INTERFACE SPECIFICATION
CELL-DYN® 1800 SYSTEM
        LIST NO. 07H79-01
           REVISION A
  ABBOTT DIAGNOSTICS DIVISION
       Abbott Laboratories
    Abbott Park, IL 60064, USA
                      INTERFACE SPECIFICATION
                             CELL-DYN® 1800 SYSTEM
Abbott Laboratories                        ©2003 Abbott Diagnostics Division
Abbott Park, IL 60064, USA                        07H79-01A—February 2003
NOTES
                                                 TABLE OF CONTENTS
1.0 Introduction..................................................................................................................... 1
2.0 Mechanical Interface ...................................................................................................... 1
3.0 Electrical Interface .......................................................................................................... 2
4.0 Data Interface ................................................................................................................. 2
5.0 Communication Protocol ................................................................................................ 3
6.0 Identification (ID) Segment ............................................................................................. 5
7.0 Results Segment ............................................................................................................ 8
CELL-DYN® 1800 System                                                                                 Interface Specification
07H79-01—February 2003                                                                                                       Page i
                          NOTES
Interface Specification            CELL-DYN® 1800 System
Page ii                           07H79-01—February 2003
1.0   Introduction
      1.1    Purpose
             This document describes the interfacing characteristics of the Abbott
             CELL-DYN® 1800 automated hematology analyzer when attached to a Host.
      1.2    Definitions and Conventions
             Host: external computer or data collection system.
             CELL-DYN 1800: Abbott CELL-DYN 1800.
             This specification follows guidelines adopted at the Biomedical Instrumentation
             Interface Standards Conference held at the University of Florida in December 1980
             and at the University of Texas in Dallas, April 1981.
             Signal designators and related characteristics follow EIA Standard RS232C as
             summarized in “The Handbook of Computers and Computing,” Seidman and Flores,
             Van Nostrad Reinhold, 1984; and “Integrated Circuits Applications Handbook,”
             Arthur H. Seidman, John Wiley & Sons, 1983.
             Numeric equivalents of characters are shown as hexadecimal (hex) values.
2.0   Mechanical Interface
      2.1    Connector arrangement
             The CELL-DYN 1800 provides a standard DB-9 male connector, labeled COM1,
             mounted on the side of the instrument.
             Pins on the DB-9 connector: (standard RS-232 9 pin assignments)
               Pin 1: Data Carrier Detect (DCD input) (monitored)
               Pin 2: Data to CELL-DYN 1800 (RD-receive data) (ACK/NAK, XON/XOFF)
               Pin 3: Data to Host (LIS, results) (TD—transmit data)
               Pin 4: Data Terminal Ready (DTR) (set true)
               Pin 5: Signal Ground
               Pin 6: Data Set Ready (DSR) (monitored)
               Pin 7: Request to Send (RTS output) (set true)
               Pin 8: Clear to Send (CTS input) (monitored, if requested in Setup)
               Pin 9: Ring Indicate (RI input) (unused)
CELL-DYN® 1800 System                                                    Interface Specification
07H79-01—February 2003                                                               Page 1 of 36
3.0   Electrical Interface
      3.1      Voltage levels and electrical characteristics are as defined by the EIA
               RS-232C specification.
      3.2      The maximum recommended cable length is 30 meters, or 100 feet. The actual
               maximum workable cable length is dependent on the environment of the site, the
               selected baud rate, and the equipment being connected together.
4.0   Data Interface
      The system can be configured for data transmission using the COMPUTER SETUP Menu.
      The configurable transmission parameters include data bits, stop bits, parity and baud rate.
      4.1      The asynchronous method of data transmission (serial by bit) is used.
      4.2      All information transmitted is in character form and is represented by 7-bit or ASCII.
      4.3      Transmitted characters consist of one (1) start bit, seven (7) or eight (8) data bits
               (least significant first), one or no parity bit, and one (1) or two (2) stop bits.
      4.4      Parity may be selected as none, odd, or even.
      4.5      The transmission speed may be selected from 1200, 2400, 4800, or 9600 bits per
               second (bps).
      4.6      With the exception of the control characters mentioned in Section 5.2, only printable
               ASCII characters (hex 20 to hex 7E) are used in a message.
Interface Specification                                                   CELL-DYN® 1800 System
Page 2 of 36                                                             07H79-01—February 2003
5.0   Communication Protocol
      5.1   Communication Modes
            5.1.1   RUN Menu Automatic Transmit Mode
                    The User may select the Automatic Transmit mode on the COMPUTER
                    SETUP Menu. This mode allows the automatic transmission of results during
                    the RUN cycle. If Automatic Transmit mode has been selected, the user may
                    also choose whether or not to transmit histograms along with the count data.
            5.1.2   DATALOG Menu Transmit Mode
                    On the DATALOG Menu, the user may select results from a single sample or
                    from multiple samples for transmission. Only count data may be transmitted
                    from the Data Log menu.
            5.1.3   DISPLAY SPECIMEN Menu Transmit Mode
                    On the DISPLAY SPECIMEN Menu, the user may request transmission of
                    results from the specimen being displayed. Count data are transmitted.
                    Histograms are also transmitted if the Automatic Transmit mode is selected in
                    the COMPUTER SETUP Menu.
      5.2   Transmission Control
            Transmission control is provided in two ways: 1) XOFF/XON protocol in which the
            Host transmits an XOFF character (hex 13) to stop transmission from the System and
            an XON character (hex 11) to re-start transmission; and 2) CTS (Clear To Send)
            hardware control.
            The XOFF/XON protocol has a 1.5 second time-out. If no XON is received for 1.5
            seconds after an XOFF, then transmission resumes anyway.
            The CTS hardware control has no time-out. That is, no data transmission will be
            started with CTS false. However, the transmission will still time-out if at any one time
            CTS remains false for longer than the time-out selected in the COMPUTER SETUP
            Menu (see Section 5.3.1 below). This is a different use and meaning for time-out, but
            it is needed to prevent the Host from indefinitely delaying the CELL-DYN 1800 from
            processing the next sample. Re-transmission requests are also supported and
            discussed in Section 5.3.
CELL-DYN® 1800 System                                                       Interface Specification
07H79-01—February 2003                                                                 Page 3 of 36
      5.3      Response from Host
               5.3.1   Between-Transmission Time-out
                       The time-out interval after transmission of one message is programmable in
                       the COMPUTER SETUP Menu from 100 milliseconds to 9.9 seconds in 100
                       millisecond increments.
                       NOTE: The timer starts just as the ETX, which ends the message, is
                             transmitted.
               5.3.2   Release for Next Message
                       The Host can release the CELL-DYN 1800 to send the next message by
                       sending an ACK (06 hex) before the time-out interval passes. Otherwise, the
                       CELL-DYN 1800 will begin transmission of the next message at the end of
                       the time-out interval in 5.3.1 above.
               5.3.3   Re-transmission
                       If for any reason the Host requires re-transmission of the message, it signals
                       the CELL-DYN 1800 by sending a NAK (15 hex) before expiration of the
                       time-out. A message will be sent by the CELL-DYN 1800 a maximum of
                       three (3) times. After that, the same conditions prevail as after a time-out.
