Astm F 2394-07
Astm F 2394-07
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
     Copyright by ASTM Int'l (all rights reserved); Mon Aug 15 06:42:44 EDT 2011
                                                                                 1
     Downloaded/printed by
     SAADET IMIK (ALVIMEDICA+MEDICAL+TECHNOLOGIES) pursuant to License Agreement. No further reproductions authorized.
                                                                   F2394 – 07
   3.1.2 crimp, v—to secure the stent on the delivery system by             3.1.15 non-recoverable movement, n—a displacement of the
radially compressing and plastically deforming the stent onto            stent relative to the balloon such that if the shearing force was
the balloon.                                                             reduced to zero, the stent would remain displaced in the
   3.1.3 delivery system, n—a system similar to a balloon                direction of the shearing force relative to the initial placement
dilatation catheter that is used to deliver and deploy a stent at        on the balloon. The force at which non-recoverable movement
the target site and then removed.                                        begins is defined as the initial displacement force (see defini-
   3.1.4 displacement force, critical distance peak, n—a stent           tion above).
securement test endpoint characterizing the maximum force                   3.1.16 pre-test treatment, n—a treatment of the stent deliv-
required to displace the stent with respect to the balloon a             ery system prior to the evaluation of securement that simulates
critical distance. This critical distance is the minimum of the          preparatory, environmental, mechanical or other conditions
following two distances. The first is the distance at which the          that may be encountered prior to or during clinical use of the
undamaged stent could overhang the balloon body resulting in             device. Examples include subjecting the devices to elevated
a clinically significant, incomplete end deployment. The sec-            shipping temperature/humidity, catheter preparation per use
ond is the length (distance) of stent compression or buckling            instructions, pre-soaking, bending treatments, tracking treat-
that could result in a clinically significant incomplete deploy-         ments (tracking fixture, see definition below) and tracking
ment of the stent against the vessel walls. (See Fig. X2.1.)             through lesion treatments (lesion fixture, see definition below).
   3.1.5 displacement force, initial, n—a stent securement test             3.1.17 pre-test treatment tracking fixture, n—a pre-test
endpoint characterizing the initial force required to displace the       treatment fixture used to simulate an anatomical vasculature.
stent with respect to the balloon such that the displacement is          Use of the fixture with a guide catheter, a guide-wire and the
a non-recoverable movement (see 3.1.15). (See Fig. X2.1.)                stent-balloon catheter delivery system is intended to simulate
   3.1.6 displacement force, initial peak, n—a stent secure-             the bending and frictional forces of tracking the device to the
ment test endpoint characterizing the first peak in force that           lesion site that may be encountered in the clinical setting. See
occurs during or after stent displacement with respect to the            the engineering diagrams in the Appendix. Note that these
balloon. (See Fig. X2.1.)                                                engineering diagrams simulate vessels with a moderately
   3.1.7 dislodgment force, peak, n—a stent securement test              difficult degree of coronary tortuousity but do not include
endpoint characterizing the peak or maximum force required to            simulated lesions.
completely dislodge the stent from the delivery system balloon.             3.1.18 pre-test treatment lesion fixture, n—a pre-test treat-
During a test, this force will occur after or coincide with the          ment fixture used to simulate an anatomical vasculature and
initial displacement force. (See Fig. X2.1.)                             lesion. Use of the fixture with a guide catheter, a guide-wire,
   3.1.8 end flaring, n—a distal or proximal outward conical             and the stent-balloon catheter delivery system is intended to
opening of the diameter of the stent on the balloon. End flaring         simulate the bending, frictional and mechanical resistance
is a contributing factor to the probability that the stent may           forces of tracking the device across the lesion site that may be
become caught during withdrawal into a guide catheter while              encountered in the clinical setting.
tracking through a lesion.                                                  3.1.19 securement test, guide-type, n—a stent securement
   3.1.9 failure mode effect analysis (FMEA), n—an analytical            test that is similar to the clinical scenario of pulling an
approach to methodically determine and address all possible              undeployed stent delivery system back into a guide catheter,
product failure modes, their associated causes, and their                arterial sheath or hemostasis valve. Examples include guides,
criticality. Used to evaluate designs, prioritize testing, and           rings, or shims ideally designed to engage the stent end or body
track risk reducing improvements to the product.                         but not the catheter balloon. The shim securement test, de-
   3.1.10 gauge length, n—the initial unstressed length of               scribed in Section 7, uses complementary thin, rigid plates with
catheter tubing between the proximal end of the stent to the             rounded “V” notches that are sized to circumferentially engage
grips which engage the catheter tubing.                                  the stent end but not the catheter balloon. See the engineering
   3.1.11 grips, n—a means of applying force to the stent and            diagrams in the Appendix.
balloon catheter to displace or dislodge the stent relative to the          3.1.20 securement test, lesion-type, n—a stent securement
balloon. In particular, grips refer to the end of a device which         test that is similar to the clinical scenario of pushing or pulling
makes the contact with the stent. Typical grips used to apply            an undeployed stent delivery system through or around a
force to the stent include shims (as used in Figs. X2.5-X2.8);           fibrous or calcified lesion. Examples include tape, nubs,
tape which sticks to the stent but not the balloon; an iris which        protrusions or sandpaper ideally designed to engage the stent
can be narrowed down to allow the balloon to slip by but not             end or body but not the catheter balloon.
the stent; or nubs which contact the stent but not the balloon.
