PRACTICAL
BE VOCAL ABOUT LOCAL
Kjeldahl Distillation
RUGGED
SMART
SAFE SIMPLE EFFICIENT
DIGESTION SYSTEMS
for Labs
Comfortable and reliable Labquest steam distillation unit
designed predominantly
for Kjeldahl analysis
Perform Kjeldahl nitrogen or protein determinations in the Food & Feed industries and
several other applications in environmental control, agricultural produce, products, soil
industries after having digested the sample accurately in the most cost-effective way
KDI050 is compliant with many official methods of steam distillation for different
applications such as the determination of ammoniacal nitrogen, protein determination,
nitrogen content (direct or indirect), nitric nitrogen (after reduction/Devarda’s Alloy) and
alcohol content.
EFFORTLESS USE ENERGY SAVING COMPETENCE SAFETY
Simple programming Start and stop Robust acid Brilliant
and straight forward cooling water resistant housing supervision of all
operation automatically Compact crucial functions
during the footprint saves
Quick start option Comprehensive
distillation useful laboratory
with favourite built-in safety
process with bench space.
method menu systems for
excellent
operator
insulation of
protection
internal parts.
2
KDI050 is the unique range solution to quantify the nitrogen/protein content with the Kjeldahl
method. The automated Kjeldahl analyzer, with premium performance offers the highest
sample throughput with increased boiler capacity and is the most productive system
available.
Sample analysis requires no monitoring.
The inbuilt N/Protein calculator with extraction of results makes the system versatile like never
before
Automatic dilution, steam, alkali, acid
and test tube drain
Anti corrosive FRP body construction
7’’ colored touch screen display
Steam regulation (10-100%)
Process kill switch to halt the process
in case of accidental door opening
Process resume from stop
Detachable bottom drip tray for extra
protection against acid attack
Anytime alkali addition during
the distillation cycle
Tube sensor that enables safe operating
conditions in event of tube damage
Reagent refill level sensor to avert
reagent dry out
One point calibration of alkali and
distilled water pumps
Two point calibration for acid pump
Real time monitoring of process parameters
SELV protection
Nitrogen reproducibility ±1 %
RSD <1%
LOD of 1-200mg with of ≥0.1mg of Nitrogen
Nitrogen recovery >99.5%
3
Technical details
Parameters KDI050
Tube Capacity (ml) 250ml/ 400ml
Minimum distillation time (mins) 5
Current Consumption (A) 10
Power consumption (W) 1500
Dimension (L X B X H)mm (480 X 470 X 820)
HSN Code 84198990
Ordering Code BLFAKDI05000000000
Accessories
Description Code
KDI Drip Tray ASM BLA0KDI000DRP0TRAY
Oven Mittes - KDI ASM BLA0OVM000KDI00000
Polypropylene Aspirator Bottle - 34101 ASM BLA0ASBPP050034101
BOROSIL LIMITED
LIMITED