Tsdtse

The document describes two models of thermal shock chambers, TSD 101 W and TSE 12 A, designed to meet international testing standards for uniform thermal stress application. These chambers feature advanced instrumentation for remote monitoring, improved temperature uniformity, and safety mechanisms to protect specimens during testing. They are available in different volumes and equipped with options for enhanced functionality, including easy wiring access and multilingual displays.

Uploaded by

ashley williams
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
6 views16 pages

Tsdtse

The document describes two models of thermal shock chambers, TSD 101 W and TSE 12 A, designed to meet international testing standards for uniform thermal stress application. These chambers feature advanced instrumentation for remote monitoring, improved temperature uniformity, and safety mechanisms to protect specimens during testing. They are available in different volumes and equipped with options for enhanced functionality, including easy wiring access and multilingual displays.

Uploaded by

ashley williams
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 16

Elevator type

Thermal Shock Chamber


TSD 101 W
TSE 12 A

CAT.No.E16133-W2403
Two-zone chamber capable of exposing specimens to
a uniform thermal stress.

These two-zone thermal shock chambers are designed to specifically meet the needs of
MIL, IEC, JASO and other international testing standards.
You can choose from 11L or 100L to fit your testing.
They come mounted with The N-instrumentation for improved operability and visibility,
making remote monitoring and control via an Ethernet connection possible from your desk.
Thermal shock chambers that apply uniform levels of thermal stress to specimens and that
can be used in a wide range of fields, from research and development through to
inspections and production.

1
To minimize our chambers potential environmental impact
R-449A is the best alternative to R-404A
64%
Reduction

GWP
GWP
3920 1397
*R-449A is available on request

11L 100L

TSE-12-A TSD-101-W
*The viewing window, casters (TSD) and recorder, emergency stop pushbutton (TSE) are optional.
*Equipped with options. 2
Features

Reduce test time with a two-zone elevator type

Model Volume Test area Short temperature recovery


TSE-12-A Air-cooled 10.9L W320×H148×D230mm time
TSD-101-W Water-cooled 100L W710×H345×D410mm
Meets international standards
Designed to comply with major envi-
TSD specimen temperature recovery example ronmental test standards such as IEC,
(based on MIL-STD-883 Method 1010.9) JEDEC, EIAJ, MIL, JASO, IPC and
Test conditions Measurement conditions SAE. (p.5〜 6)
Hot exposure: 150 for 30 min. Thermocouples were embedded in 10 ICs
Cold exposure: 65 for 30 min. placed on two two levels in each corner and
Specimens: ICs, 10 kg in the center of a specimen basket.
(Specimens embedded with thermocouples

200
were placed beneath other ICs.)
Transfer time within 10sec.
Less than 15 min.
150 Test area transfer time between hot/cold
)

100 ch a mb e r me e t s i nt e r n at ion a l t e st
Temperature (

50 standards requirement.
0

−50

−100 Less than 15 min.


Improved temperature
290 300 310 320 330 340 350 360 370 380 390 400 410
Time (min.)
uniformity
Uniform airflow in the test area allows
outstanding temperature uniformity.
Uniform thermal stress is applied to
TSD temperature uniformity example each specimen, minimizing variation in
Test conditions
Hot exposure: 150 for 30 min.
Measurement conditions
Thermocouples were attached to the surface of test results.
Cold exposure: 65 for 30 min. 10 ICs placed on two levels in each corners and
Specimens: ICs ( 10) in the center of a specimen basket.
200

150
0.7 Smooth specimen transfer
“Soft move mode” is automatically
)

100
Temperature (

50 a c t ivat e d whe n s p e ci me n s move


0 between the hot and cold chambers to
−50
reduce vibration and shock.
0.5
−100
330 340 350 360 370 380 390 400 410 420 430 440
Time (min.)

