Agilent 7700 Series ICP-MS
MassHunter Workstation
Quick Start Guide
This guide describes how to use the Agilent 7700 Series ICP- MS
MassHunter Workstation.
                   Agilent Technologies
    Startup                                                                  3
    Preparing for Analysis                                                   4
      Things to check before analysis                                        4
      Prepare tuning solutions                                               4
      Ignite the plasma                                                      5
      Check tuning                                                           6
    Setting Quantitative Method Parameters                                  12
      Set up a Quantitative Method                                          12
      Access Method parameters directly                                     19
      Use the Method Wizard to create a method                              19
    Running the Samples                                                     21
      Set up the Sequence                                                   21
      Analyze the samples                                                   23
    Quantitative Data Analysis                                              25
      Typical scenarios for Data Analysis                                   25
      Open an existing batch for data analysis                              26
      Creating a new batch for data acquisition                             26
      Create the Data Analysis (DA) method for Quantitative data analysis   28
      Process the batch                                                     32
      View analysis results                                                 33
      Save analysis results                                                 34
      Generate a report                                                     34
    Semi-Quantitative Analysis                                              35
      Prepare for SemiQuant Analysis                                        35
      Create a SemiQuant acquisition method                                 35
      Analyze the samples                                                   37
      Create a new batch or open a new batch for SQ data analysis           37
      Create the Data Analysis (DA) method for Quantitative data analysis   28
      Process the batch                                                     40
      Correcting the Semiquant Factors                                      41
      Change the SemiQuant Factors                                          42
2         G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide
Startup
                    To start the Agilent 7700 Series ICP- MS MassHunter Workstation software:
                    • Double- click the ICP MS Top desktop icon              , or
                    • Select Programs > ICP- MS MassHunter Workstation > ICP- MS Top from
                      the Windows Start menu.
                    The ICP- MS Top window opens:
         Toolbar    The following shortcuts are available on the ICP- MS MassHunter Workstation
                    toolbar:
                    Button   Action                             Button   Action
                             Display Instrument Status Panel             Display Data Acquisition Panel
                             Display Tuning Panel                        Display Data Analysis Panel
                             Run Current Method                          Run Current Sequence
                             Run Method Wizard                           Edit Sequence
                             Edit Method                                 Load Sequence
                             Load Method                                 Save Sequence
                             Save Method                                 Display online help window
G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide                                             3
Preparing for Analysis
              The following topics will help you prepare the instrument for analysis with
              Agilent 7700 Series ICP- MS MassHunter Workstation.
     Things to check before analysis
              Utilities
                 •   Argon gas pressure: 500 to 700 kPa
                 •   Cell gas (Helium): 90 to 130 kPa
                 •   Cell gas (Hydrogen): 20 to 60 kPa
                 •   Exhaust duct (on)
                 •   Cooling water (Chiller or heat exchanger on)
                 •   Drain and rinse tank (not full)
              Peristaltic pump tubing
                 • Sample, drain, and internal standard lines
     Prepare tuning solutions
              You will need the following standards and solutions for ICP- MS operation:
              • Tuning solution (1ppb Li, Co, Y, Ce, Tl)
              • Internal standard solution
              • Dilute the P/A factor tune solutions 75x, then mix the two solutions 1:1 to
                prepare a P/A factor tune solution for use in the lab. PA factor tuning is
                used to achieve linear functionality across the detector’s pulse and analog
                modes. Pulse mode is used to quantitate relatively low concentrations (ppb
                range). Analog mode is used to quantitate relatively high concentrations
                (ppm range).
4                                 G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide
        Ignite the plasma
                    1 Open the Instrument Control panel in either of the following ways:
                      • Select Instrument Control from the Instrument menu, or
                      • Click the Instrument Control     icon.
                      The ICP- MS Instrument Control panel opens:
                    2 Confirm that the instrument is in Standby mode by checking the title bar of
                      the Instrument Control panel.
                    3 Ignite the plasma by clicking the Ignite Plasma       icon in the toolbar.
                    4 Once the plasma is on and the instrument goes from standby to analysis
                      mode, check tuning as described in the following section.
G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide                                  5
     Check tuning
                 The following diagram labels the parts of the ICP- MS instrument. You can
                 refer to this diagram during the discussion of tuning parameters in this
                 section.
                                              Interface       Octopole Reaction System (ORS)
                                    Skimmer Cone                                        Quadrupole Mass Analyzer
                                                                          Octopole
               Ion Source                       Ion                     Reaction Cell
                                               Lens                                     Hyperbolic Quadrupole Rod Assembly
                               RF Generator              Reaction Gas
    Sample Introduction                                                                      Electron Multiplier
                                                                                                              Detector
                AMFC                                                                       Turbo-
                                                                                          Molecular
                                                                                           Pump
               Ar Gas                                                                                    Vacuum System
                        Sample
                          Sampling Cone       Foreline
                                               Pump
                 Open the Tuning panel by selecting Tune from the Instrument menu,
                 or by clicking the Tune    icon on the ICP- Top toolbar.
6                                             G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide
         Toolbar    The following items (shortcuts) are available on the Tuning panel toolbar:
                    Button   Action                             Button   Action
                             Load tuning parameters from file            Save tuning parameters to file
                             Generating tuning report                    Print current graph
                             Tune sensitivity                            Tune Oxide Ion
                             Tune Doubly Charged Ion                     Resolution and Axis
                             Select Mode & start Autotune                Set Acquisition parameters for tuning
                             View meter control panel                    Display online help window
             Tip    See the “Tuning 7700” in online help for more information on individual
                    tuning parameters.
                    Plasma Correction
                    Execution of plasma correction is necessary in the following cases:
                    • When the instrument is installed
                    • When the nebulizer has been changed
                      (However, if the same nebulizer unit is reused repeatedly by detaching and
                      reattaching, it is not necessary to conduct the correction each time as long
                      as the nebulizer is installed correctly.)
                    • When the sampling cone is changed
                    • When the torch is changed
                    • When the matrix tolerance has changed after the instrument has been used
                      for 6- 12 months and normal maintenance procedures have been followed
                       Once plasma correction is performed, there is no need to conduct it again in
                       most cases, provided the instrument is used for normal measurement.
G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide                                                    7
    Operating Method
    When performing plasma adjustment, use a 1pbb tuning solution containing
    Ce. The following explains the method, refer to the online help.
    Settings for Preset Mode
    Select the nebulizer and tuning type, and set to the Preset mode. Under the
    Preset mode, parameters such as RF power, sampling position, carrier gas flow
    rate, makeup gas flow rate, and nebulizer pump speed are fixed to the preset
    values.
