HS5x2 Oscilloscopes
User's Guide
HScope HS5x2 Oscilloscopes User's Guide v.250130
Notices
© Martinloren®. 2005-2025
No part of this manual may be Technology License Safety Notices
reproduced in any form or by
any means (including electronic The hardware and/or software CAUTION
storage and retrieval or described in this document are
translation into a foreign furnished under a license and
A CAUTION notice denotes a
language) without prior may be used or copied only in
hazard. It calls attention to an
agreement and written consent accordance with the terms of
operating
from Martinloren®, as governed such license.
procedure, practice, or the like
by Chinese and international
that, if not correctly performed
copyright laws.
or adhered to, could result in
damage to the product or loss of
Edition important data. Do not proceed
First edition, October 2024 beyond a CAUTION notice until
Current edition, January 2025 the indicated conditions are fully
understood and met.
Warranty
WARNING
The material contained in this
document is provided "as is,"
and is subject to being A WARNING notice denotes a
changed, without notice, in hazard. It calls attention to an
future editions. Further, to the operating procedure, practice,
maximum extent permitted or the like that, if not correctly
by applicable law, performed or adhered to,
Martinloren® disclaims all could result in personal injury
warranties, either express or or death. Do not proceed
implied, with regard to this beyond a WARNING notice
manual and any information until the indicated conditions
contained herein, including are fully understood and met.
but not limited to the implied
warranties of merchantability
and fitness for a particular
purpose.
Martinloren® shall not be
liable for errors or for
incidental or consequential
damages in connection with
the furnishing, use, or
performance of this document
or of any information
contained herein.
Should Martinloren® and the
user have a separate written
agreement with warranty
terms covering the material in
this document that conflict
with these terms, the warranty
terms in the separate
agreement shall control.
HScope HS5x2 Oscilloscopes User's Guide 2
HScope HS5x2 Oscilloscopes - At a Glance
HS512 MAX
HS502
HS 502 (2024 Edition) HS 512 Max
Low level signals Wider input range
Indicated for
(standard vehicles, sensors, mV measures) (standard and heavy duty vehicles)
No Input Channels 2
Min Input Range ± 500mV ±100mV
Max Input Range ±16V ±80V
230V (for 30 seconds)
Max Supported Voltage
50Vp Protected against high
(protection limit)
input voltages spikes
Input Ranges 5 steps 12 steps
Input Coupling AC / DC
Input Impedance 1 Mohm (1.010 kOhm) 1 Mohm (1.030 kOhm)
8 - 12 (depending on rate)
ADC bits 300KSa/s - 4.8MSa/s (12 bit)
7 - 13MSa/s (8 bit)
Sampling Rate 1kSa/s - 13MSa/s
1KSa – 300KSa (2 channels)
Real-time Rates
1KSa – 600KSa (1 channel)
1.6 MHz (2Vpp signal @5MSa/s)
BW
3 MHz (2Vpp signal @13MSa/s)
32k samples @ 12-16bit (half when enable both channels)
Memory Buffer
64k samples @ 8bit (7-13MSa/s, half when enable both channels)
Hardware, real-time, Hardware, real-time,
Trigger
position within first 500 samples variable position
HScope HS5x2 Oscilloscopes User's Guide 3
The HScope HS5x2 oscilloscopes deliver these features:
Operate with almost all Android devices (check requirements).
High refresh rate.
All operation with Android touch screen and gesture.
Many built-in measurements.
Waveforms filtering function (low-pass, high-pass, band-pass)
Advanced Math: dynamic frequency, duty-cycle.
Long continuous recording with Automotive Module, up to 600KSa/s with single channel.
Long recording in paging mode for top rates.
Serial decode with optional Digital Module for: UART/RS232, CAN, LIN.
Low USB power consumption for long use.
Easy data saving and sharing. Open file format.
For more information about HScope oscilloscopes, see:
www.Martinloren.com/hscope
HScope HS5x2 Oscilloscopes User's Guide 4
1 Getting Started
This chapter describes the steps you take when using the oscilloscope for the first time.
Inspect the Package Contents
Power-On the Oscilloscope
Connect Probes to the Oscilloscope
Input a Waveform
Use Autoscale & Basic Gesture
Compensate Passive Probes / Attenuators
HScope HS5x2 Oscilloscopes User's Guide 5
Inspect the Package Contents
Inspect the shipping container for damage.
