Performance Access Dual
Performance Access Dual
459800220281
Revision A
This document and the information contained in it is proprietary and confidential information of Philips Healthcare ("Philips") and may not be reproduced, copied in
whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department. Use of this document
and the information contained in it is strictly reserved for current Philips personnel and Philips customers who have a current and valid license from Philips for use by
the customer’s designated in-house service employee on equipment located at the customer’s designated site. Use of this document by unauthorized persons is strictly
prohibited. Report violation of these requirements to the Philips Legal Department. This document must be returned to Philips when the user is no longer licensed
and in any event upon ’Philips first written request.
Warranty Disclaimer
PHILIPS PROVIDES THIS DOCUMENT WITHOUT WARRANTY OF ANY KIND, IMPLIED OR EXPRESSED,
INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE.
Limitation of Liability
PHILIPS HAS TAKEN CARE TO ENSURE THE ACCURACY OF THIS DOCUMENT. HOWEVER, PHILIPS
ASSUMES NO LIABILITY FOR ERRORS OR OMISSIONS AND RESERVES THE RIGHT TO MAKE CHANGES
WITHOUT FURTHER NOTICE TO ANY PRODUCTS HEREIN TO IMPROVE RELIABILITY, FUNCTION, OR
DESIGN. PHILIPS MAY MAKE IMPROVEMENTS OR CHANGES IN THE PRODUCT(S) OR PROGRAM(S)
DESCRIBED IN THIS DOCUMENT AT ANY TIME.
Password Notice
THE PASSWORD IS THE PROPERTY OF PHILIPS HEALTHCARE AND IS PROVIDED FOR THE EXCLUSIVE
PURPOSE OF PROVIDING ACCESS TO SELECTED SERVICE UTILITIES WHICH ARE DESCRIBED IN THIS
SERVICE MANUAL AS BEING ASSOCIATED WITH THE PASSWORD. USE OF THIS PASSWORD FOR ANY
PURPOSE OTHER THAN FOR THE ACCESS TO THE SELECTED SERVICES UTILITIES IS STRICTLY
PROHIBITED.
Symbol Descriptions
Electrical-grounding Instructions
The equipment must be grounded to an earth ground by a separate conductor. The neutral side of the line is not to be considered the earth ground. On equipment
provided with a line cord, the equipment must be connected to properly grounded, three-pin receptacle. Do not use a three-to-two pin adapter.
Contents
Access Dual Performance Manual ................................................................................................... 1
Introduction ............................................................................................................................................................ 1
Scope................................................................................................................................................................ 1
Overview ................................................................................................................................................................ 1
Equipment Required ......................................................................................................................................... 2
Performance Test Criteria Matrix...................................................................................................................... 2
Safety ..................................................................................................................................................................... 3
Prerequisites for Performance Testing................................................................................................................... 4
Automatic Tests ..................................................................................................................................................... 5
Position the System Test Phantom................................................................................................................... 5
Start the Automatic Test ................................................................................................................................... 6
Automatic Performance Test ............................................................................................................................ 7
Before Beginning the Automatic Performance Test .................................................................................... 7
Automatic Acceptance Test ............................................................................................................................ 12
Automatic Constancy Test.............................................................................................................................. 15
Automatic QA Test.......................................................................................................................................... 16
459800220281 CSIP Level 1 © 2013 Koninlijke Philips Electronics N.V All Rights Reserved Refer to Front Cover i
Attention: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material, without permission, is a violation of law.
Access Dual Performance Test Instructions Revision A Access Dual Performance
Introduction
This document provides system performance tests, test methods, and performance specifications for the Access Dual
System. Personnel using this document and performing these tests are expected to be familiar with CT system test
methods and the operation of the Access Dual System. A Access Dual System meeting these performance
specifications is a necessary but not sufficient condition for acceptable system performance and acceptable image
quality for system shipment to customers and subsequent clinical use.
