Repeatability 0.2%
Accuracy ±1% of reading at rates>0.6 ft/s. ±0.5% with on-site calibration
Response Time 0-999 seconds, user-configurable
Velocity ±0.03 ~ ±105 ft/s (±0.01 ~ ±30 m/s), bi-directional
Pipe Size 0.5” ~ 240” (15 ~ 6,000mm)
Meter, Feet, Cubic Meter, Liter, Cubic Feet, USA Gallon, Imperial Gallon,
Rate Units Oil Barrel, USA Liquid Barrel, Imperial Liquid Barrel, Million USA Gallons.
User configurable.
Totaliser 7-digit totals for net, positive and negative flow
Liquid Types Virtually all liquids
2.0 MEASUREMENT
The instrument can operate either from the built-in Ni-H rechargeable battery, which will last
over 10 hours of continuous operation when fully charged, or from an external AC/power
supply from the battery charger.
The battery charging circuit employs both constant-current and constant-voltage charging
methods. It has a characteristic of fast charging at the beginning and very slow charging when
the battery approaches to full charge. Generally, when the green LED is on, the battery is
nearly 95% charged, and when the red LED is off, the battery is nearly 98% charged.
Since the charging current becomes tapered when the battery charging is nearly completed,
i.e. the charging current becomes smaller and smaller, therefore, there should be no over-
charging problem. This also means the charging progress can last very long. The charger can
be connected to the handset all the time when an around-the-clock measurement is required.
When fully charged, the terminal voltage reaches around 4.25V. The terminal voltage is
displayed on window M07. When the battery is nearly consumed, the battery voltage drops to
below 3V. The approximate remaining working time is indicated in this window as well.
Notice that the battery remaining working time is estimated based on the current battery
voltage. It may have some errors, especially when the terminal voltage is in the range from
3.70 to -3.90 volts.
For Battery maintenance and replacement, please refer to Appendix A.
2.2 Power On
Press ON key to turn on the power and press OFF to turn off the power.
Av da Invernada 12 - Vila Congonhas - CEP 04612-060 - São Paulo - SP
Fone / fax 011 5096 4654
Para honra e glória de ΙҲΘγΣ
Page 8 of 66
Once the flow meter is turned on, it will run a self-diagnostic program, checking first the
hardware and then the software integrity. If there is any abnormality, corresponding error
messages will be displayed.
Generally, there should be no display of error messages, and the flow meter will go to the
most commonly used Menu Window #01 (short for M01) to display the Velocity, Flow Rate,
Positive Totaliser, Signal Strength and Signal Quality, based on the pipe parameters
configured last time by the user or by the initial program.
The flow measurement program always operates in the background of the user interface. This
means that the flow measurement will keep running regardless of any user menu window
browsing or viewing. Only when the user enters new pipe parameters will the flow meter
change measurement to reflect the new parameter changes.
When new pipe parameters are entered or when the power is turned on, the flow meter will
enter into a self-adjusting mode to adjust the gain of the receiving circuits so that the signal
strength will be within a proper range. By this step, the flow meter finds the best receiving
signals. The user will see the progress by the number 1, 2, or 3, located on the lower right
corner of the LCD display.
When the user adjusts the position of the installed transducers, the flow meter will re-adjust
the signal gain automatically.
Any user-entered configuration value will be stored in the NVRAM (non-volatile memory), until
it is modified by the user.
2.3 Keypad
Key ▲/+ is the going UP key when the user wants to go to the
upper menu window. It also works as + key when entering
numbers.
Key ▼/- is the going DOWN key when the user wants to go to the
lower menu window. It also works as the ‘–‘ key when entering
numbers.
Key MENU is the key for the direct menu window jump over. Whenever the user wants to
proceed to a certain menu window, the user can press this key followed by a 2-digit number.
The user interface of this flow meter comprises about 100 different menu windows that are
numbered by M00, M01, M02 … M99.
(2) Press ▲/+ or ▼/- key. Each time of the ▲/+ key pressing will lead to the lower-numbered
menu window. For example, if the current window is on M12, the display will go to window
M11 after the ▲/+ key is pressed once.
(1) Menu windows for number entering, e.g., M11 for setting up pipe outer diameter.
(2) Menu windows for option selection, e.g., M14 for the selection of pipe materials.
(3) Results display windows, e.g., window M00 for displaying Velocity, Flow Rate, etc.
For number entering windows, the user can directly press the digit keys if the user wants to
modify the value. For example, if the current window is on M11, and the user wants to enter
219.2345 as the pipe outer diameter, then, the flowing keys should be pressed: 2 1 9 . 2
3 4 5 ENT.
For option selection windows, the user should first press the ENT key to get into option
selection mode. Then, use ▲/+ , ▼/- , or digit key to select the right option. Consequently,
press the ENT to make the selection.
For example, assume your pipe material is stainless steel and you are currently on menu
window M14 which is for the selection of pipe materials (if you are on a different window, you
need press MENU 1 4 first in order to enter into the M14 window.) You need to press the
ENT key to get into the option selection mode. Then, either press the ▲/+ and ▼/- keys to
make the cursor on the line that displays “1. Stainless Steel”, or press the 1 key directly. At
the end, press ENT again to make the selection.
Generally, the ENT key must be pressed to get into the option selection mode for option
modifications. If the “Locked M47 Open’ message is indicated on the lowest line of the LCD
display, it means that the modification operation is locked out. In such cases, the user should
go to M47 to have the instrument unlocked before any further modification can be made.
M00~M09 windows for the display of the instantaneous flow rate, net totalizer value, positive
totalizer value, negative totalizer value, instantaneous flow velocity, date time, battery voltage
and estimated working hours for the battery.
M10~M29 windows for entering system parameters, such as pipe outer diameter, pipe wall
thickness, liquid type, transducer type / model, transducer installation method, etc. Transducer
installation spacing is also displayed on one of the windows.
M30~M38 windows for flow rate unit selection and totalizer configuration. User can use these
windows to select flow rate unit, such as cubic meter or liter, as well as to turn on / off each
totalizer, or to zero the totalizers.
M40~M49 windows for setting response time, zeroing / calibrating the system and changing
password.
M50~M53 windows for setting up the built-in logger.
M60-M78 windows for setting up time-keeper and displaying software version, system serial
number ESN and alarms.
M90~M94 windows for displaying diagnostic data. Those data are very useful when doing a
more accurate measurement.
M97~M99 are not windows but commands for window copy output and pipe parameter output.