TSD TSE

Test area anti-drop mechanism


TSD specimen temperature recovery example to protect specimens
(based on MIL-STD-883 condition C) The test area's drive unit is equipped
Test conditions
Hot exposure: 150 for 30 min.
Measurement conditions
Thermocouples were embedded in 10 ICs w it h a b r a k i ng d ev ic e t o p r eve nt
Cold exposure: 65 for 30 min.
Specimens: ICs, 10 kg
placed on two two levels in each corner and
in the center of a specimen basket.
specimens from falling the test area
(Specimens embedded with thermocouples
were placed beneath other ICs.)
under any abnormal situations.
200 For testing of small test samples, the
Less than 15 min.
150
s p e c i m e n p r ot e c t ive m e s h wh ich
covered entire test area can be added.
)

100
(Option)
Temperature (

50

−50 TSE hot exposure TSE cold exposure


−100 Less than 15 min.
290 300 310 320 330 340 350 360 370 380 390 400 410
3 Time (min.)
Features

Easy wiring access


A cable port on right side allows for
easy wiring for specimen measurement.

Specimen Temperature
Trigger (STT)
With up to two sensors attached to
specimen(s), the STT function begins
counting the exposure time once the
specimen reaches a set temperature, or
promptly a ct ivat e s mov i ng of t he
specimen for the next exposure. This
reduces overall testing time and ensures TSD-101-W Specimen temperature measurement

( )
Specimen temperature sensors Standard: 2 sensors
a c c u r at e s p e ci me n t e mp e r at u r e s. Optional: additional 3 sensors
Temperature readings can be recorded
for each specimen and test area by
connecting a temperature recorder.
(TSD)

Safe specimen handling


thanks to ambient
temperature recovery
The ambient temperat ure recover y
feature intakes external air to return the (
TSD-101-W Test areas Top: hot chamber
Bottom: cold chamber )
test area to an ambient temperature
af ter test i ng has f i n ished or been
paused. (TSD)

Comprehensive safety system


A double safety system ensures that any
t r a n sfe r b et we e n t e st a r e a s st op s
automatically when the door is open,
and that the door locks while transfer is
in progress.

4
TEST STANDARD AND COMPATIBLE MODELS

Temperature setting Number of


Test standard Soak time Recovery time Model *
Hot (℃) Cold (℃) cycles

+70 ± 2 3 hours
−5 ±3
+ 85 ± 2 2 hours
−10 ± 3
+100 ± 2 1 hour
IEC 60068-2-14 Na − 25 ± 3 TSD
+125 ± 2 30 min. 10% of soak time 5
( JIS C 60068−2−14 Na) − 40 ± 3 TSE
+155 ± 2 10 min.
− 55 ± 3
+175 ± 2 3 hours if not specified in
− 65 ± 3
+ 200 ± 2 product specifications

A + 85(+10,− 0 ) − 55(+ 0,−10 ) Specimen 5 to 14 min.

B +125(+15,− 0 ) − 55(+ 0,−10 ) Specimen 5 to 14 min.


IEC 60749-25 Not
C +150(+15,− 0 ) − 65(+ 0,−10 ) 1/ 5/ 10/ 15 min. Specimen 5 to 29 min. TSD
(JESD22-A104F) specified
H +150(+15,− 0 ) − 55(+ 0,−10 ) Specimen 5 to 14 min.

M +150(+15,− 0 ) − 40(+ 0,−10 ) Specimen 5 to 15 min.

− 50 ± 3
+100 ± 2 − 45 ± 3
+ 95 ± 2 − 40 ± 3 3 hours
+ 90 ± 2 − 35 ± 3 2 hours
+ 85 ± 2 − 30 ± 3 1 hour
IEC 61747-5 Na 10% of soak time TSD
+ 80 ± 2 − 25 ± 3 30 min. 5・10
(EIAJ ED-2531A Na) TSE
+75 ± 2 − 20 ± 3 10 min.
+70 ± 2 −15 ± 3 3 hours if not specified in
+ 65 ± 2 −10 ± 3 product specifications
+ 60 ± 2 −5 ±3
−0 ±3

15g and below: at least 10 min.