                       The Host may (it is desirable but not mandatory) acknowledge the re-
                       transmission with ACK or NAK. If ACK is sent, the CELL-DYN 1800 will
                       immediately start processing the next record (if any). If NAK is sent, the
                       record will NOT be transmitted a third time. Receipt of either the ACK or
                       NAK will allow the CELL-DYN 1800 to move to the next record more quickly.
      5.4      Message Format
               The message format consists of the following elements:
                       STX, ID SEGMENT, RESULT SEGMENT, CHECK SUM, ETX.
               5.4.1   There are four (4) types of messages, each distinguished by its ID segment:
                          WBC Histogram Message
                          RBC Histogram Message
                          PLT Histogram Message
                          Count Data Message
               5.4.2   The four messages taken together represent all the results of testing a single
                       specimen.
               5.4.3   Each message begins with STX (hex 02).
               5.4.4   The Identification (ID) segment is described fully in Section 6.
               5.4.5   The Result segment is described fully in Section 7.
               5.4.6   The Check Sum is always provided and may be optionally processed by the
                       Host to verify correct transmission. It is generated by taking the modulo-256
                       sum of all the characters in the message except the STX and ETX
                       characters. The two-digit hexadecimal representation of the Check Sum byte
                       is placed immediately before the ETX as two ASCII characters.
               5.4.7   The ETX character (hex 03) is the last character of the message.
Interface Specification                                                   CELL-DYN® 1800 System
Page 4 of 36                                                             07H79-01—February 2003
              5.4.8   Message Length
                      Each of the three Histogram messages is 1204 characters long. The Count
                      Data message is 406 characters long. These message lengths count all
                      characters from STX to ETX inclusive.
              5.4.9   Data Representation
                      Numeric data are transmitted in fields of fixed length with zeros used to fill empty
                      spaces on the left. Out-of-range numeric values are represented by strings of
                      ">" characters (hex 3E), and undefined numeric values are represented by
                      strings of "–" characters (hex 2D). Alphanumeric data are transmitted in fields
                      of fixed length enclosed in double quotation marks (hex 22). Within the
                      quotation marks, the data are right-justified and blanks (hex 20) are used to fill
                      empty spaces. Fields are separated by commas (hex 2C).
6.0     Identification (ID) Segment
6.0.1   For counting purposes, the STX mentioned in Section 5.4.3 is taken as byte 1.
        6.1      The Identification segment of each message identifies the type of message and the
                 specimen the message represents.
        6.2      Message Type—Field 1
                 The Message Type field identifies the message. There are four message types:
                         WBC Histogram Message— Type "WBC"
                         RBC Histogram Message— Type "RBC"
                         PLT Histogram Message—         Type "PLT"
                         Count Data Message—            Type "__ _" (3 blanks)
        6.3      Instrument Type—Field 2
                 The Instrument Type is an alphanumeric field of seven (7) characters enclosed in
                 double quotation marks. The string for the CELL-DYN 1800 is sent as
                 " CD1800".
        6.4      Serial Number—Field 3
                 The Instrument Serial Number is an alphanumeric field of twelve (12) characters
                 enclosed in double quotation marks. It is not available in this release, so the field is
                 transmitted as "– – – – – – – – – – – –" (twelve dashes).
        6.5      Sequence Number—Field 4
                 The Sequence Number is a numeric field of four (4) characters with a value ranging
                 from 0 to 9999.
CELL-DYN® 1800 System                                                            Interface Specification
07H79-01—February 2003                                                                      Page 5 of 36
      6.6      Specimen Type—Field 5
               The Specimen Type is the fifth field in the Identification Segment. It occupies a
               numeric field of two characters whose value identifies the specimen type as follows:
                    00:   Patient type              "PATIENT"
                    01:   Replicate 1               "REPLIC 1"
                    02:   Replicate 2               "REPLIC 2"
                    03:   Replicate 3               "REPLIC 3"
                    04:   Replicate 4               "REPLIC 4"
                    05:   Replicate 5               "REPLIC 5"
                    06:   Replicate 6               "REPLIC 6"
                    07:   Replicate 7               "REPLIC 7"
                    08:   Replicate 8               "REPLIC 8"
                    09:   Replicate 9               "REPLIC 9"
                    10:   Low Control               "LOW CTRL"
                    11:   Normal Control            "NOR CTRL"
                    12:   High Control              "HI CTRL"
                    13:   Background                "BACKGRD"
                    14:   Electrical Background     "ELEC BKGD"
                    15:   Calibrator                "CALIBRATR"
                    16:   Gain Adjustment           "GAIN ADJ"
                    17:   Auto Calibration          "AUTO CAL"
      6.7      Operator ID—Field 6
               The Operator ID is a alphanumeric field of three (3) digits enclosed in double
               quotation marks. If no Operator ID is specified, the field is transmitted as "– – –".
      6.8      Specimen Date—Field 7
               The Specimen Date, giving the date on which the specimen was run, is an
               alphanumeric field of eight (8) characters enclosed in double quotation marks. The
               default format of the date is MM/DD/YY, where MM represents the month in two
               digits, DD represents the day of the month, and YY represents the year. The date
               format can be changed in the DATE/TIME Menu.
      6.9      Specimen Time—Field 8
               The Specimen Time is a alphanumeric field of five (5) characters enclosed in double
               quotation marks. It gives the time at which the specimen was run in standard 24-
               hour format.
      6.10     Specimen ID—Field 9
               The Specimen ID is an alphanumeric field of sixteen (16) characters enclosed in
               double quotation marks. If the specimen is a patient specimen, the Specimen ID is
               entered by the operator. Otherwise, the Specimen ID is generated by the
               CELL-DYN 1800 to identify special types of samples, such as controls. If not
               specified, the Specimen ID is transmitted as "– – – – – – – – – – – – – – – – "
               (16 dashes). See Section 6.6 (Field 5).
Interface Specification                                                  CELL-DYN® 1800 System
Page 6 of 36                                                            07H79-01—February 2003
     6.11   Specimen Name—Field 10
            The Specimen Name is the tenth field in the Identification Segment. It occupies a
            text field of sixteen (16) characters enclosed in double quotation marks, and it is
            undefined except for patient samples.
     6.12   Specimen Sex—Field 11
            The Specimen Sex field consists of one (1) character “M” or “F” enclosed in double
            quotation marks. If not entered by the operator, Specimen Sex is transmitted as a
            blank " ".
     6.13   Specimen Date of Birth (DOB)—Field 12
            The Specimen Date of Birth field has an identical format to Field 7, except that Field
            7 is automatically generated by the CELL-DYN 1800. If DOB is not entered by the
            operator, it is transmitted as "bb/bb/bb" where the “bb”s are actually 2 ASCII blanks
            each.
     6.14   Dr Name—Field 13
            The Dr Name field is an alphanumeric field of twenty two (22) characters enclosed in
            double quotation marks.
     6.15   Collection Date—Field 14
            The Collection Date field is an alphanumeric field of five (5) characters enclosed in
            double quotation marks. The format is MM/DD, where MM represents the month in
            two digits and DD represents the day. The year is not included. It gives the date at
            which the specimen was collected. If the Collection Date is not entered, it is
            transmitted as "bb/bb" where the “bb”s are 2 ASCII blanks as in Field 12 above.
