AS 2853 1986: Enclosure-Temperature-controlled-performance Testing and Grading
AS 2853 1986: Enclosure-Temperature-controlled-performance Testing and Grading
com
1. Introduction
This manual is made as a guide for the harmonization of oven calibration
procedures carried out by calibration laboratories accredited by KAN and
other laboratories that require it.
This guide refers primarily to documents AS 2853 1986: Enclosure-
Temperature-controlled-performance testing and grading, with some
simplifications, and other relevant documents.
Oven calibration is carried out to determine the uniformity of the temperature in the
oven and as a reference to determine the suitability of the oven for certain purposes.
2. Scope
What is meant by an oven is an engineering facility that allows the
realization of selectively determined values of air temperature in a closed
volume within a working range.
This guide applies to the determination of the temperature characteristics of ovens used in
The working area is the part of the oven whose temperature characteristics will be
determined. For most ovens, the workspace is the entire interior space of the oven.
3. Reference documents
4. Calibration principle
4.1. Temperature measurements are carried out at certain measuring point locations in the
oven working space for 1 hour or 5 cycles after a stable temperature is reached. The total
variation is calculated from the maximum and minimum temperature values of all
measuring point locations. The total variation is an indicator of the performance of the
calibrated oven.
4.2. The determination of the number of measuring point locations and the number of temperature sensors
4.3.During the calibration process takes place the ambient temperature should not change more than 10
oC. The power supply should not be changed by more than 10% or as
required by the oven manufacturer.
4.4. If the temperature reading and/or recording is done manually, the reading should be taken
so that short-term temperature variations (due to the influence of the temperature control
5. Temperature Sensor
5.1. The temperature sensor used must be calibrated within the appropriate range for the
usage range. The selected temperature sensor has an uncertainty of not more than 1/3
5.2. The resolution of the temperature sensor indicator used should not be more than 1/10
5.3. The temperature sensor must have a faster response time than the temporal variation
of the oven.
oven wall.
6. Calibration procedure
6.1. Determination of the number of measuring point locations and the number of temperature sensors
used. To calculate the number of measuring point locations and the required number of temperature
6.1.1. Determination of the nominal calibration temperature and its temperature tolerance, t.
The temperature tolerance value can be obtained in one of the following ways. 6.1.1.1.From the
6.1.1.2. If there is no information from the manual, tolerances can be requested from the customer.
In this case the temperature tolerance is defined as the desired tolerance value, which is not
necessarily achieved, or
6.1.1.3. The temperature tolerance is 1.5x the uncertainty of the temperature sensor used.
6.1.2. Determine the value of the theoretical total variation,Ro, with Equation 1
Ro = 2 t (1)
1 11.3 15 23 45 120
2 8.0 10 16 32 88
3 5.66 7.4 11 23 62
4 4.00 5.2 8.0 16 44
5 2.83 3.7 5.7 11 31
6 2.00 2.6 4.0 8.0 22
7 1.41 1.8 2.8 5.7 16
8 1.00 1.3 2.0 4.0 11
9 0.707 0.92 1.4 2.8 7.8
10 0.500 0.65 1.0 2.0 5.5
11 0.354 0.46 0.71 1.4 3.9
12 0.250 0.33 0.50 1.0 2.8
13 0.177 0.23 0.35 0.71 2.0
14 0.125 0.16 0.25 0.50 1.4
6.1.5. Determine the number of measuring point locations and the required number of
Where
N = number of measuring point
The temperature sensor is placed at the location of the measuring point according to
the sequence in Figure 1 for a rectangular oven and Figure 2 for a cylindrical oven.
