LOGIQ P5 Premium
TruScan Imaging Technology
Internet: http://www.gemedicalsystems.com
Rev4.0.1
October 13, 2010
Product Description
User Interface
Operator Keyboard
The LOGIQ P5 Premium is a high
performance, highly mobile and easy to
use multipurpose color Doppler
Ultrasound imaging system designed for
Obstetrical, Gynecological, Cardiac,
Musculoskeletal, Vascular, Urological,
Small Parts, Superficial, Pediatric,
Neonatal, Transcranial, Abdominal and
other applications.
 Keyboard Width: 43cm
Keyboard Height: 84cm
 Alphanumeric Keyboard
 Ergonomic Hard Key Operations
 Indicator Lights Identify Activated Keys
 Integrated Recording Keys for Remote
Control of Up to 2 Peripheral Devices
and DICOM Devices
 8 TGC Pods, with Re-mapping
Functionality at Any Depth
System Architecture
TruScan Architecture  GEs
innovative, software-intensive platform
provides exceptional computational
power, image-manipulation capabilities,
real time 4D capabilities, workflow
flexibility and product upgradeability.
TruAccess is the GE advanced, raw-data
processing technology that has helped
transform ultrasound imaging. By
accessing raw data, TruScan achieves
excellent image quality and ensures
exceptional image management
capabilities.
SmartScan utilizes new advances in
operating algorithms and system
operations to aid in improving image
acquisition and patient throughput while
helping to increase diagnostic
confidence and exam consistency.
ComfortScan, our advanced userfriendly designed software suite helps to
maximize productivity. It simplifies every
exam you perform.
General Specification
Dimensions and Weight
Height: 135/141 cm
Width: 43cm
Depth: 64cm
Weight: approx. 75 kg (165 lb.)
LOGIQ P5
GE Medical Systems  Americas: 888-202-5528
P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.
GE Ultrasound - Europe: +49-212-28020
P.O. Box 1105 60 D-42665 Solingen
GE Ultrasound - Asia: 65-291-8528
Singapore: 298 Tiong Bahru Road, #15-01/06
Tiong Bahru Plaza, Singapore 168730
Monitor
Electrical Power
 Voltage: 100-120Vac or 220-240Vac
 Frequency: 50/60 Hz
 Power: Max. 750 VA with Built-in and
On-Board Peripherals
 Maximum Thermal Output: 2200
BTU/hr.
Console Design
3 Active Probe Ports
Integrated HDD (Capacity: 250GB)
Integrated DVD-R/W Drive
On-board Storage for Peripherals (Max
3 peripherals)
Wheels
- Wheel diameter: 12.5cm
- Integrated locking mechanism that
provides rolling lock and optional
caster swivel lock
Probe Holders, Removable for Cleaning
and Washing
Gel Holder, Removable for Cleaning
and Washing
Air Filters, Easily Removable
Front Handle
Rear Handle (Option)
Drawer (Option)
Probe Cable Hanger (Option)
Urology Probe Holder (Option)
DOC0518685
 15 inch TFT LCD
 XGA Format:
- Display size: 1024 x 768
- Recording area: 800 x 600
 Adjustable Tilt/Swivel
 Digital Brightness/Contrast Adjustment
System Overview
Applications
Abdominal
Obstetrical
Gynecological
Cardiac
Musculoskeletal
Vascular
Urological
Small Parts and Superficial
Pediatric and Neonatal
Transcranial
Scanning Methods
Electronic Sector
Electronic Convex
Electronic Linear
Mechanical Volume Sweep
Transducer Types
 Sector Phased Array
 Convex Array
Page 1 of 14
Microconvex Array
Linear Array
Doppler Pencil
Bi-plane Microconvex Array
Real time 4D volume probes
Operating Modes
B-Mode
M-Mode
Crossbeam
SRI
Elastography (option)
Anatomical M Mode (option)
Color Flow Mode (CFM)
Power Doppler Imaging (PDI) with
Directional Map
PW Doppler with High PRF
B-Flow (option)
Steered CW Doppler (option)
Dedicated CW Doppler (option)
Coded Contrast Imaging (option)
3D/4D Volume Modes (option)
System Standard Features
 Hard Disk partition of 50GB for image
storage
 Without compression:
 Raw DICOM: HDD capacity 33.000
images
 DICOM:
- Image Only: HDD capacity 71.000
images
 CINE Memory Frames (60 sec) (192MB)
>1000 depending on probe, imaging
depth, line density and field of view
 Real-time Triplex Mode at any Depth
and PRF
 Automatic Optimization
- Auto Tissue Optimization: ATO
- Auto Spectrum Optimization: ASO
- Auto Color Optimization: ACO
 Coded harmonic Imaging
 Coded Excitation
 Virtual Convex
 Patient Information Database
 Image Archive on Hard Drive and DVD
 Easy Backup for Media
 Report Designer
 Vascular Calcs
 Cardiac Calcs
 OB Calcs
 Fetal Trending
 Multi Gestational Calcs
 Hip Dysplasia Calcs
 Gynecological Calcs
LOGIQ P5
Urological Calcs
Renal Calcs
Real-time Auto Doppler Calculations
TruAccess, Raw Data Processing
MpegView
Key Macro
Network Storage
Auto Focus
Wide Field of View on convex
Simultaneous Display on BE9CS probe
Quick Save
B-Steer
Virtual Convex with X-Beam
ED Report
Flexible Monitor Arm
System Options
 Auto IMT
 Elastography
 TVI
 Stress Echo
 DICOM 3.0 Connectivity
 ECG
 Coded Contrast Imaging
 CW Doppler
 Anatomical M-Mode
 LOGIQView
 B-Flow
 Easy 3D (Baby Face, Vascular)
 Advanced 3D, with 3D Landscape
 Real time 4D
 3rd. probe port
 3-Pedal Foot Switch, with
Programmable Functionality
 Rear Handle
 Drawer
 Probe Cable Hanger
 Urology Probe Holder
 1-Pedal Foot Switch
 Remote Control Switch
 PDF report conversion
Media & Peripheral Options
 Integrated Mounting Kits and Remote
Controls Provided for
- B/W Digital thermal printer
- Color Digital thermal printer
- DVD Video Recorder
Display Modes
Live and Stored Display Format: Full
size and split screen - both w/
thumbnails. For Still and CINE
Review Image Format: 4x4, and
"thumbnails". For Still and CINE
Simultaneous Capability
B/PW
B/CFM or PDI
B/M
B + CFM/M
Realtime Triplex Mode
(B + CFM or PDI/PW)
B-Flow + PW (option)
Dual B (B/B)
Dual B + CFM or PDI
Selectable Alternating Modes
B/M
B/PW
B + CFM/M
B + CFM (PDI)/PW
3D  Mode
3D  Color Mode (option)
B/CW (option)
B + CFM (PDI)/CW (option)
Multi Image Split Screen
Live and/or frozen
B + B/CFM or PDI
Independent CINE playback
Quad screen format
Zoom: Write/Read/Pan
Colorized Image
Colorized B
Colorized M
Colorized PW
Colorized CW (option)
Time line Display
Independent Dual B/PW Display
Display Formats:
Top/ Bottom selectable format (Size:
1/2:1/2; 1/3:2/3; 2/3:1/3)
Side/Side selectable format
(1/2:1/2; 1/3:2/3; 0:1)
Switchable after freeze
TVI
Stress Echo
Elastography
Display Annotation
1st. Patient Name: First, Last, &
Middle name each store 27
characters. Up to 62 total characters
displayed.
2nd. Patient ID: 31 characters. Up to
27 total characters displayed.
Age, Sex and Birth Date (optional)
Hospital Name: 23 characters.
