ADS1000 User Manual (1.3)
ADS1000 User Manual (1.3)
Contents in this Manual are not allowed to copy, extract and translate before being
                                          I
ADS1000 Series Digital Storage Oscilloscope Brief
Introduction
ADS1000 Series Models:
      Model:
         ADS1000C+ Series: ADS1102C+, ADS1062C+, ADS1042C+, ADS1022C+
         ADS1000CA Series: ADS1152CA, ADS1102CA, ADS1062CA
         ADS1000CM Series: ADS1152CM ,ADS1102CM, ADS1062CM, ADS1042CM
Characteristic:
500MSa/s(ADS1000C+ series);
40Kpts(ADS1000CA series)
2Mpts(ADS1000CM Series)
● Pass/Fail function.
● Two groups reference waveforms and twenty groups capture waveforms and
twenty groups setups internal save/recall function and USB flash drive save/recall
function.
● Cursor measure covers Manual mode, Track mode and Auto mode.
                                               II
         ● Menu display in the form of pop-up in order to convenient for users using it.
● Standard setup interface: USB Host :Support USB flash drive storage/recall
function and update firmware; USB Device: Support PictBridge compatible printer
          ● Certification.
          ● guaranty card
          ● CD(including EasyScope3.0 computer software system)
● User Manual
● USB cable
                                             III
ADS1000 User Manual                                                                        (V1.3)
Review the following safety precautions to avoid injury and prevent damage to this
product or any products connected to it. To avoid potential hazards, use this product
only as specified.
Use Proper Power Cord. Use only the power cord specified for this product and
Ground the Product. This product is grounded through the grounding conductor of
the power cord. To avoid electric shock, the grounding conductor must be connected
to earth ground. Before making connections to the input or output terminals of the
Connect the Probe Properly. The probe ground lead is at ground potential. Do not
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings
and marking on the product. Consult the product manual for further ratings
Do Not Operate Without Covers. Do not operate this product with covers or panels
removed.
Use Proper Fuse. Use only the fuse type and rating specified for this product.
                                              IV
                                                                        General safety summary
Avoid Exposed Circuitry. Do not touch exposed connections and components when
power is present.
Do Not Operate With Suspected Failures. If you suspect there is damage to this
DANGER: Indicates an injury hazard immediately accessible as you read the marking.
WARNING: Indicates an injury hazard not immediately accessible as you read the
marking.
         This symbol is used where caution is required. Refer to the accompanying information
         or documents in order to protect against personal injury or damage to the instrument.
                                                V
ADS1000 User Manual                                                                       (V1.3)
          This symbol shows that the switch is a standby(power) switch located on the front of the
oscilloscope. Pressing this button toggles the oscilloscope’s state between operation and standby
mode. This switch is not a disconnect device. The instrument can only be placed in a complete
power off state by flipping the main power switch to the off (Zero) postion or by removing the
power cord .
                                               VI
Preface
This manual contains operating information for the ADS1000 Series Digital Storage
◆ “Accidence” chapter describes the front panel; display area of the oscilloscope;
◆ “Service and Support” chapter introduce warranty and technology of ADS series
products.
series oscilloscopes
◆ “AppendixB: Default Setup” chapter contains a list of the menus and controls with
the default (factory) settings that are recalled when you push the DEFAULT SETUP
front-panel button.
◆ “AppendixC: Daily Maintain and Cleaning” chapter describes how to take care of
the oscilloscope.
                                         VI
Catalogue
Chapter 1 Accidence................................................................................... 3
                                                                1
                 2.13.3 Print ..........................................................................................................84
                 2.13.4 Self Calibration .........................................................................................87
                 2.13.5 Self Test ....................................................................................................88
                 2.13.6 Update Firmware ......................................................................................90
                 2.13.7 Pass/Fail....................................................................................................91
                 2.13.8 Waveform Record .....................................................................................95
                 2.13.9 Recorder....................................................................................................98
                 2.14 Online Help Function.................................................................................101
Index....................................................................................................... 125
                                                                 2
      Chapter 1 Accidence
ADS1000 Series Digital Storage Oscilloscope is minitype and portable bench type
◆Probe compensation
◆Self Calibration
It is important for you to understand the DSO’s front panel before operating it.
The following contents are the brief introduction for the front panel function, which
is useful to be familiar with the operation of the ADS1000 series Digital Storage
convenience users to operate them, the panel contains knobs and buttons. There is a
list of five ashen buttons as menu operational buttons on the right of display screen.
