GC Manual Book
GC Manual Book
Gas Chromatograph
Operation Manual
 Agilent Technologies
Notices
 Agilent Technologies, Inc. 2013              Warranty                                   Safety Notices
No part of this manual may be reproduced       The material contained in this docu-
in any form or by any means (including
electronic storage and retrieval or transla-
                                               ment is provided as is, and is sub-         CAUTION
                                               ject to being changed, without notice,
tion into a foreign language) without prior
                                               in future editions. Further, to the max-   A CAUTION notice denotes a hazard. It
agreement and written consent from
                                               imum extent permitted by applicable        calls attention to an operating
Agilent Technologies, Inc. as governed by
United States and international copyright      law, Agilent disclaims all warranties,     procedure, practice, or the like that, if
laws.                                          either express or implied, with regard     not correctly performed or adhered to,
                                               to this manual and any information         could result in damage to the product
Manual Part Number                             contained herein, including but not        or loss of important data. Do not
                                               limited to the implied warranties of       proceed beyond a CAUTION notice
G3430-90054                                    merchantability and fitness for a par-     until the indicated conditions are fully
                                               ticular purpose. Agilent shall not be      understood and met.
Edition                                        liable for errors or for incidental or
                                               consequential damages in connection
First edition, January 2013                                                                 WA R N I N G
                                               with the furnishing, use, or perfor-
Printed in USA or China                        mance of this document or of any           A WARNING notice denotes a hazard.
Agilent Technologies, Inc.                     information contained herein. Should       It calls attention to an operating
2850 Centerville Road                          Agilent and the user have a separate       procedure, practice, or the like that, if
Wilmington, DE 19808-1610 USA                  written agreement with warranty            not correctly performed or adhered to,
Agilent Technologies, Inc.                     terms covering the material in this        could result in personal injury or
412 Ying Lun Road                              document that conflict with these          death. Do not proceed beyond a
Waigoaqiao Freed Trade Zone                    terms, the warranty terms in the sep-      WARNING notice until the indicated
Shanghai 200131 P.R.China                      arate agreement shall control.             conditions are fully understood and
                                                                                          met.
Contents
              1    Introduction
                                  Chromatography Using a GC                    10
                                  The Operating Panel 11
                                     The display 12
                                     Status lights 12
                                     Beeping instrument 13
                                     Blinking setpoint 13
                                     The keypad 14
              2    Operating Basics
                                  Overview       16
                                  Instrument Control            17
                                  To Start Up the GC            18
                                  To Shut Down the GC for Less Than a Week               19
                                  To Shut Down the GC for More Than a Week               20
                                  Correcting Problems            21
              3    Keypad Operation
                                  The Run Keys        24
                                  The GC Component Keys                   25
                                  The Status Key           26
                                  The Info Key        27
                                  The General Data Entry Keys                  28
                                  The Supporting Keys            29
                                  Method Storage and Automation Keys                30
                                  Keypad Functionality When the GC Is Controlled by an Agilent Data
                                     System 31
                                  The Service Mode Key               31
                                  About GC Status    32
                                    Status board 32
                                    Alert tones 32
                                    Error conditions 33
                                    Blinking setpoint 33
                                  About Logs 35
                                    Maintenance log              35
Operation Manual                                                                                      3
    4   Methods and Sequences
                     What Is a Method?     38
                     What Is Saved in a Method?     38
                     What Happens When You Load a Method?         39
                     Creating Methods 40
                        To load a method 41
                        To store a method 41
                     What Is a Sequence?     43
                     Creating Sequences     43
                     Automating Data Analysis, Method Development, and Sequence
                        Development 47
    6   Chromatographic Checkout
                     About Chromatographic Checkout       54
                     To Prepare for Chromatographic Checkout      55
                     To Check FID Performance     57
                     To Check TCD Performance      62
                     To Check NPD Performance       67
                     To Check uECD Performance       72
                     To Check FPD+ Performance (Sample 5188-5953)        77
                        Preparation 77
                        Phosphorus performance 78
                        Sulfur performance 82
4                                                                               Operation Manual
                                To Check FPD+ Performance (Sample 5188-5245, Japan)       84
                                   Preparation 84
                                   Phosphorus performance 85
                                   Sulfur performance 89
                                To Check FPD Performance (Sample 5188-5953)       91
                                   Preparation 91
                                   Phosphorus performance 92
                                   Sulfur performance 96
                                To Check FPD Performance (Sample 5188-5245, Japan)       99
                                   Preparation 99
                                   Phosphorus performance 100
                                   Sulfur performance 104
              7    Resource Conservation
                                Resource Conservation 108
                                   Sleep Methods 108
                                   Wake and Condition Methods         110
                                To Set the GC to Conserve Resources         112
                                To Edit an Instrument Schedule       115
                                To Create or Edit a Sleep, Wake, or Condition Method    116
                                To Put the GC to Sleep Now      117
                                To Wake the GC Now      118
Operation Manual                                                                                     5
     9   GC-MS Features
                         GC/MS Communications 132
                           Venting the MSD 132
                           MS Shutdown events 132
                         To Set Up a Vent Method    134
                         To Manually Prepare the GC for Venting the MS       135
                         To Manually Exit the MS Vent State    136
                         To Use the GC When the MSD is Shut Down           137
                         To Enable or Disable MS Communications       138
    10   Configuration
                         About Configuration 140
                           Assigning GC resources to a device 140
                           Setting configuration properties 141
                         General Topics 142
                           To Unlock the GC Configuration     142
                           Ignore Ready = 142
                           Information displays 143
                           Unconfigured: 143
                         Oven 144
                            To configure the oven for cryogenic cooling      145
                         Front Inlet/Back Inlet 147
                            To configure the Gas type 147
                            To configure the PTV or COC coolant      147
                            To configure the MMI coolant 149
                         Column # 151
                            To view a summary of column connections         154
                         Composite Columns 160
                           To configure composite columns      161
                         LTM Columns 162
                            LTM Series II column modules      162
                         Cryo Trap   163
                         Front Detector/Back Detector/Aux Detector/Aux Detector 2      165
                            To configure the makeup/reference gas 165
                            Lit offset 165
                            To configure the FPD heaters 165
                            To ignore the FID or FPD ignitor 166
6                                                                                  Operation Manual
                             Analog out 1/Analog out 2       167
                               Fast peaks        167
                             Valve Box 168
                                To assign a GC power source to a valve box heater    168
                             Thermal Aux        169
                             PCM A/PCM B/PCM C           172
                             Pressure aux 1,2,3/Pressure aux 4,5,6/Pressure aux 7,8,9      174
                             Status     175
                             Time      176
                             Valve #      177
                             Front injector/Back injector      178
                             Sample tray (7683 ALS)      180
                             Instrument       181
                             Using the Optional Barcode Reader 182
                                Barcode reader power 182
                                Installing the barcode reader 183
                                To scan configuration data using the G3494B RS-232 barcode
                                    reader 183
                                To scan configuration data using the G3494A USB barcode reader   184
                                To uninstall the RS-232 barcode reader 184
             11    Options
                             About Options       186
                             Calibration 186
                                To zero a specific flow or pressure sensor     188
                                Column calibration 188
                             Communication 193
                               Configuring the IP address for the GC     193
                             Keyboard and Display      194
Operation Manual                                                                                   7
8   Operation Manual
Agilent 7890BGas Chromatograph
Operation Manual
1
Introduction
Chromatography Using a GC 10
The Operating Panel 11
                     Agilent Technologies              9
1    Introduction
Chromatography Using a GC
                      Chromatography is the separation of a mixture of
                      compounds into individual components.
Inlet
Detector
Oven
                       Power switch
                       Operating panel
10                                                                      Operation Manual
                                                                       Introduction   1
                      Display
                      Shows status, setpoints, current
                      activity, and messages.
                      Status lights
                      LEDs indicate general status, run
                      state, program state, external
                      control, and maintenance due.
                      Keyboard
                      Use to enter settings and program the
                      GC.
Operation Manual                                                                      11
1    Introduction
                    The display
                                    The display shows details of what is currently happening in
                                    the GC and allows you to make changes to parameters as
                                    necessary.
                                                          OVEN
                                        Temperature           150.0      150.0
                                        Initial time                     1.000<
                                        Rate 1                          20.000
                                                               Actual    Setpoint
                                    The cursor, <, shows the current active line. Use the scroll
                                    keys            to select a different line in the display and
                                    to view additional lines in the display.