Date: 3 Types selectable
MM/DD/YY,
DD/MM/YY,
YY/MM/DD
Time: 2 types selectable
24 hours, 12 hours
Gestational Age from
LMP/EDD/GA/BBT
Page 2 of 14
Probe Name
Gray Map names
Probe Orientation
Depth Scale Marker
Lateral Scale Marker
Focal Zone Markers
Image Depth
Zoom Depth
B-Mode
Gain
Dynamic Range
Imaging Frequency
Edge Enhance
Frame Averaging
Gray Map
ATO On/Off
M-Mode
Gain
Dynamic Range
Time Scale
Doppler Mode
Gain
Angle
Sample Volume Depth and Width
Wall Filter
Velocity and/or Frequency Scale
Spectrum Inversion
Time Scale
PRF
Doppler Frequency
Color Flow Mode
Line Density
Frame Averaging
Packet Size
Color Scale
Color Velocity Range and Baseline
Color Threshold Marker
Color Gain
PDI
Color Scale Inversion
Color Doppler Frequency
TGC Curve
Acoustic Frame Rate
Cine Frame Number
VCR Counter
VCR Status
VCR Playback Counter
Body Pattern
Application Name
Measurement Results
Operator Message
Displayed Acoustic Output
TIS: Thermal Index Soft Tissue
TIC: Thermal Index Cranial (Bone)
TIB: Thermal Index Bone
MI: Mechanical Index
% of Power output
Biopsy Guide Line and/or Zone
Heart Rate
LOGIQ P5
General System Parameters
System Setup
Pre-programmable Categories
User Programmable Preset
Capability
Factory Default Preset Data
Languages: English, French, German,
Spanish, Italian, Portuguese, Russian,
Greek, Swedish, Danish, Dutch, Finnish,
Norwegian, Japanese, Chinese
OB Report Format: 5 Types, Tokyo
Univ., Osaka Univ., USA, Europe, and
ASUM
EFBW: 10 types, Japan, USA and
Europe (Tokyo Univ., Osaka Univ., Tokyo
Shinozuka, JSUM, German, Shephard,
Merz, Hadlock/Shephard, Williams,
Brenner)
348 Pre-defined Annotations and
User Programmable
Libraries/Annotations
Body Patterns: 140 human types
plus 14 animal types
Customized Comment Home
Position
Complete User Manual available on
board through Help (F1)
User Manual and Service Manual are
included on CD with each system. A
printed Manual is available upon request.
CINE Memory/Image Memory
CINE Memory: 192MB
Dual Image CINE Display
Quad Image CINE Display
CINE Gauge and CINE Image
Number Display
CINE Review Loop
CINE Review Speed: 20 steps (10, 20,
30, 40, 50, 60, 70, 80, 90, 100, 150, 200,
300, 400, 500, 600, 700, 800, 900,
1000%)
Selectable CINE Sequence for CINE
Review
Measurements, Calculations and
Annotations on CINE Playback
Scrolling Timeline Memory
Cine Capture Function
Image Storage
On-board database of patient
information from past exams
Storage Format: DICOM/Raw Data
DICOM Still Image Storage Size:
Gray Image: ~300K to ~1.3 MB
Color Image: ~900K to 1.9 MB
Multiframe
Display Format: Full size, 4x4, and
"thumbnails"
Live image and stored image sideby-side display
CD-R storage: 650, 700 MB
DVD storage: -R 4,7GB
Conversion to MPEG, JPEG, AVI
(SaveAs) and WMV (MPEGvue) file
formats
Internal Hard Drive partition of
50GB for Image Storage
External USB 2.0 Hard Drive support
for Import, Export, DICOM Read, SaveAs
and MPEGVue
USB 1.1 / 2.0 Memory Stick support
for SaveAs and MPEGVue
Network Storage support for Import,
Export, DICOM Read, SaveAs, MPEGVue
Connectivity
Ethernet network connection
RS-232 serial data output (need a
converter cable)
DICOM 3.0 (option)
Verify
Print
Store
Modality Worklist
Storage Commitment
Modality Performed Procedure
Step (MPPS)
Media Exchange
Off network / mobile storage
queue
Query/Retrieve; supported on
Centricity and other compatible
vendors
Structured Reporting; compatible
with LOGIQworks
Public SR Template
Media Store of SR
iLinq capability
Scanning Parameters
Digital Beamformer
2560 processing channels
Frame Rate: > 700 depending on
preset
Displayed Imaging Depth: 0  30 cm
Minimum Depth of Field: 0 2cm
(probe dependent)
Maximum Depth of Field: 0  30 cm
(probe dependent)
Transmission Focus
1  8 Focus Points Selectable (probe
and application dependent)
Focal Zone Position, B-Mode 16
steps, CFM 64 steps
Multi-Frequency/Wideband
Page 3 of 14
Technology
Frequency Range 1 to 13MHz
256 Shades of Gray
Up to 197 db processing dynamic
range
16,777,216 Hues of Color
Adjustable Field of View (FOV) up to
170 degree depending on probe
Image Reverse: Right/Left
Image Rotation: 4 steps
Rotation: 0, 90, 180, 270
B-Mode
B/M Acoustic Output: 0  100%,
2% step
Image Reverse: On/Off
B Color: 10 types
Thermal Index: TIC, TIS, TIB
Softener: 4 steps
Focus Number: 8 steps
Focus Width: 3 types
Range Focus: On/Off
Compression: 0.5  1.5, 0.1 step
Line Density: 4 steps
Line Density Zoom: 4 steps
Suppression: 6 steps
Frame Average: 8 steps
Edge Enhance: 6 steps
Scanning Size (FOV or Angle):
probe dependent, see probe
specifications
Gray Scale Map: 23 types
Clear Map: 23 types
Tint Map: 10 types
Gain: 0  98 dB, 2 dB step
Dynamic Range: 30  120 dB, 3 dB
step
Depth: 1  30 cm, 1 cm step,
depend on probe.
Rejection: 6 steps
Frequency: Up to 5 steps, depend
on probe
Auto Line Density: On/Off presettable
Steered Linear: +/- 15
Color Flow Mode
Base Line: 0  100 %, 10 % step
Invert: On/Off
CF/PDI Focus Depth: default presettable for 0  100 % of ROI in depth,
10 % step
CF/PDI Flash Suppression: 2 steps
CF/PDI Acoustic Output: 0  100%,
10% step
CF/PDI Angle Steer: 0, +/- 20
Packet Size: 5  16, dependent on
probe/application
LOGIQ P5
Line Density: 5 steps
Line Density Zoom: 5 steps
Frame Average: 7 steps
PRF: 280 Hz  19600 Hz
Spatial Filter: 6 steps
Gain: 0  40 dB, 0.5 dB step
Wall Filter:4 steps depend on
probe/application
Scanning Size (FOV or Angle): probe
dependent
CF/PDI Vertical Size (mm) of ROI:
default pre-settable
CF/PDI Center Depth (mm) of ROI:
default pre-settable
CF/PDI Frequency: Up to 3 steps,
depend on probe
Color Map: 20 types depend on
application
Transparent: 5 steps
Color Threshold: 0  100 %, 5 % step
Auto Line Density: On/Off presettable
PW/CF Ratio: 1, 2, 4
Accumulation: 8 steps
Power Doppler Imaging
PDI Map: 13 types
CF/PDI Flash Suppression: 2 steps
CF/PDI Focus Depth: default presettable for 0  100 % of ROI in depth,
10 % step
CF/PDI Acoustic Output: 0  100%,
10% step
CF/PDI Angle Steer: 0, +/- 20
Packet Size: 5  16, dependent on
probe/application
Spatial Filter: 6 steps
Frame Average: 7 steps
PRF: 280 Hz  19600 Hz
Power Threshold: 0  100 %, 5 %
step
Gain: 0  40 dB, 0.5 dB step
Wall Filter: 7 steps depend on
probe/application
CF/PDI Frequency: Up to 3 steps,
depend on probe
Auto Line Density: On/Off presettable
Transparent: 5 steps
Invert: On/Off
Accumulation: 8 steps
M-Mode
Sweep Speed: 8 steps
M Color: 10 types
M/PW Display Format: V-1/3B, V1/2B, V-2/3B, H-1/2B, H-1/4B, TL only
B/M Acoustic Output: 0  100 %, 2 %
step
Rejection: 6 steps
Dynamic Range: 30  120 dB, 3 dB
step
Edge Enhance: 6 steps
Gray Scale Map: 23 types
Clear Map: 23 types
M Gain: 0  98 dB, 2 dB step
Anatomical M-Mode
M-mode cursor adjustable at any
plane
Can be activated from a CINE loop
from a live or stored image
Available with Color Flow Mode
M & A capability
PW/CW-Mode
Maximum and Minimum Velocity
Scales
Max: 10 m/sec
Min: 5 cm/sec
Gray Scale Map: 4 types
Dynamic Range: 24  60, 4 dB step
Base Line: 5  95 %, 11 steps
SV Gate: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 12,
14, 16 mm
Angle Correct: +/- 90, 1 step
Spectral Color: 6 types
PW Sweep Speed: 8 steps
Invert: On/Off
M/PW Display Format: V-1/3B, V1/2B, V-2/3B, H-1/2B, H-1/4B, TL only
Duplex: On/Off(PW only)
PW Acoustic Output: 0  100 %,
10 % step
Spectral Averaging: 4 steps
Time Resolution: 4 steps
PW/CF Ratio: 1, 2, 4
Rejection: 15 steps
Gain: 0  32 dB, 1 dB step
Wall Filter: 5  1500 Hz, 22 steps,
depend on probe/application
PW Angle Steer: 0, +/- 5, 10, 15, 20
PRF: 640  29800 Hz with PW, 50000
Hz with CW
Sample Volume Depth: 29 steps
default pre-settable
CW-Mode (option) is Available on
the Following Probes
3S
5S
7S
3Sp
5Sp
P2D
P6D
Coded Harmonic Imaging
Available on the Following Probes
3.5C
3.5CS
Page 4 of 14
3CRF
4C
5CS
E8C
E8CS
8C
ERB
BE9CS
3S
5S
7S
3Sp
5Sp
8L
9L
11L
i12L
i739
T739
4D3C-L
4DE7C
4D8C
Softener: 4 steps
Line Density: 4 steps
Line Density Zoom: 4 steps
Suppression: 6 steps
Edge Enhance: 6 steps
Gray Scale Map: 23 types
Clear Map: 23 types
Tint Map: 10 types
Gain: 0  98 dB, 2 dB step
Dynamic Range: 30  120 dB, 3 dB
step
Rejection: 6 steps
Auto Line Density: On/Off presettable
Frequency: Up to 5 steps, depend on
probe
-
Coded Excitation
Available on the following probes
the Following Probes
11L
12L
8C
E8C
E8CS
BE9CS
ERB
Virtual Convex
Provides a Convex Field of View for
Convex, Linear and Sector probes
Available on the Following Probes
8L
9L
10L
11L
12L
i12L
LOGIQ P5
T739
I739
3S
5S
7S
3Sp
5Sp
ERB - Linear
4C
5CS
4D3C-L
3CRF
Automatic Optimization
Optimize B-Mode, B-Flow image to
improve contrast resolution.