You can set different options of the current menu in virtue of them. Other buttons are
function buttons, you can enter different function menus or obtain given function
                                           3
                                                             Common Function
                                       Menu Universal        Button
                                       button Knob
                                                                                   Run
                                                                                   Control
                                                                                  AUTO Button
                                                                                   Horizontal
Option Button                                                                      Control
                                                                                   Trigger Control
                                                                                   Vertical
                                                                                   Control
                                                                                   Probe
                                                                                   component
                                      4
1. Trigger status
     Armed:The oscilloscope is acquiring pre-trigger data. All triggers are ignored
       in this state.
     Ready: All pre-trigger data has been acquired and the oscilloscope is ready to
     accept a trigger.
     Trig’d: The oscilloscope has seen a trigger and is acquiring the posttrigger data.
     Stop: The oscilloscope has stopped acquiring waveform data.
     Auto: The oscilloscope is in auto mode and is acquiring waveforms in the absence
of triggers.
scan mode.
7. On-screen markers show the ground reference points of the displayed waveforms.
                                         5
1.2 Function checking
When you check whether or not the oscilloscope could work smoothly, please operate
as following:
Press “DEFAULT SETUP” to show the result of the self check. The probe default
attenuation is 1X.
Picture 1- 3
2. Set the switch to 1X on the probe and connect the probe to channel 1 on the
oscilloscope. To do this, align the slot in the probe connector with the key on the CH
1 BNC, push to connect, and twist to the right to lock the probe in place. Connect the
Picture 1-4
3.Press “AUTO” to show the 1KHz frequency and about 3V peak-peak square wave
in couple seconds
                                           6
                              Picture 1-5
4. Press “CH1” two times to cancel the channel 1, Press“CH2” to change screen
                                  7
1.3 Probe
1.3.1 Probe Safety
       A guard around the probe body provides a finger barrier for protection from electric
shock.
Picture 1-6
Connect the probe to the oscilloscope and connect the ground terminal to ground
            Note: ● To avoid electric shock when using the probe, keep fingers behind the guard
                        on the probe body.
                    ● To avoid electric shock while using the probe, do not touch metallic
                        portions of the probe head while it is connected to a voltage source.
                        Connect the probe to the oscilloscope and connect the ground terminal to
                        ground before you take any measurements.
Probes are available with various attenuation factors which affect the vertical
scale of the signal. The Probe Check function verifies that the Probe attenuation
You can push a vertical menu button (such as the CH 1 MENU button), and
select the Probe option that matches the attenuation factor of your probe.
Be sure that the attenuation switch on the probe matches the Probe option in the
                                                 10
      oscilloscope. Switch settings are 1X and 10X.
Note. When the attenuation switch is set to 1X, the probe limits the bandwidth of the
Picture1.3-2
1. Set the Probe option attenuation in the channel menu to 10X. Set the switch to 10X
on the probe and connect the probe to channel 1 on the oscilloscope. If you use the
probe hook-tip, ensure a proper connection by firmly inserting the tip onto the
probe.
2. Attach the probe tip to the PROBE COMP~3V connector and the reference lead
to the PROBE COMP Ground connector. Display the channel and then push the
“AUTO” button.
                                                11
              Over compensated         Compensated correctly        under compensated
                                        Picture 1.3-3
The self calibration routine lets you optimize the oscilloscope signal path for
maximum measurement accuracy. You can run the routine at any time but should
To compensate the signal path, disconnect any probes or cables from the front-panel
input connectors. Then, push the UTILITY button→ “ Next Page 1/3” → “Do Self
                                           12
Chapter 2 Functions Instruction and Operation
To use your oscilloscope effectively, you need to learn about the following
oscilloscope functions:
◆ Connector
◆ Auto Setup
◆ Default Setup
◆ Universal knob
◆ Vertical System
◆ Horizontal System
◆ Trigger System
◆ Display System
◆ Utility System
◆ Storage System
                                       11
2.1 Menu and control button
                                                 Picture 2-1
      All models:
      ■ CH1、CH2: Channel 1, channel 2 menu control button.
      ■ MATH: MATH function control button.
      ■ REF: Reference waveforms control button.
      ■ HORI MENU: Horizontal control button.
      ■ TRIG MENU: Trigger control button.
      ■ SET TO 50%: Set the trigger level to midpoint of the signal amplitude.
      ■ FORCE: Use the FORCE button to complete the current waveform acquisition
      whether the oscilloscope detects a trigger or not. This is useful for Single acquisitions
      and Normal trigger mode.
      ■ SAVE/RECALL: Display the Save/Recall Menu for setups and waveforms.
      ■ ACQUIRE: Display the Acquire Menu.