                                    A blinking asterisk (*) prompts you to press [Enter] to store
                                    a value or [Clear] to abort the entry. You cannot perform any
                                    other task until this is done.
                    Status lights
                                    The status lights provide a basic look at what is currently
                                    happening inside the Agilent 7890B GC.
12                                                                                  Operation Manual
                                                                                Introduction   1
               Beeping instrument
                               A single beep means that a problem exists, but the problem
                               will not prevent the GC from executing the run. The GC will
                               emit one beep and display a message. The GC can start the
                               run and the warning disappears when a run starts.
                               A series of warning beeps sound if the GC encounters a
                               more serious problem. The GC starts with one beep. The
                               longer the problem persists, the more the GC beeps. For
                               example, a series of beeps sound if the front inlet gas flow
                               cannot reach setpoint. The message Front inlet flow shutdown is
                               briefly displayed. The flow shuts down after 2 minutes. Press
                               [Off/No] to stop the beeping.
                               A continuous beep sounds if a hydrogen flow is shut down
                               or a thermal shutdown occurs. Press [Clear] to stop the beep.
                               Fault messages indicate hardware problems that require user
                               intervention. Depending on the type of error, the GC will
                               beep once or not at all.
               Blinking setpoint
                               If a gas flow, multiposition valve, or the oven is shut down
                               by the system, Off or On/Off will blink on the appropriate
                               line of the components parameter listing.
Operation Manual                                                                               13
1    Introduction
                    The keypad
                                 All of the parameters required to operate the Agilent 7890B
                                 GC can be entered through the GCs keypad. Normally,
                                 however, most of these parameters are controlled using an
                                 attached data system, such as Agilents OpenLAB CDS or
                                 MassHunter software.
                                 When an Agilent data system is controlling your 7890B GC,
                                 it is possible for the data system to disable editing of the
                                 GCs current method from the keypad.
GC Component keys
Supporting keys
14                                                                              Operation Manual
Agilent 7890B Gas Chromatograph
Operation Manual
2
Operating Basics
Overview 16
Instrument Control 17
To Start Up the GC 18
To Shut Down the GC for Less Than a Week 19
To Shut Down the GC for More Than a Week 20
Correcting Problems 21
                     Agilent Technologies                   15
2    Operating Basics
Overview
                        Operating the GC involves the following tasks:
                         Setting up the GC hardware for an analytical method.
                         Starting up the GC. See To Start Up the GC.
                         Preparing the automatic liquid sampler. Install the
                          method- defined syringe; configure solvent and waste
                          bottle usage and syringe size; and prepare and load
                          solvent, waste, and sample vials.
                           For the 7693A ALS, see its Installation, Operation, and
                            Maintenance manual.
                           For the 7683 ALS, see manual Operating the 7683B
                            ALS on a 7890 Series GC.
                         Loading the analytical method or sequence into the GC
                          control system.
                           See the Agilent data system documentation.
                           For standalone GC operation see To load a method
                            and To load a stored sequence.
                         Running the method or sequence.
                           See the Agilent data system documentation.
                           For standalone GC operation, see To manually inject a
                            sample with a syringe and start a run, To run a
                            method to process a single ALS sample, and To start
                            running a sequence.
                         Monitoring sample runs from the GC control panel or the
                          Agilent data system program. See About GC Status or
                          the Agilent data system documentation.
                         Shutting down the GC. See To Shut Down the GC for
                          Less Than a Week or To Shut Down the GC for More
                          Than a Week.
16                                                                       Operation Manual
                                                               Operating Basics   2
Instrument Control
                     The Agilent 7890B GC is typically controlled by an attached
                     data system such as Agilent OpenLAB CDS. Alternately, the
                     GC can be controlled entirely from its keypad, with output
                     data being sent to an attached integrator for report
                     generation.
                     Agilent data system users  Please refer to the online help
                     included in the Agilent data system for details on how to
                     load, run, or create methods and sequences using the data
                     system.
                     Standalone GC users  If you are running your GC without
                     an attached data system, see the following for details on
                     loading methods and sequences from the keypad:
                      To load a method
                      To load a stored sequence
                     For details on running methods and sequences from the
                     keypad see:
                      To manually inject a sample with a syringe and start a
                       run
                      To run a method to process a single ALS sample
                      To start running a sequence
                     For details on how to create methods and sequences using
                     the GC keypad, see Methods and Sequences.
Operation Manual                                                                  17
2    Operating Basics
To Start Up the GC
                        Successful operation begins with a properly installed and
                        maintained GC. The utility requirements for gases, power
                        supply, venting of hazardous chemicals, and required
                        operational clearances around the GC are detailed in the
                        Agilent GC, GC/MS, and ALS Site Preparation Guide.
                        1 Check gas source pressures. For required pressures, see
                              the Agilent GC, GC/MS, and ALS Site Preparation Guide.
                        2 Turn on the carrier and detector gases at their sources
                              and open the local shutoff valves.
                        3 Turn on the cryo coolant at its source, if used.
                        4 Turn on the GC power. Wait for Power on successful to
                              display.
                        5 Install the column.
                        6 Check that the column fittings are leak free. See the
                              Troubleshooting manual.
                        7 Load the analytical method. See To load a method.
                        8 Wait for the detector(s) to stabilize before acquiring data.
                              The time required for the detector to reach a stable
                              condition depends on whether the detector was turned off
                              or its temperature was reduced while the detector
                              remained powered.
18                                                                                    Operation Manual
                                                                                  Operating Basics   2
                                  3 Turn off all gases, except the carrier gas, at their sources.
                                     (Leave the carrier gas on to protect the column from
                                     atmospheric contamination.)
                                  4 If you are using cryogenic cooling, turn off the cryo
                                     coolant at the gas source.
                                  5 Reduce detector, inlet, and column temperatures to
                                     between 150 and 200 C. If desired, the detector can be
                                     turned off. See the following table to determine if it is
                                     advantageous to shut down the detector for a short time
                                     period. The time required to return the detector to a
                                     stable condition is a factor. See Table 1.
Operation Manual                                                                                     19
2    Operating Basics
20                                                                                   Operation Manual
                                                                 Operating Basics   2
Correcting Problems
                      If the GC stops operation because of a fault, check the
                      display for any messages. Press [Status] and scroll to view
                      any additional messages.
                      1 Use the keyboard or data system to stop the alert tone.
                        Press [Off/No] on the keyboard or turn off the offending
                        component in the data system.
                      2 Resolve the problem, for example, by changing gas
                        cylinders or fixing the leak. See the Troubleshooting
                        Guide for details.
                      3 Once the problem is fixed, you may need to either power
                        cycle the instrument, or use the software keyboard or
                        data system to turn the problem component off, then on
                        again. For shutdown errors, you will need to do both.
Operation Manual                                                                    21
2    Operating Basics
22                      Operation Manual
Agilent 7890B Gas Chromatograph
Operation Manual
3
Keypad Operation
The Run Keys 24
The GC Component Keys 25
The Status Key 26
The Info Key 27
The General Data Entry Keys 28
The Supporting Keys 29
Method Storage and Automation Keys 30
Keypad Functionality When the GC Is Controlled by an Agilent Data
System 31
The Service Mode Key 31
About GC Status 32
About Logs 35
                      Agilent Technologies                          23
3    Keypad Operation
24                                                                    Operation Manual
                                                                    Keypad Operation    3
Operation Manual                                                                        25
3    Keypad Operation
26                                                                  Operation Manual
                                                          Keypad Operation   3
Operation Manual                                                             27
3    Keypad Operation
28                                                                        Operation Manual
                                                                    Keypad Operation   3
Operation Manual                                                                       29
3    Keypad Operation
30                                                                      Operation Manual
                                                                      Keypad Operation   3
Operation Manual                                                                         31
3    Keypad Operation
About GC Status
                               When the GC is ready to begin a run, the display screen
                               shows STATUS Ready for Injection. Alternately, when a
                               component of the GC is not ready to begin a run, the Not
                               Ready LED is lit on the status board. Press [Status] to see a
                               message explaining why the GC is not ready.
Status board
                Alert tones
                               A series of warning beeps sounds before a shutdown occurs.
