Osciloscopio
Osciloscopio
com/Measurement/App_Notes/
XYZs/intro.html
Introduction
The oscilloscope is an essential tool if you plan to design or repair electronic equipment. It lets you "see" electrical
signals.
Energy, vibrating particles, and other invisible forces are everywhere in our physical universe. Sensors can convert these
forces into electrical signals that you can observe and study with an oscilloscope. Oscilloscopes let you "see" events that
occur in a split second.
If you are using an oscilloscope for the first time, read this book to get a solid understanding of oscilloscope basics.
Then, read the manual provided with your oscilloscope to learn specific information about how to use it in your work. After
reading this book, you will be able to:
If you come across an unfamiliar term in this book, check the glossary in the back for a definition.
This book serves as a useful classroom aid. It includes vocabulary and multiple choice written exercises on oscilloscope
theory and controls. You do not need any mathematical or electronics knowledge. This book emphasizes teaching you
about oscilloscopes - how they work and how to make them work for you.
Up To: Glossary
Up To: Index
PERMISSION TO USE:
Permission to use, copy and distribute documents delivered from this World Wide Web server and related graphics is
hereby granted, provided that the above copyright notice appear in all copies and that both that copyright notice and this
permission notice appear. All other rights reserved.
Use, duplication, or disclosure by the United States Government is subject to the restrictions set forth in DFARS
252.227-7013(c)1(ii) and FAR 53.337-19.
The name Tektronix, Inc. or the Tektronix logo may not be used in advertising or publicity pertaining to the distribution of
this information without prior, specific written permission. Tektronix, Inc. makes no representations about the suitability of
this information for any purpose. It is provided "AS IS" without express or implied warranty, including, but not limited to, the
implied warranties of merchantability, fitness for a particular purpose, or non-infringement.
In no event shall Tektronix, Inc. be liable for any special, indirect or consequential damages or any damages whatsoever
resulting from loss of use, data or profits, whether in an action of contract, negligence or other tortious action, arising out
of or in connection with the use or performance of this information.
Tektronix, Inc. assumes no responsibility for errors or omissions in this publication or other documents which are
referenced or linked to this publication. Nor does Tektronix make any commitment to update the information contained
herein. This publication and other documents are subject to change without notice.
Table of Contents
Introduction
Why Read This Book?
For More Information
The Oscilloscope
What Can You Do With It?
Analog and Digital
How Does an Oscilloscope Work?
Oscilloscope Terminology
Measurement Terms
Types of Waves
Waveform Measurements
Performance Terms
Setting Up
Grounding
Setting the Controls
Probes
Compensating the Probe
The Controls
Display Controls
Vertical Controls
Horizontal Controls
Trigger Controls
Acquisition Controls for Digital Oscilloscopes
Other Controls
Measurement Techniques
The Display
Voltage Measurements
Time and Frequency Measurements
Pulse and Rise Time Measurements
Phase Shift Measurements
What's Next?
Written Exercises
Written Exercise Answers
Glossary
Index
The Oscilloscope
What is an oscilloscope, what can you do with it, and how does it work? This section answers these fundamental
questions.
The oscilloscope is basically a graph-displaying device - it draws a graph of an electrical signal. In most applications the
graph shows how signals change over time: the vertical (Y) axis represents voltage and the horizontal (X) axis represents
time. The intensity or brightness of the display is sometimes called the Z axis. (See Figure 1.) This simple graph can tell
you many things about a signal. Here are a few:
An oscilloscope looks a lot like a small television set, except that it has a grid drawn on its screen and more controls than
a television. The front panel of an oscilloscope normally has control sections divided into Vertical, Horizontal, and Trigger
sections. There are also display controls and input connectors. See if you can locate these front panel sections in Figures
2 and 3 and on your oscilloscope.
The usefulness of an oscilloscope is not limited to the world of electronics. With the proper transducer , an oscilloscope
can measure all kinds of phenomena. A transducer is a device that creates an electrical signal in response to physical
stimuli, such as sound, mechanical stress, pressure, light, or heat. For example, a microphone is a transducer.
An automotive engineer uses an oscilloscope to measure engine vibrations. A medical researcher uses an oscilloscope
to measure brain waves. The possibilities are endless.
Oscilloscopes also come in analog and digital types. An analog oscilloscope works by directly applying a voltage being
measured to an electron beam moving across the oscilloscope screen. The voltage deflects the beam up and down
proportionally, tracing the waveform on the screen. This gives an immediate picture of the waveform.
In contrast, a digital oscilloscope samples the waveform and uses an analog-to-digital converter (or ADC) to convert the
voltage being measured into digital information. It then uses this digital information to reconstruct the waveform on the
screen.
For many applications either an analog or digital oscilloscope will do. However, each type does possess some unique
characteristics making it more or less suitable for specific tasks.
People often prefer analog oscilloscopes when it is important to display rapidly varying signals in "real time" (or as they
occur).
Digital oscilloscopes allow you to capture and view events that may happen only once. They can process the digital
waveform data or send the data to a computer for processing. Also, they can store the digital waveform data for later
viewing and printing.
Analog Oscilloscopes
When you connect an oscilloscope probe to a circuit, the voltage signal travels through the probe to the vertical system of
the oscilloscope. Figure 6 is a simple block diagram that shows how an analog oscilloscope displays a measured
signal.
Depending on how you set the vertical scale (volts/div control), an attenuator reduces the signal voltage or an amplifier
increases the signal voltage.
Next, the signal travels directly to the vertical deflection plates of the cathode ray tube (CRT). Voltage applied to these
deflection plates causes a glowing dot to move. (An electron beam hitting phosphor inside the CRT creates the glowing
dot.) A positive voltage causes the dot to move up while a negative voltage causes the dot to move down.
