OM Performa
OM Performa
GE Healthcare
Operation Instructions
37094-IMG rev 3
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
Copyright Code: 37094-IMG rev 3 Date: 10/2006
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
Documentation, trademark and the software are copyrighted with all
rights reserved. Under the copyright laws the documentation may not be
copied, photocopied, reproduced, translated, or reduced to any electronic
medium or machine readable form in whole or part, without the prior
written permission of GE Healthcare.
This manual applies to Performa units with software version 7.71a and up.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
Table of Contents
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
1 Introduction .............................................................................................................. 1
1.1 General .................................................................................................................................................................1
1.2 Radiation protection.......................................................................................................................................2
1.3 Conventions in this manual ........................................................................................................................3
2 Operation controls................................................................................................... 5
3 Operation procedure............................................................................................... 9
3.1 Preparations.......................................................................................................................................................9
3.1.1 Switching on the unit .................................................................................................................................... 9
3.1.2 Replacement of the accessories ............................................................................................................. 9
3.2 Selection of the technique factors........................................................................................................14
3.3 Automatic compression release............................................................................................................17
3.4 C-Arm adjustments......................................................................................................................................18
3.4.1 Vertical adjustment......................................................................................................................................18
3.5 Selection of the image angle...................................................................................................................19
3.5.1 Compression ...................................................................................................................................................19
3.5.2 X-ray field illumination light .....................................................................................................................20
3.5.2.1 Beam filter selection............................................................................................................................. 20
3.6 Patient positioning........................................................................................................................................21
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
6.2.1 Operating Parameters................................................................................................................................52
6.3 Tube loading chart for the M113SP tube...........................................................................................54
6.4 Electromagnetic Compatibility (EMC) tables ....................................................................................55
7 M113SP X-ray tube data sheets...........................................................................59
7.1 M113SP tube data sheet ...........................................................................................................................59
7.2 M113SP Filament emission charts........................................................................................................59
7.3 Anode heating and cooling curves.......................................................................................................60
7.4 Housing heating and cooling curves...................................................................................................60
7.5 Single load ratings........................................................................................................................................61
7.5.1 3 Ø Constant Potential 50 Hz ..................................................................................................................61
7.5.2 3 Ø Constant Potential 60 Hz ..................................................................................................................62
7.6 Description .......................................................................................................................................................63
7.7 General data ...................................................................................................................................................63
7.8 B-112 Housing................................................................................................................................................63
7.9 Filaments ..........................................................................................................................................................64
7.10 Stator 50/60 hz ..............................................................................................................................................64
7.11 Inherent filtration ..........................................................................................................................................65
7.12 Rayproofing .....................................................................................................................................................65
1 Introduction
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
1.1 GENERAL
Performa is a high quality x-ray mammography unit intended for both
screening and diagnostic examinations. The main features are the
rotating dual-angle anode tube, Tuned High Frequency (THF) generator
and easy positioning of the patient.
The vertical adjustment is motorized, and the C-Arm rotation angle ranges
from +135° to -180° degrees. The motorized compression of the breast is
controlled by pedals, and the manual compression by manual control. If
wanted, the compression can be released automatically after exposure. In
power failure the compression is released automatically.
The THF generator is integrated with the unit. The control panel can be
mounted on either side of the unit or on the wall of the examination room.
The exposure button can be situated beside the control panel or outside
the examination room.
NOTE!
Move the foot pedals away from the working area to prevent any
unwanted activations of compression.
WARNING!
Never use the equipment if it is suspected that any electrical, or radiation-
emitting component is defective or inoperative.
WARNING!
Only trained, licensed persons are allowed to operate the Performa unit.
WARNING!
USE LIMITATION : External components: The use of accessories,
transducers, and cables other than those specified may result in
degraded ELECTROMAGNETIC COMPATIBILITY of the EQUIPMENT and/or
SYSTEM.
WARNING!
If you suspect any electro-magnetical interference affecting or caused by
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
the unit, call service. Portable and mobile RF communications equipment
can interfere with operation of the medical electrical equipment.
This device emits X-ray radiation for medical purposes. X-ray tube has
molybdenum anode. X-rays are molybdenum or rhodium filtered.
Minimum HVL 0,3 mmAl without compression paddle. Performa is CE-
marked in accordance with MDD (93/42/EEC). Performa fulfills the EMC
CAUTION!
The maximum radiation that the system is capable of producing
corresponds to 5% of the typical CT-radiographic exposure. However, the
user should wear a dose recording device and follow local regulations for
radiation protection.
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
NOTE!
Contains useful information for the reader about the unit and its use.
CAUTION!
