Intelligent Vacuum Pump
User Manual
V700/V900/V1200
Contents
Disclaimer ···························································································1
Product Features ···············································································2
Technical Parameters ·········································································2
Service Parts ······················································································3
Operation Guide ···············································································4
Care and Maintenance ·······································································9
App Operating Instructions ······························································11
Disclaimer
• The vacuum pump is heavy. Be careful while handling to prevent personal injury.
• Use only ISO 46 mechanical oil to avoid vacuum pump damage.
• Do not cover the exhaust port during vacuum pump operation.
• Do not expose the suction port continuously to atmosphere for more than 3                 minutes.
• The air inlet pressure should not exceed 27. 5 bar to avoid sensor damage.
Warning Signs
    Sign           Name                                          Content
                Check the Oil    Add ISO 46 mechanical oil before the first use and keep the oil level
                   Level         between the maximum and minimum level
                                 Wear goggles when working with refrigerants . Direct contact with
               Wear Goggles
                                 refrigerants may cause injury
               Beware of High
                                 The pump surface becomes hot during normal operation. Do not
                Temperature
                                 touch the pump body or motor during operation
                   Burns
                                 Improper use may cause electrical shock hazards
              Avoid Electrical   Read and follow the instructions carefully and take precautions to
              Shock Hazards      avoid electrical shock hazards . Confirm that all associated devices
                                 are grounded correctly before power on
                                 Please remove the exhaust cap before starting up
              Danger
                                 Check the oil level and prohibit running without oil
                                                -1-
Product Features
•Intelligent Control of Motor and Solenoid Valve
•Leakage Level Judgment
•Vacuum Change in Graph
•4" Touch Screen Display
•Estimated Job Completion Time
• Data Logging and Storage via App
Technical Parameters
                 Models        V700                       V900                   V1200
       Stage                                            2 stages
   Power Supply                               110V/60Hz; 220V/50Hz
       Motor                                    AC induction motor
 Speed Regulation                                     Fixed speed
                                  1-10000microns           ±10% of Reading / ±10microns
 Vacuum Accuracy
                                  10000-19000 microns      ±20% of Reading
Operating Temperature                          41°F~104°F (5°C~40°C)
   Transmission
                                                       30ft / 10m
      Distance
   Limit Vacuum                                         15 microns
   Pumping Rate           7 CFM    (3L/S)            9 CFM    (4L/S)        12 CFM   (5L/S)
   Motor Power            3/4 HP ( 550 W )            1 HP ( 750 W )         1 HP ( 750 W )
   Fuel Volume            22oz /(650ml)              22 oz /( 650ml)        25.3 oz( 750ml)
 Overall Dimension            18.5in×10.5in×14.8in     / 4    70mm*267mm*380mm
      Weight              34.2lbs / 14.8kg           35.3lbs / 15.3kg        36.8lbs / 16kg
     Air Inlet                                      1/4 SAE ;3/8 SAE
                                              -2-
Service Parts
                            4   3
                                                      2
                    5
                                                           1
           6
                                                                    9
                                                      11
               10
                                                               12
                                                               13
       ① Display screen                   ⑧ Motor
       ② Handle                           ⑨ Motor housing
       ③ Air inlet                        ⑩ Power button
       ④ Solenoid valve                   ⑪ Vacuum sensor
       ⑤ Catcher                          ⑫ Oil window
       ⑥ Oil tank                         ⑬ Oil drain plug
       ⑦ Base
                                    -3-
Operation Guide
Conditions that must be met
1. Place on a flat surface.
2. Confirm the voltage and frequency at the outlet match with the vacuum pump
  specifications.
3. Confirm the oil level is within the Min and Max level.
4. Remove the air inlet cap (1/4 SAE fitting as shown in the
                                                                          1/4 SAE
  right image ), connect the pumped system (make sure the pipe
  fitting is fit with the air inlet fitting). Tighten the air inlet and
  make sure the system and hoses are sealed with no leakage.
5. Plug in the power cord and open the exhaust port.
Interface Introduction
                                           -4-
(1) The status bar on top of the page shows the product model, time, Bluetooth status.
① Model numbers varies depending on the product models.
②        Bluetooth not connect;        Bluetooth connected.
③ During operation,        appear when temperature exceeds the upper alarm limit,
    and       appear when temperature below the lower alarm limit.
④ Record: close means no data is recorded. Record Open: means that data is
     being recorded.
