Blood Collection Mixer
Blood Collection Mixer
Model CM745
Centron Corporation
SERVICE MANUAL
Model CM745
No. CAT.CM745SM1.6Ee
ISSUE 7 AUG 2015
Centron Corporation
44 Beolmal-ro, Dongan-gu, Anyang
Gyeonggi-do, 431-804 Korea
Tel.: +82-31.422.7890
Fax : +82-31.422.7866
www.centrons.com
Table of Contents
-------------------------------------------------------------------------------------------------------------
Page
Table of Contents i
List of Figures ii
Safety Symbols iii
Notice iii
Important Notes iv
1. Introduction 1-1
1-1. General 1-1
1-2. Features 1-2
1-3. Accessories 1-3
1-4. Specifications 1-4
2. Appearance and Functions 2-1
2-1. Front View 2-1
2-2. Rear View 2-2
2-3. Control Panel 2-3
2-4. LCD Display Panel 2-4
2-5. Tube Clamp Module Assembly 2-4
i
Page
5. Checks and Adjustments 5-1
5-1. Test Mode Subroutines 5-2
5-2. Tray Leveling 5-3
5-3. Weight Calibration 5-4
6. Troubleshooting 6-1
6-1. Precautions 6-1
6-2. Test Equipment 6-1
6-3. Preparation 6-1
6-4. Troubleshooting Procedures 6-2
List of Figures
Page
[Figure 2-1] Front View 2-1
[Figure 2-2] Rear View 2-2
[Figure 2-3] Control Panel 2-3
[Figure 2-4] Display Panel 2-4
[Figure 2-5] Tube Clamp Module 2-4
[Figure 3-1] CPU Board Functional Block Diagram 3-2
[Figure 3-2] Power Board Functional Block Diagram 3-4
[Figure 4-1] Battery Recharging 4-8
[Figure 4-2] Moving Clamp Module Assembly 4-9
[Figure 6-1] Troubleshooting Flow Charts 6-3
[Figure 7-1] System Wiring Diagram 7-2
[Figure 8-1] Exploded View of CM745 8-2
[Figure 8-2] CPU Board Layout 8-3
[Figure 8-3] Power Board Layout 8-4
[Figure 8-4] LCD Backlight Board Layout 8-5
[Figure 8-5] Battery Contact Board Layout 8-5
[Figure 8-6] Battery Terminal Board Layout 8-6
[Figure 8-7] Hall Sensor Board Layout 8-6
ii
Safety Symbols
WARNING
indicates that inappropriate use of device might result serious injury to users and
damage to the device. Users must follow the procedures as specified in this manual.
CAUTION
indicates the presence of hazard that could result in minor injury or property damage
if ignored.
NOTE
describes information for the installation, operation, or maintenance which is
important for proper use.
Notice
This equipment has been tested and found to comply with the limits of EMC
requirements in EN61326-1:2006. These limits are designed to provide
reasonable protection against harmful interference from adjacent devices.
Do not put excessive weight on the tray or push the tray downward in a
forceful manner. (This avoids damage to the weight-sensitive device
CAUTION under the tray.) Do not place weights over 2 kg (4.4 lbs) on the tray.
iii
Important Notes
Power Source
The device operates on either a battery pack or an AC power adapter, or both.
Battery Capacity
Turn power off to save the battery when the device is not to be used for hours. Device
power will turn off automatically if no control key has been touched for 30 minutes.
The device runs longer than 15 hours continuously with a fully charged battery pack.
Full charging may take 4 hours.
Battery Recharging
Recharge the battery pack prior to the first use. To recharge the battery pack, insert it in
the device and plug the AC power adapter in the jack on the rear side.
The battery pack, although it is rechargeable, is a consumable item. Replace the pack with
a new one when it seems the pack may have been degraded too much.
Backup Battery
This device is equipped with a backup battery to prevent any data loss in case of
accidental black out or battery discharge during blood collection. The built-in battery will
keep the data in process for 20 minutes.
Do not keep operating the device only with the backup battery.
Storage
Avoid leaving the device in a location where temperature is higher than 40 C (104 F).
If the equipment is not to be used for a long time, it is recommended to protect it from
dust, dirt or humidity.
