0% found this document useful (0 votes)
401 views4 pages

XT110C - XT111C - XT110D - XT111D: Single Stage Digital Controller With Multi Probe Input

This document provides instructions for installing and operating a single stage digital controller with multi-probe input. It describes how to use the LED indicators, view and change the setpoint, enter programming menus, and change parameter values. The programming menus ("Pr1" and locked "Pr2") contain configuration parameters, and parameters can be moved between the menus. Security code "321" is required to access the locked "Pr2" menu.

Uploaded by

HoàngViệtAnh
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
401 views4 pages

XT110C - XT111C - XT110D - XT111D: Single Stage Digital Controller With Multi Probe Input

This document provides instructions for installing and operating a single stage digital controller with multi-probe input. It describes how to use the LED indicators, view and change the setpoint, enter programming menus, and change parameter values. The programming menus ("Pr1" and locked "Pr2") contain configuration parameters, and parameters can be moved between the menus. Security code "321" is required to access the locked "Pr2" menu.

Uploaded by

HoàngViệtAnh
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

 ,QVWDOOLQJ DQG 2SHUDWLQJ ,QVWUXFWLRQV 1592002211

4.1 USE OF LEDS


XT110C - XT111C - XT110D - XT111D A series of light points on the front panels is used to monitor the loads controlled by the instrument. Each
LED function is described in the following table.
Single Stage Digital Controller
with Multi Probe Input LED MODE FUNCTION
ON Output relay enabled
LED1 Flashing - Programming Phase (flashing with LED2)
1. GENERAL WARNING LED2 Flashing - Programming Phase (flashing with LED1)
E.S. ON Energy saving activated by digital input
1.1 PLEASE READ BEFORE USING THIS MANUAL ON - ALARM signal
• This manual is part of the product and should be kept near the instrument for easy and quick reference. - In “Pr2” indicates the parameter is also present in “Pr1”
• The instrument shall not be used for purposes different from those described hereunder. It cannot be used as
a safety device. 4.2 TO SEE THE SETPOINT
• Check the application limits before proceeding. 1. Push and release the SET key to see the Set point value;

1.2 SAFETY PRECAUTIONS 2. To come back to the normal display push again the SET key or wait 10s.
• Check the supply voltage is correct before connecting the instrument.
• Do not expose to water or moisture: use the controller only within the operating limits avoiding sudden 4.3 TO CHANGE THE SETPOINT
temperature changes with high atmospheric humidity to prevent formation of condensation 1. Hold pushed the SET key for 3s to change the Set point value;
• Warning: disconnect all electrical connections before any kind of maintenance.

9 8
The instrument must not be opened. 2. The value of the set point will be displayed and the LED1 & 2 start blinking;
• In case of failure or faulty operation send the instrument back to the distributor or to “DIXELL S.p.A.” (see 3. To change the Set value push the or arrows within 10s.
address) with a detailed description of the fault. 4. To memorise the new set point value push the SET key again or wait 10s.
• Consider the maximum current which can be applied to each relay (see Technical Data).
• Ensure that the wires for probes, loads and the power supply are separated and far enough from each other, 4.4 TO ENTER THE PARAMETERS LIST “PR1
without crossing or intertwining.
8
To enter the parameter list “Pr1” (user accessible parameters) operate as follows:
• In case of applications in industrial environments, the use of mains filters (our mod. FT1) in parallel with
inductive loads could be useful. 1. Push for 3s the SET + keys (LED1 & 2 start blinking).
2. The controller will display the first parameter present in the Pr1 menu..
2. GENERAL DESCRIPTION
The XT110C, XT111C and XT110D, XT111D (DIN RAIL format) are single-stage ON/OFF controllers for
temperature, humidity and pressure applications with direct or inverse action, user-selectable. The
analogue input type can be set by parameter between the following, according to the model: 4.5 TO ENTER THE PARAMETERS LIST “PR2”
- PTC, NTC; The “Pr2” parameter list contains the configuration parameters. A security code is required to enter it.
- PTC, NTC, Pt100, Thermocouple J, K, S;
- 4÷20mA, 0÷1V, 0÷10V. 1. Enter the “Pr1” level, see above paragraph.
2. Select “Pr2” parameter and press the “SET” key.
3. FIRST INSTALLATION
9 8
3. The “PAS” flashing message is displayed, shortly followed by “0 - -” with a flashing zero.
4. Use or to input the security code in the flashing digit; confirm the figure by pressing “SET”.
3.1 PROBE SETTING
The security code is “321”.
The pre-set probe type is written on the label of the instrument,
see picture. If it is different from the probe that has be used, set 5. If the security code is correct the access to “Pr2” is enabled by pressing “SET” on the last digit.
the probe following procedure below