(2) The middle part of the page is the sub-interface button.
①       : Real-time measurement interface
    Measure
②       : Graph
     Graph
③       : Vacuum parameter setting
    Settings
④       : System parameter display and setting
     System
(3) The motor is off by default after power-on. Click "   " to start the motor. Click "   "
    to stop the motor.
When the motor starts, the solenoid valve automatically opens instantly. The solenoid
valve closes automatically after the motor stops to avoid the oil be suck to vacuum sensor.
1. Real time measurement interface
Click the“Measure” button on the main interface to enter the real-time measurement
interface.
                                            -5-
(1) Motor Status: show the current status of the motor.
(2) Motor Temp/Limit: show the surface temperature of the pump chassis and set
    the alarm temperature.
(3) Oil Temp/Limit: show vacuum pump oil temperature and set the alarm temperature.
(4) Vacuum Status: display the current vacuum value status, "Decrease", "Stable",
    "Rise", "--".
(5) Estimated End Time: when the vacuum is stable, it shows the estimated job
    completion time including holding time.
(6) Vacuum Reading Display: Vacuum readings shows when the vacuum of the
     system in real-time is below 8000Pa. The waiting time for the vacuum reading
     is longer for a larger system.
Note: The vacuum reading varies due to the sensor placement different. The reading
variances is caused by the uneven air flow during the vacuuming process. The farther
the air from the pump, the higher the air density is thus the vacuum reading is higher.
(7) Leakage level: when the vacuum is stable, check the leakage possibility from 0 to
    100 levels. 0 represent uncertain leakage. The larger the number, the greater the
    possibility of the leakage.
(8)     Key: return to the main interface.
2. Graph
Click the “Graph” button on the main interface to view the vacuum change in real time.
                                                                  STOP
(1) When the vacuum value drops below 8000 Pa, the vacuum change graph is displayed.
(2) The X axis represent time. Y axis represent vacuum. The vacuum value is from 0 Pa
    to 8000 Pa.
(3)     Key: return to the main interface.
                                          -6-
3. Settings
Click the “Settings”button on the main menu to enter the setting interface.
UnitP/T
Select Unit to set the unit for vacuum and temperature. The selected units are in blue.
Vacuum
Set the minimum vacuum value, logging interval, and recording status.
(1) Minimum Vacuum: set the target vacuum value. The range of the minimum vacuum
    value is 1Pa-100Pa. Click the change button to switch the minimum vacuum value;
    the display unit changes according to the selected unit.
(2) Logging Interval (S): 1, 5, 10, 30, 60, 120, 300.
(3) Recording Status: The logging can be turned on and off. The upper right corner
    shows the logging/recording status.
                                          -7-
Time
(1) Set and save the time then it shows at the top of the screen.
(2)      : Change key.
(3)      : return to the main interface.
4. System
In the main menu, click the “ System” button to enter the system interface.
(1) Running Time: refer to the accumulated running time of the motor.
(2) Holding Time: refer to the time for the vacuum pump to continue vacuuming
    after the target vacuum value is stable.
(3) Flow Rate: refer to the pumping speed of the vacuum pump.
(4)     : Change key.
(5)     : return to the main interface.
                                         -8-
Other functions
1. Temperature alarm: When the oil temperature or motor chassis temperature exceeds
   the upper limit alarm temperature set by the APP, the motor and solenoid valve are
   closed, the temperature alarm icon "      " is displayed at the top of the screen; When
   the temperature decrease within the alarm temperature, the alarm will be cleared, and
   the vacuum work continues . If temperature exceeds the offline alarm temperature, the
   temperature alarm icon " " will be displayed on the top of the vacuum pump screen,
   the motor operation will not be affected.
5. After using the vacuum pump
   1. Close the valve between the pump and the pumped container.
   2. Turn off the power switch on the pump, unplug the power plug, and remove the
      connecting pipe.
   3. Finally, close the air inlet cap to prevent dirt or floating particles from entering the
      pump cavity.
Care And Maintenance
Precautions
 1. When the vacuum is started, the air pressure is relatively high and the pump
     exhaust volume is large, thus oil mist will be overflowing with a large amount of
     gas. This is a common phenomenon for a rotary vane vacuum pump.
     When the vacuum drops to a very low level, the pump exhaust volume is small,
     and there will be no oil mist. Therefore, user should regularly monitor the oil level
     via the oil window to avoid running out of oil.