   3.1.12 guide catheter, n—a tube designed to transport the             4. Significance and Use
guide-wire and the stent delivery system into the target vessel.            4.1 The securement of the endovascular stent on the balloon
   3.1.13 guide-wire, n—a wire designed to aid in balloon,               is a critical parameter to ensure that the stent is safely delivered
ultrasound, atherectomy, or stent placement during endovascu-            to or from the treatment site.
lar procedures.                                                             4.2 This guide is intended for use by researchers and
   3.1.14 mandrel, n—a wire that may be used as an alternative           manufacturers for the development and selection of pre-test
to the intended guide-wire to provide support for the catheter           treatments, tests and test endpoints to measure stent secure-
guide-wire lumen for some test procedures.                               ment (displacement distances and dislodgment forces).
    Copyright by ASTM Int'l (all rights reserved); Mon Aug 15 06:42:44 EDT 2011
                                                                                2
    Downloaded/printed by
    SAADET IMIK (ALVIMEDICA+MEDICAL+TECHNOLOGIES) pursuant to License Agreement. No further reproductions authorized.
                                                                   F2394 – 07
  4.3 This guide may be used to investigate which practical                 6.2.1.2 Physician surveys for clinical relevance and prob-
combinations of in vitro tests best characterize clinical sce-           lems with comparable devices.
narios.                                                                     6.2.1.3 Mechanical understanding of the tests’ clinical rel-
  4.4 This guide should be used with discretion in choosing              evance and limitations.
securement tests and evaluating results due to the myriad                   6.2.1.4 Mechanical and statistical understanding of the test
possible combinations of clinical conditions, failure modes,             reproducibility limitations due to device variation, pre-test
and stent delivery system designs.                                       treatments, various grips, and test conditions.
  4.5 This guide may be of use for developing a test for                    6.2.1.5 Ability to set accept/reject criteria by physician
meeting parts 2 and 3 of the requirements of EN 14299,                   evaluation, by historical comparisons, or by other rational
Section 7.3.4.4 on Trackability.                                         means.
  4.6 This guide may be of use for developing a test to meet                6.2.2 The final securement test(s) selected must ultimately
section VII-C-8 of CDRH Guidance document.                               satisfy internal manufacturer quality standards. These stan-
                                                                         dards may include clinical relevance, FMEA analysis, statisti-
5. Clinical Scenarios                                                    cal assurance of characteristics, and challenge assurance of
   5.1 There are two failure modes—the stent is dislodged                characteristics.
from the catheter or the stent is displaced or deformed on the              6.2.3 The final securement test(s) must also satisfy external
catheter such that balloon inflation delivery would not produce          regulatory body standards. For example, the FDA QSR 21 CFR
an acceptable stent shape at the proper location. Based on               Part 820, Oct. 7, 1996 states that each test used in the process
reported clinical incidents, there are three causes for these two        of design and manufacturing of finished devices “is suitable for
types of failures:                                                       its intended purposes and is capable of producing valid
   5.1.1 Displacement or dislodgment of the stent while at-              results.” For the statistical capability evaluation, Guide E1488
tempting to track through or position in tortuous bends, fibrous         is very helpful.
or calcified lesions, or previously implanted stents, or combi-             6.3 Pre-Test Treatments:
nation thereof.                                                             6.3.1 Pre-test treatments may be conducted prior to the
   5.1.2 Displacement or dislodgment of the stent on with-               evaluation of securement to simulate preparatory, environmen-
drawal of the undeployed stent delivery system back into the             tal, mechanical, or other conditions that may be encountered
guide catheter, introducer sheath, or hemostasis valve. This             prior to or during clinical use of the device.