Air 5 min. or
Max. storage Min. storage 15 to 150g: at least 30 min.
EIAJ ED−4701 10% of soak time, 10 TSD
temp. temp. 150 to 1,500g: at least 60 min.
whichever is longer
More than 1,500g: individually specified

A +125
(± 3 ) − 65
(± 3 )

B +100(± 3 ) − 65
(± 3 )
5 cycles
Air 5 min. or
unless
EIAJ ED−4702 C +100(± 3 ) − 55
(± 3 ) 30 min. 10% of soak time, TSD
otherwise
whichever is longer
specified
Mounted Mounted
D printed circuit board printed circuit board
max. operating temp. min. operating temp.

A +125 ± 5 − 25 ± 5

B +125 ± 5 − 40 ± 5
7 min. after specimen
EIAJ ET−7407 — — TSD
C + 80 ± 5 − 30 ± 5 temperature attainment

Max. operating Min. operating


D
temp. temp.

5
TEST STANDARD AND COMPATIBLE MODELS

Temperature setting Number of


Test standard Soak time Recovery time Model *
Hot (℃) Cold (℃) cycles

A + 85(+10,− 0 ) − 55(+ 0,−10 )

B +125(+15,− 0 ) − 55(+ 0,−10 )


Specimen
MIL-STD-883 10 min. or longer TSD
C +150(+15,− 0 ) − 65(+ 0,−10 ) less than At least 10
Method 1010.9 after transition start TSE
15 min.
D + 200
(+15,− 0 ) − 65(+ 0,−10 )

F +175(+10,− 0 ) − 65(+ 0,−10 )

Within 5 min. after solder


Type joint temp. reaches ±2℃ of
+ 85 − 40
1 preset temp. Or,
0.2kg and below: 0.5 hours
JASO−D902 Air 5 min. 200 TSD
0.2 to 0.8kg: 1 hour
Type 0.8 to 1.5kg: 1.5 hours
Depends on parties involved More than 1.5kg: 2 hours
2
preset temp.

A +125
(+ 3,− 0 ) − 65
(+ 0,− 5)
IPC-TM-650 2.6.6 30 min. — 5 TSD
B + 85
(+ 3,− 0 ) − 55
(+ 0,− 5)

10 min. or longer Specimen less than


SAE J1879 +150 − 55 1000 TSD
after transition start 15 min.
* The test results may not meet specifications depending on the quantity of specimens or the setting method.
■ For futher information, please contactus.

6
LAN Tablet

Wireless router

Controller TSD
Smartphone
N-Instrumentation

An easy-to-use, easy-to-read touch panel.

Tabbed interface
High resolution 7-inch LCD. Tabs at the
bottom make for quick and easy flipping
between screens. Touching an icon
displays the menu label which, touched,
makes flipping between screens easier.

Multilingual display
Use the language icon at the top of the
Email alert display to change the display language
from Japanese to English, Simplified
Chinese, Traditional Chinese or Korean
on any screen.
Ethernet Slide label
Intranet PC
Touching(company
an icon LAN)
displays the menu Quick access button
label.
Mail server For added convenience, the star ( ★ )
TSD icon can have quick access functionality
Mobile phone
assigned, such as for jumping to a
certain screen or directly launching a
saved test pattern.
Program copy and computer editing
Test data records
Temperature settings and measurements
can be stored in the internal memory and
Copy exported with the use of USB f lash
drives. This enables them to be displayed
as graphs on web browsers and stored for
back-up purposes.
TSD TSE Test data can also be recorded in real
time to a USB flash drive.
Edit * USB flash drives not included.
* Some items may not be copied between different models and chambers with
different options.
Register up to 40 test patterns

Download edit programs via


online
The Patter n Manager Lite software
installed on your PC, edit programs
according to your testing needs, and
upload them with a USB.

The Pattern Manager Lite software allows


you to edit programs for your chamber,
view and edit data as graph, etc.
USB memory port The software can be downloaded from
the Test Navi website.