Scope
This document applies to all completely assembled and fully calibrated Access Dual CT Systems.
Overview
The purpose of the Performance, Constancy, Acceptance and QA tests are as follows:
• Performance Test
Run this test after major component replacement such as, X-ray tube or DAS. This tests the general image quality
of the scanner.
• Acceptance Test
The Acceptance test is run in the field for the newly installed scanner or if major changes were made in the scanner
version and/or scanner hardware. The result obtained during the Acceptance test is the base line for the next
Constancy test.
• Constancy Test
The Constancy test is run during Planned Maintenance or component replacement. Constancy measures the
performance and stability of the system relative to the results received in the Acceptance test.
• QA Test
The QA test runs scans to compare the CT number Uniformity, and Noise of the scan with the allowed tolerance.
459800220281 CSIP Level 1 © 2013 Koninlijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 1
Attention: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material, without permission, is a violation of law.
Access Dual Performance Test Instructions Revision A Access Dual Performance
Equipment Required
System Test Phantom 455012001021
459800220281 CSIP Level 1 © 2013 Koninlijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 2
Attention: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material, without permission, is a violation of law.
Access Dual Performance Test Instructions Revision A Access Dual Performance
Safety
!
WARNING Collision Hazard! When mounting the phantom make sure that it is positioned
so that it will not come in contact with the gantry covers or the rotor.
!
CAUTION Collision hazard! Before carry out performance testing, lift the vertical moveable couch to
340cm to avoid collision with the Gantry.
!
CAUTION Collision hazard! When mounting the system phantom make sure that it is positioned so
that it will not come in contact with the gantry covers or the Rotor.
!
WARNING Close all Gantry covers before performing tests. Check that the temperature
regulation of the gantry and the DMS function properly. Wait for the
temperature to stabilize within ±2 ℃ .
!
CAUTION Ensure that the phantom is completely vertical (Tilt = 0) before performing any calibration
or performance tests; tilt or misalignment of the phantom may result in inaccurate
calibration and performance tests.
459800220281 CSIP Level 1 © 2013 Koninlijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 3
Attention: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material, without permission, is a violation of law.
Access Dual Performance Test Instructions Revision A Access Dual Performance
459800220281 CSIP Level 1 © 2013 Koninlijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 4
Attention: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material, without permission, is a violation of law.
Access Dual Performance Test Instructions Revision A Access Dual Performance
Automatic Tests
459800220281 CSIP Level 1 © 2013 Koninlijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 5
Attention: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material, without permission, is a violation of law.
Access Dual Performance Test Instructions Revision A Access Dual Performance
NOTE During performance tests, verify that the images are free of rings and other artifacts.
459800220281 CSIP Level 1 © 2013 Koninlijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 6
Attention: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material, without permission, is a violation of law.
Access Dual Performance Test Instructions Revision A Access Dual Performance
459800220281 CSIP Level 1 © 2013 Koninlijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 7
Attention: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material, without permission, is a violation of law.
Access Dual Performance Test Instructions Revision A Access Dual Performance
4. An Instructions dialog box opens (see Figure 5 on page 9). Follow the on-screen instructions to positon the System
Phantom, and once it is in position, click OK to continue.
a. Mount the System Phantom to the Phantom Support. The Phantom must be leveled.
b. Set the laser beam to the Water Layer of the Phantom.
c. Adjust the System Phantom in the center of the Gantry.
459800220281 CSIP Level 1 © 2013 Koninlijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 8
Attention: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material, without permission, is a violation of law.
Access Dual Performance Test Instructions Revision A Access Dual Performance
Water Layer
WARNING Make sure there is no one in scanner room, and the scanner room door is
closed.
7. Click OK to continue.
459800220281 CSIP Level 1 © 2013 Koninlijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 9
Attention: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material, without permission, is a violation of law.