Selectable amount of contrast
resolution improvement (low, medium,
high)
Auto-TGC in B-Mode and Color 
adjusts overall and axial gain
Optimize Spectral Waveform 
adjusts baseline, invert, PRF (on live
image), and angle correction
Algorithm works on focal zone/
number and depth changes
Available on stored or live image
 Available in B-Mode, B-Flow, PW
Doppler, Color
B-Flow (option)
Available on 8L, 9L, 10L, 11L, i739
and T739 Probes
Background: On/Off
Sensitivity/PRI: 14 steps
Line Density: 4 steps
Edge Enhance: 6 steps
Frame Average: 8 steps
Gray Scale Map: 23 types
Clear Map: 23 types
Tint Map: 10 types
Dynamic Range: 30  120 dB, 3 dB
step
Rejection: 6 steps
Gain: 0  98 dB, 2 dB step
Auto Line Density: On/Off presettable
Accumulation: 8 steps
Coded Contrast Imaging (option)
Coded Harmonic Angio: Available on
3.5C, 3.5CS, 4C and 5CSprobes
Tissue Background Selection: 4
steps
TruAgent Detection: Available on
3.5C, 3.5CS, 4C and 5CS probes
2 frequencies on 3.5C and 3.5CS, 1
on 4C and 5CS probes
Coded Phase Inversion:
2 types, Available on 3.5C and 3.5CS
probes
3 type, Available on 4C, 5CS probes
Tissue Background Selection: 4
steps
Time Trigger Scan: 0.3, & 0.5  10
sec, 0.5 sec step
Time Intensity Curve Analysis
Accumulation: 8 steps
 SRI-HD
The LOGIQ P5 is designed for
compatibility with commercially
available ultrasound contrast agents.
Because the availability of these agents
is subject to government regulation and
approval, product features intended for
use with these agents may not be
commercially marketed nor made
available before the contrast agent is
cleared for use. Contrast related product
features are enabled only on systems for
delivery to an authorized country or
region of use. GE Medical Systems
makes no claims concerning the safety
or effectiveness of contrast agents.
Elastography Mode (option)
Elasto Acoustic Output: 0  100%
Line Density
Elasto Frequency
Frame Average
Axial Smoothing
Later Smoothing
Window
Frame Reject
Noise Reject
Transparency Map
Color and Gray Map
Soft Compress
Hard Compress
Quality Graph : On/Off
Available on the following Probes
11L
Auto IMT (option)
Automated measurement of the
intimae media thickness of common
carotid artery
Result window displayed on screen
TVI (option)
Myocardial Doppler Imaging with
color overlay on tissue image
Available on all sector probes
Tissue color overlay can be
removed to show just the 2D image,
still retaining the tissue velocity
information
Page 5 of 14
Anatomical M-mode: free
placement of M-mode generated from
the cursor independent from the axial
plane
Q-Analysis: Multiple Time -Motion
trace display from selected points in
the myocardium
CrossBeam
Stress Echo Package (option)
Advanced and flexible stress-echo
examination capabilities
Provides exercise and
pharmacological protocol templates
6 default templates
Template editor for user
configuration of existing templates or
creation of new templates
Reference scan display during
acquisition for stress level comparison
(dual screen)
Baseline level/Previous level
selectable
Raw data continuous capture
Over 180 sec available
Wall motion scoring (bulls-eye and
segmental)
LOGIQView (option)
Available on all probes
Extended Field of View Imaging
For use in B-Mode
LOGIQView Status
Auto detection of scan direction
Pre or post-process zoom up to 10X
Rotation
Auto best fit on monitor
Measurements in B-Mode
Up to 60 cm scan length
Easy 3D (option)
Colorize
Threshold (Opacification)
Render
Texture
Gray Surface
Scalpel
Auto Movie
Undo
Reset
Advanced 3D (option)
Acquisition of Color data
Automatic rendering
3D Landscape technology
3D Movie
Main Mode
Provides 3, 5 or 7 angles for
Compounding on Linear probes, 3 or 5
angles on Convex probes
Compatible with Side by Side
Display
Compatible with: Color mode,
Timeline mode, SRI-HD, Coded
Harmonic Imaging, Virtual Convex
SRI-HD
High Definition Speckle Reduction
Imaging
Provides 6 levels of speckle
reduction
Side by Side Display
Compatible with ALL linear, convex
and sector transducers
Compatible with ALL scanning
mode
Compatible with Side by Side
Display
Real Time 4D (option)
Acquisition Modes:
Realtime 4D Bmode
Static 3D Bmode
Visualization Modes:
3D Rendering (diverse surface and
intensity projection modes)
Sectional Planes (3 Section planes
perpendicular to each other)
Render Mode:
Surface texture, Surface Smooth,
max-, min- and X-ray (average
intensity projection), Gradient,
Inversion, Glass Body, Mix Mode of
two render Modes
Curved 3point Render start
3D Movie
-
LOGIQ P5
Provides Spatial Compounding
Available on the linear and convex
probes
3CRF
4C
5CS
8C
E8C
E8CS
8L
9L
11L
I12L
I739
T739
BE9CS
4D3C-L
4DE7C
4D8C
ERB
Scalpel: 3D Cut tool
Display Format:
Quad: A-/B-/C-Plane/3D
Dual: A-Plane/3D
Single: 3D or A- or B- or C-Plane
Pre-Processing
Write Zoom up to 8x
B/M-Mode
Gain
TGC
Acoustic Output
Transmission Focus Position
Transmission Focus Number
Transmission Focus Width
Imaging Frequency
Edge Enhancement
Line Density Control
Live Anatomical M-mode
PW/CW-Mode
Gain
Dynamic Range
Acoustic Output
Doppler Frequency
Velocity Scale (PRF)
PW/CF Ratio
Wall Filter
Time Resolution
Sample Volume Gate for PWMode
Length
Depth
Color Flow Mode
CFM Gain
CFM Velocity Scale (PRF)
Acoustic Output
CFM Frequency
Wall Filter
Packet Size
CFM Spatial Filter
CFM Line Density
CFM Regression Filter
Accumulation
Post-Processing w/ TruAccess (Raw
Data)
SRI-HD  Selectable level of Speckle
Reduction Imaging
Max Read Zoom to 16x
B/M-Mode
ATO (Auto Tissue Optimization)
Image Reverse
Image Rotation
Gray Map
Colorized B and M
Gain
Dynamic Range
TGC
Page 6 of 14
Compression
Rejection
Frame Averaging
B Softener
Suppression
Sweep Speed for M-Mode
Anatomical M Mode
PW/CW-Mode
ASO (Auto Spectral Optimization)
Base Line Shift
Gray Map
Post Gain
Compression
Rejection
Spectral Averaging
Colorized D
Display Format
Sweep Speed
Color Flow Mode
ACO (Auto Color Optimization)
Base Line Shift
Color Map
Post Gain
Frame Averaging
CFM Display Threshold
Angle Correct (PW mode)
Quick Angle Correct (PW mode)
Auto Angle Correct (PW mode)
Spectral Invert for Color and
Doppler
3D reconstruction from a stored
CINE loop
Anatomical M-Mode
Accumulation
Cine Capture
-
