Humastar 100
>> System Specifications
System Overview Cat. No.
Analyzer type
Reagent system
Troughtput (max. And typical)
Operation modes
Barcode reader
ISE
Reaction/ reading system
Language available
User interface
Tests Measuring modes
Analysis modes
Calculation modes
Calculation algorithms
# of programmable test
#of programmable profiles
#of standards per test
Smple /Control /
Calibrator
Sample type
Sample Vessels
Sample positions
Sampling volume
Loading
Pre-Dilution
Post-Dilution
Reagent Reagent positions
Reagent Cooling
Reagent 1-2 volume range
Reaction unit Reactionwells / positions for incubation
Reaction volume (minimum)
Reaction volume (maximum)
Cuvette washing system
Incubation temperature
Incubation time
Liquid Handling Liquid transportation
Liquid level sensors
Collusion detection for probe
Reading Optical system
Readings
Light source
Spectral range
Wavelength pre-installed
Max. # of wavelengths installed
Wavelength error ( accuracy)
Wavelength bandwidth (precision)
Detector
Absorbance range (linearity)
Data Processing Memory for
Memory capacity
Reports for
Quality control module
Max. Number of control levels
Test statistic
Warnings
Printer
LIS
General PC min. Requirements
Physical dimensions (W x D x H)
Mains fuse
Electrical requirements
Environmenta
Wash / waste tank
16890
Automatic analyzer
Open (Hunab pre-installed)
100 tests/hr maximum
100 tests/hr typical
(applies to most parameters
Random-acces with STAT
Internal sample barcode reader
None
Reaction cuvetters with mulicycle washing station
English, French, Spanish
External PC, touch screen recommended
Rountinechemistry, turbidimetry
Endpoint (bichromatic) with self blank
Differential endpoint with self blank
Fixed time
Kinetic (bichromatic)
Factor
Multi standard
Single standard
Linear (factor, linear, linear regression, average)
Non linear (cubic spline, poly-linear,logit-log four parameters)
50 actve methods on istrument
Unlimeted
Up to 8
Serum, plasma, urine,CSF
Primary tubes ( 12.5 x100 mm), optional tray ( 16x 100 mm)
Sample cups ( 10 mm), optional tray ( Hitachi 3.5 ml)
Total of 60 position for samples, optional tray 20+20 positions
2 to 300l /tes
Continuous
Automatic (In-needlr or reaction cuvette)
Automatic (abnormal levels,excessive substrate
consumption and/or lack of linearity)
30 reagent positions (20 ml, 50 ml,adapter for tube and
cups, removable tray)
Approx.9C below room temp. At bottom of reagent, independent switch
5 to 350 l
80 washable Bionex cuvettes
210L
350L
8- step washtation, 6 needles
2 bottles :systemic- and special wash solution
35C 0.2C heat transfer by air
792 sec. Max. Incubation + reading time
Pipetting arm
Capacitive
Yes ( vertical)
Interference filters
Mono- or bichromatic
Halogen lamp (Philips, 6V, 10W)
340 to 900 nm
340,405,505,546,600,650,700 nm
9
2 nm on peak wavelength
Haif bandwidth 10 mm
Silicon photodiode
0.25 to 2.5 OD
Sample result, Calibration, patient data, QC data, error-log, absorbance curves
Unlimeted
Patient, (HDD)
single test, complete sample, work sheet,
method and QC's, calibration curves, kinetics, continuos
printing
Levey-Jennings, Westgard multirules, SD, CV%
Up to 3 levels per test
Number executed, SD, CV%, Mean
Analytical limits, reagent and reaction integrity check
(blank, linearity, substrate depletion, Reaction OD etc.)
