Cardinal Laboratories, LLC
Standard Operating Procedure
     Procedure No.:                         4010                               Revision Date:
01/2008
     Subject: Biological Oxygen Demand, 5 Day (Including
Carbonaceous):
                (SM 5210 B, EPA Method 405.1 )
         -------------------------------------------------------------------------------------------------
-----
       Revision #/Date: Approved By:                                                   Date
Revised:
       -------------------------------------------------------------------------------------------------
------
       00-8/2005
       -------------------------------------------------------------------------------------------------
------
       00-01/2008
       -------------------------------------------------------------------------------------------------
------
         -------------------------------------------------------------------------------------------------
------
                                                                                         Revised 01/2008
         -------------------------------------------------------------------------------------------------
------
                           Biological Oxygen Demand, 5 Day
                               (Including Carbonaceous)
                                   EPA Method: 5210 B
Scope and Application
This method determines the BOD of a water sample.
Description
         This method is used to determine BOD5 & CBOD5 in surface waters,
         treated effluents and untreated effluents. The test measures the
         oxygen utilized during a specified period for the biochemical
         degradation of organic material (carbonaceous). It also may measure
         oxidation of sulfides and ferrous iron as well as a reduction of nitrogen
         unless an inhibitor is used.
Precautions:
         General Requirements (See procedure 0010)
         Corrosives (See procedure 0020)
Sample Preservation-Preparation:
       Samples must be unpreserved. Collect in plastic or glass bottles. If
the analysis is   performed more than 2 hours after collection, refrigerate
to 4 C. Maximum holding      time is 48 hours.
Interferences:
        Oxidation of reduced forms of nitrogen exerts nitrogenous demand.
This can be          prevented by addition of nitrification inhibitor. If inhibitor
is not added the measured        oxygen demand is the sum of carbonaceous
and nitrogenous demand. Poor quality           dilution water will appear as
sample BOD, resulting in a positive bias that is amplified by the dilution
factor.
                                                                                         Revised 01/2008
Apparatus
      - 300 ml incubation bottles:
            Wash with 1% bleach solution and rinse three times with ultra
pure water.
            Do not acid clean. Invert to drain.
      - Air incubator:
              Thermostatically controlled to 20 C  1 C. Sealed to prevent
light entering.
       - Stir plate
       - Stir bar
       - 25ml buret
       - Volumetric pipettes, various
       - Graduated cylinders, various
     - Dissolved oxygen meter:
              Orion meter using membrane electrode to take direct oxygen
measurements.
     - Tubing and glass siphon (from dilution water to BOD bottles)
Reagents
      - Sulfuric Acid Solution (H 2SO4) 1.0N:
             Slowly add 28ml concentrated sulfuric acid to ultra pure water
and
               dilute to 1L.
      - Sodium Hydroxide solution (NaOH) 1.0N:
             Dissolve 40g of NaOH in ultra pure water and dilute to 1L.
      - Nitrification inhibitor: commercially available.
              Hach product #2533-34
      - Manganous Sulfate (MnSO4) solution: commercially available.
      - Alkali-Iodide-Azide solution:
                 Dissolve 500g NaOH (or 700g KOH) and 135g NaI (or 150g KI)
in
                 ultra pure water and dilute to l liter. Add 10g NaN 3 dissolved
in
                 40 ml of ultra pure water.
                 CAUTION: this solution is highly corrosive.
        - Sodium Thiosulfate (Na2S2O3  5H2O) .0250N: commercially
available.
          - Starch indicator: commercially available.
          - Polyseed-TM: commercially available.
                                                                    Revised 01/2008
       - Sodium Sulfite solution: Dissolve 1.575g Na2SO3 in 1-L ultrapure
water. Prepare              daily.
        - Sulfuric acid (H2SO4) concentrated: ACS grade or better.
       - BOD Standard: commercially available.
      - BOD Nutrient buffer pillows: Commercially available.
      - Potassium Iodide - (KI)
      - Chlorine powder pillows: commercially available.
Procedure
1.0 Preparation of dilution water
      1.1 Add two BOD nutrient buffer pillows to 19L of ultra pure water:
      (Cat. # 14863-98          to 19L / Cat. #14862-98 to 6 L). Aerate for
      15-30 minutes using lab grade compressed air.
      1.2 Before use bring dilution water temperature to 20C.
2.0 Standardization of dilution water
             2.1 Pour off three incubation bottles with the freshly prepared
dilution water.
                 Cap and set one aside.
              2.2 To the remaining two bottles, add 1 ml MnSO 4 solution, cap
and invert
                  to mix.
             2.3 Carefully add 1 ml alkali-iodide-azide to each bottle, cap and
invert to mix.
              2.4 Carefully add 1 ml concentrated H 2SO4 to each bottle and
cap. Invert
                 several times of mix well.
              2.5 Pour off 100ml from the incubation bottle, discard.
                                                                     Revised 01/2008
               2.6 Titrate remaining 200ml with 0.0250N sodium thiosulfate to a
light straw
                   color. Add 1-2 ml of starch indicator and continue titration to
a clear
                   endpoint.
               2.7 The DO of the dilution water should be between 7.5 - 9.0
mg/L.
                   (1 ml sodium thiosulfate = 1 mg DO/ liter).
3.0 Calibration of DO meter
               3.1 Put DO probe into bottle previously capped and set aside
(step 2.1)
               3.2 Turn knob from off position to C. Let reading stabilize about
15
                   minutes.
               3.3 Turn to MG/L CAL and enter the DO of the dilution water. Use
the
                   three buttons on the display set to scroll to the correct
numbers.
               3.4 Turn knob to MG/L. After the meter has stabilized, it will give
two beeps
                   and a reading. It is then ready to measure DO results for
BOD/CBOD.