     6.16   Collection Time—Field 15
            The Collection Time field is an alphanumeric field of five (5) characters enclosed in
            double quotation marks. It gives the time at which the specimen was collected. If
            the Collection Time is not entered, it is transmitted as "bb:bb" (“bb”s are 2 ASCII
            blanks each and “:” is the middle byte).
     6.17   Comment—Field 16
            The Comment field is an alphanumeric field of sixteen (16) characters enclosed in
            double quotation marks.
            NOTE: Fields 1 to 16 above are common in both format AND actual content for a
            given sample and, as such, are transmitted multiple times when histogram
            transmission is requested (first—if sent—with all three histograms and then again
            with the count record). Given today’s typical transmission speeds and the lengths of
            the rest of the fields, the convenience of having the header information combined
            with each data group has been chosen over the absolute smallest or fastest
            transmission. Also, in the time frame of even normal sample processing, the
            transmissions (at 9600 baud) should take less than 4 seconds out of the 60 second
            (nominal) cycle time.
CELL-DYN® 1800 System                                                      Interface Specification
07H79-01—February 2003                                                               Page 7 of 36
7.0   Results Segment
      7.1      The results for each specimen are sent in the format described. All numeric fields
               are integers, and some need to be scaled by the Host.
      7.2      Histogram Messages
               Refer to Table 1, Histogram Messages, and Table 2, Histogram Record Dump.
               7.2.1   Scale Factor—Field 17
                       The Scale Factor is a numeric field of five (5) characters. It is not
                       implemented at this time and is therefore always transmitted as 00000 (5
                       ASCII zeros without the surrounding double quotes).
               7.2.2   Channel Data—Fields 18 through 273
                       The Channel Data fields are numeric fields of three (3) characters each,
                       giving normalized counts for every channel of the designated histogram
                       (WBC, RBC, or PLT). Data for each channel will occupy 4 bytes total since
                       there will be a comma between each 3-digit value. Each value is right-
                       justified with ASCII zero “0” (not null) characters to keep the fields a fixed
                       length. So a channel value of binary zero is transmitted as three ASCII zero
                       characters (000).
                       NOTE: A comma (,) follows the last field which immediately precedes the
                       CRC characters.
      7.3      Count Data Message
               Refer to Table 3, Count Data Messages, and Table 4, Count Record Dump.
               The parameters reported by the instrument may be represented in any of four
               different sets of measurement units as follows:
                       Set 1—Standard USA
                       Set 2—SI
                       Set 3—Modified SI (HGB/MCHC in mmol/L, MCH in fmol)
                       Set 4—Modified SI (HCT/PCT in %)
               NOTE: The numeric values transmitted to the host are always sent in Standard USA
                     units even if the unit of measure, also being transmitted, is Set 2, 3, or 4.
               To convert the integer transmitted in the field corresponding to a particular
               parameter to the correct value for that parameter in the units being used, the
               decimal point must be moved leftward from its implied position to the right of the
               integer. The description of each field gives a shift count indicating how many places
               to move the decimal point in each case. If, for example, the decimal point is to be
               moved two places to the left, a field value of 00123 becomes 1.23. An integer
               representing the Units Set currently in effect is transmitted in the Units Set field
               (Section 7.3.46).
               For Units Set 3, the HGB, MCH, and MCHC values must be multiplied by 0.6206
               after the decimal point has been positioned.
               The units of measure associated with the four sets are explained in the
               CELL-DYN 1800 Operator’s Manual.
Interface Specification                                                  CELL-DYN® 1800 System
Page 8 of 36                                                            07H79-01—February 2003
     7.3.1   WBC Count—Field 17
             The WBC Count is in a numeric field five (5) characters long.
             Units Set          Shift Count             Units Label
                1                    1                     K/µL
                2–3                  1                     G/L
                4                    1                     10E9/L
     7.3.2   Spare Field—Field 18
             The Spare Field is a numeric field of five (5) characters. It has a constant value
             of zero (0) at this time, but is transmitted as five “-“ (dashes) without surrounding
             double quotes, since it is a numeric field.
     7.3.3   LYM Count—Field 19
             The LYM Count is in a numeric field five (5) characters long.
             Units Set          Shift Count             Units Label
                1                    1                     K/µL
                2–3                  1                     G/L
                4                    1                     10E9/L
     7.3.4   MID Count—Field 20
             The MID Count is in a numeric field five (5) characters long.
             Units Set          Shift Count             Units Label
                1                    1                     K/µL
                2–3                  1                     G/L
                4                    1                     10E9/L
     7.3.5   GRAN Count—Field 21
             The GRAN Count is in a numeric field five (5) characters long.
             Units Set          Shift Count             Units Label
                1                    1                     K/µL
                2–3                  1                     G/L
                4                    1                     10E9/L
CELL-DYN® 1800 System                                                    Interface Specification
07H79-01—February 2003                                                              Page 9 of 36
          7.3.6   Spare Field—Field 22
                  The Spare Field is a numeric field of five (5) characters. It has a constant value
                  of zero (0) at this time, but is transmitted as five “-“ (dashes) without surrounding
                  double quotes, since it is a numeric field.
          7.3.7   RBC Count—Field 23
                  The RBC Count is in a numeric field five (5) characters long.
                  Units Set          Shift Count             Units Label
                     1                    2                     M/µL
                     2–3                  2                     T/L
                     4                    2                     10E12/L
          7.3.8   HGB Value—Field 24
                  The HGB Value is in a numeric field, five (5) characters long.
                  Units Set          Shift Count             Units Label
                     1                    1                     g/dL
                     2                    0                     g/L
                     3                    1 (x 0.6206)          mmol/L
                     4                    0                     g/L
          7.3.9   HCT Value—Field 25
                  The HCT Value is in a numeric field five (5) characters long.
                  Units Set          Shift Count             Units Label
                     1                    1                       %
                     2–3                  3                       L/L
                     4                    1                       %
          7.3.10 MCV Value—Field 26
                  The MCV Value is in a numeric field five (5) characters long.
                  Units Set          Shift Count            Units Label
                     1–4                  1                       fL
Interface Specification                                                   CELL-DYN® 1800 System
Page 10 of 36                                                            07H79-01—February 2003
        7.3.11 MCH Value—Field 27
              The MCH Value is in a numeric field five (5) characters long.
              Units Set         Shift Count             Units Label
                 1-2                 1                       pg
                 3                   2 (x 0.6206)            fmol
                 4                   1                       pg
        7.3.12 MCHC Value—Field 28
              The MCHC Value is in a numeric field five (5) characters long.
              Units Set         Shift Count             Units Label
                 1                   1                     g/dL
                 2                   0                     g/L
                 3                   1 (x 0.6206)          mmol/L
                 4                   0                     g/L
        7.3.13 RDW Value—Field 29
              The RDW Value is in a numeric field five (5) characters long.
              Units Set         Shift Count             Units Label
                 1–4                 1                       %
        7.3.14 PLT Count—Field 30
              The PLT Count is in a numeric field five (5) characters long.
              Units Set         Shift Count             Units Label
                 1                   0                     K/µL
                 2–3                 0                     G/L
                 4                   0                     10E9/L
        7.3.15 MPV Value—Field 31
              The MPV Value is in a numeric field five (5) characters long.
