Sonotrax Basic
Sonotrax Basic
                        Statement
                        This manual will help you understand the operation and maintenance of the product better. It
                        is reminded that the product shall be used strictly complying with this manual. User’s
                        operation failing to comply with this manual may result in malfunction or accident for which
                        Edan Instruments, Inc. (hereinafter called EDAN) can not be held liable.
                        EDAN owns the copyrights of this manual. Without prior written consent of EDAN, any
                        materials contained in this manual shall not be photocopied, reproduced or translated into
                        other languages.
                        Materials protected by the copyright law, including but not limited to confidential information
                        such as technical information and patent information are contained in this manual, the user
                        shall not disclose such information to any irrelevant third party.
                        The user shall understand that nothing in this manual grants him, expressly or implicitly, any
                        right or license to use any of the intellectual properties of EDAN.
                        EDAN holds the rights to modify, update, and ultimately explain this manual.
                        Product Information
                        Product Name: Ultrasonic Pocket Doppler
                        Model: SONOTRAX Lite, SONOTRAX Basic, SONOTRAX Basic A, SONOTRAX Pro,
                               SONOTRAX II, SONOTRAX II Pro
II
III
IV
                               This unit is internally powered equipment, and it is an IEC/EN 60601-1 Type B applied
                               part. Type B protection means that the connection between the equipment and personnel
                               complies with permitted leakage currents and dielectric strength of IEC/EN 60601-1.
                      WARNING and CAUTION messages must be observed. To avoid the possibility of injury,
                      observe the following precautions during the operation of the device.
                                                                 WARNING
                      1    This device is not intended for treatment.
                      2    This device is not explosion-proof and can not be used in the presence of flammable
                           anaesthetics.
                      3    Do not touch the signal input/output connector and the patient simultaneously.
                      4    We recommend that exposure to ultrasound should be kept as low as reasonably
                           achievable. This is considered to be good practice and should be observed at all
                           time.
                      5    Only use probes provided by the manufacturer.
                      6    The stretching length of probe cable should be less than two meters in order to avoid
                           the cable breaking away from the probe socket.
-1-
                                                               WARNING
                      7    Check if the equipment has visible evidence of damage that may affect the
                           personnel’s safety or the device’s capability before use. If the damage is evident,
                           replace the defective part(s) before use.
                      8    Accessory equipment connected to the analog and digital interfaces must be
                           certified according to the respective IEC/EN standards (e.g. IEC/EN 60950 for data
                           processing equipment and IEC/EN 60601-1 for medical equipment). Furthermore all
                           configurations shall comply with the valid version of the system standard IEC/EN
                           60601-1-1. Anybody who connects additional equipment to the signal input
                           connector or signal output connector to configure a medical system must ensure that
                           the system complies with the requirements of the valid version of the system
                           standard IEC/EN 60601-1-1. If in doubt, consult our technical service department or
                           your local distributor.
                      9    The device shall only be used when the battery cover is closed.
                      10 Do not throw batteries in fire as this may cause explosion.
                      11 Do not attempt to charge normal alkaline batteries. They may leak, catch fire or even
                         explode.
                      12 Replacement of the battery shall be done at least 1.5 meters away from patients.
                      13 Charge the lithium-ion polymer battery with the special charger supplied by the
                         manufacturer.
                                                                CAUTION
                      1    Federal (U.S.) law restricts this device to sale by or on the order of a physician.
                      2    Refer servicing to qualified personnel.
                      3    The main unit is designed for continuous operation and is ‘ordinary’. Do not immerse
                           it in any liquid (i.e. not drip or splash-proof).
                      4    Electromagnetic Interference - Ensure that the environment in which the device is
                           operated is not subject to any source of strong electromagnetic interference, such as
                           radio transmitters, mobile telephones, etc.
                      5    SONOTRAX Series Ultrasonic Pocket Doppler is a tool to aid the healthcare
                           professional and should not be used in place of normal fetal monitoring.