M+0~M+8 windows for some additional functions, including a scientific calculator, display of
the total working time, and display of the time and the flow rate when the device is turned on
and turned off.
Other menu windows such as M88 have no functions, or functions were cancelled because
they are not applied to this version of the software.
The major reason why the menu windows are arranged in the above way is to make this
version be compatible with previous versions. This will make life easier for the former version
users.
In order to make the hand held flow meter work properly, the user must follow the following
steps to configure the system parameters:
Example: For standard (commonly used) pipe materials and standard (commonly measured)
liquids, the parameter configuration steps are as following:
(1) Press keys MENU 1 1 to enter into M11 window. Input the pipe outer diameter through
the keypad and press ENT key.
(2) Press key ▼/- to enter into M12 window. Input the pipe thickness through the keypad and
press ENT key.
(3) Press key ▼/- to enter into M14 window. Press ENT key to get into the option selection
mode. Use keys ▲/+ and ▼/- to scroll up and down to the proper pipe material, and then
press ENT key.
(4) Press key ▼/- to enter into M16 window. Press ENT key to get into the option selection
mode. Use keys ▲/+ and ▼/- to scroll up and down to the proper liner material, and then
press ENT key. Select “No Liner”, if there is no liner.
(5) Press key ▼/- to enter into M20 window. Press ENT key to get into the option selection
mode. Use keys ▲/+ and ▼/- to scroll up and down to the proper liquid, and then press ENT
key.
(6) Press key ▼/- to enter into M23 window. Press ENT key to get into the option selection
mode. Use keys ▲/+ and ▼/- to scroll up and down to the proper transducer type, and then
press ENT key.
(7) Press key ▼/- to enter into M24 window. Press ENT key to get into the option selection
mode. Use keys ▲/+ and ▼/- to scroll up and down to the proper transducer mounting
method, and then press ENT key.
(8) Press key ▼/- to enter into M25 window. The transducer installation distance will be
displayed on the window. Based on this distance, install the transducers on the pipe now. After
installation is completed, press ENT key to go to M01 window to check if the measurement
result is good.
The first-time users may need some time to get familiar with the operation. However, the user
friendly interface of the instrument makes the operation quite easy and simple. You will soon
find that it is actually very quick to configure the instrument with very little key pressing, since
the interface allows the user to go to the desired operation directly without any extra steps.
(2) When the current window is one between M00 to M09, pressing ENT key will lead to
window M90 for displaying diagnostic data. Press ENT key again to return to the previous
window. Press the . key to go to window M11.
When the current window is M25, pressing ENT key will lead to window M01.
The first step in the installation process is to select an optimal location for installing the
transducers in order to make the measurement reliable and accurate. A basic knowledge
about the piping and its plumbing system would be advisable.
An optimal location would be defined as a long straight pipe line full of liquid that is to be
measured. The piping can be in vertical or horizontal position. The following table shows
examples of optimal locations.
The transducers used by the ultrasonic flow meter are made of piezoelectric crystals both for
transmitting and receiving ultrasonic signals through the wall of liquid piping system. The
measurement is realized by measuring the travelling time difference of the ultrasonic signals.
Since the difference is very small, the spacing and the alignment of the transducers are critical
factors to the accuracy of the measurement and the performance of the system. Meticulous
care should be taken for the installation of the transducers.
Extra care should be taken to avoid any sand or dust particles left
between the pipe surface and the transducer surface.
Horizontally lined pipes could have gas bubbles inside the upper
part of the pipe. Therefor, it is recommended to install the
transducers horizontally by the side of the pipe.
*Note: It is recommended to use the Conductive Gel product from Livingstone, as the
ultrasonic couplant for safety considerations. Other couplants, such as grease, gel, and
Vaseline, can be used as alternatives, but at your own risk.
The spacing value shown on menu window M25 refers to the distance of inner spacing
between the two transducers (see the following figure). The actual distance of the two
transducers should be as close as possible to this spacing value.
After completion of the transducer installation, the user should check the following items: the
receiving signal strength, the signal quality Q value, the delta time (traveling time difference
between the upstream and the downstream signals), the estimated liquid sound speed, the
Signal strength indicates the amplitude of receiving ultrasonic signals by a 3-digit number.
[000] means there is no signal detected and [999] refers to the maximum signal strength that
can be received.
Although the instrument works well when the signal strength ranges from 500 to 999, stronger
signal strength should be pursued, because a stronger signal means a better result. The
following methods are recommended to obtain strong signals:
(1) If the current location is not good enough for a stable and reliable flow reading, or if the
signal strength is lower than 700, relocate to a more favorable location.
(2) Try to polish the outer surface of the pipe, and apply more couplant to increase the signal
strength.
(3) Tenderly adjust the position of the two transducers, both vertically and horizontally, while
checking the signal strength. Stop at the position where the signal strength reaches to
maximum. Then, check the transducer spacing to make sure it is the same as or very close to
what window M25 shows.
Signal quality is indicated as the Q value in the instrument. A higher Q value would mean a
higher Signal to Noise Ratio (short for SNR), and accordingly a higher degree of accuracy able
to be achieved. Under normal pipe condition, the Q value is in the range of 60-90, the higher
the better.
Causes for a lower Q value could be:
1. Interference from other instruments and devices nearby, such as a power frequency
transverter which could cause strong interference. Try to relocate the flow meter to a new
place where the interference can be reduced.
2. Bad sonic coupling between the transducers and the pipe. Try to polish the pipe surface
again, clean the surface and apply more couplant, etc.
3. The selected pipe section is difficult to conduct the measurement. Relocate to a more
favourable pipe line.
The total transit time (or travelling time) and the delta time are displayed on menu window
M93. They are the primary data for the instrument to calculate the flow rate. Therefore, the
measured flow rate will vary as the total transit time and delta time vary.
The total transit time should remain stable or vary in a very small range.
This ratio is usually used to check whether the transducer installation is good and whether the
entered pipe parameters are in consistency with their actual values. If the pipe parameters are
correct and the transducers are installed properly, the transit time ratio should be in the range
of 100±3. If this range is exceeded, the user should check:
Generally speaking, when ‘R’ is displayed in the lower right corner of the LCD display, the
instrument is working properly.
If an ‘H’ flashes instead, the received signal could be poor. Please refer to the chapter on
diagnosis for more information.
If a ‘J’ is displayed, it means that the hardware of this instrument could be out of order. Refer
to the chapter on diagnosis.
