KRÜSS K100 Tensiometers
NEWEST TECHNOLOGY FOR HIGHEST MEASURING PERFORMANCE
Advancing Surface Science
We set the standards
KRÜSS K100 TENSIOMETERS
With the tensiometers of the K100 series, KRÜSS has defined the standard for the measurement Measuring Methods
of the surface and interfacial tension of liquids and solids. The Processor Tensiometers K100MK2
LIQUIDS
and K100C as well as the Single Fibre Tensiometer K100SF represent the perfect combination of
Du Noüy Ring Method
our latest research technology with functional and modern design. The K100 tensiometers have
Classical method for measuring surface
been specially designed for versatile and demanding applications in research, development, and and interfacial tension. Non-critical even with
quality assurance. difficult wetting behaviour. The force acting
on a liquid film raised by a ring is measured.
Rapid, precise, user-friendly, with high measuring accuracy and flexible software control -
Wilhelmy Plate Method
the K100 tensiometers are convincing in all their fields of application: Universal method, particularly suitable
for surface tension measurements over a
long period of time. The force resulting from
Determining the effectiveness and efficiency of surfactants by
the wetting of a vertically suspended plate is
CMC measurements measured.
Wetting behaviour of tablets, pharmaceutical active Density Determination
substances, and auxiliaries Determining the density of liquids
Wetting of lacquers and paints
Sedimentation and
Degradation product content in oils
Sediment Resistance
Tank release and cleaning validation in the food Determining the rate of sedimentation and
industry the sediment resistance of dispersions
Wetting and adhesion of coatings
Developing cosmetic products SOLIDS
Wetting properties of printing inks Dynamic Wilhelmy Method
For determining the advancing and
Wetting of fibre bundles, textiles or single
receding contact angles on solid samples
fibres up to a minimum diameter < 10 µm with a defined geometry.
Sedimentation and ductility of dispersions Single Fibre Wilhelmy Method
Checking surface modifications For determining the advancing and
receding contact angles on single fibres.
Powder Contact Angle Method
Measuring contact angles and the
adsorption behaviour on powders and other
porous substances. The increase in weight is
measured as a function of time.
Processor Tensiometer K100MK2
THE MULTI-TALENT
The K100MK2 is the universal instrument among tensiometers. The measuring pro- "Extended" CMC -
cesses have been optimized and made more exact by a high degree of automa- Measurement
tion and the latest technical analysis standards. The measuring accuracy of the
K100MK2 can be additionally increased by the use of the built-in ionizer. KRÜSS is
without any limits
the only manufacturer on the market that can offer this feature, for which an appli-
The proverbial last drop that causes the glass
cation for a patent has been made. The high accuracy of the results and the flexible to overflow - for KRÜSS this is no longer a
software control make the K100MK2 the top-class tensiometer. problem! The newly developed measuring
method "Extended CMC" defines the
determination of the Critical Micelle
FEATURES
Concentration in a completely new way:
Surface and interfacial tension thanks to the two dosing units the volume
added is aspirated off again immediately
measurement of liquids
after mixing. This considerably extends the
Dynamic contact angle measurements concentration range that can be used so that
Extended CMC determination with the CMC is always recorded - a great plus
for proper routine surfactant studies.
new innovative double dosing system
Surface energy calculations on solids,
powders, pigments, fibres, etc.
Measuring the density of liquids
Sedimentation and sediment
resistance measurements
Built-in ionizer for eliminating
electrostatic charges
Controlled by LabDesk™ software
Comparison of the concentration ranges of the two
CMC methods with two different surfactant samples
NO CHANCE FOR ELECTROSTATICS – Extended CMC-
MEASUREMENTS WITH THE IONIZER Measurement with two
dosing units
When conventional tensiometers are used to make sensitive measurements on small
samples, single fibres or insulating plastics electrostatic forces of attraction occur
between the solid and the liquid surface - the measured force or the detected surface height is incorrect.
