0 ratings0% found this document useful (0 votes) 480 views62 pagesHP435B
HP435B Operating and Service Manual
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content,
claim it here.
Available Formats
Download as PDF or read online on Scribd
ne
Operating and Service Manual
HP 435B Power Meter :'
: ““Gnctuding Options 001,002, 003, and 004)
‘SERIAL NUMBERS
Attached to the rear panel of the instrument is a serial number-plate.
‘The serial number is in the form: 0000A00000. The first four
and the letter are the serial number prefix. The last five digits are the
suffix. The prefix is the samie for identical instruments; it changes
only when a configuration change is made to the instrumentt, The
suffix, however, is assigned sequentially and ié different for each
instrument.
‘This manual applies directly to instruments with serial numbers
prefixed 23424 and 2342U.
With changes described in Chapter 7, this manual also applies to
instruments with serial numbers prefixed 2005A, 2041U, 2238A,
24410 and above.
For additional important information about serial numbers, see
“Instruments Covered By Manual” in Chapter 1.
py HEWLETT
PACKARD
HP Part No, 00435-90055
@ HEWLETT-PACKARD COMPANY 1980, 1983, 1984, 1991
1501 Page Mill Road, Palo Alto, CaltorniaNOTICE
‘The information contained in this document is subject to change without notice.
HEWLETT-PACKARD MAKES NO WARRANTY OF ANY KIND WITH REGARD TO
‘THIS MANUAL, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. Hewlett-Packard
shall not be liable for errors contained herein or direct, indirect, special, incidental or
consequencial damages in connection with the furnishing, performance, or use of this material.
WARRANTY
A copy of the specific warranty terms applicable to your Hewlett-Packard product and
replacement parts can be obtained from your local Sales and Service Office.
Herstellerbescheinigung,
Hiermit wird bescheinigt, da dieses Gerat/System in Ubereinstimmung mit den Bestimmungen
von Postverfiigung 1046/84 funkentstart ist.
Der Deutschen Bundespost wurde das Inverkehrbringen dieses Gerites/System angezeigt und die
Berechtigung zur Uberpriifung der Serie auf Einhaltung der Bestimmungen eingeriumt.
‘Zusatzinformation fir Me8- und Testger
Werden Mef- und Testgerate mit ungeschirmten Kabeln und/oder in offenen Mefanfbauten
verwendet so ist vom Betreiber sicherzustellen, dafdie Funkentstérbedingungen unter
Betriebsbedingungen an seiner Grundstiicksgrenze eingehalten werden.
Manufacturer's Declaration
This is to certify that this equipment is in accordance with the Radio Interference Requirements
of Directive FTZ 1046/1984. The German Bundespost was notified that this equipment was put
into circulation, and has been granted the right to check the equipment type for compliance with
these requirements.
Note: If test and measurement equipment is operated with unshielded cables and/or used for
measurements in open setups, the user must ensure that under these operating conditions, the
radio frequency interference limits are me at the border of his premises.Contents
1. General Information
Introduction a: ee od
Tnstruments Covered by Manual |)... a
Manual Changes Supplement ......... 1
Replaceable Parts 2... 2... eee 1
Description 13
16
16
Tape Ontpat|Optionsy es ne ern 16
Accessories Supplied... 2... 0.0. 16
Equipment Required But Not Supplied ..... 16
Equipment Available... 2... ..-.- 16
Recommended Test Equipment . . . eed
Safety Considerations... .... boooue 17
2. Installation
Introduction... ee 2
Initial Inspection... 22... ee ee 21
Preparation for Use... 2. 21
Storage and Shipment... ......-... 28
3. Operation
Introduction ©. 2 ee 31
Panel Features 2... eee eH
Operator’s Checks 2. 2-22 eee 31
Operating Instructions ©... ........ 31
Power Measurement Accuracy ......... 32
Operator’s Maintenance... - 2... 35
Operator’s Checks . . eo vee 8B
Operating Instructions 2.0... ee. 30
Specified Uncertainty Calculation. |... 312
Calculating Measurement Uncertainty... . #13
Calculating Measurement Uncertainty (Uncertainty in
aBKnown). 2.2... ee ee 316
Contents-14.
5
8.
Contents-2
Performance Tests
Introduction... . . .
Equipment Required...
Test Record... . . See ee eee ee
Performance Tests.
Power Reference Level Test ....-..~
Zero Carryover Test :
Instrumentation Accuracy Test with Calibrator | | |
Calibration Factor Test ©. 2... ee ee ee
Adjustments
Introduction... .
Safety Considerations .
Equipment Required
Factory Selected Components .... 2...
Adjustment Locations. 2... ......-.
Power Reference Oscillator Level Adjustment
‘Typical Calculations
Multivibrator Adjustment
Power Meter Adjustments with 509 Power Sensor
Power Meter Adjustments with Calibrator .
Replaceable Parts
Introduction.
Abbreviations. .
Replaceable Parts List 11... Boone
Factory Selected Parts (*).. 2. 2 ee
Ordering Instructions 2s
Parts Provisioning :
Manual Changes
Introduction. 2. ee
Manual Changes... 2... eee
Manual Change Instructions... . 5
Service
Introduction... 2...
Safety Considerations |) lll
Service Sheets 2.2... ee ee
Recommended Test Equipment... ......
Repair oe
General Service Information |)... 8
Service Sheet 1... .
Service Sheet2. 0.2... 2c ee eee
Service Sheet 3.2... eee
Service Sheet 4.0.2.2... 02. ae
Service Sheet 5
41
41
41
41
42
45
ar
49
Bl
Bl
52
52
52
53
BB
36
58
S11
61
G1
61
62
62
62
Ta
aw
Ta
gl
a1
82
&4
84
S11
817
827
835
8-41a4.
21
22,
23.
24.
25.
34.
32.
33.
41.
42.
43.
44,
BL.
52.
53.
B4,
55.
B1.
82.
83.
84.
85.
86.
87.
88.
89.
8:10.
S11.
8:12.
8:13.
814.
815.
* Changing Range Switch Scale
HP 435B Power Meter and Accessories Supplied
Line Voltage Selection. . . . .
Power Cables and Line (Mains) Plug Options
Battery Installation Bees
Power Meter with Battery Installed... . . .
Front Panel Controls, Connectors and Indicators
Rear Panel Controls, Connectors and Indicators
‘Worst Case Effects of Specified and Mismatch
Uncertainties (Values From Examples on Page
3-12 and 3-14)
Power Reference Level Test Setup... ....
Zero Carryover Test Setup. . .
Instrumentation Accuracy Test Setup with Calibrator
Calibration Factor Test Setup... .
Power Reference Oscillator Level Adjustment Setup
Multivibrator Adjustment Setup... 2...
220 Hz Zero Crossover...
Power Meter Adjustment Setup with 509 Power
Sensor
Power Meter Adjustment Setup with Calibrator
‘Ad Assembly Extended for Service... . .
Non-lnverting Amplifier (Gain
Non-inverting Amplifier (Gain =
Inverting Amplifier (Gain
Schematic Diagram Notes .... 2.2...
Troubleshooting Block Diagram ........
Multivibrator/Detector Waveforms
AS Mother Board Component Locations . .
P/O Assembly (AC Ampl/Sync Detector)
Component Locations... .... ..
P/O Ad Assembly (AC Ampl/Sync Detector)
Schematic Diagram 2... 0... oe
Al Switch Assembly Component Locations
P/O A4 Assembly (DC Ampl/Auto Zero)
Component Locations... 1.2... --
P/O Assembly (DC Ampl/Auto Zero) Schematic
Diagram oe ee eee
P/O Ad Assembly (Power Supply) Component
Locatios...- 2... - ee ee ee
P/0 Ad Assembly (Power Supply) Schematic
it Wepeceodobouans .
