0% found this document useful (0 votes)
42 views16 pages

DataSheet MobilettMiraMax

The Mobilett Mira Max is a compact and powerful mobile X-ray system designed for flexible imaging in hospitals, featuring advanced MAX assistance and detection technologies. It includes wireless detectors, automatic recharging, and a high-performance generator for optimal image quality. The system is user-friendly with a digital workflow, secure access, and various options for enhanced functionality.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
42 views16 pages

DataSheet MobilettMiraMax

The Mobilett Mira Max is a compact and powerful mobile X-ray system designed for flexible imaging in hospitals, featuring advanced MAX assistance and detection technologies. It includes wireless detectors, automatic recharging, and a high-performance generator for optimal image quality. The system is user-friendly with a digital workflow, secure access, and various options for enhanced functionality.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 16

s

Data Sheet

Mobilett Mira Max


Your mobile imaging companion

siemens.com/healthcare
Mobilett Mira Max
Your mobile imaging companion

MAX assistance: MAX detection:


Mobilett Mira Max is your mobile Your intelligent personal assistant that Applies Intelligent Innovation at every
­imaging companion that lets you proactively works for you, supports you step of the image formation.
bring a broad range of unique MAX and protects you – even in the most Enabling a unique Siemens way of provid-
­advantages to the entire hospital – ­challenging situations. ing a comprehensive imaging technology.
in a compact, powerful X-ray system For the best possible image with your
that is just always ready. MAX wi-D
­system every time.
• Large-format 35 cm x 43 cm wireless
Flexible to meet your challenges. MAXreach detector with ergonomic handle
• Four-axis counterbalanced arm design • Only 3.3 kg light and just 19 mm thin
Mobilett Mira Max always assists you in
with 213.5 cm SID and 124.5 cm • Automatic charging in the detector
finding the perfect imaging position:
­horizontal extension holder
Being lightweight and equipped with
• Lateral arm rotation of ± 90 degrees
the MAXreach tube arm, it lets you MAX mini
• Exceptional positioning flexibility
­handle any X-ray task with the • Small-format 24 cm x 30 cm wireless
• Fully integrated cables
­necessary precision – even in detector
­challenging situations. Mobility • Only 1.6 kg light and just 16 mm thin
• Ergonomic drive handle for effortless • The right size for incubators, pediatric,
MAX image quality in every situation. maneuvering orthopedic and trauma cases
With impressive imaging power of • Smartly designed detector holder
MAXswap
35 kW and our handy high-performance ­offering large foot space
• Motor-assisted driving • MAXswap is the right way to share,
MAX wi-D and MAX mini detectors, allowing you to swap detectors between
Mobilett Mira Max delivers brilliant • Compact and light-weight system
design multiple systems so you always have the
results in seconds – giving you optimal right detector when and where you
diagnostic certainty anywhere in the Always ready by design need it
hospital. • Automatic recharging of MAX wi-D in • Easy one-click registration
the detector holder • Available for MAX wi-D and MAX mini
Always ready to assist you. • If batteries are depleted, immediate
X-ray generation
Like a real friend, Mobilett Mira Max X-ray generation via mains possible
• High-frequency generator with auto-
never lets you down: • Allows manual movement in case bat-
matic exposure control
Detector sharing via MAXswap, teries are depleted
• High-performance 35 kW X-ray tube
automatic detector recharging, and • Switch to analog mode possible in case
flexible power supply are just three of detectors are not available Imaging system
the many reasons to rely on your Security mechanisms • High-performance PC with flat panel
imaging ­companion. display and syngo FLC user interface
• Secure access via key or numeric keypad
• DiamondView Plus, an enhanced multi-
(option)
scale post-processing method for
• Easy lock function for preventing access
improved contrast detail and reduced
to patient data, motor function, and
image noise
X-ray release – and seamless resump-
• Pre-set organ programs containing gen-
tion of work upon return, as the system
erator and image processing parameters
is not turned off while being locked
• Comprehensive DICOM functionality
• Additional, sophisticated user manage-
ment and data protection (option) Options (selection):
Operation • Remote control for safe system
­operation
• Digital one-stop workflow from patient
• High-power battery pack for an even
registration to image documentation
longer operating time
with syngo FLC
• External Wi-Fi connectivity
• DAP measurement chamber
• HIPAA security package
• Giraffe design
System specification