Contains important instructions. If these instructions are not observed,
malfunction of the unit or damage to the unit or other property may
occur.
WARNING!
Contains warnings and instructions about the safety of the unit. If these
warnings are not respected, serious risks and injury may be caused to the
patient and operator.
GE Healthcare
37094-IMG rev 3
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
2 Operation controls
2 Operation controls
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
11
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
A B C D E
Fig 2.2. C-Arm display (A: Breast Thickness (cm), B: Compression force (N), C: C-Arm angle
(°), D: Filter, E: Filter selection)
Fig 2.4. The ECS Pedal for compression and X-ray field illumination light switch
1 2 3
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
4 5
6 7
9
8
14
12 13
10
11
1 Ready indicator
2 Radiation indicator
NOTE!
When “grid” light is displayed, the large focal spot is selected. When
“magn” light is displayed, the small focal spot is selected.
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
1
2
3
NOTE!
In the Great Britain (GB) model toggle switch (2) and return button (3) are
functional only when compression is fully released.
1 Balance indicator (if illuminated the C-Arm is not in balance and the
C-Arm brake can not be released, press return button (3))
MINIMUM
3 Operation procedure
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
3.1 PREPARATIONS
3.1.1 Switching on the unit
! WARNING !
32458
When the light is on parts
Notice the warning label next to the power switch. Use the circuit breaker
of the system are energized located in the back of the carriage In order to switch the power off
although power switch is off. completely.
! ATTENTION !
Lorsque ce voyant est allumé, Switch the power ON using the mains switch. The timer display now
cela indique que certaines parties shows r.ES, kV display Pr, mAs display shows the software version and
du système sont sous tension bien que
l ' alimentation électrique soit coupée. density display shows the country code. Resetting must be acknowledged
by pushing the auto exp button, which starts the internal testing routine.
The following accessories are available for the unit (for more information,
refer to ACCESSORIES LIST 31581):
– 18 x 24 Semicircular diaphragm
– 18 x 24 cassette holders
– 18 x 24 Bucky devices
– 24 x 30 semicircular diaphragm
– 24 x 30 cassette holders
– 24 x 30 Bucky devices
– 10 x 24 cm compression paddle
– 21 x 24 cm compression paddle
– 21 x 24 cm XXL compression paddle
– 24 x 30 cm compression paddle
– 24 x 30 cm XL compression paddle for extra large breasts
– 24 x 30 cm XXL compression paddle for extra large breasts
– 21 x 20 cm compression paddle
– spot magnification tunnel (1.6 X)
– magnification kit (1.6 X, 1.8 X, 2.0 X and MammoSpot 1,6 X)
– magnification diaphragm
– magnification tunnel
– magnification compression paddle
– spot magnification compression paddle
– spot diaphragm
– spot kit for contact use
– biopsy and needle localization compression paddles
NOTE!
If the selected diaphragm is larger than the film, the exposure is
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
prevented and a C.CO error code is displayed
To install a special diaphragm, slide the plate into the groove in the
direction of the patient, and snap the plate into place. Refer to figure
below.
Insert the desired paddle into the groove and push the paddle upwards
until it locks.
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
Fig 3.2. Replacement of compression paddle
NOTE!
When using AA-mode with XL or XXL compression paddles, select higher
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
kv than Performa automatically proposes.
Align the guides of the desired cassette holder with the grooves on the C-
Arm. Push the cassette holder to its position. The Bucky devices and
magnification tunnel are replaced in the same way.
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
Patient positioning and exposure techniques are similar to those
described in sections 3.6, Patient positioning and 3.7, Exposure procedure
When using the Bucky device kV-values must be 2-3 kV higher compared
to exposures with cassette tunnel.
There is a separate bucky for both film formats 18x24 and 24x30 and for
The purpose is to maximize the image quality by having the object to film
distance as short as possible which reduces blurring and by being able to
use a low kV-setting because of low attenuation in the cassette holder
itself.
There is a cassette tunnel for both film formats, 18x24 and 24x30 cm and
for both cassette types, normal cassette and daylight cassette.
NOTE!
See more information of accessories in section Special exposures.
Face shield
Face shield is used to keep patient’s face out of x-ray beam during
exposure.
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
Face shield is installed by sliding it in along the grooves until the magnetic
lock is set. Remove the face shield by pulling it out along the grooves until
the magnetic lock is released.
Select the mode with the button for automatic exposure button. The light
at the corner of the button indicates that the AEC operation is selected,
mAs display indicates AA if Auto-kV is selected and A if semi automatic
AEC is selected, refer to Fig.3.4, Control panel in Auto kV or semi AEC
mode.