12
22
23
24
26
27
36
37
3-15
42
45
47
49
53
56
57
511
8:25
8:25
831
8:33,
8:33
8:39
8:398-16. A3 Power Reference Assembly Component Locations 8-45
8-17. A3 Power Reference Assembly Schematic Diagram 8-45
8-18. HP 435B Front and Rear Panel Views... . . « S47
8-19, HP 435B Top and Bottom Views. ....... 848
8-20. HP 435B Right and Left Views . 849
Tables
Jl. Specifications 2... 22. ee 14
1-2. Recommended Test Equipment . 18
41. Performance Test Record (1 of 2) 411
5-1. Factory Selected Components 52
7-1. Manual Changes by Serial Number... .. . T
8-1. Etched Circuit Soldering Equipment ../.. . 8-5
82. Assembly, Chassis and Adjustable Components
Locations. ©... ee ee wae 8-43
ntents-4HP 4358
Safety Considerations
Safety
Considerations
Warning @
This product and related documentation must be reviewed for
familiarization with safety markings and instructions before
operation.
‘This product is a Safety Class I instrument (provided with a
protective earth terminal).
Before Applying Power
Verify that the product is set to match the available line voltage and
the correct fuse is installed.
Safety Earth Ground
‘An uninterruptible safety earth ground must be provided from the
main power source to the product input wiring terminals, power cord,
or supplied power cord set.
Any interruption of the protective (grounding) conductor (inside or
outside the instrument) or disconnecting the protective earth terminal
will cause a potential shock hazard that could result in personal injury.
(Grounding one conductor of a two conductor outlet is not sufficient
protection.) In addition, verify that a common ground exists between
the unit under test and this instrument prior to energizing either unit.
Whenever it is likely that the protection has been impaired, the
instrument must be made inoperative and be secured against any
unintended operation.
If this instrument is to be energized via an autotransformer (for voltage
reduction) make sure the common terminal is connected to neutral
(that is, the grounded side of the mains supply).
Servicing instructions are for use by service-trained personnel only.
To avoid dangerous electric shock, do not perform any servicing
unless qualified to do so.
Adjustments described in the manual are pertormed with power
‘supplied to the instrument while protective covers are removed.
Energy available at many points may, if contacted, result in personal
injury.
Capacitors inside the instrument may still be charged even if the
instrument has been disconnected from its source of supply.
For continued protection against fire hazard, replace the fine fuse(s)
only with 250V fuse(s) of the same current rating and type (for
example, normal blow, time delay, etc.). Do not use repaired fuses or
short circuited fuseholders.Safety Considerations,
Waring gy
Caution e
HP 4358
Safety Symbols
AX, issiraction manual symbol: The product will be marked
‘with this symbol when it is necessary for the user to refer
+o the instruction manual (see Table of Contents for page
references).
Indicates hazardous voltages.
abe Indicates earth (ground) terminal
‘The WARNING sign denotes a hazard. It calls attention to a
procedure, practice, or the like, which, if not correctly performed or
adhered to, could result in personal injury. Do not proceed beyond a
WARNING sign until the indicated conditions are fully understood and
met.
‘The CAUTION sign denotes a hazard. It calls attention to an
operating procedure, practice, or the like, which, if not correctly
performed or adhered to, could result in damage to or destruction of
part or all of the product. Do not proceed beyond a CAUTION sign
until the indicated conditions are fully understood and met.General Information
Introduction
Instruments Covered by
Manual
Manual Changes
‘Supplement
Replaceable Parts
‘This manual provides information for the installation, operation,
testing, adjustment and maintenance of the HP 435B Power Meter.
Figure 1-1 shows the power meter with accessories supplied.
An operating manual is shipped with the instrument. This is simply a
copy of the first three sections of this manual. The operating manual
should be kept with the instrument for use by the operator. Additional
copies of the operating manual may be ordered separately through your
nearest Hewlett-Packard office. The part number is listed on the title
page of this manual.
Instrument specifications are listed in Table 1-1. These specifications
are the performance standards or limits against which the instrument
may be tested.
Options 001, 002, 003 and 004 of the power meter are documented in
this manual. The differences are noted in the appropriate location such
as Options in Chapter 1, the Replaceable Parts List, and the schematic
diagrams.
This instrument has a two-part serial number. The first four digits and
the letter comprise the serial number prefix. The last five digits form
the sequential suffix that is unique to each instrument. The contents
of this manual apply directly to instruments having the same serial
number prefix(es) listed under “SERIAL NUMBERS” on the title page.
‘An instrument manufactured after the printing of a manual may have
a serial number prefix not listed on the title page. Unlisted serial
number prefixes indicate that the manual for such an instrument has
deen amended with a distinctive yellow Manual Changes supplement
containing updated technical information.
In addition to updated information, Manual Changes supplements
may also provide corrections to errors in manuals. Manual Changes
supplements are keyed to a manual’s print date and part number, both
of which appear on the title page.
In the U.S.A., it is better to order directly from the HP Parts Center.
Contact your nearest HP sales office for more information. Also, your
nearest HP office can supply toll-free telephone numbers for ordering
parts and supplies.General information HP 4358
MODEL 4358 e
ALIGNMENT TOOL
POWER CABLE POWER SENSOR CABLE
Figure 1-1. HP 4358 Power Meter and Accessories SuppliedHP 4358.
Description
General Information
‘The power meter and a compatible power sensor are interconnected
with the power sensor cable to form a power measurement system. The
system power level range, frequency response, and load impedance are
dependent on the power sensor.
Accuracy of the power measurement system is ensured by the following
power meter characteristics:
a. An internal automatic zeroing circuit which removes error due to the
ambient temperature output of the power sensor’s power sensing
device.
, A calibration factor-adjustment which accounts for error due to the
frequency response of the power sensing device.
¢. An internal calibration reference which has an output of
1 mW £0.7% (500).
13General Information
HP 4358
Table 1-1. Specifications
FREQUENCY
Frequency Range: 100 ki to 26.5 GHz (depending on power sensor used).
POWER
Power Range: (Meter calibrated in watts and dBm.)
‘With HP 8481B or 84828 sensors:
‘With HP 8481H or 8482H sensors:
‘With HP 8481A, 8482A, 8483A.
or 8485A sensors:
44 aB with 9 full scale ranges of 5, 10, 15, 20, 25, 80, 35, 40 and 45
dBm (1 mW to 25W).
45 4B with 9 full scale ranges of
4Bm (80 AW to 3W).
50 4B with 10 full scale ranges of ~25,
15 and 20 dBm (3 pW to 100 mW).
0, 5, 10, 15, 20, 25, 30 and 85
20, —15,
10, -5, 0, 5, 10,
‘With HP 8484A sensor: 50 dB with 10 full scale ranges of ~65, ~60, 55, 50, ~45, —40,
. 85, —80, ~25 and —20 dBm (300 pW to 10 wW).
ACCURACY
Instrnmentation!: % of fll scale on all ranges.
Zero: Automatic, operated by front-panel switch.
Zero Set: £0.5% of full seale on most sensitive range, typical.
Zero Carryover: £0.59 of full seale when zeroed on the most sensitive range.
Noise: (Typical, at constant temperature, peak change over any
Jone-minute interval.)
HP 84844 20 pW
HP 8481A, 8482A, 8483A, 84854. 40 aw
HP 84818, 84821 4 Ww
HEP 84818, 84828 40 pW
Drift: (1 hour, typical, at constant temperature after 24-hour warm-up)
HP 8484A 40 pW
HIP 8481A, 8482A, 8483A, 84854 15 0W
[HP 84818, 84821 15 uW
HP 84818, 84828 15 ew
1 Includes sensor non-linearity. Add 42,-4% on top two ranges when using the HP S481A, S482A, 453A and S4S5A power sensors:
dd 24.0% on the top two ranges when using the HP 8481H and 8482H power sensors.
14HP 4358 General Information
Table 1-1. Specifications (continued)
Power Reference: ‘Internal 50 MHz oscillator with Type N Female connector on front
[panel (or rear panel, Option 008 only).
Power output: 1.00 mW
Factory set to 20.7% traceable to the National Bureau of
Standards.