3.2 How to check the liquid flow direction
Check the flow rate display. If the value is POSITIVE, the direction of the flow will be from the
RED transducer to the BLUE transducer; if the value is NEGATIVE, the direction will be from
the BLUE transducer to the RED transducer.
Use menu window M30 for the selection of units systems, either English or in Metric.
Use menu window M31 to select the flow rate unit as well as the corresponding time unit.
Use window M33 to select a proper multiplying factor for the totaliser multiplier. Make sure that
the rate of the totaliser pulse is not too fast, neither too slow. A speed of several pulses per
minute is preferable.
If the totaliser multiplying factor is too small, the output pulse will be very fast and there could
be a loss of pulses. The designed minimum pulse period is 500 milliseconds.
If the totaliser multiplying factor is too large, the output pulse will be very slow, which might be
a problem if the master device requires fast response.
The flow meter has three totaliser functions, generally you will only need the Pos totaliser set
as most pipes will have flow in one direction only.
Use M34, M35 and M36 to turn on or turn off the POS, NEG, or NET totaliser, respectively.
This operation will erase all the parameters entered by the user and setup the instrument with
factory default values.
The damper acts as a filter for a stable reading. If ‘0’ is entered in window M40, that means
there is no damping. A bigger number brings a more stable effect. But bigger damper numbers
will prevent the instrument from acting quickly.
The number displayed in window M41 is called the zero-cut-off value. When the absolute
value of the measured flow rate is less than the zero-cut-off value, the measured flow rate will
be replaced with ‘0’. This is to avoid any invalid accumulation when the actual flow is below the
zero-cut-off value.
The zero-cut-off operation does not affect the flow measurement when the actual flow is
greater than the zero-cut-off value.
When the flow in a pipe is absolutely stopped, the flow meter could still give a small non-zero
flow rate reading. In order to make the measurement accurate, it is necessary to remove this
“zero point” reading.
Window M42 allows us to take care of this issue. At first, the user should make sure that the
liquid in the pipe is totally stopped (no velocity). Then, go to window M42 and press the ENT
key to start the zero point setup function.
A scale factor (SF) is the ratio between the ‘actual flow rate’ and the flow rate measured by the
flow meter. It can be determined by calibration with a standard flow calibration equipment. To
change the SF, press M45, then the ENT key, enter the new SF, and press ENT again.
When the system is locked, the user can still browse menu windows, but cannot make any
modifications on the windows.
The password locking / unlocking is done in window M47. The system can be locked without a
password or with a password consisted of 1 to 4 digits.
For no-password locking / unlocking, just press ENT key in window M47.
CAUTION!
If the password is forgotten, after being locked no further access will be allowed, please write
down the password and store in a safe location.
The built-in data logger has a space of 24K bytes of memory, which will hold about 2000 lines
of data.
Use M50 to turn on the logger and to select the items that are going to be logged.
Use M51 to set up the starting time, time interval, and the duration each logging lasts.
Use M52 to select the data storage direction. Data can be stored in a logger buffer or directed
to the RS-232C interface without being stored into the logger buffer.
User needs to go to window M52 to clear the logging data remained in the RS-232C interface
and in the logger buffer.
The flowmeter will produce a pulse output with every unit of liquid flow. This pulse could be
used by an external pulse counter to accumulate the flow rate.
Refer to section 3.4 and 3.5 for the setup of the totalizer units and multiplier.
The totalizer pulse output can only be connected to OCT devices or BUZZER hardware
devices.
For example, assume that the POS totalizer pulse output is needed, and every pulse
represents 0.1cubic meter of liquid flow. Assume also that the pulse output is connected to an
internal Buzzer. With every 0.1 cubic meter of flow, we need the BUZZER to beep for a while.
In order to achieve this, the following steps must be performed:
The flowmeter will produce a pulse output with every unit of liquid flow. This pulse could be
used by an external pulse counter to accumulate the flow rate.
Refer to section 3.4 and 3.5 for the setup of the totalizer units and multiplier.
The totalizer pulse output can only be connected to OCT devices or BUZZER hardware
devices.
For example, assume that the POS totalizer pulse output is needed, and every pulse
represents 0.1cubic meter of liquid flow. Assume also that the pulse output is connected to an
internal Buzzer. With every 0.1 cubic meter of flow, we need the BUZZER to beep for a while.
In order to achieve this, the following steps must be performed:
(1) Select the Cubic Meter (m3) unit in window M32.
(2) Select the Multiplier factor as ‘2. X0.1’ in window M33.
(3) Select the output option ‘9. POS INT Pulse’ in window M77. (INT stands for totalized )
There are 2 types of hardware alarm signals that are available with this instrument. One is the
Buzzer, and the other is the OCT output.
The triggering sources of the alarming events for both the Buzzer and the OCT output could
be:
There are two alarms in this instrument, #1 alarm and #2 alarm. They can be configured in
windows M73, M74, M75 and M76.
For example, assume we need the Buzzer to start beeping when the flow rate is less than 300
m3/h and greater than 2000m3/h. The following setup steps would be recommended.
(1) Enter flow rate lower limit 300 in M73 for #1 alarm,
(2) Enter flow rate upper limit 2000 in M74 for #1 alarm,
(3) Select item ‘6. Alarm #1’ in M77.
The built-in buzzer is user-configurable. It can be used as an alarm. Use M77 for setups.
The OCT output is on/off type. It is user-configurable. For example, you can set the OCT
output to be a pulse signal for flow accumulation.
Use M77 for the setup.
Notice that the Frequency Output shares the same OCT hardware.
The OCT output is wired to pin 6 (for positive) and pin 5 (for ground) of the RS-232 connector.
Refer to section 6.1 for more details.
No modification on the built-in calendar will be needed in most cases. The calendar consumes
insignificant amount of power. Modification will be needed only when the battery is totally
exhausted, or when the replacement of the batteries takes a long time so that the original
clock data get lost.
Press the ENT key in M61 for Modification. Use the dot key to skip over these digits that need
no modification.
Use M70 to adjust the LCD contrast. The adjusted results will be stored in the EEPROM so
that the MASTER ERASE (factory default restoration) will make no effect on the contrast.
Use M82 to view the daily totaliser, the monthly totaliser and the yearly totaliser.
Use the working timer to check the time that has passed with a certain kind of operation. For
example, use it as a timer to show how long a fully-charged battery will last.
In window M72, press ENT key and select YES to reset the working timer.
Use M82 to view the daily totaliser, the monthly totaliser and the yearly totaliser.