failure type is usually associated with failure to cross tortuous           6.3.2 Pre-test treatments may include subjecting the devices
bends, fibrous or calcified lesions, or previously implanted             to shelf life testing, sterilization, elevated shipping
stents, or combination thereof. It is sometimes associated with          temperature/humidity, removal of the delivery system from the
less-than-ideal seating or angled placement of the guide                 carrier tube, and other catheter preparation per use instructions,
catheter tip in the ostium of the vessel.                                pre-soaking, bending treatments, tracking treatments, and
   5.1.3 Displacement or dislodgement of the stent due to                tracking through lesion treatments.
improper catheter preparation including mishandling or partial              6.3.3 Tracking treatments are intended to clinically simulate
balloon inflation during preparation. This has been identified in        the bending and frictional forces of tracking the device through
a few cases where the loose, displaced, or dislodged stent was           the guide catheter and vasculature to the lesion site. Consider-
observed prior to use but may conceivably play a role in a               ations for tracking treatments include: tracking medium (for
small percentage of cases where dislodgment occurs in pa-                example, air, water, water with lubricants, saline, blood) and
tients.                                                                  temperature; guide catheter and guide-wire selection; simu-
                                                                         lated vessel material (for example PMMA, silicone, glass,
6. Test Method Considerations                                            PTFE), tortuousity, dimensionality, length and diameter; speed
   6.1 Flowchart—See Fig. 1.                                             of tracking; and number of repetitions to track. Angiograms or
   6.2 Development and Evaluation of Securement Tests:                   autopsies of human or similar animal vasculature may be
   6.2.1 Securement test development and selection is ideally            particularly useful in developing alternative anatomical mod-
begun through the initial use of a battery of tests measuring a          els. An example of a tracking treatment is given in Section 7.
variety of failure modes. These test methods may vary from a             Two examples of tracking fixtures are given in engineering
simple intuitive tactile impression of the securement forces             diagrams in the Appendix. Note that these engineering dia-
through manipulation to clinically modeled situations with               grams simulate vessels in 2-D with a moderately difficult
guide catheters and stenosis models to in vivo animal studies            degree of coronary tortuousity but do not include simulated
with representative anatomy and physician handling. From a               lesions.
safety-risk perspective, consider how securement challenges                 6.3.4 Tracking through lesion treatments are intended to
may occur in clinical situations, what may result from loss of           clinically simulate the bending, frictional and mechanical
securement, what the severity of the outcome is to the patient,          resistance forces of tracking the device across the lesion site.
what the frequency of these situations are, and then how to test         Considerations include those of tracking treatments in addition
to detect these occurrences. Factors to consider in evaluating           to simulating the lesion material, type, and morphology. Lesion
securement tests include the following:                                  material encountered clinically may include calcium, fibrin,
   6.2.1.1 Review of the MAUDE database for reported prob-               collagen, fat, cholesterol, endothelial cells, smooth muscle
lems with comparable devices.                                            cells, red blood cells, platelets, dead white blood cells, and
    Copyright by ASTM Int'l (all rights reserved); Mon Aug 15 06:42:44 EDT 2011
                                                                                3
    Downloaded/printed by
    SAADET IMIK (ALVIMEDICA+MEDICAL+TECHNOLOGIES) pursuant to License Agreement. No further reproductions authorized.
                                                                   F2394 – 07
FIG. 1 Flowchart
macrophages. Lesion types may be calcified, fibrous, lipidic,            adjunct to force measuring tests. In particular, the measurement
thrombotic, or grumous. Lesion morphologies may be totally               of end flaring may be useful. While end flaring may increase
or partially occluded, concentric or eccentric, focal or diffuse,        the chance the stent will catch on a guide, previously placed
and of many possible shapes.                                             stents, lesions, and so forth, end flaring does not, in itself
   6.3.5 Post-Track Test Evaluation—To understand the effects            determine the force it takes to displace, dislodge or deform the
of tracking on the probability that the stent delivery system            stent. End flaring and other geometrical changes also provide a
may encounter securement challenges, the measurement of                  measure of the effects of different track and lesion geometries
stent-balloon geometry changes post tracking may be a useful             and materials. End flaring may be measured by measuring the
    Copyright by ASTM Int'l (all rights reserved); Mon Aug 15 06:42:44 EDT 2011
                                                                                4
    Downloaded/printed by
    SAADET IMIK (ALVIMEDICA+MEDICAL+TECHNOLOGIES) pursuant to License Agreement. No further reproductions authorized.
                                                                       F2394 – 07
change in peak proximal and distal end heights relative to the              consistency and simplicity by executing the remaining testing
longitudinal continuation of the nominal stent diameter.                    with the optimal selection.
   6.4 Securement Test and Grip Methods:                                       6.5.3 The gauge length chosen may influence the pull rate
   6.4.1 There are two main categories of stent securement                  transmitted to the stent. A shorter gauge length will have less
tests: guide-types and lesion-types.                                        longitudinal material deformation and transmit a less variable
   6.4.2 A guide-type securement test simulates the clinical                pull rate closer to the set pull rate. A 25 mm gauge length is a
scenario of pulling an undeployed stent delivery system back                typical value.