Physiological Input Panel (Option)
Physiological Input
ECG, 1 channel
Dual R-Trigger
Pre-settable ECG R Delay Time
Pre-settable ECG Position
Adjustable ECG Gain Control
Automatic Heart Rate Display
Measurements / Calculations
General B-Mode
Depth & Distance
Circumference (Ellipse / Trace)
Area (Ellipse / Trace)
Volume (Ellipsoid)
% Stenosis (Area or Diameter)
Angle between two lines
General M-Mode
M-Depth
Distance
Time
Slope
LOGIQ P5
Heart Rate
General Doppler
Measurements/Calculations
Velocity
Time
A/B Ratio (Velocities / Frequency
Ratio
PS (Peak Systole)
ED (End Diastole)
PS/ED (PS/ED Ratio)
ED/PS (ED/PS Ratio)
AT (Acceleration Time)
ACCEL (Acceleration)
TAMAX (Time Averaged Maximum
Velocity
Volume Flow (TAMEAN and Vessel
Area)
Heart Rate
PI (Pulsatility Index)
RI (Resistivity Index)
Real-time Doppler Auto
Measurements / Calculations
PS (Peak Systole)
ED (End Diastole)
MD (Minimum Diastole)
PI (Pulsatility Index)
RI (Resistivity Index)
AT (Acceleration Time)
ACC (Acceleration)
PS/ED (PS/ED Ratio)
ED/PS (ED/PS Ratio)
HR (Heart Rate)
TAMAX (Time Averaged Maximum
Velocity)
PVAL (Peak Velocity Value)
Volume Flow (TAMEAN and Vessel
Area)
OB Measurements/Calculations
Gestational Age by:
GS (Gestational Sac)
CRL (Crown Rump Length)
FL (Femur Length)
BPD (Biparietal Diameter)
AC (Abdominal Circumference)
HC (Head Circumference)
APTD x TTD (Anterior/Posterior
Trunk Diameter by Transverse Trunk
Diameter)
LV (Length of Vertebra)
FTA (Fetal Trunk Cross-sectional
Area)
HL (Humerus Length)
BD (Binocular Distance)
FT (Foot Length)
OFD (Occipital Frontal Diameter)
TAD (Transverse Abdominal
Diameter)
-
TCD (Transverse Cerebellum
Diameter)
THD (Thorax Transverse Diameter)
TIB (Tibia Length)
ULNA (Ulna Length)
Estimated Fetal Weight (EFW) by:
AC, BPD
AC, BPD, FL
AC, BPD, FL, HC
AC, FL
AC, FL, HC
AC, HC
Calculations and Ratios
FL/BPD
FL/AC
FL/HC
HC/AC
CI (Cephalic Index)
AFI (Amniotic Fluid Index)
Measurements / Calculations by:
Jeanty, Merz, Tokyo University, Mercer,
Hansmann, Erickson, Hill, Shephard,
Hadlock, Hohler, Campbell
Fetal Graphical Trending
Growth Percentiles
Multi-Gestational Calculations (4)
Fetal Qualitative Description
(Anatomical survey)
Fetal Environmental Description
(Biophysical profile)
Programmable OB Tables
Over 20 selectable OB Calcs
Expanded Worksheets
-
GYN Measurements/Calculations
Right Ovary Length, Width, Height
Left Ovary Length, Width, Height
Uterus Length, Width, Height
Ovarian Volume
ENDO (Endometrial thickness)
Ovarian RI
Uterine RI
Follicular measurements
Summary Reports
Vascular
Measurements/Calculations
SYS DCCA (Systolic Distal Common
Carotid Artery)
DIAS DCCA (Diastolic Distal Common
Carotid Artery)
SYS MCCA (Systolic Mid Common
Carotid Artery)
DIAS MCCA (Diastolic Mid Common
Carotid Artery)
SYS PCCA (Systolic Proximal
Common Carotid Artery)
DIAS PCCA (Diastolic Proximal
Common Carotid Artery)
Page 7 of 14
SYS DICA (Systolic Distal Internal
Carotid Artery)
DIAS DICA (Systolic Distal Internal
Carotid Artery)
SYS MICA (Systolic Mid Internal
Carotid Artery)
DIAS MICA (Diastolic Mid Internal
Carotid Artery)
SYS PICA (Systolic Proximal Internal
Carotid Artery)
DIAS PICA (Diastolic Proximal
Internal Carotid Artery)
SYS DECA (Systolic Distal External
Carotid Artery)
DIAS DECA (Diastolic Distal External
Carotid Artery)
SYS PECA (Systolic Proximal External
Carotid Artery)
DIAS PECA (Diastolic Proximal
External Carotid Artery)
VERT (Systolic Vertebral Velocity)
SUBCLAV (Systolic Subclavian
Velocity)
Summary Reports
Mean IMT Measurement Tools
Cardiac
Measurements/Calculations
B-Mode Measurements
 Aorta
Aortic Root Diameter (Ao Root Diam)
Aortic Arch Diameter (Ao Arch Diam)
Ascending Aortic Diameter (Ao Asc)
Descending Aortic Diameter (Ao Desc
Diam)
Aorta Annulus Diameter (Ao Annulus
Diam)
Aorta Isthmus (Ao Isthmus)
Aorta *** (Ao st junct)
 Aortic Valve
Aortic Valve Cusp Separation (AV Cusp)
Aortic Valve Area Planimetry (AVA Pl
ani met ry)
*** ( Trans AVA)
 Left Atrium
Left Atrium Diameter (LA Diam)
LA Length (LA Major)
LA W idth (LA Minor)
Left Atrium Diameter to AoRoot
Diameter Ratio (LA/Ao Ratio)
Left Atrium Area (LAA(d), LAA(s))
Left Atrium Volume, Single Plane,
Method of Disk (LAEDV A2C, LAESV
A2C ) (LAEDV A4C, LAESV A4C)
 Left Ventricle
LOGIQ P5
Left Ventricle Mass (LVPWd, LVPWs)
Left Ventricle Volume, Teichholz/Cubic
(LVIDd, LVI Ds)
Left Ventricle Internal Diameter (LVIDd,
LVI Ds)
Left Ventricle Length (LVLd, LVLs)
Left Ventricle Outflow Tract Diameter
(LVOT Diam)
Left Ventricle Posterior Wall Thickness
(LVPWd, LVPWs)
Left Ventricle Length (LV Major)
Left Ventricle Width (LV Minor)
Left Ventricle Outflow Tract Area (LVOT)
Left Ventricle Area, Two Chamber/Four
Chamber/Short Axis (LVA (d), LVA (s))
Left Ventricle Endocardial Area, Width
(LVA (d), LVA(s))
Left Ventricle Epicardial Area, Length
(LVAepi (d), LVAepi (s))
Left Ventricle Mass Index (LVPWd,
LVPWs)
Ejection Fraction, Teichholz/Cube
(LVIDd, LVIDs)
Left Ventricle Posterior Wall Fractional
Shortening (LVPWd, LVPWs)
Left Ventricle Stroke Index,
Teichholz/Cube (LVIDd, LVIDs, and
Body Surface Area)
Left Ventricle Fractional Shortening
(LVIDd, LVIDs)
Left Ventricle Stroke Volume,
Teichholz/Cubic (LVIDd, LVIDs)
Left Ventricle Stroke Index, Single Plane,
Two Chamber, Method of Disk (LVI Dd,
LVIDs, LVSd, LVSs)
Left Ventricle Stroke Index, Single Plane,
Four Chamber, Method of Disk (LVI Dd,
LVIDs, LVSd, LVSs)
Left Ventricle Stroke Index, Bi-Plane,
Bullet, Method of Disk (LVAd, LVAs)
Interventricular Septum (IVS)
Left Ventricle Internal Diameter (LVI D)
Left Ventricle Posterior Wall Thickness
(LVPW)
 Mitral Valve
Mitral Valve Annulus Diameter (MV Ann
Diam)
E-Point-to-Septum Separation (EPSS)
Mitral Valve Area by Pressure Half Time
(MVA By PHT)
Mitral Valve Area Planimetry (MVA
Planimetry)
 Pulmonic Valve
Pulmonic Valve Area (PV Planimetry)
Pulmonic Valve Annulus Diameter (PV
Annulus Diam)
Pulmonic Diameter (Pulmonic Diam)
Right Atrium
Right Atrium Diameter, Length (RAD
Ma)
Right Atrium Diameter, Width (RAD Mi)
Right Atrium Area (RAA)
Right Atrium Volume, Single Plane,
Method of Disk (RAAd)
Right Atrium Volume, Systolic, Single
Plane, Method of Disk (RAAs)
Right Ventricle
Right Ventricle Outflow Tract Area
(RVOT Planimetry)
