Brief Operating Instructions
Proline t-mass 65
Thermal Mass Flow Measuring System
6
8
9
MODBUS
These Brief Operating Instructions are not intended to replace the
Operating Instructions provided in the scope of supply.
Detailed information is provided in the Operating Instructions and
the additional documentation on the CD-ROM supplied.
The complete device documentation consists of:
 These Brief Operating Instructions
 Depending on the device version:
 Operating Instructions and the Description of Device Functions
 Approvals and safety certificates
 Special safety instructions in accordance with the approvals for
the device (e.g. explosion protection, pressure equipment
directive etc.)
 Additional device-specific information
KA021D/06/en/06.10
71115118
Table of contents
Proline t-mass 65
Table of contents
1 Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.1 Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.2 Installation, commissioning and operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3 Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.4 Safety conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.1 Transporting to the measuring point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2 Installation conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.3 Installation instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.4 Post-installation check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3 Wiring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.1 Connecting the various housing types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.2 Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.3 Post-connection check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4 Hardware settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
4.1 Device address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
4.2 Terminating resistors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
5 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.1 Switching on the measuring device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.2 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.3 Navigating within the function matrix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.4 Calling the Commission Quick Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
5.5 Software settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5.6 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Endress+Hauser
Proline t-mass 65
Safety instructions
1.1
Designated use
Safety instructions
 The measuring device works based on the thermal measuring principle and is to be used
exclusively for measuring the mass flow of gases (e.g. kg, Nm3, Scf).
 The measuring device can be configured to measure pure gases (e.g. oxygen, nitrogen, carbon
dioxide, argon etc.) or gas mixtures (e.g. natural gas).
 Any use other than that described here compromises the safety of persons and the entire
measuring system and is, therefore, not permitted.
 The manufacturer is not liable for damage caused by improper or non-designated use.
1.2
Installation, commissioning and operation
 The measuring device must only be installed, connected, commissioned and maintained by
qualified and authorized specialists (e.g. electrical technicians) in full compliance with the
instructions in these Brief Operating Instructions, the applicable norms, legal regulations and
certificates (depending on the application).
 The specialists must have read and understood these Brief Operating Instructions and must
follow the instructions they contain. If you are unclear on anything in these Brief Operating
Instructions, you must read the Operating Instructions (on the CD-ROM). The Operating
Instructions provide detailed information on the measuring device.
 The measuring device should only installed in a de-energized state free from outside loads or
strain.
 The measuring device may only be modified if such work is expressly permitted in the
Operating Instructions (on the CD-ROM).
 Repairs may only be performed if a genuine spare parts kit is available and this repair work is
expressly permitted.
 If performing welding work on the piping, the welding unit may not be grounded by means
of the measuring device.
1.3
Operational safety
 The measuring device is designed to meet state-of-the-art safety requirements, has been
tested, and left the factory in a condition in which it is safe to operate. Relevant regulations
and European standards have been observed.
 The manufacturer reserves the right to modify technical data without prior notice. Your
Endress+Hauser distributor will supply you with current information and updates to these
Operating Instructions.
 The information on the warning notices, nameplates and connection diagrams affixed to the
device must be observed. These contain important data on the permitted operating conditions,
the range of application of the device and information on the materials used.
 If the device is not used at atmospheric temperatures, compliance with the relevant marginal
conditions as specified in the device documentation supplied (on CD-ROM) is mandatory.
Endress+Hauser
Safety instructions
Proline t-mass 65
 The device must be wired as specified in the wiring and connection diagrams. Interconnection
must be permitted.
 All parts of the device must be included in the potential equalization of the system.
 Cables, certified cable glands and certified dummy plugs must be suitable to withstand the
prevailing operating conditions, such as the temperature range of the process. Housing
apertures that are not used must be sealed with dummy plugs.