      ■ MEASURE: Display the automated measurements menu.
      ■ CURSORS: Display the Cursor Menu. Vertical Position controls adjust cursor
      position while displaying the Cursor Menu and the cursors are activated. Cursors
      remain displayed (unless the “Type” option is set to “Off”) after leaving the Cursor
      Menu but are not adjustable.
      ■ DISPLAY: Display the Display Menu.
      ■ UTILITY: Display the Utility Menu.
      ■ DEFAULT SETUP: Recall the factory setup.
      ■ HELP: Enter the online help system.
      ■ AUTO: Automatically sets the oscilloscope controls to produce a usable display of
                                            12
the input signals.
■ RUN/STOP: Continuously acquires waveforms or stops the acquisition.
Note:If waveform acquisition is stopped (using the RUN/STOP or SINGLE button),
the SEC/DIV control expands or compresses the waveform.
■ SINGLE: Acquire a single waveform and then stops.
                                 13
2.2 Connector
Picture 2-2
■ EXT TRIG: Input connector for an external trigger source. Use the Trigger Menu
            Note:If you connect a voltage source to a ground terminal, you may damage the
            oscilloscope or the circuit under test. To avoid this, do not connect a voltage source to
            any ground terminals.
                                                14
2.3 Auto setup
      ADS1000 Series Digital Storage Oscilloscopes have auto setup function, which can
identify the type of waveform and adjust controls to produce a usable display of the
input signal.
        Option                     Introduction
                                   Auto set the screen and display several
        (Multi-cycle sine)         cyc signal.
                                   Set the screen and auto display single
        (Single-cycle sine)        cyc signal.
● If multiple channels have signals, channel with the lowest frequency signal.
invoked
channel 1.
Input a signal to Channel 1, press the “Auto” button (See picture 2-3):
                                           15
                          Picture 2-3
  Function               Setting
  Acquire Mode           Adjusted to Sampling
  Display Format         Y-T
  Display Type           Set to Dots for a video signal, set to Vectors for
                         an FFT spectrum; otherwise, unchanged
  Vertical Coupling      Adjusted to DC or AC according to the input
                         signal
  Bandwidth Limit        Off(full)
  V/div                  Adjusted
  VOLTS/DIV              Coarse
  adjustability
  Signal inverted        Off
  Horizontal position    Center
  S/div                  Adjusted
  Trigger type           Edge
  Trigger source         Auto detect the channel which has the input signal
                            16
2.4 Default setup
The oscilloscope is set up for normal operation when it is shipped from the factory.
This is the default setup. To recall this setup, press the DEFAULT SETUP button.
The options, buttons and controls that change settings when you press the DEFAULT
The DEFAULT SETUP button does not reset the following settings:
● Language option
● Display contrast
● Calibration data
                                           17
2.5 Universal knob
                                                Universal knob
                           Picture 2-4
      ADS1000 Series Digital Storage Oscilloscope has a special knob-the
“Universal” knob, Using this knob can adjust the holdoff time, move cursors, set
the pulse width, Set the Video Linage, adjust frequency upper limit or
frequency lower limit of the digital filter , adjust x mask and y mask in pass/fail
function and adjust the record frame or play back frame in waveform record
menu etc. You can turn the “Universal” knob to adjust the storage position of
setups, waveforms, pictures to save or recall them, and you can adjust the
                                           18
2.6 Vertical System
      Showing as follow picture, the buttons and knobs are useful in the vertical system
Volt/div knob
                                                             Vertical POSITION
                                                             knob
                                 Picture 2-5
      The vertical control, which each channel has, could be used for displaying waveform,
                                          19
    Table 2-4 CH1 & CH2 function menu 2:
■ Fine Resolution:The vertical scale readout displays the actual Volts/Div setting
while in the fine resolution setting. Changing the setting to coarse does not change
■ Wave Cancel:Press the menu button for the channel to display its vertical menu.
                                     20
          Note:● The oscilloscope vertical response rolls off slowly above its bandwidth (60
                   MHz, 100 MHz) , which depending on the model, or 20 MHz when the
                   Bandwidth Limit option is set to “On”. Therefore, the FFT spectrum can
                   show valid frequency information higher than the oscilloscope bandwidth.
                   However, the magnitude information near or above the bandwidth will not
                   be accurate.
                   ● If the channel is set to DC coupling,you can quickly measure the DC
                      component of the signal by simply nothing its distance from the ground
                      symbol.
                  ● If the channel is set to AC coupling,the DC component of the signal is
                   blocked allowing you to use greater sensitivity to display the AC
                   component of the signal.
