0% found this document useful (0 votes)
1K views6 pages

Skalar Methods: Analysis: Ortho Phosphate Range: 5 200 G P/liter Sample: Sea Water SAN

This document provides instructions for using a Skalar method to analyze orthophosphate levels in sea water samples using colorimetric analysis. The method utilizes a reaction of ammonium molybdate and potassium antimony tartrate in an acidic medium to form a complex with phosphate that is reduced to an intensely blue color. The complex is measured at 880nm. Reagents, standards, equipment, and procedures are described to prepare samples and calibrate the system to quantify phosphate from 5-200μg/L in sea water samples. Safety precautions are also noted.

Uploaded by

Bruno Pereira
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
1K views6 pages

Skalar Methods: Analysis: Ortho Phosphate Range: 5 200 G P/liter Sample: Sea Water SAN

This document provides instructions for using a Skalar method to analyze orthophosphate levels in sea water samples using colorimetric analysis. The method utilizes a reaction of ammonium molybdate and potassium antimony tartrate in an acidic medium to form a complex with phosphate that is reduced to an intensely blue color. The complex is measured at 880nm. Reagents, standards, equipment, and procedures are described to prepare samples and calibrate the system to quantify phosphate from 5-200μg/L in sea water samples. Safety precautions are also noted.

Uploaded by

Bruno Pereira
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 6

SKALAR METHODS 

  ANALYSIS:  ORTHO‐PHOSPHATE 
  RANGE:  5 ‐ 200 µg P/liter 
  SAMPLE:  SEA WATER 
  SAN++ 
Catnr. 503‐010w/r issue 032613/MH/99279552 
 
 
PRINCIPLE 
 
The  automated  procedure  for  the  determination  of  ortho‐Phosphate  is  based  on  the  following  reaction; 
ammonium  heptamolybdate  and  potassium  antimony(III)  oxide  tartrate  react  in  an  acidic  medium  with 
diluted solutions of phosphate to form an antimony‐phospho‐molybdate complex. This complex is reduced 
to an intensely blue‐coloured complex by L(+)ascorbic acid. The complex is measured at 880 nm. 
 
 
LABORATORY FACILITIES 
 
1. Maximum power consumption depending on the analyser configuration, 2000 VA. Check voltage at the 
back of the instrument before installation. 
2. Facilities for chemical wastes. Check environmental regulations for proper disposal of waste. 
 
 
PROCEDURE SAMPLE PREPARATION 
 
Water no. 1.1.3 
 
Field of application 
 
Sample preparation for the determination of total alkalinity, aluminium, amino acids, ammonia, anionic 
surfactants (MBAS), bromide, calcium, C.O.D., chromium, conductivity, D.O.C., ethanol, hardness, 
magnesium, manganese, methanol, nitrate + nitrite, nitrite, nonionics, ortho‐phosphate, potassium 
permanganate, potassium, total‐N (UV digestion), protein, silicate, sodium, volatile acids and urea in water. 
 
Principle 
 
The sample is not preserved and analysed as soon as possible, at least within 24 hours after collection. 
Store the sample at 4°C till analysis. 
 
Note : The sample preparation in this method is only meant to be an example procedure. It requires 
validation on its correct procedure for all types of samples to be analyzed, which is the responsibility of the 
user. 
 
 
 
 
 
 
 
 
 

catnr. 503‐010w/r    1 of 6 
REAGENTS 
 
A. Ammonium heptamolybdate solution (1 liter) 
 
Required chemicals: Potassium antimony(III)  Preparation: Dissolve the potassium antimony(III)   
oxide tartrate ............ 230 mg. oxide tartrate in ± 800 ml distilled water. 
K2(SbO)2C8H4O10.3H2O  Add carefully, while constant swirling and 
Sulfuric acid # ............ 69.4 ml. cooling the sulfuric acid. Add the 
H2SO4 (95‐97%)  ammonium heptamolybdate and dissolve. 
Ammonium  Fill up to 1 liter with distilled water, add 
heptamolybdate .............. 6 g. the FFD6 and mix. 
(NH4)6Mo7O24.4H2O   
Distilled water  Note:  Do not use metal spoons for ammonium 
H2O  heptamolybdate. The sensitivity can be 
FFD6 ............................... 2 ml. increased by 50% by using 35 ml 
concentrated sulfuric acid instead of 69.4 
ml. With the 35 ml the interference of the 
silicate is 10% for 300 ppb Si and 10% for 
10 ppm Si in a range of 100 µg P/liter. The 
end pH must be < 1. Solution is stable for 
5 days. Store at 4°C when the solution is 
not used. 
 