 The device should only be used for fluids to which all the wetted parts of the device are
sufficiently resistant. With regard to special fluids, including fluids used for cleaning,
Endress+Hauser will be happy to assist in clarifying the corrosion-resistant properties of
wetted materials. However, minor changes in temperature, concentration or in the degree of
contamination in the process may result in variations in corrosion resistance. For this reason,
Endress+Hauser does not accept any responsibility with regard to the corrosion resistance of
wetted materials in a specific application. The user is responsible for the choice of suitable
wetted materials in the process.
 Hazardous areas
Measuring devices for use in hazardous areas are labeled accordingly on the nameplate.
Relevant national regulations must be observed when operating the device in hazardous areas.
The Ex documentation on the CD-ROM is an integral part of the entire device documentation.
The installation regulations, connection data and safety instructions provided in the Ex
documentation must be observed. The symbol on the front page indicates the approval and
certification body (0 Europe, 2 USA, 1 Canada).
The nameplate also bears the documentation number of this Ex documentation
(XA***D/../..).
1.4
#
"
!
Safety conventions
Warning!
"Warning" indicates an action or procedure which, if not performed correctly, can result in injury
or a safety hazard. Comply strictly with the instructions and proceed with care.
Caution!
Caution indicates an action or procedure which, if not performed correctly, can result in
incorrect operation or destruction of the device. Comply strictly with the instructions.
Note!
"Note" indicates an action or procedure which, if not performed correctly, can have an indirect
effect on operation or trigger an unexpected response on the part of the device.
Endress+Hauser
Proline t-mass 65
Installation
Installation
2.1
Transporting to the measuring point
 Transport the measuring device to the measuring point in the original packaging.
 The covers or caps fitted on the process connections prevent mechanical damage to the
sensors during transport and storage. For this reason, do not remove the covers or caps until
immediately before installation.
Use slings around the process connections.
Warning!
#
Risk of injury! The device can slip.
The center of gravity of the measuring device may be
higher than the holding points of the slings.
Always ensure that the device cannot slip or turn around
its axis.
A0007408
Do not lift measuring devices by the transmitter housing
or the connection housing in the case of the remote
version. Do not use chains as they could damage the
housing.
A0007409
2.2
Installation conditions
 The thermal dispersion principle is sensitive to disturbed flow conditions. Therefore, the
requirements and conditions for installation in this section are especially important
 Take measures to reduce or avoid condensation (e.g. install a condensation trap, thermal
insulation, etc.).
2.2.1 Dimensions
For the dimensions of the measuring device  see associated Technical Information on the
CD-ROM.
Endress+Hauser
Installation
Proline t-mass 65
2.2.2 Pipework requirements
Good engineering practice should be followed at all times.
Further information is provided in ISO Standard 14511.
Note!
Measuring errors will result due to a mismatch of the pipes or gaskets.
Avoid
A0007520
Pipe diameter one is not equal to pipe
diameter two
A0007521
Incorrectly sized gaskets
A0007522
Incorrectly aligned Flanges and
gaskets
2.2.3 Orientation
Make sure that the direction arrow on the sensor matches the direction of gas flow through the
pipe.
Flanged version
A0007514
A0007512
A0007515
compact
remote
compact
remote
compact
remote
Insertion version
A0007516
A0007517
A0007518
compact
remote
compact
remote
compact
remote
mn
 = Recommended orientation
 = Orientation recommended in certain situations
mo = See following description
Endress+Hauser
Proline t-mass 65
Installation
m In the case of saturated or unclean gases, upward flow in a vertical pipe section is preferred to minimize
condensation/contamination.
n Not recommended if the vibrations are too high or if the installation is unstable.
o Only suitable for clean/dry gases. On horizontal pipes do not mount the sensor from the bottom, if the gas is very damp
or saturated with water (e. g. Bio Gas). Mount the sensor in a position as indicated below ( =135).
A0009897
2.2.4 Inlet/outlet runs
The thermal dispersion principle is sensitive to disturbed flow conditions.
As a general rule, the installed thermal flow sensor should always be installed as far away as
possible from any flow disturbances. Further information is provided in ISO Standard 14511.
