Testforgasesbyorsatand Chromatographicmethods-Methods: Indian Standard
Testforgasesbyorsatand Chromatographicmethods-Methods: Indian Standard
-
Gurgaon(compliance@farelabs.com) 182.66.121.30.(Reaffirmed 2015)
IS 13270 : 1992
(Reaffirmed 2014)
(Reaffirmed 2019)
(Reaffirmed 2013)
(Reaffirmed 2010)
UDC 543’27 : 543’54
(Reaffirmed 2009)
(Reaffirmed 2008)
(Reaffirmed 2007)
(Reaffirmed 2006)
(Reaffirmed 2005)
@ BIS 1992
FOREWORD
This lcdian Standard wzs sdopted by the Bureau of Indian Standards, after the draft finalized by
the Environmental Protection Sectional Committee had been approved by the Chemical Division
Council.
Orsat analysis and chromatographic gas analysis are ccmmonly used. Each one has some
advantages and disadvantages. These ar’e listed below:
Orsat Analysis Chromatographic Analysis
Advantages
Gasometric ( volumetric procedures ) Gas chromatographic analysis
1 The equipment required is relatively simple. 1 It has a great advantage of speed.
2 It does not require any calibration. 2 A gas analysis can be completed in few
minutes.
3 When the analysis is done on an infrequent 3 It can be used for low range.
basis, it is very useful.
4 Simple to operate. 4 The method is suitable for continuous
analysis, as the instrument needs calibra-
tion before use.
Disadvantages
1 Errors may be due to collection storage and 1 Instrument must have been previously
handling of samples. calibrated for each gas of interest.
2 Unless special care is taken in the collec- 2 The oven must have reached a constant
tion of samples contamination by air temperature and the detector must be
occurs. giving stable response.
3 Mercury is an ideal confining liquid/fluid 3 It is very difficult to carry instrument to
because of the solubility of all gases in it. the site. If the sample is collected in gas
Rut practically it cannot be used due to holder or any other equipment, collection,
great density and cost. Hence saturated storage or handling becomes a problem.
salt/water is used for ordinary purposes.
4 It cannot measure concentrations of gases 4 It requires inert gas cylinder.
below 0 2 percent.
In reporting the result of a test or analysis made in accordance with this standard, if the final
value, cbserved or calculated, is to be rounded off, it shall be done in accordance with IS 2 : 1960
‘Rules for rounding off numerical values ( revised )‘.
Free Standard provided by BIS via BSB Edge Private Limited to FARE LABS PRIVATE LTD. -
Gurgaon(compliance@farelabs.com) 182.66.121.30.
IS 13270 : 1992
Indian Standard
TESTFORGASESBYORSATAND
CHROMATOGRAPHICMETHODS-
METHODS
1 SCOPE 4.3 Interferences
IS132:0:1992
4.4.3.6
A distributing tip pipette containing 4.5.6 Sodium Hydrate Asbestos Absorbent
alkaline pyrogallol solution ( Duplicate
of 4.4.3.3 ). Used when an unusually accurate measure of
carbon dioxide is required. This absorbent may
also be used to remove sulphur dioxide for
4.4.3.7 These pipettes must possess smooth a more accurate determination of unsaturated
surface which will not entrap gas bubbles, acd hydrocarbons.
shall be so sufficient as to absorb the following
gases from the sample after the required number 4.5.7 Saturated Salt Solution - 75 percent.
of contacts with reagents.
Contains 30 g of sodium chloride or sodium
Gas Number of Contacts sulphate or both, 5 ml of hydrochloric acid,
with Reagents 2 drops of methyl red per 100 ml of distilled
Oxygen 4 to 5 water.
Carbon dioxide 3
4.6 Procedure
Unsaturated hydrocarbons 3
4.6.1 Analysis with Portable Apparatus for
4.4.4 Manometer Carbon Dioxide Oxygen and Carbon Monoxide
The apparatus shall be of reproducibility of The portable orsat apparatus is fitted with a
measurement of 0’02 ml per single contact or metal or wooden carrying case and uses a
0’05 ml on three successive contacts using the shortened form of burette with three gas absorb-
same reference gas or air. ing pipettes. In order, starting from the burette,
the pipettes are filled with potassium hydroxide,
4.5 Reagents pyrogallol and cuprous chloride solution res-
pectively. After filling the above pipettes to the
4.5.1Quality of Reagents engraved mark with the above solutions and
before starting the test adjust the level of each
Unless specified, otherwise, pure chemicals and to atmospheric pressure using the levelling bulb.
distilled water ( see IS 1070 : 1977 ) shall be Open the stopcock of the burette to the atmos-
employed in tests. phere. Raise the levelling bulb until the burette
fills to the stopcock with salt water ( saturated).