8 keys together for 3s: the Pr2 menu


Another possibility is the following:
After switching ON the instrument, within 30 seconds, push SET +
will be entered.
8
3.1.1 How to set the probe.
1. Enter the programming menu by pressing the SET+ for 3s.
2. Select the Pbc (Probe configuration) parameter and push the SET key. 4.6 HOW TO MOVE A PARAMETER FROM THE “PR2” MENU TO “PR1” AND
3. Set the kind of probe: VICEVERSA.
a. Controller for temperature: Pt= Pt100, J = J thermocouple, c = K thermocouple, S = S
Each parameter present in “Pr2” MENU can be removed or put into “Pr1”, user level, by pressing “SET +
b.
thermocouple; Ptc = PTC; ntc = ntc.
Controller with current or voltage inputs: cur=4÷20mA, 0-1= 0÷1V, 10= 0÷10V 8”. In “Pr2” when a parameter is present in “Pr1” the LED is on.
4. Push the SET key to confirm it.
5. Switch the controller off and on again. 4.7 HOW TO CHANGE A PARAMETER
To change a parameter value operates as follows:
NOTE: Before proceeding check and, if necessary; set with appropriate values the Minimum Set Points 1. Enter the Programming mode
(LS1 e LS2) and Maximum Set Points (US1 e US2). See also the paragraphs concerning the 2. Select the required parameter.
programming. 3. Press the “SET” key to display its value.
4. Use “UP” or “DOWN” to change its value.
5. Press “SET” to store the new value and move to the following parameter.
4. FRONT PANEL COMMANDS
TO EXIT: Press SET + UP or wait 15s without pressing a key.
NOTE: the set value is stored even when the procedure is exited by waiting the time-out to expire.

SET: To display and modify target set point; 4.8 HOW TO LOCK THE KEYBOARD
in programming mode it selects a
parameter or confirm an operation.
TO SWITCH THE INSTRUMENT
ON/OFF: If the function is enabled (par.
onF=yES), by pressing the SET key for
1. Keep pressed for more than 3 s the the 9 and 8keys.
2. The “POF” message will be displayed and the keyboard will be locked. At this point it will be possible
more than 4s the controller is switched only to see the set point or the MAX o Min temperature stored
OFF. To switch the instrument on again
3. If a key is pressed more than 3s the “POF” message will be displayed.
9
press the SET key.
LE D2 LE D1
E.S . UP: in programming mode it browses the
4.9 TO UNLOCK THE KEYBOARD
9 and 8 keys, till the “Pon” message will be displayed.
parameter codes or increases the
displayed value. Hold it pressed for a Keep pressed together for more than 3s the the

8
faster change
DOWN: in programming mode it browses 4.10 ON/OFF FUNCTION
the parameter codes or decreases the TO SWITCH THE INSTRUMENT ON/OFF: If the function is enabled (par. onF=yES), by pressing the
displayed value. Hold it pressed for a SET key for more than 4s the controller is switched OFF. To switch the instrument on again press the
faster change SET key.