Accuracy: £1.2% worst case (0.9% rss) for one year (0 to 55°C)
‘Response Time: (0 to 99% of reading, five time constants.)
Range 1 (most sensitive) -<10.0 seconds
Range 2 <8.8 seconds
Range 3 <1 seconds
Ranges 410 <500 milliseconds
‘Typical, measured at recorder output.
Cal Factor: 16-position switch normalizes meter reading to account for
calibration factor or effective efficiency.
Range 85% to 100% in 19 steps.
Cal Adjustment: Front panel adjustment provides capability to adjust gain of meter
to match power sensor in use.
Recorder Output: Proportional to indicated power with 1 volt corresponding to full
scale; 1 KS output impedance; BNC connector.
‘RF Blanking Output: Provides a contact closure to ground when autosero mode is
engaged.
‘Power Consumption: 100, 120, 220, or 240V +5%, ~10%
100 and 120 volts, 48 to 66 Hz and 360440 Hz
220 and 240 volts, 48 to 66 Bz
20 VA maximum
Weight: Net, 2.7 kg (5.9 Ibs)
155 mm high (63; inches)
130 mm wide (5 inches)
279 ma deep (11 inches)
15General Information
Options
Input-Output Options
Accessories Supplied
16
Equipment Required
But Not Supplied
Equipment Available
HP 4358.
Battery
The HP 435B Power Meter, Option 001 is supplied with a rechargeable
battery that provides up to 16 hours continuous operation from a full
charge.
If the power meter was purchased without the battery option, it may
be ordered in kit form under HP part number 00435-60012. The kit
includes the battery, the battery clamp, a 6-32 by } inch pan head
machine screw and installation instructions.
Option 002
A rear panel input connector is connected in parallel with the front
panel input connector.
Option 003
A rear panel input connector is connected in parallel with the front
panel input connector. A rear panel POWER REF OUTPUT connector
replaces the standard front panel connector.
Option 004
The 1.5 meter (5 ft.) power sensor cable is not shipped with the power
meter.
‘The accessories supplied with the power meter are shown in Figure 1-1.
a. The HP 11730A, 1.5 meter (5-foot) power sensor cable is used to
couple the power sensor to the power meter. The 1.5 meter cable is
omitted when Option 004 is ordered.
b. The line power cable may be supplied in several configurations.
Refer to the paragraph entitled “Power Cable” in Chapter 2.
‘To form 2 complete RF power measurement system a power sensor such
as the HP 8481A must be connected to the power meter via the power
sensor cable.
‘The HP 11683A Range Calibrator is recommended for performance
testing, adjusting and troubleshooting the power meter. The power
meter’s range-to-range accuracy and auto-zero operation can easily be
verified with the calibrator. It also has the capability of supplying @
full-scale test signal for each range.
An extender board (HP part number 5060-0630) may be used to place
the Ad assembly printed circuit board in a position that allows easy
access to test points and components.HP 4358
‘The following table lists the power sensor cable accessories and their
General Information
lengths that are available for use with the power meter. Order option
004 if the standard 1.5 meter cable is not desired with a cable accessory.
Power Sensor Cable Accessory Cable Length
HP 117308 3.1m (10 ft)
HP 117300 6.1m (20 ft)
HP 11730D 15.2m (50 ft)
‘HP 11730E 0.5m (100 ft)
BP 11730F 61.0m (200 ft)
Recommended Test The test equipment shown in Table 1-2 is recommended for use during
Equipment performance testing, adjustments and troubleshooting. To ensure
optimum performance of the power meter, the specifications of a
substitute instrument must equal or exceed the critical specifications
shown in the table.
Safety Considerations The power meter is a Safety Class I instrument (provided with
a protective earth terminal). This instrument has been designed
according to international safety standards and has been supplied in
safe condition.
+7Beneral information HP 4356
Table 1-2. Recommended Test Equipment
Instrument Type Critical Specifications Suggested Use!
Model
Digital Voltmeter | Function: DC, Resistance HP 34554 PA,T
Ranges:
Resistance: 2002
‘Vae: 100 mV, 1000 mV, 10V, 100V
10 MQ input impedance
54 digit resolution
Accuracy:
40.05% of reading
0.028% of range
Frequency Counter | Frequency Range: 200 Hz ~ 50 MHz EP 53144 A
pee OO ee
Accuracy: 0.01%
Oscilloscope Bandwidth: de to 50 MHz HP 17404 PAT,
‘Vertical sensitivity: 0.2 V/division
Horizontal sensitivity: 1 ms/division
Power Meter ‘Range: capability to measure 1 mW HP 4324, PA
‘Transfer Accuracy (input to output): 0.2%
Power Sensor | Range: capability to measure 1 mW HP 8481A/H or HP 8482A/H | P, A
Range Calibrator EP 116834.
/Thermistor Mount | SWR: 1.05 at 50 Mz HP 478A - H75 or 478A - 762] P, A
Accuracy: £0.5% at 50 MHz
18
1 P= Performance Tests;
2 Traceable so NIST.
= Adjustments; T = Troubleshooting,Installation
2
Introduction
Initial Inspection
Warning G
Preparation for Use
This section includes information on the initial inspection, preparation
for use, and storage and shipment instructions for the power meter.
To avoid hazardous electrical shock, do not perform electrical tests when
there are signs of shipping damage to any portion of the outer enclosure
(covers and panels).
Inspect the shipping container for damage. If the shipping container or
cushioning material is damaged, it should be kept until the contents of
‘the shipment have been checked for completeness and the instrument
has been checked mechanically and electrically. The contents of the
shipment should be as shown in Figure 1-1. Procedures for checking
electrical performance are given in Chapter 4. Ifthe contents are
incomplete, if there is mechanical damage or defect, or if the instrument
does not pass the electrical performance tests, notify the nearest
Hewlett-Packard office. If the shipping container is damaged, or the
cushioning material shows signs of stress, notify the carrier as well as
the Hewlett-Packard office. Keep the shipping materials for the carrier's
inspection.
Meter Zeroing
With the LINE switch set to OFF, the meter pointer should be
positioned directly over zero. If necessary, insert a screwdriver into
‘the mechanical meter zero control (beneath the meter) and align the
pointer with zero. Back the adjustment off slightly. The backlash in the
control ensures against a meter indication error caused by jarring the
instrument.
Range Switch Scale Selection
‘The RANGE switch has three scales on two removable rings which
correspond to the measurement capabilities of compatible power
sensors. The range scales are 3 W to 0.3 mW (+35 to -5 dBm),
100 mW to 3 :.W(+20 to -25 dBm) and 10 pW to 0.3 nW (—20 to
~65 dBm). Each scale listed indicates the maximum and minimum full
scale meter readings.
To select the correct RANGE switch knob assembly scale (see
Figure 2-1):
24Warning @
HP 4358
Figure 2-1. Changing Range Switch Scale
a. Unscrew the outer (black) knob by turning it counterclockwise.
‘Then, remove the outer knob.
b. Remove the two scale rings.
c. Determine which of the three scales is to be used.
4. Place the other scale ring on the knob assembly.
e. Place the selected ring on the knob assembly with the selected scale
out.
£, Line up the tabs of the scale rings with the slot in the knob
assembly.
g. Hold the scale rings in place with your fingers. Thread the outer
knob onto the knob assembly. Lightly tighten the knob.
For the following HP instruments (8484, 8481D, 8485D, and 8487D)
ting replacement, order HP part no. 0350-0148. For HP instrument
(8482B), order HP part no. 0350-0153.
Power Requirements
‘The power meter requires @ power source with an output of 100,
120, 220, or 240 Vac +5%, -10% single phase, 100 and 120 volts, 48
to 66 Hz and 360 to 440 Hz, 220 and 240 volts, 48 to 66 Hz. Power
consumption is 20 VA maximum.
If this instrument is to be energized via an external autotransformer,
make sure the autotransformer common terminal is connected to the
earth terminal of the power source.HP 4353.
