Closed Suction – Suction Technique and Guidance
1. Looking towards the baby’s head, circle the colour in front of the calculated number
      (ie: where ET tube is cut add 5cm)
        To Determine Suction Depth: Observe in area directly from the lavage port connector
        Number on ET tube nearest the endotracheal adapter tip _____cm
        Add 5cm to the suction at ET tube tip              + _____cm
                                                        Total _____cm
SUCTION TECHNIQUE
   2. Add 10% to normal unit suction pressure
   3. When suctioning SUPPORT & HOLD the ETT & Y-adapter
   4. PASS the catheter down the ET tube 1-2cm a time until you see the appropriate
      colour (as calculated above) in the OBSERVATION AREA
   5. Depress the thumb valve and count for 1 second to transfer the suction to the end of
      the suction catheter - if your baby will tolerate this
   6. Slowly pull back the suction catheter keeping the SUCTION ON
   7. Pull back until you see the BLACK MARK in the RINSE CHAMBER
To Rinse / Clean the Suction Catheter
    1. Attach syringe to the saline flush port
    2. FIRST DEPRESS the thumb valve to TURN ON SUCTION and the slowly introduce the
       saline ensuring your suction is on at all times. This should be done after each
       suctioning procedure
    3. When you have finished, disconnect the syringe and replace flush port cap
To Lavage
    1. Attach syringe to saline flush port
    2. Gently HOLD the Y-adapter & ET tube as VERTICAL as possible
    3. INTRODUCE the SALINE in 0.1 - 0.2ml increments
    4. REMEMBER: When 1st setting up you will need to prime the flush port with 0.5ml
                       Closed Suction Catheter Size Selection
       Size (fr)              ET tube (mm)         Closed Suction           Open Suction
           5                       2.5                  Grey                   Grey
           6                       3.0                 Green                   Green
           7                       3.5                 Yellow                 Purple
           8                       4.0                  Blue                    Blue
          10                       5.0                  Black                  Black
                                Y-Adapter Size Selection
Each ‘Y’ Adapter has a number printed on it.Match the number on the ‘Y’ to the size of the
ET tube or half a size smaller.
Eg:
                           ET Tube                        Size of ‘Y’
                           2.5 mm                          2.5 / 2.0
                           3.0mm                           3.0 / 2.5
                           3.5mm                           3.5 / 3.0
                           4.0mm                           4.0/ 3.5
                                 Trouble Shooting Guide
Ventilator displays leak                 Check all connections are tightly in place and secure
Catheter bag has blown up                Remove the catheter from the ‘Y’adpter and
                                         squeeze air out of the catheter bag and replace
                                         immediately
Catheter is not suctioning               Pull catheter out of airway until black marker is in
                                         the rinse chamber.
                                         With the suction on, flush the catheter with saline
                                         until cleared.
                                         Secreation plugs on the tip can be removed by
                                         disconnecting the catheter and wiping with a sterile
                                         wipe.
                                         Recconect.