Flanged version
15 x DN
2 x DN
15 x DN
A0007523
20 x DN
2 x DN
A0007526
Endress+Hauser
2 x DN
15 x DN
A0007524
35 x DN
2 x DN
A0007527
2 x DN
A0007525
50 x DN
2 x DN
A0007528
Installation
Proline t-mass 65
Insertion version
20 x DN
5 x DN
20 x DN
A0007529
25 x DN
5 x DN
5 x DN
20 x DN
5 x DN
A0007530
40 x DN
A0007532
5 x DN
A0007531
50 x DN
5 x DN
A0007564
A0007534
Inlet run
15 x DN for the flanged version and 20 x DN for the insertion version
Outlet run
2 x DN for the flanged version and 5 x DN for the insertion version
Note!
 Where two or more flow disturbances are located upstream of the meter, the longest indicated
inlet length should prevail.
For example if a control valve and a bend are mounted upstream of the flowmeter, the
recommended inlet length should be according to the control valve (50  DN).
 For very light gases such as Helium and Hydrogen all upstream distances should be doubled.
 A specially designed perforated plate flow conditioner can be installed if it is not possible to
observe the inlet runs required.
Please refer to the Operating Instructions on the CD-ROM for special details on the perforated
plate flow conditioner
2.2.5 Mounting conditions for the insertion version
Mounting conditions for welding sockets
"
Caution!
When mounting the fitting to a thin wall duct, use a
suitable support bracket for the sensor.
D
90
90
D =  31.0 mm  0.5 mm ( 1.22"  0.019")
A0011843
Endress+Hauser
Proline t-mass 65
Installation
Adjustment of the insertion version
Insert the sensor into the socket and tighten the lower nut of
compression fitting first by hand and then tighten it 1 revolutions
using a wrench (42 mm).
Caution!
"
 NPT thread: use a thread sealing tape or paste
 G1A thread: the sealing ring supplied must be installed
A0011888
Insertion depth calculation
Note!
All the guidelines and information on the insertion depth refer to a standard welding nozzle
provided by Endress+Hauser.
For detailed remarks on calculation refer to Technical Information on the CD-ROM.
a.
9
230
220
210
200
190
b.
Thickness of the pipe wall or duct wall
c.
Depth of the welding nozzle at the pipe or duct including the
sensor pipe union and Hot tap mounting set (if used).
a.
Calculated insertion depth:
(0.3  a) + b + c + 2 mm or
(0.3  a) + b + c + 0,079 inch
180
Internal diameter for round pipes.
Duct height for a duct if the sensor is to be installed vertically
or the duct width if it is to be installed horizontally.
(a = min. 80 mm or 3 inch)
A0011844
Insertion depth calculation with Quick Setup "Sensor"
The following data are needed for insertion depth calculation:
Round pipes
Duct
Pipe standard (DIN, ANSI or others)
Nominal diameter
Outer diameter
Thickness
Internal diameter (min. 80 mm or 3")
Endress+Hauser
Duct height
Duct width
Thickness
Mounting orientation (installed vertically or
horizontally)
Installation
Proline t-mass 65
Aligning the calculated insertion depth
a.
Tighten the upper nut of compression fitting such that
the sensor can still be adjusted.
b.
Align the scale to the calculated insertion depth.
230
220
210
200
190
180
A0011848
Aligning the insertion version to flow direction
90 (3)
90(7)
A0007536
A0007537
Check and ensure that the sensor is aligned vertically at a 90 angle on the pipe/duct.
Turn the sensor so that the arrow marking matches the direction of flow.
10
Endress+Hauser
Proline t-mass 65
Installation
To fix the insertion version
1.
Tighten the compression fitting (1) by
hand to secure the position of the sensor.
Then, using an open-ended wrench
(36 mm), tighten another 1 revolutions
in a clockwise direction.
2.
Fix the two securing screws (4)
(Allen key 3 mm; 1/8").
Warning!
#
Observe torque:
4 Nm (2.95 lbf ft)
2
3.
Check that the sensor and transmitter do
not turn.
4.
Check the measuring point for leaks at
the maximum operating pressure.
a0010114
2.2.6 Heating
For information on the heating, please see the Operating Instructions on the CD-ROM.