5. PROBES AND MEASURING RANGE


Probe Down Scale Full Scale
KEY COMBINATIONS: NTC -40°C/-40°F 110°C / 230 °F
9 8+ To lock & unlock the keyboard.
PTC -50°C / -58°F 150°C / 302°F

8 Pt100 -200°C / -328°F 600°C / 1112°F

9
SET + To enter in programming mode. TcK 0°C / 32°F 1300°C / 1999°F
SET + To return to the room temperature display. TcJ 0°C / 32°F 600°C / 1112°F
TcS 0°C / 32°F 1400°C / 1999°F

1592002211 XT110-111C-D GB R.1.0 15.12.2004.doc ;7& ² ;7&  ;7' ² ;7' 1/4


 ,QVWDOOLQJ DQG 2SHUDWLQJ ,QVWUXFWLRQV 1592002211

6. LIST OF PARAMETERS Ptb Parameters table: (read only) Shows the code of the parameters map.
rEL Software release: (read only)
REGULATION Pr2 To access the Pr2 parameter programming menu.
Hy1 Differential: (-Full Sc. / Full Sc.) Intervention differential for set point. It can be set with positive value
or with negative value. The kind of action (direct or inverse) depends on the S1C parameter (in or
di). 7. INSTALLATION AND MOUNTING
LS1 Minimum set point: (Down Sc.÷ Set) Sets the minimum acceptable value for the set point. Instrument XT110C and XT111C shall be mounted on vertical panel, in
US1 Maximum set point: (Set÷ Full Sc.) Sets the maximum acceptable value for set point. a 29x71 mm hole, and fixed using the special brackets supplied.
S1C Action type: S1C=in inverse action (heating/ humidifying /increase pressure); S1C=dir direct Instruments XT110D, XT111D, shall be mounted on an omega DIN rail
action (cooling / dehumidifying /decrease pressure). (3). To obtain an IP65 protection grade use the front panel rubber
AC Anti-short cycle delay: (0÷250 sec) Minimum time between the switching off and the following gasket (mod. RG-C) as shown in figure.
switching on
The temperature range allowed for correct operation is 0÷60 °C. Avoid
on Minimum time a stage stays switched ON (0÷250 sec)
places subject to strong vibrations, corrosive gases, excessive dirt or
ono: Minimum time between 2 following switching ON of the same load (0÷120 min).
humidity. The same recommendations apply to probes. Let air circulate
ALARMS by the cooling holes.
ALC Temperature alarms configuration: it determines if alarms are relative to set point or referred to
absolute values.
rE relative to set point; Ab absolute temperature 8. ELECTRICAL CONNECTIONS
ALL Minimum alarm:
The instruments are provided with screw terminal block to connect cables with a cross section up to 2,5
with ALC=rE: relative to set point, (0÷|Down Sc.-Set|) this value is subtracted from the set point.
mm2. Before connecting cables make sure the power supply complies with the instrument’s requirements.
The alarm signal is enabled when the probe values goes below the “SET-ALL” value.
Separate the input connection cables from the power supply cables, from the outputs and the power
with ALC=Ab absolute value, minimum alarm is enabled when the probe values goes below the
connections. Do not exceed the maximum current allowed on each relay, in case of heavier loads use a
“ALL” value.
suitable external relay.
ALU Maximum alarm:
with ALC=rE: alarm relative to set point, (0÷|Full Sc.-Set|) Maximum alarm is enabled when the
probe values exceeds the “SET+ALU” value.
9. SERIAL CONNECTIONS
with ALC=Ab: absolute alarm, (Set÷Full Sc.) Maximum alarm is enabled when the probe values All models can be connected to the monitoring and supervising system XJ500 using the serial port. The
exceeds the “ALU” value. external XJ485 serial module to interface the instrument with the monitoring and supervising system
ALH Differential for alarm recovery: (0,1÷Full scale) the alarm recovers when probe value is higher XJ500 is required.