                      6    Keep the device in a clean environment and avoid vibration during storage.
                      7    The device and accessories are to be disposed of according to local regulations after
                           their useful lives. Alternatively, they can be returned to the dealer or the
                           manufacturer for recycling or proper disposal. Batteries are hazardous waste. Do
                           NOT dispose them together with house-hold garbage. At the end of their life hand
                           the batteries over to the applicable collection points for the recycling of waste
                           batteries. For more detailed information about recycling of this product or battery,
                           please contact your local Civic Office, or the shop where you purchased the product.
-2-
1.3 Symbols
                                             The symbol indicates that the device complies with the European Council
                                             Directive 93/42/EEC concerning medical devices.
                                             The symbol indicates that the device should be sent to the special agencies
                                             according to local regulation for separate collection after its useful life.
Stand-by
CE mark
Class II equipment
Part Number
Serial Number
Date Of Manufacture
Manufacturer
Rx only (U.S.) Federal (U.S.) Law restricts this device to sale by or on the order of a physician
-3-
                      2.1 Features
                      There are six different models available: SONOTRAX Lite, SONOTRAX Basic, SONOTRAX
                      Basic A, SONOTRAX Pro, SONOTRAX II and SONOTRAX II Pro.
                      SONOTRAX Lite is for simple auscultation (intermittent listening). SONOTRAX Basic,
                      SONOTRAX Basic A, SONOTRAX Pro, SONOTRAX II and SONOTRAX II Pro not only
                      detect fetal heart sound; they also display the fetal heart rate on a LCD screen.
                      The features of the Dopplers are listed in the following chart:
                                          Model
                                                    SONOTRAX   SONOTRAX    SONOTRAX       SONOTRAX    SONOTRAX    SONOTRAX
                       Function                        Lite     Basic           Basic A     Pro           II        II Pro
LCD Display √ √ √ √ √
LCD Backlight √ √ √ √
Probe Detecting √ √ √ √ √ √
Probe Identifying √ √ √ √ √
Built-in Speaker √ √ √ √ √ √
Attached Earphone √ √ √ √ √ √
Volume Adjustable √ √ √ √ √ √
Modes Switching √ √ √ √ √
                       Rechargeable NI-MH
                                                           √      √               √          √
                       Battery (8.4V)
                       Lithium-ion     Polymer
                                                                                                          √           √
                       Battery
                       Battery Charged        in
                                                                                                          √           √
                       Machine
                       Low Battery Detecting
                                                           √      √               √          √            √           √
                       & indicating
Auto Shutdown √ √ √ √ √
Vascular Examining √ √ √ √ √ √
-4-
                      2.2 Appearance
                      NOTE:
                      The pictures and interfaces in this manual are for reference only.
-5-
                                                               Playing                          Communication
                                Working Mode
                              Battery Indicator
                                                                                                          Probe Type
                      Working Mode:
                      It indicates the mode in which the Doppler is working.
                      FHR Refresh Frequency:
                      When the examination starts, the heart shape symbol flashes, and its frequency indicates the FHR
                      refreshing frequency.
                      Battery Indicator:
                      For SONOTRAX Basic, a battery symbol appears on the LCD after it’s powered on. When the
                      battery is low, this symbol flashes; the Doppler will shut down automatically after 7 seconds.
                      For SONOTRAX Basic A, SONOTRAX Pro, SONOTRAX II and SONOTRAX II Pro, there
                      is a battery symbol on the LCD. The green panes in it indicate the battery electric energy. The
                      green panes disappear gradually with the energy consumption. When the electric energy is low,
                      the empty battery symbol flashes. Approximately five minutes later, the Doppler shuts down
                      automatically.
                      PLAYING:
                      PLAYING is lit when the Doppler is playing the recorded fetal heart sound.
                      RECORDING:
                      RECORDING is lit when the Doppler is recording fetal heart sound.
-6-
                      FHR Displaying:
                      This area displays the fetal heart rate value, and its unit is bpm (beat per minute).
                      Probe Type:
                      The Doppler can detect and identify the probe: when the probe is not connected or poorly
                      connected, the “---” symbol and “MHz” symbol on the LCD screen flash; when the probe is well
                      connected, the flashing stops, and the probe type is shown on the LCD screen.
