                 Instrument Presentation
                 Typical laser wavelengths used for Raman spectroscopy
                 Laser wavelengths ranging from ultra-violet through visible to near infra-red can be used for Raman spectroscopy. Typical
                 examples include (but are not limited to):
                   Ultra-violet: 244 nm, 266 nm, 320 nm, 355 nm, 405 nm
                   Visible: 458 nm, 473 nm, 515 nm, 532 nm, 594 nm, 633 nm, 660 nm
                   Near infra-red: 785 nm, 830 nm, 980 nm, 1064 nm
                 The choice of laser wavelength has an important impact on experimental capabilities:
                 Sensitivity
                 Raman scattering intensity is  proportional  to  λ-4  where λ  is the laser wavelength. Thus an infra-red laser results in a
                 decrease in scattering intensity by a factor of 15 or more, when compared with blue/green visible lasers.
                 Spatial resolution                      
                 The diffraction limited laser spot diameter can be calculated according to the equation, diameter = 1.22 λ/NA (where λ is
                 the wavelength of the laser, and NA is the numerical aperture of the microscope objective being used). For example, with
                 a 532 nm laser, and a 0.90/100x objective, the theoretical spot diameter will be 0.72 µm – with the same objective, a
                 785 nm laser would yield a theoretical spot diameter of 1.1µm. Thus, achievable spatial resolution is partially dependent
                 on choice of laser.                                                                                                                                                             
                                        
                 Optimization of resulting based on sample behavior
                 For example:
                    Blue or green lasers can be good for inorganic materials and resonance Raman experiments (e.g., for carbon
                    nanotubes and other carbon materials) and Surface Enhanced Raman Scattering (SERS).
                    Red or near infra-red (660-830 nm) are good for fluorescence suppression.
                    Ultra-violet lasers for resonance Raman on bio-molecules (such as proteins, DNA, and RNA), and fluorescence
                    suppression.
                 Advantages and disadvantages of ultra-violet (UV) lasers for Raman
                 Ultra-violet (UV) lasers for Raman spectroscopy typically include laser wavelengths ranging from 244 nm through to 355
                 nm.
                                                                                                                                                                        
https://www.horiba.com/en_en/ramanspectrometerpresentation/                                                                                                                                         1/7
12/15/2018                                                                 Instrument Presentation - HORIBA
                 Theoretically UV Raman spectroscopy is not different from standard analysis using visible laser wavelengths. However, in
                 practice there are a number of practical difficulties and disadvantages which must be considered.
                 Advantages
                     With certain samples, UV laser excitation   can interact in ways not possible when using visible laser sources. For
                     example, in semiconductor materials, the penetration depth of UV light is typically in the order of   a few nanometers,
                     and thus UV Raman can be used to selectively analyze from a thin top surface layer (as is commonly found in silicon
                     on insulator SOI materials). In another example, UV excitation can give rise to specific resonance enhancement with
                     biological moieties, particularly protein, DNA and RNA structures. Specific analysis of these materials within tissue can
                     be difficult using visible laser wavelengths.
                     Fluorescence suppression can often be assisted using UV lasers, by spectrally separating the Raman and fluorescence
                     signatures. With visible lasers it is common that Raman and fluorescence are superimposed, and the incomparable
                     strength of the fluorescence is what can perturb (or completely mask) the Raman spectrum. With UV excitation, the
                     Raman spectrum lies close to the laser line, whereas the fluorescence is often slightly removed to higher wavelengths.
                     Thus, they no longer overlap, and the fluorescence is no longer an issue.
                     Increased sensitivity can result from UV excitation, since Raman scattering efficiency is proportional to λ-4 , where λ  is
                     the laser wavelength. Thus, Raman scattering at 325 nm is a factor of 14 times more efficient than that at 633 nm.
                  