into a guide catheter, arterial sheath or hemostasis valve.                    6.5.4 The pull rate selected should allow meaningful dis-
Examples of grip methods used to simulate guide-type secure-                crimination and measurement of the chosen test endpoint(s)
ment tests include guide catheters, rings, or shims ideally                 and, as possible, model the clinical situation. Slower rates may
designed to engage the stent end or body but not the catheter               help visualizing pull-off mechanics in video taped tests. Faster
balloon. The shim securement test, described in Section 7, uses             rates may be more repeatable. In a test modeling lesion
complementary thin, rigid plates with rounded “V” notches to                dislodgment, set the crosshead extension speed (pull rate) to a
circumferentially engage the stent end but not the catheter                 slow rate such as at 0.5 in./min. In a test modeling withdrawal
balloon. Engineering diagrams are included in the Appendix.                 into the catheter, set the crosshead extension speed to a faster
   6.4.3 A lesion-type securement test simulates the clinical               rate such as 10 to 30 in./min. Adjust the data sampling rate as
scenario of pushing or pulling an undeployed stent delivery                 appropriate; for example, increase the sampling rate for high
system through or around a fibrous or calcified lesion. Ex-                 pull speeds; have a high sampling rate for measuring peak
amples of grip methods used to simulate lesion-type secure-                 forces.
ment tests include tape, nubs, protrusions, or sandpaper ideally               6.5.5 The direction of pull (proximally, distally, or angled)
designed to engage the stent end or body but not the catheter               chosen may depend on factors such as stent and delivery
balloon.                                                                    system design and clinical relevance. The product design may
   6.4.4 For both types of tests, it may increase the understand-           be more at risk for either proximal or distal displacement.
ing of the test to measure a baseline reference force by testing            Clinically, peak pull forces transmitted to the stent will exceed
delivery systems without or with poorly secured stents. Such                push forces. Also, the stent forces encountered clinically are
tests may provide measurements of slip forces, balloon and                  seldom oriented purely proximally or distally tangent to the
catheter deformation mechanics, and the lower limit to mea-                 undeformed stent axis. Pulling in a non-tangential direction
surable securement forces.                                                  adds in normal forces which tend to increase the displacement
   6.5 Test Conditions:                                                     and dislodgement forces. Thus, typically, the most challenging
                                                                            securement test pulls purely in the tangential direction of the
   6.5.1 Stent securement test conditions particularly impor-
                                                                            undeformed stent. This is best confirmed through systematic
tant for hygroscopic balloon materials, whose stent securement
                                                                            variation of the pull direction.
properties may change with varying exposure to water or other
fluids, include the following:                                                 6.6 Test Endpoints:
                                                                               6.6.1 There are two main test endpoints for stent securement
   6.5.1.1 Tracking and test temperature,
                                                                            evaluations: displacement force and dislodgment force.
   6.5.1.2 Tracking and test medium (for example, air, water,
                                                                               6.6.2 The displacement force is the force required to dis-
saline, blood),
                                                                            place the stent with respect to the balloon in a non-recoverable
   6.5.1.3 Tracking and test soak time (for fluid mediums), and             movement. Clinically, displacement may result in the follow-
   6.5.1.4 Time between track and test.                                     ing:
  NOTE 1—Justification for the selection of the parameters should be part      6.6.2.1 Improper positioning upon deployment due to the
of the test history file.                                                   movement of the stent with respect to the markerbands,
                                                                               6.6.2.2 Incomplete stent end deployment due to movement
   6.5.2 Guide-wire selection (or substituted test mandrel) may
                                                                            of the stent off the body of the balloon, and
particularly influence pre-test track treatments and securement
                                                                               6.6.2.3 Incomplete or poor apposition of the stent to the
tests that include simulated tracking. A stiffer guide-wire or test
                                                                            vessel walls due to compression or buckling damage to the
mandrel may place the stent delivery system under greater
                                                                            stent.
normal and frictional forces. Or, it may span and round out
curvature in a track model, especially flexible models, and                   NOTE 2—There are at least four possible endpoints to measure displace-
reduce the bending of the balloon and stent. Additionally,                  ment force: initial, critical distance peak, initial peak, and maximum
guide-wire stiffness varies along the length (for example,                  sustained force. See Fig. X2.1 for graphical representation of the first three
flexible 50-mm tip, moderate 50-mm stent/balloon support, and               of these listed endpoints. The maximum sustained force (not seen in Fig.
                                                                            X2.1) may be relevant for some stent, balloon, and test combinations.
stiff 500-mm shaft); guide-wire materials may have strong
non-linear response and recovery from bending, and each                        6.6.2.4 The initial displacement force is the initial force
different guide-wire surface, typically chosen for lubricity and            required to displace the stent with respect to the balloon. This
durability, may respond to different catheter lumen surfaces.               is often difficult to distinguish from movement of the stent
Completing ruggedness evaluations using a variety of wires                  within but not out of its balloon imprint or movement of the
may be important to determine the optimal wire choice for the               intact stent and balloon with respect to, for example, the
test. However, after choosing the optimal wire within the                   guide-wire lumen and markerbands within the balloon. There
guidelines of the Instructions for Use for your stent, maintain             may be operator variability in the assessment of the force at
    Copyright by ASTM Int'l (all rights reserved); Mon Aug 15 06:42:44 EDT 2011
                                                                                5
    Downloaded/printed by
    SAADET IMIK (ALVIMEDICA+MEDICAL+TECHNOLOGIES) pursuant to License Agreement. No further reproductions authorized.
