Technical Specifications:
Principles Species
Flow Cytometry (FCM), Semi-conductor Laser scatter, chemical dye Dog, Cat, Horse, Rat, Mouse, Rabbit, Monkey, Pig, Goat, Camel
method and independent Basophil channel Aniaml1 to Animal20.
Advanced DC impedance method for RBC and PLT counting
Cyanide free reagent for hemoglobin test Data Storage Capacity
Up to 40,000 results including numeric and
Parameters graphic information
23: WBC, Neu%, Lym%, Mon%, Eos%, Bas%, Neu#, Lym#, Mon#,
Eos#, Bas#, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, Communication
RDW-SD, PLT, MPV, PDW, PCT LAN port supports HL7 protocol
Performance Operating Environment
Parameter Linearity Range Precision ( CV % ) Temperature: 15C~30C
WBC ( 109/ L ) 0.00 - 99.99 3.0 ( 6.00-15.00 ) Humidity: 30 % ~85 %
RBC ( 1012 / L ) 0.00 - 16.99 2.0 ( 3.50 -10.00 )
Power Requirement
HGB ( g / L ) 0 - 250 2.0 ( 110 -180 )
A.C 100V-240V, 300 VA, 50/60 Hz
MCV(fL) 1.0 ( 60 -95 )
PLT ( 109 / L ) 0 - 2999 6.0 ( 150-500
Carryover
WBC , RBC 0.5%HGB 0.6%PLT 1 %
Sample Volume
CBC+DIFF 20 L
CBC 20 L
Whole blood and prediluted mode are supported
Throughput
Whole Blood 55 samples per hour
Prediluted 45 samples per hour
BC-5300 Vet
Auto Hematology Analyzer
is a trademark of Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Mindray Building, Keji 12th Road South,
2013 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights reserved. Specifications subject to changes without prior notice.
High-tech Industrial Park, Nanshan, Shenzhen 518057, P.R. China
P/N:ENG-BC5300Vet-210285x4-20130303
Tel: +86 755 8188 8998 Fax: +86 755 26582680
E-mail: intl-market@mindray.com www.mindray.com
Mindray is listed on the NYSE under the symbolMR
BC-5300 Vet V-53 Reagent Kit
Auto Hematology Analyzer An innovatively designed reagent kit to sustain one month use
(3 samples per day). The time-pack design ensures you lower
BC-5300Vet Auto Hematology Analyzer pampers your need to get a reliable and accurate leftover and easy replacement.
5-part differentiation hematology report for dog, cat, horse, rat, mouse, rabbit, pig, monkey,
Goat, Camel. With the state-of-art laser scatter, flow cytometry and chemical dye method,
microsampling technology, friendly PC software and economical reagent kit, it is
Friendly and Powerful Software
unprecedentedly easy and reliable to work with BC-5300Vet.
The well organized software design gives you easy control for the patient records,
sample results, diagnostic comments and report customizations. 40,000 sample results
Microsampling can be stored and you can easily search records and export statistics. Versatile report
formats facilitate your printing. Automatic maintenance programs can be executed by
Only 20ul of blood (both whole blood mode and prediluted mode)
double-clicking while operational errors can be removed by single clicking.
is enough to start the CBC+DIFF test on BC-5300Vet. By single
clicking, you can receive the complete report in
approximately1minute.
40,000
5-part Differention
The semi-conductor laser system collects
forward scatter and side scatter
information of white blood cells in a flow
cell and differentiates lymphocytes,
monocytes, neutrophils and eosinophils
according to the cell size and granule
complexity. Specific chemical dye
WBC/BASO
treatment to eosinophils can separate
from neutrophils in DIFF scattergram. In
0 100 200 fL addition, basophils are measured in an
impedance channel upon lysing action
on the RBC and other WBC groups.