Every set of the flow meters utilizes a unique ESN to identify the meter. The ESN is an 8-digit
number that provides the information of version and manufacturing date.
The user can also employ the ESN for instrumentation management.
Use M+1 to view the total working time since the instrument was shipped out of the
manufacturer.
Use M+4 to view the total number of times the instrument has been turned on and off since the
instrument was shipped out of the manufacturer.
Use M07 to check how long the battery will last. Also please refer to section 2.1 for further
details.
Menu
Function
window No.
Display POS (positive), NEG (negative) and NET (net) totalizer values.
M00
Display signal strength, signal quality and working status
Display POS totalizer, instantaneous flow rate, velocity, signal strength,
M01
signal quality and working status
M02 Display NEG totalizer, instantaneous flow rate, velocity, signal strength,
Av da Invernada 12 - Vila Congonhas - CEP 04612-060 - São Paulo - SP
Fone / fax 011 5096 4654
Para honra e glória de ΙҲΘγΣ
Page 22 of 66
signal quality and working status
Display NET totalizer, instantaneous flow rate, velocity, signal strength,
M03
signal quality and working status
Display date and time, instantaneous flow rate, signal strength, signal quality
M04
and working status
Display date and time, velocity, signal strength, signal quality and working
M05
status
M06 Display the wave shape of the receiving signal
M07 Display the battery terminal voltage and its estimated lasting time
M08 Display all of the detailed working status, signal strength, signal quality
Display today’s total NET flow, velocity, signal strength, signal quality and
M09
working status
M10 Window for entering the outer perimeter of the pipe
Window for entering the outer diameter of the pipe
M11
Valid range: 0 to 6000mm.
M12 Window for entering pipe wall thickness
Window for entering the inner diameter of the pipe. If pipe outer diameter and
M13 wall thickness are entered correctly, the inner diameter will be calculated
automatically, thus no need to change anything in this window.
Window for selecting pipe material
Standard pipe materials (no need to enter the material sound speed) include:
M14 (0) carbon steel (1) stainless steel (2) cast iron (3) ductile iron
(4) copper (5) PVC (6) aluminum (7) asbestos
(8) fiberglass
M15 Window for entering the sound speed of non-standard pipe materials
Window for selecting the liner material. Select none for pipes without any
liner.
Standard liner materials (no need to enter liner sound speed) include:
M16
(1) Tar Epoxy (2) Rubber (3) Mortar (4) Polypropylene
(5) Polystryol (6)Polystyrene (7) Polyester (8) Polyethylene
(9) Ebonite (10) Teflon
M17 Window for entering the sound speed of non-standard liner materials
M18 Window for entering the liner thickness, if there is a liner
M19 Window for entering the roughness coefficient of the pipe inner surface
Window for selecting fluid type
For standard liquids (no need to enter liquid sound speed) include:
(0) Water (1) Sea Water (2) Kerosene (3) Gasoline
M20
(4) Fuel oil (5) Crude Oil (6) Propane at -45C (7) Butane at 0C
(8)Other liquids (9) Diesel Oil (10)Caster Oil (11)Peanut Oil
(12) #90 Gasoline (13) #93 Gasoline (14) Alcohol (15) Hot water at 125C
M21 Window for entering the sound speed of non-standard liquids
M22 Window for entering the viscosity of non-standard liquids
M23 Window for selecting transducer type
Av da Invernada 12 - Vila Congonhas - CEP 04612-060 - São Paulo - SP
Fone / fax 011 5096 4654
Para honra e glória de ΙҲΘγΣ
Page 23 of 66
There are 14 different types of transducers for selection.
If the π type spool-piece transducers are used, the user needs to configure
the 3 transducer parameters.
Otherwise, the user needs to configure the 4 transducer parameters.
Window for selecting the transducer mounting methods
M24 Four methods can be selected:
(0) V-method (1) Z-method (2) N-method (3) W-method
M25 Display the transducer mounting spacing or distance
Entry to store the pipe parameters into the internal NVRAM (non-volatile
M26
memory)
M27 Entry to read the previously saved pipe parameters
Entry to determine whether or not to keep the last correct value when poor
M28
signal condition occurs. YES is the factory default
Window to set the threshold below which the receiving signal is defined as
M29
poor. Valid number: from 000 to 999. 0 is the factory default
Window for selecting unit system. ‘Metric’ is the factory default. The
M30 conversion from English to Metric or vice versa will not affect the unit for
totalisers.
Window for selecting flow rate unit system.
Flow rate can be in
0. Cubic meter short for (m3)
1. Liter (l)
2. USA gallon (gal)
3. Imperial Gallon (igl)
M31 4. Million USA gallon (mgl)
5. Cubic feet (cf)
6. USA liquid barrel (bal)
7. Imperial liquid barrel (ib)
8. Oil barrel (ob)
The flow unit in terms of time can be per day, per hour, per minute or per
second. So there are 36 different flow rate units in total for selection.
M32 Window for selecting the totalisers’ unit
Window for setting the totaliser multiplying factor
M33
The multiplying factor ranges from 0.001 to 10000
M34 Turn on or turn off the NET totaliser
M35 Turn on or turn off the POS totaliser
M36 Turn on or turn off the NEG totaliser
(1) Totaliser reset
(2) Restore the factory default settings. Press the dot key followed by the
M37
backspace key. Attention, it is recommended to make notes on the
parameters before doing the restoration.
Manual totaliser used for calibration. Press any key to start and press the key
M38
again to stop the totaliser.
M39 Language selection, Chinese or English.
M40 Flow rate damper setup. The damping parameter ranges from 0 to 999
When powered on, the ultrasonic flowmeter automatically starts the self-diagnosis process to
find if there are any hardware and software problems. If a problem is identified, an error
message will be displayed. The following table shows the possible error messages, the
corresponding causes and their solutions.
Error message Causes Solutions
ROM Testing Error (1)Reboot the system
Problem with the software
Data Testing Error (2)Contact the manufacturer.
When this message is displayed,
Data Storing Error User-entered parameters get lost. press ENT key to restore the
default configuration.
System Clock Problem with the system clock or (1)Power on again
Slow or Fast Error the crystal oscillator. (2)Contact the manufacturer
Initialize the calendar in menu
Date Time Error Problem with the system calendar
window M61
Reboot repetitively Hardware problems Contact the manufacturer
The ultrasonic flow meter will show an Error Code (a single letter like I, R, etc.) in the lower
right corner on menu windows M00, M01, M02, M03, M90 and M08. When any abnormal Error
Code shows, counter-measures should be taken.