attached to the wall of the oven. A tool holder can be used to help place the temperature
10 17E 15F
6A 3A
19E
8F
4F
13A
20E
9A
1C 12F
11F
14A
18E
2A 5A
16E
7F
Position code:
C = Center
F = Center of face
A = Corner
Figure 1. Location of the temperature sensor placement measuring point for a rectangular oven
Diameter
7 1 12
2 11 20 6 9 13 4 10 5 19 8 3
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 (part)
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
180 315 225 90 0 135 225 90 0 270 45 315 180 270 45 135 0 (level)
Figure 2. Location of the measuring point for placing the temperature sensor for a cylindrical oven
The oven temperature is set according to the nominal calibration temperature by using the buttons
provided on the oven. If there is no temperature indicator in the oven, the temperature setting button
Table 2. Example of measurement data recording table for 9 measuring point locations
1
2
3
4
5
6
7
8
9
Indicator
Room temperature
6.5. If calibration is carried out for more than one nominal temperature value,
then steps 6.3 to 6.4 are carried out again for the next nominal temperature
value.
on the difference between the maximum and minimum temperatures and the mean
value for each temperature sensor. Next, is calculated to determine other parameters
as follows:
The difference between the indicator temperature (7.1) and the measured oven temperature (7.2) must be less
than Ro.
7.4. Table 3 can be used to summarize the calculation results with the number of rows depending on the
model as follows
'
K = t 'Correct tindicator
with:
K= correction that must be given to the oven indicator indication to obtain the
The uncertainty values and coverage factors are obtained from the calibration certificate
of the equipment.
U95
ub=
1
k
b. Drift temperature sensor and indicator
The drift uncertainty of the temperature sensor and indicator is obtained from the old value (n
long) and new (nnew) on the calibration certificate assuming the distribution is rectangular.
1 2(nold n )new
ub=
2
3
c. The measured temperature variation consisting of
1 2 NST
ub5 = for analog indicators.
6
e. Variation of the indication of the oven temperature indicator
The standard uncertainty of the indicator variation is obtained from the difference between the
largest and smallest values in the instrument readings from each measuring point assuming the
distribution is rectangular.
f. Combined uncertainty
U95 = uc .k
9. Determination of CMC
For the purpose of determining the CMC of a calibration laboratory, the uncertainty originating from
the spatial variation and the empirical variation of the oven is eliminated so that it becomes as
follows.
10. Recalibration
The calibration interval depends on the use of the oven, but recalibration is carried out once
every 5 years at most. Recalibration must also be carried out if repairs or modifications have
The 10 mg weight shall meet the performance check criteria of the mass value i.e. 0.1% of actual mass value. E.g. For all the 10 measurements of
10 mg weight, variation in the weighing cannot exceed 0.01mg.
Performance check
After the Auto calibration put 1 mg, 2 mg, 5 mg, 10 mg and 20 mg weights individually. The measurement shall be within the 0.1% of the actual mass
value of the individual weight as given in the performance check log. (Attachment no. 02)
Measurement uncertainty check
The measurement of uncertainty shall be carried out by using 10 mg weight.
Put the external weight of 10 mg on the pan and note the 10 measurements. (Attachment no. 03)
Calculate the measurement of uncertainty as follows.
Measurement Uncertainity = Standard Deviation X 2
Actual Mass Value
Measurement of uncertainty shall be Not More Than 0.10.
Calibration frequency
It shall be performed daily, after any maintenance, after the relocation and after the power failure of the balance.
Each manufacturer supplies the maximum loading condition of the balance. The calibration of the balance shall be designed in such a way that the
performance check weights cover the entire loading range of the balance. Lower and higher load limits shall be checked for the performance check.
Before calibration make sure that the level bubble is in the center of the indicator and if internal calibration facility is available in the balance then it
shall also be performed daily before performance check of the balance.
Always use a check weight appropriate for the particular balance. E.g. a 100 – mg check weight might be selected for a balance that has a load limit
of 150 mg. Thus the operator must know the maximum capacity of the balance to select the correct check weight.
https://www.pharmaguideline.com/2011/01/analytical-balance-calibration.html 1/4
20/03/2023, 11:50 Analytical Balance Calibration (Updated) : Pharmaguideline
https://www.pharmaguideline.com/2011/01/analytical-balance-calibration.html 2/4
20/03/2023, 11:50 Analytical Balance Calibration (Updated) : Pharmaguideline
Date Std. Wt. Std. Wt. Observation of Uncertainty Weight Done Checked Remark
Cal. No. Less than or equal box no. by by
to 0.001 and
valid
upto
10 mg
Calculation
Measurement = Standard Deviation X 2
uncertainty Actual Mass Value
Ankur Choudhary is India's first professional pharmaceutical blogger, author and founder of pharmaguideline.com, a widely-read pharmaceutical
blog since 2008. Sign-up for the free email updates for your daily dose of pharmaceutical tips.