Caution e
Warning gy
Installation
Line Voltage Selection
BEFORE SWITCHING ON THIS INSTRUMENT, make sure the
instrument is set to the voltage of the power source.
Figure 2-2 provides instructions for line voltage and fuse selection. The
line voltage selection card and proper fuse are factory installed for 120
‘Vac operation.
Fuses may be ordered as:
Fuse HP Part Number
0.1 A, 250 V slow blow (100/120 Vac) 2110-0234
0.062'A, 250 V slow blow (220/240 Vac) 22110-0311
Operating voltage is shown
in module window.
SELECTION OF OPERATING VOLTAGE
1. Open cover door, pull the FUSE PULL lever and rotate to
eft. Remove the fuse.
2. Remove the Line Voltage Selection Card. Postion the ard
fo the ine voltage appears at topeft commer, Puth the card
firaly ino the sot.
‘3. Rotate the FUSE PULL lever to its normal potion. Insert
‘fuse of he correct vali the holder Close the cover
oor.
Figure 2-2. Line Voltage Selection
To avoid the possibility of hazardous electrical shock, do not operate this
instrument at line voltages greater than 126.5 Vac with ine frequencies
greater than 66 Hz (leakage currents at these line settings may exceed
3.5 mA),
Power Cable
In accordance with international safety standards, this instrument
is equipped with a three-wire power cable. When connected to an
appropriate ac power receptacle, this cable grounds the instrument
cabinet. The type of power cable plug shipped with each instrument
depends on the country of destination. Refer to Figure 2-3 for the part
numbers of the power cable plugs available.
23Installation
24
Warning G
HP 4358,
BEFORE SWITCHING ON THIS INSTRUMENT, the protective earth
terminals of this instrument must be connected to the protective
‘conductor of the (Mains) power cord. The Mains plug shall only be
inserted in a socket outlet provided with a protective earth contact. The
protective action must not be negated by the use of an extension cord
(power cable) without a protective conductor (grounding).
‘Option 908
warTeD incon
Option 383
Option 312 option sie
Figure 2-3. Power Cables and Line (Mains) Plug Options
Interconnections
‘The power meter and a power sensor are integral parts of this
measurement system. Before measurements can be performed, the
power meter and sensor must be connected together with the power
sensor cable. (The cable is supplied with the power meter.)
‘The power sensor cable couples the dc supply and sampling gate drive
from the power meter to the power sensor and the 220 Hz ac output
signal from the power sensor to the power meter.HP 4356
Caution e
Installation
‘The maximum voltage which may be safely coupled to the power meter
input from the power sensor is 18 mVrms.
Operating Environment
‘The operating environment should be within the following limitations:
‘Temperature . 0 to 55°C
Humidity . ++ <95% relative at 40°C
Altitude. . 4570 meters (15 000 feet)
Bench Operation
The instrument cabinet has plastic feet and a foldaway tilt stand for
convenience in bench operation. (The plastic feet are shaped to ensure
self aligning of the instruments when stacked.) The tilt stand raises the
front of the instrument for easier viewing of the control panel.
Rack Mounting
Instruments that are narrower than full rackwidth may be
rack-mounted using Hewlett-Packard adapter frames or combining
cases.
‘Adaptor Frames. Hewlett-Packard accessory adaptor frames are an
economical means of rack mounting instruments that are narrower than
full rack-width. A set of spacer clamps, supplied with each adaptor
frame, permits instruments of different dimensions to be combined and
rack mounted as a unit. Accessory blank panels are available for filling
unused spaces.
Combining Cases. HP 1051A and 1052A Combining Cases are
metal enclosures that allow combinations of one-third and one-half
rack-width instruments to be assembled for use on a workbench or for
mounting in a rack of standard 19-inch spacing. Each case includes
a set of partitions for positioning and retaining instruments and a
rack mounting kit. No tools are required for installing the partitions.
For bench use the cases have the same convenient features as full
rackwidth instruments, (for example, fold-away tilt stands and specially
designed feet for easier instrument stacking). Accessories available for
the combining cases include blank filler panels and snap-on full width
control panel covers.
Battery Operation
To operate the power meter on battery power, the battery must be
installed and charged, the line power cable must be disconnected, and
the LINE switch must be ON.
251stallation
Warning y
HP 4358
Battery Installation,
This task should be performed only by service trained persons who are
aware of the potential shock hazard of working on an instrument with
protective covers removed.
To avoid hazardous electrical shock, the line (Mains) power cable should
be disconnected before attempting to install the battery.
Do not short the battery terminals. This may result in overheating which
‘can cause burns or increase risk of fire.
Do not incinerate or mutilate the battery. It might burst or release toxic
materials causing personal injury.
‘The battery is installed in the power meter as follows (see Figure 2-4):
Figure 2-4. Battery InstallationHP 4358
Installation
a. Remove the top cover.
b. Hold the battery above the power meter, parallel to printed circuit,
board A4. The battery terminal lugs must face the circuit board.
¢. Loosen the lugs. Move the battery down into place and guide the
Ings into the slots on the circuit board. The battery should now rest
on the aluminum deck.
a. Place the battery clamp over the battery and secure it. The two
prongs fit into slots on the rear panel and the & by } inch pan head
machine screw holds the forward end of the clamp in place.
e. Tighten the battery terminal lugs by hand.
Figure 2-5 shows the power meter with battery installed.
Figure 2-5. Power Meter with Battery Installed
Battery Charging. The battery is being charged if the battery has been
installed, the line power cable is connected to the available line power,
and the LINE switch is ON. In the fully charged condition, (24-hour
charge time), the battery will supply power for a minimum of 16 hours.
27nstallation
Storage and Shipment
28
HP 4358
Environment
‘The instrument should be stored in a clean, dry environment. The
following environmental limitations apply to both storage and shipment:
55 to +75°C
<95% relative at 40°C
<15 300 meters (50 000 feet)
‘Temperature
Humidity .
Altitude
Packaging
Tagging for Service. If the instrument is being returned to
Hewlett-Packard for service, please complete one of the blue repair tags
located at the end of this manual and attach it to the instrument.
Original Packaging. Containers and materials identical to those used in
factory packaging are available through Hewlett-Packard offices. If the
instrument is being returned to Hewlett-Packard for servicing, attach
a tag indicating the type of service required, return address, model
number and full serial number. Mark the container FRAGILE to ensure
careful handling. Refer to the instrument by model number and full
serial number in any correspondence.
Other Packaging. The following general instructions should be used for
re-packaging with commercially available materials:
a. Wrap the instrument in heavy paper or plastic. (If shipping to a
Hewlett-Packard office or service center, attach a tag indicating the
type of service required, return address, model number and full serial
number.)
b. Use a strong shipping container. A doublewall carton made of
2.4 MPa (350 pound) test material is adequate.
c. Use a layer of shock-absorbing material 70 to 100 mm (3 to 4 inches)
thick around all sides of the instrument to provide firm cushioning
and prevent movement inside the container. Protect the control
panel with cardboard.
4. Seal the shipping container securely.
e. Mark the shipping container FRAGILE to ensure careful handling.
{. In any correspondence, refer to the instrument by model number and
full serial number.Operation
Introduction
Panel Features
Operator's Checks
Note g
Operating Instructions
Warning G
This section provides complete operating instructions for the power
meter. The instructions cover:
panel features
1 operator’s checks
= operating instructions
m power measurement accuracy
= operator’s maintenance
Front and rear panel features of the power meter are described in
Figure 3-1 and Figure 3-2. These figures contain a detailed description
of the controls, indicators and connectors.
If the instrument does not operate properly and is being returned to
Hewlett-Packard for service, please complete one of the blue repair tags
located at the end of this manual and attach it to the instrument.
Upon receipt of the instrument, or to check the power meter for an
indication of normal operation, follow the operational procedures in the
“Operator’s Checks” section. These checks are designed to familiarize
the operator with the power meter and to provide an understanding of
the operating capabilities.
General operating instructions are listed in the “Operating
Instructions” section. The instructions will familiarize the operator
‘with the basic practices used when operating the power meter.
