DAR-9500f
 Installation Manual
This document is for the use of specialist service personnel.
People untrained should refrain from using this document.
Thank you for purchasing a SHIMADZU medical system. Before using the system,
please read this manual thoroughly and use the system correctly.
      The precautions and prohibitions seen through the manual are classified as
      follows:
DISCLAIMER
     ● Images and illustrations in this manual may be different from the actual
       display. All images and illustrations are provided as examples.
     ● Some parts of illustrations may be omitted.
     ● The contents of this manual are subject to change without notice due to
       continual improvements.
     ● The contents of this manual are made assurance doubly sure on the creation,
       however, if there is any chance of mistakes or omission of recording, they
       may not be modified immediately.
                             HISTORY
REV      DATE                            CONTENT OF CHANGE
 A    2011/08/12                         ---------------------------------
 B    2012/10/18   Chapter 3 applies to Fujitsu PC.
                   Chapter 4 applies to Windows 7.
                   Chapter 12 applies to Ghost 15.0.
 C    2013/07/11   Applies to Bi-plane system.
                   Chapter 5 applies to Ver.5.2.X
                   Chapter 13 applies to PC Backup S/W
                   Add chapter 15 Stent View (Option).
                   Add appendix.
D     2013/11/11   Chapter 3 and 4 apply to DVI-MAIN3.
E     2013/12/20   Chapter 3, 4 and 8 change input channel of REF PC.
F     2014/06/04   Chapter 3, 4 and 15 applies to SCORE PRO Advance.
                   Add chapter 3 OMRON UPS.
G     2014/11      Applies to M730 (Chapter 3, 4, 11, 15 and Appendix D)
                   Add description of DVI MAIN3 in Chapter 3 and 4.
H     2015/2       Chapter 3 and 4 apply to DVI-OUT BOX.
J     2015/3       Applies to DAR Ver.6 (Chapter 3, 5, 10, and Chapter 11)
                   Add chapter 16 UserQC (Option)
K     2015/3       Delete Appendix A. Updating ACQ System S/W
                   Delete Appendix B. Updating REF/VRS System S/W
L     2015/6       Chapter 5 applies to Geometry Configuration
M     2015/7       Add description of Geometry Configuration in chapter 5.
                   Add description of System Backup Instruction in chapter 13.
                   Delete Appendix A I-SYS Update.
                   Add Appendix A Apply UPG1 for PC Replacement.
N     2015/12      Add Appendix C Troubleshooting of FPD.
This page is intentionally left blank.
                                     TABLE OF CONTENTS
1. INTRODUCTION ......................................................................................................... 1-1
2. INSTALLATION/ENVIRONMENTAL CONDITION .................................................. 2-1
   2.1 POWER SOURCE ................................................................................................................. 2-1
   2.2 TEMPERATURE AND HUMIDITY ......................................................................................... 2-1
   2.3 GROUNDING ...................................................................................................................... 2-1
   2.4 GENERATED HEAT QUANTITY ............................................................................................ 2-1
3. INSTALLATION OF DAR-9500F................................................................................. 3-1
   3.1 INSTALLATION TOOLS ....................................................................................................... 3-1
   3.2 SYSTEM CONFIGURATION OF DAR-9500F (ACQ3 CABINET) ......................................... 3-2
     3.2.1 Configuration of Single-plane System .................................................................... 3-2
     3.2.2 Configuration of Optional PC (3D Work Station etc.) System .............................. 3-4
   3.3 NAMES OF EACH UNIT AND COMPONENT ......................................................................... 3-5
     3.3.1 Constraction of ACQ3 Cabinet ................................................................................ 3-6
   3.4 CONFIRMATION OF IVR NEO TYPES ................................................................................ 3-8
   3.5 CONFIRMATION OF DVI-MAIN TYPE.............................................................................. 3-11
   3.6 CHECK UPS TYPE ........................................................................................................... 3-11
   3.7 INSTALLATION OF DAR-9500F ........................................................................................ 3-12
     3.7.1 Installation of ACQ3 Cabinet ................................................................................ 3-12
     3.7.2 Installation of REF Cabinet .................................................................................. 3-13
     3.7.3 Installation of DVI OUT (In case of MTA-40C) .................................................... 3-13
     3.7.4 Installation of DVI OUT (In case of Monitor Cart 2) ........................................... 3-14
     3.7.5 Installation of DVI OUT (In case of DVI OUT Box) ............................................ 3-15
     3.7.6 Installtion of FPD ................................................................................................... 3-15
   3.8 CABINET FIXING ............................................................................................................. 3-16
   3.9 CONNECTION DPENDING ON SUPPLY VOLTAGE .............................................................. 3-17
   3.10 CABLE CONNECTION (EXTERNAL CONNECTION) ..................................................... 3-18
     3.10.1 Connection Diagram ............................................................................................ 3-18
     3.10.2 Cable Connection (Power / Earth) ....................................................................... 3-19
     3.10.3 Cable Connection (DVI / Monitor) ....................................................................... 3-20
     3.10.4 Cable Connetion (REF Cabinet) .......................................................................... 3-27
     3.10.5 Cable Connection (FPD / FPD Chiller) ............................................................... 3-28
     3.10.6 Cable Connection (JOINT BOX 2/System Display) ........................................... 3-29
     3.10.7 Cable Connection (Information Display) ............................................................ 3-30
     3.10.8 Cable Connection (IVR NEO) .............................................................................. 3-31
     3.10.9 Cable Connection (D150GC-40) ........................................................................... 3-34
     3.10.10 Cable Connection to Polygraph ......................................................................... 3-35
   3.11 ACQ3 CABINET INTERNAL CABLE CONNECTION .......................................................... 3-36
     3.11.1 Conncetion Diagram ............................................................................................ 3-36
     3.11.2 Internal Connection (CP2) ................................................................................... 3-39
     3.11.3 Internal Connection (SYSDISP PC) .................................................................... 3-39
     3.11.4 Internal Connection (Maintenance PC(MPC)) ................................................... 3-39
     3.11.5 Internal Conneciton (HUB/ROUTER) ................................................................ 3-40
     3.11.6 Internal Connection (UPS) .................................................................................. 3-40
   3.12 INTERNAL CABLE CONNECTION (INFORMATION) .......................................................... 3-41
     3.12.1 ACQ3 CABINET ................................................................................................... 3-41
     3.12.2 REF CABINET ..................................................................................................... 3-43
   3.13 CONFIRM DIP SW SETTING .......................................................................................... 3-44
     3.13.1 I-SYS BOARD ....................................................................................................... 3-44
     3.13.2 DVI MAIN2 BOARD ............................................................................................ 3-46
     3.13.3 DVI MAIN3 BOARD ............................................................................................ 3-49
                                                                     i                                                    M517-E610
    3.13.4 IVR NEO I/F, PERIP-SH BOARD ...................................................................... 3-55
    3.13.5 IVR NEO I/F2, PERIP-SH2 BOARD .................................................................. 3-57
    3.13.6 IVR I/F5 BOARD .................................................................................................. 3-60
  3.14 INSTALLATION OF LIVE MONITOR OPTION .................................................................. 3-61
    3.14.1 LIVE Monitor in Examination Room(Optional) ................................................. 3-61
    3.14.2 LIVE Monitor in Control Room (Optional) ......................................................... 3-65
  3.15 LAYOUT DRAWING OF PCB............................................................................................ 3-69
    3.15.1 I-SYS BOARD ....................................................................................................... 3-69
    3.15.2 DVI MAIN2 BOARD ............................................................................................ 3-70
    3.15.3 DVI MAIN3 BOARD ............................................................................................ 3-71
    3.15.4 IVR NEO I/F / PERIP-SH BOARD ..................................................................... 3-72
    3.15.5 IVR NEO I/F2 / PERIP-SH2 BOARD ................................................................. 3-73
  3.16 LED .............................................................................................................................. 3-74
    3.16.1 I-SYS BOARD ....................................................................................................... 3-74
    3.16.2 DVI MAIN2 BOARD ............................................................................................ 3-76
    3.16.3 DVI MAIN3 BOARD ............................................................................................ 3-78
    3.16.4 IVR NEO I/F / PERIP-SH BOARD ..................................................................... 3-80
    3.16.5 IVR NEO I/F2 / PERIP-SH2 BOARD ................................................................. 3-81
4. INSTALLATION OF DAR-9500F BI-PLANE SYSTEM (ACQ3L CABINET) ............. 4-1
  4.1 SYSTEM CONFIGURATION OF DAR-9500F......................................................................... 4-2
    4.1.1 Configuration of ACQ3 CABINET .......................................................................... 4-4
    4.1.2 Configuration of ACQ3 CABINET .......................................................................... 4-5
  4.2 INSTALLATION OF DAR-9500F .......................................................................................... 4-6
    4.2.1 Installation of ACQ3 Cabinet .................................................................................. 4-6
    4.2.2 Installation of ACQ3L Cabinet ................................................................................ 4-6
    4.2.3 Installation of DVI OUT (In case of MTA-40C) ...................................................... 4-7
    4.2.4 Installation of DVI OUT (In case of DVI OUT Box) .............................................. 4-7
    4.2.5 Installtion of FPD ..................................................................................................... 4-7
  4.3 CABINET FIXING ............................................................................................................... 4-8
  4.4 CONNECTION DPENDING ON SUPPLY VOLTAGE ................................................................ 4-9
  4.5 CABLE CONNECTION (EXTERNAL CONNECTION) ....................................................... 4-10
    4.5.1 Connection Diagram .............................................................................................. 4-10
    4.5.2 Cable Connection (Power / Earth) ......................................................................... 4-11
    4.5.3 Cable Connection (DVI / Monitor) ......................................................................... 4-12
    4.5.4 Cable Connection (FPD / FPD Chiller) ................................................................. 4-18
    4.5.5 Cable Connection (ACQ3 CABINET-ACQ3L CABINET) .................................... 4-19
    4.5.6 Cable Connection (JOINT BOX 2/System Display) ............................................. 4-20
    4.5.7 Cable Connection (D150GC-40)............................................................................. 4-20
    4.5.8 Arcnet Cable Connection Diagram........................................................................ 4-22
    4.5.9 Bi-Plane Network Connection Diagram ............................................................... 4-22
  4.6 ACQ3L CABINET INTERNAL CABLE CONNECTION.......................................................... 4-23
  4.7 ACQ3L CABINET INTERNAL CABLE CONNECTION (CONNECTED AT FACTORY SETTING) 4-25
  4.8 ACQ3/REF CABINET INTERNAL CABLE CONNECTION (REFERENCE).......................... 4-25
  4.9 CONFIRMATION OF DIP SW SETTING ............................................................................. 4-26
    4.9.1 I-SYS BOARD (L) ................................................................................................... 4-26
    4.9.2 DVI MAIN2 BOARD (L) ........................................................................................ 4-28
    4.9.3 DVI MAIN3 BOARD(L) ......................................................................................... 4-31
    4.9.4 IVR NEO I/F, PERIP-SH BOARD ........................................................................ 4-35
    4.9.5 IVR NEO I/F2, PERIP-SH2 BOARD .................................................................... 4-36
  4.10 INSTALLATION OF LIVE MONITOR OPTION .................................................................. 4-39
    4.10.1 LIVE Monitor in Examination Room(Optional) ................................................. 4-39
    4.10.2 LIVE Monitor in Control Room (Optional) ......................................................... 4-43
5. REF PC AND ACQ PC CONFIGURATION ................................................................. 5-1
M517-E610                                                            ii
  5.1 DAR-9500F NETWORK CONNECTION PRE-CONFIRMATION SHEET................................. 5-1
  5.2 HOW TO STARTUP AS CONFIGURATION MODE .................................................................... 5-3
    5.2.1 Logon as super, serviceapp ...................................................................................... 5-3
  5.3 PC CONFIGURATION .......................................................................................................... 5-7
    5.3.1 Date and time configuration .................................................................................... 5-7
    5.3.2 Computer name configuration ................................................................................. 5-9
    5.3.3 IP address configuration ........................................................................................ 5-11
    5.3.4 Regional and language configuration ................................................................... 5-16
    5.3.5 Configuration of remote desktop to MPC ............................................................. 5-25
  5.4 CONFIGURATION OF OUTPUT FILE DESTINATION FOR IMAGES AND LOOPS ...................... 5-28
  5.5 CONFIGURATION OF DAR-9500F .................................................................................... 5-30
    5.5.1 Executing application ............................................................................................ 5-30
    5.5.2 Bi-plane option ....................................................................................................... 5-67
  5.6 PRINTER DEFINITION FILE ............................................................................................. 5-70
    5.6.1 About a Tool for creating Printer Definition File ................................................. 5-70
    5.6.2 How to use this tool ................................................................................................ 5-71
    5.6.3 Configuration Direction ......................................................................................... 5-75
  5.7 PRINT SMPTE IMAGE ..................................................................................................... 5-78
  5.8 DICOM PRINTER CONFIGURATION (PRINTING INFORMATION ON THE FILM) ................ 5-80
  5.9 CONFIGURATION OF MPPS ............................................................................................. 5-81
  5.10 CONFIGURATION OF STORAGE COMMITMENT ............................................................... 5-83
    5.10.1 Configuration of the port number of DAR9500f ................................................. 5-83
    5.10.2 Configuration of Inbound Settings ...................................................................... 5-84
  5.11 SYNERGY CONFIGURATION............................................................................................ 5-86
    5.11.1 Setting ACQ PC F ................................................................................................ 5-86
    5.11.2 Setting ACQ PC L (For Bi-plane System) .......................................................... 5-89
    5.11.3 REF PC ................................................................................................................. 5-92
  5.12 SETTING THE IMAGE FOR AUTO TRANSFER................................................................... 5-99
    5.12.1 Reference Image ................................................................................................... 5-99
  5.13 TROUBLE SHOOTING ................................................................................................... 5-102
    5.13.1 DICOM Connection Test .................................................................................... 5-102
    5.13.2 Investigation Procedure for Troubles ............................................................... 5-104
  5.14 CUSTOMIZED ITEMS .................................................................................................... 5-106
    5.14.1 Setting the Image Type for Auto Transfer ....................................................... 5-106
    5.14.2 White Line During Printing .............................................................................. 5-106
    5.14.3 Display or Hide DSA Tool .................................................................................. 5-106
6. ROUTER CONFIGURATION ...................................................................................... 6-1
  6.1 LOGIN................................................................................................................................ 6-1
  6.2 TIME CONFIGURATION ...................................................................................................... 6-2
  6.3 WAN IP ADDRESS CONFIGURATION ................................................................................. 6-3
  6.4 FIREWALL CONFIGURATION METHOD ............................................................................... 6-4
  6.5 INITIALIZATION METHOD .................................................................................................. 6-8
  6.6 LAN IP ADDRESS CONFIGURATION .................................................................................. 6-9
  6.7 INDIVIDUAL SETTINGS .................................................................................................... 6-11
7. CONFIGURATION OF IP ADDRESS OF I-SYS AND CP2 ........................................ 7-1
  7.1 CONFIGURATION OF I-SYS IP ADDRESS ........................................................................... 7-2
  7.2 CONFIGURATION OF CP2 IP ADDRESS .............................................................................. 7-4
  7.3 CONFIGURATION OF I-SYS F/W WRITER .......................................................................... 7-7
8. DVI MAIN CONNECTION CONFIGURATION .......................................................... 8-1
  8.1 DVI MAIN CONNECTION CONFIGURATION ...................................................................... 8-1
  8.2 SETTING LIVE MONITOR OPTION .................................................................................. 8-11
  8.3 CHAGE THE SETTING OF I-SYS BOARD ......................................................................... 8-12
                                                                     iii                                                     M517-E610
9. FPD COOLING SYSTEM (HEC002-A5B) .................................................................... 9-1
   9.1 OUTLINE ........................................................................................................................... 9-1
   9.2 SPECIFICATION.................................................................................................................. 9-1
   9.3 INSTALLATION ................................................................................................................... 9-2
   9.4 DAILY CHECK ITEM ............................................................................................................ 9-7
   9.5 MAINTENANCE .................................................................................................................. 9-8
   9.6 CONNECTING CHART ......................................................................................................... 9-9
   9.7 EXTERNAL FIGURE .......................................................................................................... 9-10
10. ADJUST OF VIRTUAL COLLIMATION ................................................................... 10-1
   10.1 OUTLINE ....................................................................................................................... 10-1
   10.2 OPERATION SPECIFICATION .......................................................................................... 10-1
   10.3 ADJUSTMENT ................................................................................................................ 10-2
     10.3.1 Adjust of H and V Shutter ................................................................................... 10-3
     10.3.2 Adjust C Shutter .................................................................................................. 10-7
   10.4 CHECK SHEET ............................................................................................................. 10-11
11. INSTALLATION DOSE METER (OPTION) .............................................................. 11-1
   11.1 OUTLINE........................................................................................................................ 11-1
   11.2 COMPONENTS ................................................................................................................ 11-1
   11.3 INSTALLATION ............................................................................................................... 11-3
     11.3.1 Cable Connections ................................................................................................ 11-3
     11.3.2 Connecting to C-arm ............................................................................................ 11-4
     11.3.3 Mounting to C-arm (8 inch) ................................................................................. 11-5
     11.3.4 Connecting to ACQ PC (Already wired at factory)............................................. 11-6
   11.4 SETTING EXPOSURE PARAMETERS TAB ......................................................................... 11-7
   11.5 OPERATION CHECK ....................................................................................................... 11-8
   11.6 DOSE RATE AND CUMULATIVE DOSE DISPLAY ................................................................ 11-9
     11.6.1 Outline .................................................................................................................. 11-9
     11.6.2 Where dose rate and cumulative dose are measured ........................................ 11-9
     11.6.3 Operating Method .............................................................................................. 11-10
     11.6.4 Rejection limit .................................................................................................... 11-11
12. INSTALLATION SCAN CONVERTER (OPTION) .................................................... 12-1
   12.1 COMPONENTS................................................................................................................ 12-1
   12.2 INSTALLATION ............................................................................................................... 12-2
     12.2.1 Cable Connections ................................................................................................ 12-2
     12.2.2 Connecting REF Cabinet ..................................................................................... 12-3
     12.2.3 Dip SW Configuration of DVI MAIN MODULE ................................................ 12-4
   12.3 CONFIGURATION OF INPUT/OUTPUT CHANNEL ............................................................ 12-5
   12.4 OPERATION CHECK ....................................................................................................... 12-8
13. SYSTEM BACKUP AND RESTORE INSTRUCTION ............................................... 13-1
   13.1 INTRODUCTION ............................................................................................................. 13-1
   13.2 CONFIRMATION OF BACKUP S/W ................................................................................ 13-3
   13.3 DAR-9500F BACKUP S/W .......................................................................................... 13-3
     13.3.1 Creating System Backup ..................................................................................... 13-3
     13.3.2 Restore of the System ........................................................................................ 13-21
     13.3.3 In Case of Replacing HDD ................................................................................. 13-33
   13.4 PC BACKUP G11 S/W ............................................................................................... 13-50
     13.4.1 Creating System Backup ................................................................................... 13-50
     13.4.2 Verify BACKUP File .......................................................................................... 13-68
     13.4.3 Restore of the System ........................................................................................ 13-75
14. STENTVIEW (OPTION)............................................................................................. 14-1
M517-E610                                                            iv
   14.1 INTRODUCTION ............................................................................................................. 14-1
   14.2 MOUTING STENTVIEW DONGLE.................................................................................... 14-1
   14.3 SETTING DUP ............................................................................................................... 14-2
   14.4 SETTING IVR NEO ....................................................................................................... 14-3
   14.5 SETTING FUNCTION KEY .............................................................................................. 14-4
   14.6 CUSTOMIZING STENTVIEW ........................................................................................... 14-5
   14.7 SETTING GATEWAY ........................................................................................................ 14-6
   14.8 OPERATION CHECK ....................................................................................................... 14-6
     14.8.1 Check for Stent View Display .............................................................................. 14-6
     14.8.2 Operation Check for Marker Detection .............................................................. 14-6
15. ADDING BI-PLANE SYSTEM ................................................................................... 15-1
   15.1 COMPONENTS OF ADDITIONAL BI-PLANE SYSTEM........................................................ 15-2
   15.2 ADDING COMPONENTS FOR BI-PLANE TO ACQ3 CABINET ............................................ 15-3
     15.2.1 Adding ACQ3 Cabinet Door................................................................................. 15-3
     15.2.2 Adding ACQ3 Cabinet Transformer Unit ........................................................... 15-4
     15.2.3 Installation of HUB/CP2 ...................................................................................... 15-5
   15.3 ACQ3 CABINET INTERNAL CABLE CONNECTION .......................................................... 15-6
     15.3.1 Internal Connection (CP2) ................................................................................... 15-7
   15.4 CHANGING FRONTAL DVI CABLE CONNECTION ........................................................... 15-8
   15.5 CHANGING INSTALLATION AND SETTING OF OTHER THAN ACQ3 CABINET .................. 15-9
     15.5.1 Changing Power Supply Cabling of FPD Cooling Unit ..................................... 15-9
     15.5.2 Adding REF Cabinet Bi-Plane System ............................................................. 15-10
     15.5.3 Setting IVR NEO I/F, PERIP-SH BOARD ....................................................... 15-11
     15.5.4 Installation of ACQ3L Cabinet .......................................................................... 15-12
   15.6 SETTING DVI CONFIG ................................................................................................. 15-13
   15.7 CHANGING REF PC SETTING ..................................................................................... 15-14
     15.7.1 Display Setting ................................................................................................... 15-14
     15.7.2 Setting Synergy .................................................................................................. 15-16
   15.8 ACQ PC L NETWORK SETTING ................................................................................... 15-21
   15.9 LATERAL CP2 / I-SYS NETWORK SETTING ................................................................. 15-21
   15.10 SOFTWARE INSTALL................................................................................................... 15-21
   15.11 DAR-9500F APPLICATION SETTING .......................................................................... 15-21
   15.12 I-SYS UPDATE SOFTWARE INSTALL .......................................................................... 15-21
16. USERQC (OPTION) ................................................................................................... 16-1
   16.1 INTRODUCTION ............................................................................................................. 16-1
   16.2 SWITCH TO USER QC MODE ......................................................................................... 16-1
   16.3 OPERATION CHECK........................................................................................................ 16-1
   16.4 SWITCH TO NORMAL MODE ........................................................................................... 16-4
APPENDIX A. APPLY UPG1 FOR PC REPLACEMENT ............................................... A-1
        INTRODUCTION ................................................................................................................ A-1
        APPLYING UPG1.............................................................................................................. A-1
APPENDIX B. HOW TO INSTALL LICENSE WHEN REPLACING PC ....................... B-1
       INRODUCTION .................................................................................................................. B-1
       DAR LICENSE.................................................................................................................. B-2
       SCORE PRO LICENSE .................................................................................................... B-3
      B.3.1 Installing License ................................................................................................... B-3
      B.3.2 Backup License File................................................................................................ B-5
APPENDIX C. TROUBLESHOOTING OF FPD ............................................................... C-1
   C.1 INTRODUCTION ................................................................................................................ C-1
   C.2 TROUBLESHOOTING ......................................................................................................... C-1
                                                                   v                                                    M517-E610
   C.2.1 In Case of CP2 Does Not Complete Normally ...................................................... C-1
   C.2.2 In Case of Error Occurs During Image Acquisition ............................................. C-1
   C.2.3 In Case of Gain Calibration Does Not Complete .................................................. C-2
 C.3 MAINTENANCE OF CP2 ................................................................................................... C-3
   C.3.1 Confirmation of Communication ........................................................................... C-3
   C.3.2 Writing CP2 F/W .................................................................................................... C-4
   C.3.3 Setting Time of CP2.............................................................................................. C-11
   C.3.4 Confirmation of CP2 Errors ................................................................................. C-15
M517-E610                                                     vi
                                                                                      =Introduction=
1. Introduction
This manual describes the method of carrying in, installing and wiring of DAR-9500f.
                                              NOTE
  Please install the device to demonstrate the function of the device enough, to prevent the
  trouble, and to operate efficiently after understanding the content of this book.
                                            WARNING
  Do not install other software in the PC. Do not connect peripheral equipment, or modify the
  PC hardware. Otherwise, the system may fail to start, because of a change in the OS
  environment or driver confliction. If these acts are executed, Shimadzu cannot guarantee the
  operation of the system.
                                            CAUTION
  Do not connect this system to the Internet. Shimadzu shall not guarantee the security against
  virus infection or leak of hospital/patient information through Internet connection. Do not
  connect this system to other computer networks that can be linked to the Internet.
                                                1-1                                      M517-E610
=Introduction=
M517-E610                         1-2
                                                                  =Installation/Enviromental Condition=
2. Installation/Environmental Condition
 2.1 Power source
      Digital Image Processing    Single-phase 200/220/230/240V          4 KVA               Class D grounding
      system                      (50/60Hz)                              24-hour
      (DAR-9500f)                 (1*) 200/220/230/240V+10%              energization
      Lateral Image Processing    Single-phase 200/220/230/240V          4 KVA               Class D grounding
      system (*2)                 (50/60Hz)                              24-hour
      (DAR-9500f Bi)              (1*) 200/220/230/240V+10%              energization
      Transformer Box for PC      Single-phase                           4KVA                Class D grounding
                                  200/220/230/240V                       24-hour
                                  (50/60Hz)                              energization
                                  (1*) 200/220/230/240V+10%
      (*1) Another power source than C-arm and X-ray High Voltage Generator needed.
      (*2) For Bi-plane system
 2.3 Grounding
  Class D grounding/Grounding resistance       :Up to 100 Ohm
  (Class C grounding/Grounding resistance      :Up to 10 Ohm is necessary when using 3 phase
  380-480V on X-ray generator, C-arm, Catheterization table and X-ray tube control cabinet.)
*The generated heat quantity from extra accessories is not included in this table.
*The values in this table are described about the Cardio Vascular System.
                                                    2-1                                             M517-E610
=Installation/Enviromental Condition=
M517-E610                                      2-2
                                                                    =Installation of DAR-9500f (ACQ3)=
3. Installation of DAR-9500f
                                                 NOTE
   The following terms are abbreviated as shown on the right for easy understanding of this manual.
                                                Abbreviation in this manual:
      Digital Reference Cabinet                 REF CABINET
      Digital Acquisition Cabinet 3             ACQ CABINET 3
      Digital Acquisition Cabinet 3 (Lateral)   ACQ CABINET 3L
      Digital Reference PC (*1)                 REF PC
      Digital Acquisition PC (*2)               ACQ PC
                                                3-1                                    M517-E610
=Installation of DAR-9500f (ACQ3)=
(REF PC)
                                              ACQ PC F
                                                                          5) MONITOR (Examination Room)
UPS UNIT
4) IVR NEO
8) SYSTEM DISPLAY
                                                                        9) JOINT BOX2
 2) MONITOR (Control Room)
                                                                  7)8/12inch FPD
                                                                         F               11) IVR SHUTTLE
                                                                                              (OPTION)
M517-E610                                         3-2
                                                                         =Installation of DAR-9500f (ACQ3)=
Components
                                                    Size (mm)        Weight
 No.                  Name                                                      Installation Location
                                               HeightxWidthxDepth     (kg)
        REF CABINET
  1)
        (Digital Reference Cabinet)
                                        522×229×647                 35          Control Room
        ACQ3 CABINET
  6)
        (Digital Acquisition Cabinet)
                                        1810x830 x 490              250         Equipment Room
        SYSTEM DISPLAY
  8)
        (X RAY CONSOLE)
                                        380x400x175                 7           Control Room
Except the components above, there are 1set of KEYBOARD and MOUSE, and LAN Cables attached to
ROUTER and HUB. These are just spare components so not be used.
                                                         NOTE
   Refer to chapter 4.1 for Biplane components.
                                                      3-3                                    M517-E610
=Installation of DAR-9500f (ACQ3)=
1) TRANSFORMER BOX
Components
                                              Size (mm)            Weight       Installation
 No.               Name
                                         HeightxWidthxDepth         (kg)         Location
        DISTRIBUTION BOX
  2)                                 70 x 400 x 165           -               Control Room
        PDB3000
                                                       NOTE
   Distribution Box can be connected up to two according to the number of Optional PC.
M517-E610                                          3-4
                                                                  =Installation of DAR-9500f (ACQ3)=
     [Examination Room]
                                                                                  Information Display
                                                                         BEDSIDE CONSOLE
                                                                             IVR NEO
[Control Room]
                                         Control Room
                                         MONITOR
                                                                DVI SUB ASSY F
                                                 SYSTEM
                                                 DISPLAY
                                           3-5                                       M517-E610
=Installation of DAR-9500f (ACQ3)=
Inside of ACQ3 CABINET is divided into 6 shelves. These are called 1F, 2F…from the bottom.