In the Auto kV mode the exposure time and the kV are automatically
selected to produce an optimum image quality.
For manual mode selection, press AUTO EXP so many times that the mAs
indicator displays a mAs value and the AUTO EXP light indicator turns off.
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
focus, 25 kV - 35 kV for small focus and 20 - 35 kV in manual mode for
both focal spots.
NOTE!
In Michigan state mode the kV is limited to 34 kV
Select the Density with the density selection buttons. Use - Density for
lighter images, and + Density for darker images. Each step (nominal and
dependent on the phototimer calibration) decreases/increases the
exposure time by 10% (default, adjustable through software between 5%
- 20%). The range of correction is from -7 to +7.
The position of the phototimer detector can be adjusted from the knob
under the cassette holder. The detector location display indicates where
the detector is located corresponding to the breast. The semicircle on the
paddle indicates the size and location of the detector.
NOTE!
For country code 12 (Great Britain) the time range is from 0.04 sec. to 6
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
sec., and maximum mAs is 600 mAs.
NOTE!
In the New York version the maximum exposure time is the time that is
needed to achieve 300 mAs.
NOTE!
If a proper optical density could not be reached, even if maximum
exposure time would be used, the Exposure monitoring will terminate the
exposure in 50 ms. Select a higher kV and make a new exposure. CHANGE
THE CASSETTE.
Fig 3.5. Control panel in the manual mode FOR TRAINING PURPOSES ONLY!
Select the mAs setting with mAs ± keys. Pressing the mAs+ button
increases the value and pressing mAs- button decreases the value as
shown in Table 3.1.
4 6 12 16 20
25 32 40 50 63
80 100 125 150 175
200 250 300 350* 400*
450* 500* 550** 600**
Table 3.1 Selection of mAs values
The maximum mAs-value is, however, limited by the selected kV and the
maximum permitted time. The maximum mAs/mA values are shown in
table 3.2, Maximum mAs/mA values depending on kV, focus size and
country code. Software version 7.71a..
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
MGF 110 0.3 MGF 110 MGF 110
focus N.Y. 0.1 focus
0.3 focus
NOTE!
The automatic release should be turned off during a needle biopsy
procedure.
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
The cassette holder vertical adjustment range is 40 cm - 144 cm (15” - 56”)
with the C-Arm in the upright position and 80 cm - 194 cm (31” - 76”) with
C-Arm at lateral position.
The vertical movement is motorized, and the control keys are placed on
both sides of the unit. There is a button also at the lower part of handle,
used when the C-Arm is tilted (The arrow towards the image receptor
indicates downward direction and the arrow pointing away from the
image receptor upward direction).
NOTE!
If country code 16 (Australia) is selected, vertical movement starts first
after compression fully released.
NOTE!
If country code 12 (G.B.) is selected, vertical movement is inhibited during
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
Release the brake by pressing on of the release buttons and rotate the C-
Arm (figure 3.7, Selection of the image angle).
The image angle can be seen on the digital display (10) located on the C-
Arm and also at the scale ring at the C-Arm axle. The angles -135°, -120°, -
90°, -60°,-45°, 0°, +45°, +60°, +90° and +120° can be felt as détente points.
The C-Arm is locked when the button is released.
NOTE!
If country code 16 (Australia) is selected, compression is released when C-
Arm brake is released.
3.5.1 Compression
Remember, care must be taken in order not to injure the breast. The
compressed breast thickness and the compression force is display in the
C-Arm display, refer to page 4.
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
Manual compression controls are turned clockwise for compression and
counterclockwise for release (Fig. 3.8, Compression setting by manual
controls). As the manual controls allow considerable compression, up to
300 N (66 lbs.), care must be taken not to injure the breast.
For large and dense breasts the Rhodium filter should be used to reduce
the exposure time and the patient dose. Manual selection of the filter
(selected from the filter button on the C-Arm) is default, but system can
be programmed to change the filter according the breast thickness. The
selected filter is indicated with a light next to the filter selection button
and on the control panel timer display when the unit is in ready position.
NOTE!
After the exposure the timer display shows the exposure time. After the
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
cassette is removed from the cassette holder, the timer display indicates
which filter is in use. mo = Molybdenum filter (0.03 mm), rh = Rhodium filter
(0.025 mm).
NOTE!
When using the Rhodium filter do not increase the kV-setting.
NOTE!
Always advise the patient not to use deodorant prior to a mammogram!
Certain deodorants can mimic as microcalcifications on the film.
The ECS - Easy Compression System adds one more movement to the
mammography system: the direction defined by the image receptor and
the X-ray tube.