Need Help: Ask Question
https://www.pharmaguideline.com/2011/01/analytical-balance-calibration.html 3/4
20/03/2023, 11:50 Analytical Balance Calibration (Updated) : Pharmaguideline
https://www.pharmaguideline.com/2011/01/analytical-balance-calibration.html 4/4
20/03/2023, 12:16 Calibration of Centrifuge Apparatus : Pharmaguideline
-->
Calibration of time
Set the time for five minutes using set parameters. Record the results by using a digital stopwatch. Repeat the test for ten minutes.
Set the RPM of rotor to 12000. Record the results by using tachometers.
Set RPM of rotor to 6000. Record the results by using tachometers
Swing out rotor
Set the RPM of rotor to 4000. Record the results by using tachometers.
Set RPM of rotor to 2000. Record the results by using tachometers.
Calibration of temperature
Fixed angle rotor
Fill the two sample tubes with ethylene glycol and place it in the sample holder. Set the parameters like temperature 5°C, RPM: 12000
and time 45 minutes.
Record the temperature of the sample solution by using digital thermometer.
Fill the two sample tubes with ethylene glycol and place it in the sample holder. Set the parameters like temperature 10°C, RPM: 12000
and time 45 minutes. Record the temperature of the sample solution by using digital thermometer
Fill the two sample tubes with ethylene glycol and place it in the sample holder. Set the parameters like temperature 20°C, RPM: 12000
Site
and time 45 minutes. Record the temperature of the sample solution by using digital thermometer Malicious
Rate
0.0
Swing out rotor
https://www.pharmaguideline.com/2011/02/calibration-of-eppendorf-centrifuge.html 1/5
20/03/2023, 12:16 Calibration of Centrifuge Apparatus : Pharmaguideline
Fill the two sample tubes with ethylene glycol and place it in the sample holder. Set the parameters like temperature 5°C, RPM: 4000 and
time 45 minutes.
Record the temperature of the sample solution by using digital thermometer.
Fill the two sample tubes with ethylene glycol and place it in the sample holder. Set the parameters like temperature 10°C, RPM: 4000
and time 45 minutes. Record the temperature of the sample solution by using digital thermometer
Fill the two sample tubes with ethylene glycol and place it in the sample holder. Set the parameters like temperature 20°C, RPM: 4000
and time 45 minutes. Record the temperature of the sample solution by using digital thermometer
Frequency
Perform the calibration after every six months. Record the results in the calibration record format.
Maintenance and cleaning
Only neutral agents may be used for cleaning and disinfection (70% isopropanol/ water mixture or an alcohol -based disinfectant)
After cleaning with detergent, the rubber seals in the rotor chamber should be rinsed well in water and lubricated with glycerin in order to
prevent the seals from becoming brittle.
If condensation water forms in the rotor chamber after freezing occurs, dry with a soft absorbent cloth.
The rotor and bucket must be cleaned regularly to prevent contamination caused by residue.
When instrument does not comply with the specified limits the instrument should be labeled as “out of calibration” and should get repaired
/serviced.
After servicing /repairing the instrument must be calibrated as per above-mentioned procedure.
https://www.pharmaguideline.com/2011/02/calibration-of-eppendorf-centrifuge.html 2/5
20/03/2023, 12:16 Calibration of Centrifuge Apparatus : Pharmaguideline
Sr. No. RPM Set RPM on Display RPM on Tachometer Tolerance RPM
1 4000 ±100
2 2000 ±100
B] Temperature Measurement
Sr. No. Temp Set (°C) Temp on Display (°C) Temp on Thermometer (°C) Tolerance (°C)
1 5 ±2
2 10 ±2
C] Time Measurement :
Sr. No. Time Set (Min.) Time on Display (Min.) Time on Stop watch (Min.) Tolerance (Min.)