              Units Set         Shift Count             Units Label
                 1–4                 1                        fL
CELL-DYN® 1800 System                                                   Interface Specification
07H79-01—February 2003                                                          Page 11 of 36
          7.3.16 PCT Value*—Field 32
                 The PCT Value is in a numeric field five (5) characters long.
                 Units Set          Shift Count            Units Label
                    1               2                          %
                    2–3             1                          mL/L
                    4               2                          %
          7.3.17 PDW Value*—Field 33
                 The PDW Value is in a numeric field five (5) characters long.
                 Units Set          Shift Count            Units Label
                    1–4                  1                     10(GSD)
                 * Clinical significance has not been established for these parameters; therefore
                   they are not reportable in the U.S.
          7.3.18 Spare Field—Field 34
                 The Spare Field is a numeric field of five (5) characters. It has a constant value
                 of zero (0) at this time, but is transmitted as five “-“ (dashes) without surrounding
                 double quotes, since it is a numeric field.
          7.3.19 LYM % Value—Field 35
                 The LYM % Value is in a numeric field five (5) characters long.
                 Units Set          Shift Count           Units Label
                    1–4                  1                     %
          7.3.20 MID % Value—Field 36
                 The MID % Value is in a numeric field five (5) characters long.
                 Units Set          Shift Count           Units Label
                    1–4                  1                     %
          7.3.21 GRAN % Value—Field 37
                 The GRAN % Value is in a numeric field five (5) characters long.
                 Units Set          Shift Count           Units Label
                    1–4                  1                     %
Interface Specification                                                  CELL-DYN® 1800 System
Page 12 of 36                                                           07H79-01—February 2003
     7.3.22 Spare Field—Field 38
           The Spare Field is a numeric field of five (5) characters. It has a constant value of
           zero (0) at this time, but is transmitted as five “-“ (dashes) without the surrounding
           double quotes, since it is a numeric field.
     7.3.23 Spare Field—Field 39
           The Spare Field is in a numeric field. It has not been implemented at this time, so it
           is always transmitted as a “-“ single dash without the surrounding double quotes.
     7.3.24 Spare Flag—Field 40
           The Spare Flag is a numeric field of one (1) character. It has a constant value of
           zero (0) at this time, but is transmitted as a single “-“ (dash) without the surround
           double quotes, since is presumed to be a spare numeric flag.
     7.3.25 R4 WBC Flag—Field 41
           The R4 WBC Flag is in a numeric field of one character. A value of 1 indicates
           that the flag is set, and a value of 0 indicates that it is clear.
     7.3.26 GR3 WBC Flag—Field 42
           The GR3 WBC Flag is in a numeric field of one character. A value of 1 indicates
           that the GRAN R3 flag is set, and a value of 0 indicates that it is clear.
     7.3.27 MR3 WBC Flag—Field 43
           The MR3 WBC Flag is in a numeric field of one character. A value of 1 indicates
           that the Mono R3 WBC flag is set, and a value of 0 indicates that it is clear.
     7.3.28 MR2 WBC Flag—Field 44
           The MR2 WBC Flag is in a numeric field of one character. A value of 1 indicates
           that the Mono R2 WBC flag is set, and a value of 0 indicates that it is clear.
     7.3.29 LR2 WBC Flag—Field 45
           The LR2 WBC Flag is in a numeric field of one character. A value of 1 indicates that
           the Lym R2 WBC flag is set, and a value of 0 indicates that it is clear.
     7.3.30 R1 WBC Flag—Field 46
           The R1 WBC Flag is in a numeric field of one character. A value of 1 indicates that
           the R1 WBC flag is set, and a value of 0 indicates that it is clear.
     7.3.31 R0 WBC Flag—Field 47
           The R0 WBC Flag is in a numeric field of one character. A value of 1 indicates that
           the flag is set, and a value of 0 indicates that it is clear.
CELL-DYN® 1800 System                                                    Interface Specification
07H79-01—February 2003                                                            Page 13 of 36
          7.3.32 Spare Flags—Fields 48 through 60
                 There are thirteen Spare Flag fields which are each a numeric field of one (1)
                 character. It has a constant value of zero (0) at this time. These are transmitted
                 as follows (no surrounding quotes): –,–,–,–,–,–,–,–,–,–,–,–,–,
          7.3.33 LRI Flag—Field 61
                 The LRI Flag is in a numeric field of two characters. A value of 16 indicates that
                 the flag is set, and a value of 00 indicates that it is clear.
          7.3.34 URI Flag—Field 62
                 The URI Flag is in a numeric field of two characters. A value of 32 indicates that
                 the flag is set, and a value of 00 indicates that it is clear.
          7.3.35 WBC Lower Meniscus Time—Field 63
                 The WBC Lower Meniscus Time is in a numeric field five (5) characters long. It
                 gives the time in units of milliseconds.
          7.3.36 WBC Upper Meniscus Time—Field 64
                 The WBC Upper Meniscus Time is in a numeric field five (5) characters long. It
                 gives the time in units of milliseconds.
          7.3.37 RBC Lower Meniscus Time—Field 65
                 The RBC Lower Meniscus Time is in a numeric field five (5) characters long. It
                 gives the time in units of milliseconds.
          7.3.38 RBC Upper Meniscus Time—Field 66
                 The RBC Upper Meniscus Time is in a numeric field five (5) characters long. It
                 gives the time in units of milliseconds.
          7.3.39 Recount RBC Lower Meniscus Time—Field 67
                 The Recount RBC Lower Meniscus Time is in a numeric field five (5) characters
                 long. It gives the time in units of milliseconds. It has a value of 0 if there was no
                 recount.
          7.3.40 Recount RBC Upper Meniscus Time—Field 68
                 The Recount RBC Upper Meniscus Time is in a numeric field five (5) characters
                 long. It gives the time in units of milliseconds. It has a value of 0 if there was no
                 recount.
          7.3.41 Limits Set—Field 69
                 The Limits Set is a numeric field of one character. It indicates which set of
                 patient limits (1 to 4) was in effect for the given patient specimen.
Interface Specification                                                 CELL-DYN® 1800 System
Page 14 of 36                                                          07H79-01—February 2003
     7.3.42 Sample Mode—Field 70
            The Sample Mode is an alphanumeric field of one character. An “O” (letter O)
            indicates that the specimen was taken in Open Sample mode; a “P” indicates Pre-
            Dilute mode.
     7.3.43 RBC Metering Fault Flag—Field 71
            The RBC Metering Fault flag is a numeric field of one character. A “0” (zero)
            indicates that there was no metering fault. A “1” (one) indicates a flow error, and a
            “2” (two) indicates a clog.
     7.3.44 WBC Metering Fault Flag—Field 72
            The WBC Metering Fault flag is a numeric field of one character. A “0” (zero)
            indicates that there was no metering fault. A “1” (one) indicates a flow error, and a
            “2” (two) indicates a clog.
     7.3.45 Sampling Error or Other Sample Processing Fault— Field 73
            This is a numeric field of one character. A “0” (zero) indicates there was no error. A
            “1” (one) indicates there was an error.
     7.3.46 Units Set Field—Field 74
            This is a numeric field of one character. The value is 1, 2, 3, or 4 as described in
            Section 7.3.