                               The GC starts with one beep. The longer the problem
                               persists, the more the GC beeps. After a short time the
                               component with the problem shuts down, the GC emits one
                               beep, and a brief message is displayed. For example, a series
                               of beeps sounds if the front inlet gas flow cannot reach
                               setpoint. The message Front inlet flow shutdown is briefly
                               displayed. The flow shuts down after 2 minutes. Press
                               [Off/No] to stop the beep.
32                                                                            Operation Manual
                                                                               Keypad Operation   3
               Error conditions
                                  If a problem occurs, a status message appears. If the
                                  message indicates broken hardware, more information may
                                  be available. Press the applicable component key (for
                                  example, Front Det, Oven, or Front Inlet).
                                  When configured for operation with an MS using Smart
                                  Technologies (for example, a 5977 MSD), the GC will display
                                  message related to the MS. In this case, check the MS for
                                  more information.
               Blinking setpoint
                                  If the system shuts down a gas flow, multiposition valve, or
                                  the oven, Off will blink on the appropriate line of the
                                  components parameter listing.
                                  If there is a detector pneumatics shutdown or failure in
                                  another part of the detector, the detector On/Off line of the
                                  detectors parameter list blinks.
                                  For any flow or pressure parameter, and for oven
                                  temperature, go to the blinking parameter, then press
                                  [Off/No] to clear the fault. Resolve the problem if possible,
                                  then press [On/Yes] on the parameter to use it again. If the
                                  problem is not fixed, the fault will recur.
Operation Manual                                                                                  33
3    Keypad Operation
34                                                                   Operation Manual
                                                                          Keypad Operation   3
About Logs
                              Three logs are accessible from the keypad: the run log, the
                              maintenance log, and the system event log. To access the
                              logs, press [Logs] then scroll to the desired log and press
                              [Enter]. The display will indicate the number of entries the
                              log contains. Scroll through the list.
               Run log
                              The run log is cleared at the start of each new run. During
                              the run, any deviations from the planned method (including
                              keypad intervention) are listed in the run log table. When
                              the run log contains entries, the Run Log LED lights.
               Maintenance log
                              The maintenance log contains entries made by the system
                              when any of the user- defined component counters reach a
                              monitored limit. The log entry contains a description of the
                              counter, its current value, the monitored limits, and which of
                              its limits has been reached. In addition, each user task
                              related to the counter is recorded in the log, including
                              resetting, enabling or disabling monitoring, and changing
                              limits or units (cycles or duration).
Operation Manual                                                                             35
3    Keypad Operation
36                      Operation Manual
Agilent 7890B Gas Chromatograph
Operation Manual
4
Methods and Sequences
What Is a Method? 38
What Is Saved in a Method? 38
What Happens When You Load a Method? 39
Creating Methods 40
   To program a method 41
   To load a method 41
   To store a method 41
   Method mismatch 41
What Is a Sequence? 43
Creating Sequences 43
   About the priority sequence 44
   To program a sequence 44
   To program a priority sequence 45
   To program an ALS subsequence 45
   To program a valve subsequence 46
   To program post sequence events 46
   To store a sequence 46
   To load a stored sequence 47
   To determine sequence status 47
Automating Data Analysis, Method Development, and Sequence
Development 47
                    Agilent Technologies                     37
4    Methods and Sequences
What Is a Method?
                             A method is the group of settings required to analyze a
                             specific sample.
                             Since every type of sample reacts differently in the GCsome
                             samples require a higher oven temperature, others require a
                             lower gas pressure or a different detectora unique method
                             must be created for each specific type of analysis.
38                                                                         Operation Manual
                                                           Methods and Sequences   4
Operation Manual                                                                   39
4    Methods and Sequences
Creating Methods
                             A method is the group of setpoints needed to run a single
                             sample on the GC, such as oven temperature programs,
                             pressure programs, inlet temperatures, sampler parameters,
                             and so forth. A method is created by saving a group of
                             setpoints as a numbered method using the [Store] key.
                             The GC also can store several specialized methods. The GC
                             stores three methods used for resource conservation, called
                             SLEEP, CONDITION, and WAKE. When configured for use with
                             an attached MS, the GC also provides a method called MS
                             VENT, used to change GC setpoints to values appropriate for
                             a safe MS venting process. See Early Maintenance Feedback
                             on page 119 and GC- MS Features on page 131 for more
                             information about these specialized methods.
                             Components for which setpoint parameters can be stored are
                             shown in Table 2.
                             Component                          Component
                             Oven                               Aux temp
                             Valve 18                          Aux EPC
                             Front and back inlet               Aux column
                             Columns 1 to 6                     Aux detector 1 and 2
                             Front and back detector            Post run
                             Analog 1 and 2                     Run table
                             Front and back injector            Sample tray
40                                                                                     Operation Manual
                                                                      Methods and Sequences   4
               To program a method
                               1 Individually select each component for which setpoint
                                   parameters are appropriate for your method. (See
                                   Table 2.)
                               2   Examine the current setpoints and modify as desired.
                                   Repeat for each component as appropriate.
                               3   Examine the current setpoints for the ALS, if
                                   appropriate, and modify as desired.
                               4   Save the setpoints as a stored method. (See To store a
                                   method on page 41.)
               To load a method
                               1 Press [Load].
                               2 Press [Method].
                               3 Enter the number of the method to be loaded (1 through
                                   9).
                               4 Press [On/Yes] to load the method and replace the active
                                   method. Alternatively, press [Off/No] to return to the
                                   stored methods list without loading the method.
               To store a method
                               1 Ensure that the proper parameters are set.
                               2 Press [Method].
                               3 Scroll to the method to store, then press [Enter].
                               4 Press [On/Yes] to store the method and replace the active
                                   method. Alternatively, press [Off/No] to return to the
                                   stored methods list without storing the method.
               Method mismatch
                               This section applies only to a standalone (not connected to a
                               data system) GC. When a data system, such as OpenLAB
                               CDS or MassHunter, controls the GC, methods are stored in
                               the data system and can be edited there. See your data
                               system documentation for more information.
                               Suppose your standalone GC is equipped with a single FID.
                               You have created and saved methods that use this detector.
                               Now you remove the FID and install a TCD in its place.
Operation Manual                                                                              41
4    Methods and Sequences
42                                                                             Operation Manual
                                                             Methods and Sequences   4
What Is a Sequence?
                      A sequence is a list of samples to be analyzed along with the
                      method to be used for each analysis.
                      Refer to Running a Method or a Sequence from the Keypad
                      and Creating Sequences for details on how to create, load,
                      modify, and save sequences using the keypad.
Creating Sequences
                      A sequence specifies the samples to be run and the stored
                      method to be used for each. The sequence is divided into a
                      priority sequence (ALS only), subsequences (each of which
                      uses a single method), and post- sequence events
                       Priority sequence  allows you to interrupt a running ALS
                        or valve sequence to analyze urgent samples. (See About
                        the priority sequence on page 44.)
                       Subsequences  contain the stored method number and
                        information that defines a set of vials (or valve positions)
                        to be analyzed using a particular method. Sampler and/or
                        valve subsequences can be used in the same sequence.
                       Post sequence  names a method to be loaded and run
                        after the last run in the last subsequence. Specifies
                        whether the sequence is to be repeated indefinitely or
                        halted after the last subsequence.
                      Samples in each subsequence are specified as either ALS
                      tray locations or sampling valve positions (gas or liquid
                      sampling valves, often with a stream selection valve).
                      Five sequences with up to five subsequences each can be
                      stored.
Operation Manual                                                                     43
4    Methods and Sequences
               To program a sequence
                               1 Press [Seq]. (Press again, if necessary, to display
                                   subsequence information.)
                               2   Create a priority sequence, if desired. (See To program a
                                   priority sequence on page 45.) If you might want to use
                                   a priority sequence, you must program it now. (Once the
                                   sequence starts, you cannot edit it without stopping it.)
                               3   Scroll to the Method # line of Subseq 1 and enter a method
                                   number. Use 1 to 9 for the stored methods, 0 for the
                                   currently active method, or [Off/No] to end the sequence.
                               4   Press [Mode/Type] to select a valve or injector type. (See
                                   To program a valve subsequence on page 46 or To
                                   program an ALS subsequence on page 45.)
                               5   Create the next subsequence or scroll to Post Sequence.
                                   (See To program post sequence events on page 46.)