KRÜSS has thought out a solution to this problem (and also applied for a patent for it): the use of an
ionizer. This ionizes the air in the sample chamber for a short time and in this manner eliminates any
electrostatic charges. The measurement can now be made without any interference with the highest
degree of accuracy.
Single Fibre Tensiometer K100SF
EXTREMELY ACCURATE
Precision in a new dimension: the PC-controlled measuring system K100SF
determines the surface and interfacial tension of liquids and solids with the highest
effectiveness and efficiency. The built-in ionizer ensures optimal measuring condi-
tions. The high accuracy of the measuring instrument guarantees the exact charac-
terization of the finest single filaments.
FEATURES Surface energy determination on
single fibre and bundles of fibres
Fully automatic measurement of contact angle and sorption,
even on single filaments
Surface energy determination on powders, small plates
and single fibres
Extended CMC determination with
new innovative double dosing system
Surface and interfacial tension
measurements of liquids Development of high-technology
synthetic fibres
Determining the density of liquids
Built-in ionizer for eliminating
electrostatic charges The Basic Version
Controlled by LabDesk™ software Processor
Tensiometer K100C
The K100C is limited to the essentials, without
APPLICATIONS neglecting the quality. The K100C offers a
wide range of measuring methods with high
Characterization of natural precision and simple handling.
and synthetic fibres
Development of carbon
and glass fibres
Optimization of hair care products
Prediction of interactions between
fibre and matrix
Measuring and
Analysis Software LabDesk™
All KRÜSS tensiometers of the K100 series are controlled by the modularly con- Export functions allow further processing of the measuring data and the incor-
structed LabDesk™ software. poration of the diagrams in other software products.
The whole range of the available Add-Ins has a uniform user interface that gua- Measuring methods adapted to be customer-specific can be programmed.
rantees the simplest operation of all KRÜSS tensiometers. Each Add-In contains
one or more measuring methods. The advantage: different Add-Ins can be used LabDesk™ has been completely
in parallel to control several, even different instruments. In this way all the tensio- developed by KRÜSS and is being
meters from KRÜSS can be operated with the same user interface. continually updated.
Measuring programs can be stored separately so that the K100 tensiometer can LabDesk™ runs under all Windows
be rapidly set up for the routine procedures of day-to-day lab work. versions from Windows 95 to XP.
The established evaluation methods for the analysis of the measuring data, such
as for calculating the CMC, the surface energy or the contact angle, are incorpo-
rated in LabDesk™.
The graphical presentations are clearly arranged
and can also be defined to be user-specific.
For a quick and accurate overview of the
measuring data.
All raw data is available for carrying out a
check or for using user-specific evaluation
methods. Gapless documentation ensures The K100
Measuring Modules Sedimentation and
GLP-conforming work. Contact Angle Penetration
The measuring modules always Measuring the dynamic contact Determining the sedimentation
Measuring report contain measuring methods from angle and the surface free energy speed and sedimentation resi-
similar fields of application. on plates, disks, powders and pig- stance of dispersions
THE DATABASES ments
A comprehensive gas, liquid and solids database provides you with a large amount of substance data. Surface and Interfacial Surfactant
This data can be entered into the corresponding input field for defining a new measurement with a Tension Characteristics Density Determination