The signal also travels to the trigger system to start or trigger a "horizontal sweep." Horizontal sweep is a term referring to
the action of the horizontal system causing the glowing dot to move across the screen. Triggering the horizontal system
causes the horizontal time base to move the glowing dot across the screen from left to right within a specific time interval.
Many sweeps in rapid sequence cause the movement of the glowing dot to blend into a solid line. At higher speeds, the
dot may sweep across the screen up to 500,000 times each second.
Together, the horizontal sweeping action and the vertical deflection action traces a graph of the signal on the screen. The
trigger is necessary to stabilize a repeating signal. It ensures that the sweep begins at the same point of a repeating
signal, resulting in a clear picture as shown in Figure 7.
In conclusion, to use an analog oscilloscope, you need to adjust three basic settings to accommodate an incoming
signal:
The attenuation or amplification of the signal. Use the volts/div control to adjust the amplitude of the signal before it
is applied to the vertical deflection plates.
The time base. Use the sec/div control to set the amount of time per division represented horizontally across the
screen.
The triggering of the oscilloscope. Use the trigger level to stabilize a repeating signal, as well as triggering on a
single event.
Also, adjusting the focus and intensity controls enables you to create a sharp, visible display.
Digital Oscilloscopes
Some of the systems that make up digital oscilloscopes are the same as those in analog oscilloscopes; however, digital
oscilloscopes contain additional data processing systems. (See Figure 8.) With the added systems, the digital
oscilloscope collects data for the entire waveform and then displays it.
When you attach a digital oscilloscope probe to a circuit, the vertical system adjusts the amplitude of the signal, just as in
the analog oscilloscope.
Next, the analog-to-digital converter (ADC) in the acquisition system samples the signal at discrete points in time and
converts the signal's voltage at these points to digital values called sample points . The horizontal system's sample clock
determines how often the ADC takes a sample. The rate at which the clock "ticks" is called the sample rate and is
measured in samples per second.
The sample points from the ADC are stored in memory as waveform points . More than one sample point may make up
one waveform point.
Together, the waveform points make up one waveform record . The number of waveform points used to make a
waveform record is called the record length . The trigger system determines the start and stop points of the record. The
display receives these record points after being stored in memory.
Depending on the capabilities of your oscilloscope, additional processing of the sample points may take place,
enhancing the display. Pretrigger may be available, allowing you to see events before the trigger point.
Fundamentally, with a digital oscilloscope as with an analog oscilloscope, you need to adjust the vertical, horizontal, and
trigger settings to take a measurement.
Sampling Methods
The sampling method tells the digital oscilloscope how to collect sample points. For slowly changing signals, a digital
oscilloscope easily collects more than enough sample points to construct an accurate picture. However, for faster
signals, (how fast depends on the oscilloscope's maximum sample rate) the oscilloscope cannot collect enough
samples. The digital oscilloscope can do two things:
It can collect a few sample points of the signal in a single pass (in real-time sampling mode) and then use
interpolation . Interpolation is a processing technique to estimate what the waveform looks like based on a few
points.
It can build a picture of the waveform over time, as long as the signal repeats itself (equivalent-time sampling
mode).
Digital oscilloscopes use interpolation to display signals that are so fast that the oscilloscope can only collect a few
sample points. Interpolation "connects the dots."
Linear interpolation simply connects sample points with straight lines. Sine interpolation (or sin x over x interpolation)
connects sample points with curves. (See Figure 10.) Sin x over x interpolation is a mathematical process similar to the
"oversampling" used in compact disc players. With sine interpolation, points are calculated to fill in the time between the
real samples. Using this process, a signal that is sampled only a few times in each cycle can be accurately displayed or,
in the case of the compact disc player, accurately played back.
Equivalent-Time Sampling
Some digital oscilloscopes can use equivalent-time sampling to capture very fast repeating signals. Equivalent-time
sampling constructs a picture of a repetitive signal by capturing a little bit of information from each repetition. (See Figure
11.) You see the waveform slowly build up like a string of lights going on one-by-one. With sequential sampling the
points appear from left to right in sequence; with random sampling the points appear randomly along the waveform.
Oscilloscope Terminology
Learning a new skill often involves learning a new vocabulary. This idea holds true for learning how to use an
oscilloscope. This section describes some useful measurement and oscilloscope performance terms.
Measurement Terms
The generic term for a pattern that repeats over time is a wave - sound waves, brain waves, ocean waves, and voltage
waves are all repeating patterns. An oscilloscope measures voltage waves. One cycle of a wave is the portion of the
wave that repeats. A waveform is a graphic representation of a wave. A voltage waveform shows time on the horizontal
axis and voltage on the vertical axis.
Waveform shapes tell you a great deal about a signal. Any time you see a change in the height of the waveform, you know
the voltage has changed. Any time there is a flat horizontal line, you know that there is no change for that length of time.
Straight diagonal lines mean a linear change - rise or fall of voltage at a steady rate. Sharp angles on a waveform mean
sudden change. Figure 1 shows common waveforms and Figure 2 shows some common sources of waveforms.
Types of Waves
You can classify most waves into these types:
Sine waves
Square and rectangular waves
Triangle and sawtooth waves
Step and pulse shapes
Sine Waves
The sine wave is the fundamental wave shape for several reasons. It has harmonious mathematical properties - it is
the same sine shape you may have studied in high school trigonometry class. The voltage in your wall outlet varies as a
sine wave. Test signals produced by the oscillator circuit of a signal generator are often sine waves. Most AC power
sources produce sine waves. (AC stands for alternating current, although the voltage alternates too. DC stands for
direct current, which means a steady current and voltage, such as a battery produces.)
The damped sine wave is a special case you may see in a circuit that oscillates but winds down over time.
The rectangular wave is like the square wave except that the high and low time intervals are not of equal length. It is
particularly important when analyzing digital circuitry.