B. L(+)Ascorbic acid solution (1 liter) 
 
Required chemicals: L(+)Ascorbic acid ............ 11 g. Preparation: Dissolve the ascorbic acid in ± 800 ml   
C6H8O6  distilled water. Add the acetone. Fill up to 
Acetone ........................ 60 ml. 1 liter with distilled water, add the FFD6 
C3H6O  and mix. 
Distilled water   
H2O  Note:  Solution is stable for 5 days. Store at  
FFD6 ............................... 2 ml. 4°C when the solution is not used. 
 
C. Rinsing liquid sampler 
 
Required chemicals: Distilled water  Note:  For sea water samples, adjust the salinity   
H2O  of the rinsing water sampler to the same 
level as the salinity of the sea water 
samples when matrix peaks reach the 
same level as the detection limit. Use ultra 
pure sodium chloride or “zero” seawater 
for adjustment. 
Refresh weekly. 
 
STANDARDS 
 
Stock solution 100 mg P/liter (1 liter) 
 
Required chemicals: Potassium dihydrogen  Preparation: Dissolve the potassium dihydrogen    
phosphate .............. 439.4 mg. phosphate in ± 800 ml distilled water. Fill 
KH2PO4  up to 1 liter with distilled water and mix. 
Distilled water   
H2O  Note:  Solution is stable for 4 weeks. Store at 
4°C when the solution is not used. 

catnr. 503‐010w/r    2 of 6 
Stock solution 10 mg P/liter 
 
Dilute 10 ml stock solution 100 mg P/liter to 100 ml with distilled water.  
 
Note: Prepare the stock solution 10 mg P/liter fresh weekly. 
 
Working standards 
 
200 µg P/liter: Dilute 2.0 ml stock solution 10 mg P/liter to 100 ml with rinsing liquid sampler. 
160 µg P/liter: Dilute 1.6 ml stock solution 10 mg P/liter to 100 ml with rinsing liquid sampler. 
120 µg P/liter: Dilute 1.2 ml stock solution 10 mg P/liter to 100 ml with rinsing liquid sampler. 
  80 µg P/liter: Dilute 0.8 ml stock solution 10 mg P/liter to 100 ml with rinsing liquid sampler. 
  40 µg P/liter: Dilute 0.4 ml stock solution 10 mg P/liter to 100 ml with rinsing liquid sampler. 
 
Note: 
1. Prepare the working standards fresh daily. 
2. The water used for the reagents and standards must be degassed properly before making up the 
reagents and standards. Especially water produced by reverse osmosis or ion exchange equipment 
contains a lot of gasses, which must be removed by degassing (degassing procedure see operational 
remarks and trouble shooting). 
 
 
CATALOGUE NUMBERS REQUIRED CHEMICALS 
 
Product  Supplier and catnr.  Danger classification 
   
Potassium antimony(III) oxide tartrate trihydrate Merck 108092 harmful, dangerous for the 
environment 
Sulfuric acid (95‐97%)  Merck 100731  corrosive 
Ammonium heptamolybdate tetrahydrate  Merck 101182 
FFD6  Skalar SC 13908
L(+)Ascorbic acid  VWR PRO   
20150.231 
Acetone  Merck 100014  highly flammable, irritant 
Potassium dihydrogen phosphate  Merck 104873
Sodium hypochlorite solution (13% active chlorine)  Acros 21925  corrosive 
 
 
RECOMMENDED OPERATIONAL SETTINGS 
 
1. System sample time: 60 sec., wash time: 60 sec., air time: 0 sec. 
2. Module sample time: 60 sec., wash time: 60 sec., air time: 0 sec. 
3. Calibration type: 1st order ISO 8466‐1. 
4. Sampler settings normal: 2, extra: 4. 
 
 
OPERATIONAL REMARKS AND TROUBLE SHOOTING 
 
1. The stabilising time of the system is approximately 20 minutes. 
2. The sensitivity of the highest standard 200 µg P/liter is ± 0.370 AU. 
 
 

catnr. 503‐010w/r    3 of 6 
3. Persons using this Skalar method should be familiar with normal laboratory practice. This Skalar 
method does not purport to address all of the safety problems, if any, associated with its use. It is the 
responsibility of the user to establish appropriate safety and health practices and to ensure compliance 
with any national regulatory conditions. 
4. The connection between the sampler and the sample pump tube is made of 5175 tube. 
5. If a digital matrix photometer 28505901, is in use, a correction interference filter of 1010 nm ± 10 nm is 
advised. 
6. If reaction coils with water heating are used, the water bath has to be cooled with tap water when the 
environment temperature is above 25°C. 
7. The reagent bottles must be rinsed thoroughly with distilled water before refilling with fresh reagents. 
This is done to remove precipitation of micro organism and other interferences. 
8. If solvents or high concentrated acids (> 2M) must be controlled by a rinsing valve, use the solvent/acid 
resistant type. 
9. Avoid any turbidity in the reagents, filter if necessary. 
10. Degassing procedures; 
a) Good: degas the water with a helium gas flow of approximately 20 liter/hour with aid of a sparger. 
Degas 5 liter water for 15 minutes. This procedure removes all dissolved air or, degas the water by 
boiling the water for 10 minutes. Cool down to room temperature before usage. 
b) Medium: degas the water with a nitrogen gas flow of approximately 20 liter/hour with aid of sparger. 
Degas 5 liter water for 15 minutes. This procedure will remove most of the dissolved gasses. 
c) Bad: degas the water by ultrasonic waves. This procedure will remove a small part of the dissolved 
gasses. 
 