Left Pulmonary Artery Area (LPA Area)
Right Pulmonary Artery Area (RPA Area)
Right Ventricle Internal Diameter
(RVIDd, RVIDs)
Right Ventricle Diameter, Length (RVD
Ma)
Right Ventricle Diameter, Width (RVD
Mi)
Right Ventricle Wall Thickness (RVAWd,
RVAWs)
Right Ventricle Outflow Tract Diameter
(RVOT Diam)
Left Pulmonary Artery (LPA)
Main Pulmonary Artery (MPA)
Right Pulmonary Artery (RPA)
System
Interventricular Septum Thickness
(IVSd, IVSs)
Inferior Vena Cava
Pulmonary Artery Diameter (MPA)
Systemic Vein Diameter (Systemic
Diam)
Patent Ductus Arterosis Diameter (PDA
Diam)
Pericard Effusion (PEs)
Patent Foramen Ovale Diameter (PFO
Diam)
Ventricular Septal Defect Diameter
(VSD Diam)
Interventricular Septum (IVS) Fractional
Shortening (IVSd, IVSs)
Tricuspid Valve
Tricuspid Valve Area (TV Panimetry)
Tricuspid Valve Annulus Diameter (TV
Annulus Diam)
M-Mode Measurements
Page 8 of 14
 Aorta
Aortic Root Diameter (Ao Root Diam)
 Aortic Valve
Aortic Valve Diameter (AV Diam)
Aortic Valve Cusp Separation (AV Cusp)
Aortic Valve Ejection Time (LVET)
 Left Atrium
Left Atrium Diameter to AoRoot
Diameter Ratio (LA/Ao Ratio)
Left Atrium Diameter (LA Diam)
 Left Ventricle
Left Ventricle Volume, Teichholz/Cubic
(LVIDd, LVI Ds)
Left Ventricle Internal Diameter (LVIDd,
LVI Ds)
Left Ventricle Posterior Wall Thickness
(LVPWd, LVPWs)
Left Ventricle Ejection Time (LVET)
Left Ventricle Pre-Ejection Period
(LVPEP)
Interventricular Septum (IVS)
Left Ventricle Internal Diameter (LVI D)
Left Ventricle Posterior Wall Thickness
(LVPW)
 Mitral Valve
E-Point-to-Septum Separation (EPSS)
Mitral Valve Leaflet Separation (D-E
Excursion)
Mitral Valve Anterior Leaflet Excursion
(D-E Excursion)
Mitral Valve D-E Slope (D-E Slope)
Mitral Valve E-F Slope (E-F Slope)
 Pulmonic Valve
QRS complex to end of envelope (Q-toPV close)
 Right Ventricle
Right Ventricle Internal Diameter
(RVIDd, RVIDs)
Right Ventricle Wall Thickness (RVAWd,
RVAWs)
Right Ventricle Outflow Tract Diameter
(RVOT Diam)
Right Ventricle Ejection Time (RVET)
Right Ventricle Pre-Ejection Period
(RVPEP)
Velocity Circumferential Fiber
Shortening (Vcf)
 System
Interventricular Septum Thickness
(IVSd, IVSs)
Pericard Effusion (PE(d))
Interventricular Septum (IVS) Fractional
Shortening (IVSd, IVSs)
LOGIQ P5
 Tricuspid Valve
QRS complex to end of envelope (Q-toTV close)
Doppler Mode Measurements
 Aortic Valve
Aortic Insufficiency Mean Pressure
Gradient (AR Trace)
Aortic Insufficiency Peak Pressure
Gradient (AR Vmax)
Aortic Insufficiency End Diastole
Pressure Gradient (AR Trace)
Aortic Insufficiency Mean Velocity (AR
Trace)
Aortic Insufficiency Mean Square Root
Velocity (AR Trace)
Aortic Insufficiency Velocity Time
Integral (AR Trace)
Aortic Valve Mean Velocity (AV Trace)
Aortic Valve Mean Square Root Velocity
(AV Trace)
Aortic Valve Velocity Time Integral (AV
Trace)
Aortic Valve Mean Pressure Gradient
(AV Trace)
Aortic Valve Peak Pressure Gradient
(AR Vmax)
Aortic Insufficiency Peak Velocity (AR
Vmax)
Aortic Insufficiency End-Diastolic
Velocity (AR Trace)
Aortic Valve Peak Velocity (AV Vmax)
Aortic Valve Peak Velocity at Point E
(AV Vmax)
Aorta Proximal Coarctation (Coarc PreDuct)
Aorta Distal Coarctation (Coarc PostDuct)
Aortic Valve Insufficiency Pressure Half
Time (AR PHT)
Aortic Valve Flow Acceleration (AV
Trace)
Aortic Valve Pressure Half Time (AV
Trace)
Aortic Valve Acceleration Time (AV Acc
Ti me)
Aortic Valve Deceleration TIme (AV
Trace)
Aortic Valve Ejection Time (AVET)
Aortic Valve Acceleration to Ejection
Time Ratio (AV Acc Time, AVET)
Aortic Valve Area according to PHT
 Left Ventricle
Left Ventricle Outflow Tract Peak
Pressure Gradient (VLOT Vmax)
Left Ventricle Outflow Tract Peak
Velocity (LVOT Vmax)
Left Ventricle Outflow Tract Mean
Pressure Gradient (LVOT Trace)
Left Ventricle Outflow Tract Mean
Velocity (LVOT Trace)
Left Ventricle Outflow Tract Mean
Square Root Velocity (LVOT Trace)
Left Ventricle Outflow Tract Velocity
Time Integral (LVOT Trace)
Left Ventricle Ejection Time (LVET)
Cardiac Output by Aortic Flow (AVA Pl
ani met ry, AV Trace)
Stroke Volume Index by Aortic Flow
(AVA Planimetry, AV Trace)
 Mitral Valve
Mitral Valve Regurgitant Flow
Acceleration (MR Trace)
Mitral Valve Regurgitant Mean Velocity
(MR Trace)
Mitral Regurgitant Mean Square Root
Velocity (MR Trace)
Mitral Regurgitant Mean Pressure
Gradient (MR Trace)
Mitral Regurgitant Velocity Time
Integral (MR Trace)
Mitral Valve Mean Velocity (MR Trace)
Mitral Valve Mean Square Root Velocity
(MR Trace)
Mitral Valve Velocity Time Integral (MR
Trace)
Mitral Valve Mean Pressure Gradient
(MR Trace)
Mitral Regurgitant Peak Pressure
Gradient (MR Vmax)
Mitral Valve Peak Pressure Gradient
(MR Vmax)
Mitral Regurgitant Peak Velocity (MR
Vmax)
Mitral Valve Peak Velocity (MR Vmax)
Mitral Valve Velocity Peak A (MV A
Velocity)
Mitral Valve Velocity Peak E (MV E
Velocity)
Mitral Valve Area according to PHT (MV
PHT)
Mitral Valve Flow Deceleration (MV
Trace)
Mitral Valve Pressure Half Time (PV
PHT)
Mitral Valve Flow Acceleration (MV
Page 9 of 14
Trace)
Mitral Valve E-Peak to A-Peak Ratio (AC and D-E) (MV E/ARatio)
Mitral Valve Acceleration Time (MV Acc
Time)
Mitral Valve Deceleration Time (MV Dec
Time)
Mitral Valve Ejection Time ((MV Trace)
Mitral Valve A-Wave Duration (MV A
Dur)
Mitral Valve Time to Peak (MV Trace)
Mitral Valve Acceleration
Time/Deceleration Time Ratio
(MVAcc/Dec Time)
Stroke Volume Index by Mitral Flow
(MVA Planimetry, MVTrace)
Mitral Valve Area from Continuity
Equation (MVAPlanimetry, LVOT Vmax,
MV Vmax)
 Pulmonic Valve
Pulmonic Insufficiency Peak Pressure
Gradient (PR Vmax)
Pulmonic Insufficiency End-Diastolic
Pressure Gradient (PRTrace)
Pulmonic Valve Peak Pressure Gradient
(PV Vmax)
Pulmonic End-Diastolic Pressure
Gradient (PR Trace)
Pulmonic Insufficiency Peak Velocity
(PR Vmax)
Pulmonic Insufficiency End-Diastolic
Velocity (Prend Vmax)
Pulmonic Valve Peak Velocity (PV
Vmax)
Pulmonic End-Diastolic Velocity (PV
Trace)
Pulmonary Artery Diastolic Pressure
(PV Trace)
Pulmonic Insufficiency Mean Pressure
Gradient (PR Trace)
Pulmonic Valve Mean Pressure
Gradient (PV Trace)
Pulmonic Insufficiency Mean Velocity
(PR Trace)
Pulmonic Insufficiency Mean Square
Root Velocity(PR Trace)