2.2.7 Thermal insulation
For information on the thermal insulation, please see the Operating Instructions on the
CD-ROM.
2.2.8 Vibrations
"
Caution!
Excessive vibration can result in mechanical damage to the measuring device and its mounting.
For information on vibrations, please see the Operating Instructions on the CD-ROM.
Endress+Hauser
11
Installation
Proline t-mass 65
2.3
Installation instructions
2.3.1 Turning the aluminum field housing
Aluminum field housing for non-Ex area
b
 180
 180
f
A0007540
Aluminum field housing for Zone 1 or Class I Div. 1
max. 360
b
Turn the transmitter housing
gently clockwise until the stop
(end of the thread).
c.
Turn the transmitter
counterclockwise (max. 360)
to the desired position.
d.
Retighten the setscrew.
Nicht-eigensichere
Stromkreise durch
IP40-Abdeckung geschtzt
p cover tight w
Kee ircuits are ali hile
ve
c
p cover tight wh
Kee ircuits are alive ile
c
Release the setscrew.
b.
er Spannu
ht unt
ng
Nic
ffnen
Non-intrinsically safe
circuits Ip40 protected
Boucles de courant
sans scurit intrinsque
protges par Ip40
Ne
a.
uvrir l
pas o te apparei
l
sous nsion
A0008036
2.3.2 Turning the local display
4 x 45
a.
Press in the side latches on the display module and
remove the module from the cover plate of the
electronics compartment.
b.
Turn the display to the desired position (max. 4 x
45 in both directions) and reset it onto the cover
plate of the electronics compartment.
a
A0007541
12
Endress+Hauser
Proline t-mass 65
Installation
2.3.3 Installing the wall-mount housing
"
Caution!
 Make sure that the ambient temperature does not exceed the permitted range.
 Always install the wall-mount housing in such a way that the cable entries point downwards.
Mounted directly on the wall
35 (1.38)
Connection compartment
2.
Securing screws M6 (max.  6.5 mm (0.25 inch);
screw head max.  10.5 mm (0.4 inch)
3.
Housing bores for securing screws
1
90 (3.54)
1.
81.5 (3.2)
mm (inch)
192 (7.56)
A0007542
Pipe mounting
"
 2070
( 0.792.75)
Caution!
mm (inch) Danger of overheating! If the device is mounted on a
warm pipe, make sure that the housing temperature does
not exceed +60 C (+140 F) which is the maximum
temperature permitted.
~155 (~ 6.1)
A0007543
Endress+Hauser
13
Installation
Proline t-mass 65
Panel mounting
+0.5 (+0.019)
0.5 (0.019)
210 (8.27)
+0.5 (+0.019)
0.5 (0.019)
245 (9.65)
mm (inch)
~110 (~4.33)
A0007544
2.4
14
Post-installation check
Is the measuring device damaged (visual inspection)?
Does the device correspond to specifications at the measuring point?
Is the serial number of sensor and the connected transmitter the same?
Are the measuring point number and labeling correct (visual inspection)?
Professional installation (correct pipe internal diameter, correctly sized gaskets)?
Is the alignment of pipe/gasket/flowmeter body correct?
Has the correct sensor orientation been selected in terms of type, fluid properties, fluid
temperature?
Does the arrow on the sensor point in the direction of the flow in the pipe?
Are sufficient inlet and outlet runs available before and after the measuring point?
Is flow conditioner correctly installed (if available)?
Is the sensor immersion depth (only insertion version) correct?
Is the measuring device protected against moisture and sunlight?
Is the measuring device protected against overheating?
Is the measuring device protected against excessive vibrations?
Have the gas conditions been checked for purity, cleanliness and dryness?
Endress+Hauser
Proline t-mass 65
"
Wiring
Wiring
Warning!
Risk of electric shock!
 Never mount or wire the measuring device while it is connected to the power supply.
 Prior to connecting the power supply, connect the protective earth to the ground terminal on
the housing.
Caution!
Risk of damaging the electronic components!