‘Any interruption of the protective (grounding) conductor (inside or
outside the instrument) or disconnecting the protective earth terminal
is likely to make this instrument dangerous. Intentional interruption is,
prohibited.eration
Power Measurement
Accuracy
HP 4358.
‘A power measurement is never free from error or uncertainty. Any
RF system has RF losses, mismatch losses, mismatch uncertainty,
instrumentation uncertainty and calibration uncertainty. Measurement
errors as high as 50% are not only possible, they are highly likely unless
the error sources are understood and, as much as possible, eliminated.
‘Sources of Error and Measurement Uncertainty
RF Losses. Some of the RF power that enters the power sensor is
not dissipated in the power sensing elements. This RF loss is caused
by dissipation in the walls of waveguide power sensors, in the center
conductor of coaxial power sensors, in the dielectric of capacitors,
connections within the sensor and radiation losses.
Mismatch. The result of mismatched impedances between the device
under test and the power sensor is that some of the power fed to the
sensor is reflected before it is dissipated in the load. Mismatches affect
the measurement in two ways. First, the initial reflection is a simple
loss and is called mismatch loss. Second, the power reflected from the
sensor mismatch travels back up the transmission line until it reaches,
the source. There, most of it is dissipated in the source impedance,
but some of it is re-refiected by the source mismatch. The re-reflected
power returns to the power sensor and adds to, or subtracts from, the
incident power. For all practical purposes, the effect the re-reflected
power has upon the power measurement is unpredictable. This effect is
called mismatch uncertainty.
Instrumentation Uncertainty. Instrumentation uncertainty describes the
ability of the metering circuits to accurately measure the dc output
from the power sensor's power sensing device. In the power meter, this
error is less than 1%}. It is important to realize, however, that a 1%
meter does not automatically give 1% overall measurement accuracy.
Power Reference Uncertainty. The uncertainty of the output level of the
power reference oscillator is +0.7%. This reference is normally used
to calibrate the system and is, therefore, a part of the system’s total
measurement uncertainty.
Cal Factor Switch Resolution Error. The resolution of the CAL
FACTOR switch contributes a significant error to the total
measurement because the switch has 1% steps. The maximum error
possible in each position is 0.5%
2 Refer to Instrument accuracy specification in Chapter 1 when using the top
‘two ranges.HP 4358,
Operation
Corrections for Error
Calibration Factor and Effective Efficiency. The two correction factors
basic to power meters are calibration factor and effective efficiency.
Effective efficiency is the correction factor for RF losses within the
power sensor. Calibration factor takes into account the effective
efficiency and mismatch losses.
Calibration factor is expressed as a percentage with 100% meaning
the power sensor has no losses. Normally the calibration factor will
be 100% at 50 MHz, the operating frequency of the internal reference
oscillator.
The power sensors used with the power meter have individually
calibrated calibration factor curves placed on their covers. To correct
for RF and mismatch losses, simply find the power sensor's calibration
factor at the measurement frequency from the curve or the table that is
supplied with the power sensor, and set the CAL FACTOR switch to
this value.
‘The CAL FACTOR switch resolution error of 0.5% may be reduced
by one of the following methods:
1. Set the CAL FACTOR switch to the nearest positions above and
below the correction factor given on the table. Interpolating between
‘the power levels measured provides the corrected power level.
2. Leave the CAL FACTOR switch on 100% after calibration. Then,
make the measurement and record the reading. Use the reflection
coefficient, magnitude and phase angle, if such a table is supplied
with the power sensor, to calculate the corrected power level.
Calculating Worst Case Uncertainty
‘Worst case uncertainty is the sum of the specified uncertainties and
mismatch uncertainty. Uncertainty calculation is outlined in the
following two subsections. Examples are worked out in the “Specified
Uncertainty Calculation” section and the “Calculating Measurement
Uncertainty” section. For a more complete explanation of measurement
uncertainty refer to HP application note AN-64-1 “Fundamentals of RF
and Microwave Power Measurement”
33'peration
HP 4358
Specified Uncertainties, The specified uncertainties which account for
part of the total power measurement uncertainty are:
a. Instrumentation +1%" or £0.05 dB.
b. Power reference +0.7% or +0.03 4B.
c. CAL FACTOR switch resolution, 0 to 40.5% (depending on Cal
Factor).
4, Zero set, 40.5% of full scale of lowest range which is 15 nW.
e. Zero Carryover, £0.5%.
£. Noise and Drift, depends on the range and type of sensor.
g. Calibration factor uncertainty, which depends on sensor type, is
Uisted in the sensor manual.
‘The “Specified Uncertainty Calculation” section gives an example of
specified uncertainty calculation.
Calculating Mismatch Uncertainty. Mismatch uncertainty is the result of
the source mismatch interacting with the power sensor mismatch. The
magnitude of uncertainty is related to the magnitudes of the source
and power sensor reflection coefficients, which can be calculated from
SWR. The “Calculating Measurement Uncertainty” section shows
how the calculations are made and Figure 3-3 illustrates mismatch
uncertainty and total calculated uncertainty for two cases. In the first,
case, the power sensor’s SWR = 1.5, and in the second case, the power
sensor’s SWR = 1.25. In both cases the source has an SWR of 2.0. The
example shows the effect on power measurement accuracy a poorly
matched power sensor will have as compared to one with low mismatch.
‘A faster, easier way to find mismatch uncertainty is to use the
HP Mismatch Error (uncertainty) Limits/Reflectometer Calculator.
The calculator may be obtained, on request, from your nearest
Hewlett-Packard office by using HP part number 5952-0948.
The method of calculating measurement uncertainty from the
uncertainty in 4B is shown in the “Calculating Measurement
Uncertainty (Uncertainty in dB Known)” section. This method would
be used when the initial uncertainty calculations were made with the
Mismatch Error/Refiectometer Calculator.
2 Refer to Instrument accuracy specification in Chapter 1 when using the top
two ranges.HP 4358,
Operator’s Maintenance
Caution 6
Warning @
Operation
‘The only maintenance responsibilities the operator should normally
perform are primary power fuse replacement, LINE switch lamp
replacement and rechargeable battery replacement.
Battery replacement is the only operation that requires tools. A.
Posidrive screwdriver is needed to remove the battery clamp.
‘The primary power fuse is found within the A6 Power Module
Assembly on the power meter’s rear panel. For instructions on how to
‘change the fuse, refer to the paragraph entitled Line Voltage Selection
in Chapter 2.
‘Make sure that only fuses with the required rated current and of the
specified type (slow blow, time delay, etc.) are used for replacement.
‘The use of repaired fuses and the short-circuiting of fuse-holders must
be avoided.
Lamp Replacement
The lamp is contained in a plastic lens which doubles for a pushbutton
on the LINE switch. When the power meter LINE switch is ON and
is being operated by the available line power, the lamp should be
illuminated. If the lamp is defective, remove the lens by pulling it
straight out. Order lamp (HP part number 3131-0434) CD6 and replace
the old pushbutton-lamp assembly with the new one, To replace the
assembly, align the pins with the notch in the receptacle and push
straight in.
Battery Replacement
Hf the meter indicates that the battery is discharged by a full downscale
reading, and after charging the battery still will only power the power
meter for a short period of time, the battery is probably defective.
‘The replacement battery, BT1 (HP part number 1420-0096), may be
ordered through the nearest Hewlett-Packard office. Refer to Battery
Installation in Chapter 2.
This task should be performed only by service trained persons who are
‘aware of the potential shock hazard of working on an instrument with
Protective covers removed.dperation
HP 4353
@ Meter. Normally indicates average RF power in
4Bm or Watts. During battery operation the meter
continuously indicates battery condition. A nor-
rmalreading indicates the battery is charged; a full
down-scale reading indicates the battery is dis
charged or is defective.
@ Moter Zero. Mechanical adjustment used to zerothe
meter when the LINE switch is OFF.
© LINE switch. Connects line or battery power to the
Power Meter circuits when the LINE switch is ON.