Keep it away from any heating source, direct sunlight, water vapor, or such
environment as poor ventilation, excessive shock or vibration.
iv
1. Introduction
1-1. General
This Service Manual describes technical information of the Blood Collection Mixer
CM745 to aid service and maintenance of the equipment.
1. Introduction
2. Appearance and Functions
3. System Descriptions
4. Standard Operating Procedures
5. Checks and Adjustments
6. Troubleshooting
7. Schematic Diagrams
8. Technical Illustrations
9. Parts List
1-1
1-2. Features
CM745 is a compact and easy-to-use device, equipped with convenient features as;
(1) Clear LCD display and LED indicators provide various operational information.
(2) The device is so flexible that users can adjust operational parameters as;
- collection target volume
- criteria on which alarms are triggered
- on/off of beep sounding
- tray rocking speed
- simple weight calibration in digital method
(3) Various audible and visual alarms work.
(4) It operates on dual power sources - the AC power adapter and the battery pack. A
large capacity Li-ion rechargeable battery pack lasting 15 hours is provided as an
option for outdoor use.
(5) Onboard battery(rechargeable) is built in for emergency power backup. It supplies
power for 20 more minutes to safely complete ongoing collection.
(6) Tube clamp module can be installed either at left side or at right side.
(7) A deeper tray may be selected as an option.
(8) A barcode scanner may be opted for label checking, which reads most barcodes
including ISBT128 and Codabar codes.
(9) The device and accessories can be stored and moved in a canvas carrying bag or
a carrying case if it is separately purchased.
1-2
1-3. Accessories
A standard set of CM745 package consists of the following;
The items below are not included in the standard set of CM745, but are available under
separate orders.
Battery Pack BL101 Battery Charger BC110 Barcode Scanner & Holder
1-3
1-4. Specifications
1-4
2. Appearance and Functions
(1) tray(detachable) - is plastic molded on which blood collection bags are loaded. Tray
should be detached when mixer is stored in the carrying/transportation bag.
(2) sample tube holder – holds samples tubes.
(3) power indication lamp – is lit when power is turned on.
(4) POWER switch - turns the power on and off.
(5) control panel – various controls buttons are located.
(6) LCD display panel - is a custom-designed display panel on which all information is
shown.
(7) target preset – displays preset target volume.
(8) tube clamp module(movable) – crimps blood bag tubing to block blood flowing into
blood bag. It can be affixed to the other side.
2-1
2-2. Rear View
(1) tube clamp module(movable)- crimps blood bag tubing to block blood flowing into
blood bag. It can be affixed to the other side.
(2) barcode scanner connector – is the connecting port of the barcode scanner to the main
unit. It is also used to download an operating software from a PC.
(3) power adapter jack - takes power directly from the AC power adapter. When the
adapter is plugged in, it supplies power to the equipment and the battery pack will be
recharged.
(4) sample tube holder - holds samples tubes.
(5) tray leveling screw - is used to level the position of the tray.
(6) tray(detachable) - is plastic molded on which blood collection bags are loaded. Tray
should be detached when mixer is stored in the carrying/transportation bag.
2-2
2-3. Control Panel
(1) START indicator lamp – It is lit in green color when power is turned on, but it will
be changed to yellow color when alarm is enabled. It also indicates in amber color when
battery is charging in case that device is powered off.
(2) [5 ml ▲] - increases target collection volume by 5 ml increment.
(3) [POWERCAL] - dual function key:
Short press (a click for 1 second)- turns device power on or off.
Short press during booting period of 10 seconds - calibration procedure starts.
(*) Remark:
If the key is pressed during a collection cycle, power does not turn off , but a
warning alarm (Wrong key alarm: 3 short beeps) sounds. This is a measure not to
make the device turned off by an unintentional or accidental press during a collection
cycle.
If the key is pressed and held for 10 seconds continuously, device power does turn
off regardless of the procedure. This is a measure to make the device turned off
during a collection cycle, in case that user wants to stop a collection before
completion.
(4) [PRESET] - displays preset target collection volumes.
(5) [5 ml ▼] - decreases target collection volume by 5 ml increments.
(6) Display panel - displays weight, volume, elapsed time and operation symbols.
(7) CLAMP lamp [ ] - turns on in amber color when the tube clamp is closed.
(8) [CLAMP] - opens or closes clamp.
(9) [END] – Pressing this key ends the collection procedure.