7
Network
Wireless LAN connection
* Requires an intranet

Remote monitor and control


(Ethernet connection) Wireless LAN connection
LAN Tablet

The chamber comes with an ESPEC Wireless router


original web application. Connecting to
the chamber Ethernet port (LAN’s port)
TSD LAN
makes it possible to control chamber Smartphone
Tablet
monitoring, pattern setting, operation
Wireless router
start/stop, and other operations from a
computer web browser. Installation of
special software is not required. All you TSD
Smartphone
ne e d is a st a nd a rd c omput e r web
browser to connect with the chamber. Watch the videos
Login privileges Remote ON/OFF Connecting Easily
Screen Chamber Pattern Run/ Operation with Ethernet
Configuration
Privileges monitor setting Stop (LAN)
Administrator

Operator

User

Edit test patterns on a web Email alert


browser
Saved test programs can be edited on a
web browser. Test programs can also be PC
downloaded to your PC. Email alertEthernet Intranet
(company LAN)

E-mail alert Mail server


Alerts such as for a test ending, for TSD Ethernet Intranet PC
(company LAN) Mobile phone
maintenance, and errors are e-mailed to
multiple recipients.
Mail server

Program
TSD copy and computer editing Mobile phone

Program copy and


Copycomputer editing
Internet
ESPEC OnlineCore (Sold separately)

Central control system recommended for


multiple environmental TSD Intranet TSE
Copy
test chambers installations
Edit
* Some items may not be copied between different models and chambers with
differentESPEC
Office options.
OnlineCore
TSD TSE
Network
camera
Measuring
Editother
Test chambers of
manufacturers
instrument

* Some items may not be copied between different models and chambers with
different options.

*Please contact ESPEC for more information, about which products can be connected.

8
TSD
100L

TSD-101-W
Model
Water-cooled
System 2-zone transition by vertical transfer of specimens
Hot exposure range + 60℃ to + 205℃
Test area Cold exposure range −77℃ to 0℃
Temp. fluctuation *2 ± 1.0℃
Hot Pre-heat upper limit + 205℃
chamber Heat-up time *3 Within 90 min from ambient temp. to + 200℃ (Setting: + 205℃)
Cold Pre-cool lower limit −77℃
Performance *1

chamber Pull-down time *3 Within 90 min from ambient temp. to −77℃ (Setting: −77℃)
・Hot exposure: +150°C (setting: +155°C 30 min)
Temp. ・Cold exposure: − 65°C (setting: − 68°C 30 min)
recovery Recovery conditions
・Sensor position: downstream
performance ・Specimen: Plastic molded ICs, 10kg
(2-zone)
Temp. recovery time Specimen temp. within 15 min
Transfer time between hot & cold chambers Within 10 seconds
・Hot exposure: +150℃ to max. + 55℃
Ambient Recovery conditions ・Ambient temp.: + 23℃
recovery ・Specimen: Plastic molded ICs, 10 kg
Ambient temp. recovery time Within 90 min
Test area Shelf brackets on 2 levels (up to 4 baskets can be installed)
System Mechanical cascade refrigeration system
Refrigeration unit Compressor 3.73kW(5HP) scroll-type × 2
Refrigerant R-404A〔 Low GWP Refrigerant R-449A is available on request〕, R-23
Cooler Plate fin cooler and cold accumulator
Elevating unit Power slider (250W)
Inside dimensions W710×H345×D410 mm
Test area capacity 100 L
Load capacity of test area *4 30 kg
Specimen basket load capacity 5kg per basket
Outside dimensions *5 W1100×H1885×D1965 mm
Weight Approx. 1100 kg
Ambient temp. range + 5℃ to + 40℃
Power supply 200V AC 208V AC 220V AC 380V AC 400/415V AC
Utility requirements

(Voltage fluctuation: rating ± 10%) 3φ 3W 50/60Hz 3φ 3W 60Hz *6 3φ 3W 60Hz 3φ 4W 50Hz 3φ 4W 50Hz *7