Mobilett Mira Max

System specification Page 4–8

Clinical workflow Page 9 – 11

Accessories Page 12

Room planning Page 13 – 15

Mobile X-ray unit


General-purpose digital mobile X-ray unit for radiography of the recumbent and seated patient
Forward/Reverse motor drive for relocating the unit, controlled at main handle
Max. incline of 7 degrees, high speed adjustable from 1.2 to 1.6 m/s
Ever ready: If the batteries are depleted, the unit can be moved manually and examinations can be performed with mains
power (with slightly reduced maximum power)
Reduced speed feature (creep mode) for fine tuning unit positioning
Counterbalanced arm can be rotated by ± 90 degrees and allows easy adjustment for all the different projections
Flexible X-ray tube
Clearly designed control panel for free setting of kV and mAs data via touch keys (analog mode)
Exposure release via hand switch
Remote control of collimator light from control panel
Automatic cable winder with 4 m (157.5“) power cable for connection to any grounded outlet,
3 m (118.1“) in countries with line voltages 100 V to 120 V
Available with either standard key or numeric keypad / PIN code (system option) for secure system access
System can be locked without requiring a shutdown

PC hardware
The X-ray system incl. digital image processing is controlled from the central, integrated image and control station
High-end PC imaging system based on Windows 7 Ultimate 64 bit with syngo-like user interface
17“ color flat display for touch screen operation
Memory capacity approx. 10.000 images

4 * Option
System specification

Mobilett Mira Max

Generator
Power output Max. 133 kV at 225 mA
Max. 450 mA at 54 kV
Max. 35 kW at 96 kV, 365 mA, 10 ms
Max. 30 kW at 99 kV, 300 mA, 100 ms (acc. to IEC 60601-2-7)
Min. 0.32 mAs
kVp range 40 kV to 133 kV in 24 or 47 steps
mAs range 0.32 mAs to 360 mAs in 27 or 53 steps
kV accuracy < ± 3 kV (40 kV to 60 kV)
< ± 5 % (60 kV to 133 kV)
mAs accuracy < ± 5 % (2 mAs to 360 mAs)
< ± 0.1 mAs (0.32 mAs to 2 mAs)
Shortest exposure time 1 ms
Max. exposure time 3.2 s (digital mode)
5 s (analog mode)
Exposure release Detachable remote control with 5 m coiled cord
Enclosure protection rating IPX0 (not protected)
Protection against electric shock Class I acc. to IEC 60601-1

X-ray tube
Tube housing (single tank with rotating-anode tube)
X-ray tube Siemens P135/30 R
Stabilization Microprocessor-based adjustment and regulation during exposure
Focal spot nominal value (IEC 336) 0.8
Anode angle 15°
Anode speed Up to 9,000 r.p.m., 150 Hz
Heat storage capacity of tube housing 800,000 J = 1,100,000 HU
Heat storage capacity of anode 90,000 J = 122,000 HU
Inherent filtration:
Collimator ≥ 1.6 mm Al (at 70 kVp)
Tube housing assembly ≥ 1.9 mm Al (at 70 kVp)
Total filtration ≥ 3.5 mm Al acc. to IEC 522/1976
Collimator Manually adjustable multileaf collimator, rotatable ± 90°
Collimator light and light field intensity LED min. 180 lux at 1 m SID
Tube housing assembly Maximum temperature 60 °C, maximum 2,000 mAs/h,
nominal continuous rating 75 W until housing temperature reaches 60 °C

* Option 5
System specification

Mobilett Mira Max

X-ray tube & generator

mAs range

Battery mode (digital) Battery mode (analog)


At 40 kV to 43 kV: 0.32 mAs to 360 mAs At 40 kV to 65 kV: 0.32 mAs to 360 mAs
At 44 kV to 49 kV: 0.32 mAs to 320 mAs At 66 kV to 73 kV: 0.32 mAs to 320 mAs
At 50 kV to 56 kV: 0.32 mAs to 280 mAs At 75 kV to 83 kV: 0.32 mAs to 280 mAs
At 57 kV to 62 kV: 0.32 mAs to 250 mAs At 85 kV to 93 kV: 0.32 mAs to 250 mAs
At 63 kV to 70 kV: 0.32 mAs to 220 mAs At 96 kV to 105 kV: 0.32 mAs to 220 mAs
At 71 kV to 77 kV: 0.32 mAs to 200 mAs At 109 kV to 117 kV: 0.32 mAs to 200 mAs
At 79 kV to 85 kV: 0.32 mAs to 180 mAs At 125 kV to 129 kV: 0.32 mAs to 180 mAs
At 88 kV to 96 kV: 0.32 mAs to 160 mAs At 133 kV: 0.32 mAs to 160 mAs
At 99 kV to 109 kV: 0.32 mAs to 140 mAs
At 113 kV to 125 kV: 0.32 mAs to 125 mAs
At 129 kV to 133 kV: 0.32 mAs to 110 mAs