(10) [START] - starts a blood collection cycle. The collection progress indicator starts
working.
(11) [TIMEBEEP] - dual function key:
Short press (one click) - displays elapsed time from start (min : sec).
Long press (hold for 3 seconds) - turns beep sound on or off. When symbol
appears, it is mute and when sound resumes, symbol disappears on the LCD.
(12) Target preset – displays preset target volume.
2-3
2-4. LCD Display Panel
2-4
3. System Descriptions
3-1. General
Blood Collection Mixer CM745 consists of a main unit, an AC power adapter, battery
pack (option), a barcode scanner (option), and a canvas carrying bag (option).
Main unit comprises a tube clamp module, six circuit board assemblies, a tray motor
assembly and a load cell. Battery pack is inserted in the compartment at the left side of
the main unit.
AC power adapter is used for both recharging the battery and for supplying power to
the main unit directly.
. CPU Board
. Power Board
. LCD Backlight Board
. Battery Contact Board
. Battery Terminal Board
. Hall Sensor Board
Following sections describe operation theory and function of each assembly in detail.
3-1
3-2. CPU Board
(1) Microprocessor;
Microprocessor U2 is of ATMEL microprocessor. It is the heart of the
equipment, controls all circuits, switches, tray, clamp, displays and alarms.
When power is on, it self-checks and, after 10 seconds, tares the weight of
hardware which are on the tray. Every time it monitors switch actions.
(2) LCD Driver and LCD
It is a LCD with backlight, which displays commands and data received from
microprocessor.
(3) Bar LED
Bar LED indicates the collection progressive status and controlled by CPU.
(4) Control Keys
Commands from the control keys are processed by CPU. Control keys are located
at the Membrane Panel.
(5) Hall Sensor;
Hall sensor data which is used for leveling the tray is fed into CPU.
3-2
(7) Load Cell Data;
CPU receives the load cell data and converts it into weight data.
(8) Load cell Amplifier and A/D Converter
Load cell Amplifier
Load cell generates very weak signal that is proportional to weight applied
on it. OP amp stages U6 and U4 amplify the weak signal by 60 dB and feed
it to A/D converter. A potentiometer VR1 adjusts offset voltage, and VR2
adjusts signal gain.
A/D Converter
A/D converter U7 converts an analog signal to a digital signal.
It communicates with the microprocessor in SPI(Serial Peripheral Interface)
format.
(9) Barcode Data;
CPU receives barcode data from the Power Board. This signal line is also used for
firmware upgrading.
(10) Power on/off control
U3 is a power on and off controller. When power switch signal is received, U3
turns on power via Q2 FET in the Power Board Ass’y. And when POWER_OFF
signal is received from CPU, it turns off power via Q2 FET in the Power Board
Ass’y.
(11) Power
There are four kinds of powers such as 14V, 14.3V, 14.7V and 15V.
14V : It is the main power which is supplied through Q2 FET when power
switch is turned on.
14.3V: It is the primitive power for activating the power control circuit which
allows the main power(14V) is feeding to circuits by the power on/off
switch.
15V: It is used for charging the spare battery.
14.7V: It is the power for lighting the START indicator lamp when the main
power is turned off.
3-3
3-3. Power Board
Power Board comprises a power control block, a barcode scanner driver and a clamp
motor controller.
3-4
3-4. Hall Sensor Board
The board has a magnetic sensor called Hall sensor. The component senses magnetic
field and generate a signal. As the tray motor assembly has magnets on its cam
structure, this sensor senses rotation angle of motor shaft and tray position.
So tray motion and position information is checked by the Hall Sensor and fed to
microprocessor.
3-5
4. Standard Operating Procedures
4-1. Starting Up
(1) Installation
a. Unpack the package and check accessories as in Section 1-3 Accessories.
b. The device can be powered by the AC power adapter or the battery pack, or both.
When the battery pack should be used, recharge it first. Refer to Section 4-6
Battery Recharging.
c. Check mounted position of tube clamp module. Users can move it to the other side.
Refer to Section 4-7.
b. Users can change the parameters. Refer to Section 4-4 Setting-up Parameters.
Parameter changes are stored in the device memory, and when power is turned on
again, the device starts working at changed setting.
4-1
4-2. Blood Collection Procedures
(2) To change the target volume preset value, press [PRESET] key.