The ECS makes it easier to elevate the breast and open the
inframammary fold. Therefore it maximizes the amount of breast tissue
visible in the image. In positioning the ECS will not replace the technologist
but helps her to do positioning more easily.
In clinical use the amount of images that have pectoral muscle visible in
cc-projections has increased from 20% (the amount prior to the use of the
ECS) to 50%.
The ECS is excellent in biopsies. If the lesion is close to the chest wall,
applying compression from down below ensures that the lesion stays
under the compression paddle. With standard compression the lesion
might may move away from underneath the paddle.
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
1 C-Arm is driven up without compressing to open the inframammary
fold
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
cassette holder.
3 Turn the C-Arm to the upright position by pressing the image angle
release button so that the scale of image angle shows zero.
4 Adjust the cassette holder to a suitable height so that the nipple is in
profile and that no skin folds are present under the breast with the
vertical drive control. See figures 3.9, Correct vertical position and
3.10, Incorrect vertical position.
5 Slightly rotate the patient medially leaning towards the stand. Refer
to Fig. 3.11, Patient positioning at cranio-caudal projection.
6 Allow the patient to grip with her arm under the cassette holder or
the hand grips, and lean towards the stand so that the cassette
holder under the breast is against the ribs.
NOTE!
If the breast does not cover the detector or the breast has an implant,
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
manual technique must be selected.
WARNING!
Keep focal spot to skin distances as large as possible in order to keep the
absorbed dose to the patient as low as reasonably achievable.
NOTE!
kV selected should produce an exposure time from 0.5 to 2 seconds. If
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
exposures are longer than 2 seconds, go up in kV. If they are shorter than
0.5sec., go down in kV.
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
holder. Turn the C-Arm to an angle of about 45°.
3 Adjust the cassette holder to a suitable height from the vertical drive
control. Place the upper edge of cassette so the axillary tail will be
visualized.
4 Ask the patient to hold the arm at a 90-degree angle and turn the
patient slightly to an oblique position. Rotate the patient towards the
cassette.
5 Place the breast on the cassette holder so that the breast totally
covers the detector of the automatic exposure control. Make sure
that the lateral tissue is pulled forward.
6 Check that the breast covers the detector.
7 Set the compression with the pedal or with the manual controls.
Make sure that the nipple is in profile and that no skin folds are
present.
8 Check the ready and auto rel lights.
9 Press exposure button to initiate an exposure, and keep it pressed
until the radiation indicator light is out and the audible sound is over.
Then release compression.
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
1 Load the film into the cassette.
2 Push the cassette lock backwards to insert the cassette into the
cassette holder.
3 Slide the cassette in until the lock returns to the normal position.
If the Ready light does not turn on 50 seconds after an exposure, read
section 5.2, Ready light not lit.
NOTE!
Check that the proper filter is selected.
Remove the cassette by pushing the cassette lock back from one side of
the cassette holder, and pushing the cassette from the other side.
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
Fig 3.16. Removing the cassette
– 36.7 x 101 mm
– 26.0 x 101 mm
For each of the four rows printed, the last 15 characters of the 40 on each
row are freely programmable. The following is an example of the text
produced by Performa and printed on to a label:
1234567890123456789012345678901234567890
B Rh 0.02s 28kV 002mAs +0 HOSPITALxNAMEx
6.5cm 200N 180^/R 0.06mGy CODExxxxxxxxxx
15 / 08 / 00 R1 12345 RADIOLOGISTxxx
_________________________ DEPARTMENTxxxx
Explanation
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
B Bucky / Cassette tunnel / Magnification
Rh Rhodium / Molybdenum filter
+0 Density correction
28kV kV
002mAs mAs
0.02s Exposure time
6.5cm Breast thickness
200N Compression force
180^/R C-Arm tilt angle, ^= degrees (°), to the /Right or
/Left
0.06mGy Average patient dose
15 /09 /04 Date of examination (day/month/year)
R1 Site of the detector (L1, L2, C1, C2, C3, C4, R1, R2)
Use digital film flasher to print the information permanently onto film.
By having the breast at a distance from the film causes the image to be
magnified. This increases the size of the smallest objects
(microcalcifications) above the grain size of the film-screen combination
thus improving the image quality.
Absorption in the air gap reduces scatter radiation thus further improving
the contrast. Good image quality requires a small focal spot though and
therefore a true 0.1mm focus is needed.
Although the object may be visualized well, this is offset by the fact that
only a small area of the breast can be imaged at a time. It therefore a
matter of opinion which magnification factor is used.
With the Performa system magnification factors 1.6x, 1.8x or 2.0x can be
used. The magnification is determined by the magnification stand. The
higher the factor is the larger the object is on the film but the area imaged
from the breast is smaller.