                 Disadvantages
                     UV Raman still remains a more sophisticated technique which requires greater expertise to handle. Reasons for this
                     include the fact that the laser beam is now invisible, and that the lasers are larger, more complex, and considerably
                     more expensive.
                     Samples are more prone to burning and degradation from the laser beam since the energy per photon is increased.
                     However, new techniques such as DuoScan™ optics allow the laser beam to be rapidly rastered over the sample and
                     thus prevent immediate burning. As an example, cellulose will burn with 325 nm excitation within a few milliseconds,
                     but with DuoScan™ it remains resilient to burning for more than five minutes.
                     Many Raman systems designed for visible and near infra- red analysis are not suitable for UV Raman. UV Raman
                     requires specific mirror coatings, microscope objectives, diffraction gratings, and a CCD detector for optimized results.
                     Modern systems such as the LabRAM HR can be configured to work efficiently from the UV through to the infra-red
                     without compromise, but nonetheless the additional requirements do come at a cost.
                 Advantages and disadvantages of near infra-red (NIR) lasers for
                 Raman
                 Near infra-red (NIR) lasers for Raman typically include a range of wavelengths greater than 700 nm, such as 785 nm, 830
                 nm, 980 nm and 1064 nm. The key reason for the use of NIR Raman is for fluorescence suppression, but there are    a
                 number of drawbacks which must be considered. Whilst at times NIR Raman is invaluable, it should certainly not be
                 considered the best solution for every sample.
                 Advantages
                     Fluorescence suppression can often be assisted using NIR lasers. Fluorescence is a two photon process, which first
                     requires absorption of a photon, and is followed by the emission of a fluorescent photon. On the other hand, Raman is
                     a one photon scattering process, which does not require absorption. While many materials absorb in the visible region,
                     fewer do so in the NIR region.Thus, in many cases NIR lasers will not give rise to fluorescence (since absorption does
                     not occur), but Raman scattering will be present as normal. In cases where samples strongly fluoresce with visible
                     excitation NIR Raman can provide a solution and allow a good Raman spectrum to still be obtained.
                  
                 Disadvantages
                     Decreased sensitivity can result from NIR excitation, since Raman scattering efficiency is proportional to λ-4 , where λ is
                     the laser wavelength. Thus, Raman scattering at 785 nm is almost a factor of 5 times less efficient than that at 532
                     nm. This problem is compounded by decreasing sensitivity of the CCD detector in the NIR Raman range. As a result,
                     measurement times are often considerably increased to get similar spectral quality to measurements made with visible
                     lasers.
                                                                                                                              
