Control Mixing

The document provides instructions for the proper mixing and handling of CELL-DYN calibrator and control vials. Key guidelines include avoiding exposure to heat, flame, or sunlight, and ensuring vials are gently inverted before sampling. It emphasizes the importance of proper storage and handling to maintain control recovery and prevent contamination.

Uploaded by

m_963934629
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
45 views1 page

Control Mixing

The document provides instructions for the proper mixing and handling of CELL-DYN calibrator and control vials. Key guidelines include avoiding exposure to heat, flame, or sunlight, and ensuring vials are gently inverted before sampling. It emphasizes the importance of proper storage and handling to maintain control recovery and prevent contamination.

Uploaded by

m_963934629
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

CELL-DYN Calibrator and Control

Mixing and Handling Instructions

- Do not expose to open


flame or direct sunlight.
- Do not heat by artificial
heat or waterbath.
15 Min - Do not microwave.
2
Allow vials to warm at room
1 temperature.
Check lot number, level and Do not mix. Ensure the cap
expiration date. is secure.

To minimize
30 Min out-of-range
7
Return each container to control recovery
2-10°C after use.
follow this 20-30 Sec
procedure every 3
Roll between palms for
time controls 20-30 seconds.
6 Do not use mechanical
After open sampling, are processed.
carefully wipe the rim of rocker.
container and inside of the
cap with a lint free wipe,
tightly cap.

8 or Do Not
More Shake.
Times 4
Do Not Mix vial by rapid inversion until all cells
5
Shake. are resuspended. Ensure the button is
Gently invert each vial immediately not present at the bottom of the vial. If
before sampling to ensure cellular the button is present continue mixing
resuspension. until resuspended. Control is ready to
be aspirated. If control is not aspirated
immediately, go to step 5 prior to
aspiration.

NOTE: Please refer to Assay or Info Sheet for additional information.

PN: 9150143B ©2010 Abbott Laboratories


REF: 09H68-01 CELL-DYN is a registered trademark of Abbott Laboratories in various jurisdictions.