                                                                        F2394 – 07
which initial displacement occurs. This endpoint is a conser-                   7.1.2 Securement-Testing Machine—A power-driven ma-
vative estimation of the force at which clinically significant               chine capable of the following:
negative outcomes may occur.                                                    7.1.2.1 Uniform crosshead speed (pull rate) within 5 % of
  6.6.2.5 The initial peak displacement force is characterized               set rate,
as the first peak in force that occurs during or after stent                    7.1.2.2 Measurement and recording of force exerted to a
displacement with respect to the balloon. A peak displacement                precision of 0.01 lbf, and
force may often occur simultaneously with stent displacement                    7.1.2.3 Measurement and recording of crosshead displace-
and would therefore be equal to the initial displacement force.              ment to a precision of 0.01 in. or time measured to a precision
However, for some stent delivery systems, the stent may move                 of 0.1 s.
a critical distance before a peak in force is reached. Therefore
                                                                                7.1.3 Crosshead Catheter Clamp—Wedge clamps or other
the initial peak displacement force is a conservative estimation
                                                                             clamps attached to the machine crosshead capable of holding
of the dislodgment force.
                                                                             the catheter beyond the securement endpoint without slipping.
  6.6.2.6 The critical distance peak displacement force is the
                                                                                7.1.4 Fixed Shim Stent Grips—An example of the fixed
maximum force required to displace the stent with respect to
                                                                             shim stent grips and associated tooling is given in engineering
the balloon a critical distance. This critical distance should be
                                                                             diagrams in the Appendix, Figs. X2.5-X2.8.
the minimum of the following two distances.
    (1) The distance at which the undamaged stent could                         7.1.5 Water Bath—Vessels used to immerse the tracking
overhang the balloon body resulting in a clinically significant,             fixture and the securement test stent grips in 37 6 2°C water.
incomplete deployment of the overhanging end of the stent,                      7.1.6 Camera and Video Recorder—Optional devices used
and                                                                          during the securement testing to help determine the stent
    (2) The length (distance) of stent compression or buckling               securement test endpoint.
that could result in a clinically significant, incomplete deploy-               7.2 Test Specimens:
ment of the deformed stent against the vessel walls.                            7.2.1 To establish a sufficiently powered test, first determine
                                                                             the standard deviation. This can be established by running a
   NOTE 3—The critical distance displacement force may often be equal to     trial to estimate the variation or by leveraging existing data.
the initial displacement force. There may be operator variability in the
assessment of the critical distance within which to determine the maxi-
                                                                             Further, calculate the required sample size with this standard
mum force. Due to this operator dependence, there may be significant         deviation, the desired minimum detectable difference, mini-
variance between operators.                                                  mum significance level and minimum power level.
   NOTE 4—The clinical significance of incomplete deployment may                7.2.2 Unless otherwise justified, all samples selected for
depend on several factors, particularly time and the location at which the   testing should be taken from finished, sterilized, clinical-
incomplete deployment occurs; for example, dislodgement during inser-        quality product. Cosmetic rejects or other non-clinical samples
tion into a coronary lesion is more critical than dislodgement on
                                                                             may be used for these tests if the cause for rejection is not
withdrawal through a hemostasis valve.
                                                                             related to securement.
   6.6.3 The peak dislodgment force is the maximum force                        7.2.3 Prepare the device in accordance with the Instructions
required to completely dislodge the stent from the delivery                  for Use.
system balloon. Clinically, if the stent is dislodged distally, this            7.3 Pre-Test Treatment Procedure:
endpoint represents the stent embolizing in the vasculature. If
                                                                                7.3.1 Set the tracking fixture, with the guide catheter and
the stent is dislodged proximally, the balloon can no longer be
                                                                             guide-wire inserted, in the 37°C water bath. Guide catheter and
used to deploy the stent. The majority of reported securement
                                                                             guide-wire selection should be consistent with the devices used
problems are associated with stent dislodgement. Though peak
                                                                             during medical procedures for the device tested.
dislodgment force is operator independent, it still may have
significant variability, particularly for stent delivery systems                7.3.2 Advance the delivery system through the guide cath-
with catch-points such as unfurled balloon cones or other                    eter and the tracking fixture. Fully track the system into the
raised profile components abutting the distal or proximal end,               fixture in 2 to 8 seconds. Withdraw it at approximately the
or both, of the stent. Since dislodgement occurs after displace-             same rate. For a worst case in loosening the stent on the
ment, the peak force is of secondary importance to the                       balloon, track the device three or more times.
dislodgement distance prior to stent deployment and of primary                  7.3.3 Remove the system from the tracking fixture and store
importance if the stent/stent delivery system is being removed.              on a catheter rack at room temperature.