6F SYS-DISP PC
                                                            CP2 L
              4F                         LCD for                                               Keyboard / Mouse
                                         maintenance
                                                            CP2 F                               for maintenance
                                                                                                     Router /
              3F                                                                                HUB (Si:1pc/Bi :2pc)
                                        MPC
                                                                                                 UPS UNIT
              2F
TRANSFORMER UNIT
1F
M517-E610                                                  3-6
                                                  =Installation of DAR-9500f (ACQ3)=
                                                                   I-SYS BOARD F
       I-SYS BOARD L
DC POWER 5V F
DC POWER 5V L
                                                                     DVI MAIN F
        PERIP-SH
         BOARD
                                            3-7                   M517-E610
=Installation of DAR-9500f (ACQ3)=
    There are two types in each IVR NEO I/F board, IVR NEO (IVR I/F) and IVR NEO cables. Available
    combinations are shown in the following table.
M517-E610                                   3-8
                                                  =Installation of DAR-9500f (ACQ3)=
                                   3-9                            M517-E610
=Installation of DAR-9500f (ACQ3)=
M517-E610                                 3-10
                                                                    =Installation of DAR-9500f (ACQ3)=
    There are two types of DVI-MAIN. Function is compatible (partially-restricted), but the board size and
    connector location are different.
    There are two types of UPS on 2nd floor of ACQ Cabinet.Installation procedures are different
    depending on APC UPS(SUA1500JB) and OMRON UPS(BN150S). For more information, please refer
    to “M517-E447 CVS Remote Maintenance Installation Manual”.
                                              3-11                                     M517-E610
=Installation of DAR-9500f (ACQ3)=
Remove the front cover of ACQ3 Cabinet and set the following components inside the cabinet.
Installation location is described as from 1F to 6F (1st Floor ~6th Floor from the bottom).
(3)Pull.
(1)Remove 2 screws.
M517-E610                                                    3-12
                                                                  =Installation of DAR-9500f (ACQ3)=
                                                NOTE
   Cooling vent is on the left side of REF Cabinet (seen from the front side), so more than 150mm of
   space is needed with other equipments when installing.
ADAPTER BASE
                                             3-13                                    M517-E610
=Installation of DAR-9500f (ACQ3)=
M517-E610                                        3-14
                                                                  =Installation of DAR-9500f (ACQ3)=
                                               NOTE
   DVI OUT Box is an optional item to contain DVT OUT ASSY in case of combining a monitor
   support other than MTA-40 series or MTA-CART series. Prepare a monitor support separately to
   mount examination room monitors and information display.
    Ref.) Dimensions
                       120                             245
        145
175
215
Refer to “Installation manual for MH-200S/MH300” for mounting FPD on C-arm and wiring cable.
                                             3-15                                    M517-E610
=Installation of DAR-9500f (ACQ3)=
    To prevent the cabinet from falling down, fix it against the wall, using 2 pieces of thesupplied body
    keep plates 3 and following screaws.
            <accessories 3 (502-28015)>
             No.                  Name                       Parts Number          PC
             1     Fixing Plate                         565-10733                   2
             2     Anchor,HKD-S M8X30                   026-50810-02                4
             3     BOLT, SST HEXSOCET M8X25             022-27157              2
             4     WASSHER, SST PLAIN M8                023-66180                  2
             5     WASSHER, SST SPRING 2 M8             023-77080              2
             6     BOLT, SST HEXSOCH SEMS P3 M6X16      022-27648                  4
M517-E610                                     3-16
                                                                 =Installation of DAR-9500f (ACQ3)=
   Change connection of the input wire to the transformer in ACQ3 cabinet and optional Trans BOX as
   follows depending on the supply votage.
                                           CAUTION
  Be sure to make the connection change mentioned here in accordance with the observed supply
  voltage at the installation site.
               [AIN]
                                                      [P-200V]
                                           3-17                                  M517-E610
             Examination Room                         Equipment Room                                                                                   Control Room
                                                                                                                                         [6-5]
                                    [9-2]                                                       [6-10]
                                                                                                                                               [6-1]
M517-E610
                                                                                                                                                               JOINT BOX2
                                    [9-1]                           [6-12]                      [6-9]
                 Information                                                                                                                   [6-3]
                 Display                              SYSDISP PC2                                          [6-8]
                                                                                                                                               [6-2]
            [9-3]                                                                     [12-4]                                                             SYSDISP
                        POLYGRAP H                                                                                                             [6-4]
                                                                                                [12-5]                                                   I/O
                                                                                      [12-1]
                                                 I-SYS                                             D150GC-
                                                                                                                                                                                3.10.1 Connection Diagram
[6-6]
                                    [7-1]
                         FPD                     CP2          DAR-9500f                  [11-2]
                                                                                                                                                 [4-3]         [4-4]
                                                            ACQ3 Cabinet
              IVR                    [10-1]                                                        [6-7]                             F-Ref
3-18
                                                                                                                       IVR Shuttle                                    F-ACQ
                                                                                                                                                                                3.10 Cable Connection (External Connection)
                                                                                                                                                 REF PC            [2-4]
              [3-2]      [3-3]                                                                                     [2-5]
                                              [1-2]        [1-1]             [11-1]                                                                              [2-3]
                                                                                      FPD
            MTA/Monitor Cart                                                          Chiller
                                               Power Distribution Board                                                         REF
            Multi Tap                [3-1]     1-phase 200V-240VAC                                                              CABINE T
                                                                                     =Installation of DAR-9500f (ACQ3)=
                                                             3-19                                    M517-E610
=Installation of DAR-9500f (ACQ3)=
 In case of DVI-MAIN2
M517-E610                                 3-20
                                =Installation of DAR-9500f (ACQ3)=
 In case of DVI-MAIN3
                         3-21                   M517-E610
=Installation of DAR-9500f (ACQ3)=
                                                  NOTE
   The above figure shows the cable connection for Single-plane system.
   For Bi-plane system, follow the label for Bi-plane and connect [5-2] [5-3] [8-1] and [8-2].
M517-E610                                       3-22
                                                                           =Installation of DAR-9500f (ACQ3)=
                                                   3-23                                     M517-E610
      =Installation of DAR-9500f (ACQ3)=
      <Control Room>
  Ref.                                          Length
                   Name            Parts No.                          From               To
  No.                                            [m]
5-6        DVI CABLE, KC-DVI-5   088-52848-70               REF CABINET        REF MONITOR F
                                                5
                                                            F [REF1 MONITOR]
5-7        DVI CABLE, KC-DVI-5   088-52848-70               REF CABINET        ACQ MONITOR F
                                                5
                                                            F [ACQ MONITOR]
4-3        MONITOR Attachment    (Attached to               REF MONITOR F      JOINT BOX2 OUTLET TAP
                                                -
           POWER CABLE           MONITOR)
4-4        MONITOR Attachment    (Attached to               ACQ MONITOR F      JOINT BOX2 OUTLET TAP
                                                -
           POWER CABLE           MONITOR)
      <Examination Room>
  Ref.                                          Length
                   Name            Parts No.                          From               To
  No.                                            [m]
8-4        MONITOR Attachment    (Attached to               MONITOR SUPPORT    MONITOR SUPPORT
           DVI CABLE             MONITOR)                   /MONITOR CART      /MONITOR CART
                                                -           /DVI OUT BOX       /DVI OUT BOX
                                                            DVI OUT ASSY F     REF MONITOR F
                                                            [REF1 MONITOR]
8-5        MONITOR Attachment    (Attached to               MONITOR SUPPORT    MONITOR SUPPORT
           DVI CABLE             MONITOR)                   /MONITOR CART      /MONITOR CART
                                                -           /DVI OUT BOX       /DVI OUT BOX
                                                            DVI OUT ASSY F     ACQ MONITOR F
                                                            [ACQ MONITOR]
3-2        MONITOR Attachment    (Attached to               MONITOR SUPPORT    MONITOR SUPPORT
           POWER CABLE           MONITOR)                   /MONITOR CART      /MONITOR CART
                                                -
                                                            /DVI OUT BOX       /DVI OUT BOX
                                                            REF MONITOR F      MULTI TAP
3-3        MONITOR Attachment    (Attached to               MONITOR SUPPORT    MONITOR SUPPORT
           POWER CABLE           MONITOR)                   /MONITOR CART      /MONITOR CART
                                                -
                                                            /DVI OUT BOX       /DVI OUT BOX
                                                            ACQ MONITOR F      MULTI TAP
      M517-E610                                      3-24
                       =Installation of DAR-9500f (ACQ3)=
DVI            POWER
INPUT          INPUT   Turn on the MONITOR power supply.
        3-25                                M517-E610
=Installation of DAR-9500f (ACQ3)=
                                              NOTE
   Refer to the following figure to check that cables (POWER CABLE) [5-4] [5-5] and [8-3] are
   connected properly.
LED
LED
DVI MAIN 3
   If LEDs blink, cables have been connected properly. If do not blink, turn off the power of
   DVI-MAIN immediately, and check that cable mark and board silk are matched.
M517-E610                                   3-26
                                                                         =Installation of DAR-9500f (ACQ3)=
 Ref.                                             Length
                  Name                Parts No.                         From                           To
  No.                                              [m]
2-3       CABLE,KB-K622K         088-52848-10              REF PC                          KEYBOARD
          KEYBOARD                                -        KEYBOARD CONNECTOR
          EXTENSION CABLE
2-4       CABLE,U2C-E10BK        088-52859-68              REF PC                          MOUSE
          MOUSE,                                  -        USB CONNECTOR
          EXTENSION CABLE
2-5       LAN CABLE              563-65570-06              REF PC                          ACQ3 CABINET 3F
                                                  20
                                                           LAN                             LAN HUB ch4
2-5
SLOT 7 6 5 4 3 2 1
                                                         REF PC REAR
                    2-3        2-4
                                             Dongle
                                             for REF License
                                                       NOTE
      Be sure to attach the dongle (562-28609-01 Dongle, REF) for DAR application license which
      provided in the pack of REF cabinet, to USB connector of REF PC. Or DAR application will not
      start.
                                                  3-27                                        M517-E610
=Installation of DAR-9500f (ACQ3)=
        Cable to FPD
  Ref.                                                Length
                    Name             Parts No.                           From                    To
  No.                                                  [m]
7-1        FPD CABLE               Attached      to            ACQ3 CABINET 4F        C-arm
                                                      40
                                   FPD                         CP2                    FPD MAIN BODY
                                                           NOTE
       For Bi-plane system, be sure to connect power supply cable [11-1] for Frontal FPD cooling unit to
       ACQ3L cabinet because of electricity capacity.
M517-E610                                             3-28
                                                                         =Installation of DAR-9500f (ACQ3)=
 <Control Room>
 Ref.                                                      Length
                    Name                Parts No.                         From                   To
 No.                                                        [m]
4-1     POWER CABLE JOINT BOX 2    562-27939-05                     JOINT BOX 2        REF CABINET F
                                                           5
                                                                    [CN7]              [CN7]
4-2     EARTH CABLE, JOINT BOX 2   562-27940-05                     JOINT BOX 2        REF CABINET
                                                           5
                                                                    EARTH BAR          EARTH BAR
6-1     CORD SET KP300C            071-60815-04                     JOINT BOX 2        SYSTEM DISPLAY
                                                                    OUTLET TAP         Touch Panel LCD
                                                           2.5
                                                                    24 hours ON        POWER SUPPLY
                                                                    [System Display]
6-2     CABLE,SYS-DISP CN1         562-27794                        SYSTEM DISPLAY     JOINT BOX 2
                                                           3        SYS-DISP SIDE      SYSDISP I/O [CN1]
                                                                    [CN1]
6-3     EARTH CABLE SYS-DISP       563-67043                        JOINT BOX 2        SYSTEM DISPLAY
                                                           3
                                                                    EARTH BAR          REAR COVER
6-4     TOUCH PANEL Attachment     (Attached to                     JOINT BOX          SYSTEM DISPLAY
        I/F CABLE/CABLE,C232N-E9   SYSTEM DISPLAY)         -        SYSDISP I/O        Touch Panel LCD
                                   /088-52859-16                    [J9]               [TOUCH PANEL I/F]
6-5     HAND SWITCH,C2U-23         571-64129                        SYSTEM DISPLAY     SYSTEM DISPLAY
                                                           -
                                                                    BACK PLANE         HAND SWITCH HOLDER
6-6     CABLE, IVR J10             503-67038-03                     JOINT BOX 2        IVR SHUTTLE
                                                           3
        (IVR SHUTTLE OPTION)                                        SYSDISP I/O[J10]   [CN1]
                                                    3-29                                    M517-E610
 =Installation of DAR-9500f (ACQ3)=
 Ref.                                                    Length
                   Name               Parts No.                         From                   To
 No.                                                      [m]
9-1     VGA CABLE,INFODISP       502-27934-02                     Examination Room   ACQ3 CABINET 6F
                                                                  INFORMATION        SYSDISP PC
                                                         30
                                                                  DISPLAY            VGA CONNECTOR 2
                                                                  VGA
9-2     CABLE,AUDIO2-ID          502-27787-02                     Examination Room   ACQ3 CABINET 6F
                                                                  INFORMATION        SYSDISP PC
                                                         30
                                                                  DISPLAY            MINI STEREO L
                                                                  AUDIO IN
9-3     DISPLAY Attachment       (Attached          to            Examination Room   Examination Room
        POWER CABLE              DISPLAY)                         INFORMATION        MTA MULTI TAP
                                                         -
                                                                  DISPLAY
                                                                  [POWER]
 M517-E610                                        3-30
                                                                               =Installation of DAR-9500f (ACQ3)=
  Ref.                                            Length
                  Name               Parts No.                          From                            To
  No.                                              [m]
10-1      IVR NEO CABLE ASSY       502-24834-20               IVR NEO I/F [CN8] ,         Examination Room
                                                  20          TRANSFORMER UNIT            BED SIDE
                                                              [EARTH BAR]                 IVR NEO
C-2 C-2
              10-1                                                10-1
         (Shield clamp)                                           (Shield clamp)
                            10-1                                                      2
                      (Protective earth
                     and shield clamp)
                                                       3-31                                     M517-E610
=Installation of DAR-9500f (ACQ3)=
<Removing Cover>
M517-E610                            3-32
                                                                               =Installation of DAR-9500f (ACQ3)=
  Ref.                                        Length
                 Name       Parts No.                                 From                            To
  No.                                          [m]
10-1     IVR NEO CABLE    565-11069-20                    IVR NEO I/F2 [J13]              Examination Room
                                              20          DOOR                            BED SIDE
                                                          [EARTH]                         IVR NEO
C-2
                                                                                        10-1
                                                                                   (Shield clamp)
                                                                         10-1
                                                                  (Protective earth
                                                                 and shield clamp)
                                                   3-33                                         M517-E610
=Installation of DAR-9500f (ACQ3)=
                                                                       12-3
                                               COM I/F 2002
                                                                                                12-2
J34 DAR
                           D150GC-40F
12-4       6-11                                                 12-1
                                                                                  FLUO-2002
12-5
UD CONT-2002
M517-E610                                              3-34
                                                                             =Installation of DAR-9500f (ACQ3)=
                                                       NOTE
         ・Connect BNC (cables in the hospital) to CH1(CN24) in Polygraph and it can be connected to ECG.
         ・Referring to [Exposure Parameters Tab] of DAR-9500f OPERATION MANUAL, and execute
          necessary configurations.
         ・(For Vpp of BNC CONNECTOR of Polygraph) The scope of available input power voltage is
          0-10V.
                                                      3-35                                    M517-E610
                   SYSDISP PC2              SLOT 4        3       2       1
                                                                                       [C-2]        IVR NEO I/F                    CP2F
                                                                                                     CN14
                              LAN
                                                                                                     ARC                             EXTERNAL SYNC        16Bit DIGITAL VIDEO             LAN
M517-E610
                                    [C-1]                                        [A-10]
                                                                                                                            Dose Meter                 [B-3]          [B-2]           [B-1]
                                                                                                            [A-12]          (option)
                       ACQ PC
                                      SLOT 7          6       5       4          3      2      1
                                                                                                        P2                       [A-2]                         CN20      CN1
                                                                                                                                           CN8
                                                                           CL0
                                                                                                                                                                                                      3.11.1 Conncetion Diagram
                                        LAN
                                                                           CL1                          P1
                                                                                                                                                                                              [D-4]
                                                                                                                                 [A-11]
                                                                                                                                                                               CN16
                                             [A-8]
                                                                                                                                                                                                                                                    =Installation of DAR-9500f (ACQ3)=
                                                                          [A-3]
                           [A-4]      [A-5]                                                                                                CN17
                                                                  [A-7]
                                                                                                             [A-1]
                                                          [A-9]                                                                            CN4 LAN
                                                                                                                                                                                                      3.11 ACQ3 Cabinet Internal Cable Connection
I-SYS
3-36
                                              [D-2]           UPS                                                                                CN3      BOARD F
[A-6] [E-2]
                                      [C-3]
            COM1     MPC
                                                                                                   TRANS         J102                       DVI MAIN
            COM2     LAN              [D-3]                                          [E-1]         UNIT                                     ASSY F
                                                                                                           CP2
                        [D-1]
                                                                                                                                                   J10[△1]
                                                                                                            [B-4]
ROUTER
                                                                  HUB
                                                                                                                        [E-4]
                            [E-3]
                                                                       =Installation of DAR-9500f (ACQ3)=
 Ref.                                                 Length
                Name                 Parts No.                         From                       To
  No.                                                  [m]
 A-1    SOLIOS-R/ISYS CABLE    562-28190-01                    ACQ PC F              I-SYS F
                                                      1.5
                                                               SLOT3 [CL0] [CL1]     [CN3]
 A-2    IO-R/I-SYS CABLE       565-10473-25                    [A-3]                 I-SYS F
                                                      1
                                                               CABLE CONNECTOR       [CN8]
 A-3    IO CONV CABLE          562-28135-03                    ACQ PC F              [A-2]
                                                      0.3
                                                               SLOT4 [J1]            CABLE CONNECTOR
 A-4    KEYBOARD               (Attached to ACQ PC)            ACQ PC F              Keyboard Holder
                                                      -        KEYBOARD
                                                               CONNECTOR
 A-5    MOUSE                  (Attached to ACQ PC)            ACQ PC F              Keyboard Holder
                                                      -
                                                               USB CONNECTOR
 A-6    ACQ PC Attachment      (Attached to ACQ PC)            ACQ PC F              UPS
                                                      -
        POWER CABLE                                            [ATX POWER SUPPLY]    [ACQ PC F]
 A-7    ACQ INTERLOCK          565-10463                       ACQ PC F              TRANSFOMAR UNIT
                                                      2
        CABLE                                                  SLOT5                 [J102],
 A-8    LAN CABLE 3 ACQ        565-10447-01                    ACQ PC F              LAN HUB
                                                      1
                                                               LAN                   [CH1]
 A-9    DVI CABLE 3 ACQ        565-10467-02                    ACQ PC F              DVI MAIN F
                                                      2        SLOT6                 DVI IN1 [J10]
                                                                                     [△1]
 A-10   DAR ARCNET ACQ-SYS3    563-75790-11                    ACQ PC F              SYSDISP PC REAR
                                                      1
                                                               SLOT2 [J6]            SLOT1 [J6]
 A-11   DAR ARCNET ISYS-ACQ    563-75790-13                    ACQ PC F              I-SYS
                                                      1.5
                                                               SLOT2 [J5]            [CN17]
 A-12   DOSEMETER,SERIAL       565-10449-01                    ACQ PC F              DOOR
                                                      1.5
        CABLE                                                  SLOT1 [P2]            SERIAL(Dose Metor)
                                                            3-37                                M517-E610
=Installation of DAR-9500f (ACQ3)=
        M720
                                                                      A-2
                               Dongle
                   A-8         for Stent View (Option)                 A-3                    A-10
SLOT 7 6 5 4 3 2 1
                                                                                    CL0
                                                                                                            P1
                                                                                    CL1
                                                                                                            P2
 M730
                                                NOTE
   Be sure to attach the dongle (562-28609-01 Dongle, ACQ F) for DAR application license which
   provided in the pack of ACQ PC, to USB connector of ACQ PC. Or DAR application will not start.
   (For M730, the dongole is mounted inside PC) And also attach the license dongle for Stent View
   (Option).
M517-E610                                     3-38
                                                                       =Installation of DAR-9500f (ACQ3)=
 Ref.                                                 Length
                Name                 Parts No.                         From                         To
  No.                                                  [m]
 B-1    LAN CABLE 3 CP2        565-10447-02                    CP2 F                   LAN HUB
                                                      1
                                                               [ETHERNET]              [CH2]
 B-2    I-SYS-CP2 LVDS CABLE   562-27740-02                    CP2 F                   I-SYS
                                                      1
                                                               [16BIT DIGITAL VIDEO]   [CN1]
 B-3    I-SYS-CP2 IO CABLE     565-10423-02                    CP2 F                   I-SYS
                                                      1
                                                               [EXTERNAL SYNC]         [CN20]
 B-4    CP2 Attached   POWER   Attached to CP2                 CP2 F                   TRANSFORMER UNIT
                                                      -
        CABLE                                                  [POWER SUPPLY]          [CP2 F]
 Ref.                                                 Length
                Name                 Parts No.                         From                         To
  No.                                                  [m]
 C-1    LAN CABLE 3 SYSDISP    565-10447-07                    SYSDISP PC              ROUTER LAN-4
                                                      2
                                                               LAN
 C-2    DAR ARCNET SYS-NEO3    563-75790-12                    SYSDISP PC              IVR NEO I/F
                                                      2
                                                               SLOT3 [J5]              [CN14]
 C-3    POWER CABLE            (Attached to SYSDISP            SYSDISP PC              UPS
                                                      -
                               PC)                             [ATX POWER SUPPLY]      [SYSDISP PC]
 Ref.                                                 Length
                Name                 Parts No.                         From                         To
  No.                                                  [m]
 D-1    LAN CABLE 3 MPC        565-10447-06                    MPC                     ROUTER LAN-3
                                                      1
                                                               LAN
 D-2    UPS CABLE              565-10448-01                    MPC                     UPS
                                                      -
        (for APC UPS)                                          [COM1]                  [Advance Port]
 D-2    UPS CABLE              565-12398-01                    MPC                     UPS
                                                      -        [COM1]                  [RS-232C]
        (for OMRON UPS)
 D-3    POWER CABLE            (Attached to MPC)               MPC                     UPS
                                                      -
                                                               [POWER SUPPLY]          [MPC]
 D-4    CABLE, SERIAL CROSS    565-10470-04                    MPC                     I-SYS
                                                      1.5      [COM2]                  [CN16]
        DEBUG
                                                            3-39                                  M517-E610
=Installation of DAR-9500f (ACQ3)=
 Ref.                                                                      Length
                       Name                      Parts No.                                               From                       To
  No.                                                                       [m]
 E-1         ROUTER Attachment             (Attached to ROUTER)                              ROUTER                In case of APC UPS
             POWER CABLE                                                                     [POWER SUPPLY]        [ROUTER]
                                                                           -                                       In case of OMRON UPS
                                                                                                                   OA TAP
                                                                                                                   [ROUTER]
 E-2         HUB Attachment                (Attached to HUB)                                 HUB                   UPS
                                                                           -
             POWER SUPPLY                                                                    [POWER SUUPLY]        [HUB F]
 E-3         LAN CABLE 3 ROUTER            565-10447-05                                      ROUTER LAN-1          LAN HUB
                                                                           1
                                                                                                                   [CH8]
 E-4         LAN CABLE 3 ISYS              565-10447-03                                      LAN HUB               I-SYS
                                                                           2
                                                                                             [CH3]                 [CN4]
<NETWORK CONNECION>
                                                                                                                          REF PC
                                                          ROUTER
       ACQ PC ASSY
                                                               4   3   2   1           WAN
                                                                                                                         LAN
       LAN
                                     MPC                                                         [E-3]
              CP2
                      LAN             LAN
                                                       [D-1]
[E-4]
                                                                               1   3    5    7
             [A-6]                                           HUB
                                                                                                                         LAN
                                                                               2   4     6   8
                                                                                                                   CN4
                             [B-1]                                                                       [2-5]            I-SYS
                                                                                                                          BOARD F
M517-E610                                                          3-40
                                                                      =Installation of DAR-9500f (ACQ3)=
  Ref.                                               Length
                 Name                 Parts No.                       From                       To
   No.                                                [m]
 Z-1      FAN ASSY,TOP 2        563-70260-10                  TRANSFORMER UNIT       Cabinet Top FAN
                                                     -
                                                              [LOUT1 LOUT2]
 Z-2      I-SYS DC CABLE        565-10444                     TRANSFORMER UNIT       I-SYS F/IOF
                                                     -
                                                              [J100 J101]            [CN6 CN21/CN3]
 Z-3      REF      INTERLOCK    565-10471-10                  TRANSFORMER UNIT       DVI MAIN F
                                                     1
          CABLE                                               [J104]                 [J9]
 Z-4      AC CABLE DVI-MAIN     565-10446                     TRANSFORMER UNIT       DOOR
                                                     1.1      [LOUT1 LOUT2]          PS3
                                                                                     [CN1]
 Z-5      EARTH CABLE           565-10472-02                  TRANSFORMER UNIT       DOOR
                                                     0.8
                                                              EARTH BAR              EARTH BAR
 Z-6      NEO IF DC CABLE       562-28022                     TRANSFORMER UNIT       IVR NEO I/F
                                                     -
                                                              [J105]                 [CN2]
 Z-7      CABLE SERIAL CROSS    565-10470-03                  I-SYS F                DVI MAIN F
                                                     0.8
          DVI3                                                [CN9]                  [J30]
 Z-8      LAN CABLE DVI         562-28024-01                  Si: I-SYS F [CN31]     DVI MAIN F
 (Note)                                              1        Bi: I-SYS L [CN31]     [J16(△7)]
 Z-9      DVI-MAIN DC CABLE     565-10464                     DVI MAIN F             DOOR
                                                     0.5      [J1]                   PS3
                                                                                     [CN2]
 Z-10     UPS Attachement       (Attached to UPS)             UPS                    TRANSFORMER UNIT
                                                     -
          Power SUPPPLY                                       [POWER SUPPLY]         [UPS]
 Z-11     I-SYS FAN ASSY F      565-12323-01                  No.[Z-12]              No.[11-2]
                                                     2.0
                                                              [j112]CONNECTOR        [J113]CONNECTOR
 Z-12     I-SYS FAN JUNCTION    565-12292-01                  TRANSUNIT TB3          No.[Z-11]
                                                     0.8
          CABLE                                               [12V-12G]              [J112]CONNECTOR
 Z-13     IOF FAN ASSY F        565-12322-01                  No.[Z-14]              No.[11-2]
                                                     0.4
                                                              [J110]CONNECTOR        [J111]CONNECTOR
 Z-14     IOF FAN JUNCTION      565-12292                     TRANSUNIT TB3          No.[Z-13]
                                                     1.3
          CABLE                                               [12V 12G]              [J110]CONNECTOR
(Note) Connect to I-SYS F for Single-plane system and I-SYS L for Bi-plane system.