4.0 Sample pretreatment: The following parameters must be checked before
    setting up
                     BODs.
       4.1 pH should be adjusted to between 6.5 - 7.5 using a minimum of 1N
Sulfuric
           acid or 1N Sodium Hydroxide.
       4.2 Samples should be checked for total residual chlorine using Hach
chlorine powder        pillows. If chlorine is present, it should be removed
by addition of sodium sulfite.
               4.2.1 To 100 mls of sample, add 1ml of 6N H 2SO4, 1g potassium
                                  iodide (KI) crystals. Stir to mix.
               4.2.2 Titrate to a pale yellow with sodium sulfite solution, add 1
ml of starch                 indicator and titrate until sample just becomes
clear.
                                                                      Revised 01/2008
             4.2.3 Add to neutralized sample a proportionate amount of
sodium                                 sulfite. Mix well.
           4.2.4 After 10-20 minutes, check for residual chlorine. Avoid
excess                        sodium sulfite, it exerts an oxygen demand
and may positively                                 bias results.
      4.3 Samples must be brought to room temperature before aliquoting.
      4.4 Dissolved Oxygen level on sample should be taken directly from
          the sample
          bottle. If D.O. is greater than 9.0, reduce by purging with organic
          free compressed air at 20 C.
       4.5. If sample is analyzed for CBOD, add 2 aliquots of the nitrification
inhibitor prior                to initial DO reading.
5.0 Dilution Technique
      5.1 Several dilutions should be made for each sample. Choose
      dilution volumes based        on historical data or follow these general
      rules:
             1. Strong Industrial Wastes:
                    Prepare an initial dilution of 10X and prepare bottles in
the                               range of 3mls to 30mls using the diluted
sample.
             2. Raw & Settled Wastewater:
                   Prepare bottles with 3 to 15mls of sample.
             3. Biologically Treated Effluents:
                    Prepare bottles using 15 to 75mls of sample.
             4. River Water:
                    Prepare bottles using 75 to 300mls of sample.
             5.1.1 For samples where content of organisms is unknown, seed
the samples                    to ensure biological degradation. Use 6 ml
per incubation bottle.
             5.1.2 To make polyseed-TM: Empty one capsule of polyseed-TM
into                      200ml of unaerated dilution water. ( This is dilution
water with the                        nutrients added but not aerated.)
Aerate gently for 1 hour. Filter thru                    #4 Whatman filter
paper. Make daily.
             5.1.3 Standard should be prepared with 3 dilutions of 3.0 mls
each and                       requires seed.
                                                                    Revised 01/2008
       5.2 Fill with dilution water by using the siphon apparatus under the
level of the               sample. Rinse after each bottle. Measure initial DO
with oxygen probe and                  replace displaced water with dilution
water. This measurement should be                          made as quickly as
possible after dilution.
      5.3 Cap tightly making sure no air bubbles are entrapped in the
incubation bottle,              cover with plastic cap and place in the
incubator. A dilution water blank should be         run with each sample set. If
samples have been seeded, a seeded dilution water                       blank
should be run. If CBODs are analyzed, a dilution water blank with
                 nitrification inhibitor should be run.
      5.4 Incubate for 5 days at 20 C  1 C.
      5.5 Measure final DO.
             5.5.1 Calibrate meter per steps 2.0-3.5.
6.0 Calculations:
      6.1 In order to have a valid dilution for calculation:
             6.1.1 The difference between the final and initial D.O. must be
                                >2.0 and
             6.1.2 The final D. O. should not be <1.0.
             6.1.3 Report results that are <10 to 2 significant figures.
Results that are                       >10 are reported to 3 significant
figures.
                            BOD CALCULATIONS
SEED VALUE:mg/L Depletion = Initial DO - Final DO
UNSEEDED SAMPLES:
                                                                    Revised 01/2008
                     BOD mg/L = (Initial DO-Final DO) X 300
                                                    mls sample
SEEDED SAMPLES:
                     BOD mg/L = (Initial DO - Final DO - Seed Value) X   300
                                                                  mls sample
DETECTION LIMIT ADJUSTMENTS:
     If all sample depletions are <2.0 the result is BDL. If the largest
sample volume used        was not 300 mls. YOU MUST ADJUST THE
DETECTION LIMIT!!!!!
      Calculation:
             (2.0) X        300
                    mls sample* * This is the largest volume of sample used.
OVER RANGE RESULTS:
     If all Final DOs are <1.0 the result will be reported as >.
     To calculate this, use the initial DO of the dilution with the *smallest
sample volume.
      Calculation: (Initial DO - 1.0) X        300
                                          mls sample*
STANDARD CALCULATION:
      True value is 200 mg/L
      Calculation: BOD mg/L = (Initial DO - Final DO - Seed Value)
                                             0.02
Waste Disposal
      Samples are disposed of in the sink with running water.
Quality Control
       *     Each analytical batch analyzed for BOD or CBOD must be
accompanied by                   a method blank. Blanks are taken through
the incubation steps.
                                                                   Revised 01/2008
             The unseeded blanks should not exceed a BOD of 0.2mg/L.
            The DO of seeded blanks should fall within the range of 0.6-
      1.0mg/L.
      *      All volumes and D.O. readings, must be entered in a notebook
form for data                   review.
      *      The current acceptable limits for QC parameters are 70-130%.
References
       Standard Methods for the Examination of Water and Wastewater 18th
edition 1992
       Method 5210B, pg 5 - 1 thru 5 - 6.
      Methods for Chemical Analysis of Water and Wastes
      EPA 600/4-79-020, Method 405.1 - pp.405.1-1 thru 405.1-2
       National Environmental Methods Index. April 1994. Method 5210B. 7
April 2004.
       <http://www.nemi.gov>
                                                                 Revised 01/2008