Dislodgement force is a vital characteristic but is not a                       7.4 Securement Test Procedure:
complete measure of securement.                                                 7.4.1 Set up the shim stent grips within the water bath at a
                                                                             temperature of 37 6 2°C.
7. Examples of Apparatus and Procedure                                          7.4.2 If used, set up the camera and video recorder to allow
  7.1 Apparatus Example:                                                     clear display of the stent/shim interface.
  7.1.1 Pre-Test Treatment Tracking Fixture—An example of                       7.4.3 Set the crosshead extension speed (pull rate) at 2
the fixture is given in the engineering diagrams in the Appen-               in./min.
dix, Figs. X2.1 and X2.2. Note that this model simulates a                      7.4.4 Insert a suitable guide-wire or test mandrel into the
coronary vessel in a 2-D plane with a moderately difficult                   catheter lumen so that it will extend at least from the catheter
degree of tortuousity but no simulated lesions.                              clamp point to the tip of the catheter.
    Copyright by ASTM Int'l (all rights reserved); Mon Aug 15 06:42:44 EDT 2011
                                                                                6
    Downloaded/printed by
    SAADET IMIK (ALVIMEDICA+MEDICAL+TECHNOLOGIES) pursuant to License Agreement. No further reproductions authorized.
                                                                     F2394 – 07
   7.4.5 Insert catheter through the open “V” of the shims and               8.4.1.4 Performance compared to historical data or against
close the shims over the folded balloon at the proximal end of            other stent delivery system designs, models, or sizes.
the stent.                                                                   8.4.2 Test scenario:
   7.4.6 Check shim size. Shims should be sized to allow the                 8.4.2.1 Simulated use conditions.
delivery system balloon to move freely while the stent remains               8.4.2.2 Comparative bench testing which emphasizes re-
blocked from moving through the shims. Remove and replace                 peatability and the ability to discriminate products and de-
shims as necessary to obtain proper fit.                                  emphasizes actual use conditions.
   7.4.7 Secure crosshead catheter clamp at 1.0 6 0.1 in.                    8.4.2.3 Challenge testing which combines 2 or more low
proximal to the proximal edge of the stent (1 in. gauge length).          probability events/situations to create an unlikely condition in
   7.4.8 Start video recorder, if used.                                   which it is likely the performance of the device will decrease.
   7.4.9 Start machine crosshead extension.
                                                                             8.4.3 Characteristics to be evaluated:
   7.4.10 Monitor test to observe chosen test endpoint.
                                                                             8.4.3.1 Stent securement performance.
   7.4.11 Record the securement endpoint(s), force and mode
(for example, stent slippage or buckling).                                   8.4.3.2 Changes in stent geometry with tracking and/or
   7.4.12 If the stent pulls through the shim ID, the peak force          securement testing.
is a false representation of stent securement and should be                  8.4.3.3 Specific stent securement or safety mechanisms.
recorded as invalid data. The stent delivery system should not               8.4.3.4 Conclusions to be made based on the results of the
be retested.                                                              testing, especially accept/reject criteria.
                                                                             8.5 References/Supporting Documents:
8. Test Report                                                               8.5.1 Test protocol (what is to be done), test method (how),
                                                                          test method development reports (how you know the test
   NOTE 5—The following is a content recommendation for stent secure-
ment test reports. The documentation should be clear enough so that the   method is suitable and capable for intended purpose), and data
study could be repeated.                                                  sheets.
                                                                             8.5.2 As required, reference equipment calibration and
   8.1 Test Report Identification:
                                                                          maintenance logs, and operator training records.
   8.1.1 Report title; author, contributors and their affiliations;
                                                                             8.6 Test Specimens/Materials/Traceability:
name of organization presenting the results; date of publica-
tion; any additional document identifiers, such as a document                8.6.1 Identification:
number, as required by the organization or reviewing bodies.                 8.6.1.1 Stent: length, diameter, coatings or treatments.
   8.1.2 Product(s) tested.                                                  8.6.1.2 Stent delivery system—If a balloon catheter, indica-
   8.2 Executive Summary:                                                 tion of balloon material characteristics (for example, compliant
   8.2.1 Summarize the nature and scope of the testing.                   versus non-compliant); monorail, over-the-wire, combination
   8.2.2 Summarize conclusions made as a result of the testing,           device, or other; and the stent delivery system coatings or
especially claims pertaining to safety.                                   treatments.
   8.3 Objective:                                                            8.6.1.3 Accessory devices used, for example, guide-wires.