Mains mode (digital) Mains mode (analog)


At 40 kV to 44 kV: 0.32 mAs to 125 mAs At 40 kV to 51 kV: 0.32 mAs to 140 mAs
At 45 kV to 50 kV: 0.32 mAs to 110 mAs At 52 kV to 57 kV: 0.32 mAs to 125 mAs
At 51 kV to 55 kV: 0.32 mAs to 100 mAs At 59 kV to 65 kV: 0.32 mAs to 110 mAs
At 56 kV to 60 kV: 0.32 mAs to 90 mAs At 66 kV to 72 kV: 0.32 mAs to 100 mAs
At 62 kV to 68 kV: 0.32 mAs to 80 mAs At 73 kV to 79 kV: 0.32 mAs to 90 mAs
At 70 kV to 77 kV: 0.32 mAs to 71 mAs At 81 kV to 90 kV: 0.32 mAs to 80 mAs
At 79 kV to 90 kV: 0.32 mAs to 63 mAs At 93 kV to 99 kV: 0.32 mAs to 71 mAs
At 93 kV to 102 kV: 0.32 mAs to 56 mAs At 102 kV to 113 kV: 0.32 mAs to 63 mAs
At 105 kV to 113 kV: 0.32 mAs to 50 mAs At 117 kV to 125 kV: 0.32 mAs to 56 mAs
At 117 kV to 125 kV: 0.32 mAs to 45 mAs At 129 kV to 133 kV: 0.32 mAs to 50 mAs
At 133 kV: 0.32 mAs to 40 mAs

6
System specification

Mobilett Mira Max

MAX wi-D
Detector technology Cesium iodide scintillator coupled to TFT matrix with amorphous silicon
technology
Dimensions (active area) 34.8 cm x 42.4 cm (13.7” x 16.7”)
Active detector matrix 2350 x 2866
Dimensions with detector housing 44 cm x 46.1 cm x 1.9 cm (17.3” x 18.1” x 0.75”)
Pixel size 148 µm
Semiconductor material Amorphous silicon (a-Si)
Scintillator Cesium iodide (Csl)
Digitization depth 16 bits
DQE in %; 2 µGy (RQA5) 70 % at 0.05 lp/mm
51 % at 1 lp/mm, (IEC 62220)
42 % at 2 lp/mm, (IEC 62220)
29 % at 3 lp/mm, (IEC 62220)
19 % at Nyquist
MTF in % (RQA5) 63 % at 1 lp/mm, (IEC 62220)
35 % at 2 lp/mm, (IEC 62220)
19 % at 3 lp/mm, (IEC 62220)
12 % at Nyquist
Data transmission WLAN 1) < 2 s preview; < 5 s full image
Thickness 19 mm (0.75”)
Weight 3.3 kg (7.3 lbs)
Max. load capacity 150 kg (330 lbs) with patient recumbent
100 kg (220 lbs) with patient standing
Battery Lithium-ion, rechargeable, exchangeable
Charging time 90 minutes for 80 % battery capacity
Battery operation time Up to 950 images
Up to 5.5 h during regular utilization
Up to 6 h in standby mode
Clip-on grid* Grid, Pb 5/85, f0 = 115 cm; Pb with aluminum interspacing
Charging location Onboard detector holder, or external battery charger*
WLAN Standard IEEE 802.11n, 2 x 2 mimo
If there is a WLAN or other wireless equipment in your working environment, please consult your Siemens representative
for optimal set-up of the wireless connection
IEC Regulations Electromagnetic compatibility: compliance with IEC 60601-1-2
This detector does not affect pacemakers that comply with
DIN EN 45502-2-1, Section 27

* Option
1)
The preview/full image transmission time depends on the quality of the WiFi link and the selected processing parameters 7
System specification