Access Dual Performance Test Instructions Revision A Access Dual Performance
8. The application scans the System Phantom and calculates the position of the System Phantom.
If the position is not accepted, the application provides instructions to adjust the position of the Phantom.
9. When the System Phantom position is accepted, the system will automatically start the Performance Test scan (see
Figure 8 on page 10).
459800220281 CSIP Level 1 © 2013 Koninlijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 10
Attention: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material, without permission, is a violation of law.
Access Dual Performance Test Instructions Revision A Access Dual Performance
NOTE If any of the sub-items failed during the Auto-performance test, repeat the same sub-item or perform it
again manually.
459800220281 CSIP Level 1 © 2013 Koninlijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 11
Attention: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material, without permission, is a violation of law.
Access Dual Performance Test Instructions Revision A Access Dual Performance
4. An Instructions dialog box appears. Follow the on-screen instructions to position the System Phantom, and once it
is in position, click OK to continue.
5. In the following window, click Start to select the desired items, and click Next to continue the Acceptance test.
WARNING Make sure no one is in the scanner room, and the scanner room door is closed.
6. Click OK to continue.
459800220281 CSIP Level 1 © 2013 Koninlijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 12
Attention: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material, without permission, is a violation of law.
Access Dual Performance Test Instructions Revision A Access Dual Performance
7. The application scans the System Phantom and calculates the position of the System Phantom.
If the position is not accepted, the application will provide instructions for adjusting position of the Phantom.
8. When the System Phantom position is accepted, the system will automatically start the Acceptance Test scan.
The test automatically runs a sequence of scans and the scans and parameters are listed in Verify Item area.
9. When the image quality of each item is found to comply with the required tolerances its State is recorded in green
as Passed. When the image quality of an item fails to comply with any required tolerance its State is recorded in red
as Failed.
10. When the Auto-acceptance test has completed, an Auto-acceptance report is generated automatically and saved.
11. Click Report to view the test results (see Figure 11 on page 14).
NOTE If any of the sub-items failed during the auto-Acceptance test, repeat the same sub-item or perform it
again manually.
12. When all the items passed and the measurements are within the required ranges, a baseline is automatically
saved,
459800220281 CSIP Level 1 © 2013 Koninlijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 13
Attention: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material, without permission, is a violation of law.
Access Dual Performance Test Instructions Revision A Access Dual Performance
459800220281 CSIP Level 1 © 2013 Koninlijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 14
Attention: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material, without permission, is a violation of law.
Access Dual Performance Test Instructions Revision A Access Dual Performance
459800220281 CSIP Level 1 © 2013 Koninlijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 15
Attention: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material, without permission, is a violation of law.
Access Dual Performance Test Instructions Revision A Access Dual Performance
Automatic QA Test
Run the Automatic QA Test according to the following procedure:
1. In the Auto-performance dialog box, highlight QA (see Figure 3 on page 6), and click Next.
2. Follow Step 2 on page 12 to Step 9 on page 13 of the Automatic Acceptance Test on page 12.
3. When the QA Test has completed, a QA Test report is generated automatically and saved. Click Report to view the
test results (see Figure 12 on page 15).
NOTE If any of the sub-items failed during QA test, repeat the same sub-item or perform it again manually.
459800220281 CSIP Level 1 © 2013 Koninlijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 16
Attention: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material, without permission, is a violation of law.
Access Dual Performance Test Instructions Revision A System Test Phantom
The test phantom used for the majority of these tests is the System Performance Phantom. This phantom is supplied as
part of every Access Dual system. The phantom consists of a combined Head and Body Phantom. The Head Phantom
is a 200 mm diameter PVC shell filled with distilled water. The Head Phantom has three sections (see detailed
description of the system phantom in Figure 14 on page 18–Figure 17 on page 21.
1. The physics layer is used for impulse response and tomographic section thickness (slice width) measurements.
2. The water layer is used to measure uniformity and noise.
3. The Multi–Pin Layer is used for low contrast scale check.
4. The body layer is used to measure CT values for body agreements and noise.
The Body Phantom part of this phantom consists of a water hole (0 ±10 CT) and air (1000 ±10 CT).