During battery operation, the lamp contained
within the LINE switch will not be illuminated
when the INSTRUMENT is ON.
AANGE Switch. Selects desired power range: keyed
to meter full-scale deflection; has three removable
scales which are changed to match the range of
the power sensor.
(© POWER REF OUTPUT. RF output of 1.00 mW 40.70%
into 500 at 50 MHz from an internal reference
oscillator. Available for system calibration.
CAL ADS. Screwdriver adjustment for calibrating
any power sensor and Power Meter as a system, t0
a known standard.
@ input Connector. tnput from the power sensor via
the power sensor cable.
© CALFACTOR Switch. changes the gain of the Power
Meter amplifier circuits to compensate for mis-
match losses and effective efficiency of the power
© zn0 switch. The ZERO switch activates a feed
back circuit, which automatically zeros the meter
pointer, and a rear panel RF blanking signal.
Figure 3-1. Front Panel Controls, Connectors and IndicatorsHP 4358.
Operation
(@ POWER SENSOR INPUT. Option 002 and 003 have a
rear panel input connector wired in parallel with
the front panel input connector.
© Power Module Assembly.
© Window. safety interlock; fuse cannot be removed
while power cable is connected to Power Meter.
Q FUSE PULL Handle. Mechanical interlock to guar-
antee fuse has been removed before Line Voltage
Selection Card can be removed.
@ Fuse. Refer to Section II for values.
@ Line Voltage Selection Card. Matches transformer
primary to available line voltage.
@ Receptacle. For power cable connection to avail-
able line voltage.
© POWER REF OUTPUT. Takes the place of the front
panel POWER REF OUTPUT connector (Option
008 only)
© RECORDER OUTPUT. Provides a linear output with
respect tothe input power. +1.00 Vde corresponds
tometer full-scale. The minimum load which may
be coupled to the output is 1 Ma.
@ FF BLANKING QUTPUT. Contact closure to ground
when ZERO switch is pressed. May be used to
remove RF input signal during automatic zeroing
operation,
@ POWER REF Switch. Opens or closes the circuit
from the power supply tothe power reference oscil-
lator. Reduces current drain during battery opera-
tion when OFF.
Figure 3-2, Rear Panel Controls, Connectors and Indicators
a7Operation HP 4358
Operator’s Checks 1, BEFORE SWITCHING ON THIS INSTRUMENT, check that the
power transformer primary is matched to the available line voltage,
the correct fuse is installed and the safety precautions are taken.
See “Power Requirements”, “Line Voltage Selection”, “Power
Cable” and associated warnings and cautions in Chapter 2.
Warning G BEFORE CONNECTING LINE POWER TO THIS INSTRUMENT, ensure
that all devices connected to this instrument are connected to the
protective (earth) ground.
BEFORE SWITCHING ON THIS INSTRUMENT, ensure that the line
power (Mains) plug is connected to a three-conductor line power outlet
that has a protective (earth) ground. (Grounding one conductor of a
two-conductor outlet is not sufficient.)
Caution Do not twist the body of the power sensor when connecting or
disconnecting it. This can cause major damage to the power sensor.
2. Set the meter indication to zero with the mechanical meter zero
control. Back the control off slightly.
3. Connect the power sensor to the power meter with the power sensor
cable.
4. Connect the power cable to the power outlet and power module
receptacles. Set the LINE switch to ON; the lamp within the switch
lens should be illuminated.
5. Change the power meter’s RANGE switch scale so it corresponds
to the range of the power sensor. Refer to “Range Switch Scale
Selection” in Chapter 2.
6. Set the power meter Controls as follows:
RANGE switch position . fully cew
CAL FACTOR switch .. 100%
POWER REF switch ... OFF
7, Press the ZERO switch and verify that the meter pointer moves to
“0” (zero) and the RF BLANKING OUTPUT is shorted to ground.
8. Set the RANGE switch to the position indicated in the following
table. Then, connect the power sensor (and adapter or attenuator
as required) to the POWER REF OUTPUT and set the rear
panel POWER REF switch to (ON). Verify that the meter reads
‘approximately the same as indicated in the table.
38HP 4358.
Operation
RANGE | Meter
Power Sensor Switch | Indication
Position
HP 8481B and 8482B (remove attenuator) 3W lw
HIP 8481A, 8482A, 84811, 84828 3mw | imw
HP 84854 (HP 1250-1250 Adapter required) 3mw_ | 1mw
HP 8483A (HP 1250-0597 Mechanical Adapter 3mw | 0.96 mw
required)
HP 8484A (HP 11708A Reference Attenuator 3. 1W
required)
9. Step the CAL FACTOR switch through its range noting a small
increase in meter reading with each successive step. Reset the CAL
FACTOR switch to 100%.
10. Set the RANGE switch to the position indicated in the table below.
‘Then, adjust the CAL ADJ control for a full-scale meter reading for
50Q power sensors and a 96% of full scale meter reading for 750.
power sensors.
Power Sensor RANGE Switch
Position
[HP 84818 and 84828 (remove attenuator) iw
HIP 8481A, 8482A, 8481H, 8482 imw.
HP 84854 (HP 1250-1250 Adapter required) 1mw
HP 8483A (HP 1250-0597 Mechanical Adapter required) 1mw
HP 8484A (HP 11708A Reference Attenuator required) LEW
11. Check at the rear panel RECORDER OUTPUT jack for an output
of #1 Vde.
12. To check operation using battery power, disconnect the power
cable from the rear power panel module receptacle and set the
LINE switch to ON (the lamp within the switch will lens not
be illuminated). When a power measurement is made, a normal
‘upscale reading indicates normal operation; a full dowan-scale
reading indicates the battery is discharged.
39'peration HP 4358,
)perating Instructions
1. BEFORE SWITCHING ON THIS INSTRUMENT, check that the
power transformer primary is matched to the available line voltage,
the correct fuse is installed and safety precautions are taken. See
“Power Requirement”, “Line Voltage Selection”, “Power Cable” and
associated warnings and cautions in Chapter 2.
Warning e BEFORE CONNECTING LINE POWER TO THE INSTRUMENT, ensure that
all devices connected to this instrument are connected to the protective
(earth) ground.
BEFORE SWITCHING ON THIS INSTRUMENT, ensure that the line
power (Mains) plug is connected to a three-conductor line power outlet
that has a protective (earth) ground. (Grounding one conductor of a
two-conductor outlet is not sufficient.)
Caution Do not twist the body of the power sensor when connecting or
disconnecting it. This can cause major damage to the sensor.
2, Set the meter indication to zero with the mechanical meter zero
control. Back the control off slightly.
3. Connect the power sensor to the power meter with the power sensor
cable.
4, Connect the power cable to the power outlet and power module
receptacles. Set the LINE switch to ON; the lamp within the switch
lens should be lit.
5. Change the power meter’s RANGE switch scale so it corresponds
to the range of the power sensor. Refer to the paragraph entitled
Range Switch Scale Selection in Chapter 2.
6. Set the power meter switches as follows:
RANGE position
CAL FACTOR ......... 100%
POWER REF ......... OFF
7. Press the ZERO switch, allow 5 seconds for the zeroing operation to
take place, and release the switch.
8, Set the RANGE switch to the position indicated in the following
table. Then, connect the power sensor (and adapter or attenuator
as required) to the POWER REF OUTPUT and set the rear panel
POWER REF switch to (ON). For 509 power sensors, adjust the
CAL ADJ control for a full-scale reading; the meter pointer should
be aligned with the CAL mark (full-scale reading) on the meter
face. For 750 power sensors, adjust the CAL ADJ control for a
96% of full scale reading; the meter pointer should be aligned with
the 0.96 mark on the meter face.
fully cow
0HP 4358
Caution e
A
Note @
Operation
Power Sensor
/HP 84818 and 8482B (remove attenuator) iW
HP 8481A, 84824, $4811, 8482 1mw.