Pulmonic Insufficiency Velocity Time
Integral (PR Trace)
Pulmonic Valve Mean Velocity (PV
Trace)
Pulmonic Valve Mean Square Root
Velocity (PV Trace)
Pulmonic Valve Velocity Time Integral
LOGIQ P5
(PV Trace)
Pulmonic Insufficiency Pressure Half
Time (PR PHT)
Pulmonic Valve Flow Acceleration (PV
Acc Time)
Pulmonic Valve Acceleration Time (PV
Acc Time)
Pulmonic Valve Ejection Time (PVET)
Pulmonic Valve Pre-Ejection Period
(PVPEP)
QRS complex to end of envelope (Q-toPV close)
Pulmonic Valve Acceleration to Ejection
TIme Ratio (PV Acc Time, PVET)
Pulmonic Valve Pre-Ejection to Ejection
Time Ratio (PVPEP, PVET)
 Right Ventricle
Right Ventricle Outflow Tract Peak
Pressure Gradient (RVOT Vmax)
Right Ventricle Systolic Pressure (RVOT
Vmax)
Right Ventricle Outflow Tract Peak
Velocity (RVOT Vmax)
Right Ventricle Diastolic Pressure (RVOT
Trace)
Right Ventricle Outflow Tract Velocity
Time Integral (RVOTTrace)
Right Ventricle Ejection Time (RV Trace)
Stroke Volume by Pulmonic Flow
(RVOTPlanimetry, RVOTTrace)
Right Ventricle Stroke Volume Index by
Pulmonic Flow (RVOT Planimetry, RVOT
Trace)
 System
Pulmonary Artery Peak Velocity (PV
Vmax)
Pulmonary Vein Velocity Peak A
(reverse) (P Vein A)
Pulmonary Vein Peak Velocity (P Vein D,
P Vein S)
Systemic Vein Peak Velocity (PDA
Diastolic, PDA Systolic)
Ventricular Septal Defect Peak Velocity
(VSD Vmax)
Atrial Septal Defect (ASD Diastolic, ASD
Systolic)
Pulmonary Artery Velocity Time
Integral (PV Trace)
Systemic Vein Velocity Time Integral
(PDA Trace)
Pulmonary Vein A-Wave Duration (P
Vein A Dur)
IsoVolumetric Relaxation Time (IVRT)
IsoVolumetric Contraction Time (IVCT)
Pulmonary Vein S/D Ratio (P Vein D, P
Vein S)
Ventricular Septal Defect Peak
Pressure Gradient (VSD Vmax)
Pulmonic-to-Systemic Flow Ratio
(Qp/Qs)
 Tricuspid Valve
Tricuspid Regurgitant Peak Pressure
Gradient (TR Vmax)
Tricuspid Valve Peak Pressure Gradient
(TV Vmax)
Tricuspid Regurgitant Peak Velocity (TR
Vmax)
Tricuspid Valve Peak Velocity (TV Vmax)
Tricuspid Valve Velocity Peak A (TV A
Velocity)
Tricuspid Valve Velocity Peak E (TV E
Velocity)
Tricuspid Regurgitant Mean Pressure
Gradient (TR Trace)
Tricuspid Valve Mean Pressure
Gradient (TV Trace)
Tricuspid Regurgitant Mean Velocity
(TR Trace)
Tricuspid Regurgitant Mean Square
Root Velocity (TR Trace)
Tricuspid Regurgitant Velocity Time
Integral (TR Trace)
Tricuspid Valve Mean Velocity (TV
Trace)
Tricuspid Valve Mean Square Root
Velocity (TV Trace)
Tricuspid Valve Velocity Time Integral
(TV Trace)
Tricuspid Valve Time to Peak (TV
Acc/Dec Time)
Tricuspid Valve Ejection Time (TV
Acc/Dec Time)
Tricuspid Valve A-Wave Duration (TV A
Dur)
QRS complex to end of envelope (Q-toTV close)
Tricuspid Valve Pressure Half Time (TV
PHT)
Stroke Volume by Tricuspid Flow (TV
Planimetry, TV Trace)
Tricuspid Valve E-Peak to A-Peak Ratio
(TV E/A Velocity)
Color Flow Mode Measurements
 Aortic Valve
Proximal Isovelocity Surface Area:
Page 10 of 14
Regurgitant Orifice Area (AR Radius)
Proximal Isovelocity Surface Area:
Radius of Aliased Point (AR Radius)
Proximal Isovelocity Surface Area:
Regurgitant Flow (AR Trace)
Proximal Isovelocity Surface Area:
Regurgitant Volume Flow (AR Trace)
Proximal Isovelocity Surface Area:
Aliased Velocity (AR Vmax)
 Mitral Valve
Proximal Isovelocity Surface Area:
Regurgitant Orifice Area (MR Radius)
Proximal Isovelocity Surface Area:
Radius of Aliased Point (MR Radius)
Proximal Isovelocity Surface Area:
Regurgitant Flow (MR Trace)
Proximal Isovelocity Surface Area:
Regurgitant Volume Flow (MR Trace)
Proximal Isovelocity Surface Area:
Aliased Velocity (MR Vmax)
-
Combination Mode Measurements
 Aortic Valve
Aortic Valve Area (Ao Root Diam,
LVOT Vmax, AV Vmax)
Aortic Valve Area by Continuity
Equation by Peak Velocity (Ao Root
Diam, LVOT Vmax, AV Vmax)
Stroke Volume by Aortic Flow (AVA Pl
ani met ry, AV Trace)
Cardiac Output by Aortic Flow (AVA
Planimetry, AV Trace, HR)
Aortic Valve Area by Continuity
Equation VTI (Ao Root Diam, LVOT
Vmax, AV Trace)
 Left Ventricle
Cardiac Output, Teichholz/Cubic (LVIDd,
LVI Ds, HR)
Cardiac Output Two Chamber, Single
Plane, Area-Length/ Method of
Disk(Simpson) (LVAd, LVAs, HR)
Cardiac Output Four Chamber, Single
Plane, Area-Length/ Method of
Disk(Simpson) (LVAd, LVAs, HR)
Ejection Fraction Two Chamber, Single
Plane, Area-Length/ Method of
Disk(Simpson) (LVAd, LVAs)
Ejection Fraction Four Chamber, Single
Plane, Area-Length/ Method of
Disk(Simpson) (LVAd, LVAs)
Left Ventricle Stroke Volume, Single
Plane, Two Chamber/Four Chamber,
Area-Length (LVAd, LVAs)
LOGIQ P5
Left Ventricle Stroke Volume, Single
Plane, Two Chamber/Four Chamber,
Method of Disk(Simpson) (LVIDd, LVIDs,
LVAd, LVAs)
Left Ventricle Volume, Two
Chamber/Four Chamber, Area-Length
(LVAd, LVAs)
Ejection Fraction, Bi-Plane, Method of
Disk (LVAd, LVAs, 2CH, 4CH)
Left Ventricle Stroke Volume, Bi-Plane,
Method of Disk (LVAd, LVAs, 2CH, 4CH)
Left Ventricle Volume, Bi-Plane, Method
of Disk (LVAd, LVAs, 2CH, 4CH)
Left Ventricle Stroke Index, Single Plane,
Two Chamber/Four Chamber, AreaLength (LVSd, LVSs, and BSA)
Left Ventricle Volume, Single Plane,
Two Chamber/Four Chamber, Method
of Disk (LVAd, LVAs)
Left Ventricle Volume, Apical View,
Long Axis, Method of Disk (LVAd, LVAs)
Stroke Volume by Aortic Flow (AVA
Planimetry, AV Trace)
 Mitral Valve
Stroke Volume by Mitral Flow (MVA
Planimetry, MV Trace)
Cardiac Output by Mitral Flow (MVA
Planimetry, MV Trace, HR)
 Pulmonic Valve
Stroke Volume by Pulmonic Flow (PV
Planimetry, PV Trace)
Cardiac Output by Pulmonic Flow (PV
Planimetry, PV Trace, HR)
 Tricuspid Valve
Cardiac Output by Tricuspid Flow (TV
Planimetry, TV Trace, HR)
- Applications: Abdomen, OB Gyn,
Urology, Vascular
- Probe Band Width: 2.0  5.0 MHz
- Number of Element: 128
- Convex Radius: 38 mmR
- FOV (Max): 68
- Physical Foot Print: 48 x 11 mm
- B-mode Imaging Frequency: 2.0, 3.0,
4.0, 5.0 MHz
- Harmonic Frequency: 4, 5, 5.2, 5.5
MHz
- Doppler Frequency: 2.5, 3.3 MHz
- Biopsy Guide Available: Single Angle,
Reusable
 3.5C Wide Band Convex Probe
- Applications: Abdomen, OB Gyn,
Urology, Vascular
- Probe Band Width: 2.0  5.0 MHz
- Number of Element: 128
- Convex Radius: 38 mmR
- FOV (Max): 68
- Physical Foot Print: 48 x 11 mm