                                                           3-41                                M517-E610
=Installation of DAR-9500f (ACQ3)=
M517-E610                            3-42
                                                                                                     =Installation of DAR-9500f (ACQ3)=
Ref.                                                                     Length
                  Name                           Parts No.                                           From                                       To
 No.                                                                      [m]
X-1    POWER SWITCH CABLE                  565-11373                                REF PC                                    REF PC
                                                                         -
                                                                                    [MOTHER BOARD]                            [CNB]
X-2    PWCNT CABLE ASSY                    562-24810                                REF PC                                    REF PC
                                                                         -
                                                                                    [REAR PANEL]                              [POWER SUPPLY/CNB]
X-3    CABLE POWER UNIT 2                  562-27272-05                             REF CABINET                               REF CABINET
                                                                         0.2
                                                                                    [TB R0/R100]                              [CN]
X-4    CABLE POWER UNIT 2                  562-27272-06                             REF CABINET                               REF CABINET
                                                                         0.05
                                                                                    [TB R100]                                 [TB R100]
X-5    CABLE POWER UNIT 2                  562-27272-07                             REF CABINET                               REF CABINET
                                                                         0.05
                                                                                    [TB R0]                                   [TB R0]
X-6    FAN CABLE                           563-62784-02                             REF CABINET                               REF CABINET
                                                                         0.65
                                                                                    [TB]                                      [FAN]
X-7    SPARK KILLER ASSY                   562-27838                                REF CABINET
                                                                         -
                                                                                    [TB]
X-8    CABLE REF CN7                       562-27942                                REF CABINET
                                                                         -
                                                                                    [TB]
X-9    CABLE POWER UNIT 2                  562-27272-08                             REF CABINET                               REF CABINET RY1
                                                                         0.23
                                                                                    [TB]                                      [23] [13]
X-10   DVI CABLE 3 REF F                   565-10467-01                             REF PC                                    DVI SUB
                                                                         1
                                                                                    SLOT6                                     [IN (J1)]
X-11   REAR PANEL INLET                    563-62782                                REF CABINET                               REF CABINET
                                                                         0.15
                                                                                    REAR PANEL INLET                          NFB LINE [L][N]
X-12   CABLE POWER UNIT 2                  562-27272-01                             REF CABINET                               REF CABINET TB
                                                                         0.13
                                                                                    NFB LOAD [L]                              RY1[23]
X-13   CABLE POWER UNIT 2                  562-27272-02                             REF CABINET                               REF CABINET
                                                                         0.16
                                                                                    NFB LOAD [N]                              [F1]
X-14   CABLE POWER UNIT 2                  562-27272-03                             REF CABINET                               REF CABINET TB
                                                                         0.11
                                                                                    [F1]                                      RY1[13]
X-15   POWER CONT CABLE 2                  562-27542                                REF CABINET RY1                           REF PC
                                                                         0.4
                                                                                    [A1][A2]                                  SLOT5
X-16   ATX POWER CABLE REF                 565-11257                                REF PC
                                                                         0.53
                                                                                    POWER SUPPLY
X-17   CABLE POWER UNIT2                   562-27272-04                             REF CABINET                               REF CABINET
                                                                         0.2
                                                                                    RY1 [24][14]                              [CN]
                                                       [X-16]
                                                                                                                                       REF PW
          REF PC                  Power
                        CNC       Supply                        DVI SUB                           [X-6]
                                                                                                                                           [X-17]
                                                                                                                               CN
                                                                                                                  [X-3]
FRONT     [X-1]           [X-2]
                  CNB
                                                                                                            TB
                                                                                                                 [X-9]
                                                                                PANEL
                                                                                                  NFB
                                                                                INLET                     [X-13]              [X-14]
                                                                                        [X-11]
[X-12]
                                                                                3-43                                                       M517-E610
=Installation of DAR-9500f (ACQ3)=
Check that I-SYS for Frontal DIP SW is set as the following table.
                                                   NOTE
   Refer to chapter 4.9 for I-SYS BOARD setting for Bi-plane Lateral.
      DIP-SW1
                  1       2      3     4       5      6     7       8
        OFF       ○                    ○              ○     ○       ○
        ON               ○      ○             ○
      DIP-SW2
                  1      2      3      4      5       6     7       8
        OFF       ○      ○      ○      ○      ○             ○       ○
        ON                                            ○
      DIP-SW3
                  1      2      3      4      5       6     7       8
        OFF       ○      ○      ○      ○      ○       ○     ○       ○
        ON
                                                                         SW1
      Push Switch
SW2
SW3
M517-E610                                      3-44
                                                                  =Installation of DAR-9500f (ACQ3)=
   SW1
      SW1                   Contents                        OFF                          ON
         1     Startup Software                   Firmware                 Maintenance Program
         2     IQP                                Rev.2.2.0 less than      Rev.2.2.0 or more
         3     FPGA Config                        JTAG Cable               Flash Memory
         4     Where to store FPGA Config file    Regular                  Backup
               during configuration by Flash.
         5     IOF ON/OFF                         OFF                      ON
         6     Push Switch                        Enable                   Disable
               FPGA reconfiguration
         7     Push Switch                        Disable                  Enable
               Software reset
         8     Select Frontal/Lateral             Frontal                  Lateral
   SW2
      SW2                  Contents                         OFF                        ON
         1
         2
         3     Unused
         4
         5
         6     DAR IO Communication               CURRENT LOOP             LVTTL
         7
               Unused
         8
   SW3
      SW3                   Contents                        OFF                        ON
         1     Unused
         2     Unused
         2     Baud rate of maintenance program   (S3_2:S3_3)
         3                                        OFF : OFF = 115200 bps
                                                  OFF : ON = 230400 bps
                                                  ON : OFF = 460800 bps
                                                  ON : OFF = 921600 bps
         4     Debug of Ethernet                  OFF                      ON
               Hardware debug
         5
         6
               Unused
         7
         8
                                                     3-45                               M517-E610
=Installation of DAR-9500f (ACQ3)=
Check that DVI-MAIN2 for Frontal DIP SW is set as the following table.
                                                  NOTE
   Refer to chapter 4.9 for DVI MAIN2 setting for Bi-plane Lateral.
      DIP-SW1
                  1      2       3     4      5      6     7      8
        OFF       ○      ○             ○      ○      ○     ○      ○
        ON                      ○
      DIP-SW2
                  1      2      3      4      5      6     7      8
        OFF       ○      ○      ○      ○      ○      ○     ○      ○
        ON
      DIP-SW7
                   1     2      3      4      5      6     7      8
        OFF              ○      ○                          ○      ○
        ON        ○                    ○      ○      ○
SW7
SW1 SW2
M517-E610                                     3-46
                                                                  =Installation of DAR-9500f (ACQ3)=
   SW1
      SW1                   Contents                        OFF                        ON
         1     Startup Software                  Firmware                  Maintenance Program
         2     Unused
         3     FPGA Config                       JTAG Cable                Flash Memory
         4     Where to store FPGA Config file   Regular                   Backup
               during configuration by Flash.
         5
               Unused
         6
         7     VCR Output                        NTSC                      PAL
         8     Unused
   SW2
      SW2                   Contents                        OFF                        ON
         1     Startup Software                  Firmware                  VGA VCR
                                                                           Maintenance Program
         2
         3
         4
         5     Unused
         6
         7
         8
                                                   3-47                                   M517-E610
=Installation of DAR-9500f (ACQ3)=
   SW7
     Signal amplification level of DVI-MAIN2 input channel (from J13 to J16) can be changed using DIP
    SW7. Amplification level can be set with 8 kinds of level (from 000 to 111).
M517-E610                                     3-48
                                                                    =Installation of DAR-9500f (ACQ3)=
Check that DVI-MAIN3 for Frontal DIP SW is set as the following table.
                                                NOTE
 Refer to chapter 4.9 for DVI MAIN3 setting for Bi-plane Lateral.
    DIP-SW1
                1       2      3     4       5     6      7       8
      OFF       ○                    ○      ○      ○      ○       ○
                                           (*1)
      ON               ○      ○
    DIP-SW2
                1      2      3      4      5      6          7     8
      OFF       ○      ○      ○      ○      ○
                                                         (Note)
      ON
    DIP-SW3
                1      2      3      4      5      6      7       8
      OFF       ○      ○      ○      ○      ○      ○      ○       ○
      ON
    DIP-SW7
                 1     2      3       4     5      6      7         8
      OFF              ○      ○                           ○
      ON        ○                    ○      ○      ○              ○
(*1) With SMART Display option and LIVE Monitor option, set the segment5 of DIP SW1 to ON.
  (Note) DIP SW is aleready set at the factory. If changing is necessary, refet to “(2) Adjusting REF PC
  startup signal and NTSC/PAL signal”.
                                                       3-49                                M517-E610
=Installation of DAR-9500f (ACQ3)=
LD18
LD36
                                LD31
                                                                                    LD17
                                                  SW1
                                                                      LD14
                                                                      LD13
                                                       SW2
                                                                      LD15
                                LD32                   SW3                          LD16
LD33 SW7
      <SW1>
       SW1                  Description                         OFF                               ON
         1      Startup Sotware                        Firmware                       Maintenance Program
         2      Selection of DVI-MAIN Type             DVI-MAIN 2                     DVI-MAIN 3
         3      FPGA Config                            JTAG Cable                     Flash Memory
         4      Where to store FPGA Config file        Regular                        Backup
                during configuration by Flash.
         5      SMART Display Option and
                                                       Disable                        Enable
                LIVE Monitor Option
         6      Push Switch                            Enable                         Disable
                FPGA reconfiguration
         7      VCR Output                             NTSC                           PAL
         8      Unused
M517-E610                                       3-50
                                                                =Installation of DAR-9500f (ACQ3)=
<SW2>
  SW2                 Description                         OFF                        ON
    1     Startup Software                     Firmware                  VGA VCR
                                                                         Maintenance Program
    2     Unused
    3     Phase shifting check mode of REF Regular                   When checking
    4     PC startup signal and NTSC/PAL
    5     signal.
    6     Selecting phase of REF PC startup 6 OFF /7 OFF /8 OFF: Phase 315˚
    7     signal and NTSC/PAL signal.       6 ON /7 OFF /8 OFF: Phase 270˚
    8                                       6 OFF /7 ON /8 OFF: Phase 225˚
                                            6 ON /7 ON /8 OFF: Phase 180˚
                                            6 OFF /7 OFF /8 ON: Phase 135˚
                                            6 ON /7 OFF /8 ON: Phase 90˚
                                            6 OFF /7 ON /8 ON: Phase 45˚
                                            6 ON /7 ON /8 ON: No phase shift
<SW3>
Signal amplification level of DVI-MAIN3 input channel (from J20 to J27) can be changed using DIP
SW3.
  SW3                  Description                     OFF                          ON
    1     Setting amplification level     of   OFF: 1
    2     DVI-MAIN3 J20 and J27.               ON: 0
    3                                          (Note 1)
    4     Setting Output Level                 1000 mV                   1200 mV
    5     Setting Enabled Channel of Output    (Note 2)
    6
    7     Unused
    8     Unused
                                                 3-51                                 M517-E610
=Installation of DAR-9500f (ACQ3)=
      <SW7>
      Signal amplification level of DVI-MAIN3 input channel (from J13 to J16) can be changed using DIP
      SW7.
        SW7                  Description                     OFF                          ON
          1      Setting amplification level      of OFF: 1
          2      DVI-MAIN3 J13 and J14.              ON: 0
          3                                          (Note)
          4      Setting amplification level      of OFF: 1
          5      DVI-MAIN3 J15 and J16.              ON: 0
          6
          7      The choice of how to boost the        Enable the setting of      Enable the value of the internal
                 input equalizer CH4IN ~ CH7IN         DIP-SW7-1 ~ 7-6            register
          8      The choice of how to boost the        Enable the setting value   Enable the value of the internal
                 output equalizer CH1A(B) ~            of the external pin        register
                 CH4A(B)
 If the trouble that REF PC doesn’t start up and image of NTSC/PAL is noisy is occurred, it can adjust REF PC
startup signal and NTSC/PAL signal by DIP SW of DVI-MAIN.
 By changing phase of signal, it can adjust REF PC startup signal and NTSC/PAL signal
 Show the relation between DIP SW and phase.
M517-E610                                       3-52
                                                                =Installation of DAR-9500f (ACQ3)=
2.   Change the switch of SW2-6, SW2-7 and SW2-8 by following the Table 3-2, and check that
     NTSC/PAL is displayed. If there is any trouble, “white” and “black” images will change
     approximately every second on the monitor. Or, the relay will keep funtioning.
     Normally, NTSC/PAL is displayed in three phases, so set the intermediate phase switch to SW2-6,
     SW2-7 and SW2-8. (For example: Select 135° if 90°, 135° and 180° are available)
2.   Change the switch of SW2-6, SW2-7 and SW2-8 by following the Table 3-2, and measure the voltage
     of coaxial connector of DVI-SUB (Fig. 3-1) with an oscilloscope or tester.
Coaxial connector
                                                   3-53                                 M517-E610
=Installation of DAR-9500f (ACQ3)=
          Normal
          Voltage is stable as the following figure.
          Abnormal
          Low and high voltage will change approximately every second as the following figure. The
         difference of both voltages is approximately 0.7V. Or, the relay will keep functioning and voltage is
         not stable.
High Voltage
Low Voltage
         Normally, voltage is stable in three phases, so set the intermediate phase switch to SW2-6, SW2-7 and
         SW2-8. (For example: Select 135° if 90°, 135° and 180° are available)
M517-E610                                         3-54
                                                                   =Installation of DAR-9500f (ACQ3)=
     DIP-SW2
                  1      2      3      4      5     6          7
                       (*1)                 (*2)
      OFF        ○      ○                    ○      ○      ○
      ON                       ○      ○
   (*1) If there is IVR Shuttle (Option), it is need to set to ON segment 2 of the DIP-SW2 on the
   PERIP-SH.
   (*2) If there is a Bi-plane system with IVR Shuttle (Option), it is need to set to ON segment 5 of the
   DIP-SW2 on the PERIP-SH.
     DIP-SW3
                 1      2      3       4
      OFF        ○      ○      ○
      ON                              ○
SW2
SW1
JP1
SW3
                                                        3-55                                M517-E610
=Installation of DAR-9500f (ACQ3)=
   SW2
      SW2                   Contents                      OFF              ON
         1     Unused
         2     IVR Shuttle                        None          When using IVR Shuttle
         3     Sound of fluoroscopy               ON            OFF
         4     Sound of radiography               ON            OFF
         5     IVR Shuttle for Biplane            None          When using IVR Shuttle for
                                                                Biplane
         6
               Unused
         7
   SW3
      SW3                  Contents                       OFF              ON
         1     Unused
         2     F/W writing                        OFF           ON
         3     Unused
         4     ARCNET communication rate          12MHz         10MHz
   Jumper
    JP1 IN: ARCNET termination ON (Normally)
       OUT: ARCNET termination OFF
M517-E610                                  3-56
                                                                      =Installation of DAR-9500f (ACQ3)=
      DIP-SW1
                   1     2      3      4      5        6      7
       OFF         ○     ○      ○      ○      ○        ○      ○
       ON
      DIP-SW1
                   1      2     3      4      5        6      7
       OFF         ○            ○      ○      ○        ○      ○
       ON                ○
(2) PERIP-SH2
      DIP-SW2
                   1      2      3      4      5       6          7
                        (*1)                 (*2)
       OFF         ○     ○                    ○        ○      ○
       ON                       ○      ○
    (*1) If there is IVR Shuttle (Option), it is need to set to ON segment 2 of the DIP-SW2 on the
    PERIP-SH2.
    (*2) If there is a Bi-plane system with IVR Shuttle (Option), it is need to set to ON segment 5 of the
    DIP-SW2 on the PERIP-SH2.
      DIP-SW3
                   1     2      3       4
       OFF         ○     ○      ○
       ON                              ○
                                                           3-57                              M517-E610
=Installation of DAR-9500f (ACQ3)=
SW2
                                                                       JP1
                      SW3
SW1
M517-E610                                   3-58
                                                              =Installation of DAR-9500f (ACQ3)=
   SW1
      SW1                  Contents                     OFF                       ON
         1     IVR NEO                         (S1_1:S1_2)
         2                                     OFF : OFF = When connecting IVR NEO (503-63818-01)
                                               OFF : ON = When connecting IVR NEO (503-63818-11)
         3
         4
         5
               Unused
         6
         7
         8
   SW2
      SW2                   Contents                    OFF                        ON
         1     Unused
         2     IVR Shuttle                     None                    When using IVR Shuttle
         3     Sound of fluoroscopy            ON                      OFF
         4     Sound of radiography            ON                      OFF
         5     IVR Shuttle for Biplane         None                    When using IVR Shuttle for
                                                                       Biplane
         6
               Unused
         7
   SW3
      SW3                  Contents                     OFF                        ON
         1     Unused
         2     F/W writing                     OFF                     ON
         3     Unused
         4     ARCNET communication rate       12MHz                   10MHz
   Jumper
    JP1 IN: ARCNET termination ON (Normally)
       OUT: ARCNET termination OFF
                                                 3-59                               M517-E610
=Installation of DAR-9500f (ACQ3)=
Check that DIP SW for IVR I/F5 is set as the following table.
      DIP-SW1
                  1      2      3      4      5       6     7
       OFF        ○      ○      ○      ○      ○       ○     ○
       ON
The following figure is showing the position of IVR I/F5 DIP SW.
SW1
(1) DIP-SW
     <SW1>
        SW2                Contents                              OFF               ON
         1      IVR NEO button lights up in yellow.   Light up         Light out
         2
         3
         4
         5      Unused
         6
         7
         8
M517-E610                                      3-60
                                                                    =Installation of DAR-9500f (ACQ3)=
3. Connect DVI OUT board on DVI OUT ASSY and optional DVI-OUT board.
                                                       3-61                                 M517-E610
=Installation of DAR-9500f (ACQ3)=
Fix 4 screws.
M517-E610                                   3-62
                                    =Installation of DAR-9500f (ACQ3)=
                             3-63                         M517-E610
=Installation of DAR-9500f (ACQ3)=
  Ref.                                                       Length
                Name                Parts No.                               From                  To
  No.                                                         [m]
  17-1   LAN DVI CABLE        502-27540-23                            MONITOR SUPPORT   ACQ3 Cabinet F Door
                                                         30           DVI OUT ASSY F    DVI MAIN F
                                                                      [③]               [⑧]
  17-2   MONITOR Attachment   (Attached to LIVE                       MONITOR SUPPORT   MONITOR SUPPORT
         DVI CABLE            MONITOR)                   -            DVI OUT ASSY F    LIVE MONITOR
                                                                      [REF2 MONITOR]
  17-3   MONITOR Attachment   (Attached to LIVE                       MONITOR SUPPORT   MONITOR SUPPORT
                                                         -
         POWER CABLE          MONITOR)                                LIVE MONITOR      MULTI TAP
M517-E610                                         3-64
                                                                                 =Installation of DAR-9500f (ACQ3)=
Removing covers
  (1) Remove four screws where indicated by         (2) Open the top cover to right and left side.
  arrows.
                                                                3-65                                               M517-E610
=Installation of DAR-9500f (ACQ3)=
2. Remove DVI OUT board and a blinder plate from DVI OUT ASSY.
3. Connect DVI OUT board on rear panel and optional DVI-OUT board.
Fix 4 screws.
M517-E610                                            3-66
                                    =Installation of DAR-9500f (ACQ3)=
                             3-67                         M517-E610
=Installation of DAR-9500f (ACQ3)=
  Ref.                                                      Length
                Name                 Parts No.                               From                  To
  No.                                                        [m]
  17-4   LAN DVI CABLE        502-27540-21                           REF Cabinet F      ACQ3 Cabinet F Door
                                                        30           [③]                DVI MAIN F
                                                                                        [⑦]
  17-5   DVI CABLE,           088-50905-16                           REF Cabinet        LIVE MONITOR
                                                        5
         KC-DVI-5                                                    F [REF2 MONITOR]
  17-6   MONITOR Attachment   (Attached to                           LIVE MONITOR F     JOINT BOX2 OUTLET TAP
                                                        -
         POWER CABLE          MONITOR)
M517-E610                                        3-68
                                                                                      =Installation of DAR-9500f (ACQ3)=
                                          I-SYS BOARD
                                          (ACQ3 CABINET DOOR)
CN20 CN11
                                                                                     CN16
                                                                 CN3
                        CN8
                                                         CN2
                                                                                    SW1
                        CN22
                                                 CN19
                                        CN13
                                CN9
                                                                                    SW2
                                                               IOF
                                                                                    SW3
                         CN
                         31
                                                                                     CN6
                         CN
                         17
CN18
                 CN15                                                               CN
                                                                                    21
                               CN4
                                           CN2
                 ECG1           CN14
                                                        ISYS
                 ECG2
                 CN25                 CN3                              CN23
Left Right
                                                                     3-69                                   M517-E610
=Installation of DAR-9500f (ACQ3)=
J30
                                                                                      LD29
Left                                                                                                     Right
             J34
              J2                                                                     LD20          J16
                                                                                             LD8
                                                                                     LD19          J15
                                                                                             LD7
                                                                                       SW7
              J1
                                     J9            LD1                               LD18          J14
                        LD30                                                                 LD6
             J31                 SW1 SW5 SW4 SW3 SW2                                 LD17          J13
                                                                                             LD5
                          LD32        SW6        LD31             LD2          LD3     LD4
M517-E610                                                  3-70
                                                                            =Installation of DAR-9500f (ACQ3)=
LD18
LD36
                               LD31
                                                                             LD17
                                                              LD14
                                                              LD13
                                                              LD15
                               LD32
                                                                             LD16
LD33
                                                            3-71                                  M517-E610
=Installation of DAR-9500f (ACQ3)=
                                                                 CN13
                                             CN1
                                                                 CN12
                                             CN2
                            PERIP-SH
CN3
                                                                 CN10
        Left                   IVR NEO I/F                              Right
                                                                 CN8
                      CN2
M517-E610                                          3-72
                                                                      =Installation of DAR-9500f (ACQ3)=
                                                     CN    CN
                                                     15    14
CN5
                                                            J14
                                  PERIP-SH2
                                              CN1
                                              CN2
                                                            J13
                Left              IVR NEO I/F2                  CN8
                                                                              Right
                            CN2
J11
                                                    3-73                                    M517-E610
                                                                                                      3.16 LED
M517-E610
                                                                                                      3.16.1 I-SYS BOARD
                          LD32
                          LD31                LD37
                          LD30                LD38
                          LD29                LD39
                          LD28
                          LD27
                                                                     Location of LED is shown below
                                       LD34
                                       LD33             LD3
                                                                                                                           =Installation of DAR-9500f (ACQ3)=
                   LD36                                 LD4
                                                        LD5
                                                        LD6
                                                        LD7
                                                        LD8
                                                        LD9
            LD26                                       LD10
            LD25                                     LD35
            LD24
            LD23
            LD22
3-74
            LD21
            LD20
            LD19
            LD18
            LD17
            LD16
            LD15                 LD2
            LD14
            LD13                                              LD40
            LD12                 LD1
            LD11
                                                              =Installation of DAR-9500f (ACQ3)=
Number                            Explanation
LD1      Output signal of interlocking power for system.
LD2      Output signal of interlocking power for CP2.
LD3
         Unused
LD4
LD5      Output signal of starting data acquisition to CP2.
LD6      Output signal of external timing to CP2.
LD7      Output signal of enabling exposure frame to CP2.
LD8
LD9
         Unused
LD10
LD11
LD12     Output signal of bust start to UD.
LD13     Output signal of command pulse fluoroscopy to UD.
LD14     Output signal of step movement for reverse to UD.
LD15     Output signal of step movement for bringing forward to UD.
LD16     Output VTR signal for recording fluoroscopy to UD.
LD17     Output blurring singal of RSM to UD.
LD18     Unused
LD19     Output signal of command moving MH toUD.
LD20     Unused
LD21     Output signal of foot switch for caliblation to UD.
LD22     Unused
LD23     Output signal of command radiography to UD.
LD24     Unused
LD25     Output signal of hand switch for unattended test to UD.
LD26     Output signal of command actuating injector to UD.
LD27     Ethernte 10M BASE communication.
LD28     Ethernte 100M BASE communication.
LD29     Ethernte 1000M BASE communication.
LD30     Specification of Ethernet communication.
         Lighting : full duplex
         Extinction : half duplex
LD31     Send data by Ethenet.
LD32     Receive data by Ethernet.
LD33     During boot up ConfigFPGA(M4).
LD34     During boot up FPGA(M1,M2,M3).
LD35     Unused
LD36     Condition of serial communication in between FPGA.
LD37     Caution signal of FPD cooling system.
LD38     Caution signal of IOF FAN
LD39     Caution signal of I-SYS FAN
LD40     LIVE display of UD.
                                              3-75                                  M517-E610
=Installation of DAR-9500f (ACQ3)=
LD29
                                                                   LD20
                                                                           LD8
                                                                   LD19
                                                                           LD7
                                         LD1                       LD18
                       LD30                                                LD6
                                                                   LD17
                                                                           LD5
                         LD32          LD31        LD2       LD3     LD4
M517-E610                                 3-76
                                                      =Installation of DAR-9500f (ACQ3)=
Number                           Explanation
LD1      FPGA Configuration is completed.
LD2      DVI In-1 input signal is detected.
LD3      DVI In-2 input signal is detected.
LD4      DVI In-3 input signal is detected.
LD5      DVI In-4 input signal is detected.
LD6      DVI In-5 input signal is detected.
LD7      DVI In-6 input signal is detected.
LD8      DVI In-7 input signal is detected.
LD9      DVI CH1A is outputting.
LD10     DVI CH1B is outputting.
LD11     DVI CH2A is outputting.
LD12     DVI CH2B is outputting.
LD13     DVI CH3A is outputting.
LD14     DVI CH3B is outputting.
LD15     DVI CH4B is outputting.
LD16     DVI CH4A is outputting.
LD17     DVI In-4 input signal is detected (H/W).
LD18     DVI In-5 input signal is detected (H/W).
LD19     DVI In-6 input signal is detected (H/W).
LD20     DVI In-7 input signal is detected (H/W).
LD21     DVI CH1A is outputting (H/W).
LD22     DVI CH1B is outputting (H/W).
LD23     DVI CH2A is outputting (H/W).
LD24     DVI CH2B is outputting (H/W).
LD25     DVI CH3A is outputting (H/W).
LD26     DVI CH3B is outputting (H/W).
LD27     DVI CH4A is outputting (H/W).
LD28     DVI CH4B is outputting (H/W).
LD29     0: Maintenance program is running.
         1: Firmware is running.
         2: VGA VCR maintenance program is running.
LD30     SUB Board is detected.
LD31     OUT1 Board is selected.
LD32     OUT2 Board is selected.
                                          3-77                              M517-E610
=Installation of DAR-9500f (ACQ3)=
LD18
LD36
                  LD31
                                                                  LD17
                                                    LD14
                                                    LD13
                                                    LD15
                  LD32
                                                                  LD16
LD33
M517-E610                                    3-78
                                                       =Installation of DAR-9500f (ACQ3)=
Ref.                                    Explanation
 No.
LD1    Lighting   Clock Detection of Input Equalizer CH4IN (H/W)
LD2    Lighting   Clock Detection of Input Equalizer CH5IN (H/W)
LD3    Lighting   Clock Detection of Input Equalizer CH6IN (H/W)
LD4    Lighting   Clock Detection of Input Equalizer CH7IN (H/W)
LD5    Lighting   Clock Detection of Output Equalizer CH1A (H/W)
LD6    Lighting   Clock Detection of Output Equalizer CH1B (H/W)
LD7    Lighting   Clock Detection of Output Equalizer CH2A (H/W)
LD8    Lighting   Clock Detection of Output Equalizer CH2B (H/W)
LD9    Lighting   Clock Detection of Output Equalizer CH3A (H/W)
LD10   Lighting   Clock Detection of Output Equalizer CH3B (H/W)
LD11   Lighting   Clock Detection of Output Equalizer CH4A (H/W)
LD12   Lighting   Clock Detection of Output Equalizer CH4B (H/W)
LD13   Blinking   Three FPGA serial communications operate normally
LD14   Lighting   Unused
LD15   Lighting   Unused
LD16   Lighting   DVI CH1 Input Signal Detection
LD17   Lighting   DVI CH2 Input Signal Detection
LD18   Lighting   DVI CH3 Input Signal Detection
LD19   Lighting   DVI CH4 Input Signal Detection
LD20   Lighting   DVI CH5 Input Signal Detection
LD21   Lighting   DVI CH6 Input Signal Detection
LD22   Lighting   DVI CH7 Input Signal Detection
LD23   Lighting   DVI CH1A Output
LD24   Lighting   DVI CH1B Output
LD25   Lighting   DVI CH2A Output
LD26   Lighting   DVI CH2B Output
LD27   Lighting   DVI CH3A Output
LD28   Lighting   DVI CH3B Output
LD29   Lighting   DVI CH4A Output
LD30   Lighting   DVI CH4B Output
LD31   Lighting   DVI-SUB Board Detection
LD32   Lighting   DVI-OUT Board 1 Detection
LD33   Lighting   DVI-OUT Board 2 Detection
LD36   Lighting   5V Input Detection
                                           3-79                              M517-E610
=Installation of DAR-9500f (ACQ3)=
   PERIP-SH
        Number                            Explanation
        LD1        Flashes during normal F/W operation.
        LD2        Flashes during ARCNET communication,
        LD3        ON during ARCNET is connected.
                   Flashes when malfunction occurs.
M517-E610                                     3-80
                                                                  =Installation of DAR-9500f (ACQ3)=
   PERIP-SH2
        Number                            Explanation
        LD1        Flashes during normal F/W operation.