  There are four sets of measurement units. A problem arises under the following conditions:
         1. Some data is collected and stored under one unit of measure.
         2. Subsequently, data is collected and stored under another unit of measure.
         3. All the data is transmitted to a Laboratory Information System.
  The numeric results data, starting in field 17, are always sent as Standard USA units even
  though that data may be displayed in a different unit of measure on the instrument. The
  displayed unit of measure, sent in field 74, is transmitted for informational purposes only and
  may or may not correspond to the Standard USA format of the transmitted numeric results data.
  NOTE: A comma (,) follows the last field which immediately precedes the CRC characters.
CELL-DYN® 1800 System                                                      Interface Specification
07H79-01—February 2003                                                              Page 15 of 36
                             TABLE 1 HISTOGRAM MESSAGES
                                                         Start at        Byte
          Field #   Description                           Byte          Length       Paragraph #
            N/A     STX (0x02)                              1              1            5.4.3
             1      Message Type                            2             3*              6.2
             2      Instrument Type                         8             7*              6.3
             3      Serial Number                           18            12*             6.4
             4      Sequence Number                         33             4              6.5
             5      Specimen Type                           38             2              6.6
             6      Operator ID                             41            3*              6.7
             7      Specimen Date                           47            8*              6.8
             8      Specimen Time                           58            5*              6.9
             9      Specimen ID                             66            16*            6.10
            10      Specimen Name                           85            16*            6.11
            11      Specimen Sex                           104            1*             6.12
            12      Specimen DOB                           108            8*             6.13
            13      Dr Name                                119            22*            6.14
            14      Collect Date                           144            5*             6.15
            15      Collect Time                           152            5*             6.16
            16      Comment                                160            16*            6.17
                    The above fields are common
                    to both this Table and Table 3
            17      Scale Factor (n/a)                    179              5             7.2.1
            18      Channel 1 Data                        185              3             7.2.2
             "      " "                                     "              "               ...
            273     Channel 256 Data                      1205             3             7.2.2
            N/A     CRC (2 hex ASCII printable)           1209             2             5.4.6
            N/A     ETX (0x03)                            1211             1             5.4.7
        *indicates characters enclosed in double quotation marks.
        Note: If a field contains 3 characters plus double quotation marks and a comma, the
              total number of bytes used is 6. See the Histogram Message Example.
Interface Specification                                              CELL-DYN® 1800 System
Page 16 of 36                                                       07H79-01—February 2003
                                     TABLE 2 HISTOGRAM RECORD DUMP
02 22 57 42 43 22 2C 22 20 43 44 31 37 30 30 22                        . "WBC", "CD1800"
2C 22 2D 2D 2D 2D 2D 2D 2D 2D 2D 2D 2D 2D 22 2C                        , "– – – – – – – – – – – –",
34 32 38 38 2C 30 30 2C 22 2D 2D 2D 22 2C 22 31                        4288,00, "– – –","1
30 2F 31 39 2F 39 34 22 2C 22 31 36 3A 34 34 22                        0/19/94", "16:44"
2C 22 2D 2D 2D 2D 2D 2D 2D 2D 2D 2D 2D 2D 2D 2D                        , "– – – – – – – – – – – – – – – –" , “
2D 2D 22 2C 22 20 20
20 20 20 20 20 20 20 20 20 20 20 20 20 31 22 2C                                1",
22 4D 22 2C 22 20 20 2F 20 20 2F 20 20 22 2C 22                        "M" , " / / " , "
20 20 20 20 20 20 20 20 20 20 20 20 20 20 20 20
20 20 20 20 20 20 22 2C 22 20 20 2F 20 20 22 2C                                " , " / ",
22 20 20 3A 20 22 22 C 22 20 20 20 20 20 20 20                                 " :","
20 20 20 20 20 20 20 20 20 22 2C 30 30 30 30 30                                  ",00000
2C 30 30 30 2C 30 30 30 2C 30 30 30 2C 30 30 30                        ,000,000,000,000
2C 30 30 30 2C 30 30 30 2C 30 30 30 2C 30 30 30                        ,000,000,000,000
2C 30 30 30 2C 30 34 32 2C 30 34 38 2C 30 35 34                        ,000,042,048,054
2C 30 35 37 2C 30 35 39 2C 30 35 35 2C 30 34 35                        ,057,059,055,045
2C 30 33 35 2C 30 32 35 2C 30 32 30 2C 30 31 36                        ,035,025,020,016
  ...
  ... (channels have been deleted)
  ...
2C 30 35 32 2C 30 35 33-2C 30 35 31 2C 30 35 30                        ,052,053,051,050
2C 30 34 39 2C 30 34 38-2C 30 34 37 2C 30 30 30                        ,049,048,047,000
2C 30 30 30 2C 30 30 30-2C 30 30 30 2C 30 30 30                        ,000,000,000,000
2C 36 35 03                                                            ,65.
Note: Each line contains 16 characters starting with 1, not zero. The “02” and “.” bolded on the
      first line refer to the STX character. The “03” and “65” bolded on the last line refer to the
      ETX and CRC characters respectively.
CELL-DYN® 1800 System                                                          Interface Specification
07H79-01—February 2003                                                                       Page 17 of 36
                            TABLE 3 COUNT DATA MESSAGES
          Field #   Description                      Start at       Byte    Paragraph #
            n/a     STX (0x02)                          1            1         5.4.3
             1      Message Type                        2            3*          6.2
             2      Instrument Type                     8            7*          6.3
             3      Serial Number                       18          12*          6.4
             4      Sequence Number                     33           4           6.5
             5      Specimen Type                       38           2           6.6
             6      Operator ID                         41           3*          6.7
             7      Specimen Date                       47           8*          6.8
             8      Specimen Time                       58           5*          6.9
             9      Specimen ID                         66          16*         6.10
            10      Specimen Name                       85          16*         6.11
            11      Specimen Sex                       104           1*         6.12
            12      Specimen DOB                       108           8*         6.13
            13      Dr Name                            119          22*         6.14
            14      Collect Date                       144           5*         6.15
            15      Collect Time                       152           5*         6.16
            16      Comment                            160          16*         6.17
                    The above fields are common
                    to both this Table and Table 1
             17     WBC Count                          179           5         7.3.1
             18     Spare Field                        185           5         7.3.2
             19     LYM Count                          191           5         7.3.3
             20     MID Count                          197           5         7.3.4
             21     GRAN Count                         203           5         7.3.5
             22     Spare Field                        209           5         7.3.6
             23     RBC Count                          215           5         7.3.7
             24     HGB Value                          221           5         7.3.8
             25     HCT Value                          227           5         7.3.9
             26     MCV Value                          233           5         7.3.10
             27     MCH Value                          239           5         7.3.11
             28     MCHC Value                         245           5         7.3.12
             29     RDW Value                          251           5         7.3.13
             30     PLT Count                          257           5         7.3.14
             31     MPV Value                          263           5         7.3.15
             32     PCT Value                          269           5         7.3.16
             33     PDW Value                          275           5         7.3.17
             34     Spare Field                        281           5         7.3.18
             35     LYM % Value                        287           5         7.3.19
             36     MID % Value                        293           5         7.3.20
             37     GRAN % Value                       299           5         7.3.21
             38     Spare Field                        305           5         7.3.22
             39     Spare Field                        311           1         7.3.23
Interface Specification                                          CELL-DYN® 1800 System
Page 18 of 36                                                   07H79-01—February 2003
                           TABLE 3 COUNT DATA MESSAGES
                                        (Continued)
          40      Spare Field                            313              1            7.3.24
          41      R4 WBC Flag                            315              1            7.3.25
          42      GR3 WBC Flag                           317              1            7.3.26
          43      MR3 WBC Flag                           319              1            7.3.27
          44      MR2 WBC Flag                           321              1            7.3.28
          45      LR2 WBC Flag                           323              1            7.3.29
          46      R1 WBC Flag                            325              1            7.3.30
          47      R0 WBC Flag                            327              1            7.3.31
        48—60     Spare Flags                            329              1            7.3.32
          61      LRI Flag                               355              2            7.3.33
          62      URI Flag                               358              2            7.3.34
          63      WBC Lower Meniscus Time                361              5            7.3.35
          64      WBC Upper Meniscus Time                367              5            7.3.36
          65      RBC Lower Meniscus Time                373              5            7.3.37
          66      RBC Upper Meniscus Time                379              5            7.3.38
                  Recount RBC Lower Meniscus
          67                                             385              5            7.3.39
                  Time
                  Recount RBC Upper Meniscus
          68                                             391              5            7.3.40
                  Time
          69      Limits Set                             397             1             7.3.41
          70      Sample Mode                            399             1*            7.3.42
          71      RBC Metering Fault Flag                403             1             7.3.43
          72      WBC Metering Fault Flag                405             1             7.3.44
                  Sampling Error/Incomplete
          73                                             407              1            7.3.45
                  Aspiration Flag
          74      Units Set Field                        409              1            7.3.46
          n/a     CRC (2 hex ASCII printable)            411              2            5.4.6
          n/a     ETX (0x03)                             413              1            5.4.7
      * indicates characters enclosed in double quotation marks
      Note: If a field contains 3 characters plus double quotation marks and a comma, the total
            number of bytes used is 6. See the Count Data Message Example.