    1 In the Tuning window, select Preset Plasma >> Select Plasma.
       The Plasma Select dialog appears.
    2 Check the Preset Mode box.
    3 Select the type of nebulizer to be used from the drop- down menu (For
       7700x).
    4 Select Robust or one of the 3 levels of Ultra Robust (For 7700x). Select Hot
       Plasma (For 7700s).
       If the matrix concentration exceeds 1%, select Ultra Robust- Level: High (if
       the desired sensitivity is not achieved, select Level Medium or Low).
    5 Click OK.
       The Tuning window becomes active.
    Checking the Tuning
    Check the tuning as follows:
    1 Check the each parameters:
      • Open the last- used tuning file (typically nogas.u, h2.u or he.u).
      • Select Acquisition Parameters from the Acq. Params menu. Set the mass
        numbers to be measured as 7 (Li), 89 (Y), 205 (Tl), and 156/140 (Ce oxide
        formation rate) and click OK.
      • Check the following values displayed on the Tuning window:
8                         G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide
                    Table 1     Typical Sensitivity Values (For 7700x)
                                    Mass                                        Sensitivity (cps/ppb)
                                                                No Gas Mode        He Gas Mode          H2 Gas Mode
                                                                                   (3.6ml/min)          (4.0ml/min)
                   7Li                                                 30,000                    ---                  ---
                   59
                        Co                                                ---               24,000                    ---
                   89Y                                                100,000                    ---           60,000
                   205Tl                                               60,000                    ---                  ---
                   Oxide ratio ≤ 1.5% (≤2% for MicroFlow Nebulizer)
                    2 If the desired sensitivity is not achieved, select Autotune from Tume menu.
                      Refer to the Help and execute Autotune.
                    3 After completing autotune, check the sensitivity parameters on the ICP- MS
                      Tuning - Sensitivity window.
                    4 Save the tune file by selecting Save from the File menu. In general you will
                      want to overwrite the tune file that was open, so that it has the updated
                      values.
                    5 (optional) Generate a tune report by selecting Generate Report from the
                      File menu. Enter optional comments if desired and click OK. A report
                      showing the following information is sent to your printer:
                         • Sensitivity data
                         • Resolution/Axis data
                         • Lens parameter settings
                         Depending on how your system is configured, tuning results will also be
                         recorded in the maintenance log.
             Tip         You can also get a hard copy of current tune parameters by clicking the
                         Stop button and selecting Print from the File menu.
G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide                                                               9
                          Reference: For 7700x
                          The recommended tuning parameter values for 7700x are shown below.
Table 2         Recommended Values for 7700x (When Using a MicroMist Nebulizer)
        Parameter                    No Gas Mode                      Cell Gas Mode                High Energy Collision Mode *2
                            Recommended Recommended Recommended Recommended Recommended Recommended
                                Value      Range        Value      Range        Value      Range
RF Power [W]                           1550            Fixed             1550             Fixed             1550           Fixed
Smpl Depth [mm]                         8.0            Fixed               8.0            Fixed               8.0          Fixed
Carrier Gas [L/min]                    1.05      1.01 to 1.11            1.05      1.01 to 1.11              1.05    1.01 to 1.11
Makeup Gas [L/min]                        0         0 to 1.11               0         0 to 1.11                0        0 to 1.11
Dilution Gas [L/min]                      0            Fixed                0             Fixed                0           Fixed
Neb Pump [rps]                          0.1            Fixed               0.1            Fixed               0.1          Fixed
S/C Temp [degC]                           2            Fixed                2             Fixed                2           Fixed
He or H2 gas [ml/min]                     0            Fixed          3.6 (He)   3.2 to 4.0 (He)          10 (He)    7 to 10 (He)
                                                                      4.0 (H2)   3.6 to 4.4 (H2)
Extract 1 [V]                             0            Fixed                0             Fixed                0           Fixed
Extract 2 [V]                          -180     -200 to -160             -180      -200 to -160             -180     -200 to -160
Omega Bias [V]                          -80       -110 to -70             -80       -110 to -70              -80      -110 to -70
Omega Lens [V]                           10           7 to 12              10           7 to 12               10          7 to 12
Cell Entrance [V]                       -30        -40 to -30             -40        -40 to -30             -130     -150 to -110
Cell Exit [V]                           -50        -60 to -40             -60        -60 to -40             -150           Fixed
Deflect [V]                              10           8 to 15               0            -5 to 4             -80        -90 to-70
Plate Bias [V]                          -40        -50 to -30             -60             Fixed             -150           Fixed
OctP RF [V]                             180       150 to 200              180       150 to 200               190      180 to 200
                     *1
OctP Bias [V]                            -8         -10 to -6             -18             Fixed             -100           Fixed
                *1
QP Bias [V]                              -3          -7 to -3             -15             Fixed              -96       -97 to -90
                          *1 Generally, set the QP Bias at least 2 to 3V higher than OctP Bias.
                          *2 High-Energy Collision mode allows for enhanced reduction of interference under the Helium mode.
                             However, as it will also result in reduced low-mass sensitivity, it should be used in accordance with
                             the purpose of the analysis.
10                                                  G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide
                          Reference: For 7700s
                          The recommended tuning parameter values for 7700s are shown below.