        LD2        Flashes during ARCNET communication,
        LD3        ON during ARCNET is connected.
                   Flashes when malfunction occurs.
                                                     3-81                               M517-E610
=Installation of DAR-9500f (ACQ3)=
M517-E610                                   3-82
4. Installation of DAR-9500f Bi-Plane System (ACQ3L CABINET)
                                                   NOTE
  Installation of DAR-9500f Bi-plane system is described in this manual.
  Install Frontal cabinet according to chapter 3 first, and then work on this chapter.
  To install Bi-plane system with hospital which already installed DAR-9500f Single-plane system,
  refer to chapter15 to add ACQ3 cabinet (Frontal) and then install Bi-plane system according to this
  chapter.
                                                         4-1                                M517-E610
=Installation of DAR-9500f (ACQ3)=
4) IVR NEO
                                  SYS-DISP PC
                                                                                     5) Examination Room Monitor
                                                                 ACQ PC L
                                      ACQ PC F
     3) Keyboard / Mouse
                                                                                           13) Distribution BOX 2
                                      UPS UNIT
8) System Display
M517-E610                                             4-2
Components
                                                       Size (mm)              Weight
  No.                 Name                                                               Installation Location
                                                  HeightxWidthxDepth           (kg)
        REF CABINET
  1)
        (Digital Reference Cabinet)
                                           522×229×647                       35          Control Room
        ACQ3 CABINET
  6)
        (Digital Acquisition Cabinet)
                                           1810x830 x 490                    250         Equipment Room
        SYSTEM DISPLAY
  8)
        (X RAY CONSOLE)
                                           380x400x175                       7           Control Room
        ACQ3L CABINET
  12)
        (Digital Acquisition Cabinet)
                                           1810x800x400                      200         Equipment Room
Except the components above, there are 1set of KEYBOARD and MOUSE, and LAN Cables attached to ROUTER and HUB. These are just
spare components so not be used.
                                                                 4-3                                             M517-E610
   =Installation of DAR-9500f (ACQ3)=
Compare with Single-plane system, the following devices are added to ACQ3 cabinet.
I-SYS + IOF L
CP2 L
HUB L
                                                                                            ADDITIONAL
                                                                                            Bi TRANS UNIT
DC POWER (5V)
   M517-E610                                    4-4
  4.1.2 Configuration of ACQ3 CABINET
DVI MAIN L
ACQ PC L
TRANS UNIT
                                         4-5      M517-E610
=Installation of DAR-9500f (ACQ3)=
Remove the front cover of ACQ3 Cabinet and set the following components inside the cabinet.
    Use two-way tap to connect the power cable of HUB with the same socket for single-plane HUB. And
    also, use an adaptor to space the earth terminal from CP2 socket.
Single-plane Bi-plane
    Make about 50mm space to the right and left side for cooling air vent of cabinet side and make about
    1m space in front of the cabinet to be able to work with DOOR opened when maintenance.
Remove the front cover of ACQ3L Cabinet and set the following components inside the cabin
M517-E610                                       4-6
4.2.3 Installation of DVI OUT (In case of MTA-40C)
ADAPTER BASE
Refer to “Installation manual for MH-400” for mounting FPD on C-arm and wiring cable.
                                                     4-7                                    M517-E610
=Installation of DAR-9500f (ACQ3)=
    To prevent the cabinet from falling down, fix it against the wall, using provided fixing plates and the
    following screws.
           <accessories 3 (502-28564)>
              No.                      Name                   Parts Number        PC
              1      Fixing Plate                        565-11472                 2
              2      Anchor,HKD-S M8X30                  026-50810-02              2
              3      BOLT, SST HEXSOCET M8X25            022-27157                 2
              4      WASSHER, SST PLAIN M8               023-66180                 2
              5      WASSHER, SST SPRING 2 M8            023-77080                 2
M517-E610                                       4-8
4.4 Connection Dpending on Supply Voltage
   Change connection of the input wire to the transformer in ACQ3L cabinet and optional Trans BOX as
   follows depending on the supply votage.
                                            CAUTION
  Be sure to make the connection change mentioned here in accordance with the observed supply
  voltage at the installation site.
                                                     ACQ3L Cabinet
                        Supply Votage                 Transformer
                                                         [AIN]
                    240V +10%,-10%                        240
                    230V +10%,-10%                        230
                    220V +10%,-10%                        220
                    200V +10%,-10%                        200
[AIN]
                                                    4-9                                 M517-E610
                                                                                            [31-4]                 Equipment Room
                                      SYSDISP PC2
                                                                                                                                                    Control Room
                                                                  [31-5]
M517-E610
                                                                                                                           [30-1]
                                                                    I-SYS
                                                                                                                                                       SYSDISP I/O
                                                                    BOARD (F)
                                                                                                             [31-6]
                                     DAR-9500f
                                                                                                                                                      JOINT BOX2
                                     ACQ3 Cabinet                                                [31-1]   D150GC-40
                                                                    I-SYS                                 Cabinet
                                                                                                 [31-3]
                                                                    BOARD (L)
                                                                                                          (L)                                        [23-3]            [23-4]
                                                                                                                                                                                        4.5.1 Connection Diagram
                                                                                                                               D150GC-40
                                                                                             [32-1]                            Cabinet
                                                                                                                               (F)         L-Ref                              L-ACQ
                                                                                        [32-2]            [31-2]                           Monitor                            Monitor
                                                                                                                                                                                                                                     =Installation of DAR-9500f (ACQ3)=
                                              CP2(L)                                                                  [31-7]
                                                                            HUB(L)                                                                   [24-6]          [24-7]
                                                Transformer
                                                Unit                                                      R-SYS ACQ CONV
                                                                            UPS                           BOARD
                                                                                                                                                                                        4.5 Cable Connection (External Connection)
4-10
                                                                                                          ACQ PC (L)
                                             [28-1]                                              [32-3]
            Examination Room                                                   [32-4]
                                                            4-11                              M517-E610
=Installation of DAR-9500f (ACQ3)=
M517-E610                                4-12
 In case of DVI MAIN3
                         4-13   M517-E610
   =Installation of DAR-9500f (ACQ3)=
   M517-E610                                            4-14
   <Examination Room to Eqipment Room>
  Ref.                                              Length
                   Name                Parts No.                       From              To
  No.                                                [m]
27-1     OUT-MAIN                    562-27540-13            MONITOR SUPPORT   ACQ3 CABINET DOOR
         CONNECTING CABLE 2                                  /DVI OUT BOX      DVI MAIN L
                                                    30
         (F MAIN ②)                                          DVI OUT ASSY L    [②]
         (F OUT ①)                                           [①]
27-2     OUT-MAIN                    562-27540-15            MONITOR SUPPORT   ACQ3 CABINET DOOR
         CONNECTING CABLE 4                                  /DVI OUT BOX      DVI MAIN L
                                                    30
         (F MAIN ④)                                          DVI OUT ASSY L    [④]
         (F OUT ②)                                           [②]
27-3     OUT-MAIN                    562-27331-06            MONITOR SUPPORT   ACQ3 CABINET DOOR
         POWER CABLE 2                                       /DVI OUT BOX      DVI MAIN L
                                                    30       DVI OUT ASSY L    [○B]
                                                             [○B]
   <Control Room>
  Ref.                                              Length
                  Name                 Parts No.                       From              To
  No.                                                [m]
24-6      DVI CABLE, KC-DVI-5        088-52848-70            REF CABINET       REF MONITOR L
                                                    5        DVI OUT ASSY L
                                                              [REF1 MONITOR]
24-7     DVI CABLE, KC-DVI-5         088-52848-70            REF CABINET       ACQ MONITOR L
                                                    5        DVI OUT ASSY L
                                                             [ACQ MONITOR]
23-3     MONITOR Attachment          (Attached to            REF MONITOR L     JOINT BOX2 OUTLET TAP
                                                    -
         POWER CABLE                 MONITOR)
23-4     MONITOR Attachment          (Attached to            ACQ MONITOR L     JOINT BOX2 OUTLET TAP
                                                    -
         POWER CABLE                 MONITOR)
   <Examination Room>
  Ref.                                              Length
                   Name                Parts No.                       From              To
  No.                                                [m]
27-4     MONITOR Attachment          (Attached to            MONITOR SUPPORT   MONITOR SUPPORT
         DVI CABLE                   MONITOR)                /DVI OUT BOX      /DVI OUT BOX
                                                    -
                                                             DVI OUT ASSY L    REF MONITOR L
                                                             [REF1 MONITOR]
27-5     MONITOR Attachment          (Attached to            MONITOR SUPPORT   MONITOR SUPPORT
         DVI CABLE                   MONITOR)                /DVI OUT BOX      /DVI OUT BOX
                                                    -
                                                             DVI OUT ASSY L    ACQ MONITOR L
                                                             [ACQ MONITOR]
25-2     MONITOR Attachment          (Attached to            MONITOR SUPPORT   MONITOR SUPPORT
         POWER CABLE                 MONITOR)       -        /DVI OUT BOX      /DVI OUT BOX
                                                             REF MONITOR L     MULTI TAP
25-3     MONITOR Attachment          (Attached to            MONITOR SUPPORT   MONITOR SUPPORT
         POWER CABLE                 MONITOR)       -        /DVI OUT BOX      /DVI OUT BOX
                                                             ACQ MONITOR L     MULTI TAP
                                                               4-15                       M517-E610
=Installation of DAR-9500f (ACQ3)=
                       DVI                  POWER
                       INPUT                INPUT   Turn on the MONITOR power supply.
M517-E610                            4-16
                                           NOTE
Refer to the following figure to check that cables (POWER CABLE) [27-4] [27-5] and [27-3] are
connected properly.
LED
DVI MAIN 2
LED
DVI MAIN 3
If LEDs blink, cables have been connected properly. If do not blink, turn off the power of
DVI-MAIN immediately, and check that cable mark and board silk are matched.
                                                 4-17                               M517-E610
=Installation of DAR-9500f (ACQ3)=
  Cable to FPD
  Ref.                                               Length
                   Name             Parts No.                           From                To
  No.                                                 [m]
28-1       FPD CABLE              Attached      to            ACQ3 CABINET 4F     Lateral C-arm
                                                     40
                                  FPD                         CP2 L               FPD MAIN BODY
                                                          NOTE
       On Bi-plane system, be sure to change the power cable for Frontal FPD cooling unit from ACQ3
       cabinet to ACQ3L cabinet because of electricity capacity.
M517-E610                                            4-18
 4.5.5 Cable Connection (ACQ3 CABINET-ACQ3L CABINET)
 Ref.                                                        Length
                    Name                    Parts No.                             From                        To
 No.                                                          [m]
32-1      R-SYS CONV CABLE             563-67036-02                       ACQ3 CABINET DOOR        ACQ3L CABINET DOOR
                                                             10           I-SYS L                  R-SYS ACQ CONV CN2
                                                                          CN8
32-2      SOLIOS-R/ISYS CABLE          562-28190-11                       ACQ3 CABINET DOOR        ACQ3L CABINET
                                                             10           I-SYS L                  ACQ PC L CL0/CL1
                                                                          CN3
32-3      LAN CABLE 3                  565-10447-11                       ACQ3 CABINET             ACQ3L CABINET
                                                             10
          HUB L 1                                                         HUB L 1                  ACQ PC LAN
32-4      ACQ 3L PC                    565-11476-01                       ACQ3 CABINET             ACQ3L CABINET
                                                             10
          POWER CABLE                                                     UPS [ACQ PC L]           ACQ PC POWER
32-5      SERIAL CABLE, ISYS L         565-11389-01                       ACQ3 CABINET DOOR        ACQ3L CABINET DOOR
                                                             10           I-SYS L                  DVI MAIN J30
                                                                          CN9
32-6      LAN CABLE DVI L              562-28024-11                       ACQ3 CABINET DOOR        ACQ3L CABINET DOOR
(Note)                                                       22           I-SYS F                  DVI MAIN
                                                                          CN31                     DVI IN7 [J16]
32-7      POWER RELAY CABLE, BI        563-71712                          ACQ3 CABINET             ACQ3L CABINET
                                                             10           TRANSFORMER UNIT         TRANSFORMER UNIT
                                                                          J103                     J303
(Note) LAN CABLE DVI (connected to DVI MAIN F) which is connected to I-SYS F CN31 should be connected to I-SYS L.
                                                                              LA-1
              32-1                                                        (Shield clamp)
         (Shield clamp)
RSYS-ACQ CONV
32-3
SLOT 7 6 5 4 3 2 1
CL0
CL1
                                                ACQ PC L 背面
  32-4                                        ACQ PC L REAR
                                                                                            32-2
                                                                   4-19                                         M517-E610
 =Installation of DAR-9500f (ACQ3)=
 M517-E610                                                     4-20
                      DAR I/F 2002
                                             J4
                                                                31-3
                                          COM I/F 2002
J34 DAR
                       D150GC-40L
31-4      30-1                                           31-1
                                                                       FLUO-2002
31-6
UD CONT-2002
                                                         4-21                      M517-E610
=Installation of DAR-9500f (ACQ3)=
                                                                                              DAR-9500f
  I-SYS L                 I-SYS F    ACQ               SYSDISP                                ACQ3 Cabinet
                                     PC F              PC
                                                                                                   IVR NEO I/F
                                    J5     J6           J6       J5
  CN18                    CN18                                                                            CN14
    CN17                    CN17
J4 J3 J4 J3 CN5
                                                                                                                           REF PC
                                                         ROUTER
         ACQ PC F ASSY
                                                             4   3    2   1               WAN
                                                                                                                         LAN
         LAN
                                     MPC
                CP2 F
                           LAN           LAN
                                                                              1       3        5      7
                                                          HUB F
                                                                                                                         LAN
 CP2 L                                                                            2       4       6       8
                                                                                                                   CN4
               LAN
                                                                                                                          I-SYS
                                                                                                                          BOARD F
         ACQ PC L ASSY
                                                                              1       3       5       7
                                                          HUB L
         LAN
                                                                              2       4       6       8
                                                                                                                         LAN
                                                                                                                   CN4
                                                                                                                          I-SYS
   ACQ3L Cabinet
                                                                                                                          BOARD L
M517-E610                                                    4-22
4.6 ACQ3L Cabinet Internal Cable Connection
 Ref.                                              Length
                Name              Parts No.                       From                     To
  No.                                               [m]
 LA-1   ACQ CONVCABLE       563-67035-02                    [LA-2]            R-SYS ACQ CONV
                                                   1.3
                                                            CABLE CONNECTOR   [CN1]
 LA-2   IO CONV CABLE       562-28135-03                    ACQ PC L REAR     [LA-1]
                                                   0.3
                                                            SLOT4             CABLE CONNECTOR
 LA-3   KEYBOARD            (Attached to ACQ PC)            ACQ PC L          Keyboard Holder
                                                   -        KEYBOARD
                                                            CONNECTOR
 LA-4   MOUSE               (Attached to ACQ PC)            ACQ PC L          Keyboard Holder
                                                   -
                                                            USB CONNECTOR
 LA-5   ACQ INTERLOCK       565-10463                       ACQ PC L          TRANSFORMER UNIT
                                                   2
        CABLE 3                                             SLOT5             [J302]
 LA-6   DVI CABLE 3 ACQ L   565-10467-12                    ACQ PC L          DVI MAIN L
                                                   2        SLOT6             DVI IN1 [J10]
                                                            DP-DVI CABLE      [△1]
 LA-7   DOSEMETER,SERIAL    565-10449-01                    ACQ PC L          DOOR
                                                   1.5
        CABLE                                               SLOT1 [P2]        SERIAL(Dosemeter)
                                                         4-23                            M517-E610
=Installation of DAR-9500f (ACQ3)=
 M720
            Keyboard     Mouse
            (LA-3)       (LA-4)
 M730
                                                                    LA-1
                                         Dongle
                         32-3            for Stent View(Option)          LA-2
SLOT 7 6 5 4 3 2 1
CL0
CL1
          32-4     キーボード
                  Keyboard      マウス
                                Mouse
                  (LA-3)
                   (LA-3)       (LA-4)
                                (LA-4)         ACQ PC
                                               ACQ PC L
                                                      L Rear
                                                        背面          LA-6        LA-5          32-2   LA-7
                                                      NOTE
   Be sure to attach the dongle (562-28609-03 Dongle, ACQ L) for DAR application license which
   provided in the pack of ACQ PC L, to USB connector of ACQ PC L. (For M730, the dongle is
   mounted insdie PC) Or DAR application will not start. And also attach the license dongle for Stent
   View (Option).
M517-E610                                          4-24
4.7 ACQ3L Cabinet Internal Cable Connection (Connected at Factory Setting)
 Ref.                                          Length
                Name              Parts No.                     From                    To
  No.                                           [m]
 LZ-1   FAN ASSY,TOP 2      563-70260-10                 TRANSFORMER UNIT   Cabinet Top FAN
                                               -
                                                         [LOUT1 LOUT2]
 LZ-2   AC CABLE,ACQ3L      565-11230                    TRANSFORMER UNIT   DC POWER PS3
                                               3.0
                                                         [LOUT5 LOUT6]      [CN1]
 LZ-3   DC12V CABLE,ACQ3L   562-28385                    TRANSFORMER UNIT   R-SYS ACQ CONV [CN3]/
                                               2.5
                                                         [J301]             DOSEMETER POWER
 LZ-4   DC5V CABLE,ACQ3L    565-11229                    DC POWER PS3       DVI MAIN L [J1]/
                                               0.25      [CN2]              R-SYS ACQ CONV
                                                                            [CN4]
 LZ-5   EARTH CABLE,ACQ3L   565-10472-03                 TRANSFORMER UNIT   IVR-NEO I/F
                                               2.0
                                                         [J105]             [CN2]
 LZ-6   I-SYS FAN ASSY L    565-12323-02                 No.[Z-12]          No.[29-2]
                                               0.6
                                                         [J212]CONNECTOR    [J213]CONNECTOR
 LZ-7   IOF FAN ASSY L      565-12322-02                 No.[Z-14]          No.[29-2]
                                               0.4
                                                         [J210]CONNNECTOR   [J211]CONNECTOR
                                                      4-25                            M517-E610
=Installation of DAR-9500f (ACQ3)=
                                                  NOTE
   I-SYS BOARD settings of Bi-plane system are described in this section. Refer to section 3.12 for
   Frontal I-SYS BOARD.
      DIP-SW1
                  1       2      3     4      5      6      7        8
       OFF        ○                    ○             ○      ○
       ON                ○      ○             ○                      ○
      DIP-SW2
                  1      2      3      4      5      6      7        8
       OFF        ○      ○      ○      ○      ○      ○      ○        ○
       ON
      DIP-SW3
                  1      2      3      4      5      6      7        8
       OFF        ○      ○      ○      ○      ○      ○      ○        ○
       ON
                                                                         SW1
      Push Switch
                                                                         SW2
SW3
M517-E610                                     4-26
(1) Contents of DIP SW
   SW1
      SW1                   Contents                        OFF                          ON
         1     Startup Software                   Firmware                 Maintenance Program
         2     IQP                                Rev.2.2.0 less than      Rev.2.2.0 or more
         3     FPGA Config                        JTAG Cable               Flash Memory
         4     Where to store FPGA Config file    Regular                  Backup
               during configuration by Flash.
         5     IOF ON/OFF                         OFF                      ON
         6     Push Switch                        Enable                   Disable
               FPGA reconfiguration
         7     Push Switch                        Disable                  Enable
               Software reset
         8     Select Frontal/Lateral             Frontal                  Lateral
   SW2
      SW2                  Contents                         OFF                       ON
         1
         2
         3     Unused
         4
         5
         6     DAR IO Communication               CURRENT LOOP             LVTTL
         7
               Unused
         8
   SW3
      SW3                   Contents                        OFF                       ON
         1     Unused
         2     Unused
         2     Baud rate of maintenance program   (S3_2:S3_3)
         3                                        OFF : OFF = 115200 bps
                                                  OFF : ON = 230400 bps
                                                  ON : OFF = 460800 bps
                                                  ON : OFF = 921600 bps
         4     Debug of Ethernet                  OFF                      ON
               Hardware debug
         5
         6
               Unused
         7
         8
                                                     4-27                               M517-E610
=Installation of DAR-9500f (ACQ3)=
                                                NOTE
   DVI MAIN BOARD settings of Bi-plane system are described in this section. Refer to section 3.12
   for Frontal DVI MAIN2 BOARD.
      DIP-SW1
                  1     2      3     4      5       6    7       8
       OFF        ○     ○            ○      ○       ○    ○       ○
       ON                      ○
      DIP-SW2
                  1     2      3     4      5       6    7       8
       OFF        ○     ○      ○     ○      ○       ○    ○       ○
       ON
      DIP-SW7
                  1     2      3      4     5       6    7       8
       OFF              ○      ○            ○       ○    ○       ○
       ON         ○                  ○
SW7
SW1 SW2
M517-E610                                    4-28
(1) Contents of DIP SW
   SW1
      SW1                   Contents                        OFF               ON
         1     Startup Software                  Firmware         Maintenance Program
         2     Unused
         3     FPGA Config                       JTAG Cable       Flash Memory
         4     Where to store FPGA Config file   Regular          Backup
               during configuration by Flash.
         5
               Unused
         6
         7     VCR Output                        NTSC             PAL
         8     Unused
   SW2
      SW2                   Contents                        OFF               ON
         1     Startup Software                  Firmware         VGA VCR
                                                                  Maintenance Program
         2
         3
         4
         5     Unused
         6
         7
         8
                                                   4-29                          M517-E610
=Installation of DAR-9500f (ACQ3)=
   SW7
     Signal amplification level of DVI-MAIN input channel (from J13 to J16) can be changed using DIP
    SW7. Amplification level can be set with 8 kinds of level (from 000 to 111).
M517-E610                                     4-30
4.9.3 DVI MAIN3 BOARD(L)
                                                    NOTE
  DVI MAIN3 BOARD settings of Bi-plane system are described in this section. Refer to section 3.12
  for Frontal DVI MAIN3 BOARD.
     DIP-SW1
                  1      2      3       4      5      6          7   8
                                             (*1)
      OFF        ○                     ○      ○      ○       ○       ○
      ON                 ○      ○
     DIP-SW2
                 1       2      3      4      5       6          7   8
      OFF        ○       ○      ○      ○      ○
                                                            (Note)
      ON
     DIP-SW3
                 1       2      3      4      5      6       7       8
      OFF        ○       ○      ○      ○      ○      ○       ○       ○
      ON
     DIP-SW7
                  1      2      3       4     5      6       7       8
      OFF                ○      ○             ○      ○       ○       ○
      ON         ○                     ○
(*1) With SMART Display option and LIVE Monitor option, set the segment5 of DIP SW1 to ON.
(Note) DIP SW is aleready set at the factory. If changing is necessary, refet to section 3.12.
                                                          4-31                                  M517-E610
=Installation of DAR-9500f (ACQ3)=
LD18
LD36
                               LD31
                                                                                  LD17
                                                SW1
                                                                    LD14
                                                                    LD13
                                                     SW2
                                                                    LD15
                               LD32                  SW3                          LD16
LD33 SW7
      <SW1>
       SW1                 Description                        OFF                               ON
         1     Startup Sotware                       Firmware                       Maintenance Program
         2     Selection of DVI-MAIN Type            DVI-MAIN 2                     DVI-MAIN 3
         3     FPGA Config                           JTAG Cable                     Flash Memory
         4     Where to store FPGA Config file       Regular                        Backup
               during configuration by Flash.
         5     SMART Display Option and
                                                     Disable                        Enable
               LIVE Monitor Option
         6     Push Switch                           Enable                         Disable
               FPGA reconfiguration
         7     VCR Output                            NTSC                           PAL
         8     Unused
M517-E610                                     4-32
<SW2>
  SW2                 Description                           OFF                          ON
    1     Startup Software                       Firmware                    VGA VCR
                                                                             Maintenance Program
    2     Unused
    3     Phase shifting check mode of REF Regular                   When checking
    4     PC startup signal and NTSC/PAL
    5     signal.
    6     Selecting phase of REF PC startup 6 OFF /7 OFF /8 OFF: Phase 315˚
    7     signal and NTSC/PAL signal.       6 ON /7 OFF /8 OFF: Phase 270˚
    8                                       6 OFF /7 ON /8 OFF: Phase 225˚
                                            6 ON /7 ON /8 OFF: Phase 180˚
                                            6 OFF /7 OFF /8 ON: Phase 135˚
                                            6 ON /7 OFF /8 ON: Phase 90˚
                                            6 OFF /7 ON /8 ON: Phase 45˚
                                            6 ON /7 ON /8 ON: No phase shift
<SW3>
Signal amplification level of DVI-MAIN3 input channel (from J20 to J27) can be changed using DIP
SW3.
  SW3                  Description                     OFF                          ON
    1     Setting amplification level       of   OFF: 1
    2     DVI-MAIN3 J20 and J27.                 ON: 0
    3                                            (Note 1)
    4     Setting Output Level                   1000 mV                     1200 mV
    5     Setting Enabled Channel of Output      (Note 2)
    6
    7     Unused
    8     Unused
                                                   4-33                                   M517-E610
=Installation of DAR-9500f (ACQ3)=
     <SW7>
     Signal amplification level of DVI-MAIN 3 input channel (from J13 to J16) can be changed using DIP
     SW7.
       SW7                  Description                     OFF                           ON
         1     Setting amplification level    of OFF: 1
         2     DVI-MAIN3 J13 and J14.            ON: 0
         3                                       (Note)
         4     Setting amplification level    of OFF: 1
         5     DVI-MAIN3 J15 and J16.            ON: 0
         6
         7     The choice of how to boost the       Enable the setting of      Enable the value of the internal
               input equalizer CH4IN ~ CH7IN        DIP-SW7-1 ~ 7-6            register
         8     The choice of how to boost the       Enable the setting value   Enable the value of the internal
               output equalizer CH1A(B) ~           of the external pin        register
               CH4A(B)
M517-E610                                    4-34
4.9.4 IVR NEO I/F, PERIP-SH BOARD
    For the system with IVR Shuttle, set ON the DIP-SW 2-5 for Bi-plane mode.
    Check that PERIP-SH DIP SW is set as the following table.
      DIP-SW2
                  1      2 (*)   3      4    5 (*)   6       7
       OFF        ○       ○                   ○      ○       ○
       ON                        ○     ○
      (*1) If there is IVR Shuttle (Option), it is need to set to ON segment 2 and 5of the DIP-SW2 on the
      PERIP-SH.
      (*2) If there is Bi-plane system with IVR Shuttle (Option), it is also need to set to ON segment 5 of
      the DIP-SW2 on the PERIP-SH2
      DIP-SW3
                  1       2      3      4
       OFF        ○       ○      ○
       ON                              ○
   SW2
      SW2                    Contents                            OFF                         ON
         1      Unused
         2      IVR Shuttle                          None                       When using IVR Shuttle
         3      Sound of fluoroscopy                 ON                         OFF
         4      Sound of radiography                 ON                         OFF
         5      IVR Shuttle for Biplane              None                       When using IVR Shuttle for
                                                                                Biplane
         6
                Unused
         7
   SW3
      SW3                   Contents                             OFF                         ON
         1      Unused
         2      F/W writing                          OFF                        ON
         3      Unused
         4      ARCNET communication rate            12MHz                      10MHz
   Jumper
    JP1 IN: ARCNET termination ON (Normally)
       OUT: ARCNET termination OFF
                                                         4-35                                 M517-E610
=Installation of DAR-9500f (ACQ3)=
       DIP-SW1
                  1       2      3      4      5      6      7
        OFF       ○       ○      ○      ○      ○      ○      ○
        ON
       DIP-SW1
                  1       2      3      4      5      6      7
        OFF       ○              ○      ○      ○      ○      ○
        ON                ○
(2) PERIP-SH2
     Check that PERIP-SH2 DIP SW is set as the following table.
       DIP-SW2
                  1     2 (*)    3      4     5 (*)   6      7
        OFF       ○      ○                     ○      ○      ○
        ON                       ○      ○
      (*1) If there is IVR Shuttle (Option), it is need to set to ON segment 2 of the DIP-SW2 on the
      PERIP-SH2.
      (*2) If there is Bi-plane system with IVR Shuttle (Option), it is also need to set to ON segment 5 of
      the DIP-SW2 on the PERIP-SH2
       DIP-SW3
                  1       2      3      4
        OFF       ○       ○      ○
        ON                              ○
M517-E610                                      4-36
The following figure is showing the position of IVR NEO I/F2 and PERIP-SH2 DIP SW.