Connect the power supply in accordance with the specifications on the diagram inside the
connection compartments cover.
Additionally for the remote version:
"
Caution!
Risk of damaging the electronic components!
 Maximum cable length: 100 m (328 ft)
 Observe the cable specifications of the connecting cable  Operating Instructions on the
CD-ROM.
Note!
Install the connecting cable securely to prevent movement.
Additionally for measuring devices with fieldbus communication:
"
Caution!
Risk of damaging the electronic components!
 Observe the cable specification of the fieldbus cable  Operating Instructions on the
CD-ROM.
 Keep the stripped and twisted lengths of cable shield as short as possible.
 Screen and ground the signal lines  Operating Instructions on the CD-ROM.
 When using in systems without potential matching  Operating Instructions on the
CD-ROM.
Additionally for Ex-certified measuring devices:
Warning!
When wiring Ex-certified measuring devices, all the safety instructions, wiring diagrams,
technical information etc. of the related Ex documentation must be observed
 Ex documentation on the CD-ROM.
Endress+Hauser
15
Wiring
Proline t-mass 65
3.1
Connecting the various housing types
Wire the unit using the terminal assignment diagram inside the cover.
3.1.1 Compact version (transmitter)
Transmitter connection:
1
2
3
4
3
4
2
Connection diagram inside the connection compartment cover
Power supply cable
Signal cable or fieldbus cable
Optional
1
A0007545
3.1.2 Remote version (transmitter)
Transmitter connection:
1
2
3
4
Connection diagram inside the connection compartment cover
Power supply cable
Signal cable
Fieldbus cable
Connecting cable connection:
2 3 4 5
Sensor/transmitter connecting cable
A0007546
3.1.3 Remote version (transmitter)
Transmitter connection:
3
4
2
1
2
3/4
p cover tight w
Kee ircuits are ali hile
ve
c
over tight whi
ep c
Ke ircuits are alive le
c
ht unter Spannun
Nic
g
ffnen
Connection diagram inside the connection compartment cover
Power supply cable
Signal cable or fieldbus cable
vri
pas ou r lappa
Ne sous tension reil
1
Connecting cable connection:
Sensor/transmitter connecting cable
A0007547
3.1.4 Remote version (sensor)
Transmitter connection:
1
Connection diagram inside the connection compartment cover
1
Connecting cable connection:
5
Sensor/transmitter connecting cable
A0007548
16
Endress+Hauser
Proline t-mass 65
3.2
Wiring
Degree of protection
The devices fulfill all the requirements for IP 67 (NEMA 4X).
After mounting in the field or service work, the following points have to be observed to ensure
that IP 67 (NEMA 4X) protection is retained:
 Install the measuring device in such a way that the cable entries do not point upwards.
 Do not remove the grommet from the cable entry.
 Remove all unused cable entries and insert blanking or certified plugs instead.
 Use cable entries and drain plugs with a long-term operating temperature range in accordance
with the temperature specified on the nameplate.
A0007549
Tighten the cable entries correctly.
3.3
A0007550
The cables must loop down before they enter the cable
entries ("water trap").
Post-connection check
Are cables or is the device damaged (visual inspection)?
Does the supply voltage match the specifications on the nameplate?
Are the power supply and signal cables correctly connected?
Do the cables used comply with the necessary specifications?
Do the mounted cables have adequate strain relief?
Cables correctly segregated by type? Without loops and crossovers?
Are all screw terminals firmly tightened?
Are all the cable entries installed, correctly tightened and firmly sealed?
Cables looped as "water traps"?
Are all housing covers installed and correctly tightened?
In addition, for measuring devices with fieldbus communication:
 Are all the connecting components (T-boxes, junction boxes, connectors, etc.) connected
with each other correctly?
 Has each fieldbus segment been terminated at both ends with a bus terminator?
 Has the max. length of the fieldbus cable and the spurs been observed in accordance with the
specifications?
 Is the fieldbus cable fully shielded and correctly grounded?