https://www.horiba.com/en_en/ramanspectrometerpresentation/                                                                                       2/7
12/15/2018                                                                 Instrument Presentation - HORIBA
                     NIR lasers often have beam qualities (e.g., beam width and divergence) which are not as well suited for microscopy.
                     The result is that spatial resolution can be somewhat compromised, and thus achievable results may not meet
                     theoretical predictions.
                 Types of laser (Rayleigh) filtering used
                 There are two main classes of filters used for Raman spectroscopy.
                 Optical filters
                 These optical components are placed in the Raman beam path, and are used to selectively block the laser line (Rayleigh
                 scatter) while allowing the Raman scattered light through   to the spectrometer and detector. Each laser wavelength
                 requires an individual filter. There are two main types of filters used, both of which can be used without user intervention
                 or optimization:
                     Edge. An edge filter is a long pass optical filter which absorbs all wavelengths up to a certain point, and then transmits
                     with high efficiency all wavelengths above this point. As an example, a 532 nm edge filter will absorb all light up to
                     533.5 nm or above (e.g., 50 cm-1 or above) including the laser emission. Above 533.5 nm it will transmit light, allowing
                     detection of the Raman spectrum (stretching from 50 cm-1 up to 3500 cm-1 or  above).  The edge filters have an ultra
                     steep edge between the absorbing and transmitting spectral region, and offer excellent blocking of the laser line.The
                     advantage of the edge filter is that it is environmentally stable with a near infinite lifetime.
                     Holographic notch. A notch filter has a sharp, discrete absorption which for Raman is chosen to coincide with a
                     specific laser wavelength. Typically the absorption will be a few nanometers wide (corresponding to a few hundred cm-
                     1). As an example, for a 532 nm laser, a notch filter will be chosen with a central absorption at 532 nm. The laser line
                     will be absorbed, but the Raman spectrum above will be transmitted. Unlike the edge filter, the notch does have a finite
                     lifetime, and will degrade with time. The advantage of the notch is that it allows measurements to be made for both the
                     Stokes and anti-Stokes Raman scattering, which is useful for certain specialized measurements.
                     Tunable filtering spectrometer with a triple monochromator instrument, it is possible to work in a ‘double subtractive,
                     single spectrograph’ mode. The first two monochromators are used as an integral pair, which first disperses both the
                     Rayleigh (laser) and Raman scattered light, physically blocks the Rayleigh, and then recombines the light. The third
                     spectrometer then acts in the normal way to disperse the Raman scattered light with subsequent detection. The
                     advantage of using a triple spectrometer in this way is that the laser filter is infinitely variable, and the one instrument
                     can be used to work with any number of laser sources. In addition, the filtering performance is excellent, and allows
                     Raman analysis down to 4-5 cm-1. However, such an instrument does require rather more expertise to operate
                     compared to the more standard filter based single monochromator systems, and therefore would rarely be considered
                     for routine analysis.
                 Description of CCD detector for Raman
                                                                A CCD (Charge Coupled Device) is a silicon-based multichannel array
                 Fig 30: CCD Detector                           detector of UV, visible and near-infra light. They are used for Raman
                                                                spectroscopy because they are extremely sensitive to light (and thus suitable
                                                                for analysis of the inherently weak Raman signal), and allow multichannel
                 operation (which means that the entire Raman spectrum can be detected in a single acquisition). CCDs are widely used,
                 not least as the sensors in digital cameras, but versions for scientific spectroscopy are of a considerably higher grade to
                 give the best possible sensitivity, uniformity and noise characteristics.
                 CCD detectors are typically one dimensional (linear) or two dimensional (area) arrays of thousands or millions of individual
                 detector elements (also known as pixels). Each element interacts with light to build up a charge – the brighter the light,
                 and/or the longer  the  interaction,  the  more  charge is registered. At the end of the measurement read out, electronics
                 pull the charge from the elements, at which point each individual charge reading is measured.
                 In a typical Raman spectrometer, the Raman scattered light is dispersed using the diffraction grating, and this dispersed
                 light is then projected onto the long axis of the CCD array. The first element will detect light from the low cm-1 edge of the
                 spectrum, the second element will detect light from the next spectral position, and so on...the last element
                                                                                                                            will
                                                                                                                                  detect
                                                                                                                                       