                      Communication
                      The Doppler can communicate with PC. This function is reserved.
                      2.4 Buttons
                      At most there are four push buttons (Power, MODE, START/STOP and REC/PLAY) and a
                      volume control button on the main unit of the Doppler. Their primary functions are as follows:
Function: Start/ stop examining (Mode 3)/ setting (Mode 4 and Mode 5).
                      (4) REC/PLAY
                          (Only for SONOTRAX Pro/ SONOTRAX II Pro)
-7-
                      2.5 Socket
                      The two sockets are located on the top panel of the Doppler.
(1) Earphone socket : the earphone or line-in cable connects to the Doppler via this socket.
                      Signal Interface
                                            1       GND            2 Signal
                              1
                         4                  3       Signal         4 Signal
                                  2
                                      3
                         5                  5       Signal
                      (2) Charge socket                : the special lithium-ion polymer battery charger connects to the
                                                        Doppler via this socket. (For SONOTRAX II and SONOTRAX II Pro
                                                           only)
                      NOTE:
                      Accessory equipment connected to the analog and digital interfaces must be certified
                      according to the respective IEC/EN standards (e.g. IEC/EN 60950 for data processing
                      equipment and IEC/EN 60601-1 for medical equipment). Furthermore all configurations
                      shall comply with the valid version of the system standard IEC/EN 60601-1-1. Anybody
                      who connects additional equipment to the signal input connector or signal output
                      connector to configure a medical system must ensure that the system complies with the
                      requirements of the valid version of the system standard IEC/EN 60601-1-1. If in doubt,
                      consult our technical service department or your local distributor.
-8-
2.6 Probes
-9-
                                                                        Jack                    Definition
                                                                        1                       Power Supply
                                                                        2                       Signal
                                                                        3                       Probe Coding 1
                                                                        4                       Probe Coding 2
                                                                        5                       GND
                                                                        6                       (Shell) GND
- 10 -
                                                                       CAUTION
                      1    Do not try to connect any other plug to the probe socket except the plug of the
                           probes mentioned above.
                      2    Do not stretch the probe cable for more than two meters long.
                      2.7 Batteries
                      SONOTRAX Lite, SONOTRAX Basic, SONOTRAX Basic A and SONOTRAX Pro are
                      powered either by 9V alkaline batteries (IEC Type No. 6LR61/6LF22 or equivalent) or by
                      rechargeable NI-MH batteries (8.4V).
                      SONOTRAX II and SONOTRAX II Pro are powered by lithium-ion polymer batteries
                      supplied by the manufacturer.
- 11 -
                      Chapter 3 Operation
                      NOTE:
                      To ensure that the Doppler works properly, please read this chapter and Chapter 1
                      Safety Guide before operation; follow the steps when connecting all the components.
                      Open the package; take out the Doppler and accessories carefully. Keep the package for possible
                      future transportation or storage. Check the components according to the packing list.
                       Check for any mechanical damage.
                       Check all the cables and accessories.
                      If there is any problem, contact us or your local distributor immediately.
- 12 -
                      For SONOTRAX II and SONOTRAX II Pro, insert the battery connector into the battery
                      socket in the compartment. (Figure 3-2 b)
(a) (b)
                                                                        CAUTION
                      1    Take out the alkaline/NI-MH battery and keep it in a cool and dry environment if the
                           Doppler is not used for an extended period.
                      2    Do not remove the lithium-ion polymer battery frequently after the initial installation. If
                           the Doppler is not used for an extended period, charge the battery at least every
                           three months.
- 13 -
                                                                    CAUTION
                      Do not take out or place the probe when the Doppler is on. Remember to take out the
                      probe before switching on the Doppler, and place the probe after switching off the
                      Doppler.
- 14 -
                                                                    CAUTION
                      Do not pull the probe cable directly.
                      NOTE:
                      Place the temporarily unused probe carefully and avoid falling off, splash or stress, etc.
                      When the Doppler is not used for a long time, it’s recommended to connect the probe to
                      the Doppler and keep them safely in the package.