SW2
JP1
SW3
SW1
                                                4-37                                 M517-E610
=Installation of DAR-9500f (ACQ3)=
   SW1
      SW1                  Contents                        OFF                        ON
         1     IVR NEO                            (S1_1:S1_2)
         2                                        OFF : OFF = IVR NEO (503-63818-01) When connecting
                                                  OFF : ON = IVR NEO (503-63818-11) When connecting
         3
         4
         5
               Unused
         6
         7
         8
   SW2
      SW2                   Contents                      OFF                         ON
         1     Unused
         2     IVR Shuttle                        None                    When using IVR Shuttle
         3     Sound of fluoroscopy               ON                      OFF
         4     Sound of radiography               ON                      OFF
         5     IVR Shuttle for Biplane            None                    When using IVR Shuttle for
                                                                          Biplane
         6
               Unused
         7
   SW3
      SW3                  Contents                       OFF                         ON
         1     Unused
         2     F/W writing                        OFF                     ON
         3     Unused
         4     ARCNET communication rate          12MHz                   10MHz
   Jumper
    JP1 IN: ARCNET termination ON (Normally)
       OUT: ARCNET termination OFF
M517-E610                                  4-38
4.10 Installation of LIVE Monitor Option
    This chapter describes installation of adding LIVE Monitor Option. Skip this chapter if there is no
    LIVE Monitor Option.
2. Remove DVI OUT board and a blinder plate from DVI OUT ASSY.
3. Connect DVI OUT board on DVI OUT ASSY and optional DVI-OUT board.
                                                      4-39                                 M517-E610
=Installation of DAR-9500f (ACQ3)=
Fix 4 screws.
M517-E610                                 4-40
(2) Cable Connection.
                             4-41   M517-E610
=Installation of DAR-9500f (ACQ3)=
  Ref.                                                          Length
                  Name                 Parts No.                               From                  To
   No.                                                           [m]
 40-1      LAN DVI CABLE         502-27540-23                            MONITOR SUPPORT   ACQ3L Cabinet F Door
                                                            30           DVI OUT ASSY L    DVI MAIN F
                                                                         [③]               [⑧]
 40-2      MONITOR Attachment    (Attached to LIVE                       MONITOR SUPPORT   MONITOR SUPPORT
           DVI CABLE             MONITOR)                   -            DVI OUT ASSY L    LIVE MONITOR
                                                                         [REF2 MONITOR]
 40-3      MONITOR Attachment    (Attached to LIVE                       MONITOR SUPPORT   MONITOR SUPPORT
                                                            -
           POWER CABLE           MONITOR)                                LIVE MONITOR      MULTI TAP
M517-E610                                            4-42
4.10.2 LIVE Monitor in Control Room (Optional)
Removing covers
  (1) Remove four screws where indicated by          (2) Open the top cover to right and left side.
  arrows.
                                                                 4-43                                               M517-E610
=Installation of DAR-9500f (ACQ3)=
2. Remove DVI OUT board and a blinder plate from DVI OUT ASSY.
3. Connect DVI OUT board on rear panel and optional DVI-OUT board.
Fix 4 screws.
M517-E610                                       4-44
(2) Cable Connection
                             4-45   M517-E610
=Installation of DAR-9500f (ACQ3)=
  Ref.                                                         Length
                  Name                  Parts No.                             From                  To
   No.                                                          [m]
 40-4      LAN DVI CABLE         502-27540-21                           REF Cabinet      ACQ3L Cabinet F Door
                                                           30           DVI OUT L        DVI MAIN F
                                                                        [③]              [⑦]
 40-5      DVI CABLE,            088-50905-16                           REF Cabinet L    LIVE MONITOR
                                                           5
           KC-DVI-5                                                     [REF2 MONITOR]
 40-6      MONITOR Attachment    (Attached to                           LIVE MONITOR F   JOINT BOX2
                                                           -
           POWER CABLE           MONITOR)                                                OUTLET TAP
M517-E610                                           4-46
                                                    =REF PC and ACQ PC Configuration=
This section describes the configuration procedure of REF PC and ACQ PC.
  Fill out the following pre-confirmation sheet before the configuration of REF PC, ACQ PC
  and Router.
  This is required for the configuration procedure.
                                          5-1                                  M517-E610
              = REF PC and ACQ PC Configuration =
              M517-E610                                                              5-2
                                                        =REF PC and ACQ PC Configuration=
*This system is equipped with three accounts (operator: normal user account, super:
user administrator mode, serviceapp: service mode). Normally when starting up, it is
operator mode. How to logon as super, serviceapp is described below.
*Unless specifically instructed, work as serviceapp for configurations in this section.
<<Logon REF-PC>>
 (1) Turn on the power supply of PC and wait for DAR application to startup. After application is
    startup, press [Ctrl]+[Alt]+[Delete] at the same time and the following dialog is displayed.
                                             5-3                                     M517-E610
= REF PC and ACQ PC Configuration =
     (3) Select user to login and enter the password. Enter the Password [voyager] when login with
        [serviceapp]. (Enter the Password [dar9k] when login with [super])
<<Logon ACQ-PC>>
(1) Move a mouse to the window of ACQ PC and click [Exit] on the side menu on application.
  M517-E610                                      5-4
                                                     =REF PC and ACQ PC Configuration=
(3) Select user to login and enter the password. Enter the Password [voyager] when login with
   [serviceapp]. (Enter the Password [dar9k] when login with [super])
                                          5-5                                      M517-E610
= REF PC and ACQ PC Configuration =
     (1) Double-click [Ultra VNC Viewer] on the REF PC desktop to display [UltraVNC Viewer]
        window. Enter VNC Server [DAR-9500ACQ-F-01] (For two or more PCs, change the last
        value as follows. DAR-9500ACQ-F-02, 03…) and then click [Connect].
(3) ACQ PC window is displayed, so click [Exit] on the side menu on application.
  M517-E610                                       5-6
                                                    =REF PC and ACQ PC Configuration=
5.3 PC configuration
PC configuration has been configured in the factory. Perform only in the needed case.
 Set [Date and Time] on MPC as the setting is synchronized with MPC when the system
 is started.
 See “Remote Maintenance Installation Manual (M517-E447)” for the detail.
 Click [Start] on the lower left corner of the window, and select [Control Panel].
 Double click [Date and Time] icon, and the following dialog will display.
 Select [Date & Time] tab and configure proper time.
Click [Change date and time] on [Date and Time] tab to set current date and time.
                                          5-7                                    M517-E610
= REF PC and ACQ PC Configuration =
     Select [Time Zone] tab and select (UTC-09:00) Osaka, Sapporo, Tokyo from the drop-
     down list.
     And click [OK] and close.
  M517-E610                                5-8
                                                   =REF PC and ACQ PC Configuration=
Click [Start] on the lower left corner of the window, and select [Control Panel].
Double-click [System] icon and the following dialog is displayed.
                                         5-9                                    M517-E610
= REF PC and ACQ PC Configuration =
  M517-E610                                   5-10
                                                   =REF PC and ACQ PC Configuration=
Click [Start] on the lower left corner of the window, and select [Control Panel].
                                         5-11                                   M517-E610
= REF PC and ACQ PC Configuration =
     Select [Internet Protocol Version4 (TCP/IPv4)] and click [Properties]. The following
     dialog is displayed.
     Check “Use the following IP address” and configure IP address, Subnet mask, and
     Default gateway as follows. (Configured before shipping)
  M517-E610                                   5-12
                                               =REF PC and ACQ PC Configuration=
ACQ-PC
IP address 192.168.100.22
Subnet mask 255.255.255.0
Default gateway 192.168.100.29
ACQ-F PC
IP address 192.168.100.22
Subnet mask 255.255.255.0
Default gateway 192.168.100.29
ACQ-L PC
IP address 192.168.100.25
Subnet mask 255.255.255.0
Default gateway 192.168.100.29
If hospital has two or more systems, set IP address, Subnet mask and Default gateway
as the following table.
When the inside hospital LAN is 192.168.100.x , change the third value such as
192.168.101.x.
                                       5-13                               M517-E610
= REF PC and ACQ PC Configuration =
<<Confirmation of Network>>
Input as follows on the command prompt, and press [Enter] key on the keyboard.
Ping X.X.X.X
     X.X.X.X is the IP address (e.g. 192.168.1.1) for such as server printer inside hospital.
     Please confirm that message of “Reply from...” is displayed as follows.
     If the correct message is not displayed, reconfirm Router configuration, LAN cable
     connection,
     Power of the Hub of inside hospital network, and entered IP address.
     When the network between patient information and image is different in the hospital
     [refer to DAR-9500f Network Connection Pre-Confirmation Sheet (Note 3)], execute
     the following one on command prompt.
     Note) If hospital has two or more systems, change the last number of route -p add
     ippaddress mask 255.255.255.255.192.168.100.28 to “48”, “68”.
  M517-E610                                   5-14
                                                    =REF PC and ACQ PC Configuration=
                                          5-15                                 M517-E610
= REF PC and ACQ PC Configuration =
     Click [Start] on the lower left corner of the window, and select [Control Panel].
     Double-click [Regional and Language…] icon, and the following dialog is displayed.
     Click [Formtas] tab and select appropriate item from dropdown list. (Corresponding
     languages are Japanese and English only.)
  M517-E610                                   5-16
                                                  =REF PC and ACQ PC Configuration=
Select appropriate item from dropdown list. Click [OK] and close the dialog.
                                       5-17                                    M517-E610
= REF PC and ACQ PC Configuration =
Select appropriate item from dropdown list. Click [OK] and close the dialog.
  M517-E610                                   5-18
                                                =REF PC and ACQ PC Configuration=
Click [OK] on [Regional and Language] dialog and close the dialog.
                                      5-19                             M517-E610
= REF PC and ACQ PC Configuration =
Press [Ctrl] + [Alt] + [Del] keys to display the following dialog, and start [Task Manager].
  M517-E610                                    5-20
                                                  =REF PC and ACQ PC Configuration=
Select [Application] tab and click [New Task...]. Open [Create New Task] and input
[intl.cpl]. And click [OK].
[Regional and Language Options] is opened. Select [Formats] tab and select
appropriate item fromdropdown list. (Corresponding languages are Japanese and
English only.)
Select [Location] tab and select appropriate item from dropdown list.
                                       5-21                                 M517-E610
= REF PC and ACQ PC Configuration =
Select appropriate item from dropdown list. And then click [OK] to close the dialog.
  M517-E610                                   5-22
                                                  =REF PC and ACQ PC Configuration=
Select appropriate item from dropdown list. And then click [OK] to close the dialog.
                                        5-23                                  M517-E610
= REF PC and ACQ PC Configuration =
Click [OK] on [Regional and Language] dialog and close the dialog.
  M517-E610                                 5-24
                                                    =REF PC and ACQ PC Configuration=
If hospital has two or more systems, configure setting of remote desktop to MPC as the
following.
                                          5-25                                 M517-E610
= REF PC and ACQ PC Configuration =
Click [Save].
  M517-E610                                 5-26
                                              =REF PC and ACQ PC Configuration=
                                      5-27                                M517-E610
= REF PC and ACQ PC Configuration =
    DAR-9500f can make an output file of images and loops in a network shared folder on
    user’s pc.
    To use this feature, it is required to configure an output destination in advance.
    Follow the instruction below for configuration.
    Logon as service user. (Refer to 5.2.1) Start up Configure Network Drive from 2.Setup
    Menu of Maintenance Menu (HTML) on Desktop.
    The following dialog is displayed. Enter the IP address desired as the computer to make
    an output file. Click [OK].
    With the column blank, if [OK] button of [Cancel] pressed, Network Drive Information
    exits making a file.
Enter the network shared folder name desired as an output file destination. Click [OK].
Enter user name accessible to the network shared folder described above. Click [OK].
  M517-E610                                   5-28
                                                      =REF PC and ACQ PC Configuration=
After completing entering step (1) to (4), confirmation dialog is displayed. Click [OK]. A
setting file is made that has information of output file destination.
                                           5-29                                  M517-E610
= REF PC and ACQ PC Configuration =
     After logon as service user (refer to 5.2.1), double-click “DAR-9500f” software icon on
     the desktop to execute application. The following window is displayed.
Side Menu
  M517-E610                                   5-30
                                                    =REF PC and ACQ PC Configuration=
                                         NOTE
Do not configure options not described in this chapter. Be sure to configure
mandatory items marked as “*”. Except for them, configure as needed.
For Bi-plane option, refer to “5.5.2 Bi-plane option”.
                                         5-31                                  M517-E610
= REF PC and ACQ PC Configuration =
REF:
ACQ:
  M517-E610                                   5-32
                                                 =REF PC and ACQ PC Configuration=
Link:
Item                        Configuration procedure
Link Mode*                  Check this item.
(Both REF and ACQ)
Linked Station*             Enter IP address “192.168.100.22” of ACQ PC (Frontal) to
(REF Only)                  REF PC.
                            If hospital has two or more systems, change IP address as
                            “192.168.100.42”, “192.168.100.62”.
Biplane                     Check if it is Bi-plane system.
A (Frontal)                 Refer to “5.5.2 Bi-plane option”.
B (Lateral)
Plane B Name
   “*” is mandatory item.
                                       5-33                                 M517-E610
= REF PC and ACQ PC Configuration =
    Primary Archive:
    Item                       Configuration procedure
    Manual Archive*            This item cannot be set.
    (REF Only)
    Primary Archive*           If not sending automatically to the image server after the
    (REF Only)                 acquisition, check this item. (If link mode is valid, it will
                               display in gray on ACQ PC)
    CD/DVD Copies*             With primary Archive on, if copying the configured number
    (REF Only)                 of CDs here, the delete-prohibit lock of the study data is
                               released.
                               (If link mode is valid, it will display in gray on ACQ PC)
      “*” is mandatory item.
      2)Click [Apply] and configure parameters.
  M517-E610                                 5-34
                                                    =REF PC and ACQ PC Configuration=
REF:
1) To add new DICOM host, choose in “Host Type”, click [New] button and configure
   parameters as follows. To know AE title and capabilities, DICOM Conformance
   document is useful.
                                         5-35                                   M517-E610
= REF PC and ACQ PC Configuration =
   Host Type*:
   Item                         Configuration procedure
   MWM Server*                  Check [MWM Server].
       “*” is mandatory item.
   Information:
   Item                         Configuration procedure
   AE Title*                    Enter AE Title of MWM Server. (Up to 16 characters)
   Host*                        Enter IP Address of MWM Server. (Up to 64 characters)
   Alias*                       Enter the name which is easy to identify. Do not enter more
                                than 11 characters.
   Port*                        Enter Port Number of MWM server.
   Packet size*                 Referring to DICOM Conformance of MWM server, if max
                                PDU size is described, enter that value. If it is not described,
                                enter 99280.
   File packets grouping*       Input “2”.
       “*” is mandatory item.
   DICOM Host*:
   Item                         Configuration procedure
   Yes/No                       Select “Yes” if the host is correspond to DICOM.
       “*” is mandatory item.
   Capabilities*:
   Item                         Configuration procedure
   Query                        Check [Query].
   Relational                   Check [Relational].
       “*” is mandatory item.
  M517-E610                                    5-36
                                                       =REF PC and ACQ PC Configuration=
 Host Type*:
 Item                       Configuration procedure
 Printer                    Check [Printer].
     “*” is mandatory item.
 Information:
 Item                      Configuration procedure
 AE Title*                 Enter AE Title of Printer. (Up to 16 characters)
 Host*                     Enter IP Address of Printer. (Up to 64 characters)
 Alias*                    Enter the name which is easy to identify. Do not enter more than
                           11 characters.
 Port*                     Enter Port Number of Printer.
 Packet size*              Referring to DICOM Conformance of Printer, if max PDU size is
                           described, enter that value. If not described, enter 99280.
 File packets grouping* Input “2”.
 Printer Model*            Enter the name of printer definition file (with exception of the
                           extension). (E.g. if the name of printer definition file is
                           “printer.pro”, enter as “printer”.
                           Printer definition file is needed to save in
                           “C\program andFiles\Voyager\Workstation\Data”.
                           *The detail of printer definition file is described in “5.6 Printer
                           Definition File”.
     “*” is mandatory item.
                                             5-37                                  M517-E610
= REF PC and ACQ PC Configuration =
REF:
   Host Type*:
   Item                       Configuration procedure
   Other                      Check [Other].
       “*” is mandatory item.
   Information:
   Item                         Configuration procedure
   AE Title*                    Enter AE Title of Image Server. (Up to 16 characters)
   Host*                        Enter IP Address of Image Server. (Up to 64 characters)
   Alias*                       Enter name which is easy to identify. Do not enter more than 11
                                characters.
   Port*                        Enter Port Number of Image Server.
   Packet size*                 Referring to DICOM Conformance of Image server, if max PDU
                                size is described, enter that value. If not described, enter
                                99280.
   File packets grouping*       Input “2”.
       “*” is mandatory item.
   Capabilities*:
   Item                         Configuration procedure
   Store*                       Check [Store].
   Storage Commit               If image server supports Storage Commitment and Users utilize
                                this function, check.
                                NOTE) Refer to 5.10 Configuration of Storage Commitment for
                                detailed configuration method.
   Query                        If image server supports query and acquisition features (query・
                                retrieve), check.
   Relational                   Check [Relational].
  M517-E610                                     5-38
                                                     =REF PC and ACQ PC Configuration=
4) To edit Hosts Configuration, pull down [AE Title] and choose [AE Title].
                                           5-39                                   M517-E610
= REF PC and ACQ PC Configuration =
Success
      Failure
                                             NOTE
     When changing packet size and file packet grouping value, communication speed
     may be improved (depends on the network environment).
     Packet size should be multiple number of 1460 and less than MAXPDU size (this is
     described in DICOM conformance statement). This is the fastest setting value in
     theory.
     Value of file packet grouping is 2.
     It is efficient to display the network tab of task manager and check when testing.
  M517-E610                                   5-40
                                                   =REF PC and ACQ PC Configuration=
REF:
ACQ:
                                         5-41                                   M517-E610
= REF PC and ACQ PC Configuration =
REF:
  M517-E610                                  5-42
                                                   =REF PC and ACQ PC Configuration=
                                        5-43                                   M517-E610
= REF PC and ACQ PC Configuration =
REF:
ACQ:
  M517-E610                                  5-44
                                                   =REF PC and ACQ PC Configuration=
                                         NOTE
 This item exists every account. Therefore, for normal user setting, configure after
 login by “OPERATOR”.
                                         5-45                                   M517-E610
= REF PC and ACQ PC Configuration =
REF:
    1) When click [Performing Physician Table], physicians will display. Each physician has
       each row to input first name and last name within section.
    2) To add a physician, click [Add] to show new row. And then input last name and first
       name.
    3) To delete the registered name, select desired physician from the table with a mouse
       and then click [Delete] button.
  M517-E610                                    5-46
                                                =REF PC and ACQ PC Configuration=
                                       5-47                                M517-E610
= REF PC and ACQ PC Configuration =
                                            NOTE
     Select the physician first and then set DUP. Set Acquisition default menu, Fluoro
     record and Acquisition for physicians individually.
  M517-E610                                  5-48
                                                   =REF PC and ACQ PC Configuration=
Fluoro Record:
Item                  Configuration procedure
Direct                When you check this item, the image will be saved only when
                      press [Direct Record] button.
Last N Seconds        When you check this item, newest fluoroscopy image will always
                      be recorded in buffer. When click [Save], images in buffer will be
                      saved.
Acquisition:
Item                  Configuration procedure
Hold Fluoro           When you check this item, hold the last image at the end of
                      fluoroscopy. When uncheck this item, the monitor reverts to
                      whatever was displayed before the fluoro acquisition.
Hold Recorded         When you check this item, hold the last image of the just-
Fluoro                recorded fluoro loop after the fluoscopy. When uncheck this item,
                      the just-recorded fluoro loop is played back.
E-Shutters:
Item                  Configuration procedure
Display E-Shutters    Set E-Shutters.
                      For item under development, do not put a check.
                      (There are some bugs. For example, displayed image may be
                      distorted if moving the collimator during fluoroscopy.)
Study Loops:
Item                  Configuration procedure
Auto Transfer         Activate [Auto Transfer] on [New Study] window.
                      When activate the [Auto Transfer], the study will be transferred to
                      the default server automatically.
                      Normally, checked this item.
IVR NEO and           Select a function and parameter associated with the function (if
Shortcut key          any) and click the [Assign] button. Repeat this process for each F
                      key.
                      To remove the function assigned a key, select the key and click
                      the [Delete] button.
                                         5-49                                  M517-E610
= REF PC and ACQ PC Configuration =
      The following functions are set to IVR NEO and ShortcutKey as default.
   IVR NEO                                     Shortcut Key
   Key         Description                     Key        Descripton
   B1          Select Fluoro Program           F1         Select Fluoro Program
   B2          Monitor Select                  F2         Select Rad Program
   B3          Select Rad Program              F3         Live
   B4          Live                            F4         Filter by C-arm
   B5          Filter by C-arm                 F5         Filter by C-arm
   B6          C-arm Reposition                F6         Start/Stop/Close Timer
   B7          Start/Stop/Close Timer          F7         Cine Area Zoom
   B8          Cine Area Zoom                  F8         Select Image Comment
   B9          Select Image Comment            F9         16-up Selector
   B10         16-up Selector                  F10        Save Fluoro
   B11         Save Fluoro                     F11        Save Still Image
   B12         Save Still Image                F12        Add Reference Image
   B13         Add Reference Image
   B14         Reference Mode
   B15         Single/Split Display
   B16         Left/Right Focus
   Coordination with
   Geometry
   Item                  Configuration procedure
   Coordination with     When checked, SIMAP Mask image and the Guide will hide when
   Geometry              changing geometry.
   Enhance Contour
   Item                  Configuration procedure
   Smoothing             The setting value will be used as default
                         value. Default value is 5.
   Thickness             Specify default value of contour thickness.
   Guide Edition:
   Item                  Configuration procedure
   Guide Dots Size       Specify dots size when placing the guide. Default value is 2.
   Line Type             Specify line style for guide line.
   Line Width            Specify line width for guide line. Default value is 5.
   Line Color            Specify line color for guide line.
   Default Line          Enable to set default line item when [Edit] window is displayed.
                         Default line is straight.
   Mask Region:
   Item                  Configuration procedure
   Initial Size (%)      Set the initial size when setting ROI. Initial size is specified by
                         percentage of the image. If the initial value is 30, the size of ROI
                         will be displayed as 30% in vertically and horizontally. The default
                         value is 30.
   Minimum Size (%)      Set minimum size of ROI. The default value is 10.
  M517-E610                                   5-50
                                                     =REF PC and ACQ PC Configuration=
Coordination with
Geometry:
Item                    Configuration procedure
Guide Display           If communication is disconnected, specify the time (sec) before
Timeout (sec)           invalid the guide automatically. If input “0”, it will not invalid
                        though the communication is disconnected. The default value is
                        30.
Oblique Angle           If Coordination with Geometry is checked and the C-arm is
Margin (degree)         changed more than input angle in an oblique direction, SIMAP,
                        Guide Line, interest region of StentView will hide. The default
                        value is 1.
Sagittal Angle          If Coordination with Geometry is checked and the C-arm is
Margin (degree)         changed more than input angle in a sagittal direction, SIMAP,
                        Guide Line, interest region of StentView will hide. The default
                        value is 1.
SID Margin (mm)         If Coordination with Geometry is checked and SID is changed
                        more than input value, SIMAP, Guide Line, interest region of
                        StentView will hide. The default value is 10.
Table Longitudinal      If Coordination with Geometry is checked and the table moves
Position Margin         more than input value in a longitudinal direction, SIMAP, Guide
(mm)                    Line, interest region of StentView will hide.
Table Transversal       If Coordination with Geometry is checked and the table moves
Position Margin         more than input value in a transversal direction, SIMAP, Guide
(mm)                    Line, interest region of StentView will hide.
Table Height Margin     If Coordination with Geometry is checked and the table height is
(mm)                    changed more than input value, SIMAP, Guide Line, interest
                        region of StentView will hide.
Ceiling Travel          If Coordination with Geometry is checked and the C-arm moves
Longitudinal Position   more than input value in a longitudinal direction, SIMAP, Guide
Margin (mm)             Line, interest region of StentView will hide.
Ceiling Travel          If Coordination with Geometry is checked and the C-arm moves
Transversal Position    more than input value in a transversal direction, SIMAP, Guide
Margin (mm)             Line, interest region of StentView will hide.
                                           5-51                                  M517-E610
= REF PC and ACQ PC Configuration =
    Check radiography or fluoroscopy, and click [Select]. Menu will be displayed, so select
    and click the item to default. (Selected item will be highlighted)
      Selecting Radiography
      REF:
      Selecting Fluoroscopy
      REF:
  M517-E610                                  5-52
                                                   =REF PC and ACQ PC Configuration=
(h) DICOM
REF:
ACQ:
    DICOM Option:
Item                      Configuration Procedure
DICOM Option              Check [Host Authentication].
                                         5-53                                 M517-E610
= REF PC and ACQ PC Configuration =
  M517-E610                                 5-54
                                                    =REF PC and ACQ PC Configuration=
REF:
Media:
Item                       Configuration procedure
Write Speed                Select the media write speed.
                           Normally, select the maximum speed.
Convert 512×512            Defines whether studies are converted to a lossless
                           compressed DICOM format (512x512 8-bit) before writing to
                           removable media.
                           If unchecked, the DICOM format (1k x 1k 12-bit) is written to
                           the media. Large uncompressed Cine (raw) studies may not
                           fit on a single CD so it is recommended to check this box if
                           using CD.
Automatically Write        When checked, writing on [CD] / [DVD] will start when study
                           is end.
Auto-accept Defaults       When checked, only the message to insert blank media is
                           displayed in a pop-up window during [CD] / [DVD] writing. All
                           other information is displayed in the status bar.
Image Processing           When checked, image processing including DSA subtraction
                           will be performed before writing to [CD] / [DVD].
Additional Files           Enable to add DICOM viewer to the media. All the files and
                           folders in a selected folder will be written in the media root.
                                         5-55                                    M517-E610
= REF PC and ACQ PC Configuration =
REF:
   Mandatory Items:
   Item                    Configuration procedure
   Patient Information     Define which items in the new study window are to be made
   Study Information       mandatory.
                           Set a check mark to indicate mandatory or uncheck the mark to
                           indicate optional.
   Accession Number:
   Item                    Configuration procedure
   Prefix                  (Available only if [Enable IHE Support] is not selected on the
                           [DICOM/IHE] tab.)
                           Used as the prefix for the auto-generated accession number
                           that will be used to identify each study loop. If the [Site] field is
                           used, the value entered there is copied into this field. (Up to 6
                           characters: alphabet, number, underbar “_”, and period “.”)
   Current                 Show and set the current auto-generated accession number.
                           If [Prefix] is used, prefix characters will be set before this auto-
                           generated [Current] value. Accession number, including prefix,
                           can be set up to 16 digits. (0 to 10,000,000)
                           Note) To avoid overlapping, do not reset the number to a lower
                           value that was used before.
   Increment               Set the increment that is added to [Current] value when
                           automatically generating the next accession number.
                           Set this value to zero if you do not want to automatically
                           generate the accession number. This can be used to set by
                           HIS/RIS system or manually. (0 to 100)
  M517-E610                                   5-56
                                                        =REF PC and ACQ PC Configuration=
Units:
Item                   Configuration procedure
Use Regional Setting s Set the patient [Height] and [Weight] measurement system
Units                  according to regional settings for New Study window.
                       (Example: Metric or U.S.)
                       If it is unchecked, user can select the desired unit.
Link Both Units        This is valid only when [Use Regional Settings Units] is not
                       selected. If selected, forcibly to use the same measurement
                       system for both patient height and weight. (Example: Metric or
                       U.S.)
Edit Patient
Information:
Item                       Configuration procedure
Allow user to modify       When checked, enable to modify study information without
patient information        input administrator’s password.
Local Worklist:
Item                       Configuration procedure
Local Worklist             If check [Local Worklist Enabled], Local Worklist Management
                           window will be displayed when starting new study. The study
                           selected from Local Worklist can be deleted as the following
                           instruction.
Pediatrics:
Item                       Configuration procedure
Show Pediatric             If check this box, warning is displayed at the start of new study
Warning                    if patient age is under the [Upper age limit].
                                            NOTE
 For this item setting, execute by account for System Administrator. Otherwise, after
 executing by account for Service, close the application, and then execute [Setup
 Menu]-[Copy Mandatory Information] on Maintenance Menu.