The magnification studies always use the smaller film format, 18x24. Both
cassette types, normal cassette and daylight cassette can be used.
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
breast which gets imaged or by applying compression only to a specific
area, spot compression.
In the whole field imaging a larger area can be studied from one image,
for example when it is not quite sure where the area of interest is in the
breast.
It also compresses the breast to thinner from this region, thereby allowing
the use of lower kV-setting resulting in a better image quality. Positioning
using the spot compression is more difficult though.
When using the whole film area the magnification diaphragm on the left
has to be used. When using the spot compression the same diaphragm
can be used or the specific spot diaphragm, on the right, can used.
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
contrast and also the patient dose.
The paddle can be used with a standard diaphragm which exposes the
whole film or with the special diaphragm which exposes only the
compressed part of the breast.
NOTE!
If the breast does not cover the detector or the breast has an implant,
manual technique must be selected.
3.9.3 Biopsy
1 Select the image size required and the corresponding diaphragm.
2 Attach the biopsy compression paddle.
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
3 Clean the compression paddle and breast area carefully. If the Laser
guide is used, read the separate operation instructions.
CAUTION!
Do not use automatic release. Press AutoRel so that the indicator is “off”.
NOTE!
In case the patient moves in the middle of a procedure or compression
thickness is changed, in order to guarantee the accuracy of the biopsy,
new image should be taken or the procedure should be started from the
beginning.
9 Develop the film. The holes in the compression paddle can be seen
on the image, and localization of the biopsy target can be made.
Perform the biopsy and release the compression.
Biopsy paddles are used in manual biopsy for Fine Needle Aspiration (FNA)
or lesion marking.
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
steel wire or with carbon or a dye (color).
The surgeon then follows either the wire
or color trail to remove the correct lesion.
The openings from the paddle can be seen in the image and with their
help the location of the lesion in this plane can be estimated. How deep in
the breast the lesion is can be estimated from images that have been
taken using another projection. These can be taken earlier or during the
same procedure.
NOTE!
Continued tapping of the footswitch is improper usage of the automatic
compression. Use the light switch to turn the collimator light on.
The needle is inserted into the breast after careful estimation and another
There are three kinds of biopsy paddles available with different openings,
holes with diameter of 18mm, holes with diameter of 9mm and a square
opening of size 50x90mm. Which one to use is decided by the doctor’s
preference.
For localization use the biopsy paddle can have a fast release arm which
makes it easier to remove the paddle while the localizing wire is left in the
breast.
When inserting the needle into the breast the cross hair diaphragm can
be used to help the insertion.
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
light onto the breast.
NOTE!
Release compression manually so that the hub of the needle does not
catch on the localizer paddle.
The spot of the lesion is marked with steel wire. It is recommended to use
the fast release needle localization paddles. The location of the wire end is
checked with a new exposure. If needed, the location of the wire is
changed and controlled with another exposure. Fix the control wire to the
breast with tape.
Spot compression
paddle (straight arm)
Normally the whole film is exposed to make it easier to read on the light
box. However, if desired the x-ray field can be collimated to a smaller
area. This reduces the patient dose because only a part of the breast gets
radiation. Also, the contrast is improved because the amount of scatter
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
radiation is reduced. However, films exposed this way are more difficult to
read because of the large unexposed area.
NOTE!
Care must be taken not to overcompress the breast and risk injury to the
patient.
GE Healthcare
37094-IMG rev 3
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
4 Integrated Quality Control test
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
The Performa has the Integrated Quality Control test to maintain the
optical film densities the same even if the film processor conditions have
changed.
NOTE!
The IQC-mode can be entered from the Performa by first selecting the
Auto kV mode (AA) or Semi AEC mode (A) and then pressing the +mAs-
button for three seconds in the Performa control panel.
NOTE!
The compression paddle drives automatically to the upper limit. This is to
prevent exposures to a patient.
The timer display indicates iqc for one second. Then for about two
seconds the correction factor from the previous IQC procedure is shown
and thereafter the text iqc shows up again.
The mAs display shows the optical density value that is followed either the
sensitometer method or the test exposure method.
NOTE!
If the compression paddle is not in the upper limit, the exposures are
inhibited, the iqc display starts to blink and paddle drives up again.
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
2 Read the optical density from the step that has been selected during
the installation.
3 Select Auto kV mode (AA) or Semi AEC mode (A). Press the +mAs
button for three seconds. Enter the measured optical density by
using the +/- density buttons. Press Autoexp-button to enter the
value.
The time display shows the calculated correction whereafter the system
returns to normal mode and is ready for use.