than Alarm value + ALH. The standard ModBus RTU protocol it is used.
ALd Alarm delay:(0÷999 min) time interval between the detection of an alarm condition and alarm NOTE: XT110C or XT111C instruments with current or voltage input and 230V or 115V supply,
signalling. cannot be connected to the XJ485 serial module.
dAO Delay of alarm at start-up: (0÷23.5h) time interval between the detection of the alarm condition
after instrument power on and alarm signalling. 10. HOW TO USE THE HOT KEY
So1 Relay status with faulty probe: So1=oFF open; So1=on closed.
tbA Status of alarm relay after pushing a key. (XT111C/ XT111D only): oFF = relay disabled; on = 10.1 HOW TO PROGRAM A HOT KEY FROM THE INSTRUMENT (UPLOAD)
relay enabled.
9
AS Alarm relay configuration (XT111C/ XT111D only): cL = 4-6 terminals open with alarm; oP = 4-6 1. Program one controller with the front keypad.
terminals closed with alarm. 2. When the controller is ON, insert the “Hot key” and push key; the "uPL" message appears
followed a by flashing “End”
PROBES AND DISPLAY
3. Push “SET” key and the End will stop flashing.
LCI Start of scale, only with current or voltage input: (with rES = in, dE, cE: -99.00÷199.00, with 4. Turn OFF the instrument remove the “Hot Key”, then turn it ON again.
rES=irE -999÷1999) Adjustment of read out corresponding to 4mA or 0V input signal.
UCI End of scale, only with current or voltage input (with rES = in, dE, cE: -99.00÷199.00, with
rES=irE -999÷1999) Adjustment of read out corresponding to 20mA or 1V or 10V input signal.
NOTE: the “Err” message is displayed for failed programming. In this case push again 9 key if you want
to restart the upload again or remove the “Hot key” to abort the operation.
oPb Probe calibration: (-999÷999) allows to adjust possible offset of the probe.
rES Resolution: select the resolution of the controller.
in= integer (-99÷199);
10.2 HOW TO PROGRAM AN INSTRUMENT USING A HOT KEY (DOWNLOAD)
dEC= 1 decimal point (-99.0÷199.0), 1. Turn OFF the instrument.
cE = 2 digits after the decimal point (-99.00÷199.00) only for current or voltage input, 2. Insert a programmed “Hot Key” into the 5 PIN receptacle and then turn the Controller ON.
irE = integer, large scale (–999÷1999) only for current or voltage input. 3. Automatically the parameter list of the “Hot Key” is downloaded into the Controller memory, the
WARNING: if rES is changed from “irE” to another value, all the parameters values expressed in “doL” message is blinking followed a by flashing “End”.
degrees: SET, Hy1, LS1, uS1, ALL, ALu, ALH, LCi, uCi, LAo, uAo, HES, have to be checked and 4. After 10 seconds the instrument will restart working with the new parameters.
set. 5. Remove the “Hot Key”..
NOTE: the decimal point selection is not available on models with thermocouple input.
UdM Measurement unit: it depends on models: NOTE the message “Err” is displayed for failed programming. In this case turn the unit off and then on if
for temperature:°C = Celsius; °F = Fahrenheit. you want to restart the download again or remove the “Hot key” to abort the operation.
with 4÷20mA, 0÷1V, 0÷10V input : 0= °C; 1= °F, 2= %RH, 3=bar, 4=PSI, 5=no measurement
unit.
PbC Probe selection: it sets the kind of probe. It depends on models 11. DIGITAL INPUT
for temperature NTC/PTC: Ptc = PTC; ntc = ntc. The controllers have 1 free contact digital input. It is programmable in 5 different configurations by the
for temperature standard: Pt= Pt100, J = J thermocouple, c = K thermocouple, S = S “i1F” parameter.
thermocouple; Ptc = PTC; ntc = ntc.
with 4÷20mA, 0÷1V, 0÷10V input : cur=4÷20mA, 0-1= 0÷1V, 10= 0÷10V. 11.1 INVERT THE KIND OF ACTION: HEATING-COOLING (I1F = C-H)