- 15 -
                      When the Doppler works in this mode, the working mode on LCD displays 1.
                      As soon as a fetal heart signal is detected, the heart shape on LCD flashes and the FHR display
                      area displays the real-time FHR value.
                      Keep pressing the MODE button until the working mode on LCD displays 2.
                      The heart shape on LCD flashes when fetal heart signal is detected, and the FHR value indicates
                      the average value of every eight beats.
                      Keep pressing the MODE button until the working mode on LCD displays 3.
                      Press the START/STOP button once and start counting immediately, viz. count one at the
                      moment when the button is pressed. The LCD displays the flashing heart shape symbol and “---”.
                      Press the button again on the 10th count (after nine beat intervals).
                      The Doppler automatically calculates and displays the average FHR over the 10 beats. This rate
                      value will not disappear until another measurement starts or the mode is changed.
                      This mode is only available with SONOTRAX Pro and SONOTRAX II Pro.
                      Keep pressing the MODE button until the working mode on LCD displays 5.
                      Press the START/STOP button, the sampling frequency switches between 4 and 8 on the LCD.
                      “4” means the frequency is 4KHz, and the longest record time is 480 seconds; while “8” means
                      8KHz, and the longest record time is 240 seconds.
- 16 -
                      NOTE:
                      1    The setting under this mode will be saved automatically after mode switches
                           or normal power off.
                      2    After changing frequency, the records related to the original frequency will be
                           cleared automatically.
                      NOTE:
                      1    The best quality records will only be obtained if the probe is placed in the optimum
                           position.
                      2    Positions with strong placental sounds or umbilical blood flow sound should be
                           avoided.
                      3    If the fetus is in the cephalic position and the mother is supine, the clearest heart
                           sound will normally be found on the midline below the umbilicus. During examining,
                           the pregnant woman’s prolonged lying in the supine position should be avoided
                           owing to the possibility of supine hypotension. Sitting up or lateral positions are
                           preferable and may be more comfortable.
                      4    It is impossible to examine FHR unless a fetal heart sound is present. The fetal pulse
                           can be distinguished from the maternal pulse by feeling the mother’s pulse during
                           the examination.
- 17 -
(the socket with the symbol “ ”) of the PC, refer to figure 3-7. If the PC has no audio input
socket, insert the plug into the microphone socket (the socket with the symbol “ ”).
- 18 -
                      Turn on PC and run the sound recorder (Click on Start > Programs > Accessories >
                      Entertainment > Sound Recorder). Refer to figure 3-8.
                      Perform FHR examination with the method described in section 3.6. When the ideal signal is
                      detected, unplug the earphone (if it’s connected) and insert the other plug of the audio cable into
                      the earphone socket on the Doppler.
- 19 -
- 20 -
                         c) If the line in or/and microphone volume control is/are not shown in the Volume Control
                            menu, click on Options > Properties, tick Line In and Microphone as shown in figure
                            3-12, click on OK:
- 21 -
- 22 -
                                                                 WARNING
                      The device is not intended for ophthalmic use. Do not use it for examining ophthalmic
                      vessels, or any other procedures which may cause the ultrasound beam to pass through
                      the eye.
                      4MHz, 5MHz or 8MHz vascular probes are to be connected to the Doppler to perform vascular
                      examination.
                      Choose the appropriate probe as required. The probe with how frequency has a deeper
                      penetration depth, while the probe with high frequency has better resolution and wider detecting
                      range. The 4MHz vascular probe is optimized for examining blood vessels; the 5MHz vascular
                      probe is optimized for examining deeper vessels, and the 8MHz vascular probe is optimized for
                      examining surface vessels.
                      Apply a liberal amount of gel on the site to be examined. Place the probe at a 45º angle on the skin
                      over the vessel to be examined. Adjust the position of the probe to obtain the loudest blood flow
                      sound. Refer to figure 3-13 for the probe sites:
- 23 -
- 24 -
                                                                   CAUTION
                      Make sure the Doppler is shut down before charging the battery or opening the battery
                      compartment.