mouse-click. Reference sources are given for each substance data, so that you, the user, can check the Surface and interfacial tension CMC measurement with Dosimat Determining the density of
quality of the data yourself. All databases are editable, so that you can add further data records at any measurements by the ring, plate control, for almost unlimited con- liquids
time. and rod methods centration ranges
Technical Data Norms & Standards
K100 SERIES K100 TENSIOMETER
K100C K100MK2 K100SF
SFT/ IFT
During the programming of the LabDesk™ control and evaluation software great
Range 1-1000 mN/m 1-1000 mN/m 1-1000 mN/m value has been placed on compliance with the GLP guidelines (Good Laboratory
Resolution 0.01 mN/m 0.001 mN/m 0.0001 mN/m
Position Practice). The report generated after a measurement meticulously protocols all the
Range >110 mm >110 mm >110 mm
measurement parameters and analysis conditions. Together with its high degree of
Resolution 20 µm 0.1 µm 0.1 µm
Speed 0.09 … 500 mm/min 0.09 … 500 mm/min 0.09 … 500 mm/min accuracy, the K100 tensiometer complies with the following Standards and Norms:
Weight Measurement
Range 120 g 210 g 3g+3g
Resolution 100 µg 10 µg 1 µg
Data Rate ~ 50 Hz ~ 50 Hz ~ 50 Hz ASTM D 0971 – 91 Standard test method for interfacial tension of oil against
Calibration/Adjustment manual (with CP0503) automatic automatic water by the ring method
Temperature Control
Range -10 to 130 °C -10 to 130 °C -10 to 90 °C ASTM D 1331 – 56 Standard test method for surface and interfacial tension of
Temperature solutions of surface active agents
Range -20 … 200 °C -60 … 450 °C -60 … 450 °C
Resolution 0.1 °C 0.01 °C 0.01 °C ASTM D 1417 – 83 Standard method of testing rubber latices-synthetic
Probes 2 2 2
Measuring Methods ASTM D 1590 – 60 Standard test method for surface tension of water
CMC (Single Dosimat) • • • DIN 53593 Prüfung von Latex: Bestimmung der Oberflächenspannung
CMC (Double Dosing System) • • •
Modified Washburn Method • • • DIN 53914 Prüfung von Tensiden; Bestimmung der Oberflächenspannung
Single Fibre Contact Angle – – •
Dynamic Wilhelmy Method • • • DIN EN 14210 Grenzflächenaktive Stoffe – Bestimmung der Grenzflächenspannung
Wilhelmy Plate Method • • • von grenzflächenaktiven Lösungen mittels Bügel- oder Ringverfahren
Du Noüy Ring Method • • •
Lenard Frame Method • • – DIN 14272 Schaummittel: Wasserfilmbildende Schaummittel zur Erzeugung
Ring Method (small samples) • • •
Plate Method (small samples) • • • von Schwerschaum für Löschzwecke
Rod Method – • • ISO 1409 – 1995 Plastics/rubber – Polymer dispersions and rubber latices (natural and
Density • • •
Sorption • • • synthetic) – Determination of surface tension by the ring method
Sedimentation • • –
Sediment Resistance • • –
ISO 6295 Determination of interfacial tension of oil against water
Sample Vessel ISO 6889 Surface active agents – Determination of interfacial tension
50 mm, 70 mm, 100 mm 50 mm, 70 mm, 100 mm 50 mm, 70 mm, 100 mm
inverse CMC inverse CMC inverse CMC by drawing up liquid films
Housing Amtsblatt der EU Oberflächenspannung
Size (L x W x H) 390 x 300 x 585 mm 390 x 300 x 585 mm 390 x 300 x 585 mm
Weight 21 kg 24 kg 24 kg L251/37 (1984)
Interface
PC RS232 (USB optional) RS232 (USB optional) RS232 (USB optional)
Auxiliary Ports 1 x RS232 2 x RS232 2 x RS232
Ionizer
– • •
Technical data are subject to change without notice.
YOUR DISTRIBUTOR:
Visit our website at:
www.kruss.de
Advancing Surface Science
KRÜSS GmbH KRÜSS GmbH KRÜSS USA
Wissenschaftliche Laborgeräte 38/40 Avenue Jean Jaurès 1020 Crews Road, Suite K
Borsteler Chaussee 85-99a F-91120 Palaiseau Matthews, NC 28105
D-22453 Hamburg
Tel.: +49 – 40 – 51 44 01-0 Tel.: +33 – 1 – 60 14 94 94 Tel.: +1 – 704 – 847 8933
Fax: +49 – 40 – 51 44 01-98 Fax: +33 – 1 – 60 14 95 48 Fax: +1 – 704 – 847 9416
E-Mail: info@kruss.de E-Mail: info@kruss.fr E-Mail: info@kruss-usa.com