Each channel has its own separate Menu. The items are set up separately
Take the CH1 for example, the tested signal is a sine wave signal with DC
deflection:
                                                                                               Set to AC
                                                                                               coupling
                                AC status
                                                    Picture 2-6
                                               21
                                                                          Set to DC
                                                                          coupling
             AC Status
                               Picture 2-7
      ● Press“CH1”→“Coupling”→“GND”, Set to GROUND mode. It disconnects
                                                                      Set to GND
                                                                      coupling
                               Picture 2-8
             GND Status
2. Set up the channel bandwidth limit
Take the CH1 for example, the tested signal is a pulse signal with the high
frequency surge:
● Press “CH1”→“BW Limit”→ “On”,Set the band width Limited to open state. The
                                  22
                                                                         Set bandwidth to
                                                                         20MHZ
                                  Picture 2-9
              Bandwidth Limit
              Symbol
● Press“CH1”→“BW Limit”→ “Off”, Set bandwidth Limited to close state, the High
Picture 2-10
3. Volts/Div settings
Vertical scale adjust have Coarse and fine two modes, Vertical sensitivity range is
makes the vertical scaling in a 1-2-5-step sequence from 2mv/div, 5mv/div, 10mv/div
to 10v/div.
                                     23
                                                                           Set to Coarse
Picture 2-11
● Press CH1”→ Volts/Div”→ Fine”. This setting changes the vertical to small
steps between the coarse settings. It will be helpful when you need to adjust the
Set to Fine
Picture 2-12
In order to assort the attenuation coefficient, you need to response in the channel
operation Menu. If the attenuation coefficient is 10:1, the input coefficient should be
set to 10X, so that the mistake of the Volts/div information and measure testing
should be forbidden.
Take the CH1 for example, when you use the 100:1 probe:
● Press“CH1”→“Probe” →“100X
                                     24
                                                Probe attenuation
                                                factor
       Status    changes
       with the probe
       attenuation
                                 Picture 2-13
5. To invert a waveform
● Press“CH1”→“Invert”→“On”:
Picture 2-14
● Press“CH1”→“Invert”→“Off”:
                                    25
                               Picture 2-15
6. Digital Filter
Press “CH1”→“Next Page”→ “Filter”, display the digital filter menu. Select “Filter
Type”, then select “Upper Limit” or “Lower Limit” and turn the “Universal” knob to
adjust them.
● Press “CH1”→“Next Page”→ “Filter” →“Off”. Turn off the Digital Filter function.
Picture 2-16
● Press “CH1”→ “Next Page”→ “Filter” → “On”. Turn on the Digital Filter function.
                                       26
                                      Picture 2-17
1. Use the Vertical “POSITION” knobs to move the channel waveforms up or down
on the screen. This button’s resolution is variety as per the vertical scale.
2. When you adjust the vertical position of channels waveforms, the vertical
position information will display on the left bottom of screen. For example
“Volts Pos=24.6mV”.
3. Press the vertical “POSITION” knob to set the vertical position to zero.
■ “Volts/div” knob
1. Use the “Volts/div” knobs to control how the oscilloscope amplifies or attenuates
the source signal of channel waveforms. When you turn the “volts/div” knob, the
oscilloscope increases or decreases the vertical size of the waveform on the screen
2. When you press the “Volt/div” Knob, you can switch “Volt/div” option between
“Coarse” and “Fine”. The vertical scale is made sure by the 1-2-5 step in the Coarse.
Increase in the clockwise, reduce in the anticlockwise. In the fine mode, the knob
changes the Volts/Div scale in small steps between the coarse settings. Increase in the
                                            27
2.6.3 Math Functions
Math shows the results after +,-,*and FFT operation of the CH1 and CH2.
Press the MATH button to display the waveform math operations. Press the MATH
MATH Waveform
Picture 2-18
                                           28
 1. FFT Spectrum Analyzer
The FFT process mathematically converts a time-domain signal into its frequency
components. You can use the Math FFT mode to view the following types of signals:
● Analyze vibration
To use the Math FFT mode, you need to perform the following tasks:
● Turn the vertical “POSITION” knob to move the YT waveform to the center
                                     29
vertically (zero divisions).
● Turn the horizontal “POSITION” knob to position the part of the YT waveform
that you want to analyze in the center eight divisions of the screen.
The oscilloscope calculates the FFT spectrum using the center 1024 points of the
time-domain waveform.
● Turn the “Volts/div” knob to ensure that the entire waveform remains on the
screen.
● Turn the “S/div” knob to provide the resolution you want in the FFT spectrum.
channel.