Maximum load current 64 A 62 A 58 A 34 A 32 A
Cooling water supply pressure *8 0.2 Mpa Gauge to 0.5 Mpa Gauge (2 kg/cm2 G to 5 kg/cm2 G)
Cooling water supply rate *9 2050L/ h (at reference water temp. + 25℃), 3400L/h (at reference water temp. + 32℃)
Piping connection size Carbon steel pipe, ID 32 mm
Cooling water temp. range + 5℃ to + 38℃
Noise level *10 Max. 65 dB
Exhaust heat rate 12600 kJ/h (3000 kcal/h)
Exhaust air volume 250 m3/h
*1 Under the conditions of a + 23℃ ambient temperature, cooling *6 This model complies with the requirements of the National Electric Code (NFPA 70) for
water temperature + 25℃, rated voltage, and no specimen inside the United States of America (NEC spec.)
the test area. *7 This model complies with the requirements of the European Community Directives (CE spec.)
*2 The performance values are based on IEC 60068-3-5:2001, *8 A pressure regulator valve is required if the pressure exceeds 0.5MPa Gauge (5kg/cm2G)
*3 When each chamber is operated independently *9 Rate depends on the cleanliness of the heat exchanger
*4 When using the test area floor or heavy-duty shelves (option) *10 Measurements are to be taken in an anechoic room at a height of 1.2m from the floor, and
*5 Excluding protrusions a distance of 1m from the front panel (ISO 1996-1: 2003.A-weighted sound pressure level)

International standards and regional standards compliance


RoHS directive
Safety of Machinery (ISO 12100), Low Voltage (IEC 60204-1), EMC* (IEC 61000-6-2, EN 55011).
* For the CE-marked only.
9
FITTINGS LOCATION TSD-101-W

Specimen temperature
input terminai

Specimen power supply


Operation lamp USB memory port Time signal output terminal
control terminal

Recorder (option)
Cable port
ID 100mm

Overheat protect /
overcool protection Hot chamber

Ethernet port

Cold chamber
Emergency stop
pushbutton
(option)

Front side Right side


Power cable is optional, not equipped as standard fitting.

SAFETY DEVICES ACCESSORIES



Leakage breaker (200, 220V AC) ●
Specimen basket
Time signal Specimen power supply

Circuit breaker (208, 380, 400/415V
Test area
AC) output
(18-8 terminal
Cr-Ni stainless steel: 5 mesh control
metal terminal
basket)

Electrical compartment door switch W700×H40×D410 mm/ load capacity 5kg 2

Hot chamber overheat protection switch

Cold chamber overheat protection switch

Hot chamber overheat protector
(Built into temperature controller)

Cold chamber overheat/ overcool protectors
(Built into temperature controller)

TestEmergency
area overheat/ overcool protectors
stop pushbutton Over protect /
(Built into temperature controller)
(option) overcool protection Ethernet port

Test area overheat/ overcool protectors Time signal output terminal
Operation lamp Shelf brackets 2 sets
Specimen power supply
Circuit breaker


Recorder (option)
Refrigerator high/ low pressure switch

Cartridge fuse (3A, 5A, 7A, 10A, 15A) control terminal 4

Compressor
USB memory built-in protector

Cable port rubber plug
Cable port 2
port ID 50mmPerforated cable port cap 1
Temperature switch for compressor


Water suspension relay

Wire fisher (specimen wiring tool) 1

Temperature switch for air circulator

Thermocouple 2

Air circulator thermal relay

Specimen temperature input connector 2

Motor inverter

3-pole socket (208V AC spec. only) 3

Motor reverse prevention relay

Nipple R1 1/4 in. (32 A) 1

Hot chamber door switch

Strainer R1 1/4 in. (32 A) 1

Cold chamber door switch

Strainer element R1 1/4 in. (32 A) 1

Door lock mechanisms

Breaker handle cover (except 208V AC) 1
Casters Operation manual 1
Cartridge fuse

Front side Warranty cardRight side 1


Specimen power supply control terminal


Cooling tower interlock terminal Power cable Approx.2.5m length from chamber

10
TSE
11L

TSE-12-A
Model
Air-cooled
System 2-zone transition by vertical transfer of specimens
Hot exposure range + 60℃ to + 200℃
Test area Cold exposure range − 65℃ to 0℃
Temperature fluctuation *2 ± 0.5℃

Hot Pre-heat upper limit + 205℃


chamber Heat-up time *3 Within 30 min. from ambient temp. to + 200℃ (Setting: + 205℃)
Performance *1