                               6   Save the completed sequence. (See To store a sequence
                                   on page 46.)
44                                                                              Operation Manual
                                                                       Methods and Sequences   4
Operation Manual                                                                               45
4    Methods and Sequences
               To store a sequence
                               1 Press [Store][Seq].
                               2   Enter an identifying number for the sequence (19).
                               3   Press [On/Yes] to store the sequence. Alternatively, press
                                   [Off/No] to cancel.
                                   A message is displayed if a sequence with the number
                                   you selected already exists.
                                    Press [On/Yes] to replace the existing sequence or
                                     [Off/No] to cancel.
                               Sequences can also be stored from within the stored
                               sequence list ([Seq]) by scrolling to the appropriate sequence
                               number and pressing the [Store] key.
46                                                                               Operation Manual
                                                                      Methods and Sequences   4
Operation Manual                                                                              47
4    Methods and Sequences
48                           Operation Manual
Agilent 7890B Gas Chromatograph
Operation Manual
5
Running a Method or a Sequence from
the Keypad
Running Methods from the Keypad 50
Running Sequences from the Keypad 51
                     Agilent Technologies                  49
5    Running a Method or a Sequence from the Keypad
                To abort a method
                                      1 Press [Stop].
                                      2 When you are ready to resume running analyses, load the
                                         appropriate sequence or method. (See "To load a method"
                                         or "To load a stored sequence".)
50                                                                                     Operation Manual
                                                  Running a Method or a Sequence from the Keypad   5
                               Ready wait
                               If a sequence is started but the instrument is not ready (due
                               to oven temperature, equilibration times, and so forth), the
                               sequence will not start until all instrument setpoints are
                               ready.
Operation Manual                                                                                   51
5    Running a Method or a Sequence from the Keypad
                Aborting a sequence
                                      When a sequence is aborted, it stops immediately without
                                      waiting for the current run to finish.
                                      The following will cause a sequence to abort:
                                       The [Stop] key is pressed.
                                       A sampler error occurs, producing an error message.
                                       The GC detects a configuration mismatch during a method
                                        load.
                                       A running sequence tries to load a method that doesnt
                                        exist.
                                       The sampler is turned off. You can correct the problem
                                        and then resume the sequence. The aborted sample run
                                        will be repeated.
52                                                                                     Operation Manual
Agilent 7890B Gas Chromatograph
Operation Manual
6
Chromatographic Checkout
                     Agilent Technologies                  53
6    Chromatographic Checkout
54                                                                             Operation Manual
                                                                        Chromatographic Checkout          6
                          For a new GC, check the installed inlet liner. The liner shipped in the inlet
                   NOTE
                          may not be the liner recommended for checkout.
Operation Manual                                                                                      55
6    Chromatographic Checkout
56                                                                                        Operation Manual
                                                             Chromatographic Checkout   6
Operation Manual                                                                        57
6    Chromatographic Checkout
58                                                                                           Operation Manual
                                                                    Chromatographic Checkout   6
Operation Manual                                                                               59
6    Chromatographic Checkout
60                                                                                          Operation Manual
                                                                         Chromatographic Checkout   6
          FID1 A, (C:\FID.D)
 pA
                                                                           C15      C16
400
350
300
250
200
150
100
                                                         C13       C14
50
 0
      0                        1   2            3              4            5             mi
Operation Manual                                                                                    61
6    Chromatographic Checkout
62                                                                                      Operation Manual
                                                                    Chromatographic Checkout   6
Operation Manual                                                                               63
6    Chromatographic Checkout
64                                                                                       Operation Manual
                                                               Chromatographic Checkout    6
Operation Manual                                                                           65
6         Chromatographic Checkout
25 uV
    70
50
40
30
     20
                         2           4                       6               8
                                         Time (min.)
66                                                                                         Operation Manual
                                                             Chromatographic Checkout   6
Operation Manual                                                                        67
6    Chromatographic Checkout
68                                                                                             Operation Manual
                                                                   Chromatographic Checkout   6
Operation Manual                                                                              69
6    Chromatographic Checkout
70                                                                                         Operation Manual
                                                                          Chromatographic Checkout   6
60 Azobenzene
50
40
30
20
                                                 Octadecane
10
1 2 3 4 5 6 7 8 9 min
Operation Manual                                                                                     71
6    Chromatographic Checkout
72                                                                              Operation Manual
                                                                      Chromatographic Checkout   6
Operation Manual                                                                                 73
6    Chromatographic Checkout
74                                                                                          Operation Manual
                                                               Chromatographic Checkout   6
                   Injection volume                    1 L
                   Syringe size                        10 L
                   Solvent A pre washes                2
                   Solvent A post washes               2
                   Solvent A wash volume               8
                   Solvent B pre washes                0
                   Solvent B post washes               0
                   Solvent B wash volume               0
                   Injection mode (7693A)              Normal
                   Airgap Volume (7693A)               0.20
                   Viscosity delay                     0
                   Inject Dispense Speed (7693A)       6000
                   Plunger speed (7683)                Fast, for all inlets except COC.
                   PreInjection dwell                  0
                   PostInjection dwell                 0
                   Manual injection
                   Injection volume                    1 L
                   Data system
                   Data rate                           5 Hz
Operation Manual                                                                          75
6        Chromatographic Checkout
          ECD1 B, (C:\ECD.D)
     Hz
                                                 Lindane
12000
                                                 (18713-60040
                                                 5183-0379)
10000
8000
6000                                                            Aldrin
                                                                (18713-60040)
4000
2000
     0
                     2              4     6          8           10             12 min
76                                                                                       Operation Manual
                                                                   Chromatographic Checkout     6
               Preparation
                             1 Gather the following:
                                Evaluation column, HP- 5 30 m  0.32 mm  0.25 m
                                 (19091J- 413)
                                FPD performance evaluation (checkout) sample
                                 (5188- 5953), 2.5 mg/L ( 0.5%) methylparathion in
                                 isooctane
                                Phosphorus filter
                                Sulfur filter and filter spacer
                                4- mL solvent and waste bottles or equivalent for
                                 autoinjector.
                                2- mL sample vials or equivalent for sample.
                                Chromatographic- grade isooctane for syringe wash
                                 solvent.
                                Inlet and injector hardware (See To Prepare for
                                 Chromatographic Checkout.)
                             2 Verify the following:
                                Capillary column adapter installed. If not, install it.
                                 Chromatographic- grade gases plumbed and configured:
                                  helium as carrier gas, nitrogen, hydrogen, and air.
                                 Empty waste vials loaded in sample turret.
                                 4- mL vial with diffusion cap filled with isooctane and
                                  inserted in Solvent A injector position.
                             3 Replace consumable parts (liner, septum, traps, syringe,
                               and so forth) as needed for the checkout. See To Prepare
                               for Chromatographic Checkout.
                             4 Verify that the Lit Offset is set appropriately. Typically, it
                               should be about 2.0 pA for the checkout method.
                             5 Install the evaluation column. (See the procedure for the
                               SS, PP, COC, MMI, or PTV in the Maintenance manual.)
                                Set the oven, inlet, and detector to 250 C and bake
                                 out for at least 15 minutes. (See the procedure for the
                                 SS, PP, COC, MMI, or PTV in the Maintenance manual.)
                                Be sure to configure the column.
Operation Manual                                                                                77
6    Chromatographic Checkout
                Phosphorus performance
                                1 If it is not already installed, install the phosphorus filter.
                                2 Create or load a method with the parameter values listed
                                    in Table 10.
78                                                                                               Operation Manual
                                                                    Chromatographic Checkout   6
Operation Manual                                                                               79
6    Chromatographic Checkout
80                                                                                          Operation Manual
                                                        Chromatographic Checkout   6
Operation Manual                                                                   81
6    Chromatographic Checkout
Methylparathion
Isooctane
                Sulfur performance
                                1 Install the sulfur filter and filter spacer.
                                2 Ignite the FPD flame if not lit.
                                3 Display the signal output and monitor. This output
                                  typically runs between 50 and 60 but can be as high as
                                  70. Wait for the output to stabilize. This takes
                                  approximately 1 hour.
                                  If the baseline output is too high:
                                   Check column installation. If installed too high, the
                                    stationery phase burns in the flame and increases
                                    measured output.
                                   Check for leaks.
                                   Bake out the detector and column at 250 C.
                                   Wrong flows set for installed filter.