459800220281 CSIP Level 1 © 2013 Koninlijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 17
Attention: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material, without permission, is a violation of law.
Access Dual Performance Test Instructions Revision A System Test Phantom
n
ectio
m m) S
ld (300
e Fie
Larg m
60 m
on
m) Secti
50 m
ld (2
m a ll Fie m m
S 116
A lu
mi n u
Sli ce Wi dt h ) L
ring
Cente
a ye
Pin
m
(25 m
r
OD)
Oblique View
Figure 14:
459800220281 CSIP Level 1 © 2013 Koninlijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 18
Attention: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material, without permission, is a violation of law.
Access Dual Performance Test Instructions Revision A System Test Phantom
Ramp
Locator Handle
Small Field (250 mm) Section
is filled with water
Aculon
Mounting
Clamp
Centering
Pin Teflon pin
Spatial Phantom
Point Spread Function Phantom
(0.18 mm copper wire) Perspex
pin
Multipin Layer
Water Layer
Ramp
Locator
Figure 15:
459800220281 CSIP Level 1 © 2013 Koninlijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 19
Attention: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material, without permission, is a violation of law.
Access Dual Performance Test Instructions Revision A System Test Phantom
Large Field (300 mm OD) Section Small Field (200 mm OD) Section
7 mm
2m
m m
6m
1 of 4 Ramp
Locators
m 5m
3m m
4 mm
Lexan Polyethelyene
(≈110 ± 15) (≈-60 ± 15)
Water
in Large Field Section
(0 ± 10)
Centering Teflon
pin in Large Field Section
(≈950 ± 50)
R Perspex Teflon
(≈130 ± 15) (≈950 ± 50)
ntal
rizo
o ho
5° t
Spatial Phantom,
at 4
Perspex (≈130 ± 15) with Lexan pins
strip
inum
alum
P
Note:
Aculon = nylon
Perspex = acrylic Figure 16:
459800220281 CSIP Level 1 © 2013 Koninlijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 20
Attention: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material, without permission, is a violation of law.
Access Dual Performance Test Instructions Revision A System Test Phantom
Handle
Water Fill
Mounting Clamp
Front View
Figure 17:
459800220281 CSIP Level 1 © 2013 Koninlijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 21
Attention: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material, without permission, is a violation of law.
Access Dual Performance Test Instructions Revision A System Test Phantom
Table 2:
459800220281 CSIP Level 1 © 2013 Koninlijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 22
Attention: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material, without permission, is a violation of law.
Access Dual Performance Test Instructions Revision A Phantom Holder
Overview
The following figure show the contents of the Access Dual phantom holder.
.
A
B
459800220281 CSIP Level 1 © 2013 Koninlijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 23
Attention: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material, without permission, is a violation of law.
Access Dual Performance Test Instructions Revision A Phantom Holder
5
1
2
7
8
4
Figure 19: Side View Figure 20: Front View
NO Name Note
1. Vertical Adjustment Micro Adjustment In Vertical
2. Tilt Adjustment Screw 1 X-Axis Rotation
3. Phantom Connection Part
4. Tilt Adjustment Screw 2 Y-Axis Rotation
5. Horizontal Adjustment Screw In Horizontal
6. Phantom Auxiliary Connection Part Used In Nonlinearity Calibration
7. Phantom Initial Mounting Position 0mm,-50mm,-105mm,-150mm,-180mm
8. Holder Fixing Screw
459800220281 CSIP Level 1 © 2013 Koninlijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 24
Attention: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material, without permission, is a violation of law.
Access Dual Performance Test Instructions Revision A Phantom Holder
459800220281 CSIP Level 1 © 2013 Koninlijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 25
Attention: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material, without permission, is a violation of law.