4. According to Nyquist law, turn the “S/div” knob to adjust the sampling rate(This
parameter is displayed behind the time base parameter) is at least double than input
signal frequency.
Press the MATH button to display the Math Menu. Use the options to select the
Source channel, Window algorithm, and FFT Zoom factor. You can display only one
FFT spectrum at a time. You can select “Full screen” or “Split” in “ Display” option
to display FFT waveform on full screen or display channel waveform and it’s FFT
                                     30
                                  Picture 2-19
                                           Frequency at the
                                           center graticule
                                           line
                                                                          FFT window
                                                                          type
waveform repeats forever. With an integral number of cycles, the YT waveform starts
and ends at the me amplitude and there are no discontinuities in the signal shape A
non-integral number of cycles in the YT waveform causes the signal start and end
points to be at different amplitudes. The transitions between the start and end points
                                     31
cause discontinuities in the signal that introduce high-frequency transients.
According to the tested options and source speciality, make sure the window you
need to use.
You can magnify and use cursors to take measurements on the FFT spectrum. The
option button to select “1X”, “2X”, “5X” or “10X”. Moreover, you also can turn the
You can take two measurements on FFT spectrums: magnitude (in dB) and frequency
(in Hz). Magnitude is referenced to 0 dB, where 0 dB equals 1 VRMS. You can use
the cursors to take measurements at any zoom factor. (Refer to cursor measure2.11.2)
                                       32
frequency.
6) Turn the “S/div” knob to adjust sampling rate (at least duble bigger than freqency
of input signal) .
7) If FFT display on full screen, press CH1 button again to remove channel
waveform display.
12) Press the “CurA” option button, turn the “Universal” knob to move Cursor A to
13) Press the “CurB”option button, turn the “Universal” knob to move Cursor B to
14) The amplitude (△T) displays on the top right of the screen.
                                    33
                               Picture 2-21
5) Press the “Cur1” option button, turn the “Universal” button to move Cursor 1 to
6) The value of Cur1 displaying on the top right of the screen is FFT highest
Picture 2-22
                                     34
2.6.4 Using Ref
                                         Picture 2-23
      Operation step:
1. Press the “REF” menu button to display the “Reference waveform menu”.
3. Turn the vertical “POSITION” knob and “Volt/div” knob to adjust the vertical
4. Press the top third option button to select “REFA” or “REFB” as storage position.
6. Press the bottom option button to select “REFA On” or “REFB On” to recall
                                           35
the reference waveform.
                                  36
2.7 Horizontal System
As follow Picture, there are one button and two knobs in the HORIZONTAL area.
S/div knob
                                                            Horizontal
                                                            “POSITION”
                                                            knob
Picture 2- 25
                                           37
2.7.1 Horizontal control knob
           You can use the horizontal controls to change the horizontal scale and position
of waveforms. The horizontal position readout shows the time represented by the
center of the screen, using the time of the trigger as zero. Changing the horizontal
scale causes the waveform to expand or contract around the screen center.
1. Adjust the horizontal position of all channels and math waveforms (the position of
the trigger relative to the center of the screen). The resolution of this control varies
2. When you press the horizontal “POSITON” Knob, you can set the horizontal
position to zero.
■“S/div” knob
1. Using to change the horizontal time scale to magnify or compress the waveform. If
waveform acquisition is stopped (using the RUN/STOP or SINGLE button), turn the
2. Select the horizontal time/div (scale factor) for the main or the window time base.
When Window Zone is enabled, it changes the width of the window zone by
When the SEC/DIV control is set to100 ms/div or slower and the trigger mode
is set to Auto, the oscilloscope enters the scan acquisition mode. In this mode, the
waveform display updates from left to right. There is no trigger or horizontal position
2.7.2 Delayed
           Use the Window Zone option to define a segment of a waveform to see more
detail. The Window time base setting cannot be set slower than the Main time base
setting.
                                            38
    You can turn the Horizontal Position and SEC/DIV controls to enlarge or minish
If you want to see a section of the waveform in detail, follow these steps:
(1) Press the “HORI MENU” button to enter the “Horizontal menu”.
(2) Turn the “S/div” knob to change the main timebase scale.
Turn the “Horizontal Position” knob (adjust window’s position) to select the
window that your need and expanded window waveform display on the below half
Picture 2-28
                                    39
2.8 Trigger System
        The trigger determines when the oscilloscope starts to acquire data and display a
Here are three buttons and one Knob in the Trigger area. Showing as Picture.2-29:
LEVEL knob
                                    Picture 2-29
      ■ “TRIG MENU” Button: Press the “TRIG MENU” Button to display the “Trigger
      Menu”.