CELL-DYN® 1800 System                                                   Interface Specification
07H79-01—February 2003                                                           Page 19 of 36
                                TABLE 4 COUNT RECORD DUMP
02 22 20 20 20 22 2C 22-20 43 44 31 37 30 30 22              . " " , " CD1800"
2C 22 2D 2D 2D 2D 2D 2D-2D 2D 2D 2D 2D 2D 22 2C              , "– – – – – – – – – – – –",
34 32 38 38 2C 30 30 2C-22 2D 2D 2D 22 2C 22 31              4288,00, "– – –","1
30 2F 31 39 2F 39 34 22-2C 22 31 36 3A 34 34 22              0/19/94", "16:44"
2C 22 2D 2D 2D 2D 2D 2D-2D 2D 2D 2D 2D 2D 2D 2D              , "– – – – – – – – – – – – – – – –" , "
2D 2D 22 2C 22 20 20
20 20 20 20 20 20 20 20-20 20 20 20 20 31 22 2C                       1",
22 4D 22 2C 22 20 20 2F-20 20 2F 20 20 22 2C 22              "M" , " /      / ","
20 20 20 20 20 20 20 20-20 20 20 20 20 20 20 20
20 20 20 20 20 20 22 2C-22 20 20 2F 20 20 22 2C                       "," / ",
22 20 20 3A 20 20 22 2C-22 20 20 20 20 20 20 20              " :","
20 20 20 20 20 20 20 20-20 22 2C 30 30 35 32 35                       ",00525
2C 2D 2D 2D 2D 2D 2C 30-30 30 31 30 2C 30 30 30              ,– – – – –,00010,000
31 36 2C 30 30 34 39 39-2C 2D 2D 2D 2D 2D 2C 30              16,00499,– – – – –,0
30 32 35 31 2C 30 30 30-37 34 2C 30 30 32 32 36              0251,00074,00226
2C 30 30 39 30 30 2C 30-30 32 39 35 2C 30 30 33              ,00900,00295,003
32 37 2C 30 30 31 38 37-2C 30 30 33 39 35 2C 30              27,00187,00395,0
30 30 36 38 2C 30 30 30-32 37 2C 30 30 31 38 38              0068,00027,00188
2C 2D 2D 2D 2D 2D 2C 30-30 30 31 39 2C 30 30 30              ,– – – – –,00019,000
33 31 2C 30 30 39 35 30-2C 2D 2D 2D 2D 2D 2C 2D              31,00950,– – – – –,–
2C 2D 2C 30 2C 30 2C 30-2C 31 2C 31 2C 30 2C 30              ,–,0,0,0,1,1,0,0
2C 2D 2C 2D 2C 2D 2C 2D-2C 2D 2C 2D 2C 2D 2C 2D              ,–,–,–,–,–,–,–,–
2C 2D 2C 2D 2C 2D 2C 2D-2C 2D 2C 30 30 2C 30 30              ,–,–,–,–,–,00,00
2C 30 34 38 33 37 2C 30-32 31 32 31 2C 30 36 34              ,04837,02121,064
31 31 2C 30 35 30 37 30-2C 30 30 30 30 30 2C 30              11,05070,00000,0
30 30 30 30 2C 34 2C 22-4F 22 2C 30 2C 30 2C 30              0000,4, "O",0,0,0
2C 31 2C 33 37 03                                            ,1,37.
Interface Specification                                      CELL-DYN® 1800 System
Page 20 of 36                                               07H79-01—February 2003
                          HISTOGRAM MESSAGE EXAMPLE
               Actual transmission          [remarks for text here]
               (no CR/LF)
               STX                          [start of text]
               "WBC",                       [message type]                  Field 1
               " CD1800",                   [instrument type]
               "– – – – – – – – – – – –",   [serial #]
               1040,                        [sequence no.]