Endress+Hauser
17
Hardware settings
Proline t-mass 65
Hardware settings
This section only deals with the hardware settings needed for commissioning. All other settings
(e.g. output configuration, write protection, etc.) are described in the associated Operating
Instructions on the CD-ROM.
Note!
No hardware settings are needed for measuring devices with HART or FOUNDATION Fieldbus
type communication.
4.1
Device address
Has to be set for measuring devices with the following communication methods:
 PROFIBUS DP/PA
 MODBUS RS485
The device address can be configured via:
 Miniature switches  see description below
 Local operation  see software settings section
Addressing via miniature switches
Warning!
Risk of electric shock! Risk of damaging the electronic components!
 All the safety instructions for the measuring device must be observed and all the warnings
heeded   3.
 ESD (Electrostatic Discharge): Use a workspace, working environment and tools purposely
designed for electrostatic sensitive devices.
Warning!
Switch off the power supply before opening the device.
W
E
N
O
1
2
3
4
Loosen the cheese head screw of the securing clamp
with an Allen key (3 mm)
b.
Unscrew cover of the electronics compartment from
the transmitter housing.
c.
Loosen the securing screws of the display module
and remove the onsite display (if present).
d.
Set the position of the miniature switches on the
I/O board using a sharp pointed object.
Installation is the reverse of the removal procedure.
W
E
N
O
1
2
3
4
W
E
N
O
1
2
3
4
W
E
N
O
1
2
3
4
a.
W
E
N
O
1
2
3
4
W
E
N
O
1
2
3
4
A0007551
18
Endress+Hauser
Proline t-mass 65
Hardware settings
PROFIBUS
Device address range: 0 to 126
Factory setting: 126
OFF ON
1
16
32
64
a.
Miniature switches for the device address
(example shown: 1+16+32 = device address 49)
b.
Miniature switches for the addressing mode:
OFF = software addressing via local
operation/operating program (factory setting)
ON = hardware addressing via miniature switches
c.
Miniature switch not assigned.
1
2
3
4
OFF ON
A0007552
MODBUS RS485
Device address range: 1 to 247
Factory setting: 247
OFF ON
1
a.
Miniature switches for the device address
(example shown: 1+16+32 = device address 49)
b.
Miniature switches for the addressing mode:
OFF = software addressing via local
operation/operating program (factory setting)
ON = hardware addressing via miniature switches
c.
Miniature switch not assigned.
a
1
16
32
64
128
2
c
3
4
OFF ON
A0007554
Endress+Hauser
19
Hardware settings
4.2
Proline t-mass 65
Terminating resistors
Note!
If the measuring device is used at the end of a bus segment, termination is required.
This can be performed in the measuring device by setting the terminating resistors on the I/O
board. Generally, however, it is recommended to use an external bus terminator and not
perform termination at the measuring device itself.
Has to be set for measuring devices with the following communication methods:
 PROFIBUS DP
 Baudrate  1.5 MBaud  Termination can be performed at the measuring device, see
graphic
 Baudrate > 1.5 MBaud  An external bus terminator must be used
 MODBUS RS485  Termination can be performed at the measuring device, see graphic
Warning!
Risk of electric shock! Risk of damaging the electronic components!
 All the safety instructions for the measuring device must be observed and all the warnings
heeded   3.
 Use a workspace, working environment and tools purposely designed for electrostatically
sensitive devices.
Setting the terminating switch SW1
on the I/O board:
ON - ON - ON - ON
W
E
N
O
1
2
3
4
W
E
N
O
1
2
3
4
W
E
N
O
1
2
3
4
+5V
SW 1
W
E
N
O
1
2
3
4
2
W
E
N
O
1
2
3
4
220 W
W
E
N
O
1
2
3
4
W
E
N
O
1
2
3
4
390 W
3
4
OFF ON
390 W
A0007556
20
Endress+Hauser
Proline t-mass 65
Commissioning
Commissioning
5.1
Switching on the measuring device
On completion of the installation (successful post-installation check), wiring (successful
post-connection check) and after making the necessary hardware settings, where applicable, the
permitted power supply (see nameplate) can be switched on for the measuring device.