P3F Third wire presence for Pt100 probe: for using 2 or 3 wires Pt100 probes: no = 2 wires probe;
yES = 3 wires probe. This function allows to invert the regulation of the controller: from direct to inverse and viceversa.
ANALOG OUPUT – ONLY FOR XT110D, XT111D - OPTIONAL
11.2 REMOTE ON/OFF (I1F = OFF)
AOC Analog output configuration (only for models with analog output):
AOC=Pb Probe reading. The analog output parameters LAO and UAO are independent and This function allows to switch ON and OFF the instrument.
correspond to the absolute read-out probe signal.
AOC=Er Probe-Set Point. the analog output parameters LAO and UAO are related to the Set. 11.3 GENERIC ALARM (I1F = EAL)
LAO Lower analog output limit: (only for models with analog output) minimum value of temperature As soon as the digital input is activated the unit will wait for “did” time delay before signalling the “EAL”
associated to the 4mA analog output. This value can be absolute or relative to the Set Point by alarm message. The outputs status don’t change. The alarm stops just after the digital input is de-
setting the AOC parameter. activated.
UAO Upper analog output limit: (only for models with analog output) maximum value of
temperature associated to the 20mA analog output. This value can be absolute or relative to the Set 11.4 SERIOUS ALARM MODE (I1F = BAL)
Point by setting the AOC parameter.
SAO Analog output safety with probe fault (only for models with analog output): determines what When the digital input is activated, the unit will wait for “did” delay before signalling the “bAL” alarm
state the analog output should assume when the probe is faulty: message. The relay outputs are switched OFF. The alarm will stop as soon as the digital input is de-
SAO = oFF; analog output = 40mA. SAO = on; analog output = 20mA activated.
DIGITAL INPUT
11.5 ENERGY SAVING (I1F = HES)
HES Set point change during during the Energy Saving cycle : (Down Sc./Full Sc.) sets the variation
of the set point during the Energy Saving cycle. The Energy Saving function allows to change the set point value as the result of the SET+ HES
i1F Digital input operating mode: configure the digital input function: c-H = to invert the kind of (parameter) sum. This function is enabled until the digital input is activated.
action: direct - reverse;
oFF = to switch the controller off.; AUS = Not used; HES = Energy Saving; EAL = generic external 12. ALARM SIGNALS
alarm; bAL = serious external alarm: it switches off the loads. Message Cause Outputs
i1P Digital input polarity: “PFo” Probe broken or absence Alarm output ON; Output according to parameter “So1”
CL : the digital input is activated by closing the contact; “PFc” Probe short circuited Alarm output ON; Output according to parameter “So1”
OP : the digital input is activated by opening the contact
“HA” Maximum alarm Alarm output ON; Other outputs unchanged.
did Digital input alarm delay: (0÷255 min) delay between the detection of the external alarm condition
“LA” Minimum alarm Alarm output ON; Other outputs unchanged.
(i1F= EAL or i1F = bAL) and its signalling.
“EAL” External alarm Output unchanged.
OTHER
“bAL” Serious external alarm Output OFF.
Adr RS485 serial address (0÷247) identifies the instrument within a control or supervising system.
onF Swithching ON/OFF enabling from keyboard: (no = disabled; yES=enabled) It permits the
switching ON/OFF of the instrument by pressing the SET key for more than 4s.
1592002211 XT110-111C-D GB R.1.0 15.12.2004.doc ;7& ² ;7&  ;7' ² ;7' 2/4
 ,QVWDOOLQJ DQG 2SHUDWLQJ ,QVWUXFWLRQV 1592002211