               00,                          [specimen type]                 Field 5
               "123",                       [operator ID]
               "08/31/03",                  [specimen date]
               "12:34",                     [specimen time]
               "123456789ABCDEFG",          [specimen ID]
               "         JOHN DOE",         [specimen name]                 Field 10
               "M",                         [specimen sex]
               "10/15/46",                  [specimen DOB]
               "         DR JANE DOE",      [doctor name]
               "08/30",                     [collect date]
               "09:50",                     [collect time]                  Field 15
               "         sample 1",         [comment]                       Field 16
               00000,                       [scale factor n/a]              Field 17
               000,                         [channel 1 data]                Field 18
               000,                         [channel 2 data]
               000,                         [channel 3 data]
               000,                         [channel 4 data]
               000,                         [channel 5 data]
               000,                         [channel 6 data]
CELL-DYN® 1800 System                                            Interface Specification
07H79-01—February 2003                                                   Page 21 of 36
                           HISTOGRAM MESSAGE EXAMPLE
                                   (Continued)
                    000,                         [channel 7 data]
                    000,                         [channel 8 data]
                    000,                         [channel 9 data]
                    000,                         [channel 10 data]
                    000,                         [channel 11 data]
                    000,                         [channel 12 data]
                    000,                         [channel 13 data]
                    000,                         [channel 14 data]
                    000,                         [channel 15 data]
                    000,                         [channel 16 data]
                    000,                         [channel 17 data]
                    000,                         [channel 18 data]
                    000,                         [channel 19 data]
                    000,                         [channel 20 data]
                    000,                         [channel 21 data]
                    000,                         [channel 22 data]
                    000,                         [channel 23 data]
                    000,                         [channel 24 data]
                    000,                         [channel 25 data]
                    034,                         [channel 26 data]
                    034,                         [channel 27 data]
                    034,                         [channel 28 data]
                    034,                         [channel 29 data]
Interface Specification                                      CELL-DYN® 1800 System
Page 22 of 36                                               07H79-01—February 2003
                         HISTOGRAM MESSAGE EXAMPLE
                                 (Continued)
               054,                            [channel 30 data]
               054,                            [channel 31 data]
               054,                            [channel 32 data]
               054,                            [channel 33 data]
               080,                            [channel 34 data]
               080,                            [channel 35 data]
               080,                            [channel 36 data]
               080,                            [channel 37 data]
               093,                            [channel 38 data]
               093,                            [channel 39 data]
               093,                            [channel 40 data]
               093,                            [channel 41 data]
               080,                            [channel 42 data]
               080,                            [channel 43 data]
               080,                            [channel 44 data]
               080,                            [channel 45 data]
               071,                            [channel 46 data]
               071,                            [channel 47 data]
               071,                            [channel 48 data]
               071,                            [channel 49 data]
               063,                            [channel 50 data]
               063,                            [channel 51 data]
               063,                            [channel 52 data]
CELL-DYN® 1800 System                                          Interface Specification
07H79-01—February 2003                                                 Page 23 of 36
                           HISTOGRAM MESSAGE EXAMPLE
                                   (Continued)
                    063,                         [channel 53 data]
                    041,                         [channel 54 data]
                    041,                         [channel 55 data]
                    041,                         [channel 56 data]
                    041,                         [channel 57 data]
                    032,                         [channel 58 data]
                    041,                         [channel 59 data]
                    029,                         [channel 60 data]
                    029,                         [channel 61 data]
                    029,                         [channel 62 data]
                    029,                         [channel 63 data]
                    029,                         [channel 64 data]
                    027,                         [channel 65 data]
                    027,                         [channel 66 data]
                    027,                         [channel 67 data]
                    027,                         [channel 68 data]
                    027,                         [channel 69 data]
                    034,                         [channel 70 data]
                    034,                         [channel 71 data]
                    034,                         [channel 72 data]
                    034,                         [channel 73 data]
                    034,                         [channel 74 data]
                    039,                         [channel 75 data]
                    039,                         [channel 76 data]
Interface Specification                                      CELL-DYN® 1800 System
Page 24 of 36                                               07H79-01—February 2003
                         HISTOGRAM MESSAGE EXAMPLE
                                 (Continued)
               039,                            [channel 77 data]
               039,                            [channel 78 data]
               037,                            [channel 79 data]
               037,                            [channel 80 data]
               037,                            [channel 81 data]
               037,                            [channel 82 data]
               037,                            [channel 83 data]
               037,                            [channel 84 data]
               037,                            [channel 85 data]
               041,                            [channel 86 data]
               041,                            [channel 87 data]
               041,                            [channel 88 data]
               041,                            [channel 89 data]
               046,                            [channel 90 data]
               046,                            [channel 91 data]
               046,                            [channel 92 data]
               046,                            [channel 93 data]
               046,                            [channel 94 data]
               046,                            [channel 95 data]
               046,                            [channel 96 data]
               046,                            [channel 97 data]
               046,                            [channel 98 data]
               046,                            [channel 99 data]
               046,                            [channel 100 data]
CELL-DYN® 1800 System                                          Interface Specification
07H79-01—February 2003                                                 Page 25 of 36
                           HISTOGRAM MESSAGE EXAMPLE
                                   (Continued)
                    046,                         [channel 101 data]
                    049,                         [channel 102 data]
                    046,                         [channel 103 data]
                    046,                         [channel 104 data]
                    046,                         [channel 105 data]
                    054,                         [channel 106 data]
                    054,                         [channel 107 data]
                    054,                         [channel 108 data]
                    054,                         [channel 109 data]
                    058,                         [channel 110 data]
                    054,                         [channel 111 data]
                    054,                         [channel 112 data]
                    054,                         [channel 113 data]
                    063,                         [channel 114 data]
                    063,                         [channel 115 data]
                    063,                         [channel 116 data]
                    063,                         [channel 117 data]
                    066,                         [channel 118 data]
                    063,                         [channel 119 data]
                    063,                         [channel 120 data]
                    063,                         [channel 121 data]
                    063,                         [channel 122 data]
                    066,                         [channel 123 data]
                    066,                         [channel 124 data]
Interface Specification                                     CELL-DYN® 1800 System
Page 26 of 36                                              07H79-01—February 2003
                         HISTOGRAM MESSAGE EXAMPLE
                                 (Continued)
               066,                            [channel 125 data]
               066,                            [channel 126 data]
               066,                            [channel 127 data]
               066,                            [channel 128 data]
               066,                            [channel 129 data]
               066,                            [channel 130 data]
               068,                            [channel 131 data]
               068,                            [channel 132 data]
               068,                            [channel 133 data]
               068,                            [channel 134 data]
               073,                            [channel 135 data]
               068,                            [channel 136 data]
               068,                            [channel 137 data]
               068,                            [channel 138 data]
               068,                            [channel 139 data]
               078,                            [channel 140 data]
               080,                            [channel 141 data]
               080,                            [channel 142 data]
               080,                            [channel 143 data]
               080,                            [channel 144 data]
               073,                            [channel 145 data]
               073,                            [channel 146 data]
               073,                            [channel 147 data]
               073,                            [channel 148 data]
CELL-DYN® 1800 System                                          Interface Specification
07H79-01—February 2003                                                 Page 27 of 36
                           HISTOGRAM MESSAGE EXAMPLE
                                   (Continued)
                    073,                         [channel 149 data]
                    073,                         [channel 150 data]
                    075,                         [channel 151 data]
                    083,                         [channel 152 data]
                    083,                         [channel 153 data]
                    083,                         [channel 154 data]
                    083,                         [channel 155 data]
                    083,                         [channel 156 data]
                    083,                         [channel 157 data]
                    083,                         [channel 158 data]
                    083,                         [channel 159 data]
                    088,                         [channel 160 data]
                    090,                         [channel 161 data]
                    090,                         [channel 162 data]
                    090,                         [channel 163 data]
                    090,                         [channel 164 data]
                    090,                         [channel 165 data]
                    090,                         [channel 166 data]
                    090,                         [channel 167 data]
                    090,                         [channel 168 data]
                    090,                         [channel 169 data]
                    095,                         [channel 170 data]