A collection of pulses travelling together creates a pulse train. Digital components in a computer communicate with each
other using pulses. Pulses are also common in x-ray and communications equipment.
Figure 6 shows examples of step and pulse shapes and a pulse train.
Waveform Measurements
You use many terms to describe the types of measurements that you take with your oscilloscope. This section describes
some of the most common measurements and terms.
Voltage
Voltage is the amount of electric potential (a kind of signal strength) between two points in a circuit. Usually one of these
points is ground (zero volts) but not always - you may want to measure the voltage from the maximum peak to the
minimum peak of a waveform, referred to at the peak-to-peak voltage. The word amplitude commonly refers to the
maximum voltage of a signal measured from ground or zero volts. The waveform shown in Figure 8 has an amplitude of
one volt and a peak-to-peak voltage of two volts.
Phase
Phase is best explained by looking at a sine wave. Sine waves are based on circular motion and a circle has 360
degrees. One cycle of a sine wave has 360 degrees, as shown in Figure 8. Using degrees, you can refer to the phase
angle of a sine wave when you want to describe how much of the period has elapsed.
Phase shift describes the difference in timing between two otherwise similar signals. In Figure 9, the waveform labeled
"current" is said to be 905 out of phase with the waveform labeled "voltage," since the waves reach similar points in their
Wednesday, March 25, 1998 Page: 4
http://www.tek.com/Measurement/App_Notes/
XYZs/terminology.html
cycles exactly 1/4 of a cycle apart (360 degrees/4 = 90 degrees). Phase shifts are common in electronics.
Performance Terms
The terms described in this section may come up in your discussions about oscilloscope performance. Understanding
these terms will help you evaluate and compare your oscilloscope with other models.
Bandwidth
The bandwidth specification tells you the frequency range the oscilloscope accurately measures.
As signal frequency increases, the capability of the oscilloscope to accurately respond decreases. By convention, the
bandwidth tells you the frequency at which the displayed signal reduces to 70.7% of the applied sine wave signal. (This
70.7% point is referred to as the "-3 dB point," a term based on a logarithmic scale.)
Rise Time
Rise time is another way of describing the useful frequency range of an oscilloscope. Rise time may be a more
appropriate performance consideration when you expect to measure pulses and steps. An oscilloscope cannot
accurately display pulses with rise times faster than the specified rise time of the oscilloscope.
Vertical Sensitivity
The vertical sensitivity indicates how much the vertical amplifier can amplify a weak signal. Vertical sensitivity is usually
given in millivolts (mV) per division. The smallest voltage a general purpose oscilloscope can detect is typically about 2
mV per vertical screen division.
Sweep Speed
For analog oscilloscopes, this specification indicates how fast the trace can sweep across the screen, allowing you to
see fine details. The fastest sweep speed of an oscilloscope is usually given in nanoseconds/div.
Gain Accuracy
The gain accuracy indicates how accurately the vertical system attenuates or amplifies a signal. This is usually listed as a
percentage error.
Sample Rate
On digital oscilloscopes, the sampling rate indicates how many samples per second the ADC (and therefore the
oscilloscope) can acquire. Maximum sample rates are usually given in megasamples per second (MS/s). The faster the
oscilloscope can sample, the more accurately it can represent fine details in a fast signal. The minimum sample rate
may also be important if you need to look at slowly changing signals over long periods of time. Typically, the sample rate
changes with changes made to the sec/div control to maintain a constant number of waveform points in the waveform
record.
Record Length
The record length of a digital oscilloscope indicates how many waveform points the oscilloscope is able to acquire for
one waveform record. Some digital oscilloscopes let you adjust the record length. The maximum record length depends
on the amount of memory in your oscilloscope. Since the oscilloscope can only store a finite number of waveform points,
there is a trade-off between record detail and record length. You can acquire either a detailed picture of a signal for a short
period of time (the oscilloscope "fills up" on waveform points quickly) or a less detailed picture for a longer period of time.
Some oscilloscopes let you add more memory to increase the record length for special applications.
Up To: Glossary
Up To: Index
Setting Up
This section briefly describes how to set up and start using an oscilloscope - specifically, how to ground the oscilloscope,
set the controls in standard positions, and compensate the probe.
Grounding
Proper grounding is an important step when setting up to take measurements or work on a circuit. Properly grounding the
oscilloscope protects you from a hazardous shock and grounding yourself protects your circuits from damage.
Grounding the oscilloscope is necessary for safety. If a high voltage contacts the case of an ungrounded oscilloscope,
any part of the case, including knobs that appear insulated, it can give you a shock. However, with a properly grounded
oscilloscope, the current travels through the grounding path to earth ground rather than through you to earth ground.
To ground the oscilloscope means to connect it to an electrically neutral reference point (such as earth ground). Ground
your oscilloscope by plugging its three-pronged power cord into an outlet grounded to earth ground.
Grounding is also necessary for taking accurate measurements with your oscilloscope. The oscilloscope needs to share
the same ground as any circuits you are testing.
Some oscilloscopes do not require the separate connection to earth ground. These oscilloscopes have insulated cases
and controls, which keeps any possible shock hazard away from the user.
Ground Yourself
If you are working with integrated circuits (ICs), you also need to ground yourself. Integrated circuits have tiny conduction
paths that can be damaged by static electricity that builds up on your body. You can ruin an expensive IC simply by walking
across a carpet or taking off a sweater and then touching the leads of the IC. To solve this problem, wear a grounding
strap (see Figure 1). This strap safely sends static charges on your body to earth ground.
Horizontal , and Trigger . Your oscilloscope may have other sections, depending on the model and type (analog or
digital).
Notice the input connectors on your oscilloscope. This is where you attach probes. Most oscilloscopes have at least two
input channels and each channel can display a waveform on the screen. Multiple channels are handy for comparing
waveforms.