- B-mode Imaging Frequency: 2.0, 3.0,
4.0, 5.0 MHz
- Harmonic Frequency: 4, 5, 5.2, 5.5
MHz
- Doppler Frequency: 2.5, 3.3 MHz
- Biopsy Guide Available: Single Angle,
Reusable
Report Designer
On-board reporting package
automates report writing
Formats various exam results into a
report suitable for printing to a
windows printer or reviewing on a
standard PC
Exam results include patient info,
exam info, measurements, calculations,
images, comments and diagnosis
Standard templates provided
Customizable templates
Probes
 3.5CS Thin Wide Band Convex Probe
4C Wide Band Convex Probe
Applications: Abdomen, OB Gyn,
Urology, Vascular
Probe Band Width : 1.4  5 MHz
Number of Element: 128
Convex Radius : 60 mmR
FOV (Max) : 58
Physical Foot Print : 60 x 18 mm
B-mode Imaging Frequency : 2.0,
3.0, 4.0, 5.0 MHz
Harmonic Frequency : 4.0, 5.0, 5.2,
5.5 MHz
Doppler Frequency : 2.5, 3.3 MHz
Biopsy Guide Available : Multi
Angle, Reusable
-
5CS Convex Probe
Applications: Abdomen, OB Gyn,
Urology
Probe Band Width : 2  6 MHz
Number of Element : 128
Convex Radius : 60 mmR
FOV (Max) : 58
Physical Foot Print : 60 x 18 mm
B-mode Imaging Frequency : 2.0,
3.0, 4.0, 5.0 MHz
-
Page 11 of 14
Harmonic Frequency : 4.0, 5.0, 5.5,
6.0 MHz
Doppler Frequency : 2.5, 3.3 MHz
Biopsy Guide Available : Multi
Angle, Reusable
8L Wide Band Linear Probe
Applications : Vascular, Small
Parts
Probe Band Width : 4  11MHz
Number of Element : 128
FOV (Max) : 38 mm
Physical Foot Print : 38 x 10 mm
B-mode Imaging Frequency : 6.0,
8.0, 10.0 MHz
Harmonic Frequency : 8.0, 10.0,
12.0 MHz
Doppler Frequency : 5.0, 6.7 MHz
Steered Angle : 0, +/- 5, 10, 15, 20
Biopsy Guide Available : Multi
Angle, Reusable
9L Wide Band Linear Probe
Applications: Vascular, Small Parts
Probe Band Width : 3  10 MHz
Number of Element : 192
FOV (Max) : 44 mm
Physical Foot Print : 44 x 6 mm
B-mode Imaging Frequency : 5.0,
7.0, 9.0 MHz
Harmonic Frequency : 8.0, 10.0
MHz
Doppler Frequency : 4.0, 5.0 MHz
Steered Angle : 0, +/- 5, 10, 15, 20
Biopsy Guide Available : Multi
Angle, Reusable
 10L Wide Band Linear Probe
- Applications: Vascular, Small Parts,
Neonatal, Pediatrics
- Probe Band Width: 4  11 MHz
- Number of Element: 192
- FOV (Max): 39 mm
- Physical Foot Print: 39 x 10 mm
- B-mode Imaging Frequency: 6.0, 8.0,
10.0 MHz
- Doppler Frequency: 5.0, 6.7 MHz
- Steered Angle: +/- 10, 20
- Biopsy Guide Available: Multi Angle,
Reusable
11L Wide Band Linear Probe
Applications : Small Parts,
Vascular, Neonatal, Pediatrics
Probe Band Width : 5  13 MHz
Number of Element : 192
FOV (Max) : 38.4 mm
Physical Foot Print : 38 x 10mm
B-mode Imaging Frequency : 7.0,
LOGIQ P5
10.0, 12.0 MHz
Harmonic Frequency : 10.0, 12.0,
13.0 MHz
Doppler Frequency : 5.0, 6.7 MHz
Steered Angle : 0, +/- 5, 10, 15, 20
Biopsy Guide Available : Multi
Angle, Reusable
 12L Wide Band Linear Probe
- Applications: Small Parts, Vascular,
Neonatal, Pediatrics
- Probe Band Width: 5  13.0 MHz
- Number of Element: 192
- FOV (Max): 39 mm
- Physical Foot Print: 45 x 10 mm
- B-mode Imaging Frequency: 7.0, 10.0,
12.0 MHz
- Harmonic Frequency: 10.0, 12.0 MHz
- Doppler Frequency: 5.0, 6.7 MHz
- Steered Angle: +/- 10, 20
Biopsy Guide Available: Single Angle,
Reusable
3CRF Wide Band Microconvex Probe
Applications: Abdomen, Urology,
Vascular
Probe Band Width : 2.0  4 MHz
Number of Element: 128
Convex Radius : 20 mmR
FOV (Max) : 80
Physical Foot Print : 28 x 16 mm
B-mode Imaging Frequency : 2.0,
3.0, 3.5, 4.0 MHz
Harmonic Frequency : 3.6, 3.8, 4.0,
4.2MHz
Doppler Frequency : 2.5, 3.3 MHz
Biopsy Guide Available : Single
Angle, Reusable
-
E8C Wide Band Microconvex Probe
Applications : OB Gyn, Urology,
Endocavity
Probe Band Width : 4  11 MHz
Number of Element : 128
Convex Radius : 11 mmR
FOV (Max) : 133
Physical Foot Print : 26 x 10 mm
B-mode Imaging Frequency : 6.0,
8.0, 10.0 MHz
Harmonic Frequency : 8.0, 10.0
MHz
Doppler Frequency : 4.0, 5.0 MHz
Biopsy Guide Available : Single
Angle, Disposable and Reusable
Number of Element : 128
Convex Radius : 9 mmR
FOV (Max) : 170
Physical Foot Print : 15 x 10 mm
B-mode Imaging Frequency : 6.0,
8.0, 10.0 MHz
Harmonic Frequency : 8.0, 10.0
MHz
Doppler Frequency : 4.0, 5.0 MHz
Biopsy Guide Available : Single
Angle, Disposable and Reusable
-
E8CS Wide Band Microconvex Probe
Applications : OB Gyn, Urology,
Endocavity
Probe Band Width : 4  11 MHz
-
8C Wide Band Microconvex Probe
Applications : Neonatal, Pediatrics
Probe Band Width : 4 11 MHz
Number of Element : 128
Convex Radius : 11 mmR
FOV (Max) : 133
Physical Foot Print : 26 x 10 mm
B-mode Imaging Frequency : 6.0,
8.0, 10.0 MHz
Harmonic Frequency : 8.0, 10.0
MHz
Doppler Frequency : 4.0, 5.0 MHz
Biopsy Guide Available : None
-
3S Wide Band Phased Array Sector
Probe
Applications : Cardiac, Trancranial,
Abdomen
Probe Band Width : 1.5  3.5 MHz
Number of Element : 64
FOV (Max) : 90
Physical Foot Print : 19 x 16 mm
B-mode Imaging Frequency : 2.0,
2.5, 3.0 MHz
Harmonic Frequency : 2.8, 3.0, 3.2,
3.6 MHz
Doppler Frequency : 1.7, 2.0, 2.2
MHz
CW Doppler Frequency : 2.0 MHz
Biopsy Guide Available : Multi
Angle, Reusable
5S Wide Band Phased Array Sector
Probe
Applications : Cardiac, Trancranial,
Abdomen
Probe Band Width : 2  5 MHz
Number of Element : 96
FOV (Max) : 90
Physical Foot Print : 19 x 11 mm
B-mode Imaging Frequency : 3.0 ,
4.0, 5.0 MHz
Harmonic Frequency : 4.0, 5.0 MHz
Doppler Frequency : 2.5, 3.3 MHz
CW Doppler Frequency : 2.5 MHz
Biopsy Guide Available: Multi Angle,
Reusable
Page 12 of 14
7S Wide Band Phased Array Sector
Probe
Applications: Cardiac, Pediatrics,
Abdomen
Probe Band Width : 3  8 MHz
Number of Element : 96
FOV (Max) : 90
Physical Foot Print : 14 x 7 mm
B-mode Imaging Frequency : 5.0,
6.0, 7.0 MHz
Harmonic Frequency : 6.0, 7.0, 8.0,
8.5 MHz
Doppler Frequency : 4.0, 5.0 MHz
CW Doppler Frequency : 4.0 MHz
Biopsy Guide Available : None
3Sp Wide Band Phased Array Sector
Probe
Applications : Cardiac, Trancranial,
Abdomen
Probe Band Width : 1.5  5MHz
Number of Element : 64
FOV (Max) : 90
Physical Foot Print : 16 x 14 mm
B-mode Imaging Frequency : 2.0,
3.0, 4.0, 5.0 MHz
Harmonic Frequency : 3.0, 3.5, 4.0,
5.0, 5.5 MHz
Doppler Frequency : 1.8, 2.0, 2.5,
3.3, 4.0 MHz
CW Doppler Frequency : 2.0 MHz
Biopsy Guide Available: Multi Angle,
Reusable