Message
Error
displayed on Causes Solutions
code
window M08
R System Normal No error
(1)Unable to receive
signals
(2)Transducers installed
improperly
(1) Adjust measuring location
(3)Loosen contact or not
(2)Polish the pipe surface and
enough couplant between
clean the spot
I No Signal trasducer and pipe outer
(3)Make sure the couplant is
surface.
enough
(4)Pipe liners are too thick
(4)Check the transducer cables
or the deposition inside of
the pipe is too thick.
(5)Transducer cables are
not properly connected
J Hardware Error Hardware problem Contact the manufacturer
Av da Invernada 12 - Vila Congonhas - CEP 04612-060 - São Paulo - SP
Fone / fax 011 5096 4654
Para honra e glória de ΙҲΘγΣ
Page 28 of 66
(1)Poor signal detected
(2)Transducers installed
improperly (1)Adjust measuring location
(3)Too much fouling (2)Polish the pipe surface and
(corrosion, deposition, clean the spot
H PoorSig Detected
etc.) (3)Make sure the couplant is
(4)The pipe liner is too enough
thick. (4)Check the transducer cables
(5)Problem with
transducer cables
The actual frequency for Check the values entered in
the Frequency Output is window M66, M67, M68 and
Q Frequ OutputOver
out of the range specified M69, and use a larger value in
by the user M69
System RAM
(1) Temporary problems
Error
with RAM, RTC (1) Turn on the power again
F Date Time Error
(2) Permanent problems (2) contact the manufacturer
CPU or IRQ Error
with hardware
ROM Parity Error
Instrument is in the
1 progress of adjusting the
2 Adjusting Gain gain for the signal, and No need for action
3 the number indicates the
progressive steps
Relocate the meter to where the
No liquid inside the pipe
K Empty pipe pipe is full of liquid
Incorrect setup in M29
Enter 0 in M29
Q: Why the instrument displays 0.0000 flow rate while the liquid in the pipe is actually flowing?
The signal strength is checked to be good (the working status is “R”) and the signal quality Q
has a satisfactory value.
A: The problem is likely to be caused by the incorrect “Zero Point” setting. The user may have
conducted the “Zero Point” setup while the flow was not standstill. To solve this problem, use
the ‘Reset Zero’ function in menu window M43 to clear the zero point.
Q: The displayed flow rate is much lower or much higher than the actual flow rate in the pipe
under normal working conditions. Why?
A:The entered offset value might be wrong. Enter ‘0’ offset in window M44.
Incorrect transducer installation. Re-install the transducers carefully.
The ‘Zero Point’ is wrong. Go to window M42 and redo the “Zero Point” setup. Make sure that
the flow inside the pipe is standstill. No velocity is allowed during this setup process.
Q: Why the battery can not work as long as the time indicated on M07?
A:The battery may have come to the end of its service life. Replace it with a new one.
Av da Invernada 12 - Vila Congonhas - CEP 04612-060 - São Paulo - SP
Fone / fax 011 5096 4654
Para honra e glória de ΙҲΘγΣ
Page 29 of 66
New battery is not compatible with the battery estimating software. The software needs to be
upgraded. Please contact the manufacturer.
The battery has not been fully charged.
There is indeed a time difference between the actual working time and the estimated one,
especially when the terminal voltage is in the range from 3.70 to 3.90 volts. Therefore, the
estimated working time is for reference only
The ultrasonic flow meter integrates a standard RS-232C communication interface and a
complete set of communication protocol.
The protocol is comprised of a set of basic commands that are strings in ASCII format, ending
with a carriage (CR) and line feed (LF). Commonly used commands are listed in the following
table.
Notes:
1. (CR) stands for Carriage Return. Its ASCII code is 0DH. (LF) strands for Line Feed. Its
ASCII code is 0AH.
2. “d” stands for a digit number of 0~9. 0 is expressed as +0.000000E+00.
3. “d” stands for a digit number of 0~9. The number before “E” is integer.
4. Working status code, 1-6 letters. Refer to Table 5.2 for error code.
5. “s” is “ON”, “OFF” or “UD’”. For instance, “TR:ON, RL:UD” means that the OCT is in
closed state and RELAY is not used.
6. @ stands for key value. For instance, value 30H means key “0”, command “M4” is
equivalent to press key “4”.
7. “a” stands for current value, a digit number of 0~20. For instance, A02.34, A00.2
8. “dddddddt” stands for 8-digit electronic serial number. “t” stands for flow meter type.
9. If there are more than one flow meters in a network, all the basic commands must be
prefixed with N or W. Otherwise, multiple flow meters may reply to the same request.
10. Adding a GSM module to the flow meter allows the user to check flow meter flow rate and
other parameters from a cell phone.
11. Used for visiting the printer buffer content.
(1) Prefix P
The prefix P can be added before any command in the above table to have the returning data
followed with two bytes of CRC check sum, which is the adding sum of the original character
string.
Take command DI+(CR) (Return POS Totaliser Value) as an example. The binary data for
DI+(CR) is 44H, 49H, 2BH and 0DH. Assume the return value of this command is
+1234567E+0m3(CR)(LF) ( the string in hexadecimal is 2BH, 31H, 32H, 33H, 34H, 35H, 36H,
37H, 45H, 2BH, 30H, 6DH, 33H, 20H, 0DH, 0AH).
Then, the P-prefixed command, PDI+(CR), would return +1234567E+0m3!F7(CR)(LF). The ‘!’
acts as the starter of the check sum (F7) which is obtained by adding up the string, 2BH+
31H+ 32H+ 33H+ 34H+ 35H+ 36H+ 37H+ 45H+ 2BH+ 30H+ 6DH+ 33H+ 20H = (2) F7H.
Please note that it is allowed to not have data entry or to have SPACES (20H) character
before the ‘!’ character.
Av da Invernada 12 - Vila Congonhas - CEP 04612-060 - São Paulo - SP
Fone / fax 011 5096 4654
Para honra e glória de ΙҲΘγΣ
Page 32 of 66
(2) Prefix W
The prefix W is used for networking commands. The format of a networking command is:
W + IDN address string + basic command.
The IDN address should have a value between 0 and 65534, except 13(0DH), 10 (0AH),
42(2AH,*), 38(26H, &).
For example, if you want to visit the instantaneous flow velocity of device IDN=12345, the
following command should be sent to this device: W12345DV(CR). The corresponding binary
code is 57H, 31H, 32H, 33H, 34H, 35H, 44H, 56H, 0DH.