When the power supply is switched on, the measuring device performs a number of power-up
checks and device self-checks. As this procedure progresses the following messages can appear
on the local display:
Display examples:
Name of the measuring device
Start-up message
START-UP
RUNNING
Name of the measuring device
Displays the current software
DEVICE SOFTWARE
V XX.XX.XX
SYSTEM OK
Beginning of operation
 OPERATION
The measuring device starts operating as soon as the startup procedure is complete.
Various measured values and/or status variables appear on the display.
Note!
If an error occurs during startup, this is displayed by an error message. The error messages that
occur most frequently when a measuring device is commissioned are described in the
Troubleshooting section   26.
Endress+Hauser
21
Commissioning
5.2
Proline t-mass 65
Operation
5.2.1 Display elements
Display lines/fields
48.25 xx/yy
3702.6 x
1
2
1.
Main line for primary measured values
2.
Additional line for additional measured
variables/status variables
3.
Current measured values
4.
Engineering units/time units
A0011838
5.2.2 Operating elements
Operating keys
Esc
E
2
3
A0007559
1.
() Minus key for entering, selecting
2.
(+) Plus key for entering, selecting
3.
Enter key for calling the function matrix, saving
When the +/ keys are pressed simultaneously (Esc):
 Exit the function matrix step-by-step:
 > 3 sec. = cancel data input and
return to the measured value display
5.2.3 Displaying error messages
1
2
XXXXXXXXXX
#000 00:00:05
3
1.
Type of error:
P = Process error, S = System error
2.
Error message type:
$ = Fault message, ! = Notice message
3.
Error number
4.
Duration of the last error that occurred:
Hours: Minutes: Seconds
5.
Error designation
4
A0007561
 List of the most common error messages during
commissioning,   26
 List of all error messages, see associated Operating
Instructions on the CD-ROM
22
Endress+Hauser
Proline t-mass 65
5.3
Commissioning
Navigating within the function matrix
Esc
Esc
Esc
>3s
E
o
E
q
n
E
E
A0007562
1.
F  Enter the function matrix (starting with measured value display)
2.
P  Select the group (e.g. OPERATION)
F  Confirm selection
3.
N Select function (e.g. LANGUAGE)
4.
P  Enter code 65 (only for the first time you access the function matrix)
F  Confirm entry
P  Change function/selection (e.g. ENGLISH)
F  Confirm selection
5.
Q  Return to measured value display step by step
6.
Q > 3 s  Return immediately to measured value display
Endress+Hauser
23
Commissioning
5.4
Proline t-mass 65
Calling the Commission Quick Setup
All the functions needed for commissioning are called up automatically with the Quick Setup.
The functions can be changed and adapted to the process in question.
1.
F  Enter the function matrix (starting with measured value display)
2.
P  Select the group QUICK SETUP
F  Confirm selection
3.
QUICK SETUP COMMISSION function appears.
4.
Intermediate step if configuration is blocked:
P  Enter the code 65 (confirm with F ) and thus enable configuration
5.
P  Go to Commission Quick Setup
6.
P  Select YES
F  Confirm selection
7.
F  Start Commission Quick Setup
8.
Configure the individual functions/settings:
 Via P-key, select option or enter number
 Via F-key, confirm entry and go to next function
 Via Q-key, return to Quick Setup Commission function
(settings already made are retained)
Note!
Observe the following when performing the Quick Setup:
 Configuration selection: Select the Actual Settings option
 Unit selection: This is not offered again for selection after configuring a unit
 Output selection: This is not offered again for selection after configuring an output
 Automatic configuration of the display: if YES is selected
 Main line = Mass flow
 Additional line = Totalizer 1
 If asked whether additional Quick Setups should be executed: select YES
All the available functions of the measuring device and their configuration options as well as
additional Quick Setups, if available, are described in detail in the "Description of Device
Functions" Operating Instructions. The related Operating Instructions can be found on the
CD-ROM.