Table 3         Recommended Values for 7700s (When Using a MicroFlow Nebulizer)
       Parameter              High Sensitivity           Cell Gas Mode          High Energy Collision         Cool Plasma
                                                                                       Mode*2
                           Recom-         Recom- Recom-             Recom- Recom-             Recom- Recom-             Recom-
                           mended         mended mended             mended mended             mended mended             mended
                             Value         Range   Value             Range   Value             Range   Value             Range
RF Power [W]                  1500            fixed     1500            fixed      1500            fixed      600           fixed
Smpl Depth [mm]                   8         7 to 10         8         7 to 10         8           7 to 10      18           fixed
Carrier Gas [L/min]             0.7           fixed       0.7           fixed       0.7            fixed      0.7           fixed
Makeup Gas [L/min]              0.5      0.3 to 0.7       0.5      0.3 to 0.7       0.5      0.3 to 0.7      0.75      0.6 to 1.2
Dilution Gas [L/min]              0           fixed         0           fixed         0            fixed        0           fixed
Neb Pump [rps]                  0.1           fixed       0.1           fixed       0.1            fixed      0.1           fixed
S/C Temp [degC]                   2           fixed         2           fixed         2            fixed        2           fixed
He or H2 gas [ml/min]             0           fixed   3.6(He) 3.2 to 4.0(He)     10(He)    7 to 10(He)          0           fixed
                                                      4.0(H2) 3.6 to 4.4(H2)
Extract 1 [V]                   4.5          3 to 7       4.5          3 to 7       4.5            3 to 7    -120     -200 to -40
Extract 2 [V]                  -100     -170 to -60      -100     -170 to -60      -100     -170 to -60        -5        -30 to 5
Omega Bias [V]                  -70     -100 to -30       -70     -100 to -30       -70     -100 to -30       -70     -120 to -30
Omega Lens [V]                   11         5 to 15        11         5 to 15        11           5 to 15       6         3 to 10
Cell Entrance [V]               -30      -40 to -30       -40      -40 to -30      -130   -150 to -110        -30      -40 to -30
Cell Exit [V]                   -50      -60 to -40       -60      -60 to -40      -150            fixed      -60           fixed
Deflect [V]                      12         8 to 15         0         -5 to 4       -80      -90 to -70         9         5 to 13
Plate Bias [V]                  -40      -50 to -30       -60           fixed      -150            fixed      -60           fixed
OctP RF [V]                     180     150 to 200        180     150 to 200       -190     180 to 200        150     100 to 200
OctP Bias [V] *1                 -8       -12 to -6       -18           fixed      -100            fixed      -18      -30 to -10
QP Bias [V] *1                   -3        -5 to -3       -15           fixed       -96      -97 to -90        -5        -5 to -3
                          *1 Generally, set the QP Bias at least 2 to 3V higher than OctP Bias.
                          *2 High-Energy Collision mode allows for enhanced reduction of interference under the Helium mode.
                             However, as it will also result in reduced low-mass sensitivity, it should be used in accordance with
                             the purpose of the analysis.
G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide                                                                  11
Setting Quantitative Method Parameters
               The following topics will help you create an acquisition method for
               quantitative analysis with the Agilent 7700 Series ICP- MS MassHunter
               Workstation.
     Set up a Quantitative Method
               Create a quantitative method for environmental analysis as described below:
               1 Select Edit Entire Method from the Methods menu. This initiates a series
                 of dialog boxes, so you can set the parameters for your method. See the
                 online help for a full description of the items on these dialog boxes.
         Tip   See also “Use the Method Wizard to create a method” on page 19.
               2 On the Edit Method dialog box, mark the following options, then click OK.
               3 On the Method Information dialog box:
                 a Enter a comment
                 b Mark the Data Analysis check box to run the data analysis automatically
                   after data acquisition. For instructions on setting up a Data Analysis
                   method, see “Open an existing batch for data analysis” on page 26.
                 c Click OK.
12                                G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide
                    4 On the Select Sample Types (or Select QC Items, if the Intelligent Sequence
                      software is installed) dialog box:
                      a Configure the list of Selected QC Items on the right to contain CalStd
                        and Sample only.
                      b Depending on the method you started with, you may have to use the
                        Add- > and <- Remove buttons to create the desired list.
                      c Click OK.
                    5 On the Interference Equations dialog box:
                      a Display interference correction equations in one of the following ways:
                      • Mark Show All Masses to display all interference correction equations.
                      • From the pull- down menu at the bottom right, select Interference
                        Equation Library, to display interference correction equations in the
                        selected library.
                      b To edit the interference equation for a mass:
                      • Click on the desired mass number in the list.
                      • Click the Edit button to open the Edit Equation dialog box.
                      • Edit the equation as described in online help and click OK to close the
                        Edit Equation dialog box.
                      c Repeat Step 5b for all masses you want to edit.
                      d (optional) Click the Save button on the Interference Equation dialog box
                        to save the edited equations for use with other methods.
                      e When finished, click OK.
                    6 On the Acquisition Mode dialog box:
                      a Check the Spectrum (Multi Tune) option.
                      b Click OK. Multi Tune allows you to use different tuning modes (e.g.
                        Hydrogen and Helium modes) for different elements in a single run.
G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide                                  13
     7 Specify the masses and element to analyze as follows:
          a On the Spectrum (Multi Tune) Acquisition Parameters dialog box, click
            the Periodic Table button as shown below:
     7a
          This opens the Masses dialog box shown below:
14                        G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide
                      b Using Table 4 on page 18 as a guide, click to select the masses to
                        monitor for each element. Selected masses appear in red in the
                        dialog box.
                        Double- click on a element to open the Select Isotopes dialog box.
                        Click the buttons for the isotopes you want to monitor. You can
                        select one or more isotopes per element. Red check marks indicate
                        the selected isotopes. To clear an isotope, click the right mouse
                        button. Click OK when you are finished selecting isotopes for that
                        mass.
                      c Repeat Step 7b to set up each mass as shown in Table 4.
                      d When you are finished, click the Mass Table button. The Mass Table
                        dialog box opens to display the masses to be monitored for each element.
                        Check that the masses listed in the Mass Table dialog box match those
                        shown in Table 4 and adjust them if necessary. This is the list of
                        masses that will be monitored in the method.
                      e When finished setting up all the required elements and masses, click OK
                         on the Mass Table and Masses dialog boxes.
                    8 Set the Tune File and Stabilization time for each step in the upper right
                      area of the Spectrum (Multi Tune) Acquisition Parameters dialog box:
                       Step      Tune File                                     Stabilization Time
                         1       Hydrogen mode                                 30 sec
                         2       Helium mode                                   30-50 sec
G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide                                       15
     9 Set the Integration time per Point for each mass:
       a Click on the row for the mass in the column for the appropriate step (i.e.
         Step 1 for Hydrogen mode, Step 2 for Helium mode) as specified in the
         Reaction Gas column in the Table 4 on page 18. You can select
         multiple masses by <Ctrl>+clicking on them.
       b Enter the Integration time per Point as specified in Table 4. The
         integration times in the table are generally relatively short, and can
         be extended as necessary.
       c Click the Enter button. The per Point integration time and calculated
         per Mass integration time will be updated in the table.
       9b
                                                                             9a
       9c
       d Repeat for each mass in the table. For masses such as 71 that are
         monitored in both steps, remember to enter integration times for both.
16                      G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide
                    10 Make the following additional selections on the Spectrum (Multi Tune)
                       Acquisition Parameters dialog box:
                      a Select Full Quant (3) for the Peak Pattern.
                      b Set the Repetition count to 3.
                                                                              10b
                                                           10a
                      c Mark the Return to First Tune Step option at the bottom of the screen,
                        and click OK.