                    095,                         [channel 171 data]
                    095,                         [channel 172 data]
Interface Specification                                     CELL-DYN® 1800 System
Page 28 of 36                                              07H79-01—February 2003
                         HISTOGRAM MESSAGE EXAMPLE
                                 (Continued)
               095,                            [channel 173 data]
               110,                            [channel 174 data]
               095,                            [channel 175 data]
               095,                            [channel 176 data]
               095,                            [channel 177 data]
               110,                            [channel 178 data]
               110,                            [channel 179 data]
               110,                            [channel 180 data]
               110,                            [channel 181 data]
               110,                            [channel 182 data]
               105,                            [channel 183 data]
               110,                            [channel 184 data]
               110,                            [channel 185 data]
               110,                            [channel 186 data]
               100,                            [channel 187 data]
               100,                            [channel 188 data]
               100,                            [channel 189 data]
               100,                            [channel 190 data]
               100,                            [channel 191 data]
               100,                            [channel 192 data]
               093,                            [channel 193 data]
               085,                            [channel 194 data]
               085,                            [channel 195 data]
               085,                            [channel 196 data]
CELL-DYN® 1800 System                                          Interface Specification
07H79-01—February 2003                                                 Page 29 of 36
                           HISTOGRAM MESSAGE EXAMPLE
                                   (Continued)
                    085,                         [channel 197 data]
                    085,                         [channel 198 data]
                    078,                         [channel 199 data]
                    078,                         [channel 200 data]
                    085,                         [channel 201 data]
                    085,                         [channel 202 data]
                    085,                         [channel 203 data]
                    085,                         [channel 204 data]
                    085,                         [channel 205 data]
                    085,                         [channel 206 data]
                    085,                         [channel 207 data]
                    083,                         [channel 208 data]
                    083,                         [channel 209 data]
                    083,                         [channel 210 data]
                    083,                         [channel 211 data]
                    073,                         [channel 212 data]
                    083,                         [channel 213 data]
                    083,                         [channel 214 data]
                    083,                         [channel 215 data]
                    083,                         [channel 216 data]
                    083,                         [channel 217 data]
                    083,                         [channel 218 data]
                    083,                         [channel 219 data]
                    083,                         [channel 220 data]
Interface Specification                                     CELL-DYN® 1800 System
Page 30 of 36                                              07H79-01—February 2003
                         HISTOGRAM MESSAGE EXAMPLE
                                 (Continued)
               066,                            [channel 221 data]
               054,                            [channel 222 data]
               054,                            [channel 223 data]
               054,                            [channel 224 data]
               054,                            [channel 225 data]
               054,                            [channel 226 data]
               054,                            [channel 227 data]
               051,                            [channel 228 data]
               051,                            [channel 229 data]
               051,                            [channel 230 data]
               051,                            [channel 231 data]
               051,                            [channel 232 data]
               051,                            [channel 233 data]
               051,                            [channel 234 data]
               046,                            [channel 235 data]
               034,                            [channel 236 data]
               034,                            [channel 237 data]
               034,                            [channel 238 data]
               034,                            [channel 239 data]
               034,                            [channel 240 data]
               034,                            [channel 241 data]
               034,                            [channel 242 data]
               034,                            [channel 243 data]
               034,                            [channel 244 data]
CELL-DYN® 1800 System                                          Interface Specification
07H79-01—February 2003                                                 Page 31 of 36
                           HISTOGRAM MESSAGE EXAMPLE
                                   (Continued)
                    034,                         [channel 245 data]
                    034,                         [channel 246 data]
                    034,                         [channel 247 data]
                    027,                         [channel 248 data]
                    027,                         [channel 249 data]
                    027,                         [channel 250 data]
                    027,                         [channel 251 data]
                    027,                         [channel 252 data]
                    027,                         [channel 253 data]
                    000,                         [channel 254 data]
                    000,                         [channel 255 data]
                    000,                         [channel 256 data]       Field 273
                                                 [note comma after
                                                  channel 256 but not
                                                  after the checksum]
                    E7                           [checksum]
                                                  NOTE: Sample checksum value may
                                                  not match this record’s data.
                    ETX                          [end of text]
Interface Specification                                       CELL-DYN® 1800 System
Page 32 of 36                                                07H79-01—February 2003
                           COUNT DATA MESSAGE EXAMPLE
           STX                                   [start of text]            Field 1
           "WBC",                                [message type]
           " CD1800",                            [instrument type]
           "– – – – – – – – – – – –",            [serial #]
           1040,                                 [sequence no.]
           00,                                   [specimen type]            Field 5
           "123",                                [operator ID]
           "08/31/03",                           [specimen date]
           "12:34",                              [specimen time]
           "123456789ABCDEFG",                   [specimen ID]
           "        JOHN DOE",                   [specimen name]            Field 10
           "M",                                  [specimen sex]
           "10/15/46",                           [specimen DOB]
           "        DR JANE DOE",                [doctor name]
           "08/30",                              [collect date]
           "09:50",                              [collect time]             Field 15
           "        sample 1",                   [comment]                  Field 16
           00083,                                [WBC count]                Field 17
           – – – – –,                            [spare field]
           00015,                                [LYM count]
           00006,                                [MID count]                Field 20
           00062,                                [GRAN count]
           – – – – –,                            [spare field]
           00476,                                [RBC count]                Field 23
CELL-DYN® 1800 System                                            Interface Specification
07H79-01—February 2003                                                   Page 33 of 36
                             COUNT DATA MESSAGE EXAMPLE
                                     (continued)
                00166,                             [HGB value]         Field 24
                00447,                             [HCT value]
                00940,                             [MCV value]
                00349,                             [MCH value]
                00371,                             [MCHC value]
                00118,                             [RDW value]
                00230,                             [PLT value]         Field 30
                00084,                             [MPV value]
                00019,                             [PCT value]
                00174,                             [PDW value]
                – – – – –,                         [spare field]
                00178,                             [LYM% value]        Field 35
                00077,                             [MID% value]
                00745,                             [GRAN% value]
                – – – – –,                         [spare field]
                –,                                 [spare field]
                –,                                 [spare field]       Field 40
                0,                                 [R4 WBC flag]
                0,                                 [GR3 WBC flag]
                0,                                 [MR3 WBC flag]
                1,                                 [MR2 WBC flag]
                0,                                 [LR2 WBC flag]      Field 45
                0,                                 [R1 WBC flag]
                1,                                 [R0 WBC flag]
Interface Specification                                    CELL-DYN® 1800 System
Page 34 of 36                                             07H79-01—February 2003
                         COUNT DATA MESSAGE EXAMPLE
                                  (continued)
           –,                                   [spare field]              Field 48
           –,                                   [spare field]              Field 49
           –,                                   [spare field]              Field 50
           –,                                   [spare field]
           –,                                   [spare field]
           –,                                   [spare field]
           –,                                   [spare field]
           –,                                   [spare field]              Field 55
           –,                                   [spare field]
           –,                                   [spare field]
           –,                                   [spare field]
           –,                                   [spare field]
           –,                                   [spare field]              Field 60
           00,                                  [LRI flag]
           00,                                  [URI flag]
           04862,                               [WBC lower meniscus time]
           02363,                               [WBC upper meniscus time]
           08107,                               [RBC lower meniscus time]
           07656,                               [RBC upper meniscus time]
           00000,                               [recount lower time]
           21845,                               [recount upper time]
           1,                                   [limits set]
           "O",                                 [sample mode]              Field 70
           0,                                   [RBC metering fault]
CELL-DYN® 1800 System                                           Interface Specification
07H79-01—February 2003                                                  Page 35 of 36
                          COUNT DATA MESSAGE EXAMPLE
                                   (continued)
                0,                               [WBC metering fault]
                0,                               [sample error/incomplete
                                                  aspiration]            Field 73
                1,                               [units set field]       Field 74
                9B                               [checksum]
                                                 NOTE: Sample checksum value
                                                 may not match this record’s data.
                ETX                              [end of text]
Interface Specification                                  CELL-DYN® 1800 System