(3) Prefix N
The prefix N is a single byte IDN network address, not recommended in a new design.
(4) Command binder &
The & command binder or connector can connect up to 6 basic commands to form a longer
command so that it will make the programming much easier.
For example, assume we want device IDN=4321 to return the flow rate, velocity and POS
totaliser value simultaneously. The combined command would be W4321DQD&DV&DI+(CR),
and the result would be:
+1.234567E+12m3/d(CR)
+3.1235926E+00m/s(CR)
+1234567E+0m3(CR)
The protocol provides the capability of virtual key-pressing. A remote RS-232C terminal can
send an ‘M’ command along with a key code to simulate the scenario that the key is pressed
through the keypad of the flow meter. This functionality allows the user to operate the flow
meter in the office far away from the testing site.
For example, the command “M1” is sent to the flow meter through the RS-232C link, the flow
meter will treat the command as if the user has pressed the 1 key through the keypad.
The ASCII codes and corresponding key values of the keypad keys are listed in the following
table.
Hexadecima
Decimal ASCII Hexadecimal Decimal ASCII
Key l Key
Key code Code Key code Key code Code
Key code
0 30H 48 0 8 38H 56 8
1 31H 49 1 9 39H 57 9
2 32H 50 2 . 3AH 58 :
3 33H 51 3 ◄ 3BH,0BH 59 ;
4 34H 52 4 MENU 3CH,0CH 60 <
7.1 Warranty
The products manufactured by our company are warranted to be free from defects in materials
and workmanship for a period of one year from the date of shipment to the original purchaser.
Our obligation should be limited to restoring the meter to normal operation or replacing the
meter, at our company’s choice, and shall be conditioned upon receiving written notice of any
alleged defect within 10 days after its discovery. It will determine if the return of the meter is
necessary. If it is , the user should be responsible for the one-way shipping fee from the
customer to the manufacturer.
Our company is not liable to any defects or damage attributable to miss usage, improper
installation,out-of-spec operating conditions, replacement of unauthorized parts and acts of
nature.Besides, fuses and batteries are not part of this warranty.
7.2 Service
For operational problems, please contact the technical support department by telephone, fax,
email or internet. In most cases, problems could be solved immediately.
For any hardware failure of the instrument, we recommend our customers to send back the
instrument for service. Please contact the technical support department with the model
number and serial number of the unit before sending the unit back to us.
Take notice that the cost for repairing can only be determined after receipt and inspection of
the instrument. A quotation will be sent to the customer before proceeding with the service.
Important Notice for Product Return
Before returning the instrument for warranty repair or service, please read the following
carefully:
1. If the return item has been exposed to nuclear or other radioactive environment, or has
been in contact with hazardous material which could pose any danger to our personnel, the
unit cannot be serviced.
2. If the return item has been exposed to or in contact with dangerous materials, but has been
certified as hazard-free device by a recognized organization, you are required to supply the
certification for the service.
3. If the return item does not have a Return Material Authorization associated, it will be sent
back without any service conducted.
When the battery is unable to power the instrument for 2 to 3 hours after it is fully recharged,
this usually indicates that the battery is near its product life and needs to be replaced. Please
consult the manufacturer for replacing the battery pack.
Classification: Copper tube is classified into four different specification types based on wall
thickness for a specific outside diameter. The tables provided below are for reference sizing
based on application:
PN 6 PN 9
PN 12 PN 18
Wall Wall
PIPE Wall Wall Conver
Thicknes I/D Thicknes I/D I/D I/D
(mm O/D Thickness Thickness t to
s (mm) s (mm) (mm) (mm)
) Min (mm) Min (mm) inches
Min (mm) Min (mm)
Max (mm) Max (mm)
Max (mm) Max (mm)
1.4 1.7
25 33.40 33.70 - - - 30.45 1.70 2.10 29.75 2.50 3.00 28.05 -
0 0
1.7 2.1
32 42.10 42.40 - - - 38.45 2.20 2.60 37.45 3.20 3.70 35.35 -
0 0
125 140.0 140.4 - - - 5.5 6.3 128.4 7.20 8.10 124.9 - - - 5"
160.0 160.5 4.2 4.2 151.2 6.3 7.1 146.8 142.6 12.0 13.6 134.6
150 8.30 9.30 6"
0 0 0 0 5 0 0 5 5 0 0 5
225.0 225.6 5.4 6.1 213.8 7.9 8.9 208.5 10.5 11.7 203.1
200 - - - 8"
0 0 0 0 0 0 0 0 0 0 0
Table A1: Standard ANSI Pipe Size Data for Carbon Steel and Stainless Steel Pipe
0.095 XS 80 80S
0.065 - - 10S
0.119 XS 80 80S
0.065 - - 10S
0.126 XS 80 80S
Av da Invernada 12 - Vila Congonhas - CEP 04612-060 - São Paulo - SP
Fone / fax 011 5096 4654
Para honra e glória de ΙҲΘγΣ
Page 38 of 66
0.065 - - 5S
0.083 - - 10S
0.187 - 160 -
0.294 XXS - -
Table A1: (Continued) Standard ANSI Pipe Size Data for Carbon Steel and Stainless Steel
Pipe
0.065 - - 5S
0.109 - - 10S
0.154 STD 40 40S
2 2.375
0.218 XS 80 80S
0.344 - 160 -
0.436 XXS - -
Table A1: (Continued) Standard ANSI Pipe Size Data for Carbon Steel and Stainless Steel
Pipe
Table A1: (Continued) Standard ANSI Pipe Size Data for Carbon Steel and Stainless Steel
Pipe
0.218 - - 10S
0.250 - 10 -
0.375 STD 20 -
0.500 XS 30 -
0.594 - 40 -
20 20.000
0.812 - 60 -
1.031 - 80 -
1.281 - 100 -
1.500 - 120 -
1.750 - 140 -
1.969 - 160 -
0.188 - - 5S
0.218 - - 10S
0.250 - 10 -
0.375 STD 20 -
0.500 - 40 -
22 22.000 0.875 - 60 -
1.125 - 80 -
1.375 - 100 -
1.625 - 120 -
1.875 - 140 -
2.215 - 160 -
0.312 - 10 10S
0.375 STD - -
30 30.000
0.500 XS 20 -
0.625 - 30 -
0.750 - 40 -
Table A1: (Continued) Standard ANSI Pipe Size Data for Carbon Steel and Stainless Steel
0.375 STD - -
32 32.000 0.500 XS 20 -
0.625 - 30 -
0.688 - 40 -
0.344 - 10 10S
0.375 STD - -
34 34.000 0.500 XS 20 -
0.625 - 30 -
0.688 - 40 -
0.312 - 10 10S
0.375 STD - -
36 36.000 0.500 XS 20 -
0.625 - 30 -
0.750 - 40 -
0.375 STD - -
0.500 XS 20 -
42 42.000
0.625 - 30 -
0.750 - 40 -
0.375 STD - -
48 48.000
0.500 XS - -
Pipe Size Outer Wall Outer Wall Outer Wall Outer Wall Thickness
Diameter Thickness Diameter Thickness Diameter Thickness Diameter (in.)