 2. During operation, storage and transportation, keep the vacuum pump clean to
     prevent pollutants such as water, mechanical impurities,etc.from entering the
     pump so as not to affect the service life of the vacuum pump and the normal
     operation of the system.
 3. The oil in the pump should be drained if it is not used for a long time. Store the
     pump in a dry and clean environment.
 4. The diameter of the pipe is larger than the inner diameter of the air inlet to avoid
     affecting the pumping speed.
 5. Check the tightness of the pipe connection to prevent leakage. Recommend using
     the vacuum grease on the connection and clamp it with a clamp to ensurethe seal.
 6. Do not use the pump to pump out gases which contain high level of oxygen,metal
     corrosive and explosive gases. In addition, do not pump in any gases that react
     with pump oil and containing large amount of water vapor.
 7. It is recommended to clean the catcher once every half year.
                                             -9-
How to replace the vacuum pump oil
1. Run the pump for about 3-5minutes until it is warmed up to runs out the pump oil.
2. When the pump is running, open the air inlet port at the same time to let the oil flow
   back to the oil tank. The pump must stop when replacing the oil.
   Stop the pump and open the oil drain plug. Then open the catcher to speed up the
   process of oil exhausting.
3. Tilt the pump body to completely drain the residual oil, and tighten the drain plug.
4. Open the catch device and add the new pump oil.
5. Cover the air inlet cap and start the pump to run-check the oil level after one
   minute.If the oil level is below the lower limit, slowly add the oil until it reaches the
   normal oil level. Finally, screw on the catcher.
Troubleshooting Guide
 Failure
            Cause of Issue                                Troubleshooting Method
 Phenomenon
                                                           Add oil between the maximum and minimum
               1. Insufficient Oil
                                                          level
               2. Pump oil Emulsified, Polluted           Replace with new oil
               3. The Oil Inlet Is Blocked or the Oil
                                                          Clean the oil inlet and filter
                  Supply is Insufficient
               4. Leakage of the Pump System and          Check the system and the connection to
 Low Vacuum
                  Connection                              prevent the leakage
                                                          Check the size of the pumped container,
               5. Improper Selection of Pump              recalculate and select the appropriate type of
                                                          pump
               6. Parts Wear and Tear due to Long
                                                          Repair or replace with a new pump
                  Time Used.
               1. The Oil Seal Is Damaged                 Replace oil seal
 Oil Leakage
               2. Loose or Damaged Tank                   Tighten the screws and replace the O -ring
                  Connections
                                                   -10-
              1. Too Much of Oil                          Drain the oil to the oil level line
Fuel
Injection     2. The Air Inlet Is Over Pressured for      Choose a pump with larger pumping speed
                 a Long Time
                                                          Open the air inlet, start the motor repeatedly
              1. Oil Temperature is Too Low
                                                          or heat up the pump oil
              2. Motor Failure                            Check and repair
 Difficulty
              3. Foreign Objects Fallen into the          Check and clean
 Starting
                 Pump
              4. Power Failure                            Check and repair
              5. The Power Supply Voltage Is Too
                                                          Check the power supply voltage
                 Low
APP Operating Instructions
1. Power on the vacuum pump.
2. Turn on the Bluetooth function of the mobile phone.
3. Open the “Elitech Gauge”APP.
                                                   -11-
4. Click "Add Device" to enter the device management interface. And the devices that
   can be connected are displayed in red. Click      on the top right of the device to
   establish a connection. After successful connection,the device bar turns green and
   the Bluetooth icon displays      .
5. For the connected device, click “ ” at the bottom right of the device to add the
   device to the work interface. The bottom of the screen shows that the device has
   been added successfully.
6. Click the return button in the upper left corner of the device management interface
   to enter the work interface.
                                         -12-
                                                                   Screenshot
Historical device
   Device name
                                                                   Real-time
                                                                   vacuum value
   Mac address
                                                                   Search to add
                                                                   a new device
The interface includes historical devices and the new devices.
(1) The history device displays the previously connected devices, click the device bar
to automatically connect the device. When the connection is successful, and "        "
is displayed in the upper left corner.
(2)                   : after clicking it will search for new devices.
(3)     : Bluetooth is not connected, click and try to connect this device.
(4)     : Bluetooth has been connected successfully, click to disconnect.
(5)     : Motor stop status, click to start running.
(6)     : Motor running status, click to stop the motor.
(7) Real-time vacuum value: display the real-time vacuum value.
