InfinityLab LC/MSD Series
Agilent LC/MSD
                                           For OpenLAB CDS and OpenLAB CDS ChemStation Edition
                                           Data Sheet
The Agilent LC/MSD (liquid chromatograph/mass selective detector) achieves sensitivity and resolution specifications with
autotune.
Parameter                               Measure                                  Specification
Mass range                                                                       102,000 m/z
SIM sensitivity, ESI positive           1 pg of reserpine injected on column,    Signal-to-noise ratio, RMS >100:1
                                        quantifying the m/z 609 ion
SIM sensitivity, APCI positive          1 pg of reserpine injected on column,    Signal-to-noise ratio, RMS >10:1
                                        quantifying the m/z 609 ion
Maximum scan speed                      u/sec (Da/sec)                           10,400 u/sec (Da/sec)
Polarity switching                      Switching time from positive to          300 msec
                                        negative polarity
Mass resolution (Autotune)              Full width at half maximum               Unit resolution (0.7 u (Da))
Mass stability (drift)                  Over 12 hours                            ~0.1 u (Da) or 100 ppm, whichever is larger
Mass accuracy                                                                    0.005 u (Da)1
Spectral accuracy                                                                99.0 % spectral accuracy1
Minimum SIM dwell time                                                           5 msec
Maximum SIM transitions                                                          100 SIM ions per time segment
Dynamic range                                                                    >6  106 (electronic)
1
 Only applicable with optional Cerno MassWorks Accurate Mass Software for LC/MSD using OpenLab CDS ChemStation
Edition. Not verified during installation.
u: Unified atomic mass unit
General system specifications
    Parameter                                               Specification
    Single point of control                                 Single-point data system method capability with full control of Agilent LCs including
                                                            the 1100 Series, 1200 Series, Infinity Series, and InfinityLab Series with the
                                                            AgilentInfinityLab LC/MSD Series single quadrupole based systems.
    Supported data systems                                  OpenLAB CDS, OpenLAB CDS ChemStation Edition, and MassHunter Data Analysis1
    Time programming                                         SIM ion selection	                Drying gas flow
                                                             Scan functions	                   Nebulizer pressure
                                                             Fragmentor voltage	               Drying gas temperature
                                                             External CAN valves	              Capillary voltage (Positive)
                                                             MS diverter valve2	               Capillary voltage (Negative)
    Wide range of ionization sources                         Electrospray (ESI)
                                                             Multimode source (simultaneous ESI and APCI)
                                                             Atmospheric pressure chemical ionization (APCI)
                                                             Atmospheric pressure photoionization (APPI)
    Detector                                                High energy conversion dynode and electron multiplier horn
    Vacuum system                                           One turbo-molecular pump with one mechanical pump
1
    With data translation
2
    OpenLAB CDS only
Ordering information
G6125BA: Agilent LC/MSD System with OpenLAB CDS ChemStation Edition
Includes Agilent LC/MSD, based on single quadrupole technology, Electrospray ion source, and Agilent OpenLab CDS
ChemStation Edition data system with PC and display. System installation and familiarization included.
G6125CA: Agilent LC/MSD System for OpenLAB CDS
Includes Agilent LC/MSD, based on single quadrupole technology, Electrospray ion source, and Agilent OpenLab CDS data
system with PC and display. System installation and familiarization included.
G6887AA: Cerno MassWorks Accurate Mass Software for LC/MSD
The above are bid specifications and are not performed at installation. See Site Preparation                       www.agilent.com/chem/lcmsd
Guide, Installation Manual, and Service Notes for additional product and specification
information.
                                                                                                                   Agilent Technologies shall not be liable for errors
                                                                                                                   contained herein or for incidental or consequential
                                                                                                                   damages in connection with the furnishing, performance,
                                                                                                                   or use of this material.
                                                                                                                   Information, descriptions, and speci cations in this
                                                                                                                   publication are subject to change without notice.
                                                                                                                    Agilent Technologies, Inc. 2017
                                                                                                                   Published in the U.S.A.
                                                                                                                   April 7, 2017
                                                                                                                   5991-7829EN