Specifications for Gas Analyzer
Component
NOx
SO2
CO
CO2
O2*1
Measurement methods
NDIR
NDIR
NDIR
NDIR
Magnetopneumatic
detection
Range*2
Standard
200~5000
ppm
200~5000
ppm
200~5000
ppm
5~25 vol%10~25 vol%
Optional
100 ppm~
50 ppm~
100 ppm~
Within a
factor of 10
Within a Within a factor of
factor of 5 2.5
Range Ratio
Within a
Within a
factor of 10 factor of 10
Repeatability
Within 0.5% of full scale (with optional range, or during O 2
measurement, 1.0% of full scale)
Linearity (indicator error)
1.0% of full scale
Zero drift
1.0% of full scale (assuming surrounding temperature is maintained
within 5 C)
(with optional range, or O2 measurement, 2.0% of full scale)
Span drift
2.0% of full scale/week (assuming surrounding temperature is
maintained within 5)
Response time*3
Within 60 seconds (Td + T90 from equipment intake area) (sample
flow 0.6 L/min.) (within 240 seconds for SO2 only)
Overall interference fom coexisting gases
2.0% of full scale/week (within standard range, with standard gas
formation)
Display
Touch panel LCD (backlight) (four usable lines)
Environment
Temperature
-5 to 40 C (away from direct sunlight and radiation heat )*4
Condition
Humidity
85% or less (no condensation)
Vibration
100 Hz, 0.3 m/S2 or less
Dust
Standard environment or better
Temperature
250 C or lower
Dust
0.1 g/Nm3 or less
Standard gas
composition*5
NO: 500 ppm or less; NO2: 15 ppm or less; SO2: 1000 ppm or less;
SO3: 50 ppm or less;
CO: 200 ppm or less; CO2: 15% volume or less; H2O: 40% volume or
less
Measuring Gas
Condition
Sampling method
Dry sampling using an electric cooler
Sample gas flow
2.5 L/min~3.0 L/min
Sample inlet tube
Teflon tubing (8/6 mm)
(1) -1.96 to 4.9 kPa
(2) 3.43 kPa
(3) -4.9 to 1.96 kPa
Sample gas pressure
4.9 kPa (three points
selected)
(with no sample gas back
pressure)
Pressure control
Pressure control uses a regulator and cylinder; Reduced pressure
sampling; Control pressure: -4.9 kPa
Output
DC 4 to 20 mA (absolute output) (DC 0 to 16 mA/DC 0 to 1 V/DC 1 to
5V optional) Max. 12 output systems
External output
Analysis alerts, analysis warnings, range display, corrections,
conservation, purging (option)
Contact capacity: DC 30 V 1 A, AC 250 V 1 A resistance load
Correction method
Dry correction, automatic correction (correction cycle: 7 days
standard, can be adjusted to between 1 and 99 days), manual
correction
Calibration gas
Zero gas
With measurement method authorization: N 2, When
there is no measurement method authorization: N2 or
ambient air
O2 carrier
gas
Ambient air
Span gas
Gas cylinder for each component measured (when there
is no measurement method authorization: O2 or ambient
air can be used)
Probe collection point filter
Flange: JIS 10K, 40 AFF; Sample probe tube length: 1000 mm;
Material: SUS-316 stainless steel;
Fil er element: SUS-304 stainless steel and 2um-pleated quartz
wool; Electric heater: 100 VA, with water droplet proof case
Power supply
AC 100 V 15 V(85 V~115 V)
Power frequency
50/60 Hz (switchable)
Power consumption
About 800 VA (heated piping 30m: +1100 VA; heater in tray: +300 VA)
Exterior dimensions/Mass
600 (W) x 1770 (H) x 300 (D) mm (high pressure gas cylinders, 3.4 L
cylinders, maximum of 3 cylinders); About 180 kg (not including
cylinders)
600 (W) x 1770 (H) x 300 (D) mm (high pressure gas cylinders, 3.4 L
cylinders, maximum of 3 cylinders); About 200 kg (not including
cylinders)
Materials in contract with
sample gas
SUS-316 stainless steel, SUS-304 stainless steel, Teflon,
polypropylene, polyethylene,
flourorubber, PVC, PVDF, and glass
Enclosure
Independent outdoor installation
Plate thickness: Main unit, doors, top plate, steel plate: 2.3 mm;
Channel base: 3.2 mm; Doors: front opening; Interface: right front
Color/Finish
Semi-gloss Munsell 5Y7/1 on all inner and outer surfaces
*1: No carrier gas cylinder is required.
*2: Up to two ranges are supported for each component.
*3: Response time may vary depending on the sampling system composition.
*4: Support is available for -15 to 40 C (cold-climate version) and 5 to 50 C.
*5: An NH3 scrubber is available as an option for cases where a combined gas includes NH 3. SO2
measurement corrected for CH4 interference is available for cases where there is CH4 in the sample
gas for SO2 measurement. CO measurement corrected for N 2O interference is available for cases
where there is N2O in the sample gas for CO measurement.