                    10c
                    11 On the Peristaltic Pump Program dialog box, set the following parameters,
                       then click OK. See online help for more information on these parameters
                       and on the optional features: Intelligent Rinse and Execute Pre- emptive
                       Rinse.
                      • Before Acquisition
                      • After Acquisition (Rinse Port)
                       • After Acquisition (Rinse Vial)
                    12 (optional) Set up pre- emptive rinsing as follows:
                      a Select Set PeriPump Program from the PeriPump menu in the
                        Acquisition window.
                      b On the Peristaltic Pump Program dialog box, mark the Execute
                        Pre- emptive Rinse option at the bottom of the screen.
                      c Enter an initial time of 15 seconds, then click OK.
                       d Adjust the time so that the analysis is completed 5 seconds before the
                         rinse solution reaches the nebulizer.
                    13 To save the method:
                      a On the Method Save Options dialog box, mark all items, and click OK.
                      b On the Save Method As dialog box, enter a method name, and click OK.
G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide                                 17
     Table 4     Set up information for elements and masses for Environmental Analysis
       Element         Mass       Confirmation       Mode              Internal       Approx. Integration
                                     Mass1                           Standard2         Time (sec/Point)
                                                                           6
          Be             9                          No gas                  Li                0.3
                                                                           6
          B              11                          No gas                 Li                0.1
          Na             23                            He                 45Sc               0.05
                                                                          45Sc
          Mg             24                            He                                    0.05
                                                                          45
          Al             27                            He                    Sc               0.3
          P              31                         No Gas        6Li (72Ge, 45Sc)3           0.1
          K              39                            He               9 or 71              0.05
          Ca             44            43              He               9 or 71               0.1
                                                                          45Sc
          V              51                            He                                     0.5
                                                                          45Sc
          Cr             52            53              He                                      1
                                                                          45Sc
          Mn             55                            He                                     0.1
          Fe             56            57              He                 45Sc                0.1
                                                                          45Sc
          Co             59                            He                                     0.1
                                                                          45Sc
          Ni             60                            He                                      1
          Cu             63            65              He                 45Sc                0.1
                                                                         72Ge
          Zn             66            64              He                                     0.1
                                                                         72Ge
          As             75                            He                                      1
          Se             78                        He or (H2)4           72Ge                  5
                                                                         72Ge
          Br             79                         No Gas                                    0.1
                                                                         103Rh
          Mo             95             98             He                                     0.1
          Ag            107            109          No Gas               115In                0.1
                                                                    115In, (209Bi)
          Cd            111            114          No Gas                                     1
                                                                 103Rh (115In, 209Bi)
          Sn            118                         No Gas                                    0.1
                                                                    115In, (209Bi)
          Sb            121                         No Gas                                    0.1
                                                                 103Rh (115In, 209Bi)
           I            127                         No Gas                                    0.1
                                                                    115In, (209Bi)
          Ba            137            135          No Gas                                    0.1
                                                                         209Bi
          Hg            201            202          No Gas                                     1
                                                                    175Lu, (209Bi)
          Tl            205                         No Gas                                    0.1
                                                                    175Lu, (209Bi)
          Pb            208                         No Gas                                    0.1
                                                                    175Lu, (209Bi)
          Th            232                         No Gas                                    0.1
                                                                    175Lu, (209Bi)
          U             238                         No Gas                                     1
     1) Can be subject to interferences.
     2) ISTD solution P/N is 5188-6525. See page 23 for more information.
     3) Alternate internal standards are shown in parentheses.
     4) When H2 option is installed (then better MDL can be achieved).
18                            G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide
        Access Method parameters directly
                    You can access some method panels directly, using the following menu items:
                    Method parameters        Window        Menu             Menu item
                    Method Information       ICP-MS Top    Methods          Edit Method Information
                    Select Sample Items,     ICP-MS Top    Methods          Select QC Items
                    Select QC Items
                    (Intelligent Sequence
                    Software is installed)
                    Interference Equation    ICP-MS Top    Methods          Edit Interference Equation
                    Acquisition Mode         Acquisition   EditParameters   Set Mode
                    Spectrum Acquisition     Acquisition   EditParameters   Set Parameters
                    Parameters
                    Peristaltic Pump         Acquisition   PeriPump         Set Peripump Program
                    Program
                    Method Save Options      ICP-MS Top    Methods          Save As
        Use the Method Wizard to create a method
                    This way of creating a method makes it easy to create methods for
                    environmental analyses, since it supplies many of the parameters that are
                    appropriate for various sample types and regulations.
                    1 Select Run Method Wizard from the Methods menu in ICP- MS Top.
   General Items    2 Enter name for the method and click the Next> button.
         Sample     3 Select the sample as follows:
                       a Select Environmental as the Application Type.
                       b Select a Sample, such as Drinking Water or High TDS. A description of
                         the current selection is displayed in the lower part of the window.
                       c Select a Regulation, such as EPA200.8 or EPA6020. Only the guidelines
                         that apply to the selected sample are displayed.
                       d Click the Next> button.
G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide                                            19
     Elements   4 The list of masses and elements that will be monitored is displayed. The
                  information in this list is based on the sample and regulation selected
                  above. Review the element information and make any changes as follows:
                  a Right- click on the element of interest in the Element List Information.
                  b Select one of the following options from the shortcut menu:
                  • Add - see Step 4c below
                  • Delete - removes the selected mass from the list
                  • Edit - see Step 4c below
                  c If you are adding or editing an element, enter the following information
                    on the Add (or Edit) Elements dialog box, then click OK.
                  •   Mass
                  •   Element Name
                  •   Integration time (sec/Point)
                  •   Tune step number
                  d (optional) Click the Edit/View Equation button to view or edit an
                    interference correction equation. See online help for more information.
                  e Click the Next> button.
       Finish   5 Review the method information, then click the Finish button.
          Tip   You can edit the method later by selecting Edit Entire Method from the
                Methods menu and marking the parts of the method to edit, or from the menus
                items shown in the previous section.
20                                 G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide
Running the Samples
                    The following topics will help you prepare a sequence, and run your samples
                    with the Agilent 7700 Series ICP- MS MassHunter Workstation.
        Set up the Sequence
                    Set up a sequence for your analysis as follows:
                    1 Select Edit Sample Log Table from the Sequence menu in the ICP- MS Top
                      window. The samples you enter will be injected in the order they are listed
                      in the table.
                    2 Enter the following information for the first sample (Row 1) of the table:
                      a Method: Double- click in the Method column of Row 1 to open the Select
                        Method dialog box. Select the desired method and click OK.
                      b Type: Click in the Type column of Row 1 to open a list of types.