82                                                                               Operation Manual
                                                         Chromatographic Checkout   6
Methylparathion
Isooctane
Operation Manual                                                                    83
6    Chromatographic Checkout
                Preparation
                                1 Gather the following:
                                   Evaluation column, DB5 15 m  0.32 mm  1.0 m
                                    (123- 5513)
                                   FPD performance evaluation (checkout) sample
                                    (5188- 5245, Japan), composition: n- Dodecane
                                    7499 mg/L ( 5%), Dodecanethiol 2.0 mg/L ( 5%),
                                    Tributyl Phosphate 2.0 mg/L ( 5%), tert- Butyldisulfide
                                    1.0 mg/L ( 5%), in isooctane as solvent
                                   Phosphorus filter
                                   Sulfur filter and filter spacer
                                   4- mL solvent and waste bottles or equivalent for
                                    autoinjector.
                                   2- mL sample vials or equivalent for sample.
                                   Chromatographic- grade isooctane for syringe wash
                                    solvent.
                                   Inlet and injector hardware (See To Prepare for
                                    Chromatographic Checkout.)
                                2 Verify the following:
                                   Capillary column adapter installed. If not, install it.
                                   Chromatographic- grade gases plumbed and configured:
                                    helium as carrier gas, nitrogen, hydrogen, and air.
                                   Empty waste vials loaded in sample turret.
                                   4- mL vial with diffusion cap filled with isooctane and
                                    inserted in Solvent A injector position.
                                3 Replace consumable parts (liner, septum, traps, syringe,
                                  and so forth) as needed for the checkout. See To Prepare
                                  for Chromatographic Checkout.
                                4 Verify the lit offset is set appropriately. Typically, it
                                  should be about 2.0 pA for the checkout method.
                                5 Install the evaluation column. (See the procedure for the
                                  SS, PP, COC, MMI, or PTV in the Maintenance manual.)
84                                                                                Operation Manual
                                                                             Chromatographic Checkout   6
               Phosphorus performance
                             1 If it is not already installed, install the phosphorus filter.
                             2 Create or load a method with the parameter values listed
                                 in Table 12.
Operation Manual                                                                                        85
6    Chromatographic Checkout
86                                                                                         Operation Manual
                                                                   Chromatographic Checkout   6
Operation Manual                                                                              87
6    Chromatographic Checkout
88                                                                               Operation Manual
                                                                        Chromatographic Checkout   6
Tributylphosphate
Isooctane t-Butyldisulfide
               Sulfur performance
                                1 Install the sulfur filter.
                                2 Ignite the FPD flame, if not lit.
                                3 Display the signal output and monitor. This output
                                   typically runs between 50 and 60 but can be as high as
                                   70. Wait for the output to stabilize. This takes
                                   approximately 2 hours.
                                   If the baseline output is too high:
                                    Check column installation. If installed too high, the
                                     stationery phase burns in the flame and increases
                                     measured output.
                                    Check for leaks.
                                    Bake out the detector and column at 250 C.
                                    Wrong flows set for installed filter
                                   If the baseline output is zero, verify the electrometer is on
                                   and the flame is lit.
                                4 If using a data system, prepare the data system to
                                   perform one run using the loaded checkout method. Make
                                   sure the data system will output a chromatogram.
Operation Manual                                                                                   89
6    Chromatographic Checkout
t-Butyldisulfide
1-Dodecanethiol
Isooctane
90                                                                                      Operation Manual
                                                                   Chromatographic Checkout     6
               Preparation
                             1 Gather the following:
                                Evaluation column, HP- 5 30 m  0.32 mm  0.25 m
                                 (19091J- 413)
                                FPD performance evaluation (checkout) sample
                                 (5188- 5953), 2.5 mg/L ( 0.5%) methylparathion in
                                 isooctane
                                Phosphorus filter
                                Sulfur filter and filter spacer
                                4- mL solvent and waste bottles or equivalent for
                                 autoinjector.
                                2- mL sample vials or equivalent for sample.
                                Chromatographic- grade isooctane for syringe wash
                                 solvent.
                                Inlet and injector hardware (See To Prepare for
                                 Chromatographic Checkout.)
                             2 Verify the following:
                                Capillary column adapter installed. If not, install it.
                                 Chromatographic- grade gases plumbed and configured:
                                  helium as carrier gas, nitrogen, hydrogen, and air.
                                 Empty waste vials loaded in sample turret.
                                 4- mL vial with diffusion cap filled with isooctane and
                                  inserted in Solvent A injector position.
                             3 Replace consumable parts (liner, septum, traps, syringe,
                               and so forth) as needed for the checkout. See To Prepare
                               for Chromatographic Checkout.
                             4 Verify that the Lit Offset is set appropriately. Typically, it
                               should be about 2.0 pA for the checkout method.
                             5 Install the evaluation column. (See the procedure for the
                               SS, PP, COC, MMI, or PTV in the Maintenance manual.)
                                Set the oven, inlet, and detector to 250 C and bake
                                 out for at least 15 minutes. (See the procedure for the
                                 SS, PP, COC, MMI, or PTV in the Maintenance manual.)
                                Be sure to configure the column.
Operation Manual                                                                                91
6    Chromatographic Checkout
                Phosphorus performance
                                1 If it is not already installed, install the phosphorus filter.
                                2 Create or load a method with the parameter values listed
                                    in Table 10.
92                                                                                         Operation Manual
                                                              Chromatographic Checkout   6
Operation Manual                                                                         93
6    Chromatographic Checkout
94                                                                                       Operation Manual
                                                        Chromatographic Checkout   6
Operation Manual                                                                   95
6    Chromatographic Checkout
Methylparathion
Isooctane
                Sulfur performance
                                1 Install the sulfur filter and filter spacer.
                                2 Make the following method parameter changes.
96                                                                               Operation Manual
                                                        Chromatographic Checkout   6
Operation Manual                                                                   97
6    Chromatographic Checkout
Methylparathion
Isooctane
98                                                Operation Manual
                                                                   Chromatographic Checkout   6
               Preparation
                             1 Gather the following:
                                Evaluation column, DB5 15 m  0.32 mm  1.0 m
                                 (123- 5513)
                                FPD performance evaluation (checkout) sample
                                 (5188- 5245, Japan), composition: n- Dodecane
                                 7499 mg/L ( 5%), Dodecanethiol 2.0 mg/L ( 5%),
                                 Tributyl Phosphate 2.0 mg/L ( 5%), tert- Butyldisulfide
                                 1.0 mg/L ( 5%), in isooctane as solvent
                                Phosphorus filter
                                Sulfur filter and filter spacer
                                4- mL solvent and waste bottles or equivalent for
                                 autoinjector.
                                2- mL sample vials or equivalent for sample.
                                Chromatographic- grade isooctane for syringe wash
                                 solvent.
                                Inlet and injector hardware (See To Prepare for
                                 Chromatographic Checkout.)
                             2 Verify the following:
                                Capillary column adapter installed. If not, install it.
                                Chromatographic- grade gases plumbed and configured:
                                 helium as carrier gas, nitrogen, hydrogen, and air.
                                Empty waste vials loaded in sample turret.
                                4- mL vial with diffusion cap filled with isooctane and
                                 inserted in Solvent A injector position.
                             3 Replace consumable parts (liner, septum, traps, syringe,
                               and so forth) as needed for the checkout. See To Prepare
                               for Chromatographic Checkout.
                             4 Verify the lit offset is set appropriately. Typically, it
                               should be about 2.0 pA for the checkout method.
Operation Manual                                                                              99
6     Chromatographic Checkout
                 Phosphorus performance
                                 1 If it is not already installed, install the phosphorus filter.
                                 2 Create or load a method with the parameter values listed
                                     in Table 12.
                                 Initial temp                     70 C
                                 Initial time                     0 min
                                 Rate 1                           10 C/min
                                 Final temp                       105 C
                                 Final time                       0 min
                                 Rate 2                           20 C/min
                                 Final temp 2                     190 C
                                 Final time 2                     7.25 min for sulfur
                                                                  12.25 min for phosphorus
                                 ALS settings (if installed)
                                 Sample washes                    2
                                 Sample pumps                     6
                                 Sample wash volume               8
                                 Injection volume                 1 L
                                 Syringe size                     10 L
                                 Solvent A pre washes             2
                                 Solvent A post washes            2
                                 Solvent A wash volume            8
                                 Solvent B pre washes             0
                                 Solvent B post washes            0
                                 Solvent B wash volume            0
                                 Injection mode (7693A)           Normal
                                 Airgap Volume (7693A)            0.20
                                 Viscosity delay                  0
                                 Inject Dispense Speed (7693A)    6000
                                 Plunger speed (7683)             Fast, for all inlets except COC.