■“LEVEL” Knob: The LEVEL knob is to set the corresponding signal voltage of
trigger point in order to sample. Press the “LEVEL” knob can set trigger level to
zero.
■“SET TO 50%” Button: Use the “SET TO 50%” button to stabilize a waveform
quickly. The oscilloscope can set the Trigger Level to be about halfway between the
minimum and maximum voltage levels automatically. This is useful when you
connect a signal to the EXT TRIG BNC and set the trigger source to Ext or Ext/5.
■“FORCE” Button: Use the FORCE button to complete the current waveform
acquisition whether the oscilloscope detects a trigger or not. This is useful for
■ Pretrigger/Delayed trigger: The data before and after trigger the trigger position
is typically set at the horizontal center of the screen, in the full-screen display the
6div data of pretrigger and delayed trigger can be surveyed. More data of pretrigger
                                            40
       The feature is very useful because you can see the events that led up the trigger point
everything to the right of the trigger point is called posttrigger information the
as a trigger. The source can be any signal connected to a channel BNC, to the EXT
TRIG BNC or the AC power line (available only with Edge triggers).
Alternative.
■ Edge Trigger
Use Edge triggering to trigger on the edge of the oscilloscope input signal at the
trigger threshold.
                                            41
     Mode        Auto           Use this mode to let the acquisition free-run in the
                                absence of a valid trigger; This mode allows an
                                untriggered, scanning waveform at 100 ms/div or
                                slower time base settings.
                 Normal         Use this mode when you want to see only valid
                                triggered waveforms; when you use this mode, the
                                oscilloscope does not display a waveform until after
                                the first trigger.
                 Single         when you want the oscilloscope to acquire a single
                                waveform, press the “SINGLE ”button.
       Set up                    Enter the “Trigger Setup Menu”(See table 2-14).
                              Picture 2-30
Operate Instruction:
1. Set up Type
                                42
    2) Press the “Type” option button to select “Edge”.
2. Set up Source
According to input signal, press the “Source” option button to select “CH1”,
3. Set up Slope
“Single”.
Auto: The waveform refresh at a high speed whether the trigger condition is
satisfied or not.
Normal: The waveform refresh when the trigger condition is satisfied and waits
for next trigger event occurring when the trigger condition is not satisfied.
a. Press the “Set up” button to enter the “Trigger Setup Menu”.
b. Press the “Coupling” option button to select “DC”, “AC”, “HF Reject” or
“LF Reject”.
                                       43
   When                                                    Select how to compare the
               (Positive pulse width less than pulse       trigger pulse relative to the
               width setting)                              value selected in the Set
                                                           Pulse Width option.
               (Positive pulse width larger than pulse
               width setting)
                                   Picture 2-31
 Table 2-16 Pulse Trigger function Menu 2:
Next Page Page 2/2 Press this button to return the first page.
                                        44
                                Picture 2-32
Operate Instruction:
1. Set up Type
2. Set up condition
“ ”or“ ”。
                                        45
 y                                         of the sync pulse.
Picture 2-33
                                46
                                                 Trigger menu”.
Picture 2-34
Operate Instruction:
1. Set up Type
2. Set up Polarity
3. Set up synchronization
1) Press the “Sync” option button to select “All Lines”, “Line Num”, “Odd Field”,
“Even Field”.
2) If you select “Line Num”, you can turn the “Universal” knob to set the appointed
line number.
4. Set up Standard
of the oscilloscope.
                                     47
Option           Setting          Instruction
Type             Slope            Trigger on positive slope of negative
                                  slope according to setup time of the
                                  oscilloscope.
Source           CH1              Select trigger source.
                 CH2
                 EXT
                 EXT/5
When                              Select trigger condition.
Picture 2-35
Mode
                 Auto             Use this mode to let the acquisition
                                  free-run in the absence of a valid
                                  trigger; This mode allows an
                                  48
                                     settings.
                                     Use this mode when you want to see
                                     only valid triggered waveforms;
                   Normal            when you use this mode, the
                                     oscilloscope does not display a
                                     waveform until after the first trigger.
                                     when you want the oscilloscope to
                    Single           acquire a single waveform, press the
                                     “SINGLE ”button.
                                     Enter the “Trigger setup menu”(See
 Set up
                                     table 2-14).
 Next Page         Page 2/2          Return the first page of slope trigger.
Picture 2-36
Operate Instruction:
“ ”, “ ”,“ ”或 “ ”.
7. Press the “Time” button, turn the “Universal” knob to adjust slope time.
8. Press the “Next Page Page 1/2” option button to enter the second page of “Slope
trigger menu”.