 
GENERAL REMARKS 
 
1. In most analyses, the first peak coming from the baseline is lower than it should be. The first peak 
coming from the baseline is rejected and will not be included in the calculation of the CV‐value. 
2. If the sample take up volume is less than 1.00 ml/min, a bypass is required to increase the sample 
stream to approximately 1.00 ml/min. For SA 5210 debubbler, when the sample is 1.00 ml/min and 
higher use 0.23 ml/min debubble tube. 
 
 
MAINTENANCE 
 
1. Daily; when daily analysis is finished rinse all reagent lines with distilled water for 30 minutes. 
2. Weekly; to decontaminate the system, rinse with 1% diluted hypochlorite solution for 30 minutes and 
with distilled water for another 30 minutes. 
3. Monthly; replace all pump tubing (or shift pump tube one colour bridge on pump). 
 
 
MODULE CONSUMABLES 
 
silicone tube  catnr. 3150  sleeves  catnr. 5406 
polyethylene tube  catnr. 3142  acid flex sleeves  catnr. 5415 
polyethylene tube  catnr. 5141  pump tube 1.00 ml/min. catnr. 3031 
polyethylene tube  catnr. 5142  pump tube 0.42 ml/min.  catnr. 3028 
sleeves  catnr. 5400  pump tube 0.32 ml/min.  catnr. 3027 
sleeves  catnr. 5401     
 
 
 
 

catnr. 503‐010w/r    4 of 6 
MODULE COMPONENTS 
 
manifold holder  catnr. 25200105  debubbler  catnr. 5207 
module  catnr. 5107  reactor  catnr. 25500952 
end block  catnr. 5109  sinkers  catnr. 5380 
inlet connector  catnr. 5216  flow cell 50 mm catnr. 6425 
inlet connector  catnr. 9242  filter 880 nm  catnr. 6609 
 
 
REFERENCES 
 
Chemical method (no. 503) 
 
1. Standard Methods for Examination of Water and Waste Water. 15th edition 1980 APHA‐AWWA‐WPCF 
page 410‐425. 
2. Boltz, D.F., Mellon, M.G.,"Spectrophotometric determination of phosphate as molydiphosphoric acid", 
Analytical chemistry, Vol. 20, No 8, August 1948, page 749‐751. 
3. Walinga, I., van Vark, W., Houba, V.J.G., van der Lee, L.L., Plant analysis Procedures, Part 7, Department 
of Soil Science and Plant Nutrition, Wageningen Agricultural University, Syllabus 1989, Page 138‐141. 
4. ISO 15681‐2, Determination of ortho phosphate and total phosphorus contents by flow analysis, Part 2: 
Method by continuous flow analysis (CFA). 
5. ISO 3696:1987, Water for analytical laboratory use. Specification and test methods. 
6. ASTM, D1193, Standard Specification for Reagent Water. 
 
Sample preparation (water no. 1.1.3) 
 
1. Environmental Protection Agency, methods for chemical analysis of water and wastes, 1983. 
2. Standard Methods for the determination of water and waste water, 17th edition, 1989. 
3. ASTM, 1990. 
4. International Organisation for Standardisation, ISO‐5667‐3. 
 
 

catnr. 503‐010w/r    5 of 6 
FLOW DIAGRAM
waste
ml/min

25500952/5303 Flow cell 50 mm


Filter 880 nm

40°C

L(+)Ascorbic acid solution 0.32


Ammonium heptamolybdate sol. 0.32 9242 *

5207
waste
Air
Sample 1.00 5216
0.42

1.00
waste

Sampler

* polythene tube catnr. 5141


 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Revision:  Rev 1  Rev 2 Rev 3 Method:
Date:        Analysis:  Phosphate 
Chemists        Range:  5‐200 µg P/liter 
Authorization      Sample: Sea water 
      First issued: 26‐03‐2013 
 

catnr. 503‐010w/r    6 of 6 

You might also like