Some oscilloscopes have an AUTOSET or PRESET button that sets up the controls in one step to accommodate a signal.
If your oscilloscope does not have this feature, it is helpful to set the controls to standard positions before taking
measurements.
These are general instructions for setting up your oscilloscope. If you are not sure how to do any of these steps, refer to
the manual that came with your oscilloscope. The Controls section describes the controls in more detail.
Probes
Now you are ready to connect a probe to your oscilloscope. It is important to use a probe designed to work with your
oscilloscope. A probe is more than a cable with a clip-on tip. It is a high-quality connector, carefully designed not to pick
up stray radio and power line noise.
Probes are designed not to influence the behavior of the circuit you are testing. However, no measurement device can act
as a perfectly invisible observer. The unintentional interaction of the probe and oscilloscope with the circuit being tested is
called circuit loading . To minimize circuit loading, you will probably use a 10X attenuator (passive) probe.
Wednesday, March 25, 1998 Page: 2
http://www.tek.com/Measurement/App_Notes/
XYZs/setting.html
Your oscilloscope probably arrived with a passive probe as a standard accessory. Passive probes provide you with an
excellent tool for general-purpose testing and troubleshooting. For more specific measurements or tests, many other
types of probes exist. Two examples are active and current probes.
Descriptions of these probes follow, with more emphasis given to the passive probe since this is the probe type that
allows you the most flexibility of use.
The 10X (read as "ten times") attenuator probe minimizes circuit loading and is an excellent general-purpose passive
probe. Circuit loading becomes more pronounced at higher frequencies, so be sure to use this type of probe when
measuring signals above 5 kHz. The 10X attenuator probe improves the accuracy of your measurements, but it also
reduces the amplitude of the signal seen on the screen by a factor of 10.
Because it attenuates the signal, the 10X attenuator probe makes it difficult to look at signals less than 10 millivolts. The
1X probe is similar to the 10X attenuator probe but lacks the attenuation circuitry. Without this circuitry, more interference
is introduced to the circuit being tested. Use the 10X attenuator probe as your standard probe, but keep the 1X probe
handy for measuring weak signals. Some probes have a convenient feature for switching between 1X and 10X
attenuation at the probe tip. If your probe has this feature, make sure you are using the correct setting before taking
measurements.
Many oscilloscopes can detect whether you are using a 1X or 10X probe and adjust their screen readouts accordingly.
However with some oscilloscopes, you must set the type of probe you are using or read from the proper 1X or 10X
marking on the volts/div control.
The 10X attenuator probe works by balancing the probe's electrical properties against the oscilloscope's electrical
properties. Before using a 10X attenuator probe you need to adjust this balance for your particular oscilloscope. This
adjustment is called compensating the probe and is further described in the next section. Figure 3 shows a simple
diagram of the internal workings of a probe, its adjustment, and the input of an oscilloscope.
Figure 4 shows a typical passive probe and some accessories to use with the probe.
Most oscilloscopes have a square wave reference signal available at a terminal on the front panel used to compensate
the probe. You compensate a probe by:
When you compensate the probe, always attach any accessory tips you will use and connect the probe to the vertical
channel you plan to use. This way the oscilloscope has the same electrical properties as it does when you take
measurements.
Up To: Glossary
Wednesday, March 25, 1998 Page: 5
http://www.tek.com/Measurement/App_Notes/
XYZs/controls.html
The Controls
This section briefly describes the basic controls found on analog and digital oscilloscopes. Remember that some
controls differ between analog and digital oscilloscopes; your oscilloscope probably has controls not discussed here.
Display Controls
Display systems vary between analog and digital oscilloscopes. Common controls include:
An intensity control to adjust the brightness of the waveform. As you increase the sweep speed of an analog
oscilloscope, you need to increase the intensity level.
A focus control to adjust the sharpness of the waveform. Digital oscilloscopes may not have a focus control.
A trace rotation control to align the waveform trace with the screen's horizontal axis. The position of your
oscilloscope in the earth's magnetic field affects waveform alignment. Digital oscilloscopes may not have a trace
rotation control.
Other display controls may let you adjust the intensity of the graticule lights and turn on or off any on-screen
information (such as menus).
Vertical Controls
Use the vertical controls to position and scale the waveform vertically. Your oscilloscope also has controls for setting the
input coupling and other signal conditioning, described in this section. Figure 1 shows a typical front panel and on-screen
menus for the vertical controls.
The volts per division (usually written volts/div) setting varies the size of the waveform on the screen. A good general
Wednesday, March 25, 1998 Page: 1
http://www.tek.com/Measurement/App_Notes/
XYZs/controls.html
purpose oscilloscope can accurately display signal levels from about 4 millivolts to 40 volts.
The volts/div setting is a scale factor. For example, if the volts/div setting is 5 volts, then each of the eight vertical divisions
represents 5 volts and the entire screen can show 40 volts from bottom to top (assuming a graticule with eight major
divisions). If the setting is 0.5 volts/div, the screen can display 4 volts from bottom to top, and so on. The maximum voltage
you can display on the screen is the volts/div setting times the number of vertical divisions. (Recall that the probe you use,
1X or 10X, also influences the scale factor. You must divide the volts/div scale by the attenuation factor of the probe if the
oscilloscope does not do it for you.)
Often the volts/div scale has either a variable gain or a fine gain control for scaling a displayed signal to a certain number
of divisions. Use this control to take rise time measurements.
Input Coupling
Coupling means the method used to connect an electrical signal from one circuit to another. In this case, the input
coupling is the connection from your test circuit to the oscilloscope. The coupling can be set to DC, AC, or ground. DC
coupling shows all of an input signal. AC coupling blocks the DC component of a signal so that you see the waveform
centered at zero volts. Figure 2 illustrates this difference. The AC coupling setting is handy when the entire signal
(alternating plus constant components) is too large for the volts/div setting.