The mAs display in the Performa control panel shows the text “P.HA” to
indicate the use of a phantom.
Place the phantom onto the cassette holder (Bucky). Having the acrylic
overlapping makes sure that no stray radiation hits the detector.
1 Acrylic/Mammographic phantom
2 Cassette holder top
3 Cassette
4 Image receptor
5 Detector
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
phantom may have moved inwards
from the cassette holder edge. Check
also the placement of the detector.
Test exposure
Feed the measured optical density to the unit with the +/- DENSITY
buttons. The Performa unit selects 1.30 as default value, but the value can
be reprogrammed by the service technician.
After the IQC test, verify that the phantom image O.D. lies within the
country-specific limits, e.g. 1.60 ±0.15.
GE Healthcare
37094-IMG rev 3
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
5 Problem situations
5 Problem situations
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
5.1 NO POWER INDICATION
A Check that power cord and display cable are connected.
B Check that the mains switch is switched on.
C Check the fuses or the circuit breaker
D If not cleared, call service.
Check codes indicated by C.xx. Check codes indicate conditions that can
be handled by the operator.
Error codes indicated by E.xx. Most error codes require a call to the
service personnel.
All possible conditions and the action required by them are listed on
tables 5.1, Performa check code guide, 5.2, Performa error code guide and
5.3, Performa AEC adjustment quick fix guide.
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
Check code in Reason for error Corrective action
timer display
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
C.bU The bucky grid is not in Switch the unit off and
starting position. on, re-attach the bucky.
If the code re-occurs
call for service.
C.Cn The bucky grid is not Switch the unit off and
moving properly. on, re-attach the bucky.
If the code re-occurs
call for service.
C.ES The exposure switch Switch off the power.
was pressed when the Check the exposure
power was turned on. switch and turn on the
power.
C.PS IQC system limit Call for film processor
reached. service.
C.br Brake released during Check that patient has
exposure. not pressed the brake
release button.
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
Error code in Reason for error Corrective action
timer display
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
AEC adjustment quick fix guide
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
Film:Kodak Min R 2000, Screen: Kodak Min R 2000 and Min R 2190
Film: Fuji AD-M, Screen: Fuji AD Mammo Fine
Film: Agfa HDR and HDR-C, Screen: Agfa HD and HD-S
< 3 cm 23 Mo 0
4 - 5 cm 24 - 26 Mo 0
6 - 7 cm 27 - 29 Mo or Rh* 0
> 8 cm 30 - 32 Mo or Rh* 0
Small breasts 16 - 40 23 - 25
NOTE!
Implant displaced views and specimens may be phototimed if sufficient
breast tissue covers the photocell closest to the chest wall.
Comments:
1 Maintain exposure times between 0.5 and 2.5 seconds. Decrease kVp
if times are less than 0.5 seconds and increase if they are more than
2.5 seconds.
2 If exposure times are within indicated times, use + or - density
settings to achieve optimal film density.
3 Your technique may vary from those indicated above depending on
film density preferences, processor or film irregularity, or different
film/screen combinations.
4 *Rhodium - use for very dense breasts, 6 cm or greater. No need to
increase kV.
5 Magnification views - increase technique by 2 - 3 kVp, if exposure
time exceeds 5 seconds.
6 Specifications
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
6.1 SYSTEM SPECIFICATION MODEL MGF-110
NOTE!
The Performa is a cord-connected device, delivered without a plug (in
some markets). Depending upon local electrical code requirements, it
may be either hard-wired or plugged into an outlet. Please consult your
electrician to determine the requirements, and if necessary to provide a
plug which matches your outlet.
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
Enclosure IP 20
Weight 265 kg (584 lbs.)
Height 1.9 m (6.23 ft)
Operation mode Continuous operation with
intermittent loading
Radiation intensity Typical range 1000-1200mR/
100mA at 28kV
Environmental specifications
Generator (THF)
Generator (THF)
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
(6s U.K.)
Focus 0.1mm 0.05 - 10 s
Tolerance: ± 0.10 + instrument
inaccuracy
Performa Filtration
Molybdenum 0.03 mm
Positioning
Compression
Recording systems
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
Contact for 18 x 24 and 24 x 30cm film
Bucky for 18 x 24 and 24 x 30cm film
Magnification for 18 x 24cm film 1.6, 1.8 & 2.0
exposed size 10 x 24
Biopsy for 18 x 24cm film
Attenuation equivalent
ROC Grid
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
Anode type Rotating dual-angle anode
Anode heat capacity 300,000 HU
Target material Molybdenum alloy, niobium
doped
Anode rotation speed 3000rpm(50Hz)/3600rpm(60Hz)
Focal spot sizes M113SP 0.1, 0.3 (IEC60336)
Port material beryllium
Reference axis 42 mm (M113SP) from chest wall
edge (tube angle 4°)
1 1 Focal spot is marked in the front part of the tube head assembly.