                                 PreInjection dwell               0
                                 PostInjection dwell              0
                                 Manual injection
                                 Injection volume                 1 L
                                 Data System
                                 Data rate                        5 Hz
Tributylphosphate
Isooctane t-Butyldisulfide
                 Sulfur performance
                                       1 Install the sulfur filter.
                                       2 Make the following method parameter changes.
t-Butyldisulfide
1-Dodecanethiol
Isooctane
7
Resource Conservation
Resource Conservation
                                 The 7890B GC provides an instrument schedule to conserve
                                 resources such as electricity and gases. Using the instrument
                                 schedule, you can create sleep, wake, and conditioning
                                 methods that allow you to program resource usage. A SLEEP
                                 method sets low flows and temperatures. A WAKE method
                                 sets new flows and temperatures, typically to restore
                                 operating conditions. A CONDITION method sets flows and
                                 temperatures for a specific run time, typically high enough
                                 to clean out contamination if present.
                                 Load the sleep method at a specified time during the day to
                                 reduce flows and temperatures. Load the wake or condition
                                 method to restore analytical settings before operating the GC
                                 again. For example, load the sleep method at the end of each
                                 day or work week, then load the wake or condition method
                                 an hour or so before arriving to work the next day.
                 Sleep Methods
                                 Create a sleep method to reduce gas and power usage during
                                 times of reduced activity.
                                 When creating a sleep method, consider the following:
                                  The detector. While you can reduce temperatures and gas
                                   usage, consider the stabilization time required to prepare
                                   the detector for use. See Table 1, Detector stabilization
                                   times, on page 18. The power savings is minimal.
                                  Connected devices. If connected to an external device
                                   such as a mass spectrometer, set compatible flows and
                                   temperatures.
                                  The columns and oven. Be sure to maintain sufficient
                                   flow to protect the columns at the temperature set for the
                                   oven. You may need to experiment to find the best
                                   reduced flow rate and temperature. Also consider whether
                                   the additional thermal cycling might loosen fittings,
                                   especially for MS transfer line connections. In this case,
                                   consider keeping the oven temperature > 110 C.
                                  The inlets. Maintain sufficient flow to prevent
                                   contamination.
                                  Cryo cooling. Devices which use cryo cooling may start
                                   immediately using cryogen if the wake method requires it.
                                 See Table 14 below for general recommendations.
GC Component                    Comment
Columns and oven                 Reduce temperature to save power.
                                 Turn off to save the most power.
                                 Maintain some carrier gas flow to protect the columns.
Inlets                          For all inlets:
                                 Reduce temperatures. Reduce temperatures to 40 C or Off to save the most
                                   power.
Split/splitless                  Use split mode to prevent diffusion of contamination from the vent line. Use
                                   reduced split ratio.
                                 Reduce pressure. Consider using current Gas Saver levels, if used.
Cool on-column                   Reduce pressure.
                                 Consider reducing septum purge flow.
Multimode                        Use split mode to prevent diffusion of contamination from the vent line. Use
                                   reduced split ratio.
                                 Reduce pressure. Consider using current Gas Saver levels, if used.
Purged packed                    Reduce pressure.
                                 Consider reducing septum purge flow.
Volatiles interface              Reduce pressure.
                                 Consider reducing septum purge flow.
Detectors
FID                              Turn off the flame. (This turns off hydrogen and air flows.)
                                 Reduce temperatures. (Keep at or above 100 C to reduce contamination and
                                   condensation.)
                                 Turn off makeup flow.
FPD+                             Turn off the flame. (This turns off hydrogen and air flows.)
                                 Reduce temperatures. (Keep at or above 100 C to reduce contamination and
                                   condensation.)
                                 Turn off makeup flow.
ECD                             Reduce makeup flow. Try using 1520 mL/min and test results.
                                 Maintain temperature to avoid long recovery/stabilization times.
NPD                              Maintain flows and temperatures. Sleep not recommended due to recovery
                                   times and also thermal cycling can reduce bead life.
TCD                              Leave filament on.
                                 Leave block temperature on.
                                 Reduce reference and makeup flows.
FPD                              Maintain flows and temperatures. Sleep not recommended.
GC Component                     Comment
Other devices
Valve box                         Reduce temperature. (Keep valve box temperature high enough to prevent
                                   sample condensation, if applicable.)
Aux thermal zones                 Reduce or turn off. Also refer to connected device manuals, for example for a
                                   connected MSD.
Aux pressures or flows            Reduce or turn off as appropriate for connected columns, transfer lines, and so
                                   forth. Always refer to connected device or instrument manuals, for example for
                                   a connected MSD, to maintain at least the minimum recommended flows or
                                   pressures.
                                 INST SCHEDULE
                         Go to sleep now.             <
                         Edit the Instrument Schedule?
                         Edit clock table?
                                            Actual   Setpoint
                                    INST SCHEDULE ( 2 of 2)
                                 Monday:                 07:30<
                                 Type:    Wake with WAKE file.
Actual Setpoint
                                    INST SCHEDULE
                            Go to sleep now.             <
                            Edit the Instrument Schedule?
                            Edit clock table?
                                              Actual   Setpoint
                               INST SCHEDULE ( 3 of 9)
                            Monday:                 05:30<
                            Type: Front det adjust offset
Actual Setpoint
                                    INST SCHEDULE
                            Delete this event?
                            YES to delete,
                            NO to cancel
                                              Actual   Setpoint
                                     INST SCHEDULE
                             Go to sleep now.             <
                             Edit the Instrument Schedule?
                             Edit clock table?
                                              Actual   Setpoint
8
Early Maintenance Feedback
Early Maintenance Feedback (EMF) 120
Default Thresholds 122
Available Counters 123
To Enable or Change a Limit for an EMF Counter 126
To Disable an EMF Counter 127
To Reset an EMF Counter 128
EMF Counters for Autosamplers 129
EMF Counters for MS Instruments 130
                 Counter types
                                   Injection counters increment whenever an injection occurs
                                   on the GC via an ALS injector, headspace sampler, or
                                   sampling valve. Manual injections do not increment counters.
                                   The GC differentiates between front and back injections, and
                                   only increments counters associated with the configured
                                   injection flow path.
                                   For example, consider the following GC:
               Thresholds
                            The EMF feature provides two warning thresholds: Service due
                            and Service warning.
                             Service Due: When the counter exceeds this number of
                              injections or days, the Service Due indicator lights and an
                              entry is made in the Maintenance Log. The Service Due limit
                              must be larger than the Service warning limit.
                             Service warning: When the counter exceeds this number of
                              injections or days, the instrument status shows a
                              reminder that the component may need maintenance
                              soon.
                            Both thresholds are set independently for each counter. You
                            can enabled one or both, as desired.
Default Thresholds
                                   Selected counters have default thresholds to use as a
                                   starting point. To view any available information for a
                                   counter:
                                   1 Navigate to the desired counter and press [Enter]. See To
                                     Enable or Change a Limit for an EMF Counter.
                                   2 Scroll to the counters Service Due entry and press
                                     [Mode/Type]. If available, the default threshold for the
                                     counter displays. Press [Clear] to return to the counter.
                                   If a default limit is not suggested, enter a conservative limit
                                   based on your experience. Use the warning feature to alert
                                   you when service is approaching, then track performance to
                                   determine if the Service Due threshold is too high or too low.
                                   For any EMF counter, you may need to adjust the threshold
                                   values based on the demands of your applications.
Available Counters
                                        Table 15 lists the most common counters available. The
                                        available counters will vary based on the installed GC
                                        options, consumables, and future updates.
9
GC-MS Features
GC/MS Communications 132
To Set Up a Vent Method 134
To Manually Prepare the GC for Venting the MS 135
To Manually Exit the MS Vent State 136
To Use the GC When the MSD is Shut Down 137
To Enable or Disable MS Communications 138
GC/MS Communications
                              When the 7890B GC and a 5977 Series MSD (or other MS
                              instrument with Smart Technologies) are configured together,
                              they communicate with and react to each other. (See the
                              Installation and First Startup manual for configuration
                              details.) The two most important events that cause
                              interaction are venting the MSD and an MS shutdown.