If your shipping container appears to be damaged, keep the shipping container or cushioning material until
you have inspected the contents of the shipment for completeness and have checked the oscilloscope
mechanically and electrically.
Verify that you received the following items and any optional accessories you may have ordered:
HScope HS5x2 oscilloscope.
USB OTG Cable (it may include micro-usb adapter).
Oscilloscope probes (one for each analog input channel) are not included in the basic offer.
Power-On the Oscilloscope
Android Device OS version: Android 7+, best with Android 9+
Requirements RAM: 512KB, best with 4GB+
Flash: at least 1GB free (for waveform saving).
App Installation You may install HScope app from Google Play Store or Huawei App Store. In the
app store just search for “HScope”.
To power-on the 1 Connect the USB cable to the rear of the oscilloscope, then to your Android
oscilloscope device. Route the power cord so that there is no pressure on the USB
connectors.
2 Check that the oscilloscope red LED is on.
Always use a grounded power cord. Do not defeat the power cord ground.
HScope HS5x2 Oscilloscopes User's Guide 6
Connect Probes to the Oscilloscope
1 Connect the oscilloscope probe to an oscilloscope channel BNC connector.
2 Connect the probe's retractable hook tip to the point of interest on the circuit or device under test. Be sure
to connect the probe ground lead to a ground point on the circuit.
CAUTION Maximum input voltage at analog inputs
50Vpk for HS502
230Vrms for 30s for HS512 MAX
CAUTION Do not float the probes ground
Defeating the ground connection and "floating" the probes ground will
probably result in inaccurate measurements and may also cause equipment
damage. If you need to measure between two live points, use a differential
probe with sufficient dynamic range.
WARNING Do not negate the protective action of the ground connection to the
oscilloscope. The oscilloscope must remain grounded through the probe
ground. Defeating the ground may creates an electric shock hazard
through the oscilloscope case or the Android device in user’s hands.
HScope HS5x2 Oscilloscopes User's Guide 7
Input a Waveform
The Probe Comp signal is used for compensating probes.
1 Connect an oscilloscope probe from channel 1 to the Probe Ref terminal on the back panel.
WARNING A voltage source should never be connected to the ground terminal of this
instrument. If, for any reason, the Protective Conductor Terminal is
disconnected or not functioning properly and a voltage source is
connected to the equipment's ground terminals, the entire chassis and
Android device will beat the voltage potential of the voltage source, and
the operator or bystanders could receive an electric shock.
HScope HS5x2 Oscilloscopes User's Guide 8
Use Autoscale & Basic Gesture
Autoscale automatically configure the screen to best display the input signals vertically.
1 Double tap on the center of the screen.
You should see a waveform on the oscilloscope's display similar to this:
X2
2 Pitch and zoom with 2 fingers to change vertical zoom or horizontal zoom (time scale) or both.
HScope HS5x2 Oscilloscopes User's Guide 9
3 Use one finger to scroll or move the view up and down. Or move the finger on the bottom bar to scroll
just left and right.
If you see the waveform, but the square wave is not shaped correctly as shown above, perform the procedure
under "Compensate Passive Probes".
If you do not see the waveform, make sure the probe is connected securely to the front panel channel input
BNC and to the Probe Ref terminal.
HScope HS5x2 Oscilloscopes User's Guide 10
Compensate Passive Probes / Attenuators
Each oscilloscope passive probe must be compensated to match the input characteristics of the oscilloscope
channel to which it is connected. A poorly compensated probe can introduce significant measurement errors.
NOTE If your probe has a configurable attenuation setting (like the P6100 probes do),
the 10:1 setting must be used for probe compensation.
1 Enable the channel to which the probe is connected (CH-1 or CH-2).
2 Input the Probe Ref signal.
3 Perform the autoscale
4 If necessary, use a nonmetallic tool (supplied with the probe) to adjust the trimmer capacitor on the probe
/ attenuator for the flattest pulse possible.
Perfectly compensated
Over compensated
Under compensated
5 Repeat the procedure for each probe.
HScope HS5x2 Oscilloscopes User's Guide 11