2 The collimator label is at the front end of the tube head assembly.
2
Measurement criteria for exposure time
Fuses
Supply Board
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
Filament Board
Inverter Board
Maximum operating voltage: 35 kVp effective focal spot 0.3 mm for large
spot 0.1 mm for small focal spot.
X-ray tube: Varian M113SP 0.3/0.1. For additional information refer to tube
specification sheets.
±4 mAs for mAs values less than or equal to 40 mAs, ±10% for mAs values
greater than 40 mAs.
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
Power Supply Requirements:
Line voltage 230 VAC ±10%, 50/60 Hz, single phase. Operating Voltage
Range 198-242 (±10%) Automatic regulation for all voltages within
Operating Voltage Range.
Anode:
Heating = kV x mAs = Joule (increased only during exposures)
Cooling = 30 Joule / sec (decreased continuously)
Housing:
Heating = kV x mAs = Joule (increased only during exposures)
Cooling = refer to diag. below (decreased continuously)
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
kV M113SP mA M113SP
(SF) (sw 7.7x) mA
(LF)
20 25 75
21 26 80
22 28 85
23 30 90
24 29 95
25 28 100
26 27 95
27 26 90
28 25 85
29 24 85
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
The Performa is suitable for use in the specified electromagnetic environment. The purchaser or user
of the Performa should assure that it is used in an electromagnetic environment as described below:
Radio-Frequency Group 1 The Performa uses RF energy only for its internal function.
Emissions Therefore, the RF emission is very low and not likely to cause
CISPR11 any interference in nearby electronic equipment.
Radio-Frequency Class B The Performa is suitable for use in all establishments,
Emissions including domestic establishments and those directly
CISPR11 connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions Not The Performa is suitable for use in all establishments,
applicable including domestic establishments and those directly
IEC 61000-3-2 connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Voltage Not The Performa is suitable for use in all establishments,
fluctuations/ applicable including domestic establishments and those directly
flicker emissions connected to the public low-voltage power supply network
The Performa is suitable for use in the specified electromagnetic environment. The purchaser or user
of the Performa should assure that it is used in an electromagnetic environment as described below:
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
Immunity Test IEC 60601-1-2 Compliance Electromagnetic Environment
Test Level Level
The Performa is suitable for use in the specified electromagnetic environment. The purchaser or user
of the Performa should assure that it is used in an electromagnetic environment as described below:
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
Immunity IEC Compliance Electromagnetic
Test 60601-1-2 Level Environment
Test Level
*Field strengths from fixed transmitters, such as base stations for cellular telephones and land mobile
radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be estimated accurately.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be performed. If the measured field strength exceeds the RF compliance level above,
observe the Performa to verify normal operation in each use location. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the Performa.
**Over the frequency range 150 kHz to 80 MHz, field strengths are less than [V1] V/m.
The Recommended Separation Distances are listed in the next table.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
Table 6.4 RF immunity of non-life-support equipment or system IEC 60601-1-2
Recommended Separation Distances for Portable and Mobile RF Communications Equipment IEC
60601-1-2
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
Frequency of 150KHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,5 GHz
Transmitter
3,5 3,5 7
Equation d =[ ] P d =[ ] P d =[ ] P
V1 E1 E1
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
7.1 M113SP TUBE DATA SHEET
Various tube charts and markings are included in the latest revision of
5708, M113SP Rotating Anode Tube by Varian Medical Systems.
GE Healthcare
ANODE HEATING AND COOLING CURVES
37094-IMG rev 3
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
37094-IMG rev 3
7.5
7.5.1
GE Healthcare
SINGLE LOAD RATINGS
3 Ø Constant Potential 50 Hz
61
7 M113SP X-ray tube data sheets
3 Ø Constant Potential 60 Hz
GE Healthcare
37094-IMG rev 3
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
7 M113SP X-ray tube data sheets
7.6 DESCRIPTION
The M113SP rotating anode X-ray tube is a 39 kVp VARIAN insert tube with
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
0.1 mm and 0.3 mm focal spot combination and 77 mm diameter bi-
angular (10°/16°) Molybdenum anode disc, cathode grounded.
The M113SP metal center section X-ray tube is for use in Mammography
systems. The use of a metal center section tube eliminates arcing due to
metal deposition on glass and allows use of a Beryllium window in the
insert for minimum inherent filtration.