The ground setting disconnects the input signal from the vertical system, which lets you see where zero volts is on the
screen. With grounded input coupling and auto trigger mode, you see a horizontal line on the screen that represents zero
volts. Switching from DC to ground and back again is a handy way of measuring signal voltage levels with respect to
ground.
Bandwidth Limit
Most oscilloscopes have a circuit that limits the bandwidth of the oscilloscope. By limiting the bandwidth, you reduce the
noise that sometimes appears on the displayed waveform, providing you with a more defined signal display.
Channel Invert
Most oscilloscopes have an invert function that allows you to display a signal "upside-down." That is, with low voltage at
the top of the screen and high voltage at the bottom.
On analog scopes, multiple channels are displayed using either an alternate or chop mode. (Digital oscilloscopes do not
normally use chop or alternate mode.)
Alternate mode draws each channel alternately - the oscilloscope completes one sweep on channel 1, then one sweep
on channel 2, a second sweep on channel 1, and so on. Use this mode with medium- to high-speed signals, when the
sec/div scale is set to 0.5 ms or faster.
Wednesday, March 25, 1998 Page: 2
http://www.tek.com/Measurement/App_Notes/
XYZs/controls.html
Chop mode causes the oscilloscope to draw small parts of each signal by switching back and forth between them. The
switching rate is too fast for you to notice, so the waveform looks whole. You typically use this mode with slow signals
requiring sweep speeds of 1 ms per division or less. Figure 3 shows the difference between the two modes. It is often
useful to view the signal both ways, to make sure you have the best view.
Math Operations
Your oscilloscope may also have operations to allow you to add waveforms together, creating a new waveform display.
Analog oscilloscopes combine the signals while digital oscilloscopes mathematically create new waveforms. Subtracting
waveforms is another math operation. Subtraction with analog oscilloscopes is possible by using the channel invert
function on one signal and then use the add operation. Digital oscilloscopes typically have a subtraction operation
available. Figure 4 illustrates a third waveform created by adding two different signals together.
Horizontal Controls
Use the horizontal controls to position and scale the waveform horizontally. Figure 5 shows a typical front panel and
on-screen menus for the horizontal controls.
The seconds per division (usually written as sec/div) setting lets you select the rate at which the waveform is drawn
across the screen (also known as the time base setting or sweep speed). This setting is a scale factor. For example, if
the setting is 1 ms, each horizontal division represents 1 ms and the total screen width represents 10 ms (ten divisions).
Changing the sec/div setting lets you look at longer or shorter time intervals of the input signal.
As with the vertical volts/div scale, the horizontal sec/div scale may have variable timing, allowing you to set the horizontal
time scale in between the discrete settings.
The delayed time base requires the setting of a delay time and possibly the use of delayed trigger modes and other
settings not described in this book. Refer to the manual supplied with your oscilloscope for information on how to use
these features.
Trigger Position
The trigger position control may be located in the horizontal control section of your oscilloscope. It actually represents "the
horizontal position of the trigger in the waveform record." Horizontal trigger position control is only available on digital
oscilloscopes.
Varying the horizontal trigger position allows you to capture what a signal did before a trigger event (called pretrigger
viewing ).
Digital oscilloscopes can provide pretrigger viewing because they constantly process the input signal whether a trigger
has been received or not. A steady stream of data flows through the oscilloscope; the trigger merely tells the oscilloscope
to save the present data in memory. In contrast, analog oscilloscopes only display the signal after receiving the trigger.
Pretrigger viewing is a valuable troubleshooting aid. For example, if a problem occurs intermittently, you can trigger on the
problem, record the events that led up to it and, possibly, find the cause.
Magnification
Wednesday, March 25, 1998 Page: 4
http://www.tek.com/Measurement/App_Notes/
XYZs/controls.html
Your oscilloscope may have special horizontal magnification settings that let you display a magnified section of the
waveform on-screen.
XY Mode
Most oscilloscopes have the capability of displaying a second channel signal along the X-axis (instead of time). This is
called XY mode; you will find a longer discussion later in this book.
Trigger Controls
The trigger controls let you stabilize repeating waveforms and capture single-shot waveforms. Figure 6 shows a typical
front panel and on-screen menus for the trigger controls.
The trigger makes repeating waveforms appear static on the oscilloscope display. Imagine the jumble on the screen that
would result if each sweep started at a different place on the signal (see Figure 7).
For edge triggering, the trigger level and slope controls provide the basic trigger point definition.
The trigger circuit acts as a comparator. You select the slope and voltage level of one side of the comparator. When the
trigger signal matches your settings, the oscilloscope generates a trigger.
The slope control determines whether the trigger point is on the rising or the falling edge of a signal. A rising edge
is a positive slope and a falling edge is a negative slope.
The level control determines where on the edge the trigger point occurs.
Figure 8 shows you how the trigger slope and level settings determine how a waveform is displayed.
Trigger Sources
The oscilloscope does not necessarily have to trigger on the signal being measured. Several sources can trigger the
sweep:
Most of the time you can leave the oscilloscope set to trigger on the channel displayed.
Note that the oscilloscope can use an alternate trigger source whether displayed or not. So you have to be careful not to
unwittingly trigger on, for example, channel 1 while displaying channel 2.
Trigger Modes
The trigger mode determines whether or not the oscilloscope draws a waveform if it does not detect a trigger. Common
trigger modes include normal and auto .
In normal mode the oscilloscope only sweeps if the input signal reaches the set trigger point; otherwise (on an analog
oscilloscope) the screen is blank or (on a digital oscilloscope) frozen on the last acquired waveform. Normal mode can
be disorienting since you may not see the signal at first if the level control is not adjusted correctly.