                      • Select CalBlk for calibration blanks.
                      • Select CalStd for calibration standards.
                      • Select Sample for unknown samples.
                      c Vial: Click in the Vial column of Row 1 and enter the vial number of the
                        first sample or standard.
                      • The large bottles for cleaning solutions are positioned at 1, 2, and 3.
                      • The small vials on the rack are positioned at 1001, 1002, 1003... etc.
                      d Data File: Leave this cell blank to assign a file name automatically at run
                        time.
                      e Sample: Enter a name for the sample in the Sample column.
                      f Comment: Enter a comment for the sample in the Comment column.
                      g Dil/Lvl: Specify a Dilution factor or Level.
                      • For standard solutions: click in the Dil/Lvl column of Row 1 and select
                        the desired level from the list.
                      • For samples: click in the Dil/Lvl column of Row 1 and type in the
                        dilution factor (1 for undiluted), or double- click to open the Calculate
                        Dilution Factor dialog box. See online help for more information.
                      h ISTD Conc: Double- click in the ISTD Conc column of Row 1 to open the
                        Set ISTD Conc dialog box. Select the desired calibration level or set a
                        ISTD Conc value, then click OK.
G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide                                    21
     3 Fill in values for the rest of your samples as follows:
       a Select the first row and the subsequent rows by dragging the mouse.
       b Right- click in the highlighted rows and select Fill Down from the
         shortcut menu.
       c The values for Method, Type, Dil/LVL and ISTD Conc are copied from
          Row 1 to the selected blank rows. Sequential Vial numbers are assigned.
     4 Modify the information for the sample rows added in Step 3 to match your
       analyses.
     5 (optional) To turn off the plasma automatically at the end of the sequence,
       append the following row to the end of the sequence:
       • Type: Keyword
       • Data File: StandBy
       • Leave all other cells blank
     6 (optional) Select the Import DA Method from Existing Batch check box to
       import an existing data analysis method from the specified data batch
       directory. You can import the DA Method Only or the DA method and Std
       Data. You can also apply a data analysis method after data acquisition.
     7 Click OK to close the Edit Sample Log Table dialog box.
     8 If errors are found in the table, a message appears.
       a Click OK and the offending cell will be highlighted in the table.
        b Fix the problem in the highlighted cell.
     9 Repeat Steps 7 & 8 until no errors are detected.
     10 (optional) Select Save from the Sequence menu in the ICP- MS Top window.
        Enter a name for the sequence you just created, and click OK. You don’t
        need to save a sequence to run it, but doing so allows you to open it later
        and use it as a starting point for creating a new sequence more quickly.
22                      G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide
        Analyze the samples
                    1 Prepare the following standards, samples, and each solutions to use for
                      analysis. The solutions should contain 0.5% HCl as well as nitric acid.
                      • Make up the following calibration standards by diluting SPEX
                        XSTC- 760A with nitric acid to approx. 1%. The concentration of the
                        undiluted solution is 100x that of drinking water quality standard.
                         Calibration     SPEX XSTC-760A                                    Na            Mg, K, Ca
                         curve ranges                                                    (ppm)            (ppm)
                         Level 1         Blank HNO3                                         0                0
                         Level 2         1/1000 dilution (1/10 of standard)                10                5
                         Level 3         1/200 dilution (1/2 of standard)                  50                25
                         Level 4         1/100 dilution (standard concentration)           100               50
                         For high-concentration Na, Mg, K, and Ca, it is safer to prepare their mixed standard
                         solutions separately instead of mixing with the SPEX mixed standard solution, so as to
                         prevent contamination with the reagent, particularly if the required concentration level is
                         1/100 of the standard.
                      • Dilute samples with nitric acid to approx. 1%
                      • Dilute Internal standard solution (P/N 5188- 6525) 200 times (for on- line
                        internal standard)
                      • Use of nitric acid (1 to 5%) is recommended for the cleaning solution, as
                         it will clean the introduction system better than pure water. If pure
                         water is used, residual memory may be washed out while introducing
                         acidic samples, affecting the quantitative measurement.
                    2 Select Run from the Sequence menu or click the       icon on the toolbar to
                      open the Start Sequence dialog box.
                    3 Set the parameters to the desired values.
G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide                                                          23
     4 The Data Batch Directory is set to a unique name based on the current
       date and time. The format is year, month, day, hour, and sequential
       numbers. For example, 08H14K01.B is the second batch directory created
       between 10AM and 11AM on August 14, 2008. The parts of the name are
       derived as follows:
        08 =    2008
        H=      August
                (month - A through L corresponds to the 12 months, in order, so A = January, etc.)
        14=     the day of the month
        K=      the hour of the day
                (A - X correspond to 0 - 23)
        01=     a sequential number for sequences run within the same hour, i.e. 00, 01, 02, and so on
        .B=     the file extension for a batch file
       If you wish, you can type in a different directory name instead. The name
       can be up to eight characters long. The batch folder will be created in
       C:\ICPMH\1\DATA\.
     5 (optional) Select the Import DA Method from Existing Batch check box to
       import an existing data analysis method from the specified data batch
       directory. You can import the DA Method Only or the DA method and Std.
       Data.
     6 Click the Edit DA Method button in the lower left of the dialog box, then
       create a Data Analysis method as described on page 28.
     7 Click the Run Sequence button to start the run. The Run Sequence Status
       dialog box appears.
     8 If desired, you can monitor acquisition as described in online help.
     9 When the sequence is complete, analyze the data as described in the
       following section.
24                        G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide
Quantitative Data Analysis
                         The following topics will help you perform data analysis on your samples with
                         the Agilent 7700 Series ICP- MS MassHunter Workstation.
         Typical scenarios for Data Analysis
                                                                   Description
 To import an existing   The data was acquired with reference to a Data Analysis (DA) method.
       method            (E.g. the “Edit DA Method” button was clicked or the “Import DA Method from Existing Batch”
                         option was selected when setting up the sequence).
                         In this scenario, the ICP-MS Data Analysis window opens automatically after acquiring the data.
                         Proceed in one of the following ways:
                         • If no changes are required to the DA method, then proceed to “View analysis results” on
                             page 33 and to “Generate a report” on page 34.
                         • If you wish to modify the existing DA method for this acquired data, then proceed to “Create
                             the Data Analysis (DA) method for Quantitative data analysis” on page 28 to open the Method
                             Editor and make changes to the DA method.