NOTE - ‘Pure chemicals’ shall mean chemicals Connect the stopcock -of the burette to the
that do not contain impurities shich affect the atmosphere to be sampled or to a sample con-
results of analysis. tainer and fill the burette with sample gas by
lowering the levelling bulb until the meniscus
4.5.2 Potassium Hydroxide Solution Saturated of the water level reads the desired volume in
the burette ( I’,). Open the stopcock connecting
In 200 ml of distilled water, dissolve solid the burette to manifold of the absorbing system
potassium hydroxide until excess potassium and also open the stopcock of the potassium
hydrcxide remains. Cool the saturated solution -hydroxide pipette. Pass the gas contained in
to at least 3°C below lowest expected tempera- the burette into the potassium hydroxid pipette
ture at which analysis will be carried out. by first raising and then lowering the levelling
Decant and store the supernate liquid. bottle. Repeat until three to five full contacts
have been made. Return the remainder of the
45.3 Activated Sulphuric Acid gas sample to the burette using the levelling
bulb-until the level of potassium hydroxide
Concentrated sulphuric acid containing silver solution returns to the engraved mark, and with
sulphate or vanadium pentoxide. the pipette stopcock closed, again, adjust the
water level in the burette to atmospheric pres-
4.5.4 Alkaline Pyrogallol Solution sure using the levelling bulb. Measure the
volume V, of the remaining gas and record the
Dissolve 17 g of pyrogallol crystals in 100 ml of percent carbon dioxide as follows:
potassium hydroxide solution ( 4.5.2 ). Store
under rcfregeration in a glass-stoppered bottle. Carbon dioxide, percentage = loo ( ‘\- Va )
1
4.5.5 Acidic Copper (II) Chloride Solution
Similarly, oxygen is removed from the remaining
Dissolve 450 g of copper (II) chloride in 2 500 ml gas volume V, by passing this gas into the
of hydrochloric acid ( relative density 1’18 ). If pyrogallol solution in the second pipette.
this solution appears greenish or black in colour Measurement of the remaining volume, V, is
after preparation strips or turnings of copper used to calculate the percent oxygen as follows:
shall be added to the solution until a straw-
yellow coloured liquid is produced on standing. 100 ( V, - Vs )
Oxygen, percent =
Store solution over copper turnings or wire. V1
2
Free Standard provided by BIS via BSB Edge Private Limited to FARE LABS PRIVATE LTD. -
Gurgaon(compliance@farelabs.com) 182.66.121.30.
IS 13270: 1992
Carbon monoxide is measured by mnnipulating 0’5 perc:nt they are best determined using
the remaining gas volume V3 as done previously reaction tubes rather than -pipettes.
to admit this sample into the pipette containing
copper (II) chloride. However, before returning 4.6.4 Combustion Analysis
the gas volume V, to the burette for measure- Prepare nitrogen to be used as a transfer gas by
ment, volume V4 % passed once into the potas- absorption of oxygen from uncontaminated air
sium hydroxide pipette to remove any hydro- by contact with alkaline pyrogallol pipette. Flush
chloric acid vapours evolved from copper (II) the manifold with this nitrogen, then transfer
chloride. approximately 40 ml of the pure nitrogen to the
Carbon monoxide, = 100 ( V, - V4 ) duplicate potassium hydroxide pipette ( 4.4.3.5 )
percent Vl
for storage V,. Transfer approximstely 95 ml
of pure cylinder oxygen V, to the burette;
4.6.2 Analysis by Constant Pressure Volumetry measure and transfer to the slow combustion
pipette for storage. Lst the inert impurities
In the laboratory bench apparatus the burette is shown by the published analysis of this cylinder
filled with mercury and enclosed in a water oxygen be represented as VT. Measure a fresh
jacket. The pipettes are connected to the 30 to 35 ml of sample gas V, through the fuming
burette by a manifold and a bubbling pipette sulphuric acid pipette prior to combustion
containing sulphuric acid ( 4.4.3.2 ) is used analysis.