The measuring device is ready for operation on completion of the following Quick Setups:
Commission, Sensor (only for 65 I), Gas, Pressure, Heat Flow (if used)
24
Endress+Hauser
Proline t-mass 65
5.5
Commissioning
Software settings
5.5.1 Device address
Has to be set for measuring devices with the following communication methods:
 PROFIBUS DP/PA
Device address range 0 to 126, factory setting 126
 MODBUS RS485
Device address range 1 to 247, factory setting 247
The device address can be configured via:
 Miniature switches  see Hardware settings section
 Local operation  see "Calling the Communication Quick Setup"
Note!
The COMMISSIONING SETUP must be executed before setting the device address.
Calling the Quick Setup Communication
1.
F  Enter the function matrix (starting with measured value display)
2.
P  Select the group QUICK SETUP
F  Confirm selection
3.
N  Select the QUICK SETUP COMMUNICATION function
4.
Intermediate step if configuration is blocked:
P  Enter the code 65 (confirm with F ) and thus enable configuration
5.
P  Go to Quick Setup Communication
6.
P  Select YES
F  Confirm selection
7.
F  Start Quick Setup Communication
8.
P  Select YES
F  Confirm selection
9.
Configure the individual functions/settings:
 Via P-key, select option or enter number
 Via F-key, confirm entry and go to next function
 Via Q-key, return to Quick Setup Commission function
(settings already made are retained)
All the available functions of the measuring device and their configuration options as well as
additional Quick Setups, if available, are described in detail in the "Description of Device
Functions" Operating Instructions. The related Operating Instructions can be found on the
CD-ROM.
The measuring device is ready for operation on completion of the Quick Setup Communication.
Endress+Hauser
25
Commissioning
5.6
Proline t-mass 65
Troubleshooting
The error messages that can occur most frequently when a measuring device is commissioned
are described here.
A complete description of all the error messages  Operating Instructions on the CD-ROM.
HART
No.
Error message / Type
Cause/remedy
351
352
S: RANGE CUR.OUTn
$: # 351352
Current output:
The actual value for the flow lies outside the set limits.
Change full scale value entered or reduce flow.
359
360
S: RANGE PULSEn
$: # 359360
Pulse output: Pulse output frequency is outside the set range.
422
Remedy:
P: FLOW LIMIT
$: # 422
1.
Increase pulse value
2.
When selecting the pulse width, choose a value that can still
be processed by a connected counter.
3.
Reduce flow.
The measured flow has exceeded the maximum limit.
Reduce the flow rate or replace the instrument with a suitable size
for the application.
! Note!
Error can be configured as a fault or notice message.
PROFIBUS
Extended
Cause/remedy
diagnostic
message in the
PROFIBUS
master
Limits
0x13 BAD Sensor
failure
Quality substatus
422 P: FLOW LIMIT
$: # 422
Quality status
No. Device status message
(local display)
Quality code (hex)
Measured value status
PROFIBUS measured value
status
Constant
Meas. flow The measured flow has exceeded the
exceeded max maximum limit.
limit
Reduce the flow rate or replace the
instrument with a suitable size for the
application.
Note!
Error can be configured as a fault or
notice message.
26
Endress+Hauser
Proline t-mass 65
Commissioning
FOUNDATION Fieldbus
No.
Error messages:
FOUNDATION Fieldbus (FF)*
(onsite display)
Analog Input function block
Error messages
Cause/remedy
422
Device status message (FF):
Measured flow exceeds max limit
 Err. No. 422
Onsite display:
P: FLOW LIMIT
!: # 422
OUT. QUALITY = UNCERTAIN
See HART table
OUT. SUBSTATUS = Non-specific
MODBUS RS485
Register 6859 Register 6821
Data type:
Data type:
Integer
String (18 byte)
No. Error message / Type
Cause/remedy
59
422 P: FLOW LIMIT
$: # 422
The measured flow has exceeded the maximum
limit.
FLOW LIMIT
Reduce the flow rate or replace the instrument
with a suitable size for the application.
Note!
Error can be configured as a fault or notice
message.
Endress+Hauser
27
www.endress.com/worldwide
KA021D/06/en/06.10
71115118
FM+SGML 6.0