Auto mode causes the oscilloscope to sweep, even without a trigger. If no signal is present, a timer in the oscilloscope
triggers the sweep. This ensures that the display will not disappear if the signal drops to small voltages. It is also the best
mode to use if you are looking at many signals and do not want to bother setting the trigger each time.
In practice, you will probably use both modes: normal mode because it is more versatile and auto mode because it
requires less adjustment.
Some oscilloscopes also include special modes for single sweeps, triggering on video signals, or automatically setting
the trigger level.
Trigger Coupling
Just as you can select either AC or DC coupling for the vertical system, you can choose the kind of coupling for the trigger
signal.
Besides AC and DC coupling, your oscilloscope may also have high frequency rejection, low frequency rejection, and
noise rejection trigger coupling. These special settings are useful for eliminating noise from the trigger signal to prevent
false triggering.
Trigger Holdoff
Sometimes getting an oscilloscope to trigger on the correct part of a signal requires great skill. Many oscilloscopes have
special features to make this task easier.
Trigger holdoff is an adjustable period of time during which the oscilloscope cannot trigger. This feature is useful when
you are triggering on complex waveform shapes, so that the oscilloscope only triggers on the first eligible trigger point.
Figure 9 shows how using trigger holdoff helps create a usable display.
Acquisition Modes
Acquisition modes control how waveform points are produced from sample points. Recall from the first section that
sample points are the digital values that come directly out of the Analog-to-Digital-Converter (ADC). The time between
sample points is called the sample interval. Waveform points are the digital values that are stored in memory and
displayed to form the waveform. The time value difference between waveform points is called the waveform interval. The
sample interval and the waveform interval may be but need not be the same. This fact leads to the existence of several
different acquisition modes in which one waveform point is made up from several sequentially acquired sample points.
Additionally, waveform points can be created from a composite of sample points taken from multiple acquisitions, which
leads to another set of acquisition modes. A description of the most commonly used acquisition modes follows.
Sample Mode: This is the simplest acquisition mode. The oscilloscope creates a waveform point by saving one
sample point during each waveform interval.
Peak Detect Mode: The oscilloscope saves the minimum and maximum value sample points taken during two
waveform intervals and uses these samples as the two corresponding waveform points. Digital oscilloscopes
with peak detect mode run the ADC at a fast sample rate, even at very slow time base settings (long waveform
interval), and are able to capture fast signal changes that would occur between the waveform points if in sample
mode. Peak detect mode is particularly useful for seeing narrow pulses spaced far apart in time.
Hi Res Mode: Like peak detect, hi res mode is a way of getting more information in cases when the ADC can
sample faster than the time base setting requires. In this case, multiple samples taken within one waveform
interval are averaged together to produce one waveform point. The result is a decrease in noise and an
improvement in resolution for low speed signals.
Envelope Mode: Envelope mode is similar to peak detect mode. However, in envelope mode, the minimum and
maximum waveform points from multiple acquisitions are combined to form a waveform that shows min/max
changes over time. Peak detect mode is usually used to acquire the records that are combined to form the
envelope waveform.
Average Mode: In average mode, the oscilloscope saves one sample point during each waveform interval as in
sample mode. However, waveform points from consecutive acquisitions are then averaged together to produce
the final displayed waveform. Average mode reduces noise without loss of bandwidth but requires a repeating
signal.
One of the greatest advantages of digital oscilloscopes is their ability to store waveforms for later viewing. To this end,
there are usually one or more buttons on the front panel that allow you to stop and start the acquisition system so you can
Wednesday, March 25, 1998 Page: 9
http://www.tek.com/Measurement/
App_Notes/XYZs/controls.html
analyze waveforms at your leisure. Additionally, you may want the oscilloscope to automatically stop acquiring after one
acquisition is complete or after one set of records has been turned into an envelope or average waveform. This feature is
commonly called single sweep or single sequence and its controls are usually found either with the other acquisition
controls or with the trigger controls.
Sampling Methods
In digital oscilloscopes that can use either real-time sampling or equivalent-time sampling as described earlier, the
acquisition controls will allow you to choose which one to use for acquiring signals. Note that this choice makes no
difference for slow time base settings and only has an effect when the ADC cannot sample fast enough to fill the record
with waveform points in one pass.
Other Controls
So far we have described the basic controls that a beginner needs to know about. Your oscilloscope may have other
controls for various functions. Some of these may include:
Measurement cursors
Keypads for mathematical operations or data entry
Print capabilities
Interfaces for connecting your oscilloscope to a computer
Look over the other options available to you and read your oscilloscope's manual to find out more about these other
controls.
Up To: Glossary
Up To: Index
Measurement Techniques
This section teaches you basic measurement techniques. The two most basic measurements you can make are voltage
and time measurements. Just about every other measurement is based on one of these two fundamental techniques.
This section discusses methods for taking measurements visually with the oscilloscope screen. Many digital
oscilloscopes have internal software that will take these measurements automatically. Knowing how to take the
measurements manually will help you understand and check the automatic measurements of the digital oscilloscopes.
The Display
Take a look at the oscilloscope display. Notice the grid markings on the screen - these markings create the graticule .
Each vertical and horizontal line constitutes a major division . The graticule is usually laid out in an 8-by-10 division
pattern. Labeling on the oscilloscope controls (such as volts/div and sec/div) always refers to major divisions. The tick
marks on the center horizontal and vertical graticule lines (see Figure 1) are called minor divisions.
Many oscilloscopes display on the screen how many volts each vertical division represents and how many seconds each
horizontal division represents. Many oscilloscopes also have 0%, 10%, 90%, and 100% markings on the graticule (see
Figure 1) to help make rise time measurements, described later.
