Function K1301 K1302 K1305A K1305B K1306 K06A K06B K06C K12A K12B K12C
Human-machine interface
X √ √ √ √ √ √ √ √ √ √
(color touch screen)
Add alkali liquid Manual Auto Auto Auto Auto Auto Auto Auto Auto Auto Auto
Add absorption liquid Manual Manual Auto Auto Auto Auto Auto Auto Auto Auto Auto
Safety door in place
X √ √ √ √ √ √ √ √ √ √
detection
Digestive tube in place
X √ √ √ √ √ √ √ √ √ √
detection
Cooling water detection
X X X X X √ √ √ √ √ √
and alarm
Steam flow control X X X √ √ √ √ √ √ √ √
Over-temperature alarm X X X X X √ √ √ √ √ √
Auto titrate X X X X √ √ √ √ √ √ √
Auto calculate result X X X X √ √ √ √ √ √ √
Auto store data X X X X √ √ √ √ √ √ √
Nitrate nitrogen
X X X √ X X √ √ X √ √
determining module
High nitrogen
X X X X X X X √ X X √
determining module
Auto waste discharge X X X X X X X X √ √ √
Reagent level detection X X X X √ √ √ √ √ √ √
USB interface X X X X X X X X √ √ √
Computer
X X X X X √ √ √ √ √ √
communication interface
Wireless LAN
X X X X X X X X √ √ √
transmission
“√” represent available; “X” represent unavailable
Parameter K1301 K1302 K1305A K1305B K1306 K06A K06B K06C K12A K12B K12C
Measuring range 0.1mg-240mg N(general nitrogen & nitrate nitrogen); 0.1mg-280mg N(high nitrogen)
Analysis time 3-8min/sample(general nitrogen); 10-18min/sample(nitrate nitrogen); 4-8min/sample(high nitrogen)
Reproducibility 0.5% relative error
≥99%(distilled
water)
Recovery ≥99.5% ≥99.8% ≥99.5% ≥99.8%
≥98%(tap
water)
Titration accuracy --- 2.0μL/step 1.0μL/step
Sample size Solid≤5g; liquid≤25ml
Cooling water
consumption in the 1.5L/min
distillation process
1000
Data storage capacity --- groups
2000 groups
Power supply AC220V±10% 50HZ
Power consumption 1800W
Net weight 30kgs 35kgs
Dimentions(mm) 420 X 365 X 720 460 X 450 X 720