Loading [a11y]/accessibility-menu.js
Effects of low charge injection densities on corrosion responses of pulsed 316LVM stainless steel electrodes [functional neuromuscular stimulation] | IEEE Journals & Magazine | IEEE Xplore

Effects of low charge injection densities on corrosion responses of pulsed 316LVM stainless steel electrodes [functional neuromuscular stimulation]


Abstract:

The safe charge injection density for pulsing of 316LVM electrodes has been reported to be 40 /spl mu/C/cm/sup 2/. However, only 20 /spl mu/C/cm/sup 2/ is available for n...Show More

Abstract:

The safe charge injection density for pulsing of 316LVM electrodes has been reported to be 40 /spl mu/C/cm/sup 2/. However, only 20 /spl mu/C/cm/sup 2/ is available for nonfaradic charge transfer and double layer charge injection. Therefore, the authors evaluated long term pulsing at 20 /spl mu/C/cm/sup 2/ with capacitor coupling.
Published in: IEEE Transactions on Biomedical Engineering ( Volume: 43, Issue: 6, June 1996)
Page(s): 660 - 663
Date of Publication: 06 August 2002

ISSN Information:

PubMed ID: 8987272

References