CALIBRATION
CALIBRATION THE
THE
TESTING
TESTING SYSTEM
SYSTEM
Mohamad Pauzi Ismail
Introduction
• Purpose of Calibration
• Standard Blocks
• Calibration with Normal Probe
• Calibration with Angle Probe
• Construction of DAC curve
• DGS diagram
• Coupling medium
Purpose of Calibration
• To produce reliable and reproducible
test results
Standard Test Blocks
• To perform
– Equipment characteristic verification
– Range/Time base calibration
– Reference level or sensitivity setting
• Calibration block?
• Reference block?
IIW-V1 block
IIW V2 block
ASME block
IOW block
Other blocks
• ASTM blocks
• Area amplitude block
• Distance amplitude block
EQUIPMENT
CHARACTERISTIC
• Horizontal linearity
• Screen height linearity
• Amplitude control linearity
• Resolution
• Maximum penetrating power
• Pulse length
CALIBRATION-Normal
probe
• Determine test range
• Min. TR at least = block thickness
• Calibration echoes – 1st and 2nd BWE
probe
w
flaw
w beam path distance
CALIBRATION-Angle
probe
• Use 100mm curvature of V1 or
25mm/50mm curvature of V2 block
• Range – min. 100mm for V1
• Range – 100mm/125mm for V2
Beam profile
• Use IOW or ASME block
CALIBRATION
• Probe index
• Probe angle
SENSITIVITY
SETTING
• Backwall echo method
• DAC
• DGS
COUPLING
• Water, glycerine, oils, petroleum
grease, silicon grease, wall paper
paste, etc.
• Water with wetting agent suitable
for smooth surface
• Grease/heavy oil for hot, vertical or
rough surface
Probe selection
• Normal probe – twin crystal for
t<30mm
• Angle probe
BS 3923 JIS Z 3060
6 ~ 15mm : 80o 6 ~ 40mm : 70o
15 ~ 35mm : 70o 40 ~ 60mm : 70o or 60o
35 ~ 60mm : 60o and 450 > 60mm : 70o and 45o
or 60o and 45o