HP 84854 (HP 1250-1250 Adapter required) 1mw
EP 84834 (HP 1250-0597 Mechanical Adapter required) imw
HIP 84844 (HP 11708A Reference Attenuator required) Law
9. Disconnect the power sensor from the POWER REF OUTPUT and
set the POWER REF switch to OFF.
10. Locate the calibration curve on the power sensor cover. find the
CAL FACTOR for the measurement frequency; set the CAL
FACTOR for the measurement frequency; set the CAL FACTOR
switch accordingly.
LL. Set the RANGE switch so that full scale is greater than the power
level to be measured.
12. Connect the power sensor to the RF source. Read the power level
in dBm or watts on the panel meter.
See “Operating Precautions” in the power sensor Operating and Service
manuals for maximum power levels which may be safely coupled to this
system. Levels which exceed the limits may damage the power sensor,
power meter, or both.
‘When the battery is being used as the power supply for the power
meter, an automatic test circuit continually monitors battery condition.
‘When the battery voltage is above a predetermined level, the meter
indicates the correct power level. When the voltage drops below the
threshold level, the meter reading is full downscale.
a1Iperation HP 4958
specified Uncertainty Calculation
Conditions:
Range —1 mW
Meter Reading — 0.7 mW
Sensor — HP 84814
Frequency — 1 GHz
CAL FACTOR — 99.5%
(FS)? Instrumentation Uncertainty =4001mW = = 40.06 dB
(R)? Power Reference Uncertainty = 40.0049 mW = 40.03 dB
(R) CAL FACTOR Switch Resolution Uncertainty = 40.0035 mW = 40.02. 4B
(R) Zero Set Uncertainty = £0.000015 mW = +0.00009 4B
(FS) Zero Carryover Uncertainty = 405% = 40.005mW — = 40.03 dB
(R) Noise 008% = 40.00025 aB
(R) Drift 002% = £0.00009 4B
(R) Cal Factor Uncertainty = 42.70% =40019mW = 40.1248
40.0425 mW
LPS = % of full scale
2R
Total Specified Uncertainties = 40.0425 mw
= 9988 (100)
= 46.07%
= Wlog 9245
B12
£0.26 dBHP 4358. Operation
Calculating Measurement Uncertainty
1. Calculate the reflection coefficient from the given SWR.
SWR-1
SWR+1
ower Sensor #1 Power Sensor = Power Souee
owns own 133 own 20
n= iit a= Hit n= it
=u ote
= 02 = 0.111 = 0.333
2. Calculate the relative power and percentage power mismatch
uncertainties from the reflection coefficients. An initial reference level of
1is assumed.
Relative Power Uncertainty
PU=[1+ (Pops)}?
PUL = {1 [( 0.2)(0.833)]}? PU2 = {1+ {(0.111)(0.383)]}?
{1 40.087}? = {140087}
{1.067}? and {0.983}?
= 1.138 and 0.871
{1.087}? and {0.963}?
1.075 and 0.927
Percentage Power Uncertainty
SPU = (PU — 1) 100%
(2.138 — 1) 100% and (0.871 ~ 1) 100%
= (0.138) 100% and (0.128) 100%
13.8% and 12.9%
PU2 = (1.075 ~ 1) 100% and (0.927 ~ 1) 100%
(0.075) 100% and (0.078) 100%
= 75% and ~13%
3-13eration
bag
3. Calculate the Measurement Uncertainty in 4B.
MU;
MU2
Measurement Uncertainty in dB
MU =10 lose (%)] 42
[10g (24%)]
10 [0.056]
4056 a3
[totes (+4)]
10 [0.081]
40.81 dB
and
and
and
and
and
and
HP 4358.
10 fon (2#2)]
10 [—-0.060]
-08 4B
10 flog (*$%)]
10 [0.033]
0.33 4BHP 4358, Operation
INDICATED POWER VERSUS RANGE OF ACTUAL POWER
INDICATED POWER TOTAL SPECIFIED WORST CASE UNCERTAINTY
UNCERTAINTIES (SPECIFIED & MISMATCH)
EXAMPLE 1:
‘SENSOR SWR=1.5
‘SOURCE SWR = 2.0
07 mW (1.55 dBm)
0. Ly 8m)
EXAMPLE 2:
‘SENSOR SWR = 1.25
SOURCE SWR=20
— sa (1.29 dBm)
0.7 mil (1.55 48)
1 18 al (2.1 8m)
Figure 3-3.
Worst Case Effects of Specified and Mismatch Uncertainties
(Values From Examples on Page 3-12 and 3-14)
315peration HP 4356
2alculating Measurement Uncertainty (Uncertainty in dB Known)
1. For this example the known values are: source SWR, 2.2 and
power sensor SWR, 1.16. From the Mismatch Error Calculator the
mismatch uncertainty is found to be +0.24, -0.25 dB.
2. Add the results from the “Specified Uncertainty Calculation” section,
(40.26 4B). The total measurement uncertainty is +0.50, -0.51 dB.
3. Calculate the relative measurement uncertainty from the following
formula:
aB
slog ()
tog()
B = le (9)
to? (88)
= 880
= 112
4, Calculate the percentage Measurement Uncertainty.
Percentage Measurement Uncertainty
MU = (P1 — Po) 100
%MU = (122-1) 100 (0.889 — 1) 100
= 412.2% = 111%
316Performance Tests
4
Introduction
Equipment Required
Test Record
Performance Tests
Note 9
‘The procedures in this section test the electrical performance of the
power meter using the specifications of Table 1-1 as performance
standards. All tests can be performed without access to the interior of
the instrument. A simpler operational test is included in Chapter 3
‘under “Operator’s Checks.”
Equipment required for the performance tests is listed in Table 1-2,
Recommended Test Equipment. Any equipment that satisfies the
critical specifications given in the table may be substituted for the
recommended model(s)
Results of the performance tests may be recorded on the test record
at the end of the test procedures. The test record lists all of the
tested specifications and their acceptable limits. Test results recorded
at incoming inspection can be used for comparison in periodic
maintenance, troubleshooting and after repairs or adjustments.
‘The performance tests given in this section are suitable for incoming
inspection, troubleshooting or preventive maintenance. During any
performance test, all shields and connecting hardware must be in place.
Perform the tests in the order given and record the data on the test,
card and/or in the data spaces provided at the end of each procedure.
The power meter must have a half-hour warmup and the line voltage
must be within +5%, —10% of nominal if the performance tests are to
be considered valid.
Each test is arranged so that the specification is written as it appears in
Table 1-1. Next, a description of the test and any special instructions
or problem areas are included. Bach test that requires test equipment
has a setup drawing and a list of the required equipment. ‘The initial
steps of each procedure give control settings required for that particular
test,
atPerformance Tests
HP 4358
?ower Reference Level Test
42
Specification
Description
Equipment
Internal 50 MHz oscillator with ‘Type-N Female connector on front
panel (or rear panel, Option 003 only). Power output: 1.00 mW.
Factory set to +0.7% traceable to the National Bureau of Standards.
Accuracy: 1.2% worst case (0.9% rss) for one year (0 to 55°C).
‘The power reference oscillator output is factory adjusted to
1 mW 40.7%. To achieve this accuracy, Hewlett-Packard employs
a special measurement system accurate to 0.5% (traceable to the
National Bureau of Standards) and allows for a transfer error of 0.2%
in making the adjustment. If an equivalent measurement system is
employed for verification, the power reference oscillator output can
be verified to 1 mW +1.9% (1.2% accuracy +0.5% verification
system error :40.0% transfer error = 1.9% maximum error). The power
reference oscillator can be set to 0.7% using the same equipment
and following the adjustment procedure in Chapter 5. To ensure
maximum accuracy in verifying the power reference oscillator output,
the following procedure provides step-by-step instructions for using
specified Hewlett-Packard test instruments of known capability. If
equivalent test instruments are used, signal acquisition criteria may
vary and reference should be made to the manufacturer's guidelines for
operating the instruments.
‘The power meter may be returned to the nearest Hewlett-Packard office
to have the power reference oscillator checked and/or adjusted. Refer to
Chapter 2, “Packaging.