Clod Pre-cool lower limit − 82℃


chamber Pull-down time *3 Within 90 min. from ambient temp. to − 80℃ (Setting: − 82℃)
・ Hot exposure: +150℃, 30 min
Temp. ・ Cold exposure: − 65℃, 30 min
recovery Recovery conditions
・ Sensor position: Upstream
performance ・ Specimen: Plastic molded ICs 2 kg
(2-zone)
Temp. recovery time Within 5 min.
Transfer time between hot & cold chambers Within 10 seconds
Test area Shelf brackets on 2 levels of fixed location
Heater Stripped wire heater
Construction

System Mechanical cascade refrigeration system


Refrigeration
Compressor Rotary 1.5 kW × 2
unit
Refrigerant R-404A〔 Low GWP Refrigerant R-449A is available on request〕, R-508A
Cooler Plate fin cooler and cold accumulator
Elevating unit Linear motor (55W)
Test area capacity W320×H148×D230mm
Inner volume of test area 10.9 L
Load capacity of test area 8 kg
Outside dimensions *4 W680×H1745×D1050mm
Weight Approx. 400kg
Ambient temp. range 0 to + 40℃
Utility requirements

Power supply 200V AC 220V AC 380V AC 400/415V AC


(Voltage fluctuation: rating ± 10%) 3φ 3W 50/60Hz 3φ 3W 60Hz 3φ 4W 50Hz 3φ 4W 50Hz *5
Maximum load current 26A 25A 17A 17A
Noise level *6 Max.60dB
Exhaust heat rate *7 17,585kJ/h (4200 kcal/h)
*1 The performance values are under the conditions of a +23℃ ambient temperature, relative humidity of 65%rh, rated voltage, and no specimen. Heat up time and pull
down time are those of single-unit operation of each chamber.
*2 The performance values are based on IEC60068-3-5:2001.
*3 Temperature heat-up/pull-down time account for performance of each temperature chamber.
*4 Excluding protrusions.
*5 Compliance with CE Marking.
*6 At 1m from front of chamber, 1.2m from floor. (ISO 1996-1:2003 A-weighted sound pressure level) depending on environment
*7 At ambient temperature +23℃.

International standards and regional standards compliance


RoHS directive
Safety of Machinery (ISO 12100), Low Voltage (IEC 60204-1), EMC (IEC 61000-6-2, EN 55011).

11
Time signal Specimen power supply
Test area output terminal control terminal

FITTINGS LOCATION TSE-12-A

Emergency stop pushbutton Over protect /


(option) overcool protection Ethernet port
Time signal output terminal
Operation lamp Specimen power supply
Recorder (option) control terminal

USB memory Cable port


port ID 50mm

Casters
Front side Right side

Power cable Approx.2.5m length from chamber

Safety precautions
Do not use explosive substances, flammable substances, or substances that
SAFETY DEVICES ACCESSORIES
contain those substances as a specimen under any circumstances.
Danger: Risk of explosion and fire

Leakage breaker (200, 220V AC) Specimen
Do not put corrosive

substancesbasket
inside the test area. If corrosive substances are

Circuit breaker (380, 400 / 415V AC) (18-8
generated from the Cr-Ni stainless
specimen, the life of steel, 5 mesh
the product's metal basket)
corrosion resistance will
W320×H35×D230mm
decrease dramatically /load capacity:
due, in particular, to corrosion of stainless2kg 2
steel, resin, and

Electrical compartment door switch
silicone.

Hot chamber overheat protection switch Do not use living organisms or items that exceed the allowable heat load as a

Cold chamber overheat protection switch specimen.

Hot chamber overheat protector Always read the operation manual before using the Product.
(Built into temperature controller)

Cold chamber overheat / overcool protectors
(Built into temperature controller)

Test area overheat and overcool protectors
(Built into temperature controller)
Safety precautions

Test area overheat / overcool protectors

Cartridge
Do not use specimens fuse
which are explosive or inflammable,

Refrigerator high pressure switch 3A,substances.
5A (200/220V AC) 1each
or which contain such To do so could be

Thermal relay for compressor 3A, lead
hazardous, as this may 5A,7A (380/400/415V
to fire or explosion. AC) 1each

Temperature switch for compressor Cable
Do not place corrosive

port rubber
materials plug If corrosive
in the chamber. 2

Temperature switch for air circulator substances or humidifying
● water(specimen
Wire fisher is used, the wiring
life of the unit
tool) 1
may be significantly shortened.