 To create a method      The data was acquired without reference to a DA method.
after completing data    In this scenario, the ICP-MS Data Analysis window opens automatically after acquiring the data,
      acquisition        but only the file name is displayed, since no analysis method has yet been applied. Proceed as
                         follows:
                         a Open the method editor as described in “Create the Data Analysis (DA) method for Quantitative
                             data analysis” on page 28.
                         b Proceed to create a DA method for Quantitative data analysis.
  To create a method     You want to create a DA method without reference to any acquired data.
before performing data   a Follow the steps in“Open an existing batch for data analysis” on page 26 to create a new batch
      acquisition           folder, since a data batch does not already exist.
                         b If desired, import samples as described in Step 3.
                         c Proceed to “Create the Data Analysis (DA) method for Quantitative data analysis” on page 28.
G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide                                                               25
     Open an existing batch for data analysis
               If the window is not already open, open the ICP- MS Data Analysis window in
               either of the following ways:
               • Select Main Panel from the DataAnalysis menu in the ICP- MS Top window,
                 or
               • Click the       button on the ICP- MS Top toolbar.
         Tip   To open an existing batch:
                 a Select Open Analysis File from the File menu to open the Open Batch
                   dialog box.
                 b Select the file of interest and click Open. This opens an Analysis file with
                   data that has been previously acquired.
                 c Then proceed to “Create the Data Analysis (DA) method for
                   Quantitative data analysis” on page 28 to create a Data Analysis
                   (DA) method (if a DA method was not applied at time of data
                   acquisition) or to modify the method, if needed (if the DA method
                   was already applied).
     Creating a new batch for data acquisition
               1 Create a new Batch Folder, which will be used to contain your data analysis
                 results:
                 a Open the New Batch Folder dialog box in either of the following ways:
                 • Click the New Batch Folder toolbar button           , or
                 • Select New Batch Folder from the File menu.
                 b Type in a name for the new folder (you may select a different location if
                   desired).
                 c Click the Create button.
26                                G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide
                    2 Import the samples as follows:
                      a Select Import Samples from the File menu.
                      b Navigate to the data file folder where the data files of interest are located
                        and click the Open button. The data files in that folder are displayed on
                        the Import Samples dialog box.
                      c To import samples from another folder, click the Browse button.
                      d Repeat Steps b and c until all samples of interest are displayed on the
                        Import Samples dialog box.
                      e Select the samples to import:
                      • <Ctrl>+click to select multiple samples from the list.
                      • Click the Select All button to select all samples in the list.
                      f Click OK.
G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide                                      27
     Create the Data Analysis (DA) method for Quantitative data analysis
               1 Open the Method Editor window in either of the following ways:
                 • Click the Edit Method toolbar button         , or
                 • Select Edit from the Method menu.
         Tip     To import existing DA method, click Import Method only or Import
                 Method and Standard Data and select an analysis file to import.
               2 Select Data Analysis Method from Method Tasks section 2 (on the left side
                 of the screen) to display the Data Analysis Method pane
               3 Set the options as follows:
                 a Mark the check box for FullQuant Analysis.
                 b Set the Analysis Mode to Spectrum.
                 c Select the correction method from the Interference Correction list.
                                                                   3a
                                                                   3b
                                                                   3c
               4 Select a Sample Template from the list.
28                                G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide
                    5 Select Analyte List from Method Tasks section 3 to display the Analyte List
                      pane.
                    .
                    To open the analyte list from the existing data:
                        • Click the     button, or
                        • Right- click in the pane and select Load list from acquired data.
             Tip        You can also create an analyte list in either of the following ways:
                        • To load an analyte list from an acquisition method that has already been
                          developed, click the       button.
                      • To add new analyte lines, click the   button, then designate the Tune
                        Step, Mass, Name and Analyte/ISTD for each new line as described in
                        Step 6.
                    6 Enter the following information for each analyte:
                        a Click the   button at the right end of the Tune Step column, and select
                          a tune step number.
                        b Click the     button at the right end of the Analyte/ISTD column and
                          select either analyte or ISTD.
G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide                                   29
     7 Select Full Quant from Method Tasks section 4 to display the FullQuant
       Method Table pane:
     8 Set the following calibration curve parameters:
       • The calibration curve method (External Calibration / Standard Addition)
       • Type of VIS correction for ISTD elements
       • Curve fit type (Linear / Quadratic / Excluded)
       • Handling of the origin (Ignore / Force / Blank offset)
       • Calibration curve weight
       • ISTD elements used for correction
       • Concentration of the calibration curve level (Min Conc). For most
         applications, enter 0.
       • Concentration unit
       • Concentration of the blank
30                     G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide
                    9 The following Advanced Info method tasks are described in online help:
                      • FullQuant Outlier
                       • Edit Report Templates
                    10 Select Validate from Method Tasks section 5. Correct any method errors
                       before proceeding.
                    11 Select Return to Batch- at- a- Glance from Method Tasks section 5.
                       Click Yes to save the Data Analysis method.
G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide                              31
     Process the batch
               1 Initiate batch processing in either of the following ways:
                 • Click the Process Batch toolbar button                   , or
                 • Select Process Batch from the Process menu.
                 Analysis is performed on the samples you added to the batch in the
                 previous topic, using parameters set in the DA method. See “Create the
                 Data Analysis (DA) method for Quantitative data analysis” on page 28..
               2 Review the analysis results that are displayed in the Batch Table and
                 Calibration Curve panes, as shown in the following examples.
        Note   The Process Batch button changes to black after processing:                  .
32                                G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide
        View analysis results
     Batch Table    1 View the concentration and count for each element in the samples in the
           pane       Batch Table pane.
                      • Concentration and count for problematic data (Outlier) are displayed
                        with a color of each Outlier background.
                      • See online help for information if you want to change the way the data is
                        displayed, such as sorting or customizing the columns
                      • Also see online help for information on changing the layout or doing auto
                        review.
       Spectrum     2 View mass spectra in the Spectrum pane. You can zoom in to identify
           pane       elements by right- clicking and dragging the mouse cursor around the
                      desired mass number. See online help for information on other ways to
                      display or process data in the Spectrum pane, such as:
                      • Switching between log and linear
                      • Switching between one- row and three- row
                      • Adding comments
                      • Identifying unknown spectra against an element database
                      • Subtracting background spectra
                      • Overlaying multiple spectra
                      • Tabulating the spectral information
Calibration Curve   3 View calibration curves for each element in the samples in the Calibration
             pane     Curve pane.
                      • Problematic calibration curves (Outliers) are displayed with a pink
                        background.