Figure 4-1. Power Reference Level Test Setup
‘HP 432A
HP 478A-H75/H76
‘HP 34554,
Power meter ..
‘Thermistor mount .
Digital voltmeterHP 4358
Procedure
Performance Tests
1. Set up the DVM to measure resistance. Connect the DVM between
‘the Var connector on the rear panel of the power meter and pin 1
of the thermistor mount end of the power meter interconnect cable.
2. Round off the DVM indication to two decimal places and record
this value as the internal bridge resistance (R) of the power meter
(approximately 200 ohms).
3. Connect the HP 432A Power Meter to the HP 435B Power Meter as
shown in Figure 4-1.
4. Set the HP 432A Power Meter LINE switch to ON (in) and the
POWER REF switch to OFF. Then, wait thirty minutes for the HP
432A Power Meter thermistor mount to stabilize before proceeding
to the next step.
5. Set the HP 432A Power Meter RANGE switch to COARSE ZERO
and adjust the front-panel COARSE ZERO control to obtain a zero
meter indication.
6. Fine zero the HP 432A Power Meter on the most sensitive range,
then set the HP 432A Power Meter RANGE switch to 1 mW.
Note Check that DVM input leads are isolated from chassis ground when
performing the next step.
7, Set up the DVM to measure microvolts and connect the positive
and negative input leads, respectively, to the Vcomp and Var
connectors on the rear panel of the HP 432A Power Meter.
8. Observe the indication on the DVM. If less than 400 microvolts,
proceed to the next step. If 400 microvolts or greater, press and
hold the power meter FINE ZERO switch and adjust the COARSE
ZERO control so that the DVM indicates 200 microvolts or less.
Then, release the FINE ZERO switch and proceed to the next step.
9. Round off the DVM indication to the nearest microvolt and record
this value as Vo.
10. Set the HP 435B Power Meter POWER REF switch to ON (in) and
record the indications observed on the DVM as V3.
11. Disconnect the DVM negative input lead from the Var connector
on the HP 432A Power Meter and reconnect it to the HP 432A
Power Meter chassis ground. Record the new indication observed
on the DVM as Veomp-
12. Calculate the power reference oscillator output level (PpF) from the
following formula:
2Voomp (Vi ~ Vo) + VE
2
4R (CALIBRATION FACTOR)
‘Where:
Pre ower reference oscillator output level
wreviously recorded value
43erformance Tests HP 4358
V-1 = previously recorded value
V.0 = previously recorded value
R = previously recorded value
CALIBRATION FACTOR = value for thermistor mount at
50 MHz (traceable to the National
Bureau of Standards)
13. Verify that the Pap is within the following limits:
Min. Actual Max.
0.988 mW 1.012 mWHP 4358.
Performance Tests
Le nn_n_ ee EEEEIEEI SEI IEINIIININISNIIEEEEESSEEISEEEESEEEEEEEEEEEEET
Zero Carryover Test
Specification
Description
Equipment
Procedure
+£0.5% of full scale when zeroed in the most sensitive range.
After the power meter is initially zeroed, the change in the meter
reading is monitored at the RECORDER OUTPUT as the instrument,
is stepped through its ranges. The meter readings take into account
noise and drift because zero carryover and the noise drift readings
cannot be separated. Refer to Table 5-1 if the results are not within the
limits.
HP 4358,
POWER
METER
pecoroer ourrur. _ DIGITAL VOLTMETER
Figure 4-2. Zero Carryover Test Setup
Digital voltmeter ....... HP 34554
Power sensor ... HP 8481A/H or 8482A/H
1. Set the DVM RANGE control to 100 mVde.
2. Set the power meter switches as follows:
CAL FACTOR, = 100%
RANGE position . «= fully ecw
POWER REF (rear panel) .. OFF
3. Connect the equipment shown in Figure 4-2.
4, Press the front panel ZERO switch and wait for the meter indicator’s
position to stabilize. Verify that the DVM reads 0 +1.0 mVdc.
Release the ZERO switch.
5. Verify that the RECORDER OUTPUT falls within the limits shown
on the table for each range. Record the readings.
45erformance Tests HP 4358
RANGE ‘Results RANGE Results
Switch Min. ‘Actual ‘Max. Switch Min. ‘Actual Max.
Position Position
mde mde mde mide mae mde
fully cow 15 +15 | S steps ow 5 +5
1 step cw “7 — #17 | 6 steps ow 5 = +
2 steps ow “4 — 414 | Tsteps ew 6 — +
3 steps ow =u — 411 | 8 steps ow 6 — +5
4 steps ew -8 — 8 fully ew 5 — +HP 4358
Performance Tests
i
Instrumentation Accuracy Test with Calibrator
Specification
Description
Equipment
Procedure
£1% of fall scale on all ranges.
Instrumentation accuracy is verified by coupling a full-scale reference
input from the HP 11683A Calibrator to the power meter on each
range. Verify that the RECORDER OUTPUT level is within +1% plus
noise and drift.
HP 4358
BOWER
METER
RANGE
CALTBRATOR RECORDER ouTpuT _DIGITAL_VOLTMETER
(REAR PANEL)
Figure 4-3. instrumentation Accuracy Test Setup with Calibrator
HP 3455A
HP 11683A,
Digital voltmeter .
Range calibrator
1. Set the range calibrator RANGE switch to ImW, the FUNCTION
switch to CALIBRATE and the POLARITY switch to NORMAL.
2. Set the power meter RANGE switch 5 steps from the fully cow
position.
3. Set the DVM RANGE switch to 1000 mvdc.
4, Connect the equipment as shown in Figure 4-3.
5. Adjust the front panel CAL ADJ control to provide a reading of
1000 +2 mVde.
‘To avoid damage to the meter, set the range calibrator’s FUNCTION
control to STANDBY while changing the RANGE control settings on
the power meter and range calibrator.
6. Set the power meter RANGE switch to each possible position in
‘turn. Set the range calibrator RANGE switch to the same position
and verify that the DVM reading, which includes noise and drift, is
within the limits shown in the following table.
47>erformance Tests HP 4358
RANGE Results RANGE Results
Switch [Min] Actual | Max | Switch [Min | actu | Max
Position Position
mde | mvae | mde mvae | mVde | mae
fully cow +975 — +1025, 5 steps cw +998 +1002
tstepen | 4078 | — | +102 | stepecw | 4900 | — | 010
2atepecw | 4981 | | 41019 | 7etepscw | 4000 | — | +2010
Satepecw | 49e4 | | 41016 | Setepsew | 4000 | — | 42010
Astepson | 4987 | | 41013, | talyew | 490 | | 42010
48HP 435B
Performance Tests
er
Calibration Factor
Test
Specification
Description
Equipment
Procedure
A 16-position switch normalizes meter reading to account for calibration
factor or effective efficiency. Range 85% to 100% in 1% steps.
After the power meter is zeroed on the most sensitive range, 2 1 mW
input level is applied to the power meter and the CAL ADJ control is
set to obtain a 1.000 mW indication. Then the CAL FACTOR switch is
stepped through its 16 positions and the meter is monitored to ensure
that the proper indication is obtained for each position.
HP 4358
POWER
RANGE METER
CALIBRATOR RecoRveR ourpur _DIGITAL VOLTMETER
Figure 4-4, Calibration Factor Test Setup
Digital voltmeter
Range calibrator
HP 34554
‘HP 116834
1. Set the range calibrator RANGE switch to 1 mW, the FUNCTION
switch to CALIBRATE and the POLARITY switch to NORMAL.
2. Set the power meter RANGE switch 5 steps from the fully ccw
position,
3. Set the DVM RANGE switch to Vdc.
4, Connect the equipment as shown in Figure 4-4.
5. Set the front panel CAL ADJ control to provide reading of
1090 +42 mVde on the DVM.
6. Set the CAL FACTOR switch to each position and verify that the
indications observed at each position are within the limits specified
in the following table.