                MS Shutdown events
                              When configured with a 5977 Series MSD, the following
                              events will cause an MS Shutdown in the GC:
                               Loss of communications with the MS, when not venting
                                the MS. (Requires no communications for a length of
                                time.)
                               MSD reports a high vacuum pump failure.
                              When the GC enters an MS shutdown:
                               The GC aborts any current run.
                               The oven is set to 50 C. When it reaches that setpoint, it
                                turns off.
                               The MS transfer line temperature is turned off.
                   CAUTION   Manually exiting the MS Vent state when the GC and MS are still
                             connected and the MS is venting or off can damage the MS if you
                             set inappropriate flows.
10
Configuration
About Configuration 140
   Assigning GC resources to a device 140
   Setting configuration properties 141
General Topics 142
   To Unlock the GC Configuration 142
   Ignore Ready = 142
   Information displays 143
   Unconfigured: 143
Oven 144
Front Inlet/Back Inlet 147
   To configure the PTV or COC coolant 147
   To configure the MMI coolant 149
Column # 151
   To configure a single column 152
   To configure multiple columns 155
Composite Columns 160
LTM Columns 162
Cryo Trap 163
Front Detector/Back Detector/Aux Detector/Aux Detector 2 165
Analog out 1/Analog out 2 167
   Fast peaks 167
Valve Box 168
Thermal Aux 169
   To assign a GC power source to an Aux thermal zone 169
   To configure a MSD transfer line heater 170
   To configure a nickel catalyst heater 170
   To configure an ion trap transfer line heater 171
PCM A/PCM B/PCM C 172
Pressure aux 1,2,3/Pressure aux 4,5,6/Pressure aux 7,8,9 174
Status 175
Time 176
Valve # 177
Front injector/Back injector 178
Sample tray (7683 ALS) 180
Instrument 181
Using the Optional Barcode Reader 182
About Configuration
                               Configuration is a two- part process for most GC accessory
                               devices that require power and/or communication resources
                               from the GC. In the first part of the configuration process, a
                               power and/or communication resource is assigned to the
                               device. The second part of the configuration process allows
                               setting of any configuration properties associated with the
                               device.
General Topics
               Ignore Ready =
                                The states of the various hardware elements are among the
                                factors that determine whether the GC is Ready for analysis.
                                Under some circumstances, you may not wish to have a
                                specific element readiness considered in the GC readiness
                                determination. This parameter lets you make that choice.
                                The following elements allow readiness to be ignored: inlets,
                                detectors, the oven, PCM, and auxiliary EPC modules.
                                For example, suppose an inlet heater is defective but you
                                dont plan to use that inlet today. By setting Ignore Ready =
                                TRUE for that inlet, you can use the rest of the GC. After the
                                heater is repaired, set Ignore Ready = FALSE or the run could
                                start before that inlets conditions are ready.
                                To ignore an element's readiness, press [Config], then select
                                the element. Scroll to Ignore Ready and press [On/Yes] to set it
                                to True.
               Information displays
                               Below are some examples of configuration displays:
               Unconfigured:
                               Accessory devices requiring GC power or communication
                               must be assigned these GC resources before they can be
                               used. To make this hardware element usable, first To
                               Unlock the GC Configuration on page 142 then go to the
                               Unconfigured parameter and press [Mode/Type] to install it. If
                               the hardware element you are configuring requires selection
                               of additional parameters, the GC asks for that selection. If
                               no parameters are required, press [Enter] at the GC prompt
                               to install that element. You are required to power the GC off
                               and then power the GC on to complete this configuration.
                               After restarting the GC, a message reminding you of this
                               change and its effect on the default method is displayed. If
                               needed, change your methods to accommodate the new
                               hardware.
Oven
                   See Unconfigured: on page 143 and Ignore Ready = on
                   page 142.
                   Slow oven cool down mode On reduces the oven fan speed
                   during the cool down cycle.
                               Cryo timeout Cryo timeout occurs, and the oven shuts off,
                               when a run does not start within a specified time (10 to 120
                               minutes) after the oven equilibrates. Turning cryo timeout
                               off disables this feature. We recommend that it be turned on
                               because cryo timeout conserves coolant at the end of a
                               sequence or if automation fails. A Post Sequence method
                               could also be used.
                   Cryo fault Shuts the oven down if it does not reach setpoint
                   temperature after 16 minutes of continuous cryo operation.
                   Note that this is the time to reach the setpoint, not the time
                   to stabilize and become ready at the setpoint. For example,
                   with a cool on- column inlet and cryo control in the oven
                   track mode, it may take the oven 20 to 30 minutes to
                   achieve readiness.
                   If the temperature goes below the minimum allowed
                   temperature (90C for liquid nitrogen, 70C for liquid
                   CO2), the oven will shut down.
                   The COC and PTV inlets must use the same cryo type as
                   configured for the oven.
                   Shutdown behavior
                   Both Cryo timeout and Cryo fault can cause cryo shutdown.
                   If this happens, the inlet heater is turned off and the cryo
                   valve closes. The GC beeps and displays a message.
                   The inlet heater is monitored to avoid overheating. If the
                   heater remains on at full power for more than 2 minutes,
                   the heater is shut down. The GC beeps and displays a
                   message.
                   To recover from either condition, turn the GC off, then on,
                   or enter a new setpoint.
                   Shutdown behavior
                   Both Cryo timeout and Cryo fault can cause cryo shutdown.
                   If this happens, the inlet heater is turned off and the cryo
                   valve closes. The GC beeps and displays a message.
                   The inlet heater is monitored to avoid overheating. If the
                   heater remains on at full power for more than 2 minutes,
                   the heater is shut down. The GC beeps and displays a
                   message.
                   To recover from either condition, turn the GC off, then on,
                   or enter a new setpoint.
Column #
Actual Setpoint
                   A simple example
                   An analytical column is attached at its inlet end to a
                   split/splitless inlet located at the front of the GC and the
                   column outlet is attached to an FID located at the front
                   detector position.
                                                Complicated example
                                                The inlet feeds the analytical column which ends at a
                                                three- way splitter. The splitter has the column effluent and
                                                makeup gas coming in, and transfer lines (non- coated
                                                columns) to three different detectors. This is a case where a
                                                sketch is necessary.
1.444 m x 0.18 mm x 0 m
                                                The oven was chosen for the MSD line since most of it is in
                                                the oven.
                                                As in the previous examples, your analytical method can
                                                control the flow of column # 1 which has a GC pressure
                                                controlled inlet and outlet.
Composite Columns
                                  A composite column is a capillary column that passes
                                  through multiple heating zones. A composite column consists
                                  of a main segment and one or more additional segments.
                                  There may be one segment on the input side of the main
                                  segment (In Segment) and up to two segments on its output
                                  side (Out Segment, Segment 2). Each of the four segments'
                                  lengths, diameters, and film thicknesses can be specified
                                  separately. Also, the zones that determine the temperatures
                                  of each of the four segments are specified separately. The
                                  three additional segments are often uncoated (zero film
                                  thickness) and, serving as connectors, are of shorter length
                                  than the main segment. It is necessary to specify these
                                  additional segments so that the flow- pressure relationship
                                  for the composite column can be determined.
                                  Composite columns differ from multiple columns because for
                                  composite columns, 100% of the column flow continues
                                  through a single column or through multiple column
                                  segments without additional makeup gas.
                       GC Inlet                                           Detector
                                                   Out segment
MSD
LTM Columns
                              See Unconfigured: on page 143 and Ignore Ready = on
                              page 142.
                              Low Thermal Mass (LTM) controllers and columns mount on
                              the front door of the GC and connect to LVDS connectors
                              [A- DET 1], [A- DET 2], or [EPC 6].
                              Press [Config][Aux Col #], enter the desired LTM column
                              number [1-4], and configure as a composite column. See
                              Composite Columns on page 160.
Cryo Trap
                                This discussion assumes that the trap is mounted in position
                                B, that you use liquid nitrogen coolant and control the trap
                                with Thermal Aux 1.
                                Configuration is in several parts:
                                 Configure the trap to the GC
                                 Configure a heater to the cryo trap.
                                 Configure the coolant.