Voltage Measurements
Voltage is the amount of electric potential, expressed in volts, between two points in a circuit. Usually one of these points
is ground (zero volts) but not always. Voltages can also be measured from peak-to-peak - from the maximum point of a
signal to its minimum point. You must be careful to specify which voltage you mean.
The oscilloscope is primarily a voltage-measuring device. Once you have measured the voltage, other quantities are just
a calculation away. For example, Ohm's law states that voltage between two points in a circuit equals the current times
the resistance. From any two of these quantities you can calculate the third. Another handy formula is the power law: the
power of a DC signal equals the voltage times the current. Calculations are more complicated for AC signals, but the
point here is that measuring the voltage is the first step towards calculating other quantities.
Wednesday, March 25, 1998 Page: 1
http://www.tek.com/Measurement/App_Notes/
XYZs/meastech.html
Figure 2 shows the voltage of one peak - V[p] - and the peak-to-peak voltage - V[p-p] -, which is usually twice V[p]. Use the
RMS (root-mean-square) voltage - V[RMS] - to calculate the power of an AC signal.
You take voltage measurements by counting the number of divisions a waveform spans on the oscilloscope's vertical
scale. Adjusting the signal to cover most of the screen vertically, then taking the measurement along the center vertical
graticule line having the smaller divisions, makes for the best voltage measurements. The more screen area you use, the
more accurately you can read from the screen.
Many oscilloscopes have on-screen cursors that let you take waveform measurements automatically on-screen, without
having to count graticule marks. Basically, cursors are two horizontal lines for voltage measurements and two vertical
lines for time measurements that you can move around the screen. A readout shows the voltage or time at their positions.
You take time measurements using the horizontal scale of the oscilloscope. Time measurements include measuring the
period, pulse width, and timing of pulses. Frequency is the reciprocal of the period, so once you know the period, the
frequency is one divided by the period. Like voltage measurements, time measurements are more accurate when you
adjust the portion of the signal to be measured to cover a large area of the screen. Taking time measurement along the
center horizontal graticule line, having smaller divisions, makes for the best time measurements. (See Figure 4.)
Standard pulse measurements are pulse width and pulse rise time . Rise time is the amount of time a pulse takes to go
from the low to high voltage. By convention, the rise time is measured from 10% to 90% of the full voltage of the pulse.
This eliminates any irregularities at the pulse's transition corners. This also explains why most oscilloscopes have 10%
and 90% markings on their screen. Pulse width is the amount of time the pulse takes to go from low to high and back to
low again. By convention, the pulse width is measured at 50% of full voltage. See Figure 5 for these measurement points.
Pulse measurements often require fine-tuning the triggering. To become an expert at capturing pulses, you should learn
how to use trigger holdoff and how to set the digital oscilloscope to capture pretrigger data, as described earlier in the
Controls section. Horizontal magnification is another useful feature for measuring pulses, since it allows you to see fine
details of a fast pulse.
horizontal axis. (On some digital oscilloscopes this is a display mode setting.) This mode of operation opens up a whole
new area of phase shift measurement techniques.
The phase of a wave is the amount of time that passes from the beginning of a cycle to the beginning of the next cycle,
measured in degrees. Phase shift describes the difference in timing between two otherwise identical periodic signals.
One method for measuring phase shift is to use XY mode. This involves inputting one signal into the vertical system as
usual and then another signal into the horizontal system. (This method only works if both signals are sine waves.) This
set up is called an XY measurement because both the X and Y axis are tracing voltages. The waveform resulting from this
arrangement is called a Lissajous pattern (named for French physicist Jules Antoine Lissajous and pronounced
LEE-sa-zhoo). From the shape of the Lissajous pattern, you can tell the phase difference between the two signals. You
can also tell their frequency ratio. Figure 6 shows Lissajous patterns for various frequency ratios and phase shifts.
What's Next?
This section has covered basic measurement techniques. Other measurement techniques involve setting up the
oscilloscope to test electrical components on an assembly line, subtracting noise from a signal, capturing elusive
transient signals, and many others that would take too much room to list. The measurement techniques you will use
depend on your application, but you have learned enough to get started. Practice using your oscilloscope and read more
about it. Soon its operation will be second nature to you.
Up To: Glossary
Up To: Index
Written Exercises
This section contains written exercises that cover information in this book. The exercises are divided into two parts, Part I
and Part II.
The Oscilloscope
Oscilloscope Terminology
Setting Up
The Controls
Measurement Techniques
Part I Exercises
The following exercises cover information presented in these sections:
The Oscilloscope
Oscilloscope Terminology
Check how well you have absorbed the information in these sections by doing this short self-test. You will probably want
to print a copy of this test to work on.
Vocabulary Exercise
Write the number of the definitions in the right column next to the correct words in the left column.
View answers.
View answers.
Part II Exercises
The following exercises cover information presented in these sections:
Setting Up
The Controls
Measurement Techniques
Check how well you have absorbed the information in these sections by doing this short self-test. Answers begin on
page.
Vocabulary Exercise
Write the letter of the definitions in the right column next to the correct words in the left column.
View answers .
View answers.
Glossary
AC
(Alternating Current) A signal in which the current and voltage vary in a repeating pattern over time.
ADC
(Analog-to-Digital Converter) A digital electronic component that converts an electrical signal into discrete binary
values.
Alternate Mode
A display mode of operation in which the oscilloscope completes tracing one channel before beginning to trace
another channel.
Amplitude
The magnitude of a quantity or strength of a signal. In electronics, amplitude usually refers to either voltage or
power.
Attenuation
A decrease in signal voltage during its transmission from one point to another.
Averaging
A processing technique used by digital oscilloscopes to eliminate noise in a signal.
Bandwidth
A frequency range.
CRT
(Cathode-Ray Tube) An electron-beam tube in which the beam can be focused on a luminescent screen and
varied in both position and intensity to produce a visible pattern. A television picture tube is a CRT.