12.1 ALARM RELAY STATUS 14.3 XT111C – 12VAC/DC OR 24VAC/DC


Input: 4÷ 20mA
Status of the instrument XT111C XT111D 0÷ 1V / 0÷ 10V
AS = CL AS= oP AS = CL AS= oP
Instrument off 4-6 closed 4-6 closed 20-21 closed 20-21 closed
Normal operating 4-6 closed 4-6 open 20-21 closed 20-21 open
   
Alarm present 4-6 open 4-6 closed 20-21 open 20-21 closed
PTC / NTC

12.2 SILENCING BUZZER / ALARM RELAY OUTPUT    


8(3)A250V 8(3)A250V
Once the alarm signal is detected the buzzer, if present, can be disabled by pressing any key.
XT111C/XT111D: the alarm relay status depends on the tbA parameter: with tbA=yES the relay is
disabled by pressing any key, with tbA=no the alarm relay remains enabled as long as the alarm lasts.
The display signal remains as long as the alarm condition remains.
       
Hot-KEY
N.O. ALARM
12.3 ALARM RECOVERY
Probe alarms “PFo”, “PFc” start few seconds after the fault in the probe; they automatically stop few Line
seconds after the probe restarts normal operation. Check connections before replacing the probe. Probe: Pt100= 7 – 9 (8); Thermocouple J, K, S = 7(+); 9(-)
Max. and min. alarms “HA” and “LA” automatically stop as soon as the variable returns to normal values. 24Vac/cd supply: 11-12
Alarms “bAL” and “EAL” recover as soon as the digital input is disabled.

13. TECHNICAL DATA 14.4 XT111C – 230V AC OR 115V AC


Housing: self extinguishing ABS. PTC / NTC
Case: XT110C, XT111C:frontal 32x74 mm; depth 60mm;
XT110D, XT111D: 4 DIN modules 70x85 mm; depth 61mm.    
Mounting: XT110C, XT111C panel mounting in a 71x29 mm panel cut-out.
XT110D, XT111D: DIN RAIL Input 4÷ 20m A
Protection: IP20. 0÷ 1V / 0÷ 10V
Frontal protection: XT110C, XT111C IP65 with frontal gasket RG-C (optional).
Connections: Screw terminal block ≤ 2,5 mm2 heat-resistant wiring.    
Power supply: 12Vac/dc, ±10% or: 24Vac/dc ± 10% only for “C” format 8(3)A250V 8(3)A250V
or 230Vac ± 10%, 50/60Hz or 110Vac, ± 10%, 50/60Hz
Power absorption: 3VA max.
Display: 3 ½ digits, red LED        
Inputs: according to the order: NTC/PTC or NTC/PTC /Pt100 /Thermocouple J, K, S or 4÷20mA/ 0÷1V / Hot-KEY
0÷10V N.O. Alarm
Relay outputs: Load relay SPDT 8(3)A, 250Vac
Alarm: (XT111C/XT111D) relay SPDT 8(3)A, 250Vac Line
Other output: buzzer (optional) Probe: Pt100=9–11 (10); Thermocouple J, K, S= 9(+) - 11(-)
Kind of action: 1B; Pollution grade: normal, Software class: A; 115Vac supply: 7-8
Data storing: on the non-volatile memory (EEPROM);
Operating temperature: 0÷60 °C (32÷140°F); Storage temperature: -30÷85 °C (-22÷185°F).
Relative humidity: 20÷85% (no condensing) 14.5 XT110D – 230V AC OR 120V AC OR 24V AC
Measuring and regulation range: according to the probe; Input: 4÷ 20mA
Controller Accuracy a 25°C: better than ±0,5% of full scale 0÷ 1V / 0÷ 10V

14. CONNECTIONS Analog


Out   
14.1 XT110C – 12V AC/DC OR 24V AC/DC
PTC
In pu t: 4÷ 2 0m A NTC
0 ÷ 1 V / 0 ÷ 1 0V

   
Hot-KEY

8(3)A250V

PTC / NTC

   
8(3)A250V LOAD 1

Line
       
Hot-KEY
Probe: Pt100=11 - 10 (12); Thermocouple J, K, S= 11(+) - 10(-)
115Vac supply: 1-2; 24Vac supply: 1-2

Line 14.6 XT111D – 230V AC OR. 120V AC OR. 24V AC


Probe: Pt100= 7 – 9 (8); Thermocouple J, K, S = 7(+); 9(-) Input: 4÷ 20mA
24Vac/cd supply: 11-12 0÷ 1V / 0÷ 10V

14.2 XT110C – 230V AC OR 115V AC Analog   


Out
PT C / N TC

   
In pu t 4÷ 20 m A
Hot-KEY

0÷ 1V / 0 ÷ 1 0V
8(3)A250V 8(3)A250V

   
8(3)A250V

LOAD 1 Alarm
        Line
Hot-KEY
Probe: Pt100=11 - 10 (12); Thermocouple J, K, S= 11(+) - 10(-)
115Vac supply: 1-2; 24Vac supply: 1-2
Line
Pt100=9 –11 (10); Thermocouple J, K, S = 9(+) - 11(-)
115Vac supply: 7-8