                                             5-57                                     M517-E610
= REF PC and ACQ PC Configuration =
REF:
ACQ:
   Display Configuration:
   Item                       Configuration procedure
   Default to Play All        Playback all loops one after the other instead of just playback
   Loops                      the selected loop continuously.
   (REF Only)                 Leave this unchecked normally.
   Display Image              Display the image-selector icons on the right edge of monitor.
   Selector Icons             Display in a double columns when click [Double Rows or
   (Both REF and ACQ)         Columns].
                              C-arm position is displayed on image selector when [Show C-
                              arm Position] is checked.
  M517-E610                                    5-58
                                                       =REF PC and ACQ PC Configuration=
                                            NOTE
  This item exists every account. Therefore, for normal user setting, configure after
  login by “OPERATOR”.
Monitor Power:
Item                        Configuration procedure
Monitor Power               When elapse without keyboard, mouse or IVR NEO activity,
(Both REF and ACQ)          monitor window switches to power-save mode. Set [After
                            (Minutes)] field at 1 to 60.
                            Windows screensaver must be invalid.
Disable Second              Not used. Normally unchecked.
Monitor
Close Study:
Item                        Configuration procedure
Set Default To              Close confirmation message window will appear when closing
(REF Only)                  the Study. Check any required items.
                                             5-59                                   M517-E610
= REF PC and ACQ PC Configuration =
REF:
   Print Options:
   Item                     Configuration procedure
   Print in Color (Non-     Check this item to print in color using non-DICOM printer such
   DICOM)                   as laser and ink jet.
   Keep the Print           Checked as a default.
   Manager open after
   printing
   Cache Integrity:
   Item                     Configuration procedure
   Verify cache integrity   Check this item upon maintenance and reboot DAR application.
   at next startup          After rebooting, the system will try to solve an error, such as
                            disable to delete the image in the image storage folder.
                            When DAR application is started, this item will be unchecked
                            automatically.
  M517-E610                                  5-60
                                                 =REF PC and ACQ PC Configuration=
                                       5-61                              M517-E610
= REF PC and ACQ PC Configuration =
      Click [Font and Colors] from [Configuration] dialog. [Font and Color] dialog is displayed.
      Configure only for REF PC.
REF:
   Colors Options:
   Item                     Configuration procedure
   Arrow                    Set the color of [Arrow] for [Annotation].
   Annotation Text          Set the color of [Text] for [Annotation].
   Line                     Set the color of [Line] for [Annotation].
   Curve                    Set the color of [Line] for [Annotation].
   Caliper                  Set the color of [Caliper] for [Annotation].
   Analysis Text            Set the color of [Text] for [QCA/LV].
   Study Text               Set the color of [Text] for [Study] of [Annotation].
   Font Options:
   Item                     Configuration procedure
   Annotation Text          Set such as the font and font size of [Annotation].
   Analysis Text            Set such as the font and font size of [QCA/LV].
   Study Text               Set such as the font and font size of [Study] for [Annotation].
   Other Settings:
   Item                     Configuration procedure
   Arrow Weight             Set the size of [Arrow] for [Annotation].
   Caliper Step             Set the size of [Caliper] for [Annotation].
  M517-E610                                   5-62
                                                     =REF PC and ACQ PC Configuration=
(2) Optional
(a) System
(b) Hosts
   Do not modify the setting of AE title [DAR9500_0x (Number 1 and up is set to “x”.
   Set when connect to DICOM host (MWM server, printer and image server).
(c) Storage
                                             5-63                               M517-E610
= REF PC and ACQ PC Configuration =
(e) Menu
(f) Database
(g) Physicians
  M517-E610                                     5-64
                                                    =REF PC and ACQ PC Configuration=
(h) DICOM
(k) Visualization
(l) Messages
Not Available.
                                           5-65                                 M517-E610
= REF PC and ACQ PC Configuration =
(m) General
(n) Logs
Not Available.
  M517-E610                              5-66
                                                  =REF PC and ACQ PC Configuration=
Several Options tabs have additional values related to the bi-plane option.
 Bi-plane options are configured on [System] tabs for two ACQ PC (ACQ PC (F) and
 ACQ PC (L)) as follows.
ACQ F:
ACQ L:
                                       5-67                                   M517-E610
= REF PC and ACQ PC Configuration =
   Link:
   Item                   Configuration procedure
   Bi-Plane               Checked.
   (Both ACQ F and
   ACQ L)
   A (Frontal)            Checked.
   (ACQ F only)
   B (Lateral)            Checked.
   (ACQ L only)
   Plane B Name           Input IP address of ACQ PC L.
   (ACQ F only)
      On the Review [Menus] tab, Bi-plane programs are configured in the [Acquisition
       Mode] option of the Camera section by selecting the [Both] item. Similarly, Frontal or
       Lateral programs can be created.
  M517-E610                                  5-68
                                                   =REF PC and ACQ PC Configuration=
                                         5-69                                M517-E610
= REF PC and ACQ PC Configuration =
     This tool assists service personnel or technical staff to provide DICOM printer definition
     file for DAR-9500f.
       Users of this tool are required to refer to DICOM Conformance statement or adequate
       documents of the target printer, and are supposed to know available printer capability
       of layout, film type and film size etc.
       Also this tool requires already existing printer definition files(.pro) as initial definition
       information.
                                               NOTE
     There is a definition file of the printer that has connected results in
     C:\Shimadzu\ConfigFiles\Printer on REF PC. (There might not be a definition file
     depending on the version of the system.)
     When a file of the printer that will try to be connected has already existed in the
     above-mentioned folder at the installation site, copy it into C:\Program
     Files\Voyager\Workstation\Data and use it as it is. If necessary, edit this file with this
     tool and use it.
     When the definition file has been created and used (operated) without trouble, it is
     not necessary to replace or edit it especially.
  M517-E610                                     5-70
                                                  =REF PC and ACQ PC Configuration=
(1) Startup
 When logging in as account “serviceapp”, an icon of Maintenance Menu.htm appears
 on the desktop. Double-click it to display Maintenance Menu window. Select “Setup
 Menu” and click “DICOM printer configuration” to execute this tool.
(2) Open
 Select Open of File menu and open printer definition file with the extension of “.pro”.
File Open dialog as below is opened. Select desired printer definition file, and click
Open button.
                                        5-71                                   M517-E610
= REF PC and ACQ PC Configuration =
      The content of printer definition file is loaded and displayed on screen as follows. With
      DICOM conformance statement or adequate document, edit each items of target
      printer definition. Configuration guideline for each item is written in the later section.
      Note) In case Open of File menu is selected again while printer definition file is open,
      warning message is displayed as below. Select Yes or No, and continue to the next
      operation.
  M517-E610                                   5-72
                                                    =REF PC and ACQ PC Configuration=
(3) Save
When the edition of the printer definition is completed or to save current contents, save
current printer definition by selecting Save of File menu. The currently opened printer
definition file will be overwritten. Note that previous contents of the printer definition file
are lost.
Note 1) In case currently opened file is read-only file, this menu is not available.
(4) Save As
To save current printer definition with other name than previous one, select Save As of
File menu.
                                         5-73                                      M517-E610
= REF PC and ACQ PC Configuration =
The dialog of Save As is opened as below. Input new file name and click Save.
      The currently opened printer definition file will be saved with new name and previous
      file is kept.
      Note 1) If the read-only file with the same file name already exists, the error message
      will be appeared and the file cannot be saved.
  M517-E610                                  5-74
                                                   =REF PC and ACQ PC Configuration=
Note) Maximum number of characters is up to 16. The available character set is ASCII.
In case more than 16 characters are input, or the unavailable ones are typed, error
message is displayed.
Note) Maximum number of characters is up to 16. The available character set is ASCII.
In case more than 16 characters are input, or the unavailable ones are typed, error
message is displayed.
Note1) The possible input number for the minimum and maximum number of copies is
from 1 to 99. And maximum value must be larger than minimum value.
                                         5-75                                   M517-E610
= REF PC and ACQ PC Configuration =
     (6) Film orientation
      Check available printer orientation of film of the target printer. At least one item should
      be selected. This enables user to select film orientation from checked ones when
      printing.
     (7) Magnification
      Check available magnification of the target printer. At least one item should be
      selected. This enables user to select magnification method from checked ones when
      printing.
  M517-E610                                    5-76
                                                 =REF PC and ACQ PC Configuration=
Note) The possible input number for the minimum and maximum columns and rows
values are from 1 to 4.
                                       5-77                                  M517-E610
= REF PC and ACQ PC Configuration =
Perform as follows to print Pattern SMPTE for print density adjustment etc.
Click [File] from the side menu to open Studies Management Window.
     Input “SMPTE” to [Patient ID], “All Time” to [Date], and then click [Search].
     SMPTE will display on the list so select and double-click it.
  M517-E610                                   5-78
                                                   =REF PC and ACQ PC Configuration=
Press [Add] button and [Print] button on the side menu. As print configuration window is
displayed, press [Print] button to print.
For print density adjustment, print without redundant gradation processing or LUT,
otherwise it is not adjusted correctly. Confirm that LUT is Linear on print configuration
window and that Auto W.Level is off and Brightness and Contrast are both set to 0 on
the side menu.
Additionally be careful because the film image might be seen different from the image
on screen if Monitor is not correctly adjusted. For Monitor adjustment, refer to Image
Quality Adjustment Manual.
                                         5-79                                   M517-E610
= REF PC and ACQ PC Configuration =
  M517-E610                                    5-80
                                                 =REF PC and ACQ PC Configuration=
                                        5-81                              M517-E610
= REF PC and ACQ PC Configuration =
  M517-E610                                 5-82
                                                       =REF PC and ACQ PC Configuration=
 (1) Click [Option] button on the Side Menu. [Configuration] dialog is displayed. (Refer to 5.5.1)
 (2) Click [Hosts] on [Configuration] dialog. [Hosts Configuration] dialog is displayed.
 (3) Choose the AE Title of configured PC.(The AE title of REF PC is selected when REF PC
    is configured, and when ACQ PC is configured, the AE title of ACQ PC is selected.)
 (4) Configure the port. Default port number is 104.If Storage commitment is enabled; Port
    number of REF PC and that of ACQ PC must be different. Setting each port numbers as
    follows.
              REF PC                             :104
              ACQ PC (ACQ-F PC)                  :105
              (For Bi-plane, additionally, the following configuration is necessary.)
              ACQ-L PC                           :106
                                            5-83                                      M517-E610
= REF PC and ACQ PC Configuration =
1). Click
2) Click
        3)
        Configure
        as picture
        4) Configure
        Server IP
        5) 
        Configure                                                                        7) Configure 105
        WAN IP of
        router
        6) Configure as
        picture                                                             8) Click
     (5) Do the same procedure as (4), and put 104,106 for procedure 7) in the picture above. So
        create 3 configurations.
  M517-E610                                      5-84
                                                       =REF PC and ACQ PC Configuration=
                                           5-85                                 M517-E610
= REF PC and ACQ PC Configuration =
                                               NOTE
     When setting Synergy, Synergy will not work after completing the setting on REF
     PC.
(2) Click [Start] on the bottom left of the window and click [All Programs]-[Synergy].
  M517-E610                                     5-86
                                                    =REF PC and ACQ PC Configuration=
                                         5-87                              M517-E610
= REF PC and ACQ PC Configuration =
     (5) Set [Name of the Server].
  M517-E610                                5-88
                                                      =REF PC and ACQ PC Configuration=
                                          NOTE
When setting Synergy, Synergy will not work after completing the setting on REF
PC.
(2) Click [Start] on the bottom left of the window and click [All Programs]-[Synergy].
                                           5-89                                     M517-E610
= REF PC and ACQ PC Configuration =
     (3) Click [Edit]-[Settings].
  M517-E610                                    5-90
                                                =REF PC and ACQ PC Configuration=
                                         5-91                          M517-E610
= REF PC and ACQ PC Configuration =
5.11.3 REF PC
(2) Click [Start] on the bottom left of the window and click [All Programs]-[Synergy].
  M517-E610                                     5-92
                                                     =REF PC and ACQ PC Configuration=
                                          5-93                              M517-E610
= REF PC and ACQ PC Configuration =
     (5) Click [Configure Server…].
(6) The following window is displayed. Double-click an icon of the setting computer name.
  M517-E610                                    5-94
                                                     =REF PC and ACQ PC Configuration=
 (8) [Server Configure] is displayed again. Double-click “DAR9500ACQ-L-XX” icon (for Bi-
   plane).
                                           5-95                                  M517-E610
= REF PC and ACQ PC Configuration =
     (9) Screen Setting] window is displayed. Input [Screen Name] (for Bi-plane).
Click [Ok].
  M517-E610                                     5-96
                                                     =REF PC and ACQ PC Configuration=
Click [OK].
                                          5-97                              M517-E610
= REF PC and ACQ PC Configuration =
     (13) [Synergy] is displayed.
    Click [Apply].
    And check that the mouse pointer is moved to ACQ PC F and ACQ PC L.
  M517-E610                              5-98
                                                   =REF PC and ACQ PC Configuration=
 If not Auto Transfer the reference registered image to the image server as a still image,
 set the image as the following procedure.
The following maintenance menu window is displayed. Click [2. Setup Menu].
                                           5-99                                M517-E610
= REF PC and ACQ PC Configuration =
     The following is displayed. Click [OFF] for "Reference image transfer ON/OFF". (Click
     [ON] for sending an image)
  M517-E610                                 5-100
                                                   =REF PC and ACQ PC Configuration=
                                       5-101                              M517-E610
= REF PC and ACQ PC Configuration =
        (b) PING is failed from our system such as DAR-9500f to other company’s system
            Check that Outbound setting of router is correct.
            Check that the other system is running and LAN is wired.
            IP address may competing or LAN cables may wired in loop when PING is
            successful or failed.
  M517-E610                                    5-102
                                                 =REF PC and ACQ PC Configuration=
                                                             ・ADICOM association
                                                             accept message will be
                                                             returned.
                                                             ・Acceptance is on
                                                             Pre.context
                                                             (Reject is NG).
               Response (RSP) of
               C-STORE.
                                      5-103                                 M517-E610
= REF PC and ACQ PC Configuration =
                          Send RQ of Release.
                          Receive RP to complete
                          the communication.
  2There is a SPY Tool as a confirmation tool on DAR-9500f. Input “zerg” on the window
  of DAR-9500f (ACQ F). And then right-click the mouse holding down ctrl, alt, and z key
  to startup.
  M517-E610                                  5-104
                                               =REF PC and ACQ PC Configuration=
                                     5-105                                M517-E610
= REF PC and ACQ PC Configuration =
       REF PC:
       Voyager-Workstation-System Options-SendAnnotation→(true/false)
       To transfer the saved image (photo file image) to the server automatically or not.
       Voyager-Workstation-System Options-PrinterSeparator→(true/false)
       To section the images frame by frame when printing.
       As the DSA tool could not display on the Review Station when shipping by mistake.
       Follow the registry key below to take action.
       HKEY_LOCAL_MACHINE\SOFTWARE\Voyager\Workstation\System
       Options\DisableableFeatures\DisableDSATool→true(disable)/false(enable)
3 Specify the file name and execute “regedit” to display the registry editor.
  M517-E610                                    5-106
                                                     =Connection and Configuration of Router=
6. Router Configuration
 6.1 Login
      1) On REF PC, login with “serviceapp” and select [Start]-[Run].
2) Enter “http:// 192.168.100.29” with English one-byte characters, and click [OK].
     3) The following dialog is displayed. Enter “manager” to user name and “friend” to
        password. And click [OK].
                                               6-1                                     M517-E610
=Connection and Configuration of Router=
3) Click [Save] button on the upper-left corner of the window to store the configuration.
M517-E610                                      6-2
                                                    =Connection and Configuration of Router=
        NOTE : If IP address of DAR-9500f is asked when connecting with image server or etc,
                 reply WAN IP address of router.
4) Click [Save] button on the upper-left corner of the window to store the configuration.
                                              6-3                                     M517-E610
=Connection and Configuration of Router=
          NOTE : If IP address of REF PC is different from the standard IP address, refer to « 6.7
          Individual Setting ».
      3) After the modification, click [Save] button on the upper-left corner of the window to
         store the configuration.
M517-E610                                      6-4
                                                   =Connection and Configuration of Router=
  2) Enter each parameter as follows and click [Add] for each setting.
     Setting Outbound
     <ACQ PC (Frontal)>
     Direction        : Outbound
     Action           : Permit
     Priority         : (Automatic setting, so do not change.)
     Source IP
      Type            : IP Address
      IP Address      : 192.168.100.22 (IP address of Frontal acquisition PC)
     Destination IP   : All
     Source Port      : All
     Destination Port : All
     Protocol         : All
      <REF PC>
      Direction          : Outbound
      Action             : Permit
      Priority           : (Automatic setting, so do not change.)
      Source IP
       Type              : IP Address
       IP Address        : 192.168.100.23 (IP address of REF PC)
      Destination IP     : All
      Source Port        : All
      Destination Port   : All
      Protocol           : All
                                             6-5                                 M517-E610
=Connection and Configuration of Router=
          Setting Inbound
          Direction          : Inbound
          Action             : Permit
          Priority           : (Automatic setting, so do not change.)
          Source IP
           Type              : IP Address
           IP Address        : (IP address of image server)
          Destination IP     : (Router WAN IP address)
          Source Port        : All
          Destination Port   : 104
          Protocol           : TCP
          NOTE : If IP address of REF PC and ACQ PC is different from standard, refer to »6.7
          Individual Setting ».
3) Click [Save] button on the upper-left corner of the window to store the configuration.
M517-E610                                       6-6
                                                =Connection and Configuration of Router=
Correct:
1) From [Access Control List] at the bottom of the window, select and click the setting to
   correct.
3) Click [Save] button on the upper-left corner of the window to store the configuration.
Delete:
1) From [Access Control List] at the bottom of the window, select and click the setting to
    remove.
2) Click [Save] button on the upper-left corner of the window to store the configuration.
                                          6-7                                   M517-E610
=Connection and Configuration of Router=
* This procedure is not necessary at installation. Do this only in the following cases:
    (a) Open the configuration window and click [System Management] - [System
        Configuration] - [Default Settings]. Back up is not necessary.
(b) Power off the switch of Router, and wait more than 5 seconds.
* Both of “Power switch” and “RESET Switch” are on the back of the Router.
    (c) Holding “RESET Switch”, turn on “Power switch”. Keep holding “RESET Switch”
        until SYSTEM LED blink on and off 3times shortly.
M517-E610                                        6-8
                                                     =Connection and Configuration of Router=
                                            NOTE
 If IP address of router is changed, it is impossible to access to the router. And menu cannot
 be selected and settings cannot be referred and changed. To recover from this conditions, set
 the same subnet mask as router, and set IP address which has same parameter in 1st, 2nd and
 3rd field as router.
 For example
 IP address:192.168.101.29
 Sub-net Mask:255.255.255.0
     1) Configure LAN IP address, Subnet Mask. Click [LAN]-[IP] from Menu. Configure IP
        address and Subnet Mask described in Configuration policy above.
                                               6-9                                    M517-E610
=Connection and Configuration of Router=
      3) Click [LAN]-[IP] from Menu. Confirm IP address and Subnet Mask configured above is
         displayed.
4) Click [Save] button on the upper-left corner of the window to store the configuration.
M517-E610                                     6-10
                                                 =Connection and Configuration of Router=
      Information Router
         WAN IP Address
           Refer to page
        Firewall Setting
           NAT setting
           Refer to page
        LAN IP Address
              IP Address         : 192.168.100.28 (LAN IP address of router)
              NetMask            : 255.255.255.0
              Directed broadcast : disable
      Image Router
        WAN IP Address
          Refer to page 6-3.
        Firewall Setting
           NAT setting
             Refer to page 6-4.
        LAN IP Address
          Refer to page 6-9.
                                            6-11                                  M517-E610
=Connection and Configuration of Router=
M517-E610                                   6-12
                                            =Configuration of IP Address of I-SYS and CP2=
                                             NOTE
  Follow this chapter only if hospital has two or more systems.
 Configure IP address of I-SYS and CP2, Subnet mask and Default gateway as the
 following table.
 <Single-plane/Bi-plane Frontal>
                                st                       nd                rd
                               1 System                2 System           3 System
      I-SYS
      IP address               192.168.100.26          192.168.100.46     192.168.100.66
      Subnet mask              255.255.255.0           255.255.255.0      255.255.255.0
      Default gateway          192.168.100.29          192.168.100.49     192.168.100.69
      CP2
      IP address               192.168.100.21          192.168.100.41     192.168.100.61
 <Bi-plane Lateral>
                                st                       nd                rd
                               1 System                2 System           3 System
      I-SYS (L)
      IP address               192.168.100.27          192.168.100.47     192.168.100.67
      Subnet mask              255.255.255.0           255.255.255.0      255.255.255.0
      Default gateway          192.168.100.29          192.168.100.49     192.168.100.69
      CP2 (L)
      IP address               192.168.100.24          192.168.100.44     192.168.100.64
                                                7-1                             M517-E610
=Configuration of IP Address of I-SYS and CP2=
                                             NOTE
  In case of single-plane system, REF image will not display during I-SYS restarting. Wait
  until I-SYS is restarted. (Approximately wait for 30 seconds)
               ==========================================
                 I - S Y S    S e r v i c e      M e n u
                   FPGA Rev.0.0.12 (2011-04-11)
               ------------------------------------------
                 1- Update FPGA/Firmware/Tables.
                 2- Board Check Program.
                 3- Debug Tools for R&D
                 4- Ethernet / RTC setting
               ------------------------------------------
                 Menu : 4
             ==========================================
               I - S Y S Ethernet/RTC T o o l s
             ------------------------------------------
               1- Ethernet Setting read
               2- MAC-Address write
               3- IP-Address write
               4- Ethernet Test Firmware Send
               5- DATE/TIME read
               6- DATE/TIME setting
               7- RTC IC Dump
               0- Quit
             ------------------------------------------
               Menu: 3
M517-E610                                     7-2
                                           =Configuration of IP Address of I-SYS and CP2=
(6) Enter each value in order of IP Address, Subnet Mask and Default Gateway. (Impossible to
    omit) When press [Enter] without entering values, they will be cancelled.
   After entering for all items, setting values will be written on FLASH memory and return to
   the original menu.
        ==========================================
          I - S Y S Ethernet/RTC T o o l s
        ------------------------------------------
          1- Ethernet Setting read
          2- MAC-Address write
          3- IP-Address write
          4- Ethernet Test Firmware Send
          5- DATE/TIME read
          6- DATE/TIME setting
          7- RTC IC Dump
          0- Quit
        ------------------------------------------
          Menu: 1
IP Address :192.168.100.046
                                              7-3                                 M517-E610
=Configuration of IP Address of I-SYS and CP2=
    (2) Connect ”SERIAL 1” on CP2 rear and COM3 port on ACQ-PC with the straight
       cable supplied with CP2.
       *COM3 is a port that the dosimeter (optional) is connected.
    (5) From [Configure] – [Communication Port] menu of [Host Download] tool, select
       COM3 and click [Connect].
M517-E610                               7-4
                                   =Configuration of IP Address of I-SYS and CP2=
(6) Check that [Host Download] tool is in connection request mode (”--Working --”
   is blinking) and turn ON the power of CP2.
(7) When “Connection Request Timeout” dialogue is displayed, check the following
   things. Click [Retry] to check that [Host Download] tool is in connection request
   mode and then turn ON the power of CP2.
Click [OK].
                                      7-5                                M517-E610
=Configuration of IP Address of I-SYS and CP2=
M517-E610                                 7-6
                                        =Configuration of IP Address of I-SYS and CP2=
Select [Network].
                                           7-7                             M517-E610
=Configuration of IP Address of I-SYS and CP2=
                             st                    nd                rd
                            1 System              2 System          3 System
      I-SYS (F)
      IP :                  192.168.100.26        192.168.100.46    192.168.100.66
      Port                  7011                  7011              7011
      I-SYS (L)
      IP :                  192.168.100.27        192.168.100.47    192.168.100.67
      Port                  7011                  7011              7011
Click [OK].
M517-E610                                 7-8
                                                              =DVI MAIN Connection Configuration=
                                                   8-1                                      M517-E610
=DVI MAIN Connection Configuration=
M517-E610                                      8-2
                                                          =DVI MAIN Connection Configuration=
       In case of DVI-MAIN2
    <Single-Plane System Standard Settings>
  Output Channel
  CH-5 (VIDEO): VGA output
  CH-6 (VIDEO-2): NTSC/PAL output
  (Reference)
  The following window will be displayed during MAP fluoroscopy.
  For Single-plane,
  REF Monitor in Examination room: LIVE image
                                                8-3                               M517-E610
=DVI MAIN Connection Configuration=
(Note) For Bi-plane system, CH1 and CH2 of Output Channel will be different from Single-plane.
M517-E610                                      8-4
                                                        =DVI MAIN Connection Configuration=
     In case of DVI-MAIN3
 <Single-Plane System Standard Settings>
NOTE
Output Channel
CH-5 (VIDEO): VGA output
CH-6 (VIDEO-2): NTSC/PAL output
(Reference)
The following window will be displayed during MAP fluoroscopy.
For Single-plane,
REF Monitor in Examination room: LIVE image
                                              8-5                               M517-E610
=DVI MAIN Connection Configuration=
(Note) For Bi-plane system, CH1 and CH2 of Output Channel will be different from Single-plane.
M517-E610                                      8-6
                                                      =DVI MAIN Connection Configuration=
                                            8-7                                    M517-E610
=DVI MAIN Connection Configuration=
M517-E610                                       8-8
             Diagram of Channel
               Showing Output Channel, Input Channel and the connector name that is printed on the board.
               Confirm the connected Input Channel and displayed monitor channel, and then set the channel.
                                                                                                                                                A:Control Room
                                                                                      Input Channel                            Output Channel   B:Examination Room
                                                                                                                                                                       (REF1
                                                                     ACQ                           In-1(ACQ-1)
                                                                                                                    CH-1[REF-1]     ①                          ①       MONITOR)
                                                                      PC                      1
                                                                                                        J10                                                                       REF1 MONITOR
                                                                                                                         J20
                                                                                                                         J21
                                                                                                                                    ②
                                                                   OPTION                     2    In-2(ACQ-2)
                                                                     PC
                                                                                                        J11
                                                                                                                     CH-2[ACQ]      ③                          ②       (ACQ
                                                                                                                                                                       MONITOR)
                                                                   OPTION                           In-3(AUX)
                                                                     PC
                                                                                              3                          J22                                                      ACQ MONITOR
                                                                                                                         J23
                                      DVI SUB ASSY
                                                                                                        J12
                                                                                                                                    ④
             REF                                                                                    In-4(REF-1)
             PC
                                     (REF1 IN)                 1                              4
8-9
                                                                                                                                                                       (REF2                     OPTION
                                                                                                        J13
                                                                                                                    CH-3[REF-2]     ⑤                          ③       MONITOR)                   LCD
                                                                                                                         J24
            OPTION                                                                                  In-5(REF-2)          J25                                                                     OPTION
              PC                     (REF2 IN)                 2                              5                                     ⑥                                                             LCD
                                                                                                        J14
                                                                       A                                                                                 POWER CABLE
                          (VIDEO          (VIDEO                   (SUB POWER)
                            OUT 2)          OUT 1)
                                  NTSC/PAL        NTSC/PAL
=DVI MAIN Connection Configuration=
NOTE
The following error message will appear when communication error occurs.
Confirm that I-SYS and DVI-MAIN are connected with a serial cable.
M517-E610                                           8-10
                                                             =DVI MAIN Connection Configuration =
                                                 8-11                                          M517-E610
=DVI MAIN Connection Configuration=
M517-E610                                           8-12
                                                       =DVI MAIN Connection Configuration =
Set “MAP/LIVE” and “Last Image Hold” as follows and click [Apply].
       MAP/LIVE:             Unchecked
       Last Image Hold:      Checked
Click [OK].
                                             8-13                               M517-E610
=DVI MAIN Connection Configuration=
Click [Exit].
             For Single-plane:     Select [Yes] and restart I-SYS. Now the setting of I-SYS BOARD is
                                   completed.
             For Bi-plane:         Select [No] and go on to step (7).
M517-E610                                         8-14
                                                    =DVI MAIN Connection Configuration =
                                             8-15                            M517-E610
=DVI MAIN Connection Configuration=
Set “MAP/LIVE” and “Last Image Hold” as follows and click [Apply].
             MAP/LIVE:             Unchecked
             Last Image Hold:      Checked
Select [OK].
M517-E610                                          8-16
                                                        =DVI MAIN Connection Configuration =
Click [Exit].