Chop Mode
A display mode of operation in which small parts of each channel are traced so that more than one waveform can
appear on the screen simultaneously.
Circuit Loading
The unintentional interaction of the probe and oscilloscope with the circuit being tested, distorting the signal.
Compensation
A probe adjustment for 10X probes that balances the capacitance of the probe with the capacitance of the
oscilloscope.
Coupling
The method of connecting two circuits together. Circuits connected with a wire are directly coupled; circuits
connected through a capacitor or a transformer are indirectly (or AC) coupled.
Cursor
An on-screen marker that you can align with a waveform to take accurate measurements.
DC (Direct Current)
A signal with a constant voltage and current.
Division
Measurement markings on the CRT graticule of the oscilloscope.
Earth Ground
A conductor that will dissipate large electrical currents into the Earth.
Envelope
The outline of a signal's highest and lowest points acquired over many repetitions.
Equivalent-time Sampling
A sampling mode in which the oscilloscope constructs a picture of a repetitive signal by capturing a little bit of
information from each repetition.
Focus
The oscilloscope control that adjusts the CRT electron beams to control the sharpness of the display.
Frequency
The number of times a signal repeats in one second, measured in Hertz (cycles per second). The frequency
equals 1/period.
Gigahertz (GHz)
1,000,000,000 Hertz; a unit of frequency.
Wednesday, March 25, 1998 Page: 1
http://www.tek.com/Measurement/App_Notes/
XYZs/
Glitch
An intermittent error in a circuit.
Graticule
The grid lines on a screen for measuring oscilloscope traces.
Ground
1. A conducting connection by which an electric circuit or equipment is connected to the earth to establish and
maintain a reference voltage level.
2. The voltage reference point in a circuit.
Hertz (Hz)
One cycle per second; the unit of frequency.
Kilohertz (kHz)
1000 Hertz; a unit of frequency.
Interpolation
A "connect-the-dots" processing technique to estimate what a fast waveform looks like based on only a few
sampled points.
Megahertz (MHz)
1,000,000 Hertz; a unit of frequency.
Megasamples per second (MS/s)
A sample rate unit equal to one million samples per second.
Microsecond
A unit of time equivalent to 0.000001 seconds.
Millisecond (ms)
A unit of time equivalent to 0.001 seconds.
Nanosecond (ns)
A unit of time equivalent to 0.000000001 seconds.
Noise
An unwanted voltage or current in an electrical circuit.
Oscilloscope
An instrument used to make voltage changes visible over time. The word oscilloscope comes from "oscillate,"
since oscilloscopes are often used to measure oscillating voltages.
Peak - V[p]
The maximum voltage level measured from a zero reference point.
Peak-to-peak - V[p-p]
The voltage measured from the maximum point of a signal to its minimum point, usually twice the V[p] level.
Peak Detection
An acquisition mode for digital oscilloscopes that lets you see the extremes of a signal.
Period
The amount of time it takes a wave to complete one cycle. The period equals 1/frequency.
Phase
The amount of time that passes from the beginning of a cycle to the beginning of the next cycle, measured in
degrees.
Probe
An oscilloscope input device, usually having a pointed metal tip for making electrical contact with a circuit element
and a flexible cable for transmitting the signal to the oscilloscope.
Pulse
A common waveform shape that has a fast rising edge, a width, and a fast falling edge.
RMS
Root mean square.
Real-time Sampling
A sampling mode in which the oscilloscope collects as many samples as it can as the signal occurs.
Record Length
The number of waveform points used to create a record of a signal.
Rise Time
The time taken for the leading edge of a pulse to rise from its minimum to its maximum values (typically measured
from 10% to 90% of these values).
Sample Point
The raw data from an ADC used to calculate waveform points.
Screen
The surface of the CRT upon which the visible pattern is produced - the display area.
Signal Generator
Wednesday, March 25, 1998 Page: 2
http://www.tek.com/Measurement/App_Notes/
XYZs/
A test device for injecting a signal into a circuit input; the circuit's output is then read by an oscilloscope.
Sine Wave
A common curved wave shape that is mathematically defined.
Single Shot
A signal measured by an oscilloscope that only occurs once (also called a transient event).
Single Sweep
A trigger mode for displaying one screenful of a signal and then stopping.
Slope
On a graph or an oscilloscope screen, the ratio of a vertical distance to a horizontal distance. A positive slope
increases from left to right, while a negative slope decreases from left to right.
Square Wave
A common wave shape consisting of repeating square pulses.
Sweep
One horizontal pass of an oscilloscope's electron beam from left to right across the CRT screen.
Sweep Speed
Same as the time base.
Time Base
Oscilloscope circuitry that controls the timing of the sweep. The time base is set by the seconds/division control.
Trace
The visible shapes drawn on a CRT by the movement of the electron beam.
Transducer
A device that converts a specific physical quantity such as sound, pressure, strain, or light intensity into an
electrical signal.
Transient
A signal measured by an oscilloscope that only occurs once (also called a single-shot event).
Trigger
The circuit that initiates a horizontal sweep on an oscilloscope and determines the beginning point of the
waveform.
Trigger Holdoff
A control that inhibits the trigger circuit from looking for a trigger level for some specified time after the end of the
waveform.
Trigger Level
The voltage level that a trigger source signal must reach before the trigger circuit initiates a sweep.
Volt
The unit of electric potential difference.
Voltage
The difference in electric potential, expressed in volts, between two points.
Waveform
A graphic representation of a voltage varying over time.
Waveform Point
A digital value that represents the voltage of a signal at a specific point in time. Waveform points are calculated
from sample points and stored in memory.
Z-axis
The signal in an oscilloscope that controls electron-beam brightness as the trace is formed.
Up To: Index