1592002211 XT110-111C-D GB R.1.0 15.12.2004.doc ;7& ² ;7&  ;7' ² ;7' 3/4


 ,QVWDOOLQJ DQG 2SHUDWLQJ ,QVWUXFWLRQV 1592002211

15. DEFAULT SETTING VALUES


COD Name Range °C/°F Lev
Set Set point LS1÷US1 0/32 -
Hy1 Differential -Full Sc./ Full Sc. -1/-2 Pr1
LS1 Minimum set point Down Sc./ Set min Pr2
US1 Maximum set point Set/ Full Sc. max Pr2
S1C Action type output in= Inverse; dir=direct in Pr2
Ac Anti-short cycle delay: 0÷250 sec 0 Pr2
on Minimum time a stage stays switched ON 0÷250 sec 0 Pr2
ono Minimum time between 2 following switching ON 0÷120 min 0 Pr2
of the same load
ALC Alarm configuration rE=relat.; Ab= absolute rE Pr2
ALL Minumum alarm (ALC=rE) 0 ÷ |Start Sc.-Set| 10.0/ 20 Pr2
(ALC=Ab) Start Sc.÷ ALu
ALU Maximum alarm (ALC=rE) 0 ÷ |Full Sc.-Set|. 10.0/ 20 Pr2
(ALC=Ab) ALL÷ Full Scale
ALH Alarm recovery differential 0÷Full scale 2.0/4 Pr2
ALd Alarm delay 0÷999 min 15 Pr2
dAO Alarm delay at start up 0÷23h 50min 1.3 Pr2
So1 Output status with faulty pr. oFF=open on=closed oFF Pr2
tbA1 Alarm relay disabling no; yES yES Pr2
AS1 Alarm relay polarity CL÷oP oP Pr2
Lci2 Start scale with current or voltage input -1999÷1999 various Pr1
Uci2 End scale with current or voltage input -1999÷1999 various Pr1
OPb Probe calibration -Full Sc./ Full Sc. 0.0 Pr1
rES Resolution in=NO; dE=0,1; cE=0,01 in Pr2
UdM Measurement unit (temp.) °C=°C; °F= °F; various Pr1
(current/voltage) 0=°C; 1=°F; 2=RH;
3=bar; 4=PSI, 5=off
PbC Kind of probe Pt=Pt100; J=tcJ; c= tck; various Pr1
S=tcS; Ptc=PTC; ntc=
NTC; 0-1=0÷1V; 10=
0÷10V; cur=0÷20mA
P3F 3rd wire presence no=2 wires; no Pr2
yES=3 wires
Aoc3 Analog output setting Pb / Er Pb Pr2
LAo3 Lower analog output limit: Down Sc./ Full Sc. 0 Pr2
uAo3 Upper analog output limit: Down Sc./ Full Sc. 0 Pr2
SAo3 Analog output safety with probe fault oFF / on oFF Pr2
HES Energy saving differential Down Sc./ Full Sc. 0.0 Pr2
i1F Digital input configuration c-H / oFF / AuS / HES / EAL Pr2
EAL / bAL
i1P Digital input polarity cL=closed; oP=open cL Pr2
did Alarm delay for dig. input 0÷120m 0 Pr2
Adr Serial address 0÷247 1 Pr2
OnF oFF function enabling no=not enabled; no Pr2
yES=enabled
Ptb Parameter table Readable only -- Pr2
rEL Software release Readable only --- Pr2
Pr2 To access the Pr2 Readable only 321 Pr1

1 Only for XT111C/XT111D;


2 Only for instrument with 4÷20mA or 0÷1V or 0÷10V;
3 Only for instruments with analog ouput

Dixell S.p.A. Z.I. Via dell’Industria, 27


32010 Pieve d’Alpago (BL) ITALY
tel. +39 - 0437 - 98 33 - fax +39 - 0437 - 98 93 13
E-mail: dixell@dixell.com - http://www.dixell.com

1592002211 XT110-111C-D GB R.1.0 15.12.2004.doc ;7& ² ;7&  ;7' ² ;7' 4/4

You might also like