                                                                  WARNING
                      DO NOT CHARGE THE ALKALINE BATTERY.
                                                                  WARNING
                      Fit the battery to the sockets of the charger properly before plugging the charger into
                      power socket.
- 25 -
                       NOTE:
                       The Doppler is not available for examining during charging.
- 26 -
                      Safety:
                                   Complies with:          IEC 60601-1:1988+A1+A2, EN 60601-1:1990+A1+A2,
                                                           IEC/EN 60601-1-2:2007, IEC/EN 61266
                      Classification:
                                   Anti-electric Shock Type:           Internally powered equipment
- 27 -
                      Display:
                           45mm*25mm LCD display
                      FHR Performance:
                           Sensitivity:         10 weeks gestation (3MHz)
                           FHR Measuring Range: 50bpm ~ 210bpm
                           Resolution:          1bpm
                           Accuracy:            ±3bpm
                      NOTE:
                      In some cases, fetal heart beats at 10 weeks gestation can not be detected due to the
                      maternal physical difference and the operator’s technique.
                      White Backlight:
                           Two Brightness Adjustable: OFF, ON
- 28 -
                                                           ≥ 300 times (after 300 circles, only 90% of the capacity can
                          Circle Life:
                                                           be restored.)
Ultrasound
- 29 -
- 30 -
                      Chapter 5 Maintenance
                      5.1 Maintenance
                      You must check that the equipment does not have visible evidence of damage that may affect the
                      patient and the operator’s safety or the Doppler’s capability before each use. If the damage is
                      evident, replacement is recommended.
                      The probe is frangible and must be handled with care.
                      Wipe the remaining gel after use to prolong the probe life.
                      The overall check of the Doppler, including safety check and function check, should be
                      performed by qualified personnel every 12 months, and each time after service.
                      5.2 Cleaning
                      Before cleaning, switch off the Doppler.
                      Keep the exterior surface of the device clean and free of dust and dirt.
                      Clean the exterior surface (display screen included) of the main unit with a dry, soft cloth. If
                      necessary, clean it with a soft cloth soaked in soap water and then wipe it dry with a dry cloth
                      immediately.
                      Wipe the remaining coupling gel off the probe. Clean it with a soft cloth dampened in solution
                      (soap water, 70% ethanol or isopropranol alcohol) and then air-dry it or wipe the moisture with a
                      soft dry cloth.
                                                                  CAUTION
                      1    Clean the main unit with soap and water only. Avoid strong solvent, such as acetone.
                      2    Never use an abrasive such as steel wool or metal polish.
                      3    The main unit is not waterproof. Do not immerse any part of it into liquid.
                      4    Avoid pouring liquids on the main unit while cleaning.
                      5    Do not remain any solution on the surface after cleaning.
                                                                  CAUTION
                      1    Pay attention not to immerse the probe socket into the disinfector.
                      2    The main unit is not waterproof, do not disinfection it.
                      3    Never try to sterilize the probe or the main unit.
- 31 -
                      6.1 Warranty
                      EDAN’s obligation under this warranty is limited to repairing, at EDAN’s option, any part which
                      upon EDAN's examination proves defective. If the product doesn’t function as warranted during
                      the warranty period, we will repair or replace it without charge.
                      Software or Firmware
                      EDAN software and firmware products which are designated by EDAN for use with a hardware
                      product, when properly installed on that hardware product, are warranted not to fail to execute
                      their programming instructions due to defects in materials and workmanship. If EDAN receives
                      notice of such defects during the warranty period that begins on the date of shipment, EDAN
                      shall repair or replace software media or firmware which does not execute their programming
                      instructions due to such defects. But EDAN doesn’t warrant that operating of the hardware,
                      software, or firmware shall be uninterrupted or free from error.
                      NOTE:
                      The charges of freight and others are excluded under warranty.
                      This unit has no parts can be repaired by users themselves. All the service should be performed
                      by authorized and qualified personnel.
