Specification: C 60

SHENZHEN COMEN MEDICAL INSTRUMENT CO.,LTD


No.2 of FIYATA Timepiece Building, Nanhuan Avenue, Gongming
sub-district, Guangming New District, Shenzhen, P.R. China
Tel: +86-755-26408879
Fax: +86-755-26431232
Email: infoatszcomen.com
Web: en.comen.com

2020 V 1.2
Patient Monitor
C 60

Standard Configuration: Battery indicator


3-lead ECG, MASIMO SpO2,NIBP, HR, Temp(Single Channel), QRS beep and alarm sound
Apnea awaking, Oxygen concentration Operating key sound
Key backlight (Optional)
Optional Configuration:
Trolley, Wall mount, Interface:
Ground wire. Dual-IBP or IBP(Single Channel) & EtCO2, RESP, Parameter cable interface
Safety Standards: AC power input socket
IEC 60601-1 IEC 60601-1-8 IEC 60601-2-27 EN 1060-3 IEC USB port
80601-2-30 IEC60601-2-34 IEC60601-2-49 ISO 80601-2-56 RJ45 port
ISO 80601-2-61 VGA output
Physical Characteristics: Data storage
Product Size: 242mm×218mm×121mm Alarm Event: 200 groups
Net Weight: 2.5kg Wave Review: 6 hours (8 waves)
Display: 8.4” color TFT touch screen NIBP Review: 2000 groups
Resolution: 800*600 Apnea awaking: 2000 groups
Trace: 6 waveforms Trend Graph: 160 hours
Ingress Protection IPX1 Trend Table: 160 hours
Operation Environment: Power-off storage: Yes
Working Alarm: User-adjustable High and Low 3-level
Temperature 5-40°C Limits;
Humidity: ≤93% Prioritized audible and visual alarm
Power Supply 100-240V~, 50/60Hz±1Hz Network: Connected to Central Monitoring System
Battery Type: Rechargeable Lithium-ion battery by hardwire/wireless
Battery Capacity: 2200mAh Respiration:
Battery Recharging Method: RA-LL Impedance Method
Time: Maximum 6 hours for charging; RR measurement
Battery backup: 2 hours for continuous working range 0-150rpm
Brightness: Manual from 10 to 100 Accuracy: 7~150rpm: ±2rpm or 2%, whichever is
Indicator: greater;
Two alarm indicators 0~6rpm: unspecified
Power indicator Resolution: ±1 rpm
RESP Apnea: 10s-40s Neonate mode 150mmHg
Alarm: Audible and visual alarm; alarm events tolerance ±3mmHg
reviewable Accuracy: ±3mmHg
Sweep Speed: 6.25, 12.5, 25mm/s Resolution: 1mmHg
Gain Selection: X0.25, X0.5, X1, X2, X4 Alarm: Systolic, Diastolic, Mean
ECG: PR from NIBP: 40bpm~240bpm
TM
Lead Type: selectable CardioTec 5-leads ECG Resolution: 1bpm
Analysis, 3-leads Accuracy: ±0.1kPa or±1mmHg (Both maximum)
Lead selection 5-lead: I; II; III; aVR; aVL;aVF; V Nellcor SpO2:
3-lead: I; II; III Measurement
Gain Selection X0.125 ,X0.25, X0.5, X1, X2, X4, Auto range 0-100%
Sweep Speed: 6.25,12.5, 25, 50mm/s Resolution: 1%
Heart Rate Range: 15-350bpm Accuracy: ±3% (70-100%, non-motion)
Resolution: 1 bpm 0-69% unspecified
Accuracy: ±1% or ±1bpm (whichever is greater) Alarm range: 0-100%
Protection: Withstand 4000VAC/50Hz voltage in PR Measurement