A default value(450 ml) will be displayed.
To select other value, just press [PRESET] key. (e.g.)
Four pre-determined default values will be displayed in turn.
The preset value can be changed by [5 ml ▲] or [5 ml ▼]
key between 10-600 ml range.
After 5 seconds, a new preset value will be saved and go
back to the previous step.
A new value will be remained unless power is turned off.
(e.g.)
(4) If a bag information is read by a barcode scanner, a preset
target volume will be automatically set and displayed for 5
seconds.
4-2
REMARK: A collection cycle begins only when the bag
weight is between 26-499 grams. If the key is pressed when
the weight on the tray is either below 26 grams or over 500
grams, an error message ("E r r") is displayed.
The display panel shows the current weight in 3 seconds.
Check the blood bag to adjust what's wrong.
Press [START] key again.
NOTE: The unit ignores an abrupt change of the weight while
the unit is in a collection cycle. Any weight change of +/ 30
grams or greater in 1 second interval will be ignored so the
display remains unchanged.
NOTE: The following alarm may sound.
- Low Flow alarm if blood flow rate is below 30 ml/min.
- T1 Time alarm if 10 minutes have elapsed.
- T2 Time alarm if 20 minutes have elapsed.
4-3
The following procedure is applied when a barcode scanner is ready to use.
NOTE
4-4
4-3. Alarms
Alarming is informed in 2 ways – beep and indication.
Alarm Audible Lamp
◙ condition remark ● action sound indication
1 End Collection Alarm:
◙ When collected volume reaches target volume; - 3 long beeps repeat per - green
● Clamp closes. 30 sec flashing
2 Low Flow Alarm: - yellow and
◙ When blood flow goes below the set limit; - 5 short beeps repeat per LOW &
Limit is variable in Section 4-4. 12 sec ALARM
flashing
3 High Flow Alarm: - yellow and
◙ When blood flow goes above the set limit; - 14 short beeps repeat per &
Limit is variable in Section 4-4. 12 sec ALARM
flashing
4 T1 Time Alarm:
◙ When elapsed time passes 10 minutes; - 20 short beeps for 10 sec.
5 T2 Time Alarm: - yellow and
◙ When elapsed time passes 20 minutes; - short beeps sound &
Clamp closes. ALARM
● To end collection, press [END] key. flashing
6 Low Battery Alarm:
◙ When battery charge is depleted; - one 5-sec long beep once - LOW
(approximately at 10 % of full charge) turns on
7 Tray Off Alarm:
◙ When tray is taken off from pedestal; - 20 short beeps for 10 sec
● Operation resumes if tray is placed on
8 Wrong Key Alarm
◙ When POWERCAL key is pressed during - 3 short beeps
blood collection cycle;
4-5
4-4. Setting-up Parameters
4. T1 Time/T2 Time;
Press [TIMEBEEP] key during booting period of 10
seconds.
T1(10 min.) can be changed by [5 ml ▲] or [5 ml ▼] key (e.g.)
in 5 minute steps. 1 : 10
Press [START] key to save the changed value.
Press [TIMEBEEP] key, then T2 value is displayed.
T2(20 min.) can be changed by [5 ml ▲] or [5 ml ▼] key (e.g.)
in 5 minute steps. 2 : 20
Press [START] key to save the changed value.
Press [POWERCAL] key to get out of the mode.
At this time the device is powered off.
NOTE:
To disable the T1 and T2 Time function, set up the value to ‘0’.
4-6
5. Motor Speed;
Press [POWERCAL] key and [START] key together during
booting period of 10 seconds. (e.g.)
Motor speed list will be displayed. 45-40-30-20-15
Select a value by [5 ml ▲] or [5 ml ▼] key.
Press [POWERCAL] key to save the new value.
At this time the device is powered off.
6. Conversion Ratio;
Press [POWERCAL] key and [END] key together during
booting period of 10 seconds. (e.g.)
Conversion ratio list will be displayed.
Select a value by [5 ml ▲] or [5 ml ▼] key. 1.051-1.052-1.053-1.054-
Press [POWERCAL] key to save the new value. 1.055-1.056-1.057-1.058-
At this time the device is powered off. 1.059-1.060-1.061
7. Sound ON/OFF;
Press [TIMEBEEP] key for 3 seconds during normal operation.
When muted, symbol appears on the display panel.