Printer connected to external PC
ASTM Bi-directional, ethernet of external PC, Polling mode on LAN
External PC ,Pentium IV 2 GHz, CD-ROM,Memory : 512 Mbytes RAM,
Hard disk capacity > 20 GBytes, USB 2.0 ports (min. 3, excl. mouse and keyboard
Operating system: Windows 7 or XP (SP3), latest Updates installed, .Net 4.0
Ethernet network adaptor , min. 900 dots vertical resolution
Recommended: 17 inch touch screen, 1280 x 1024 resolution
Instrument without any components: 69 x 76 x 52 cm
Space required for routine use: 150 x 90 x 100 cm
Packaging: 92 x 86 x 79 cm +
50 x 50 x 50 cm
Weight: Gross: 90.0 kg, net: 51.0 kg
T2 A / 250V, 5 x 20 mm
110...120, 220...240 VAC; 50/60 Hz; < 200 VA
Operating: temperature 16...30 C, humidity < 80 % rel. non condensing
Transport: temperature 0...50C, humidity < 85% rel. non condensing
20 l systemic, 2 l special wash, 20 l waste, level sensors,
water quality min. de-ionized < 10 s
Humastar 100
>> System Specifications
System Overview Cat. No.
Analyzer type
Reagent system
Troughtput (max. And typical)
Operation modes
Reaction/ reading system
Tests Measuring modes
Analysis modes
Smple /Control /
Calibrator Sample type
Sample Vessels
Sample positions
Sampling volume
Loading
Pre-Dilution
Post-Dilution
Reagent Reagent positions
Reagent Cooling
Reagent 1-2 volume range
Reaction unit Reactionwells / positions for incubation
Reaction volume (minimum)
Reaction volume (maximum)
Cuvette washing system
Incubation temperature
Incubation time
General PC min. Requirements
Physical dimensions (W x D x H)
Mains fuse
Electrical requirements
Environmenta
Wash / waste tank
16890
Automatic analyzer
Open (Hunab pre-installed)
100 tests/hr maximum
100 tests/hr typical
(applies to most parameters
Random-acces with STAT
Reaction cuvetters with mulicycle washing station
Rountinechemistry, turbidimetry
Endpoint (bichromatic) with self blank
Differential endpoint with self blank
Fixed time
Kinetic (bichromatic)
Serum, plasma, urine,CSF
Primary tubes ( 12.5 x100 mm), optional tray ( 16x 100 mm)
Sample cups ( 10 mm), optional tray ( Hitachi 3.5 ml)
Total of 60 position for samples, optional tray 20+20 positions
2 to 300l /tes
Continuous
Automatic (In-needlr or reaction cuvette)
Automatic (abnormal levels,excessive substrate consumption
and/or lack of linearity)
30 reagent positions (20 ml, 50 ml,adapter for tube and cups,
removable tray)
Approx.9C below room temp. At bottom of reagent, independent switch
5 to 350 l
80 washable Bionex cuvettes
210L
350L
8- step washtation, 6 needles
2 bottles :systemic- and special wash solution
35C 0.2C heat transfer by air
792 sec. Max. Incubation + reading time
External PC ,Pentium IV 2 GHz, CD-ROM,Memory : 512 Mbytes RAM,
Hard disk capacity > 20 GBytes, USB 2.0 ports (min. 3, excl. mouse and keyboard
Operating system: Windows 7 or XP (SP3), latest Updates installed, .Net 4.0
Ethernet network adaptor , min. 900 dots vertical resolution
Recommended: 17 inch touch screen, 1280 x 1024 resolution
Instrument without any components: 69 x 76 x 52 cm
Space required for routine use: 150 x 90 x 100 cm
Packaging: 92 x 86 x 79 cm +
50 x 50 x 50 cm
Weight: Gross: 90.0 kg, net: 51.0 kg
T2 A / 250V, 5 x 20 mm
110...120, 220...240 VAC; 50/60 Hz; < 200 VA
Operating: temperature 16...30 C, humidity < 80 % rel. non condensing
Transport: temperature 0...50C, humidity < 85% rel. non condensing
20 l systemic, 2 l special wash, 20 l waste, level sensors,
water quality min. de-ionized < 10 s