                                     49
9. Press the “Vertical” option button to select trigger level that can be adjusted.
■ Alternative trigger
The trigger signal come from two vertical channels when you use alternative trigger.
In this mode ,you can observe two irrelative signals at the same time. You can select
different trigger types for two vertical signals, and selected types cover edge, pulse,
video and slope trigger. Trigger information of two channel signals display on the
Picture 2-37
                                      50
    Type          Alternative    The trigger signal come from two vertical
                                 channels when you use alternative trigger. In this
                                 mode, you can observe two irrelative signals at
                                 the same time.
                  CH1            Set trigger type information for CH1 signal
    Source
                  CH2            Set trigger type information for CH2 signal
             Setting              Instruction
 Option
 Type        Alternative          The trigger signal come from two vertical channels
                                  when you use alternative trigger. In this mode ,you
                                  can observe two irrelative signals at the same time.
                                    51
    Sync        Line Num
                All lines     Select appropriate video sync.
                Odd field
                Even Field
    Standard    NTSC          Select the video standard for sync and line number
                Pal/Secam     count.
    Next
                Page 1/2      Enter the second page of Alternative trigger menu.
    Page
    Next
                Page 1/2      Enter the second page of the alternative trigger.
    Page
Operate Instruction:
                                   52
         2. Press the AUTO button.
2.8.3 Coupling
       Use the “Coupling” to make sure the signal that passes the trigger circuit. It is useful
If you use the trigger coupling, you should press the “TRIGGE” button and then
2.8.4 Position
       The horizontal position control establishes the time between the trigger position and
the screen center. You can adjust the horizontal “POSITION” knob control to view
waveform data before the trigger, after the trigger, or some of each. When you
change the horizontal position of a waveform, you are changing the time between the
trigger and the center of the display actually. (This appears to move the waveform to
The Slope and Level controls help to define the trigger. The Slope option (Edge
trigger type only) determines whether the oscilloscope finds the trigger point on the
                                                 53
The TRIGGER LEVEL knob controls where on the edge the trigger point occurs.
Picture 2-38
   Note: ● Press the SINGLE button when you want the oscilloscope to
               acquire a single waveform.
         ●. Trigger coupling affects only the signal passed to the trigger
               system. It does not affect the bandwidth or coupling of the
               signal displayed on the screen.
         ● Normal Polarity Sync triggers always occur on negative-going
             orizontal sync pulses. If the video waveform has positive-going
             horizontal sync pulses, use the Inverted Polarity selection.
Trigger position
Trigger level
Holdoff time
Picture 2-39
                                                                                 54
2. Press the “Type” option button to select trigger type.
3. Press the “Set up” option button to enter the “Trigger setup menu”.
4. Press the “Holdoff” option button,turn the “Universal” knob to change the
                                     55
2.9 Acquiring Signals system
       Showing as the follow picture,the “ACQUIRE” button for Acquiring Signals system
     is at the menu.
                                       Acquire button
Picture 2-40
When you acquire a signal, the oscilloscope converts it into a digital form and
displays a waveform. The acquisition mode defines how the signal is digitized and
the time base setting affects the time span and level of detail in the acquisition.
■ Sampling: In this acquisition mode, the oscilloscope samples the signal in evenly
spaced intervals to construct the waveform. This mode accurately represents signals
                                            56
Disadvantage: This mode does not acquire rapid variations in the signal that may
occur between samples. This can result in aliasing may cause narrow pulses to be
missed. In these cases, you should use the Peak Detect mode to acquire data.
signal Finds highest and lowest record points over many acquisitions.
Advantage: In this way, the oscilloscope can acquire and display narrow pulses,
                                    57
                      Picture 2-43 Average mode
■   Equivalent Time Sampling:
■ Real Time Sampling: ADS1000 has the highest Real-time sampling rate
up to 1GSa/s.
■ “RUN/STOP” Button: Press the RUN/STOP button when you want the
oscilloscope to acquire waveforms continuously. Press the button again to stop the
acquisition.
■ “SINGLE” Button: Press the SINGLE button to acquire a single waveform. Each
time you push the SINGLE button, the oscilloscope begins to acquire another
waveform. After the oscilloscope detects a trigger it completes the acquisition and
stops.
When you push the RUN/STOP or SINGLE buttons to start an acquisition, the
1). Acquire enough data to fill the portion of the waveform record to the left of the
2). Continue to acquire data while waiting for the trigger condition to occur.