Mobilett Mira Max

MAX mini
Detector technology Cesium iodide scintillator coupled to TFT matrix with amorphous silicon
technology
Dimensions (active area) 22.5 cm x 28.4 cm (8.9” x 11.2” )
Active detector matrix 1520 x 1920
Dimensions with detector housing 26.9 cm x 32.9 cm x 1.6 cm (10.6” x 13” x 0.63”)
Pixel size 148 µm
Semiconductor material Amorphous silicon (a-Si)
Scintillator Cesium iodide (Csl)
Digitization depth 16 bits
DQE in %; 2 µGy (RQA5) 66 % at 0.05 lp/mm
50 % at 1 lp/mm, (IEC 62220)
40 % at 2 lp/mm, (IEC 62220)
24 % at 3 lp/mm, (IEC 62220)
17 % at Nyquist
MTF in % (RQA5) 61 % at 1 lp/mm, (IEC 62220)
31 % at 2 lp/mm, (IEC 62220)
15 % at 3 lp/mm, (IEC 62220)
12 % at Nyquist
Data transmission WLAN 1) < 1.5 s preview; < 3.5 s full image
Thickness 16 mm (0.63”)
Weight 1.6 kg (3.5 lbs)
Max. load capacity 150 kg (330 lbs) with patient recumbent
100 kg (220 lbs) with patient standing
Battery Lithium-ion, rechargeable, exchangeable
Charging time 4 h in battery charger
Battery operation time Up to 950 images
Up to 5.5 h during regular utilization
Up to 6 h in standby mode
Charging location Battery charger
WLAN Standard IEEE 802.11n, 2 x 2 mimo
If there is a WLAN or other wireless equipment in your working environment, please consult your Siemens representative
for optimal set-up of the wireless connection
IEC Regulations Electromagnetic compatibility: compliance with IEC 60601-1-2
This detector does not affect pacemakers that comply with
DIN EN 45502-2-1, Section 27

* Option
8
1)
The preview/full image transmission time depends on the quality of the WiFi link and the selected processing parameters
Clinical workflow

Mobilett Mira Max

Patient data administration


Patient registration Retrieval of patient list and examination data from the hospital/radiology
information system (HIS/RIS)
Emergency patient registration
Patient, study and image data administration
Configuration functions
Security Package* (HIPAA) Advanced security functions incl. user management and data access
protection

Examination preparation
Workflow Prior to acquisition, the patient data is entered via the patient management
system (HIS/RIS) or at the control console
The exposure parameters are selected via the organ programs
Then the patient and X-ray unit are positioned and the exposure is released
The exposure released at the central system control is read by the detector
within a few seconds
The image is shown on the control display for orientation and is available
at the imaging system output for sending in DICOM format, e.g. to reporting
workstations, imaging networks, laser cameras, etc.
Clinical Assurance Program (CAP): Collection of deleted images, studies and
patient data including evaluation options
Key-protected by either standard key or PIN code (system option)

* Option 9
Clinical workflow

Mobilett Mira Max

Image acquisition/display/processing
Acquisition and preprocessing Selection and configuration of organ programs
Selection of generator and postprocessing parameters
Parameterization of image preprocessing Gain, harmonization, edge enhancement and look-up tables (LUT)
Display of the current exposure within 3 seconds (preview); complete
image in 10 seconds max.
Display of image labels (L/R, a.-p./p.-a.)
DiamondViewPlus multi-scale technique for image postprocessing with
higher detail contrast and reduced noise
Startup time < 60 s
DiamondView is a multi-scale technique, i.e. filter size and strength are weighted differently and serve for adaptation to
the total image content
DiamondView Enhances the signal exploitation of the dynamic range and improves the
organ-specific detail contrast (soft tissue and bone)
Image processing functions Rotation
Vertical and horizontal image flip
Zoom
Windowing for contrast/brightness
Black/white image inversion

10 * Option
Clinical workflow

Mobilett Mira Max

Data transfer and documentation

DICOM network interfaces


DICOM Send/StC Transmission of images to a DICOM network for viewing and archiving
Confirmation from the image archive (StC = Storage Commitment)
DICOM Print Printing of images to a DICOM laser camera via virtual film sheet
DICOM Worklist/MPPS Get Worklist function for importing patient data from a data management
system (RIS/HIS). XRF, CR and DX worklist entries supported, configurable
Modality Performed Procedure Step (MPPS) function for sending
examination statistics and dose information to a data management system

Documentation
Image data management Transmission of images to network
Automatic and selective printing with virtual film sheet
Up to 3 network nodes at the same time and one laser camera configurable
Export of image data (12 bit) to CD/DVD recorder in DICOM or TIFF format
Export to USB device in DICOM or TIFF format
Recycle bin This feature can be enabled or disabled
Stores rejected and deleted images that are not archived and not printed in a
separate folder
Clinical Assurance Program (CAP) Provides statistics of rejected images
Exposure index (EXI) monitoring Provides minimum and maximum EXI value for export
Printer connection* For paper printing to a Level 2 PostScript printer