                      • See online help for information on other changes you can make to the
                        calibration curve, such as excluding a calibration curve level, changing
                        the concentration of a calibration curve level, changing the calibration
                        curve type, changing the handling of the origin, or changing the
                        weighting of calibration curve.
   ISTD Stability   4 View the percent recovery of each ISTD element in the ISTD Stability Graph
     Graph pane       pane. See online help for more information.
G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide                                  33
     Save analysis results
              Be sure to always save the analysis results in one of the following ways:
              As a new file: Select Save Analysis File As from the File menu.
              To overwrite an existing results file with the same name:
                • Click the Save toolbar button           , or
                • Select Save from the File menu.
     Generate a report
              1 Select Generate from the Report menu in the MassHunter ICP- MS Analysis
                window.
              2 Mark the desired report options and click OK (see example below).
              3 Click OK to generate the report.
34                                 G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide
Semi-Quantitative Analysis
                    Semi- quantitative analysis allows you to quickly scan and obtain information
                    about the concentrations of all elements that are present in an unknown
                    sample. It is useful when you want to obtain concentration information for a
                    large number of elements (>70) without external calibration, such as screening
                    prior to quantitative analysis on target elements. Typically, semi- quantitative
                    analysis is accurate to +/- 30% or better on completely unknown samples.
        Prepare for SemiQuant Analysis
                    Prepare the following samples and solutions for analysis:
                    • Samples (dilute with nitric acid to approx. 1%)
                    • Tuning solution (Li, Co, Y, Ce, Tl, 1ppb each, with approx. 1% nitric acid)
                    • Cleaning solution (1 to 5% nitric acid)
        Create a SemiQuant acquisition method
                    1 Select Edit Entire Method from the Methods menu. This initiates a series
                      of dialog boxes, so you can set the parameters for your acquisition method.
                      See the online help for a full description of the items on these dialog boxes.
                    2 On the Edit Method dialog box, mark the following options, then click OK.
                    3 On the Method Information dialog box, set the following parameters, then
                      click OK:
                      a Enter a comment
                      b Be sure that the Data Analysis check box is checked; this is required to
                        run the data analysis automatically after data acquisition.
G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide                                     35
     4 On the Select Sample Types (or Select QC Items, if the Intelligent Sequence
       software is installed) dialog box, set the following parameters, then click
       OK:
       a Configure the list of Selected QC Items on the right to contain Sample,
         SQStd, SQBlk, and SQISTD.
       b Depending on the method you started with, you may have to use the
         Add- > and <- Remove buttons to create the desired list.
     5 Set the Data Acquisition method parameters:
       a On the Acquisition Mode dialog box, mark the Spectrum option and click
         OK.
       b On the Spectrum Acquisition Parameters dialog box, click the Mass
         Scale button.
       c On the Masses dialog box:
       • Click the Clear All button.
       • Select all mass numbers by double- clicking on the top, middle, and
         bottom rows, then click OK.
       d Make the following additional selections on the Spectrum Acquisition
         Parameters dialog box, then click OK:
       • Select Semi Quant (6) for the Peak Pattern.
       • Clear the Set every Mass option, then set the per Point Integration
         time to 0.05 or 0.1 sec.
       • Set the Repetition count to 1.
       e On the Peristaltic Pump Program dialog box, set the following
         parameters, then click OK.
       • Before Acquisition: Uptake Speed: 0 rps, Uptake Time: 0 sec,
         Stabilization Time: 70 sec
       • After Acquisition (Rinse Port): Rinse time: 30 sec
       • After Acquisition (Rinse Vial)
       See online help for more information on these parameters and on the
       optional features: Intelligent Rinse and Execute Pre- emptive Rinse.
     6 To save the method:
       a On the Method Save Options dialog box, mark all items, and click OK.
       b On the Save Method As dialog box, enter a name for the method, and
         click OK.
36                      G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide
        Analyze the samples
                    1 In the sequence table, set up a sequence with samples in the following
                      order:
                      a Tuning solution (1 ppb Li, Co, Y, Ce, Tl) - Sample Type is “SQStd”
                      b Blank (i.e. nitric acid) - Sample Type is “SQBlk”
                      c Sample 1 - > Sample 2 - > Sample 3 ... - Sample Type is “Sample”
                    2 Run the sequence:
                      a Select Run from the Sequence menu or click the                         icon on the toolbar
                        to open the Start Sequence dialog box.
                      b Click the Run Sequence button to start the run. The Run Sequence
                        Status dialog box appears.
                      c If desired, you can monitor acquisition as described in online help.
                     The same scenarios apply for offline Semi-quantitative data analysis as described in
     NOTE            “Typical scenarios for Data Analysis” on page 25.
        Create a new batch or open a new batch for SQ data analysis
                    Refer to “Open an existing batch for data analysis” on page 26 and
                    “Creating a new batch for data acquisition” on page 26.
G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide                                                   37
     Create the Data Analysis (DA) method for SemiQuant data analysis
               1 Open the Method Editor window in either of the following ways:
                 • Click the Edit Method toolbar button        , or
                 • Select Edit from the Method menu.
         Tip     To import existing DA method, click Import Method only or Import
                 Method and Standard Data and select an analysis file to import.
               2 Select Data Analysis Method from Method Tasks section 2 (on the left side
                 of the screen) to display the Data Analysis Method pane.
               3 Set the options as shown below:
                 a Mark the check box for SemiQuant Analysis.
                 b Set the Analysis Mode to Spectrum.
                 c Select the correction method from the Interference Correction list.
                                                                   3a
                                                                   3b
                                                                   3c
               4 Select a Sample Template from the list.
               5 Select SemiQuant from Method Tasks section 3 to display the SemiQuant
                 pane.
38                                G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide
             Tip    You can select elements to show in the pane by clicking Add/Remove
                    Standard Element.
     6a
                    6 On the SemiQuant Method Table pane:
                      a Mark the check box for Advanced Parameters.
                      b Set the concentration of the elements in the standard sample being used
                        for the SemiQuant factor correction.
                      c Select the ISTD correction method.
                      For setting details, refer to “SemiQuant Pane” in the “Commands
                      Reference” section of online help.
                    7 Select Validate from Method Tasks section 5. Correct any method errors
                      before proceeding.
                    8 Select Return to Batch- at- a- Glance from Method Tasks section 5.
                      Click Yes to save the Data Analysis method.
G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide                                39
     Process the batch
              1 Initiate batch processing in either of the following ways:
                • Click the Process Batch toolbar button                   , or
                • Select Process Batch from the Process menu.