Click [No].
                                             8-17                                M517-E610
=DVI MAIN Connection Configuration=
      (12) Press I-SYS BOARD F/L Push Switch and restart both I-SYS BOARD Frontal and Lateral at
          the same time.
Push Switch
M517-E610                                     8-18
                                                =Installation of FPD Cooling System (HEC002-A5B)=
Level meter
9.2 Specification
                                      Table 1. Specification table
        Device type name                     HEC002-A5B
        Temperature adjustment range         From 10 to 60 degrees
        Temperature stability                From ±0.01 to 0.03 degrees
        Reservoir tank capacity              1.2 liters
        (capacity of the circulation         (5.2 liters including the hose)
        water)
        Circulating water                    Antifreeze solution
                                             (Principal component: propylene glycol)
        Flow rate of circulating solution    1.2 liter/minute
        Heat removal ability                 230W
        Input voltage                        AC100V
        Maximum current consumption          8A
        Weight                               17.5kg (18.7kg after filling solution)
                                                    9-1                                       M517-E610
=Installation of FPD Cooling System (HEC002-A5B)=
9.3 Installation
1. Installation environment
    It is necessary to follow the installation environment below to keep the temperature of FPD at a
    prescribed FPD to prevent dew condensation.
       Examination room: Temperature from 10 to 35 degrees, humidity from 15 to 75%
       Equipment room: Temperature from 10 to 35 degrees, humidity from 15 to 75%
    Do not install the system where the temperature and humidity is higher and with corrosive substances.
    An air Cooling System is installed in this device. Let the air in from the left side and out from the right
    side. For sufficient ventilation, more than 20 cm of space is required for both side of the system, or the
    cooling capacity may degrading and it may damage the compressor in some cases.
    Avoid installation of this equipment in dusty area and make a schedule for periodic cleanup. (Refer to
    section “9.5 Maintenance for schedule)
2. Connection of hose
    Two 35m hoses are equipped. Each hose has a thick part and a thin part. Thick hose should be
    connected with the Cooling System and thin hose should be connected with FPD. It is recommended
    to cut extra part of hose and make it short as possible to increase the cooling capacity. Please be
    careful not to break the hose. The minimum bend radius for a thick hose is 95mm and thin hose is
    60mm.
    This Cooling System has a mechanism that it keeps the preset temperature even if the ambient
    temperature falls below the setting temperature. So, there is no problem when the equipment room
    temperature can be kept in the above-mentioned value. However, if there is a possibility of an
    equipment room temperature becomes below that, it becomes easier to keep the preset temperature by
    wrapping the heat insulator around hoses in the equipment room.
3. Wiring
    Please refer to section “9.6”. Connect the power supply code with the outlet of the following voltages
    in the ACQ3/ACQ3L Cabinet.
       50/60Hz: 100V ±10%
    For connection of the signal cable from the flow switch, refer to section “9.6 Connecting chart”.
M517-E610                                             9-2
                                                 =Installation of FPD Cooling System (HEC002-A5B)=
                                                      9-3                                         M517-E610
=Installation of FPD Cooling System (HEC002-A5B)=
5. Operation method
1)   Switch and display
     The following pictures show the external figure of Cooling System.
                                  External Figure of Cooling System
Display panel
                                                                             Current water
                                                                              temperature
                                                                             Setting water
                                                                              temperature
Mode
     When turn on the switch, software information is displayed for a second. And then the current and
     setting temperature, and mode are displayed. The mode is “Normal” for basic usage.
M517-E610                                           9-4
                                                =Installation of FPD Cooling System (HEC002-A5B)=
                                                    9-5                                  M517-E610
=Installation of FPD Cooling System (HEC002-A5B)=
6. Trial operation
    Keep operating the system for 10-30 minutes to confirm whether an actual measurement value of
    water temperature becomes 30 degrees (the time which reaches to 30 degrees depends on the starting
    temperature of operation). Please check that there is no solution leakage around FPD, hose joint, and
    back of the Cooling System after it starts operating. Also, confirm that water temperature display
    becomes 30 degrees. Water temperature had been set to 30 degrees at the factory setting.
M517-E610                                          9-6
                                                 =Installation of FPD Cooling System (HEC002-A5B)=
NOTE
1. Confirm that the temperature display in front of the Cooling System should be 30±1 degrees.
    Customer has to contact with our service department when it is not so.
2. Confirm that antifreeze solution is filled enough in the tank in the Cooling System. Please replenish
    the specified distilled water about once a month. (Part Name: DISTILLED WATER,20L
    P/N:017-40513-05). There is the low and high marker for guiding the distilled water in the tank.
    The appropriate solution level is the level between those two markers. You cannot operate the
    Cooling System if level is below the low marker.
3. The temperature and humidity of the examination room and the equipment room must be in the
    standard value of the installation environment condition. Adjust the set temperature of
    air-conditioning because condensation might occur if it is not so.
                                                     9-7                                       M517-E610
=Installation of FPD Cooling System (HEC002-A5B)=
9.5 Maintenance
WARNING
Unplug the power supply cord when removing the panel and cover to avoid getting an electric shock.
M517-E610                                            9-8
                                                   =Installation of FPD Cooling System (HEC002-A5B)=
              (1)
                                 Water flow
FPD
Flow switch
Filter
                       Cooling
                       System
                    (HEC002)
                                                   (3)
                                                                              ACQ3 Cabinet
J8
I-SYS
AC100V
                                                                     100V
                                                 (2)
             <SBi-Plane/Lateral>
                Part Number                     Part name                       Remark
            1     502-26532                   Hose Assy,RTE-7                     35m
            2   565-11475-02            CORD SET, WATER CHILLER L                 10m
            3   565-11390-02                Cooler Warning Cable                  10m
NOTE
   On Bi-plane system, be sure to change the power cable for Frontal FPD cooling unit from ACQ3
   cabinet to ACQ3L cabinet because of electricity capacity.
Connect the flow switch of Lateral FPD cooling unit to I-SYS L of ACQ3 Cabinet.
                                                         9-9                                 M517-E610
    =Installation of FPD Cooling System (HEC002-A5B)=
                                                                                         270
                                      436
                                                                                         210
                                                                 Filter
                                                                 フィルタ
          393
                                                                   (Wall)                               (Wall)
                                                                          200                         200
Circulation solution OUT                                 Drain
                                                       ドレイン
                             Circulation solution IN                Installed distance         Installed distance
Unit: mm
    M517-E610                                               9-10
                                                                          =Adjust of Virtual Collimation=
CAUTION
    Please follow “M526-E313 F-50 Installation Manual” to adjust virtual collimation after adjusting
    collimator.
C-Shutter H-Shutter
V-Shutter
                                                    10-1                                      M517-E610
=Adjust of Virtual Collimator=
10.3 Adjustment
    1.   Start up the system and login to REF PC and ACQ PC with “serviceapp” to start running the
         application.
    2.   Click [Option] on side menu on REF PC to display [Configuration] window.
    3.   Click [Exposure Parameters] tab.
    4.   To adjust Single-plane/Bi-plane Frontal, set FOV to [Frontal]. And set FOV to [Lateral] for
         adjusting Bi-plane Lateral.
M517-E610                                          10-2
                                                                      =Adjust of Virtual Collimation=
  Reference value for the configuration is listed in the following Table 1. Adjust based on the
  reference value.
                                   H-Shutter                                   V-Shutter
   FOV
                         Open                   Close                  Open                 Close
  12 (Large)               -5                    100                      0                  100
10 (Medium)                -8                    100                     -2                  100
  8 (Small)                -9                    100                     -5                  100
 6 (Smallest)             -12                    100                     -8                  100
4.5 (FOV 5th)             -14                    100                    -15                  100
                                      (a) 12 inch FPD Frontal
                                   H-Shutter                                   V-Shutter
   FOV
                         Open                   Close                  Open                 Close
  12 (Large)               -5                    100                      0                  100
10 (Medium)                -5                    100                     -2                  100
  8 (Small)                -8                    100                     -5                  100
 6 (Smallest)             -12                    100                     -5                  100
4.5 (FOV 5th)             -12                    100                    -15                  100
                                      (b) 12 inch FPD Lateral
                                                10-3                                       M517-E610
=Adjust of Virtual Collimator=
                                        H-Shutter                                    V-Shutter
        FOV
                              Open                    Close                  Open                 Close
      8 (Large)                -15                     100                    -15                  100
     7 (Medium)                -10                     100                     -5                  100
      6 (Small)                -15                     100                     -5                  100
    4.5 (Smallest)             -15                     100                     -5                  100
                                            (c) 8 inch FPD Frontal
                                        H-Shutter                                    V-Shutter
        FOV
                              Open                    Close                  Open                 Close
      8 (Large)                -15                     100                    -10                  100
     7 (Medium)                -15                     100                    -10                  100
      6 (Small)                -15                     100                    -15                  100
    4.5 (Smallest)             -15                     100                    -15                  100
                                            (d) 8 inch FPD Lateral
    1. Set 12 inch ([Large]) field of view size first. Confirm that field of view is set to 12 inch on the
         system display and etc. Confirm that FOV on the setting window for REF PC is set to [Large] as
         well.
M517-E610                                            10-4
                                                                      =Adjust of Virtual Collimation=
2. Adjust V-Shutter first. Close the V-Shutter halfway and perform fluoroscopy. And then close the
    C-Shutter and the virtual collimation for V-Shutter is displayed. Set value of [Open] to align
    fluoroscopy image and the position of virtual collimation for V-Shutter. Click [Apply] and that the
    setting value of virtual collimation is reflected.
NOTE
Disable to enter “-“ with numerical keypad. Use “-“ under the Function keys.
                                                  10-5                                      M517-E610
=Adjust of Virtual Collimator=
    4. Adjust FOV [Medium] (12 inch, 10/8 inch, 7 inch) next. Confirm that field of view size is 10 inch
         and FOV is [Medium].
6. Adjust other field of view as well. For Bi-plane system, set each FOV of Lateral.
M517-E610                                           10-6
                                                                     =Adjust of Virtual Collimation=
 Reference value for the configuration is listed in the following Table 2. Adjust based on the
 reference value.
                                                    C-Shutter
                       FOV
                                            Open              Close
                      12 (Large)             -10               75
                    10 (Medium)              -10               70
                      8 (Small)              -10               62
                     6 (Smallest)            -17               50
                    4.5 (FOV 5th)            -17               32
                                     (a) 12 inch FPD Frontal
                                                    C-Shutter
                       FOV
                                            Open              Close
                      12 (Large)              0                75
                    10 (Medium)               0                70
                      8 (Small)               0                62
                     6 (Smallest)             0                50
                    4.5 (FOV 5th)             0                32
                                     (b) 12 inch FPD Lateral
                                              10-7                                        M517-E610
=Adjust of Virtual Collimator=
                                                           C-Shutter
                            FOV
                                                   Open              Close
                         8 (Large)                  -10               55
                        7 (Medium)                  -10               55
                         6 (Small)                  -10               50
                       4.5 (Smallest)               -10               30
                         8 (Large)                  -10               55
                                            (c) 8 inch FPD Frontal
                                                           C-Shutter
                            FOV
                                                   Open              Close
                         8 (Large)                  -10               55
                        7 (Medium)                  -10               50
                         6 (Small)                  -10               50
                       4.5 (Smallest)               -10               30
                         8 (Large)                  -10               55
                                            (d) 8 inch FPD Lateral
    1. Set 12 inch ([Large]) field of view size first. Confirm that field of view is set to 12 inch on the
         system display and etc. Confirm that FOV on the setting window for REF PC is set to [Large] as
         well.
M517-E610                                            10-8
                                                                       =Adjust of Virtual Collimation=
2. Adjust [Close] value. Fully-close the C-shtter and perform fluoroscopy, and close H-shutter.
    Virtual collimation of C-shutter will display. Set [Close] value to align fluoroscopy image and
    virtual collimation of C-shutter. Enter the value and click [Apply], the setting value of virtual
    collimation will be reflected.
3. Adjust value of [Open]. Close the C-shutter half-way and perform fluoroscopy. Close the
    H-shutter, and the virtual collimation of C-shutter will display. Set [Open] value to align the
    fluoroscopy image and virtual collimation of C-shutter.
4. Adjust other field of view. For Bi-plane system, set each FOV of Lateral.
                                                10-9                                          M517-E610
=Adjust of Virtual Collimator=
NOTE
M517-E610                                             10-10
                                                                           =Adjust of Virtual Collimation=
                                                     10-11                                    M517-E610
=Adjust of Virtual Collimator=
M517-E610                                       10-12
                                                                  =Installation of Dose Meter Option=
11.2 Components
DOSE METER ASSY 2 for 12 inch Single Plane (502-27975-01)
 Ref.
       Parts No.       Name                         Quantity                Remarks
 No.
 <1>   572-46000       VACUDAP(158-00-15)               1                   DOSE METER
                                                                            (CHANBER)
 <2>     502-27976-01         DOSE METER CABLE F                   1        CABLE (30m)
                                                  11-1                                         M517-E610
=Installation of Dose Meter Option=
DC PLUG
RS-232C
12 inch CHAMBER
M517-E610                                       11-2
                                                                  =Installation of Dose Meter Option=
11.3 Installation
     11.3.1 Cable Connections
ACQ3 Cabinet
                                                                               <1>DOSE METER
                    DOSE METER
                    SERIAL CABLE
                    (Attached to Cabinet)
                                                   11-3                                       M517-E610
=Installation of Dose Meter Option=
Collimator F-50
M517-E610                                         11-4
                                                        =Installation of Dose Meter Option=
Screws
MH-200S/300 (Fix with M3x 4pc) MH-400 (Fix with M3x 3pc)
                                             11-5                               M517-E610
=Installation of Dose Meter Option=
 M720
SLOT 7 6 5 4 3 2 1
                                                                            CL0
                                                                                                        P1
                                                                            CL1
                                                                                                        P2
 M730
SLOT 7 6 5 4 3 2 1
CL0
CL1
                                          ACQ PC L 背面
                                      ACQ PC L Rear
M517-E610                                     11-6
                                                                  =Installation of Dose Meter Option=
Follow the instruction below and open [Exposure Parameters] tab on option configuration.
 (2) The GUI starts in Administration mode, so click [Option] on the side menu. And click [Exposure
   Parameters] to open the following window.
 (3) Choose a parameter in the list and then check any available combination of [Fluoro], [Rad], [Review]
   or [Displayed]. Each description of item is as follows.
Item Description
Parameter Description
                                                     11-7                                    M517-E610
=Installation of Dose Meter Option=
NOTE
    In case dGycm2      is displayed but the value is not displayed, there must be some problem with
    RS-232 communication.
    Check the connections described in 11.3 Installation.
    In case the value does not increase, there must be some problem with the chamber connections.
    Check the connection between the chamber cable and the main body described in11.3 Installation.
M517-E610                                          11-8
                                                                    =Installation of Dose Meter Option=
CAUTION
    Make sure to complete the collimator and the virtual collimation adjustment because AK and AKR
    are calculated from the data of collimator shutter aperture. When the collimator is readjusted, it is
    necessary to readjust in this section.
    11.6.1 Outline
     The dose rate at the reference location, i.e. 15cm from the isocenter toward the X-ray source
     along the beam axis, is displayed on the LCD during X-ray irradiation. The dose rate is
     acquired by the DAR-9500f dosimeter.
     The cumulative dose at the reference point is always displayed on the LCD that is also
     acquired by the DAR-9500f dosimeter.
     Measure in the modes of 15pps/H fluoroscopy, because these are same tendency in all
     modes.
FPD
                                                                              Tabletop
                                                           Collimator Cover
X-Ray Tube
                                                    11-9                                        M517-E610
=Installation of Dose Meter Option=
CAUTION
    Make sure to insert Pb mask plate for FPD protection to the grid slot of FPD cover, which is an
    accessory of C-arm.
NOTE
    If dose cannot be measured because of over range of dosimeter and etc., measure and check at least
    each field of view (both Frontal and Lateral for Bi-plane), a single X-ray condition.
M517-E610                                           11-10
                                                            =Installation of Dose Meter Option=
NOTE
If readjust settings related to the collimator for DAR-9500f,    check the operation of virtual
collimation.
    (1)    Copy [WORKSHEET for Adjusting Dose Rate Display.xls] that is displayed on the
           MPC desktop or in ServicePack CD (\Documents\EN\Dose Adjustment Sheet) to
           PC with Excel installed.
    (2)    Start up the system and login to REF PC and ACQ PC with “serviceapp” to start
           running the application.
    (3)    Click [Option] on side menu on REF PC to display [Configuration] window.
                                            11-11                                      M517-E610
=Installation of Dose Meter Option=
         (5)    Change “FOV” setting of [Shutter Positions] which did not fall within the rejection
                limit.
                If “Open” of H-Shutter, V-Shutter, and C-Shutter is different and different settings
                   for each field of view, input the same value as H-Shutter to V and C Shutters.
                   (Vary depending on field of view is acceptable)
                Set “Close” for H-Shutter and V-Shutter as “100”. Set to “50” if “Close” for
                   C-Shutter is “100”.
                Input “Open” for each field of view and “Close” for C-Shutter to Excel sheet.
                   (From B7 to B10 and C21 to C24)
         (6)    Make sure to confirm that the setting value is also set to ACQ PC (for
                Frontal/Lateral (Bi-plane option)). Move the mouse cursor to ACQ PC and open the
                same menu to confirm.
                If the setting value is not set, uncheck “Link Mode” on [System] tab and then set.
                Check “Link Mode” when setting is completed.
         (7)    Fully-open the collimator and C leaf to measure dose (dose rate). Dose
                measurement condition is “90kv” for fluoroscopy, and use the same DUP in
                11.6.3”Operating Method “. Input the measured dose value to Excel sheet from D7
                to D10 and the monitor display value (AKR) from E7 to E10. And then set the “Open
                (HVC Shutter)” for DAR-9500f to the calculated results (from F7 to F10).
         (8)    Make sure to confirm that the setting value is also set to ACQ PC (for
                Frontal/Lateral (Bi-plane option)). Move the mouse cursor to ACQ PC and open the
                same menu to confirm.
                If the setting value is not set, uncheck “Link Mode” on [System] tab and then set
                “Open”. Check “Link Mode” when setting is completed.
         (9)    Adjust “Close” of the C Shutter.
               Close (full close) the C Shutter and measure dose (dose rate). Input the measured
               dose value to Excel sheet D21 to D24 and monitor displayed value (AKR) to E21 to
               E24. And then set calculation result on Excel Sheet (from G21 to G24) to C Shutter
               of DAR-9500f.
         (10) Make sure to confirm that the setting value is also set to ACQ PC (for
                Frontal/Lateral (Bi-plane option)). Move the mouse cursor to ACQ PC and open the
                same menu to confirm.
                If the setting value is not set, uncheck “Link Mode” on [System] tab and then set
                “Close”. Check “Link Mode” when setting is completed.
         (11) Reboot the system and measure dose.
                Reboot the system and check the dose display.
M517-E610                                          11-12
                                                                =Installation of Dose Meter Option=
TESTED BY DATE
                                        Fluoroscopy (AKR)
                                                 Displayed
                                    Measured
                                                  dose on                Rejection    Result
          Mode      FOV      kV       dose                   Error (%)
                                                 DAR-9500f                 limit     (G・NG)
                                    (mGy/min)
                                                 (mGy/min)
                             70
                             90
                            120
                             70
                             90
                            120
                             70
                             90
         15pps/H                                                           30%
                            120
70
90
120
70
90
120
SIGNED
                                                11-13                                      M517-E610
=Installation of Dose Meter Option=
TESTED BY DATE
                                           Fluoroscopy (AK)
                                                   Displayed
                                       Measured     dose on                Rejection    Result
          Mode      FOV      kV                                Error (%)
                                      dose (mGy)   DAR-9500f                 limit     (G・NG)
                                                     (mGy)
                             70
                             90
                            120
                             70
                             90
                            120
                             70
                             90
         15pps/H                                                             30%
                            120
70
90
120
70
90
120
SIGNED
M517-E610                                          11-14
                                                                =Installation of Dose Meter Option=
TESTED BY DATE
                                        Fluoroscopy (AKR)
                                                 Displayed
                                    Measured
                                                  dose on                Rejection    Result
          Mode      FOV      kV       dose                   Error (%)
                                                 DAR-9500f                 limit     (G・NG)
                                    (mGy/min)
                                                 (mGy/min)
                             70
                             90
                            120
                             70
                             90
                            120
                             70
                             90
         15pps/H                                                           30%
                            120
70
90
120
70
90
120
SIGNED
                                                11-15                                      M517-E610
=Installation of Dose Meter Option=
TESTED BY DATE
                                           Fluoroscopy (AK)
                                                   Displayed
                                       Measured     dose on                Rejection    Result
          Mode      FOV      kV                                Error (%)
                                      dose (mGy)   DAR-9500f                 limit     (G・NG)
                                                     (mGy)
                             70
                             90
                            120
                             70
                             90
                            120
                             70
                             90
         15pps/H                                                             30%
                            120
70
90
120
70
90
120
SIGNED
M517-E610                                          11-16
                                                        =Installation of Scan Converter Option=
12.1 Components
<1> <2>
Parts List
                                                              Quantity
 Ref.
             Name                Parts No.                Single Bi-pla             Notes
 No.
                                                          -plane ne
 <1>    070-41401-10   Cable ASSY,BNC-P3-5.0              2       4         5m
        070-44024-11   Connector, RCAP-BNCJ               3       6         BNC to RCA Pin
 <2>
                                                                            converter connector
                                             12-1                                    M517-E610
=Installation of Dose Meter Option=
12.2 Installation
     12.2.1 Cable Connections
                                       VIDEO
                                       OUT1              REF Cabinet
                                       [CH6]
                                                         REF PC
                                       VIDEO
                                               (Note2)
                                       OUT2
                                       [CH5]
<1>
(Note2) VIDEO OUT2 is an option. Add DVI-SUB ASSY (502-27193) to use VIDEO OUT2.
M517-E610                                                12-2
                                                            =Installation of Scan Converter Option=
Connect <1>
 Connect <1> cable to VIDEO OUT connector on the REF Cabinet rear.
  (Note)   In case of Bi-plane, connect to the panel part in which F is printed for Frontal side, and
           the panel part in which L is printed for Lateral side.
                                               12-3                                     M517-E610
=Installation of Dose Meter Option=
SW1
DVI MAIN
<DipSW S1>
       Set DipSW 1-7 as follows.
M517-E610                                       12-4
                                                           =Installation of Scan Converter Option=
                                                    12-5                               M517-E610
=Installation of Dose Meter Option=
        (Note) CH5 [VIDEO] is also used for maintenance LCD output setting (VGA). Therefore, CH5
                [VIDEO] input channel must be set as “In-1 [ACQ-1]”even if DVI-SUB ASSY is added.
M517-E610                                       12-6
                                                 =Installation of Scan Converter Option=
                                        12-7                                 M517-E610
=Installation of Dose Meter Option=
M517-E610                                       12-8
                                                            =Backup and Restore Instruction=
   By using [DAR-9500F BACKUP SW/PC BACKUP S/W], backup of creating system and site
   installation and after the adjustment of the system backup can be saved in HDD and DVD
   media.
   Also, digital system (Digital Image Processing System DAR-9500f) can be restored as “backup
   creating mode” by using backup media that was created with [DAR-9500F BACKUP/SW PC
   BACKUP S/W].
   DAR-9500f is configured with Acquisition PC (ACQ PC) and Reference PC (REF PC), so
   backup and restore on each computers.
   When back up REF PC, an image as the following Figure 1 might be displayed. Normal image
   will be displayed if you turn off the power of ACQ PC or turn on the power of DVI-MAIN
   again.
                                              13-1                                 M517-E610
=Backup/Restore Instruction=
       When backup and restore on Acquisition PC (ACQ PC), use maintenance keyboard, mouse and
       LCD in ACQ3 cabinet. After power off ACQ PC, connect cables as follows.
SLOT 7 6 5 4 3 2 1
                                                                                       Maintenance LCD
                                                                                         (8inch LCD)
 Ref.                                                          Length
                  Name                  Parts No.                                      From                      To
  No.                                                           [m]
 1(*)     DVI CABLE                (Attached to                            ACQ PC                    Maintenance LCD
                                   Maintenance LCD)          -             SLOT6                     [DVI]
                                                                           DVI CONNECTOR
 2        USB CABLE                (Attached to                            ACQ PC                    Maintenance LCD
                                                             -
                                   Maintenance LCD)                        USB CONNECTOR             [USB]
 A-4      KEYBOARD                 (Attached to ACQ                        ACQ PC                    Keyboard Holder
                                   PC)                       -             KEYBOARD
                                                                           CONNECTOR
 A-5      MOUSE                    (Attached to ACQ                        ACQ PC                    Keyboard Holder
                                                             -
                                   PC)                                     USB CONNECTOR
                                                    NOTE
     After backup and restore on ACQ PC, reconnect DVI cable.
M517-E610                                             13-2
                                                            =Backup and Restore Instruction=
                                             13-3                                M517-E610
=Backup/Restore Instruction=
       (3) [Windows is loading files…] window is displayed, and then Windows Startup
           window is displayed.
Click [Accept].
M517-E610                                13-4
                                     =Backup and Restore Instruction=
                              13-5                        M517-E610
=Backup/Restore Instruction=
Click [Next>].
M517-E610                                    13-6
                                                  =Backup and Restore Instruction=
Put a check mark on [system Reserved (C:)] and [System (E:)]. Click [Next>].
Click [Browse…].
                                     13-7                              M517-E610
=Backup/Restore Instruction=
M517-E610                                 13-8
                                                    =Backup and Restore Instruction=
    Input an optional name (We recommend created date and DAR revision such
    as “20121019_Rev510”) and click [OK].
                                        13-9                             M517-E610
=Backup/Restore Instruction=
M517-E610                                  13-10
                                                      =Backup and Restore Instruction=
    Check [Divide into smaller files to simplify archiving] and set [file size (MB)] to
    “3857”. And click [OK].
Click [Next>].
                                        13-11                                M517-E610
=Backup/Restore Instruction=
Click [Close].
M517-E610                                     13-12
                                              =Backup and Restore Instruction=
Click [Exit].
                                    13-13                          M517-E610
=Backup/Restore Instruction=
M517-E610                              13-14
                                                   =Backup and Restore Instruction=
(17) Open “F:\Ghost-M720-XXX (XXX would be “REF” for REF PC and “ACQ” for
     ACQ PC)\*** (*** would be folder made by (6))” with Explorer. And check that
     backup file is created.
       System_Reserved_C_Drive.v2i
       System_E_Drive.v2i
       System_E_Drive_s01.v2i
                                     13-15                              M517-E610
=Backup/Restore Instruction=
                                        NOTE
  On DAR-9500F BACKUP S/W, backup image cannot be written directly to DVD.
  Therefore, using the function of GHOST 15.0 dividing into smaller files, backup
  image to write to DVD.
                                        NOTE
  As backup image is too large capacity, do not write to CD but DVD.
M517-E610                                13-16
                                                   =Backup and Restore Instruction=
(3) Open “F:\Ghost-M720-XXX (XXX would be “REF” for REF PC and “ACQ” for
    ACQ PC)\*** (*** would be the file made by 13.3.1.2(6))” with Explorer. If there
    would be several backup files, files have “_s01”, “_s02” for the file name. Save
    each file to DVD.
   Check [With a CD/DVD player] and put an optional name to the [Disc title]. (We
   recommend created date and disk number such as “10192012_01” for the
   [Disc title])
   Click [Next].
                                     13-17                               M517-E610
=Backup/Restore Instruction=
M517-E610                                 13-18
                                                   =Backup and Restore Instruction=
Click [Next].
                                        13-19                            M517-E610
=Backup/Restore Instruction=
       (9) When completed, “You have successfully burned your files to the disc” is
           displayed.
Click [Finish].
M517-E610                                13-20
                                                            =Backup and Restore Instruction=
                                          NOTE
If HDD is broken and replaced, refer to “13.3.3 In Case of Replacing HDD” for
restoring.
(2) When [Press any key to boot from CD or DVD..] is displayed, press any key.
                                             13-21                                 M517-E610
=Backup/Restore Instruction=
       (3) [Windows is loading files...] window is displayed, and Windows startup window
           is displayed after a while.
Click [Accept].
M517-E610                                13-22
                                      =Backup and Restore Instruction=
                              13-23                        M517-E610
=Backup/Restore Instruction=
Click [Next>].
M517-E610                               13-24
                                                   =Backup and Restore Instruction=
Click [OK].
Click [Browse…].