https://www.horiba.com/en_en/ramanspectrometerpresentation/                                                                                       3/7
12/15/2018                                                                  Instrument Presentation - HORIBA
                 light from the high cm-1 edge of the spectrum.
                 CCDs require some degree of cooling to make them suitable for high grade spectroscopy. Typically this is done using
                 either peltier cooling (suitable for temperatures down to -90ºC), and liquid nitrogen cryogenic cooling. Most Raman
                 systems use peltier cooled detectors, but for certain specialized applications, liquid nitrogen cooled detectors still have
                 advantages.
                 Description of EMCCD detector for Raman
                                                                An Electron Multiplying CCD (EMCCD) is a special type of CCD detector,
                                                                which uses the latest technology to enhance the spectrum quality when
                                                                extremely low signal levels are present. This enhancement is particularly
                 valuable when the Raman signal is very weak, since the electron multiplication process can result in good spectrum
                 quality, unlike the conventional CCD where only a few of the stronger features can just be observed above the noise. The
                 benefits of EM gain are clearly obvious in fast Raman spectral imaging,where the necessary short integration times can
                 often result in signals which are barely visible above the noise when measured with a conventional CCD.
                 The EMCCD has two readout registers on the chip – a conventional register and an electron multiplying (EM) register. In
                 the EM register, the clocking voltages used are higher than for conventional clocking, causing the electrons to acquire
                 sufficient energy that impact ionization can occur. At this point, extra electrons are produced and stored in   the next
                 pixel. There is only a small probability of electrons acquiring sufficient energy for impact ionization to occur (thus creating
                 additional electrons), but since the readout register has many elements within it, significant gain factors are possible (up
                 to ~1000x). The key benefit of an EMCCD is that the amplification occurs before readout of the signal, which means that
                 the signal is not readout noise limited. In other words, through amplification, the signal is raised well above the noise floor
                 which is largely determined by the noise of the readout electronics (pre-amplifier and A/D converter).
                 Spectral resolution: definition and use
                 Spectral resolution is the ability to resolve spectral features and bands into their separate components.The spectral
                 resolution required by the analyst or researcher depends upon the application involved. For example, routine analysis for
                 basic sample identification typically requires low/medium resolution. In contrast, characterization of polymorphs and
                 crystallinity often requires high resolution, since these phenomena exhibit only very subtle changes in the Raman
                 spectrum, which would not be visible in a low resolution experiment.
                 Spectral resolution is an important experimental parameter. If the resolution is too low, spectral information will be lost,
                 preventing correct identification and characterization of the sample. If the resolution is too high, total measurement time
                 can be longer than necessary. What makes resolution “too low” or “too high” depends upon the particular application,
                 and what information is desired from the experiment.
                 Typically, low/medium resolution is suitable for basic chemical identification, and distinguishing different materials. Higher
                 resolution becomes necessary to characterize more subtle spectral features – for example, minor changes in the shape
                 or position of a peak. There are a number of chemical phenomena which cause such subtle spectral changes:
                     Crystallinity – In general Raman peaks become sharper and more intense as a material moves from an amorphous to
                     crystalline structure. High spectral resolution allows even very small changes in crystallinity to be characterized.
                     Polymorphism – Polymorphs are materials which have the same chemical formula but differing solid state forms.
                     Because the chemical formula is identical, their overall Raman spectral profiles are similar, but the influence of the solid
                     form on individual bond vibrations causes subtle changes in the spectrum. Thus, it is not unexpected to observe minor
                     changes in peak shapes and positions throughout the spectrum.
                     Intrinsic stress/strain – Some materials (most notably semiconductors) display changes in their Raman/
                     photoluminescence spectra when subjected to stress and strain. In the case of  semiconductors,  stress/ strain is
                     carefully induced within the material to produce semiconducting and/or luminescent properties necessary for use of
                                                                                                                              
https://www.horiba.com/en_en/ramanspectrometerpresentation/                                                                                       4/7
12/15/2018                                                                  Instrument Presentation - HORIBA
                     that material in working devices.Raman is used as a vital tool to characterize the stress/strain, but often the effect on
                     the spectrum is subtle – thus, high resolution is necessary.
                     Hydrogen bonding – The weak inter- and intramolecular interactions caused by hydrogen bonding can cause small
                     changes in a Raman spectrum, and with high resolution capability, the spectrum can be used to investigate such
                     interactions.
                     Protein folding – A protein’s primary structure will, of course, have a significant impact on the resulting Raman
                     spectra, but the secondary and tertiary structures comprising localized and more widespread folding cause sufficient
                     perturbation to the vibrational modes to affect the spectrum. Once again, the effect will be subtle, and only high
                     resolution analysis will allow its characterization.
                 Factors affecting spectral resolution in a Raman spectrometer
                 Spectral resolution in a dispersive Raman spectrometer is determined by four
                 main factors. In the discussions below, the effect of each factor is considered               Fig 32: Czerny Turner design
                 under the assumption that all other factors remain unchanged. In real life all of
                 these factors can exist in many varied permutations, which makes direct
                 comparison of a system’s performance and capabilities difficult.
                 Spectrometer focal length
                 The longer the focal length (e.g., the distance between the dispersing grating and detector) of the spectrometer, the
                 higher the spectral resolution. Typical Raman spectrometers have focal lengths ranging from 200 mm (for low/medium
                 resolution) through to 800 mm and higher (for high resolution). It is sometimes forgotten that a long focal length
                 spectrometer is not limited to high resolution work only – With a suitable choice of gratings (see below), a high resolution
                 spectrometer can be run in a low reso-lution mode. In this way, it is ideally suited for low/medium resolution analysis for
                 routine screening, and yet can also offer high resolution analysis for more specialized applications.
                 Diffraction grating
                 The higher the groove density of the grating (typically measured as number of grooves per millimeter), the higher the
                 spectral resolution. Typical gratings used for Raman vary from perhaps 300 gr/mm (low resolution) through to 1800
                 gr/mm (high resolution). More specialized gratings (including 2400 gr/mm and 3600 gr/mm) are also available, but have
                 certain limitations, and should not be considered general purpose. The use of higher groove density gratings cannot be
                 applied ad infinitum to increase spectral resolution, since they will have fixed practical and physical limits linked with the
                 spectrometer itself. Thus, gratings provide an initial way to improve resolution, but once their limit is reached, it is
                 necessary to move to a longer focal spectrometer.
                 Laser wavelength
                 The dispersing power of a grating/spectrometer pair can usually be considered constant in terms of wavelength.
                 However, Raman spectra use an energy related unit (Raman shift, or wavenumber, cm-1) which means that the spectral
                 resolution decreases as the laser excitation is changed from infrared to visible to ultra-violet wavelengths. As an example,
                 if a 600 gr/mm grating is used with an infrared laser, a 1200 gr/mm or 1800 gr/mm will be required with a green laser to
                 achieve a similar resolution.
                 Detector
                 Most systems have a single detector, so practically the user does not have control of this factor. However, it should be
                 noted that different detectors can be configured with different pixel sizes. The smaller the pixel the higher the achievable
                 spectral resolution.
                                                                                                                                            