                                 Configure the user- configurable heater.
                                 Reboot the GC.
               Reboot the GC
                               Press [Options], select Communications, and press [Enter]. Select
                               Reboot GC and press [On/Yes] twice.
               Lit offset
                               The GC monitors the difference between the detector output
                               with the flame lit and the output when the flame is not lit.
                               If this difference falls below the setpoint, the GC assumes
                               that the flame has gone out and tries to reignite it. See the
                               Advanced Operation manual for details on how to set the Lit
                               Offset:
                                 FID
                                 FPD
                               If set too high, the lit detector baseline output can be below
                               the Lit Offset setpoint, causing the GC to erroneously try to
                               reignite the flame.
               Fast peaks
                            The GC allows you to output analog data at two speeds. The
                            faster speedto be used only with the FID, FPD, and
                            NPDallows minimum peak widths of 0.004 minutes (8 Hz
                            bandwidth), while the standard speedwhich can be used
                            with all detectors allows minimum peak widths of 0.01
                            minutes (3.0 Hz bandwidth).
                            To use fast peaks:
                            1 Press [Config][Analog out 1] or [Config][Analog out 2].
                            2 Scroll to Fast peaks and press [On/Yes].
Valve Box
                               See Unconfigured: on page 143 and Ignore Ready = on
                               page 142.
                               The valve box mounts on top of the column oven. It may
                               contain up to four valves mounted on heated blocks. Each
                               block can accommodate two valves.
                               Valve positions on the blocks are numbered. We suggest that
                               valves be installed in the blocks in numeric order.
                               All heated valves in a valve box are controlled by the same
                               temperature setpoint.
Thermal Aux
                              See Unconfigured: on page 143 and Ignore Ready = on
                              page 142.
                              The auxiliary thermal controllers provide up to three
                              channels of temperature control. These controllers are
                              labeled Thermal Aux 1, Thermal Aux 2, and Thermal Aux 3.
               To configure a PCM
                              1 Press [Config][Aux EPC #], scroll to the PCMx and press
                                [Enter].
                              2 Scroll to Gas type, press [Mode/Type], make a selection and
                                    press [Enter].
Status
                                The [Status] key has two tables associated with it. You switch
                                between them by pressing the key.
Time
                               Press [Time] to open this function. The first line always
                               displays the current date and time, and the last line always
                               displays a stopwatch. The two middle lines vary:
                               During Post Run     Show last run time and remaining Post Run
                               time.
Valve #
                               Up to 4 valves can be mounted in a temperature- controlled
                               valve box and are normally wired to the valve box bracket
                               V1 through V4 plugs, located inside the electrical
                               compartment. Additional valves or other devices (4 through
                               8) can be wired using the plug labeled EVENT on the back of
                               the GC.
               To configure a valve
                               1 Press [Config][Valve #] and enter the number (1 to 8) of
                                  the valve you are configuring. The current valve type is
                                  displayed.
                               2 To change the valve type, press [Mode/Type], select the
                                  new valve type, and press [Enter].
                               Valve types
                                Sampling Two- position (load and inject) valve. In load
                                 position, an external sample stream flows through an
                                 attached (gas sampling) or internal (liquid sampling) loop
                                 and out to waste. In inject position, the filled sampling
                                 loop is inserted into the carrier gas stream. When the
                                 valve switches from Load to Inject, a run starts if one is
                                 not already in progress. See the Advanced Operation
                                 manual for details.
                                Switching Two- position valve with four, six, or more
                                 ports. These are general- purpose valves used for such
                                 tasks as column selection, column isolation, and many
                                 others. For an example of valve control, see the Advanced
                                 Operation manual.
                                Multiposition Also called a stream selection valve. It
                                 selects one from a number of gas streams and feeds it to
                                 a sampling valve. The actuator may be ratchet- (advances
                                 the valve one position each time it is activated) or
                                 motor- driven. An example that combines a stream
                                 selection valve with a gas sampling valve is provided in
                                 the Advanced Operation manual.
                                Remote start Available selection when configuring valve
                                 #7 or #8 only. Use this selection when wires controlling
                                 an external device are attached to an internal pair of
                                 contacts controlled by the GC.
                                Other    Something else.
                                Not installed   Self- explanatory.
Instrument
                   1 Press [Config]. Scroll to Instrument and press [Enter].
                   2 Scroll to Serial #. Enter a serial number and press [Enter].
                      This function can only be done by Agilent service
                      personnel.
                   3 Scroll to Auto prep run. Press [On/Yes] to enable Auto prep
                      run, [Off/No] to disable it. See the Advanced Operation
                      manual for details.
                   4 Scroll to Zero Init Data Files.
                       Press [On/Yes] to enable it. When it is On, the GC
                        immediately begins to subtract the current detector
                        output from all future values. This applies only to
                        digital output, and is useful when a non- Agilent data
                        system has problems with baseline data that is
                        non- zero.
                       Press [Off/No] to disable it. This is appropriate for all
                        Agilent data systems.
                   5 Scroll to Require Host Connection. Set On to consider
                      whether or not the remote host reports Ready as part of
                      GC readiness.
                   6 Press [Clear] to return to the Config menu or any other
                      function to end.
11
Options
About Options 186
Calibration 186
   Maintaining EPC calibrationinlets, detectors, PCM, and AUX 186
   Auto flow zero 187
   Auto zero septum purge 187
   Zero conditions 187
   Zero intervals 188
   To zero a specific flow or pressure sensor 188
   Column calibration 188
Communication 193
   Configuring the IP address for the GC 193
Keyboard and Display 194
About Options
                             The [Options] key is used for a group of functions that are
                             usually set on installation and seldom changed afterward. It
                             accesses this menu:
                             Calibration
                             Communication
                             Keyboard and Display
Calibration
                             Press [Calibration] to list the parameters that can be
                             calibrated. These include:
                              Inlets
                              Detectors
                              ALS
                              Columns
                              Oven
                              Atmospheric pressure
                             In general, you will only need to calibrate the EPC modules
                             and capillary columns. ALS, oven, and atmospheric pressure
                             calibration should only be performed be trained service
                             personnel.
                             Flow sensors
                             The split/splitless and purged packed inlet modules use flow
                             sensors. If the Auto flow zero feature (see page 186) is on,
                             they are zeroed automatically after each run. This is the
                             recommended way. They can also be zeroed manuallysee
                             To zero a specific flow or pressure sensor.
                   Pressure sensors
                   All EPC control modules use pressure sensors. They must be
                   zeroed individually. There is no automatic zero for pressure
                   sensors.
                   Zero conditions
                   Flow sensors are zeroed with the carrier gas connected and
                   flowing.
                   Pressure sensors are zeroed with the supply gas line
                   disconnected from the gas control module.
Zero intervals
             Column calibration
                             As you use a capillary column, you may occasionally trim off
                             portions, changing the column length. If measuring the
                             actual length is impractical, and if you are using EPC with a
                             defined column, you can use an internal calibration routine
                             to estimate the actual column length. Similarly, if you do not
                             know the column internal diameter or believe it is
                             inaccurate, you can estimate the diameter from related
                             measurements.
                             Calibration modes
                             There are three ways to calibrate the column length and/or
                             diameter:
                              Calibrate using an actual measured column flow rate
                              Calibrate using an unretained peak time (elution time)
                              Calibrate both length and diameter using flow rate and
                               elution time
                   CAUTION   When you measure the column flow rate, be sure to convert the
                             measurement to normal temperature and pressure if your
                             measurement device does not report data at NTP. If you enter
                             uncorrected data, the calibration will be wrong.
             CAUTION   When you measure the column flow rate, be sure to convert the
                       measurement to normal temperature and pressure if your
                       measurement device does not report data at NTP. If you enter
                       uncorrected data, the calibration will be wrong.
                                  Unretained peak
                                  Calculated length or Calculated diameter
                                  Not calibrated
                             9   Scroll to Measured flow and enter the corrected column
                                 flow rate (in mL/min) from the run performed above.
                             10 When you press [Enter], the GC will estimate the column
                                 length or diameter based on the elution time input and
                                 will now use that data for all calculations.
                   CAUTION   When you measure the column flow rate, be sure to convert the
                             measurement to normal temperature and pressure if your
                             measurement device does not report data at NTP. If you enter
                             uncorrected data, the calibration will be wrong.
Communication