Thermal relay for air circulator ●
Breaker handle stopper (200/220V AC only)
Do not place life forms or substances that exceed allowable

Motor inverter ●
Operation manual 1
heat generation.

Motor reverse prevention relay ●
Warranty
Be sure to read the operationcard
manual before operation. 1

Hot chamber door switch

Cold chamber door switch

Test area hold

Door lock mechanisms

Fuse
Safety precautions

Specimen power supply control terminal
Do not use specimens which are explosive or inflammable, or which contain such
substances. To do so could be hazardous, as this may lead to fire or explosion.
Do not place corrosive materials in the chamber. If corrosive substances or
humidifying water is used, the life of the unit may be significantly shortened.
Do not place life forms or substances that exceed allowable heat generation.
Be sure to read the operation manual before operation.

12
OPTIONS TSD-101-W / TSE-12-A

Power cable Additional cable port TSD Paperless recorder

200V, 220V AC only Provided in addition to the standard Records temperature of each section such
・5 m cable port. (right side) as the temperature inside the chamber.
・10 m Location: Left side of the main unit
Display: 5.7inch color touch panel
Internal diameter: 100 mm
〈TSD〉 Number of inputs (Initial setting):
* If this option is not specified, the chamber does ・1 (5 more channels can be turned ON)
not come with a power cable.
Data saving cycle: 1 sec
Cable port rubber plug
・3 (3 more channels can be turned ON)
Data saving cycle: 1 sec
Prevents air leakage from the cable
Viewing window TSD ・3 (3 more channels can be turned ON)
port.
Data saving cycle: 5 sec
Used for observation of the specimens ・5 (1 more channels can be turned ON)
inside the chamber. Interface Data saving cycle: 1 sec
Dimensions: W190×H340 mm ・5 (1 more channels can be turned ON)
Chamber lamp: Halogen lamp (×1) Data saving cycle: 5 sec
・RS-485C
・6 Data saving cycle: 1 sec
・RS-232C
・6 Data saving cycle: 5 sec
・GPIB
Temperature range: −100℃ to +220℃
Internal memory: 8MB
External memory media:
Communication cables
CF memory card (256 MB)
External memory function: USB port
・RS-485C 5m/10m/30m Language support: ENG/ JPN
・GPIB 2m/4m * Select either built-in or portable type. (TSD)

Temperature recorder (digital)

Specimen basket/ shelf brackets −100℃ to +220℃ /100 mm


・RK-61: 1 pen
Equivalent to standard accessory. ・RK-63: 3 pens
Material: Stainless steel (5 mesh) ・RK-64: 6 dots
〈TSD〉
・Basket Recorder wiring
・Shelf brackets
〈TSE〉 Preparation of a power cable, tem-
・Basket perature sensor, and a grounding wire
for a dd it ional i nst allat ion i n t he
future.
Heavy-duty shelf TSD

Used to hold heavy specimen exceed- Recorder terminal


ing the load capacity of the standard
specimen basket. Used to output the temperature within
Load capacity: 15 kg test area, hot chamber, cold chamber.
* Equally dist ributed load, not included shelf
brackets and specimen baskets.

13
OPTIONS TSD-101-W / TSE-12-A

Thermocouple Emergency stop pushbutton Anchoring fixtures

Attached to specimens to measure Stops the chamber immediately. Used to bolt the chamber to the floor.
specimen temperature.
〈TSD〉
Thermocouple type T without ball Chamber dew tray
(Copper/ Copper-Nickel)
P revent s water lea k age f rom t he
〈TSE〉 chamber onto the floor.
T JIS C1602 with ballattached With cover With guard * The use of casters is recommended to facilitate
・2 m operation. (TSD)
・4 m *To prevent damage in the event of water leakage,
other preventive measures are also available.
・6 m

STT 3-point expansion TSD Casters TSD

Additional 3 points of measuring the Installed for mobility.