   8.3.1 Identify the types of specimens being tested:                       8.6.1.4 Traceability—Provide sufficient information so that
   8.3.1.1 Labeled stent length and diameter.                             processing records can be found for the test specimens—as full
   8.3.1.2 Monorail, Over-the-Wire, or other type of delivery             assemblies, partial assemblies, the assembly processes, and
system.                                                                   sterilization/conditioning records, as appropriate.
   8.3.1.3 Devices with special characteristics or histories.                8.6.2 Processing and conditioning of the stent, the delivery
   8.3.1.4 Actual or representative finished goods or special             system, and the finished device.
builds.                                                                      8.6.2.1 Pilot line or production line manufacturing.
   8.3.2 Identify the type(s) of testing being performed:                    8.6.2.2 Nominal or limit processing conditions, repeated
   8.3.2.1 Product Specification Testing—Determine if the                 cycles, or challenge conditions.
product performance meets the requirements of the product                    8.6.2.3 Actual or simulated handling, shipping, or storage
specification.                                                            conditions.
   8.3.2.2 Manufacturing Control Testing—Determine if the                    8.6.2.4 Time between critical processes, if applicable.
performance of the product is within normal tolerance and if
                                                                             8.6.2.5 Special processes not typical to the design.
there is no significant trend towards being out of control.
   8.3.2.3 Other Types of Testing—Predicate device compari-                  8.6.2.6 Rationale for choosing such test specimens if not
son testing; design alternative testing, observational testing,           obvious.
test method development.                                                     8.7 Sample Size:
   8.4 Scope:                                                                8.7.1 Sample size(s) and the sample size rationale based on
   8.4.1 Identify what the test will demonstrate:                         significance level, confidence level, and power level as appro-
   8.4.1.1 Stent securement as a stand-alone characteristic;              priate.
   8.4.1.2 The ability to retract the stent delivery system intact           8.7.2 Sampling technique—such as per a specified random
after stent securement is challenged;                                     process, first and last product in each sequential lot, special
   8.4.1.3 The ability to deploy the stent after stent securement         built.
is challenged; and/or                                                        8.8 Experimental Method:
    Copyright by ASTM Int'l (all rights reserved); Mon Aug 15 06:42:44 EDT 2011
                                                                                7
    Downloaded/printed by
    SAADET IMIK (ALVIMEDICA+MEDICAL+TECHNOLOGIES) pursuant to License Agreement. No further reproductions authorized.
                                                                   F2394 – 07
   8.8.1 Test method overview with emphasis on specific                    8.10 Conclusions:
choices made within the referenced test method or deviations               8.10.1 As appropriate, was the acceptance criteria met?
to the referenced test method.                                             8.10.2 Do the results support claims made about the device
   8.8.2 Data collected during the test, noting which data are           with regard to performance?
objective and which are observer-dependent.                                8.10.3 Was anything notable or unexpected observed?
   8.8.3 Retrospective information detailing the process for               8.10.4 Recommendations for future tests or future testing,
processing, analyzing, and presenting data, and if there were            as appropriate.
any deviations from the test method.                                       8.11 Appendices:
   8.9 Results:                                                            8.11.1 Test protocol, test method, individual data, and/or
   8.9.1 Data table(s) including max, min, means, standard               calculations, as appropriate.
deviations, and number of samples, as appropriate.
   8.9.2 Graphs—plots of actual measured data and, as appro-             9. Keywords
priate, means and standard deviations.                                     9.1 angioplasty; artery; dislodgment; displacement; medical
   8.9.3 Discussion of results.                                          device; retention; securement; stent; vascular
APPENDIXES
(Nonmandatory Information)
X1. RATIONALE
  X1.1 Beginning with an understanding of the clinical                   relevant as well as repeatable and reliable. The guide asks the
conditions which may lead to stent dislodgement or stent                 user to consider the many product life-history pre-test treat-
deformation, this guide describes a development process and              ments of the test specimen that may affect the results. The
many considerations in stent securement testing. The guide               guide also discusses some details of the test conditions and the
shows a method to develop one or more tests that are clinically          detail of how to interpret numerical results from the test.
    Copyright by ASTM Int'l (all rights reserved); Mon Aug 15 06:42:44 EDT 2011
                                                                                8
    Downloaded/printed by
    SAADET IMIK (ALVIMEDICA+MEDICAL+TECHNOLOGIES) pursuant to License Agreement. No further reproductions authorized.
                                                                  F2394 – 07
FIG. X2.2 Bottom Plate of the Test Tracking Fixture for Conditioning the Stent, Balloon, and Stent Securement System
   Copyright by ASTM Int'l (all rights reserved); Mon Aug 15 06:42:44 EDT 2011
                                                                               9
   Downloaded/printed by
   SAADET IMIK (ALVIMEDICA+MEDICAL+TECHNOLOGIES) pursuant to License Agreement. No further reproductions authorized.
