Mamual

The document provides an overview of the Model RF301B, an instrument designed for measuring the power output of electrosurgery units, highlighting its applications, operating instructions, and safety precautions. It includes sections on establishing test criteria, service information, specifications, and calibration checks. The RF301B is noted for its portability and suitability for both technical and non-technical personnel in various settings.

Uploaded by

Rene
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
3 views5 pages

Mamual

The document provides an overview of the Model RF301B, an instrument designed for measuring the power output of electrosurgery units, highlighting its applications, operating instructions, and safety precautions. It includes sections on establishing test criteria, service information, specifications, and calibration checks. The RF301B is noted for its portability and suitability for both technical and non-technical personnel in various settings.

Uploaded by

Rene
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
You are on page 1/ 5

TABLE OF CONTENTS

1.0 INTRODUCTION ••••••••••••••••..••••••....•••••• 1

1.1 Application • •••••••••••••••••••••••••••••• 1

1.2 Establishing Test Criteria. . . . . . . . . . . . . . . 1

1.3 Incorning Inspection • •••••••••••••••••• 1

2.0 DESCRIPTION•••••••••••••••••••••••••••••••••• 2

3.0 OPERATING INSTRUCTIONS. •••••••••••••••••••• 6

3.1 Safety Precautions ----------------------------------- 6

3.2 Ranges • ••••••••••••••••••••..•• 6

3.3 Set-Up Measurernents for Variable Load Output. 6

3.4 Output Measurernent --------------------------------------- 7

3.5 Heavy Load Test ---------------------------------------------- 7

3.6 Low Freguency Output test -------------------------------- 8

3.7 Isolation Check ----------------------------------------------- 8

3.8 Measuring

14

14

the

the

the

the

of

of

Electrosurgery Units.

Electrosurgery Units.

Ground in Milliarnpheres • ••••••

Bipolar Electrosurgery Devices.

of Low Power Electrosurgery

14

16

R.F.
Leakage to

Output of

Power Output

25 watts)

Output . ••..••

Set-Up Measurernents for Variable

Output Measurernent . ••....••••

Heavy Load Test • ........

Low Freguency Output Test.

Isola t ion Check • ......••••••••••••••

Measuring

Measuring

Measuring

Measuring

Measuring the

Devices ( OO

to

Oscilloscope

3 ..44

3.5

3.6

3.7

3.8

3.9

3.10

3.11

3.12

3.13

4.0 THEORY •
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
••••••••••••••••••••••••••••••••• 16
5.0 SERVICE INFORMATION •
••••••••••••••••....••••••••••••••••••••••••••..••••••••••..••••••....•••••••••••••
• 17

5.1

5.2

5.3

Blown Fuse • ••••••••••••••••....

Faulty Therrnocouple • ..••

Warranty Repair • ••••..••

17

17

18

6.0 SPECIFICATIONS .
••..••..••••............••••••••••..••••••..••••....••..••••••........................••••.... 19

7.0 CALIBRATION ANO INSTRUMENTS CHECK


• ..••••..••••••••••..••••••••••••••..••••••••••••••..••••

20

7.1 Calibration Check

. 20

BLOCK DIAGRAM • ..

SCHEMATIC • ••

PARTS LIST • ••••••..••••••••

B-1

S-1

. . . .

P-1

1 ..00

INTRODUCTION

The Model RF301B

is designed to measure the power output of

electrosurgery units. The instrument is accurate and simple to


operate. Test data can be used for calibration,:quality

assurance diagnosis and reference for establishing trends in the

operating performance of electrosurgery units.

1.1

Application

The Model RF301B

can be used by technical and non-technical

personnel to test all electrosurgery units. The proper operation

of an electrosurgery unit and calibration are easily accomplished

using the Model RF301B.

The power output can be recorded for each

setting of the electrosurgery unit on the RF3R

tag and attached to

the front panel so that personnel using the electrosurgery unit

can depend on the consistancy of the power output.

The Model RF301B

can be used by professional people in the

operating room or by engineering and maintenance people during

routine inspections. It is portable and easily transported from

facility to facility. The rugged case makes the Model RF301B

ideal for field service work.

1.2

Establishing Test Criteria

The acceptable output and standard for performance should include

the manufacturer's specifications with respect to power output at

a specified load a nnddaa specified frequency. The manufacturer's

specifications are bench marks to use in establ1shing acceptable

performances.

Safety standards should be established for electrosurgery units as

well as all other devices used in the hospital. (See electrical

Safety Prograro Guide; contact the factory.)


1.3

Incoming Inspection

Check the shipping carton fer the following accessories which are

shipped with every model.

red 5' lead with banana plug on both encts

black 5' lead as above

1 white 5' lead as above

green 5' leac as above

1 large clamp with banana jack

3 alligator clips

100 RF3R

inspection labels

2 instruction manuals

ground adapter

5 fuses