To turn on, press the key and hold for 3 seconds again.
It is recommended that the weight calibration procedure be performed prior to the first donor
collection of the day. It is also recommended when collecting on a mobile vehicle, to repeat the
weight calibration procedure each time the vehicle is moved to a new location.
CAUTION
Keep the calibration weight from moving.
Don't touch any part of the unit.
4-7
If ‘500’ and ‘CAL’are flashing, remove the calibration
weight from the tray.
When ‘0’ gram and ‘rEP’ are flashing in turn, press the
[START] key.
It goes to the step (2) above.
CAUTION
Before placing the standard weight, wait for 2 seconds.
If "E r r" is displayed, remove the standard weight from
the tray and restart the step (2) above.
NOTE: Even when LOW symbol is initially displayed, the battery pack will run the
unit for approximately 2 hours.
B. Recharging Battery:
The battery pack can be recharged in the mixer unit.
(1) Plug the AC power adapter into the jack on the rear side of the main unit.
(2) Connect the power cord plug to a wall outlet.
(3) Leave it for hours. It may take up to 4 hours to get full charged.
* When the adapter is connected, battery is under recharging even if device power
is turned off.
4-8
Recharging a battery for more than 4 hours will not be harmful to the battery.
NOTE 1
4-9
5. Checks and Adjustments
This Chapter 5 describes how to check, adjust and calibrate several important parameters
of the equipment.
All procedures described in this chapter can be performed without opening the upper
cover. More professional fault isolation and troubleshooting procedures that should be
done with upper cover open are described in Chapter 6.
5- 1
5-1. Test Mode Subroutines
For service personnel to check major key components quickly without using any tool,
some software subroutine programs are prepared in the equipment.
One can check , button key failures, the load cell and its circuit performance with these
special test mode subroutines.
If any failure or malfunction is detected at these test mode procedures, fault isolation
and troubleshooting shall be done in accordance with the procedures in Chapter 6.
(3) To finish the test mode, press [POWERCAL] key, and then it ends as power turns
off.
(3) To finish the test mode, press [POWERCAL] key, and then it ends as power turns
off.
5- 2
C. Weight Adjustment
Designed tolerance of weight measurement is +/- 1 gram. For example, if
a standard weight of 500 grams is put on the tray, the display should show a figure
between ‘499’ and ‘501’. Otherwise, the equipment may be subject to calibration
by users.
5- 3
5.3. Weight Calibration
It is recommended that the weight calibration procedure be performed prior to the first donor
collection of the day. It is also recommended when collecting on a mobile vehicle, to repeat the
weight calibration procedure each time the vehicle is moved to a new location.
CAUTION
Keep the calibration weight from moving.
Don't touch any part of the unit.
CAUTION
Before placing the standard weight, wait for 2 seconds.
If "E r r" is displayed, remove the standard weight from
the tray and restart the step (2) above.
5- 4
6. Troubleshooting
6-1. Precautions
(1) Power should be turned off when subassemblies are plugged in or taken out.
(2) Use proper test equipment and tools.
(3) Attention should be taken when checking AC mains voltage.
(4) Excessive force should not be applied on the tray and load cell structures.
(5) When assembling upper cover case after troubleshooting, be sure that wiring
harnesses would not touch load cell, motor or any moving parts.
Wiring harnesses should not be too tight.
6-3. Preparation
Before carrying out troubleshooting procedures in Section 6-4 to locate a faulty assembly
or components, follow the procedures below first.
6- 1
6-4. Troubleshooting Procedures
The troubleshooting flow charts in the [Figure 6-1] will guide service personnel to a
faulty assembly or components that should be replaced or corrected.
When a faulty assembly is isolated, it should be replaced with a new or good assembly.
In case that service personnel have good knowledge in electronics, proper components
and test equipment, the personnel can troubleshoot to isolate a single faulty component.
Technical information is prepared in the Chapters 7, 8 and 9.