Large: 0.3 mm
Maximum Anode Load: 3-phase, 180Hz, 0.1 second
Small Focal Spot: 1320 W
Large Focal Spot: 4.9 kW
Class: I
Type: B
Maximum Peak Voltage: 52 kV
Anode to Ground: 52 kV
Cathode to Ground: 0 kV
Housing Heat Storage Capacity: 370 KJ (500 kHU)
Housing Cooling Rate: 100 Watts (135HU/sec)
Weight: 10.2 kg (22.5 lbs)
Length: 33 cm (13”)
Diameter: 13.8 cm (5.44”)
Focal Spot - Port Spacing: Small 47.0mm +/- 1mm
Large 41.4 +/- 1mm
High Voltage Cable, Special Design 55 kV Rating
(Non-Federal Standard):
Acoustic noise: 50dB Max. (A-weighted) at 30 cm
Leakage Technique Factor M113SP: 52 kV, 2mA
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
Maximum Operation Temperature: +40°C (+104°F)
Low Voltage Cable:
Pin Position Function
1 Phase Shift
2 Phase
3 Common
4 Housing Ground
5 Grid
6 Not used
7 Small Filament
8 Filament Common
9 Large Filament
7.9 FILAMENTS
Amperes 4 8
Seconds to 1 0.5
start
Duty Intermittent Intermittent
Running 45 V 90 V
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
operation of the X-ray tube to 39 kV, a minimum of 0.5mm Aluminum
equivalent must be added to the X-ray beam.
7.12 RAYPROOFING
Stray radiation complies with the requirements of FDA 21 CFR, Subchapter
J for medical diagnostic-type protective X-ray tube housings.
Measured at 25kV, 80mA with one degree star pattern or a slit focal.
Maximum focal spot size = 0.45W x 0.58L (mm).
GE Healthcare
37094-IMG rev 3
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
8 System maintenance
8 System maintenance
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
8.1 CLEANING
WARNING!
Before cleaning and disinfecting, remove power from the equipment.
ALL parts of the unit which come in contact with the patient must be
cleaned and disinfected prior to each examination.
DO NOT clean surfaces on parts which contact the patient (i.e., face shield,
cassette holders, Bucky devices, compression paddles) with solvent-
based cleaners (containing benzene, alcohol, or stain removers).
As a general rule, disinfectant sprays may not be used directly. The mist
generated by spraying may penetrate into the unit, and this may damage
electronic components. To use a disinfectant agent in spray form, first
apply the agent to a clean cloth or pad, and then use the pad to apply the
agent to the surface requiring disinfectant.
8.2 CARE
Handle the plastic accessories carefully, and inspect them frequently. If
they begin to crack, discontinue their use and replace them.
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
the equipment.
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
and trained service technicians are allowed to carry out the maintenance
on this equipment. Manufacturer can supply at request necessary
documents to Instrumentarium Imaging approved service technician for
repair of such parts that the manufacturer has indicated to be repairable.
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
distributor.
Distributor:
Contact person: Tel:
PERFORMA serial Installation
number *): date:
Last
maintenance
performed:
Exposure counter
number:
Last
maintenance
performed:
Exposure counter
number:
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
Film Processor Daily Thermometer Radiographer/
Densitometer Technologist
Sensitometer
Image Quality Weekly RMI 156D Radiographer/
Phantom Technologist
Physician
Phototimer Monthly Acrylic Radiographer/
Reproducibility phantom Technologist
Physician
Film/Screen Semi-annually Contact test Physicist or
Contact tool Service
Technician
kVp Semi-annually Dynalyzer Physicist
Dose/Patient Semi-annually Ion Chamber Physicist
Exposure
Half-Value Semi-annually Aluminum HVL Physicist
NOTE!
Refer to local regulations and instructions concerning the testing of the
unit.
GE Healthcare
37094-IMG rev 3
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
9 Graphical symbols used
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
Radiographic control
Earth (ground)
Type B equipment
Dangerous voltage
ON (Power)
OFF (Power)
GE Healthcare
37094-IMG rev 3
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
FOR TRAINING PURPOSES ONLY!
General Electric Company reserves the right to make changes in specification and features shown herein, or discontinue the product described at any time without notice or obligation. Contact your GE
representative for the most current information.
World Headquarters
GE Healthcare
P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.
European Headquarters
GE Healthcare - Europe:
Paris, France - Fax +33-1-30-70-94-35
Asia Pacific
GE Healthcare -Asia:
Tokyo, Japan - Fax +81-425-85-5490
Hong Kong - Fax: +852-2559-3588
www.gehealthcare.com
Performa