                      Limit of Warranty
                      The warranty is void in the case of
                               ♦    Assembly, extensions, readjustments of any parts;
                               ♦    Modification and repair by unauthorized persons;
                               ♦    Subsequent damage caused by improper use or maintenance;
                               ♦    Replacement or remove of Serial number label and manufacturer label;
- 32 -
                                                                     CAUTION
                      Only the parts supplied by the manufacturer should be used with the Doppler.
Accessory
Earphone 01.14.104751
Consumable
- 33 -
                       The SONOTRAX Series Ultrasonic Pocket Doppler is intended for use in the electromagnetic
                       environment specified below. The customer or the user of the device should assure that it is used in such
                       an environment.
                       The SONOTRAX Series Ultrasonic Pocket Doppler is intended for use in the electromagnetic
                       environment specified below. The customer or the user of the device should assure that it is used in such
                       an environment.
- 34 -
                       NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
                       NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
                       absorption and reflection from structures, objects and people.
                       a
                               Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
                               and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
                               predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
                               transmitters, an electromagnetic site survey should be considered. If the measured field strength in
                               the location in which the SONOTRAX Series Ultrasonic Pocket Doppler is used exceeds the
                               applicable RF compliance level above, the SONOTRAX Series Ultrasonic Pocket Doppler should be
                               observed to verify normal operation. If abnormal performance is observed, additional measures may
                               be necessary, such as reorienting or relocating the SONOTRAX Series Ultrasonic Pocket Doppler.
                       b
                               Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
- 35 -
                       The SONOTRAX Series Ultrasonic Pocket Doppler is intended for use in an electromagnetic environment
                       in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent
                       electromagnetic interference by maintaining a minimum distance between portable and mobile RF
                       communications equipment (transmitters) and the SONOTRAX Series Ultrasonic Pocket Doppler as
                       recommended below, according to the maximum output power of the communications equipment.
                       For transmitters rated at a maximum output power not listed above, the recommended separation distances
                       in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P
                       is the maximum output power rating of the transmitter in watts (W) according to the transmitter
                       manufacturer.
                       NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
                       NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
                       absorption and reflection from structures, objects and people.
- 36 -
                                                                 T    20   4.8   4.0     -
                                 50              45.7                                              0     57.6        186                    94                5.93                        109.2
                                                                 Ba   40   9.6   8.0     -
                                                                 T    20   4.8   3.4     -
                                 75              45.7                                              0     56.4        175                    90                5.78                        107.8
                 1.58                                            Ba   40   9.6   6.8     -
              A=45.7dB@
                2MHz                                             T    20   4.8   3.4     -
                                 100             45.7                                              0     56.4        174                    89                5.82                        107.9
                                                                 Ba   40   9.6   6.8     -
                                                                 T    20   4.8    -      -
                                 200             45.7                                              0     49.6        173                    90                5.68                        100.9
                                                                 Ba   40   9.6    -      -
- 37 -
           1.58                                            T    10      4       -     -
           A=44.5dB@               50                                                             0   61.5       179              90                   5.97                      111.9
           3MHz                                            Ba   43.6   17.9     -     -
                                                           T    10      3       -     -
                                   75                                                             0   57.2       186              94                   5.93                     107.63
                                                           Ba   43.6   13.5     -     -
                                                           T    10      3       -     -
                               100                                                                0   57.2       173              88                   5.87                      107.5
                                                           Ba   43.6   13.5     -     -
                                                           T    10      1       -     -
                               200                                                                0   50.2       171              89                   5.67                      100.3
                                                           Ba   43.6   6.5      -     -
           2.38                                            T    10      4       -     -
           A=42.0dB@               50                                                             0   61.5       194              88                   6.73                      110.2
           3MHz                                            Ba   43.6   17.9     -     -
                                                           T    10      3       -     -
                                   75                                                             0   57.2       172              87                   5.92                      105.1
                                                           Ba   43.6   13.5     -     -
                                                           T    10      2       -     -
                               100                                                                0   53.2       165              85                   5.76                      100.9
                                                           Ba   43.6   9.6      -     -
                                                           T    10      1       -     -
                               200                                                                0   50.2       160              84                   5.59                      97.8
                                                           Ba   43.6   6.5      -     -
- 38 -