                                     13-25                              M517-E610
=Backup/Restore Instruction=
M517-E610                                 13-26
                                                  =Backup and Restore Instruction=
Click [Next>].
Click [Add…].
                                        13-27                          M517-E610
=Backup/Restore Instruction=
Click [Browse…].
M517-E610                                  13-28
                                                 =Backup and Restore Instruction=
                                      13-29                           M517-E610
=Backup/Restore Instruction=
Click [OK].
Check [Verify recover point before recovery] and then click [Next >].
M517-E610                                  13-30
                                                 =Backup and Restore Instruction=
Click [Yes].
                                    13-31                             M517-E610
=Backup/Restore Instruction=
       (15) Restore is started. Since the backup is created in a number of media, the
           following requested dialog for changing media displays during restore.
            After requesting, insert the next media and press [OK] button after a few
            seconds.
M517-E610                                  13-32
                                                          =Backup and Restore Instruction=
(2) When [Press any key to boot from CD or DVD..] is displayed, press any key.
    (3) [Windows is loading files...] window is displayed, and Windows startup window
        is displayed after a while.
                                            13-33                              M517-E610
=Backup/Restore Instruction=
Click [Accept].
M517-E610                              13-34
                                                  =Backup and Restore Instruction=
                                          13-35                        M517-E610
=Backup/Restore Instruction=
Check [Load previously saved MBR from file] and then click [Next >].
M517-E610                                 13-36
                                              =Backup and Restore Instruction=
                                   13-37                           M517-E610
=Backup/Restore Instruction=
Click [Finish].
M517-E610                                 13-38
                                                    =Backup and Restore Instruction=
Click [Next>].
                                       13-39                             M517-E610
=Backup/Restore Instruction=
Click [OK].
Click [Browse…].
M517-E610                                 13-40
                                                 =Backup and Restore Instruction=
                                      13-41                           M517-E610
=Backup/Restore Instruction=
Click [Next>].
Click [Add…].
M517-E610                                 13-42
                                              =Backup and Restore Instruction=
Click [Browse…].
                                      13-43                        M517-E610
=Backup/Restore Instruction=
M517-E610                                 13-44
                                                =Backup and Restore Instruction=
Click [OK].
Click [Add…].
                                     13-45                           M517-E610
=Backup/Restore Instruction=
M517-E610                              13-46
                                                  =Backup and Restore Instruction=
Click [OK].
Check [Verify recovery point before recovery] and then click [OK].
                                     13-47                              M517-E610
=Backup/Restore Instruction=
Click [Yes].
M517-E610                                 13-48
                                                   =Backup and Restore Instruction=
(18) Restore is started. Since the backup is created in a number of media, the
    following requested dialog for changing media displays during restore.
   After requesting, insert the next media and press [OK] button after a few
   seconds.
                                     13-49                              M517-E610
=Backup/Restore Instruction=
(2) When [Press any key to boot from CD or DVD..] is displayed, press any key.
       (3) [Windows is loading files…] window is displayed, and then Windows Startup
           window is displayed.
M517-E610                                13-50
                                                =Backup and Restore Instruction=
Click [OK].
                                   13-51                             M517-E610
=Backup/Restore Instruction=
(2) “Select local source drive by clicking on the drive number” is displayed.
M517-E610                                   13-52
                                                     =Backup and Restore Instruction=
 In case of 3 partitions
    Select [Volume Label] with “System Reser” and “System” Partition (Part) and
    click [OK].
 In case of 4 partitions
                                        13-53                             M517-E610
=Backup/Restore Instruction=
            Select “Ghost-M720-XXX (XXX would be “REF” for REC PC and “ACQ” for
            ACQ PC).
M517-E610                                 13-54
                                              =Backup and Restore Instruction=
                                     13-55                         M517-E610
=Backup/Restore Instruction=
Click [Fast].
M517-E610                                  13-56
                                                =Backup and Restore Instruction=
Click [Continue].
                                       13-57                         M517-E610
=Backup/Restore Instruction=
Click [Yes].
M517-E610                                13-58
                                              =Backup and Restore Instruction=
                                      13-59                        M517-E610
=Backup/Restore Instruction=
                                            NOTE
  As backup image is too large capacity, do not write to CD but DVD.
(1) Remove [PC BACKUP G11 S/W] from CD/DVD drive and set a blank DVD.
(3) “Select local source drive by clicking on the drive number” is displayed.
M517-E610                                   13-60
                                                     =Backup and Restore Instruction=
 In case of 3 partitions
    Select [Volume Label] with “System Reser” and “System” Partition (Part) and
    click [OK].
 In case of 4 partitions
                                        13-61                             M517-E610
=Backup/Restore Instruction=
                                          NOTE
  “******” for [@CD-R1 ******] will be different depending on PC.
Click [Save].
M517-E610                                 13-62
                                                =Backup and Restore Instruction=
Click [Fast].
Click [No].
                                   13-63                             M517-E610
=Backup/Restore Instruction=
Click [Yes].
M517-E610                                  13-64
                                                =Backup and Restore Instruction=
(11) “New CD-R disk needed (51902)” is displayed if the media of two or more is
    required.
Click [Continue].
                                   13-65                             M517-E610
=Backup/Restore Instruction=
M517-E610                                  13-66
                                                    =Backup and Restore Instruction=
Click [Yes].
                                      13-67                              M517-E610
=Backup/Restore Instruction=
(2) When [Press any key to boot from CD or DVD..] is displayed, press any key.
       (3) [Windows is loading files…] window is displayed, and then Windows Startup
           window is displayed.
M517-E610                                  13-68
                                                =Backup and Restore Instruction=
Click [OK].
                                   13-69                             M517-E610
=Backup/Restore Instruction=
M517-E610                                13-70
                                                         =Backup and Restore Instruction=
                                        NOTE
Do not select [@CD-R1 ****] (**** indicates the model name of CD/DVD). If
accidentally select it, Ghost application may be shut down after a while. In this case,
start over “13.4.1.1 Startup GHOST”.
                                           13-71                               M517-E610
=Backup/Restore Instruction=
Click [Continue].
M517-E610                                  13-72
                                                 =Backup and Restore Instruction=
Click [Yes].
                                    13-73                             M517-E610
=Backup/Restore Instruction=
M517-E610                                 13-74
                                                           =Backup and Restore Instruction=
(2) When [Press any key to boot from CD or DVD..] is displayed, press any key.
    (3) [Windows is loading files…] window is displayed, and then Windows Startup
        window is displayed.
                                            13-75                                 M517-E610
=Backup/Restore Instruction=
Click [OK].
M517-E610                               13-76
                                                      =Backup and Restore Instruction=
                                        13-77                              M517-E610
=Backup/Restore Instruction=
                                          NOTE
  Do not select [@CD-R1 ****] (**** indicates the model name of CD/DVD). If
  accidentally select it, Ghost application may be shut down after a while. In this case,
  start over “13.4.3.1 Startup GHOST”.
M517-E610                                  13-78
                                                    =Backup and Restore Instruction=
(4) “Select Local destination drive by clicking on the drive number” is displayed.
 In case of 2 partitions
                                      13-79                               M517-E610
=Backup/Restore Instruction=
 In case of 3 partitions
M517-E610                                    13-80
                                                 =Backup and Restore Instruction=
                                    13-81                             M517-E610
=Backup/Restore Instruction=
       (2) Right-click on Windows task bar [Start]-[My Computer] and click [Manage] on
           the pop-up menu.
M517-E610                                     13-82
                                                 =Backup and Restore Instruction=
                                   13-83                              M517-E610
=Backup/Restore Instruction=
Click [Next].
Click [Next].
M517-E610                                13-84
                                                      =Backup and Restore Instruction=
Check that [Assign the following drive letter:] is “D” and click [Next].
    Select [Format this volume with the following settings:] and set parameters as
    follows. And check [Perform a quick format] and then click [Next].
       File system:            NTFS
       Allocation unit size:   8192
       Volume label:           Cache
                                       13-85                                   M517-E610
=Backup/Restore Instruction=
Click [Finish].
M517-E610                               13-86
                                                  =Backup and Restore Instruction=
(10) The system condition would be as follows after setting the volume.
(11) Reboot the system. Login with “operator” and check the operation of
    radiography.
                                    13-87                              M517-E610
=Backup/Restore Instruction=
No Text
M517-E610                      13-88
                                                                    =StentView (Option)=
14. StentView (Option)
14.1 Introduction
Installation procedure of StentView Option (here after called StentView) for DAR-9500f is
described in this chapter.
StentView applies real-time image processing during radiography and display the enhanced
stent image. Automatically render the markers on both ends of balloon and use the position
information for image processing.
ACQ PC REAR
2) Startup DAR-9500f application and display new study. Check that [StentView] button is
   displayed on the side menu of REF and ACQ PC.
                                             14-1                             M517-E610
=StentView (Option)=
Set Enable/Disable of using StentView for each DUP. If StentView is set to disable, you
cannot select [StentView] button on the side menu and normal radiography wil operation
even though pressing Stent View button on IVR-NEO.
Radiography DUP contains [Cardiac] category enables StentView setting upon factory
setting. Change settings if necessary.
(1)
                                               NOTE
     Do not select StentView except DUP for Cardiac DA. Stent View will operate properly only in
     Cardiac DA radiography mode.
(2)
M517-E610                                    14-2
                                                                    =StentView (Option)=
(3)
After the setting, put IVR NEO sticker of [StentView] button on the IVR NEO button.
                                             14-3                              M517-E610
=StentView (Option)=
To assign ON/OFF of StentView to function key, set an optional function key to [Activate
StentView/ValveView] on [Physicians] tab-[IVR-NEO and Keyboard Shortcuts] ((4)).
“F” is indicated on the head of function key. In the following example figure, set StentView
for [F1] key.
(4)
After the setting, put function key sticker of [StentView] button on the keyboard.
M517-E610                                     14-4
                                                                      =StentView (Option)=
                                                                                  Enable/
       Item                           Configuration Procedure
                                                                                  Disable
Coordination with        Check this item to disable ROI when chaning            Enable
Geometry                 geometry.
1-ROI Initial Size       Specify the initial size of ROI when 1-ROI is          Enable
(%)                      selected during study. Default is “30”.
2-Proximal ROI           Specify the initial size of Proximal ROI when 2-ROI    Enable
Initial Size(%)          is selected during study.
2-Distal ROI             Specify the initial size of ROI when Distal 2-ROI is   Enable
Initial Size(%)          selected during study.
ROI Minimum Size         Specify the minimum size of ROI. Default value is      Disable
(%)                      10% and normally do not change it.
Hold the last ROI        Check this check box to keep the mode of position      Enable
Status                   and size of ROI used last. Uncheck to set ROI in the
                         center of window and select with initial size.
                         Check box is checked for the default setting.
Auto-Replay       last   Check this check box and ROI is selected on the        Enable
Fluoro                   LastImageHold, the last 2 seconds of fluoroscopy
                         replay automatically.
                         Check box is checked for the default setting.
StentView Scale          Do not change value. Leave the default value as        Disable
                         “1.3”.
                                               14-5                               M517-E610
=StentView (Option)=
When sending Stent View image to server, image processing with Gateway is necessary.
See “M517-E487 DAR-9400f Gateway Installation Manual” to set Gateway filter.
        While selecting enabled Stent View DUP, place 9B-Cu and etc. on the tabletop,
        press Stent View button, and press radiography switch as it is set to Stent View
        radiography mode. ACQ monitor layout should be changed and indicated as
        follows. At this point, Dynamic View area should be an intermediate brightness
        image like DSA.
        Moving the pre-used Stent package in front of FPD and perform Stent View
        radiography. Check that moving Stent/Catheter is displayed on Dynamic View area
        and Stent neighborhood is displayed on Stent View area.
M517-E610                                  14-6
                                                                    =Adding Bi-Plane System=
                                            NOTE
  Installation of Bi-plane system with hospital which already installed DAR-9500f Single-plane
  system is described in this chapter.
                                                15-1                                 M517-E610
=Adding Bi-Plane System=
                                     I-SYS/IOF/FAN/POWER
 4     ADITIONAL DOOR 3 Bi                                               1 set      ACQ3 Cabinet
                                     SUPPLY/etc.
       ADDITIONAL
 5     TRANSFORMER UNIT 3            ADDITIONAL KIT                      1 set      ACQ3 Cabinet
       Bi
       ADDITIONAL ACQ3 Bi
 6                                   ACQ3 CABLE                          1 set      ACQ3 Cabinet
       WIRING
M517-E610                                          15-2
                                                                       =Adding Bi-Plane System=
    According to the following figure, mount I-SYS IOF ASSY/5V POWER SUPPLY (PS4)/FAN.
    Mount it in the direction of receiving the wind to the I-SYS IOF side. Connect the cable from FAN
    to L0/L100 of Transformer Unit.
                                                  15-3                                  M517-E610
=Adding Bi-Plane System=
          To Transformer Unit3
          TB2
            To Transformer Unit3
            TB3
Additional Label
Additional Unit
M517-E610                               15-4
                                                              =Adding Bi-Plane System=
  Use two-way table top to connect HUB POWER CABLE to the same socket as Sinle-plane HUB.
  And use an adaptor to space the earth terminal for CP2 socket.
                                              UPS
 UPS
                                                                            HUB L
                HUB F
                                                                            HUB F
Single-Plane Bi-Plane
                                            15-5                              M517-E610
=Adding Bi-Plane System=
[B-1] [B-2]
                                                                      CN16
                    [C-1]         DVI-LAN
                                CN31
ARC CN3
                                                                      CN6
                                CN17
                                                                                                                   ARC
                                                                                                            CN18
                                                 IOF(L)
                                                                      CN21
                                CN4 LAN
                                                                                                                                I-SYS
                                                        I-SYS                                                  CN2       CN14
                                       CN2      CN14                                                                            BOARD F
                                                        BOARD L
                                                                                                    [C-4]
                                                                                                              [C-3]
                               [C-2]
                                                                                                                                      [D-1]
                                                                                                                                      [D-2]
[A-1]
        [A-2]                                                   [A-5]
                                        PS4
                                 CN1
CN2
                                        5V10A
                                                                     J200                 UPS
                       [A-3]
                                                                        [A-6]                        [A-4]
                                                                                                                                          HUB L
                                                                    J107
  CP2 L                                          TRANS
                LOUT1/LOUT2                      UNIT                                                                                     HUB F
M517-E610                                                                            15-6
                                                                                  =Adding Bi-Plane System=
 Ref.                                                   Length
                  Name                   Parts No.                          From                        To
 No.                                                     [m]
A-1     CP2 Attachment POWER         Attached to CP/               CP2 L                    TRANSFORMER UNIT
        CABLE/ADOPTER EARTH          071-60813-01                  [POWER SUPPLY]           [CP2 L]
                                                             -
        KPR-13
        (Triode-Diode Connector)
A-2     FAN CABLE,DOOR               565-10731                     ACQ 3 DOOR               TRANSFORMER UNIT
                                                            1.3
                                                                   I-SYS L REARA FAN        [LOUT1/LOUT2]
A-3     AC CABLE,PS4                 565-11263                     ACQ 3 DOOR               TRANSFORMER UNIT
                                                            1.05
                                                                   POWER PS4 [CN1]          [LOUT1/LOUT2]
A-4     TABLE TAP,TAP-KE2-1          071-60337-04                  UPS [HUB]                HUB F Attachment POWER
                                                                                            CABLE/
                                                             1
                                                                                            HUB L Attachment POWER
                                                                                            CABLE
A-5     I-SYS L DC CABLE             565-11262                     ISYS L [CN21] /          POWER CABLE PS4 CN2 /
                                                             1
                                                                   IOF L [CN3]              [J200]
A-6     I-SYS L CN6 CABLE            565-11264                     I-SYS L                  TRANSFORMER UNIT
                                                            1.65
                                                                   [CN6]                    [J107]
 Ref.                                                   Length
                  Name                   Parts No.                          From                        To
 No.                                                     [m]
B-1     I-SYS-CP2 IO CABLE           565-10423-12                  CP2 L                    I-SYS L
                                                        1
                                                                   [EXTERNAL SYNC]          [CN20]
B-2     I-SYS-CP2 LVDS CABLE         562-27740-12                  CP2 L                    I-SYS L
                                                        1          [16BIT DIGITAL           [CN1]
                                                                   VIDEO]
 Ref.                                                   Length
                  Name                   Parts No.                          From                        To
 No.                                                     [m]
C-1     LAN CABLE DVI                   562-28024-01                 DVI MAIN F             I-SYS L
                                                            1
        (Connection-Changeable)                                      IN6(J15)               [CN31]
        *Change the cable connected to I-SYS F [CN31] to I-SYS L [CN31].
C-2     DAR ARCNET ISYS-ISYS            563-75790-14                 I-SYS L                I-SYS F
                                                            0.8
                                                                     [CN17]                 [CN18]
C-3     I-SYS FL ECG CABLE              565-11200-01                 I-SYS L                I-SYS F
                                                            0.8
                                                                     [CN14]                 [CN14]
C-4     I-SYS FL CABLE                  562-27844-01                 I-SYS L                I-SYS F
                                                            0.8
                                                                     [CN2]                  [CN2]
C-5     CABLE,SERIAL                    565-10470-05                 I-SYS L                MPC
                                                            1.7
        CROSS L-MPC                                                  [CN16]                 [COM5]
 Ref.                                                   Length
                  Name                   Parts No.                          From                        To
 No.                                                     [m]
D-1     LAN CABLE 3 HUB L2           565-10447-12                  CP2 L                    LAN HUB L
                                                        1
                                                                   [ETHERNET]               [CH2]
D-2     LAN CABLE 3 HUB L3           565-10447-13                  I-SYS L                  LAN HUB L
                                                        2
                                                                   [CN4]                    [CH3]
D-3     LAN CABLE 3 HUB L8           565-10447-14                  LAN HUB L                LAN HUB F
                                                        1
                                                                   [CH8]                    [CH7]
                                                        15-7                                    M517-E610
=Adding Bi-Plane System=
                                                        Change theこの4本を接続変更
                                                                   connection of these 4 cables.
J30
LD29
J34
                              J2                                                          LD20          J16
                                                                                                  LD8
                                                                                          LD19          J15
                                                                                                  LD7
                                                                                            SW7
                              J1
                                           J9            LD1                              LD18          J14
                                   LD30                                                           LD6
                           J31            SW1 SW5 SW4 SW3 SW2                             LD17          J13
                                                                                                  LD5
                                    LD32    SW6         LD31           LD2          LD3     LD4
M517-E610                                              15-8
                                                                    =Adding Bi-Plane System=
                                                 NOTE
   For Bi-plane system, be sure to connect power supply cable for Frontal FPD cooling unit
   from ACQ3 Cabinet to ACQ3L cabinet because of electricity capacity.
                                                    15-9                           M517-E610
=Adding Bi-Plane System=
DVI-SUB
      SCREW
      PAN HEAD
      M2.6x5 2pc
          DVI-OUT
                                                                        After removing the nut of BNC,
                                                                        mouting board.
Mount the lock nut 2pc to rear DVI connector (CN1) of DVI-SUB board.
                                          Length
      Name              Parts No.                             From                         To
                                           [m]
DVI CABLE 3 REF L   565-10467-11      1              REF PC VIDEO CARD         DVI-SUB L
                                                     DISPLAY PORT
                                                     - DVI CONVERSION CABLE
M517-E610                                          15-10
                                                                   =Adding Bi-Plane System=
                                              15-11                             M517-E610
=Adding Bi-Plane System=
                                                NOTE
   Factory setting of IP address for ACQ PC L / CP2 L/ I-SYS L are shown as follows.
         ACQ PC L: 192.168.100.25
         CP2 L:       192.168.100.24
         I-SYS L:     192.168.100.27
   For more than two systems, if there is a rival IP address with IP address above in the same
   network, separate the network cable with other system before applying the power, and
   connect that cable again after changing IP address.
M517-E610                                       15-12
                                                                         =Adding Bi-Plane System=
                                                  15-13                               M517-E610
=Adding Bi-Plane System=
M517-E610                                       15-14
                                                                    =Adding Bi-Plane System=
Setting Display 2
4. Click [Apply] and check that screen is displayed on REF L monitor. And click [OK] to exit.
                                                 15-15                                M517-E610
=Adding Bi-Plane System=
2. Click [Start] at the left bottom of screen and click [All Programs]-[Synergy].
M517-E610                                        15-16
                                                       =Adding Bi-Plane System=
                                          15-17                     M517-E610
=Adding Bi-Plane System=
6. Double-click “Unnamed”.
M517-E610                                      15-18
                                                                =Adding Bi-Plane System=
Click [OK].
                                                 15-19                       M517-E610
=Adding Bi-Plane System=
9. [Synergy] is displayed.
         Click [Apply].
         And check that the mouse pointer is moved to ACQ PC L.
M517-E610                                    15-20
                                                                         =Adding Bi-Plane System=
                                                   15-21                                   M517-E610
=Adding Bi-Plane System=
M517-E610                                 15-22
                                                                             =UserQC (Option)=
Click [Yes].
Click [No].
                                               16-1                                  M517-E610
=UserQC (Option)=
Enter user name (your name) and password (UserQC), then click [OK].
M517-E610                                   16-2
                                                               =UserQC (Option)=
Click [No].
                                        16-3                          M517-E610
=UserQC (Option)=
Click [No].
M517-E610                                 16-4
                                                       =Apply UPG1 for PC Replacement=
    Introduction
This chapter describes instructions of applying UPG1 when replacing PC.
Apply UPG1 according to this chapter when replacing REF PC and ACQ PC.
This is not necessary for usual installation.
    Applying UPG1
Refer to the document in “565-13165 TRINIAS ALEXA UPG1”, please apply the UPG1 to the
replacement PC.
                                                A-1                         M517-E610
=Apply UPG1 for PC Replacement=
M517-E610                                  A-2
                                                 =How to install license when replacing PC=
   Inroduction
   How to install DAR license and SCORE PRO license is described in this manual.
   When replacing REF PC or ACQ PC, install DAR licnese according to this chapter.
   And when replacing ACQ PC, install SCORE PRO license according to this chapter.
   This installation is not necessary for normal installation.
                                            NOTE
  For M730,dongle of DAR license is installed inside PC.
                  ATX
                  Power Supply
                                                  B-1                                M517-E610
= How to install license when replacing PC =
    DAR License
1) Login to replaced PC (REF PC or ACQ PC) with service mode.
                                              NOTE
   For the system with Stent View option, a root of DAR license dongle may be “H:\” on
   ACQ PC.
3) Double-click “Licenses.reg”.
Click [Yes].
Click [OK].
M517-E610                                      B-2
                                                     =How to install license when replacing PC=
   If SCORE-PRO license dongle is installed, dongle information will be displayed in the list as above
   figure.
   “No key found” in [Status] means that the license is not registered. And if nothing is displayed in the
   list, dongle is not installed.
5) Select [File] – [Import License File…] menu and import the license file (***.def) in CD.
      For Bi-plane system, there are two files in CD. Import a license file appropriate to the last 4
      digits below [Hostid].
                                                      B-3                                   M517-E610
= How to install license when replacing PC =
    6) To input license key but not a license file, click and select dongle information displayed in the
       list, and input license key on [CVLicense-Set License] window displayed in [Edit] – [Set
       License Key for selected product…] menu.
Input provided license key in [Key], and click [Set Key] button.
M517-E610                                          B-4
                                                 =How to install license when replacing PC=
   1) According to the previous chapter, startup “CVLicence.exe” as in the mode that license
      registration is completed.
   2) Select [File] – [Export License File…] menu to write license information to “CVLicense.def”,
      and copy to CD.
                                                  B-5                                M517-E610
= How to install license when replacing PC =
M517-E610                                      B-6
                                                                       =Troubleshooting of FPD=
Appendix C. Troubleshooting of FPD
C.1 Introduction
This chapter describes troubleshooting of FPD.
C.2 Troubleshooting
NOTE
   If there is the time lag or no confirmation, use the tool (“Viva” or “HostDown.exe”) for setting
   the time of CP2. (Refer to “C.3.3 Setting Time of CP2”)
            Enable to acquire the image normally (with an object), but if defects (points
             and lines) are visible, use “Defect Map Editor” of “Viva” to register defects. (Refer
             to “8.10 FPD Defective Registration” in “M517-E657 Image Quality Adjustment
             Manual”)
            Enable to acquire the image normally (with an object), but if the image quality is not
             normal, check the clamp of LVDS cable and connection condition of connector of ACQ-
             PC Camera Link cable.
    2) Check the color of light of Fiber LED for CP2. A green light indicates normal
         mode and a red/orange light indicates error mode between CP2 and Receptor.
3) Check the number of 7-segment LCD for CP2. (Refer to Table C-3)
                                              C-1                                        M517-2452
=Troubleshooting of FPD=
    C.2.3 In Case of Gain Calibration Does Not Complete
    1) Use “Viva” to check the time setting. (Refer to “C.3.3 Setting Time of CP2)
           For IBS Calib Time out error, set “IBS Calibration” OFF or load backup gain calibration
            data to CP2 using “CP2 Calib Tool”, and execute gain calibration again.
           If Median value is lower or higher than default, check that “IBS Calib” is executed. And
            if it is executed, check H/W related to IBS.
M517-2452                                     C-2
                                                                =Troubleshooting of FPD=
C.3 Maintenance of CP2
                                          C-3                                   M517-2452
=Troubleshooting of FPD=
    C.3.2 Writing CP2 F/W
  If startup of CP2 does not complete and error “32” and “43” is displayed alternately,
  rewrite CP2 F/W with the following instructions.
2) Click [Edit]-[Preferences…].
M517-2452                               C-4
                                                     =Troubleshooting of FPD=
3) [ViVA-Preferences] is displayed.
Click [OK].
                                    C-5                              M517-2452
=Troubleshooting of FPD=
NOTE
M517-2452                                    C-6
                                                              =Troubleshooting of FPD=
7) [ViVA-Transmit Files] is displayed.
    Put a check mark on [Transmit This File] of [Receptor Configuration Data File] and click
    [Browse].
                                      C-7                                      M517-2452
=Troubleshooting of FPD=
    9) [ViVA-Transmit Files] is displayed again.
Click [Transmit].
Click [No].
M517-2452                              C-8
                                                        =Troubleshooting of FPD=
11) [ViVA Message Box] is displayed.
Click [Yes].
Click [OK].
13) [ViVA Message Box--Error] is displayed. (For SFD-0808 and SFD-0808AF only)
Click [OK].
                                   C-9                               M517-2452
=Troubleshooting of FPD=
    14) [ViVA Message Box--Data Loss Warning] is displayed. (For SFD-0808 and SFD-
        0808AF only)
Click [Yes].
        Turn OFF the power supply switch of CP2 rear and turn ON the power supply after
        approximately 10 seconds.
M517-2452                                 C-10
                                                  =Troubleshooting of FPD=
C.3.3 Setting Time of CP2
1) Execute the following program on REF-PC.
      In case of FPD is SFD-0808/SFD-1212
    [viva.exe] in [C:\Program Files(x86)\Varian\PaxScan_K05_Rel x.x\Viva]
    folder.
2) Click [Edit]-[Preferences…].
                                  C-11                         M517-2452
=Troubleshooting of FPD=
    3) [ViVA-Preferences] is displayed.
Click [OK].
M517-2452                                 C-12
                                                                 =Troubleshooting of FPD=
NOTE
                                          C-13                                      M517-2452
=Troubleshooting of FPD=
    8) Click [File]-[Exit] and exit “ViVA”.
M517-2452                                C-14
                                                                   =Troubleshooting of FPD=
  C.3.4 Confirmation of CP2 Errors
If error related to FPC occurs, obtain startup log of CP2 and check error number displayed on CP2.
Table C-3 is the error code list displays on 7-segment LCD of CP2.
                                           C-15                                      M517-2452
=Troubleshooting of FPD=
       No.                              Errors                   Remarks
      31     Temperature Monitor Error
      32     Receptor Error
      33     IPS Error
      34     Receptor Medium Low
      35     Receptor Medium High
      36     Flash File CRC Does Not Match to Calculated Value
      37     Receptor Firmware Error
      38     Gain Image data copy to IPU memory fails
      39     Problem sending the Error Log to the Host.
      40     Problem Clearing the Error Log.
      41     Could not read the error log sector in flash.
      42     Could not write to the error log sector in flash.
      43     Link Ready Error Detected.
      44     No Flash Memory Available
      45     CP2 Clock is NOT set
      46     Black Image Error Detected
      47     ABS Voltage in Undefined State
      48     Task Lock Failed
      49     Internal Timer Error
M517-2452 C-16