              Analysis is performed on the samples you added to the batch in the
              previous topic, using parameters set in the Data Analysis method, as
              described in “Create the Data Analysis (DA) method for SemiQuant data
              analysis” on page 38.
              2 View/check the analysis results as follows:
                a Batch Table pane- - Displays the concentration and count for each
                  element in the samples.
                b Spectrum pane- - Displays mass spectra. You can zoom in to identify
                  elements by right- clicking and dragging the mouse cursor around the
                  desired mass number.
                c Semiquant Factor pane- - Displays the SemiQuant factor graph for each
                  sample. Place the cursor near a point on a graph to display element
                  names and SemiQuant factor correction details. Use the arrow keys to
                  move between samples. If necessary, the factors can be changed as
                  described in “Change the SemiQuant Factors” on page 42.
                  If the SemiQuant Factor pane is not currently displayed, select
                  Calibration Curve from the View menu.
                d ISTD Stability Graph pane- - Displays the percent recovery of each ISTD
                   element in the samples. See online help for more information.
              3 Save results using Save Analysis File As or Save from the File menu.
              4 Select Generate from the Report menu, select report options, and click OK.
40                               G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide
        Correcting the Semiquant Factors
                    On MassHunter Workstation, the SemiQuant factors are predefined for each
                    element. The greater the SemiQuant factor, the more efficiently that element is
                    ionized in the plasma, yielding a higher sensitivity. Use the following
                    procedure to correct the SemiQuant factors, as necessary.
                    1 Select and measure standard sample containing three or four elements
                       whose concentration are known. Make sure that the concentrations of the
                       elements in the standard sample cover the low, medium, and high ranges of
                       the mass spectrum, or use the tuning solution that contains Li, Co, Y, Ce
                       and TI as the standard sample.
                    2 In the ICM- MS Data Analysis window, load the data for the standard
                       sample into the batch folder.
                    3 Set the sample type of the loaded Standard Sample to SQStd as shown in
                       the example below:
                    4 Process the batch in either of the following ways:
                      • Click the Process Batch toolbar button                   , or
                      • Select Process Batch from the Process menu.
                      The SemiQuant factor for all elements will be corrected based on the
                      elements of known concentration in the standard sample (SQStd).
G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide                                    41
     Change the SemiQuant Factors
             You can configure the SemiQuant factor, Base mass, and Reported mass
             (isotopes).
             1 Select SemiQuant Basic Parameters from the Global drop- down menu in
                the ICP- Analysis window:
               The SemiQuant Basic Parameters dialog box opens.
             2 Make the change by either:
               • Right- clicking within the table and selecting from the context menu; or
               • Selecting a cell and directly editing its content.
               Refer to the Help for more details.
42                              G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide
Notices
© Agilent Technologies, Inc. 2009               Technology Licenses
No part of this manual may be reproduced in
                                                                                                     CAUTION
                                                The hardware and/or software described in
any form or by any means (including elec-       this document are furnished under a license
tronic storage and retrieval or translation
                                                                                                 A CAUTION notice denotes a haz-
                                                and may be used or copied only in accor-
into a foreign language) without prior agree-                                                    ard. It calls attention to an operat-
                                                dance with the terms of such license.
ment and written consent from Agilent                                                            ing procedure, practice, or the like
Technologies, Inc. as governed by United
States and international copyright laws.
                                                Restricted Rights Legend                         that, if not correctly performed or
                                                If software is for use in the performance of a   adhered to, could result in damage
Manual Part Number                              U.S. Government prime contract or subcon-        to the product or loss of important
G7200-90210
                                                tract, Software is delivered and licensed as     data. Do not proceed beyond a
                                                "Commercial computer software" as                CAUTION notice until the indicated
                                                defined in DFAR 252.227-7014 (June 1995),
Edition                                         or as a "commercial item" as defined in FAR      conditions are fully understood and
Rev. A, June 2009                               2.101(a) or as "Restricted computer soft-        met.
                                                ware" as defined in FAR 52.227-19 (June
Printed in USA                                  1987) or any equivalent agency regulation or
Agilent Technologies, Inc.                      contract clause. Use, duplication or disclo-
Hewlett-Packard-Strasse 8                       sure of Software is subject to Agilent Tech-            NOTE
76337 Waldbronn, Germany                        nologies' standard commercial license
                                                terms, and non-DOD Departments and               A NOTE contains helpful informa-
Warranty                                        Agencies of the U.S. Government will             tion on the usage - it does not
                                                receive no greater than Restricted Rights as
The material contained in this docu-            defined in FAR 52.227-19(c)(1-2) (June           denote a hazard.
ment is provided "as is," and is sub-           1987). U.S. Government users will receive
ject to being changed, without notice,          no greater than Limited Rights as defined in
in future editions. Further, to the max-        FAR 52.227-14 (June 1987) or DFAR                Agilent products must only be used in the
imum extent permitted by applicable             252.227-7015 (b)(2) (November 1995), as          manner described in the Agilent product
law, Agilent disclaims all warranties,          applicable in any technical data.                user guides. Any other use may result in
either express or implied, with regard                                                           damage to the product or personal injury.
                                                Safety Notices                                   Agilent is not responsible for any damages
to this manual and any information
                                                                                                 caused, in whole or in part, by improper use
contained herein, including but not
                                                                                                 of the products, unauthorized alterations,
limited to the implied warranties of                                                             adjustments or modifications to the prod-
merchantability and fitness for a par-             WA R N I N G                                  ucts, failure to comply with procedures in
ticular purpose. Agilent shall not be                                                            Agilent product user guides, or use of the
liable for errors or for incidental or          A WARNING notice denotes a                       products in violation of applicable laws,
consequential damages in connec-                hazard. It calls attention to an                 rules or regulations.
tion with the furnishing, use, or per-
formance of this document or of any
                                                operating procedure, practice, or
information contained herein. Should            the like that, if not correctly per-
Agilent and the user have a separate            formed or adhered to, could result
written agreement with warranty                 in personal injury or death. Do not
terms covering the material in this             proceed beyond a WARNING
document that conflict with these
terms, the warranty terms in the sep-
                                                notice until the indicated condi-
arate agreement shall control.                  tions are fully understood and
                                                met.
www.agilent.com
In this Book
This Quick Start Guide
presents first steps to
using the Agilent 7700
Series ICP- MS MassHunter
Workstation.
© Agilent Technologies, Inc. 2009
Printed in USA
Rev. A, June 2009
*G7200-90210*
G7200-90210
                       Agilent Technologies