5Sp Wide Band Phased Array Sector
Probe
Applications : Cardiac, Trancranial,
Abdomen
Probe Band Width : 2.0  8.0 MHz
Number of Element : 64
FOV (Max) : 90
Physical Foot Print : 10 x 10 mm
B-mode Imaging Frequency : 4.0,
5.0, 6.7, 8.0 MHz
Harmonic Frequency : 5.0, 6.0, 8.0,
10.0 MHz
Doppler Frequency : 2.7, 3.0, 3.3
4.0, 5.0 MHz
CW Doppler Frequency : 2.5 MHz
Biopsy Guide Available: Multi Angle,
Reusable
i12L Intraoperative Wide Band
Linear Probe
Applications : Intraoperative, Small
Parts, Vascular, Pediatrics
Probe Band Width : 4  12 MHz
Number of Element : 96
FOV (Max) : 25 mm
Physical Foot Print : 29 x 10 mm
B-mode Imaging Frequency : 6.0,
LOGIQ P5
ERB Convex: 5 - 9MHz
- Number of Element: 128
- FOV (Max) :
ERB Linear: 51mm
ERB Convex: 123
- Physical Foot Print:
ERB Linear: 51 x 5 mm
ERB Convex: 17 x 5.5 mm
- B-mode Imaging Frequency: 6.0, 8.0,
10.0 MHz
- Harmonic Frequency: 8.0, 10.0 MHz
- Doppler Frequency: 5.0, 6.7 MHz
Biopsy Guide Available: Multi
Angle, Grid Reusable
8.0, 10.0 MHz
Harmonic Frequency : 10.0,
12.0MHz
Doppler Frequency : 5.0, 6.7 MHz
Steered Angle : +/- 5, 10, 15, 20
Biopsy Guide Available : None
t739 Intraoperative Wide Band
Linear Probe
Applications : Intraoperative, Small
Parts, Vascular, Pediatrics
Probe Band Width : 4 -10 MHz
Number of Element: 192
FOV (Max) : 39 mm
Physical Foot Print : 39 x 10 mm
B-mode Imaging Frequency : 6.0,
8.0, 10.0 MHz
Harmonic Frequency : 8.0, 10.0,
12.0 MHz
Doppler Frequency : 5.0, 6.7 MHz
Steered Angle : +/- 5, 10, 15, 20
Biopsy Guide Available : Multi
Angle, Reusable
i739 Intraoperative Wide Band
Linear Probe
Applications : Intraoperative, Small
Parts, Vascular, Pediatrics
Probe Band Width : 4  10 MHz
Number of Element: 192
FOV (Max) : 39 mm
Physical Foot Print : 39 x 10 mm
B-mode Imaging Frequency : 6.0,
8.0, 10.0 MHz
Harmonic Frequency : 8.0, 10.0,
12.0 MHz
Doppler Frequency : 5.0, 6.7 MHz
Steered Angle : +/- 5, 10, 15, 20
Biopsy Guide Available : None
BE9CS Wide Band Biplane
Microconvex Probe
Applications: Urology, Endocavity
Probe Band Width : 4  11MHz
Number of Element : 96 x 2
Convex Radius : 9 mmR
FOV (Max) : 127
Physical Foot Print : 20 x 10 mm
B-mode Imaging Frequency : 6.0,
8.0, 10.0 MHz
Harmonic Frequency : 8.0, 10.0
MHz
Doppler Frequency : 4.0, 5.0 MHz
Biopsy Guide Available : Single
Angle Reusable (stainless steel),
Disposable
 ERB Wide Band Biplane Probe
- Applications: Urology
- Probe Band Width:
ERB Linear5  10 MHz
4D3C-L Convex Volume Probe
Applications : Abdomen, OB Gyn
Probe Band Width : 2  5 MHz
Number of Element: 192
Convex Radius : 39.1 mmR
Volume Sweep Radius: 19.8 mmR
FOV: 87 Volume 85 x 80
Physical Foot Print : 60 x 13 mm
B-mode Imaging Frequency : 3.0,
4.0, 5.0 MHz
Harmonic Frequency : 4.0, 4.5, 4.8,
.0 MHz
Doppler Frequency : 2.5, 3.3 MHz
Biopsy Guide Available : Single
Angle, Reusable (stainless steel,
plastic)
4D8C Micro Convex Volume Probe
Applications : Neonatal, Pediatrics
Probe Band Width : 4  11 MHz
Number of Element: 192
Convex Radius : 14.0 mmR
Volume Sweep Radius: 80 mm
FOV: 120 Volume 37.4 mm x 29
Physical Foot Print : 30 x 50 mm
B-mode Imaging Frequency : 6.0,
8.0, 10.0 MHz
Harmonic Frequency : 8.0, 10.0,
1.0 MHz
Doppler Frequency : 4.0, 5.0 MHz
Biopsy Guide Available : Single
Angle Reusable(stainless steel, plastic)
4DE7C Convex Volume Probe
Applications : OB Gyn, Urology
Probe Band Width : 4  11 MHz
Number of Element: 192
Convex Radius: 10.1 mmR
Volume Sweep Radius: 11.6 mmR
FOV: 133 Volume 146 x 90
Physical Foot Print : 27 x 6 mm
B-mode Imaging Frequency : 6.0,
8.0, 10.0 MHz
Harmonic Frequency : 8.0, 10.0,
11.0 MHz
Doppler Frequency : 4.0, 5.0 MHz
Page 13 of 14
Biopsy Guide Available : Single
Angle, Reusable (stainless steel)
P2D Non-imaging Single CW
Doppler Pencil Probe
Applications : Cardiac
Frequency: 2.0 MHz
P6D Non-imaging Single CW
Doppler Pencil Probe
Applications : Cardiac, Vascular,
Pediatric
Frequency: 5.0 MHz
UG7C Gastro Probe  Europe Only
Applications Transesophageal,
Stomach
Probe Band Width : 4.5  10MHz
Number of Element: 128
Convex Radius: 8.5 mmR
FOV (Max) : 125
Physical Foot Print : 19 x 5 mm
B-mode Imaging Frequency : 6.0,
8.0, 10.0 MHz
Harmonic Frequency : 8.0, 10.0
MHz
Doppler Frequency : 5.0, 6.7 MHz
Biopsy Guide Available: None
* available only in Europe
-
 CE Marked to Council Directive
93/42/EEC on Medical Devices
 Conforms to the following standards
for safety:
-IEC/EN 60601-1 Medical Electrical
equipment, General requirement for
Safety
-IEC/EN 60601-1-1 Safety
requirements for Medical Electrical
Systems
-IEC/EN 60601-1-2 Electromagnetic
compatibility
-IEC/EN 60601-1-4 Programmable
medical systems
-IEC/EN 60601-2-37 Particular
requirements for the safety of
ultrasonic medical diagnostic and
monitoring equipment
-IEC 61157 Declaration of acoustic
output
-ISO 10993 Biological evaluation of
medical devices
NEMA UD3 Acoustic output display (MI,
TIS, TIB, TIC)
GE Ultraschall Deutschland GmbH & Co KG
Beethovenstr. 239, D-42655 Solingen
Tel.: (+49) 212-2802-0
Fax: (+49) 212-2802-28
Internet : www.gehealthcare.com
GE, GE Monogram and LOGIQ, CrossXBeam ,
TruScan , LOGIQView and Inte are trademarks
of General Electric Company.
GE Medical Systems Ultrasound & Primary
Care Diagnostics, LLC a General Electric
company doing business as GE Healthcare
 2008 General Electric Company - All rights reserved.
Inputs and Outputs
 Video In
- S-Video
 Video Out
- S-Video
 Audio Stereo In
 VGA out
 Connectors
- Footswitch
- USB (6)
- Ethernet
- Power for Peripherals
Not all features or specifications
described in this document may be
available in all probes and/or modes.
General Electric Company reserves the
right to make changes in specifications
and features shown herein, or
discontinue the product at any time
without notice or obligation. Contact GE
Representative for the most current
information.
Safety Conformance
The LOGIQ P5 is:
GE Healthcare. - America: 888-202-5528 P.O
 Listed to UL 60601-1 by a Nationally
Recognized Test Lab
 Certified to CAN/CSA C22.2 601.1 by an
SCC accredited Test Lab
GE Healthcare , GE Asia Technology Park
3rd Floor, Building 1,1 Hua Tuo Road ,
Shanghai  201203
Tel : +86 21 38774000
LOGIQ P5
Box 414, Milwaukee , Wisconsin 53201 USA
Page 14 of 14