* Option 11
Accessories

Mobilett Mira Max

Remote control*
Infrared remote control unit for exposure release for optimal radiation protection
Operating range 10 m (33 ft), 180°
Battery For 30,000 exposures
Remote finder function Locates the remote control when it is not in park position

WLAN connectivity*
Industrial-strength WLAN module for increased reach and improved connectivity
Supported WLAN standards 802.11 a/b/e/g/h/i/n
Supported frequency bands 2.4 / 5 GHz
Security / Authentication 802.11 i, 802.1x, WPA/WPA2
WPA2 Enterprise supplicants EAP-TLS, EAP-TTLS (MSCHAPv2), EAP-PEAP
(MSCHAPv2)
Supports certificates and private key upload / storage (multiple)
Data encryption TKIP, AES

Dose Area Product (DAP) measurement chamber*


Integrated dose area product measurement system
Automatic transfer of the dose value measured to the imaging system
Part of the DICOM header when image information is transferred
Typical filtration < 0.4 mm Al equivalent
DAP resolution 0.01 µGym2
Maximum measurable DAP 1 x 105 µGym2

12 * Option
Room planning

Mobilett Mira Max

Operating data
Power supply Grounded safety power socket (L/N/Gnd)
Overcurrent release An on-site 15 A/characteristic C overcurrent release (circuit breaker) is
recommended by DIN VDE 0100-710 for power supply voltages of
230 V/240 V
Country-specific requirements apply for the fuse values to be used for other
power supply voltages
Power supply voltage 100 V (12 A) to 240 V (5 A), 50/60 Hz, 1.2 kW
Standard battery pack (65 Ah) Battery Sealed lead acid batteries (65 Ah), rechargeable
Charging time 5 h (80 %)
Battery operation time 6 h/200 exposures at 70 kV/20 mAs
Standby time (On state) 12 h
Standby time (Off state) 48 h
High power battery pack (90 Ah)* Battery Sealed lead acid batteries (90 Ah), rechargeable
Charging time 6.5 h (80 %)
Battery operation time 8 h/270 exposures at 70 kV/20 mAs
Standby time (On state) 16 h
Standby time (Off state) 64 h

* Option 13
Room planning

Mobilett Mira Max

Environmental conditions
Operation
Temperature range + 10 °C to + 35 °C
Relative humidity 30 % to 75 %, non-condensing
Barometric pressure 800 hPa to 1060 hPa

Transport/storage
Temperature range – 10 °C to + 55 °C
Relative humidity 10 % to 70 % (up to 15 weeks, if FD packed)
Barometric pressure 800 hPa to 1060 hPa

Dimensions and weight


Footprint (l x w x h) 113 cm x 59.5 cm x 157 cm (44.5“ x 23.4“ x 61.8“)
Tube positioning:
Max. focal spot height 213.5 cm (84“)
Min. focal spot height 47 cm (18.5“)
Max. horizontal ext. 124.5 cm (49“)
Min. horizontal ext. 39.5 cm (14.4“)
Weight (without optional accessories) Approx. 375 kg (825 Ibs)

14 * Option
Room planning

Mobilett Mira Max

Dimensions in cm (inches)

15
On account of certain regional limitations of sales rights and service In the interest of complying with legal requirements concerning the
availability, we cannot guarantee that all products/services/features environmental compatibility of our products (protection of natural
included in this brochure are available through the Siemens sales resources and waste conservation), we may recycle certain
organization worldwide. Availability and packaging may vary by ­components where legally permissible.
­country and are subject to change without prior notice. For recycled components we use the same extensive quality assurance
measures as for ­factory-new components.
The information in this document contains general descriptions of the
technical options available and may not always apply in individual Any technical data contained in this document may vary within
cases. defined tolerances. Original images always lose a certain amount of
detail when reproduced.
Siemens reserves the right to modify the design and specifications
contained herein without prior notice.
Please contact your local Siemens sales representative for the most Mobilett Mira Max 144 43 282
current information. Mobilett Mira Max 144 49 046

Siemens Healthcare Headquarters Legal Manufacturer


Siemens Healthcare GmbH Siemens Healthcare GmbH
Henkestr. 127 Henkestr. 127
91052 Erlangen 91052 Erlangen
Germany Germany
Phone +49 9131 84-0
siemens.com/healthcare

Order No. A91XP-40012-3T3-7600 | Printed in Germany | HC DI XP MK 0317pdf | © 03.2017 Siemens Healthcare GmbH

siemens.com/healthcare

You might also like