together with a slow combustion pipette (4.4.3.4)
which is equipped with a separate levelling bulb. With the combustion gas sample contained in
Pipettes (4.4.3.5 and 4.4.3.6 ) are added to the the burette adjust the pressure in the combustion
of both absorption and pipette and the burette to atmospheric and with
system. Results
the platinum wire glowing dull red op:n the
combustion analyses are reported.
combustion pipette and slowly admit the gas
4.6.3 Removal of Gases by Absorption Analysis sample over the hot platinum wire. Allow a
full 15 minutes for the first pass of sample into
Transfer 95 to 100 ml of the sample gas $0 the the bomb combustion tubs. When all of the
burette allowing 2 to 3 minutes for attaining gas sample has been transferred to the combus-
temperature and humidity equilibrium. Using tion pipette, displace the gas contained in the
the leveling bulb bring the sample volume to manometer arm through the distributer into the
atmospheric pressure and read the exact combustion pipette. Over a period of about
volume VI. 5 minutes, return the contents of the combustion
pipette to the burette until the mercury level is
4.6.3.1 Removal of carbon dioxide ( or acid gases ) just below the platinum spiral, then return the
gas slowly to the combustion pipette and repeat
Displace the gas sample into the manifold and the slow combustion three times. Allow the
then transfer into the potassium hydroxide final pass of sample gas in the combustion
pipette. Return the sample gas to the burette. pipette to cool before returning to the burette.
Then contact the potassium hydroxide pipette Measure this residue and record as V,.
twice, finally returning the sample to the burette
and allowing 2 to 3 minutes before equilibrating 4.6.4.1 Removal of carbon dioxide produced by
the sample to atmospheric pressure with the combustion
leveling bulb. Then read the volume V,.
Displace the gas from the manometer and the
4.6.3.2 Removal of amsaturated hydrocarbon contents sample residue in the burette into the
duplicate potassium hydroxide pipette ( 4.4.3.5 )
Displace the gas sample from the manometer three times. Return this sample volume to the
arm and pass twice in and out of the pipette combustion pipette and then repeat contact with
containing activated sulphuric acid. Transfer the potassium hydroxide solution before return-
the sample to the potassium hydroxide pipette ing to the burette for measurement as VIO.
return to the sulphuric acid pipette for two
successive contacts. Finally return to the burette 4.6.4.2 Removal of excess oxygen after
for measurement F’s of the gas after standing combustion
3 minutes in the burette.
Dissolve the sample gas from the manometer and
contact 4 times in the duplicate alkaline pyro-
4.6.3.3 Re.moval of oxygen
gall01 pipette ( 4.4.3.6 ). Then contact once the
Displace the gas sample from the manometer duplicate potassium hydroxide pipette (4.4.3.5 )
and transfer twice to the pipette containing and pass once through the slow combustion
alkaline pyrogallol, then transfer to the potas- pipette before returning -to the alkaline pyro-
sium hgdroxide pipette and the sulphuric acid gall01 pipette. Transfer this residue to the
pipette in sequence. Finally transfer twice to burette and measure VI1.
the alkaline pyrogallol pipette and return to the 4.7 Calibration and Standard
burrette for measurement of the residual
volume ( V4). When acid gases, oxygen and Calibration shall be performed using commer-
.unsaturated hydrocarbons occur at levels below cially purchased oxygen, nitrogen, carbon
3
Free Standard provided by BIS via BSB Edge Private Limited to FARE LABS PRIVATE LTD. -
Gurgaon(compliance@farelabs.com) 182.66.121.30.
IS13270:1992
4
Free Standard provided by BIS via BSB Edge Private Limited to FARE LABS PRIVATE LTD. -
Gurgaon(compliance@farelabs.com) 182.66.121.30.
IS 13270 : 1992
IS 13270 : 1992
_ _
CONDUCTIVITY CELL
SAMPLING
COLUMN 7 COLUMN 1
6
Free Standard provided by BIS via BSB Edge Private Limited to FARE LABS PRIVATE LTD. -
Gurgaon(compliance@farelabs.com) 182.66.121.30.
IS 13270 : 1992
adjusted to the values specified for the instru- air diluted with pure nitrogen both samples and
ments by its manufacturer. About thirty minutes standards contain argon and no correction is
are required for instrument stabilization. The necessary.
recorder is turned eon and zeroed before samples
are introduced. 5.6.1 The standard curves should be checked
periodically.