specimens' temperat ures used for Auxiliary cooling injector (LCO2) Casters: 6
S p e c i m e n Te m p e r a t u r e Tr i g g e r levelling-feet: 4
function. Used to reduce the temperature recovery
(2 points are equipped as standard.) time of low temperature exposure by
injecting liquefied carbon dioxide at
beginning of exposure.
Exposure signal output TSD

A signal is output via a contact switch Auxiliary cooling injector (LN2)


when test area temperature is within
the user-selected range. This signal Used to reduce the temperature recovery
can be used to cont rol per ipheral time of low temperature exposure by
Operation manual
instruments, like applying a voltage to injecting liquefied nitrogen at beginning
specimens only during hot exposure, of exposure.
or monitoring test operation from a ・CD
remote point. ・Booklet

Total cycle counter Reports & certificates

Indicates cycle counts. ・Testing and inspection report


Display range: 1-99999999 ・Test data
External alarm terminal
(with resetting function) ・Calibration report
・Calibration certificate
If the safety device of the chamber is
・Traceability system chart
activated, the external alarm terminal
・Traceability certificate
will notify it to a remote point.

Additional overheat protector

Additional preventive measures can be TSD


taken for excessive temperature rise in
the chamber, in addition to the standard
equipped double overheat protector.

■ Some photographs presented in this catalog contain Japanese display.

14
https://www.espec.co.jp/english
Head Office
3-5-6, Tenjinbashi, Kita-ku, Osaka 530-8550, Japan
Tel : 81-6-6358-4741 Fax : 81-6-6358-5500

ESPEC NORTH AMERICA, INC.


Tel : 1-616-896-6100 Fax : 1-616-896-6150
ESPEC EUROPE GmbH
Tel : 49-211-361850-0
ESPEC ENVIRONMENTAL CHAMBERS
SALES AND ENGINEERING LTD. STI. (Turkey)
Tel : 90-212-438-1841 Fax : 90-212-438-1871
ESPEC ENVIRONMENTAL EQUIPMENT (SHANGHAI) CO., LTD.
Head Office
Tel : 86-21-51036677 Fax : 86-21-63372237
BEIJING Branch
Tel : 86-10-64627025 Fax : 86-10-64627036
GUANGZHOU Branch
Tel : 86-20-83317826 Fax : 86-20-83317825
SHENZHEN Branch
Tel : 86-755-83674422 Fax : 86-755-83674228
SUZHOU Branch
Tel : 86-512-68028890 Fax : 86-512-68028860
TIANJIN Branch
Tel : 86-22-26210366 Fax : 86-22-26282186
XI’AN Branch
Tel : 86-29-88312908 Fax : 86-29-88455957
CHENGDU Branch
Tel : 86-28-88457756 Fax : 86-28-88474456
WUXI Branch
Tel : 86-510-82735036 Fax : 86-510-82735039
ESPEC TEST TECHNOLOGY (SHANGHAI) CO., LTD.
Tel : 86-21-68798008 Fax : 86-21-68798088
ESPEC ENGINEERING (THAILAND) CO., LTD.
Tel : 66-3-810-9353 Fax : 66-3-810-9356
ESPEC ENGINEERING VIETNAM CO., LTD.
Mail : EEV_support@espec-vn.com

ISO 9001 JIS Q 9001


Quality Management System Assessed
and Registered
ESPEC CORP. has been assessed by and registered in the
Quality Management System based on the International MS
CM001
Standard ISO 9001:2015 (JIS Q 9001:2015) through the JIS Q 9001
JSA Solutions Co.,Ltd. JSAQ 004

The organization of these certificates is


ESPEC CORP. Japan.

ISO 27001 JIS Q 27001


Quality Management System Assessed
and Registered
The organization of these certificates is
ESPEC CORP. Japan.
JIS Q 27001
JSAI 165

ISO 14001 (JIS Q 14001)


Environmental Management System Assessed
and Registered
The organization of these certificates is MS
CM021
ESPEC Group Japan.
EC99J2046 0051

Specifications are subject to change without notice due to design improvements.


T3B06C00 (The contents of this catalog is as of March, 2024.) Corporate names and trade names mentioned in this catalog are trademarks or registered trademarks.