1 5 ±2
2 10 ±2
https://www.pharmaguideline.com/2011/02/calibration-of-eppendorf-centrifuge.html 3/5
20/03/2023, 12:16 Calibration of Centrifuge Apparatus : Pharmaguideline
Make : Model :
https://www.pharmaguideline.com/2011/02/calibration-of-eppendorf-centrifuge.html 4/5
20/03/2023, 12:16 Calibration of Centrifuge Apparatus : Pharmaguideline
Performed by : Checked by :
Date : Date :
Ankur Choudhary is India's first professional pharmaceutical blogger, author and founder of pharmaguideline.com, a widely-read pharmaceutical
blog since 2008. Sign-up for the free email updates for your daily dose of pharmaceutical tips.
Need Help: Ask Question
Site
Malicious
Rate
0.0
https://www.pharmaguideline.com/2011/02/calibration-of-eppendorf-centrifuge.html 5/5
20/03/2023, 12:16 Calibration of Glassware (Class A and Class B) : Pharmaguideline
-->
Tolerance limit
Volumetric flasks
Capacity in ml 5 10 25 50 100 200 250 500 1000
Tolerance Class A 0.02 0.02 0.03 0.04 0.06 0.08 0.1 0.15 0.2
(±ml) Class B 0.04 0.04 0.06 0.08 0.15 0.2 0.2 0.3 0.8
One mark pipette
Capacity in ml 1 2 5 10 15 20 25 50 100
Tolerance Class A Strategic
0.006 0.006 Consulting
0.01 Solutions
0.02 0.02- Full 0.03 0.03 0.04 0.06
(±ml) Class B 0.02 0.02 0.03 0.04 0.04 0.05 0.06 OPEN 0.08 0.12
Burettes
Phase Support Strategy
US & EU regulatory consulting experts propharmagroup.com
https://www.pharmaguideline.com/2011/02/calibration-of-glassware.html 1/2
20/03/2023, 12:16 Calibration of Glassware (Class A and Class B) : Pharmaguideline
Capacity in ml 10 25 50 100
Subdivision in ml 0.05 0.05 0.1 0.1
Tolerance Class A 0.01 0.03 0.05 0.1
(±ml) Class B 0.02 0.06 0.1 0.2
Frequency : Initially when volumetric glassware is purchased.
Note: When the volumetric glassware does not comply with the requirement as specified above, the volumetric glassware should be returned back to
the party.
REFERENCE
Indian Pharmacopoeia.
United States Pharmacopeia.
Ankur Choudhary is India's first professional pharmaceutical blogger, author and founder of pharmaguideline.com, a widely-read pharmaceutical
blog since 2008. Sign-up for the free email updates for your daily dose of pharmaceutical tips.
Need Help: Ask Question
https://www.pharmaguideline.com/2011/02/calibration-of-glassware.html 2/2
20/03/2023, 12:17 Calibration of Heating Block : Pharmaguideline
-->
5. Set the heating block for five different temperature points firstly 30°C second 37°C third 50°C fourth 70°C and fifth 90°C.
6. After attaining the heating block temperature, noting down the temperature for 1 hour at the interval of 10 minutes for each set temperature points.
7. Compile the data and find out the temperature uniformity throughout the 1 hour period. It should be ± 1°C.
8. After calibration, immediately put the “CALIBRATION” tag duly filled with necessary information.
Ankur Choudhary is India's first professional pharmaceutical blogger, author and founder of pharmaguideline.com, a widely-read pharmaceutical
blog since 2008. Sign-up for the free email updates for your daily dose of pharmaceutical tips.
Need Help: Ask Question
Site
Malicious
Rate
0.0
https://www.pharmaguideline.com/2011/01/calibration-of-heating-block.html 1/1
20/03/2023, 12:18 Calibration of Micropipettes : Pharmaguideline
-->
Calibration of Micropipettes
Learn how to calibrate the Micropipettes using distilled water.