(in.)
(in.) (in.) (in.) (in.) (in.) (in.) (in.)
3
4
6 7.22 0.58 7.22 0.61 7.38 0.65 7.38 0.69
8 9.42 0.66 9.42 0.66 9.60 0.75 9.60 0.80
10 11.60 0.74 11.60 0.80 11.84 0.86 11.84 0.92
12 13.78 0.82 13.78 0.89 14.08 0.97 14.08 1.04
14 15.98 0.90 15.98 0.99 16.32 1.07 16.32 1.16
16 18.16 0.90 18.16 1.08 18.54 1.18 18.54 1.27
18 20.34 1.07 20.34 1.17 20.78 1.28 20.78 1.39
Table A5: Sound Speed in Water at atmosphere pressure. Unit T (Deg C) V (m/s)
T V T V T V T V
Chemical Kinematic
Substance Specific Sound Speed v/ºC
-6
Formula Viscosity×10
Gravity
2 2
m/s ft/s m/s/ºC m /s ft /s
1.082
Acetic anhydride(22) (CH3CO)2O 1,180 3,871.4 2.5 0.769 8.274
(20 ºC)
1.082
Acetic acid,anhydride(22) (CH3CO)2O 1,180 3,871.4 2.5 0.769 8.274
(20 ºC)
Acetic acid,ethyl ester(33) C4H8 O2 0.901 1,085 3,559.7 4.4 0.467 5.025
1.156 12.438
Acetylen tetrachloride(47) C2H2CL4 1.595 1,147 3,763.1
(15 ºC) (59ºF)
1,729
5,672.6 0.292 3.141
Ammonia(35) NH3 0.771 (-33 ºC) 6.68
(-27ºF) (-33 º C) (-27ºF)
(d)
962.6 3158.2
Amorphous Polyolefin 0.98 26,600 286.000
(190 º C) (374ºF)
Av da Invernada 12 - Vila Congonhas - CEP 04612-060 - São Paulo - SP
Fone / fax 011 5096 4654
Para honra e glória de ΙҲΘγΣ
Page 53 of 66
t-Amyl alcohol C5H12O 0.81 1,204 3,950.1 4.374 47.064
Chemical Kinematic
Substance Sound Speed v/ºC -6
Formula Specific Viscosity×10
Gravity 2 2
m/s ft/s m/s/ºC m /s ft /s
0.529 5.692
1-Chloro-butane(22,46) C4H9CL 0.887 1,140 3,740.2 4.57
(15ºC) (59ºF)
Chloro-diFluoromethane 1.491 893.9 2,932.7
CHCLF2 4.79
(3)(Freon 22) (-69ºC) (-50ºC) (-58ºF)
1.26 13.55
Decane(46) C10H20 0.730 1,252 4,107.6
(20ºC) (68ºF)
Chemical Kinematic
Substance Sound Speed v/ºC -6
Formula Specific Viscosity×10
Gravity 2 2
m/s ft/s m/s/ºC m /s ft /s
8.5
Diamylamine C10H23N 1.256 4,120.7 3.9
(68°F)
0.79
1,2Dibromo-ethane(47) C2H4Br2 2.18 995 3,264.4
(20ºC)
trans-1,2-Dibromoethene
C2H2Br2 2.231 935 3,067.6
(47)
Dichloeodifluoromethane 1.516
CCl2F2 774.1 2,539.7 4.24
(3)(Freon12) (40ºC)
cis1,2-Dichloro-ethene
CHCl2F 1.284 1,061 3,481
(3,47)
trans1,2-Dichloro-ethene
C4Cl2F6 1.257 1,010 3,313.6
(3,47)
Dichloro-fluoromethane 1.426 891 2,923.2
C4H8Cl2 3.97
(3)(Freon21) (0ºC) (0ºC) (32°F)
1-2-Dichlorohexafluoro-
CClF2-CClF2 1.654 669 2,914.9
cyclobutane(47)
Diethylene glycol
C4H9(NF2)2 0.988 1,458 4,783.5
Monoethyl ether
1,2-bis(DiFluoramino)
C10H23N 1.216 1,000 3,280.8
butane(43)
1,2-bis(DiFluoramino)-
C2H4Br2 1.213 900 2,952.8
2-methylpropane(43)
1,2-bis(DiFluoramino)
C2H2Br2 1.265 960 3,149.6
propane(43)
2,2-bis(Difluoromino
C3H6(NF2)2 1.254 890 2920
propane(43)
17.208 185.158
Ethylene glycol C2H6O2 1.113 1,658 5439.6 2.1
(20ºC) (68°F)
Gravity 2 2
m/s ft/s m/s/ºC m /s ft /s
0.125 183 600.4
Helium(45) He4 0.025 269
(-269ºC) (-269ºC) (-452°F)
0.684 0.598 6.434
Heptane(22,23) C7H16 1,131 3,710.6 4.25
(209ºC) (209ºC) (68°F)
0.684
n-Heptane(29,33) C7H16 1,180 3,871.3 4.0
(20ºC)
Hexachloro-
C5Cl6 1.7180 1,150 3,773
Cyclopentadiene(47)
0.773 4.32 46.483
Hexadecane(23) C16H34 1,338 4,389.8 3.71
(20ºC) (20ºC) (68°F)
70.69 760.882
Hexalin C6H12O 0.962 1,454 4,770.3 3.6
(17ºC) (63°F)
0.649
n-Hexane(29,33) C6H14 1,079 3,540 4.53
(20ºC)
2,5Hexanedione C6H10O2 0.729 1,399 4,589.9 3.6
0.62
lsopentane(36) C5H12 980 3,215.2 4.8 0.34 3.658
(20ºC)
0.758 1,170 3,838.6
lsopropano(46) C3H8O 2.718 29.245
(20ºC) (20ºC) (68°F)
0.758 1,170 3,838.6
Lsopropyl alcohol(46) C3H8O 2.718 29.245
(20ºC) (20ºC) (68°F)
2,485 8,152.9
Lithium fluoride(42) LiF 1.29
(900ºC) (1652°F)
1,449 4,753.9
Mercury(45) Hg 13.594 0.114 1.226