                                         -13-
7. Click“   ” in the top left corner of the working interface to enter the menu.
   (Only some functions related to vacuum pump are described here).
(1) REPORTS
① Job Settings: Click the work button to enter the work list. Add a work list at the top right.
   Add work information and save the work.
                                             -14-
② Reporting: Click the report button to enter the report interface.
   Click the New report button, select data information by work or time, click the
   chart analysis button to view historical graph, click the “Export" button at the
   bottom to export data in Excel or PDF.
   Click the saved report to view historical Excel and PDF reports,and long press the
   report to delete.
③ Screenshot      Archive   save the screen capture file of the work interface. The file
   type is PDF.
                                           -15-
(2) CONFIGURATION
① SettingsClick the “Settings”button to enter the system settings
Keep Screen On: control the backlight of the screen.
Switch language: select language English or Other Languages.
General Alarm Preferences: single sound, single vibration or sound vibration alarm
at the same time.
② Help: Click the "Help" button to enter the help interface; click the ”About” button
to obtain version information; click "User manual" to enter the download interface,
and click the "Start" button to download the vacuum pump user manual.
③ ExitClick the“Exit”button to exit the APP
                                        -16-
8. Device Detail
   Click the connected device icon to enter the detailed interface of the device, display the
   vacuum pump status, and control the vacuum pump.
                                                                    Graph
Expected completion
                                                                    Motor control button
              time
                                                                    Temperature status of
   Real-time vacuum
                                                                    oil and shell
      Vacuum status                                                 Edit device name
        Leakage level
 Vacuum unit switch                                                 Set logging interval
                                                                    Recording status
  Read historical data
(1) Graph: when the vacuum value drops below 8000Pa, the real-time vacuum graph
    will be displayed.
(2) Expected Completion time: when the vacuum status is close to stable, the remaining
    vacuuming time will be displayed including the holding time.
(3) Motor Control: motor stop status displays "       ", click to run the motor; the motor
    running status displays"     ", click to stop the motor
(4) Vacuum Value: display the real-time vacuum value.
(5) Temperature Status Of Oil And Shell: According to the setting alarm temperature,
    display the temperature status "high", "normal", and "low" .
(6) Extraction Status: display the current vacuum value status, "Decrease", "Stable",
    "Rise", "--".
                                            -17-
(7) Leakage Rate: under the stable state of the vacuum value, make a judgment on
     the possibility of leakage of the current system, which is represented by 0-100;
     0 means the uncertain of the leakage status. The larger the number, the
     possibility of leakage is larger.
(8) Device Name: modify the device name, support up to 10 characters and numbers,
     click OK to confirm.
(9) Vacuum Unit: select vacuum unit, inHg , Torr, mbar, mTorr, Pa, micron, kPa.
(10) Record Switch: vacuum pump data recording switch, closed by default.
(11) Interval Time: select logging interval, 1S, 5S, 10S, 30S, 1Min, 2 Min, 5 Min.
(12) Read Historical Data: read the value saved by the vacuum pump while recording.
     Before reading the data, it will remind whether to delete the data after reading.
9. Setting
(1) Click Settings to set parameters and alarm values of vacuum pump.
                                        -18-
(2) Basic Settings
① Device Name: modify the device name, support up to 10 characters and numbers,
   click OK to save the device name.
② Vacuum Unit: select vacuum unit, inHg, Torr, mbar, mTorr, Pa, micron, kPa.
③ Temperature Unit: select the temperature unit, ºC, ºF.
④ Keep Time: the continue working time after reaching the target vacuum, the
   setting range is 0 to 200 minutes.
(3) Alarm Settings
① Vacuum Min: when the vacuum reaches the alarm value, the mobile phone will
   alarm and vibrate.
② Temperature Alarm Setting: set the max and minimum oil temperature and chassis
   temperature alarm. The temperature range of vacuum pump is -50°C to 85°C.The
   minimum temperature shall not exceed the max temperature. When the alarm
   temperature is triggered, the mobile phone will alarm with sound and vibration.
(4) Record Settings
① Record Switch: set vacuum pump data recording status, closed by default.
② Interval Time: set the logging interval , 1S, 5S, 10S, 30S, 1Min, 2 Min, 5 Min.
iOS system: Enter the APP Store and search for "Elitech Gauge" to download.
Android system: Enter Google Play search for “Elitech Gauge” to download or scan
QR code below to download.
                 Android system                       iOS system
                                         -19-