6- 2
[Figure 6-1] Troubleshooting Flow Charts (Page 1 of 6)
6- 3
[Figure 6-1] Troubleshooting Flow Charts (Page 2 of 6)
6- 4
[Figure 6-1] Troubleshooting Flow Charts (Page 3 of 6)
6- 5
[Figure 6-1] Troubleshooting Flow Charts (Page 4 of 6)
6- 6
[Figure 6-1] Troubleshooting Flow Charts (Page 5 of 6)
6- 7
[Figure 6-1] Troubleshooting Flow Charts (Page 6 of 6)
6- 8
7. Schematic Diagrams
7-1
[Figure 7-1] System Wiring Diagram
7-2
8. Technical Illustrations
8-1
[Figure 8-1] Exploded View of CM745
8-2
Top View
Bottom View
8-3
Top View
Bottom View
8-4
Top View
Bottom View
Top View
Bottom View
8-5
Top View
Bottom View
8-6
9. Parts List
Each component shall be identified with manufacturer's part number in the Parts List.
Location of a part and component can be found by a Find Number in the [Figure 8-1] Exploded
View or in the printed circuit board layouts in the [Figure 8-2] ~ [Figure 8-7].
Prefix 'A' in a find number denotes that the part is not a single component but an assembly, and
a lower-level parts list of which is presented at another place.
An order for service spare parts to distributors or to manufacturer should be accompanied with
its part number or with a designation number.
9-1
BLOOD COLLECTION MIXER CM745
FIND NO. DECSCRIPTION PART NO. Q’TY
1 TRAY 300-247 1
2 SCREW MCC04-08S 2
3 LABEL TRAY MOUNT 502-024 1
4 TRAY MOUNT 401-190 1
5 HANDLE 401-466 1
6 TOP CASE ASS’Y 400-076 1
7 BKT MOUNT 401-469 4
8 SCREW TRL30-06I 14
9 SPRING HANDLE 402-028 2
10 SCREW MPC03A06I 2
11 SCREW MPC03-16I 1
12 WASHER PL WPL03-00I 3
13 WASHER NYLON WPL03-00NY 1
14 HALL SENSOR BOARD ASS’Y 600-004 1
15 CAM FLANGE ASS’Y 400-024-1 1
16 SCREW MPC03S12I 2
17 SCREW MPC04-10K 2
18 WASHER SPRING WSB04-00K 2
19 MOVING BLOCK ASS’Y 300-051-1 1
20 SPACER SENSOR 302-010 1
21 MOTOR MOUNT B 301-033 1
22 MOVING MOTOR 102-008-5 1
23 MOTOR MOUNT PLATE 301-109 1
24 LOADCELL SPACER 401-434 1
25 LOADCELL 102-100-3 1
26 LOADCELL MOUNT BLOCK 301-086-1 1
27 WASHER SPRING WSB04-00S 4
28 SCREW MPC04-22I 4
29 SCREW MPC03S08I 2
30 LOADCELL STOPER 401-330-1 1
31 SCREW MCC06-16S 1
32 SCREW MCC04-10S 2
33 SCREW MPC03A16I 2
34 SCREW TRL03-08I 7
35 POWER BOARD ASS’Y 600-088 1
36 BKT PCB 401-467 1
37 CPU ASS’Y 600-150-1 1
38 SCREW MPC03S06I 4
39 BATTERY DOOR 201-102 1
40 BOTTOM CASE ASS’Y 400-078 1
41 SCREW MPC03A08I 4
42 PLATE BATTERY ASS’Y 400-098 1
43 BATTERY, LH025-H7C, 250mA, 8.4V 102-105 1
44 PLUG COVER 16.5 101-080 1
45 BATTERY CONTACT BOARD ASS’Y 600-086 1
46 SCREW MPC04S10I 4
47 BKT CLAMP 201-101 1
9-2
BLOOD COLLECTION MIXER CM745
FIND NO. DECSCRIPTION PART NO. Q’TY
48 SCREW MCC03-06I 2
49 TUBE CLAMP MODULE ASS’Y 300-038-6 1
50 SCREW MCC03-30I 3
51 COVER OUTER WHITE 201-034-3 1
52 SCREW MPC03S08I 4
53 SERVO MOTOR 102-001-1 1
54 MOTOR MOUNT BRACKET 401-195-1 1
55 GUIDE PLATE 401-196-1 1
56 COVER INNER WHITE 201-035C-2 1
57 SCREW TCQ30-20S 4
58 ARM 301-071 1
59 RUBBER FOOT 204-018 1
60 SCREW TRL26-08I 1
61 CAM FLANGE 201-036-1 1
62 SCREW TCQ30-10S 1
9-3