                                                               F2394 – 07
FIG. X2.3 Top Plate of the Test Tracking Fixture for Conditioning the Stent, Balloon, and Stent Securement System
Copyright by ASTM Int'l (all rights reserved); Mon Aug 15 06:42:44 EDT 2011
                                                                            10
Downloaded/printed by
SAADET IMIK (ALVIMEDICA+MEDICAL+TECHNOLOGIES) pursuant to License Agreement. No further reproductions authorized.
                                                               F2394 – 07
Copyright by ASTM Int'l (all rights reserved); Mon Aug 15 06:42:44 EDT 2011
                                                                            11
Downloaded/printed by
SAADET IMIK (ALVIMEDICA+MEDICAL+TECHNOLOGIES) pursuant to License Agreement. No further reproductions authorized.
                                                               F2394 – 07
Copyright by ASTM Int'l (all rights reserved); Mon Aug 15 06:42:44 EDT 2011
                                                                            12
Downloaded/printed by
SAADET IMIK (ALVIMEDICA+MEDICAL+TECHNOLOGIES) pursuant to License Agreement. No further reproductions authorized.
                                                                               F2394 – 07
   X3.1 Sampling—Since the securement test irreversibly                               nificant variations in test results occur with different operators,
changes the test samples such that the same samples cannot be                         different time points, a range of devices, and/or with other
used more than once, the variation in the individual results                          sources of random variation. The range of devices may include
cannot distinguish between uncontrolled variation in the test                         carefully selected devices to minimize part-part variation,
samples and uncontrolled variations associated with the equip-                        devices or model devices such as rings on rods. Along with
ment, operators, or environment. However, with sufficient                             this, if more than one machine is used, or if fixtures change,
numbers of random samples from the same population and a                              measure the machine variability and the effects on repeatability
good test design, the distributions of results can be used to                         and reproducibility.
distinguish the effects of controllable factors. Thus, prior to
testing, plan your test to efficiently and confidently resolve                           X3.5 Capability—Determine the operating range of the test.
differences in results due to controllable factors. In particular,                    For example, determine a range of diameters, lengths, materi-
determine the number of samples required for the level of                             als, designs, or other factors which are suitable for evaluation
confidence and power desired. The power is the probability                            by this test. A secondary outcome of this determination may be
that a specific mean difference will be detected. For example,                        the characterization of an entire family of stent systems and the
if the test objective is to determine whether a normally                              identification of a worst-case for further analysis.
distributed mean is greater than a specific benchmark, the
sample size may be calculated as a function of four inputs: (1)                          X3.6 Bias—Since it is unlikely there will be an accepted
the estimated standard deviation, (2) the desired minimum                             reference value of securement, it is recommended that exter-
detectable difference between the mean and the benchmark, (3)                         nally circulated test reports provide comparisons between
the desired significance level, and (4) the desired power. A                          existing products (predicate devices) and the device of interest.
significance level of 0.05 to 0.10 and a power level of 80 to                         However, new modes of failure may occur with new design
90 % are frequently used.                                                             features, for example, change from bare to coated stents, so old
                                                                                      test methods, mean values and distributions may all appropri-
  X3.2 Suitability—If the test is to be clinically predictive,                        ately change making comparison with predicate devices unre-
demonstrate that the securement test models simulate either                           liable and not useful.
access through a lesion or a withdrawal into a guide catheter.
Alternatively, the purpose of the test may be to assess manu-                            X3.7 Validation—The validation activities for a test method
facturing repeatability.                                                              depend on the criticality and the complexity of the test method.
                                                                                      That is, the more complex the test, the more work will be
   X3.3 Ruggedness—Using Guide E1169, determine the sen-
                                                                                      required to understand the method, understand the sources of
sitivity of the test results to deliberate small variations in the
                                                                                      variation (test or specimen), and choose appropriate test
test conditions and environmental factors. If any effects are
                                                                                      parameters. The more critical the need for the test (that is, is a
statistically significant relative to inherent measurement vari-
                                                                                      measured test characteristic merely annoying or potentially
ability then modify the test method to reduce these effects, that
                                                                                      life-threatening compounded by the frequency of occurrence),
is, make the test more robust.
                                                                                      the more thorough the demonstration that the test method is
  X3.4 Repeatability and Reproducibility—Determine if sig-                            relevant and predictive.
                   ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
               in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
               of infringement of such rights, are entirely their own responsibility.
                   This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
               if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
               and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
               responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
               make your views known to the ASTM Committee on Standards, at the address shown below.
                 This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
               United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
               address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
               (www.astm.org).
    Copyright by ASTM Int'l (all rights reserved); Mon Aug 15 06:42:44 EDT 2011
                                                                                13
    Downloaded/printed by
    SAADET IMIK (ALVIMEDICA+MEDICAL+TECHNOLOGIES) pursuant to License Agreement. No further reproductions authorized.