1.0 Equipment Required
Micropipettes
Micro tips
Calibrated balance
2.0 Material Required
Distilled water
3.0 Procedure
3.1 Calibrate the micropipettes in a vibration-free room at a constant temperature of water between 20 to 25°C.
3.2 Calibrate the Micropipettes for a volume of 100 and 1000 ml, and for 20 and 200 ml.
3.3 Set the desired volume using the volume-setting knob.
3.4 Carefully fit tip into tip cone and flush with distilled water by pipetting the adjusted volume at least 3 – 4 times.
3.5 Carefully aspirate the liquid and pipette the sucked volume of distilled water into a tare 10 ml volumetric flask.
3.6 Repeat at least ten times and record the result.
3.7 Calculate the pipetted volume, taking the temperature into account.
3.8 After calibration, immediately put the “CALIBRATION” tag duly filled with necessary information
4.0 Precaution
4.1 Use an analytical balance with a readability of 0.1 mg.
5.0 Frequencies
Quarterly
6.0 Limits/ Acceptance Criteria
Micropipettes ID Volume in ml Tolerance Limit in ml
Min Max
A 100 99.9 100.1
1000 999.0 1001.0
Site
B 20 19.98 20.02 Malicious
Rate
200 199.8 200.2 0.0
7.0 Abbreviation
SOP: Standard Operating Procedure
https://www.pharmaguideline.com/2011/01/calibration-of-micropipettes.html 1/2
20/03/2023, 12:18 Calibration of Micropipettes : Pharmaguideline
Ankur Choudhary is India's first professional pharmaceutical blogger, author and founder of pharmaguideline.com, a widely-read pharmaceutical
blog since 2008. Sign-up for the free email updates for your daily dose of pharmaceutical tips.
Need Help: Ask Question
Site
Malicious
Rate
0.0
https://www.pharmaguideline.com/2011/01/calibration-of-micropipettes.html 2/2
20/03/2023, 12:16 Calibration of Vernier Caliper : Pharmaguideline
-->
Repeat the above procedure for the remaining standard lengths (50 mm, 75 mm, 100 mm and 125mm).
Tolerance: + 0.05 mm
Frequency: Once in a month
Maintenance & repairs:
When the instrument does not comply with the requirement specified above; the instrument should be labeled as “Out of Calibration” and should get
repaired/ serviced.
After repairing/ servicing the instrument before taking for use, the instrument must be calibrated as per the above-mentioned procedure.
1 25 24.95 25.05
2 50 49.95 50.05
3 75 74.95 75.05
4 100 99.95 100.05
5 125 124.95 125.05
https://www.pharmaguideline.com/2011/02/calibration-of-vernier-caliper.html 1/2
20/03/2023, 12:16 Calibration of Vernier Caliper : Pharmaguideline
Ankur Choudhary is India's first professional pharmaceutical blogger, author and founder of pharmaguideline.com, a widely-read pharmaceutical
blog since 2008. Sign-up for the free email updates for your daily dose of pharmaceutical tips.
Need Help: Ask Question
https://www.pharmaguideline.com/2011/02/calibration-of-vernier-caliper.html 2/2
20/03/2023, 12:17 Calibration of Viscometer : Pharmaguideline
-->
Calibration of Viscometer
Learn how to calibrate the Viscometer using different viscosity standards.
Calibration Procedure
1. Ensure that instrument is ready for use.
2. Calibrate the instrument by using different calibration standards. (i.e. 50, 500, 1000 and 5000 cps, Use “BROOKFIELD CALIBRATION FLUID”)
3. Select the spindle as per requirement, by pressing the “SELECT SPINDLE” key, on the display “S” line will begin to blink for four seconds.
4. Press “UP or DOWN” (↑↓) arrow keys while blinking, two digit character values will change accordingly. (UP for increase in spindle no. and DOWN
for Decrease in spindle no.) Release the arrow key to halt further scrolling and re-press the “SELECT SPINDLE” key.