isolation; Range: 20-300bpm
Against electrosurgical interference and Resolution: 1bpm
defibrillation; Accuracy: ±3bpm (20-250bpm);
Bandwidth: MON Mode:0.5 Hz~40 Hz Unspecified (251-300bpm)
OPE Mode:1 Hz~20 Hz Alarm Range: 0~350bpm
Diagnostic mode: 0.05~150Hz; Masimo SpO2:
Arrhythmia Measurement &
analysis: Available alarm range 1~100%
Pacemaker Resolution: 1%
detection: Detectable Accuracy: ±3% (70-100%, non-motion)
NIBP: 1-69% unspecified
Method Automatic oscillometric PR Measurement
Work mode: Manual / Automatic Range: 25~240bpm
Measurement time: Adjustable(1-720min) Resolution 1bpm
Maximum Accuracy: ±3bpm(non-motion)
measurement time 85s ±5bpm(motion)
Measurement Unit: mmHg / kpa selectable Alarm Range: 20~350bpm
Measurement
types: Systolic, Diastolic, Mean PI value: 0.02~20%
Range of systolic Resolution: 0.01% (0.02%~9.99%)
pressure: 40-135mmHg 0.1% (10.0%~20.0%)
Range of diastolic Accuracy: unspecified
pressure: 10-100mmHg Comen SpO2:
Range of mean Measurement
pressure: 20-110mmHg & alarm range 0~100%
Over-pressure Resolution: 1%
protection: Both hardware and software Accuracy: ±3% (70-100%, non-motion)
over pressure protection Unspecified (0-69%)
PR measure Stimulation time: 5s (working 3s, stopping 2s)
Range: 20-250bpm Response time 0~20s
Resolution: 1bpm IBP
Accuracy: ±1bpm Channel: 2 Channels
Alarm Range: 0~350bpm Measured Pressure: ART, PA, CVP, RAP, LAP, ICP, LV, AO, UAP,
PI value: 0.02~20% BAP, FAP, UVP, IAP, P1, P2, P3, P4
Resolution: 0.01% (0.02%~9.99%) Measurement Unit: mmHg/ kPa selectable
0.1% (10.0%~20.0%) Measurement
Accuracy: unspecified Range: ART: 0~300mmHg
Temperature (Dual Channel) PA: -6~120 mmHg
Range: 0-50°C CVP: -10~40mmHg
TEMP sensor: Skin/rectal TEMP sensor RAP: -10~40mmHg
Resolution: 0.1°C LAP: -10~40mmHg
Accuracy: ±0.1°C (exclusive of error of sensor) ICP: -10~40mmHg
Channel: T1, T2, TD (Temperature Difference) LV: 0~300mmHg
EtCO2 (Sidestream) AO: 0~300mmHg
Unit: mmHg, kPa UAP: 0~300mmHg
Measurement BAP: 0~300mmHg
range: 0mmHg~150mmHg FAP: 0~300mmHg
Resolution 1mmHg or 0.1kPa or 0.1% UVP: -10~ 40mmHg
Accuracy 0mmHg ~40mmHg should be±2mmHg; IAP: -10~40mmHg
41mmHg ~70mmHg should P1, P2, P3, P4: -50~300mmHg
be±5%×reading; Accuracy: ±2% or ±1mmHg, whichever is greater
71mmHg ~100mmHg should Resolution: 0.1kPa or 1mmHg
be±8%×reading; (-50mmHg~300mmHg)
101mmHg~150mmHg should Alarm Range: -50mmHg~300mmHg
be±10%×reading PR from IBP: 25bpm~350bpm
Oxygen Resolution: 1bpm
compensation 0~100 mmHg Accuracy: ±1% or ±1bpm, whichever is greater
Oxygen concentration
Measurement & *Notice: Specifications subject to changes without prior
alarm range: 0%~100% notice. All rights reserved by Comen
Accuracy: ±1%
Response time <15s (21% Air to 100% Oxygen)
Apnea awaking
Stimulus mode: Asphyxiator vibration