https://www.horiba.com/en_en/ramanspectrometerpresentation/                                                                                          5/7
12/15/2018                                                                        Instrument Presentation - HORIBA
                 Description and interest of confocal Raman Microscope
                 A Raman microscope combines a Raman spectrometer with a standard optical microscope. The excitation laser beam  
                 is focused through the microscope to create a micro-spot with a diameter in the order of 0.5-10 µm. The Raman signal
                 from the sample is collected from a similar area, passes back through the microscope into the spectrometer and is there
                 analyzed for spectral information.
                 The Raman microscope allows Raman spectroscopy to be performed with microscopic spatial resolution. Thus it opens
                 up a new dimension in chemical analysis:
                     Analysis and identification of individual particles with dimensions down to 0.5 µm.
                     Characterization of sample features with dimensions down to 0.5 µm.
                     Location and identification of microscopic contaminants.
                     Raman mapping (imaging) of sample features to show distribution of components, with a spatial resolution down to
                     0.5 µm.
                  
                 Simply adding a microscope assists in giving lateral (XY) spatial resolution, but does not give depth (Z) spatial resolution.
                 For this confocal optics are required. There are several methods in use today, some truly confocal, others pseudo
                 confocal, which work with varying success. For a true confocal design (which incorporates a fully adjustable confocal
                 pinhole aperture) micron depth resolution is possible, allowing individual layers of a sample to be discretely analyzed.
                 Raman microscopes and confocal optics
                 Some Raman microscopes do not have confocal optics. Simply adding a
                 microscope assists in giving lateral (XY) spatial resolution, but does not give   Fig 33: Spectra of ibuprofen with different spectral
                 depth (Z) spatial resolution. For this, confocal optics are required. There are   resolution
                 several methods in use today, some truly confocal, others pseudo confocal,
                 which work with varying success. With a true confocal design (which
                 incorporates a fully adjustable confocal pinhole aperture) micron depth resolution is possible, allowing individual layers of
                 a sample to be discretely analyzed.
                                                               REQUEST FOR INFORMATION
                                                      Do you have any questions or requests? Use this form to contact our
                                                                                 specialists.
                 Name *                                                                    Message *
                 Company or Institution *
                 Email address *                                                                                                      
https://www.horiba.com/en_en/ramanspectrometerpresentation/                                                                                               6/7
12/15/2018                                                                           Instrument Presentation - HORIBA
                                                                                                        I have read and agree to the HORIBA privacy policy
                 Country
                      Afghanistan
                                                                              Submit request
                                                              Terms and Conditions     Privacy Notice   Cookies   Imprint
https://www.horiba.com/en_en/ramanspectrometerpresentation/                                                                                                  7/7