5.5.2 Injection of Sample
5.6.2 A severe loss in resolution of the carbon
5.5.2.1 Sampling valve
dioxide composite peaks and/or of the nitrogen/
oxygen peaks indicates the need for replacement
The sample loop is flushed with several volumes of columns. The polar gas chromatography
of calibration standard or sample gas. The column continuously looses stationary liquid
handle is then turned to divert the sample to the phase through volatilization. These vapours are
chromatograph. adsorbed on the molecular sieve column along
with the carbon dioxide, which leads to slow
5.5.2.2 Syringe deterioration of the performance of that column.
Normally this will happen slowly over a long
Sample is withdrawn from the sample vessel and period of time.
quickly injected, guarding against blow-back
of the plunger. 5.7 Calculations
5.5.3 Repetitive Analysis Concentrations are determined directly from the
calibration plots. The following conversion
A new sample may be analyzed immedintely factors apply at 76 mm Hg and 25°C.
after the last peak if the sample has emerged.
Samples should be analyzed in duplicate. Gas ( mglm3 Yupm
Carbondioxide 1’80
5.6 Calibration and Standards Oxygen 1’31
A standard curve of peak height or peak area Nitrogen 1’14
vs volume percent is prepared for each constitu- Carbon monoxide 0’,654
ent of interest by analyzing the calibration Methane 1’14
standards. The calibration should bracket the
sample concentrations. Linear plots should 5.8 Precision and Accuracy
result. However, in the presence of 5 to 7 per-
cent carbon dioxide, the calibration for oxygen Accuracy depends upon the availability of
is not linear up to 20 percent, but the calibration accurate calibration standards. These may be
plot may still be used. If the sample source is obtained with a certificate of analysis from com-
natural air, the result for oxygen may need mercial suppliers. Precision is controlled by the
correction for argon present in the sample but mode of sample introduction, primarily, gas of
not separated from the oxygen. If the oxygen & 0.3 percent. Reproducibility with a 1 ml
calibration standard mixtures contain pure gas-tight syringe is about & 1’5 percent. Pre-
oxygen dilute with pure nitrogen, the apparent cision is also affected by detector drift, which in
volume percent of conductivity detector molar turn depends upon the control of carrier gas,
response factors for oxygen are sufficiently close flow rate and system, temperature. Standard
that no appreciable error will result from assum- commercial gas chromatographic equipment
ing identical relative responses. If the oxygen capable of detector temperature control to rt 0’5
calibration standard mixtures contain natural and flow rate control to f 1 percent is adequate.
7
Free Standard provided by BIS via BSB Edge Private Limited to FARE LABS PRIVATE LTD. -
Gurgaon(compliance@farelabs.com) 182.66.121.30.
I
Standard Mark
The use of the Standard Mark is governed by the provisions of the Bureau of Indian
Standards Act, 1986 and the Rules and Regulations made thereunder. The Standard Mark on
products covered by an Indian Standard conveys the assurance that they have been produced
to comply with the requirements of that standard under a well defined system of inspection,
testing and quality control which is devised and supervised by BIS and operated by the pro-
ducer. Standard marked products are also continuously checked by BIS for conformity to
that standard as a further safeguard. Details of conditions under which a licence for the use
of the Standard Mark may be granted to manufacturers or producers may be obtained from
the Bureau of Indian Standards.
Free Standard provided by BIS via BSB Edge Private Limited to FARE LABS PRIVATE LTD. -
Gurgaon(compliance@farelabs.com) 182.66.121.30.
BIS is a statutory institution established under the B-creau ofIndian Standards Act, 1936 to promote
harmonious development of the activities of standsrdizstion, marking and quality certification of
goods and attending to connected matters in the country.
Copyright
BIS has the copyright of all its publications. No psrt of these publications mzy be reproduced in
any form without the prior PErmission in writing of BIS. This do:s not preclude the free use, in
the course of implem:nting the standsrd, of necessary details, such as symbols and sizes, type or
grade designations. Enquiries relating to copyright b: addressed to the Director ( Publications >, BIS.
Indian Standards are reviewed periodically and revised, when necessary and amendments, if any,
are issued from time to time. Users of Indian Standards should ascertain that they are in
possession of the latest amendments or edition. Comments on this Indian Standard may be sent
to BIS giving the following reference :