5. Select the speed as per requirement, by pressing the “SET SPEED” key, on the display “0.0” line will begin to blink for four seconds.
6. Select the rpm speed by pressing the “UP or DOWN” (↑↓) arrow key, the current rpm speed will display on the bottom line of the screen and RPM
will blink. [UP for increase the RPM speed & DOWN for Decrease the RPM speed.] Re-press the “SET SPEED” key.
7. Attach the selected spindle to the lower shaft, lift the shaft slightly, holding it firmly with one hand while screwing the spindle on with the other hand.
8. Put the test solution beaker below the spindle center and insert the spindle until the fluid level is at immersion groove in the spindle shaft middle of
the “SPINDLE MARK LINE”. (Ensure that no air bubble is trapped to the surface of the spindle).
9. Press the “MOTOR ON/ OFF” key to run the test.
10. Allow time to stabiles the reading, (The time required for the stabilization is depend on the speed of running & characteristics of the sample fluid.)
11. Record the stabilized reading in CPS unit in calibration log.
12. Press the “MOTOR ON/ OFF” key to stop the test and put ‘off’ the instrument mains.
13. Clean the spindle & instrument after use.
14. Affix the calibration label on the instrument.
Related: Determination of Viscosity by Using Ostwald-type, Suspended-level and Rotating Viscometer
CALIBRATION LOG
Instrument:____________ Code No:______________
Make :_____________ Model: _______________
https://www.pharmaguideline.com/2011/01/viscometer-calibration.html 1/2
20/03/2023, 12:17 Calibration of Viscometer : Pharmaguideline
2 100 cps
3 500 cps
4 1000 cps
5 5000 cps
Ankur Choudhary is India's first professional pharmaceutical blogger, author and founder of pharmaguideline.com, a widely-read pharmaceutical
blog since 2008. Sign-up for the free email updates for your daily dose of pharmaceutical tips.
Need Help: Ask Question
https://www.pharmaguideline.com/2011/01/viscometer-calibration.html 2/2
20/03/2023, 12:10 Difference among Calibration, Validation & Qualification : Pharmaguideline
-->
Calibration:
The set of operations that establish, under specified conditions, the relationship between values indicated by an instrument or system for measuring
(for example, weight, temperature and pH), recording, and controlling, or the values represented by a material measure, and the corresponding is
known values of a reference standard.
Limits for acceptance of the results of measuring should be established. Always remember any reference standard is always used in calibration.
Validation:
The action of proving and documenting that any process, procedure or method actually and consistently leads to the expected results.
It can better understand that the validation is a documented evidence to and done to prove the consistency of the expected results of any process,
procedure or method.
Qualification:
The action of proving and documenting that any premises, systems and equipment are properly installed, and/or work correctly and lead to the
expected results. Qualification is often a part (the initial stage) of validation, but the individual qualification steps alone do not constitute process
validation. Qualification is a part of validation.
https://www.pharmaguideline.com/2011/04/difference-among-calibration-validation.html 1/2
20/03/2023, 12:10 Difference among Calibration, Validation & Qualification : Pharmaguideline
Ankur Choudhary is India's first professional pharmaceutical blogger, author and founder of pharmaguideline.com, a widely-read pharmaceutical
blog since 2008. Sign-up for the free email updates for your daily dose of pharmaceutical tips.
Need Help: Ask Question
https://www.pharmaguideline.com/2011/04/difference-among-calibration-validation.html 2/2
20/03/2023, 12:13 Handling of Out of Calibration Instruments and Equipment : Pharmaguideline
-->
https://www.pharmaguideline.com/2011/07/handling-of-out-of-calibration.html 1/2
20/03/2023, 12:13 Handling of Out of Calibration Instruments and Equipment : Pharmaguideline
Ankur Choudhary is India's first professional pharmaceutical blogger, author and founder of pharmaguideline.com, a widely-read pharmaceutical
blog since 2008. Sign-up for the free email updates for your daily dose of pharmaceutical tips.
Need Help: Ask Question
https://www.pharmaguideline.com/2011/07/handling-of-out-of-calibration.html 2/2