(24ºC) (75°F)
0.902 9.705
Methy-chloroform(47) C2H3Cl3 1.33 985 3,231.6
(20ºC) (68°F)
Gravity 2 2
m/s ft/s m/s/ºC m /s ft /s
Chemical Kinematic
Substance Sound Speed v/ºC -6
Formula Specific Viscosity×10
Gravity 2 2
m/s ft/s m/s/ºC m /s ft /s
1,125 3,691
Phthalardione C8H4O3
(152ºC) (306°F)
1,125 3,691
Phthalic acid,anhydride C8H4O3
(152ºC) (306°F)
1,125 3,691
Phthalicanhydride C8H4O3
(152ºC) (306°F)
1,099.8 3,608.4
PolyterpeneResin 0.77 39,000 419,500
(190ºC) (374°F)
Chemical Kinematic
Substance Sound Speed v/ºC -6
Formula Specific Viscosity×10
Gravity 2 2
m/s ft/s m/s/ºC m /s ft /s
0.992 10.673
Pyridne C6H5N 0.982 1,415 4,642.4 4.1
(20 ) (68°F)
828.3 2,717.5
Refrigerant11(3,4) CCl3F 1.49 3.56
(0ºC) (32°F)
1.516 774.1 2,539.7
Refrigerant12(3) CCl2F2 4.24
(-40ºC) (-40ºC) (-40°F)
1.75 875.24 2,871.5
Refrigerant14(14) CF4 6.61
(-150ºC) (-150ºC) (-238°F)
1.426 891 2,923.2
Refrigerant21(3) CHCl2F 3.97
(0ºC) (0ºC) (32°F)
1.491 893.9 2,932.7
Refrigerant22(3) CHClF2 4.79
(-69ºC) (50ºC) (122°F)
Av da Invernada 12 - Vila Congonhas - CEP 04612-060 - São Paulo - SP
Fone / fax 011 5096 4654
Para honra e glória de ΙҲΘγΣ
Page 63 of 66
783.7 2,571.2
Refrigerant113(3) CCl2F-CClF2 1.563 3.44
(0ºC) (32°F)
665.3 2,182.7
Refrigerant114(3) CClF2-CClF2 1.455 3.73
(-10ºC) (14°F)
656.4 2,153.5
Refrigerant115(3) C2ClF5 4.42
(-50ºC) (-58°F)
1.62 574 1,883.2
RefrigerantC318(3) C4F8 3.88
(-20ºC) (-10ºC) (41°F)
1,072 3,517.1
Selenium(8) Se 0.68
(250ºC) (482°F)
2,082 6,830.7
Sodiumfluoride(42) NaF 0.877 1.32
(1000ºC) (1832°F)
1.884 1,763.3 5,785.1 1.37 14.74
Sodiumfluoride(48) NaNO3 0.74
(336ºC) (336ºC) (637°F) (336ºC) (637 ºF)
1.805 1,876.8 6,157.5
Sodiumfluoride(48) NaNO2
(292ºC) (292ºC) (558°F)
1,177 3,861.5
Sulfur(7,8,10) S -1.13
(250ºC) (482°F)
991 3,251.3
Tellurium(7) Te 0.73
(450ºC) (842°F)
Table A6: (Continued) Sound Speed in Liquids
Chemical Kinematic
Substance Sound Speed v/ºC -6
Formula Specific Viscosity×10
Gravity 2 2
m/s ft/s m/s/ºC m /s ft /s
1,1,2,2-Tetrabromo-
C2H2Br4 2.966 1,027 3,369.4
ethane(47)
1,1,2,2-Tetrachloro- 1.156 12.438
C2H2Cl4 1.595 1,147 3,763.4
ethane(67) (15ºC) (59°F)
1.553 1,170 3,838.6
Tetrachloroethane(46) C2H2Cl4 1.19 12.804
(20ºC) (20ºC) (68°F)
Tetrachlor-Methane 1.595
CCl4 926 3,038.1 0.607 6.531
(33,47) (20ºC)
0.763 1,331 4,366.8 2.86 30.773
Tetradecane(46) C14H3O
(20ºC) (20ºC) (68°F) (20ºC) (68°F)
2.89
Tribromo-methane(46,47) CHBr3 918 3,011.8 0.645 7.037
(20ºC)
1,1,1-Trichloro- 0.902 9.705
C2H3Cl3 1.33 985 3,231.6
ethane(47) (20ºC) (68°F)
Chemical Kinematic
Substance Sound Speed v/ºC -6
Formula Specific Viscosity×10
Gravity 2 2
m/s ft/s m/s/ºC m /s ft /s
1,1,1-Trifluoro-2-
C2HClBrF3 1.869 693 2,273.6
Chloro-2-Bromo-Ethane
1,2,2-Trifluorotrichloro- 783.7 2,571.2
CCl2-CClF2 1.563 3.44
ethane(Freon113) (0ºC) (32°F)
d-1,3,3-
C10H16O 0.947 1,320 4,330.7 0.22 2.367
Trimethylnorcamphor
1,610 5,282.2
Trinitrotoluene(43) C7H5(NO2)3 1.64
(81ºC) (178°F)
Water,distilled(49,50)
H2O 0.996 1,498 4,914.7 -2.4 0.695 10.76
Water,sea
0.791
m-Xylene(46) CH4O 1,076 3,530.2 0.695 7.478
(20ºC)
630 2,067
o-Xylene(29,46 Xe
(-109ºC) (-164°F)
0.868 1,343 4,406.2 0.749 8.059
P-xylene(46) C8H10
(15ºC) (20ºC) (68°F) (15ºC) (59°F)
0.897 0.903 9.716
Xylenehexafluoride C8H10 1,331.5 4,368.4 4.1
(20ºC) (20ºC) (68°F)
1,334 4,376.6
Zinc(7) C8H10 0.662 7.123
(20ºC) (68°F)
1,1,1-Trifluoro-2-Chloro-
C8H4F6 1.37 879 2,883.9 0.613 6.595
2-Bromo-Ethane
1,2,2-Trifluorotrichloro- 3,298 10,820.2
Zn
ethane(Freon113) (450ºC) (842°F)