                                       58
■ Base time: The oscilloscope digitizes waveforms by acquiring the value of an
input signal at discrete points. The time base allows you to control how often the
To adjust the time base to a horizontal scale that suits your purpose, use the S/div
knob.
 Aliasing occurs when the oscilloscope does not sample the signal fast enough to
 construct an accurate waveform record. When this happens, the oscilloscope
 displays a waveform with a frequency lower than the actual input waveform, or
 triggers and displays an unstable waveform.
Sampled points
Picture 2-44
Operate Introduction:
You can press the “Acquisition” option button or turn the “Universal” knob to
2. Set up Averages
When you select “Average” format, you can press the “Averages” option button
                                     59
        “On” is sin interpolation
                              Picture 2-44-1
       “Off”is linear interpolation.
Picture 2-44-2
 Press the “Mode” option button to select “Real Time” or “Equ Time”.
5. Set up Sampling rate
 In “Real time” mode, turn the “S/div” knob to display the real time sampling
  rate at corresponding time base scale; In “Equ time” mode, turn the “S/div”
  knob to display the equ time sampling rate at corresponding time base scale”.
                                60
2.10 Display System
                                     Display System
                                        Picture 2-45
Picture 2-46
                                               61
Table 2-30 Display system function menu 2:
                             Picture 2-47
   Table 2-31 Display system function menu3:
     Optio          Setting            Introduction
     n
     Skin           Classical          Set up screen style.
                    Modern
                    Tradition
                    Succinct
     Next           Page 3/3           Press this button to return the
     Page                              first page.
                                 62
                                  Picture 2-48
Operate Introduction:
2. Set up Persist
Press the “Persist” option button to select “Off”, “1 Sec”, “2 Sec”, “5Sec”
or “Infinite”. You can use this option to observe some especial waveforms.
Picture 2-49
3. Set up Intensity
1) Press the “Intensity” option button and turn the “ Universal” knob to
4. Set up Brightness
1) Press the “Brightness” option button and turn the “Universal” knob to
                                    63
               adjust grid brightness.
1) Press the “Next Page” option button to enter second display menu.
5. Set up Screen
Press the “Screen” option button to select “Normal” or “Inverted” to set the
6. Set up Grid
Press the “Menu Display” option button to select “2 sec”, “5sec”, “10sec”,
8. Set Skin
Press the “skin” option button or turn the “Universal” knob to select
Lissajous patterns. The format plots the voltage on channel 1 against the voltage on
channel 2, where channel is the horizontal axis and channel 2 is the vertical axis.
The oscilloscope uses the untriggered Sample acquisition mode and displays data as
dots.
adjusted
                                                64
Operation steps :
and position.
and position.
200MSa/s).
● Cursor
● Trigger Control
● Scan Display
                                      65
2.11 Measure System
The Oscilloscope displays the voltage in relation to time and test the wave form
might look at a waveform amplitude and determine that it is a little more than 100
mV. You can take simple measurements by counting the major and minor graticule
divisions involved and multiplying by the scale factor. For example, if you counted
five major vertical graticule divisions between the minimum and maximum values of
a waveform and knew you had a scale factor of 100mV/div, then you could easily
Cursors Button
Picture 2-50
The cursor measurement has three modes: Manual, Track and Auto Measure.
                                           66
 Manual Mode
cursors to measure voltage or time.You can move the cursor by turning the
“ Univeral ” knob. Before using cursors, you should make sure that you havre set the
■ Voltage Cursor: Voltage cursors appear as horizontal lines on the display and
■ Time Cursor: Time cursors appear as vertical lines on the display and measure the
horizontal parameters.
■ Cursor Moving: Use the “universal” knob to move cursor1 and cursor2 .They
could be moved when the corresponding cursor option are selected, and cursor value
will display on the bottom left and top right of the screen when you move the cursor.
4. Press the “Source” option button to slect “CH1”, “CH2”, “MATH”, “REFA” or
                                    67
     If the measurement type is set to “ Voltage”, the values are:
                                 Picture 2-51
     Track mode
on the waveform automatically. You could adjust cursor’s horizontal postion on the
waveform by turning the “ Universal ” knob. The oscilloscope displays the values on
                                     68
the top right of the screen.
3. Press the “Cursor A” option button to select the input signal channel.
4. Press the “Cursor B” option button to select the input signal channel.
5. Select “Cur A”, turn the “Universal” knob to move Cursor A horizontally.
6. Select “Cur B”, turn the “Universal” knob to move Cursor B horizontally.
A→T: The horizontal postion of Cursor A ( Time cursor cented around the
midpoint of screen).
B→T: The horizontal postion of Cursor B ( Time cursor cented around the
midpoint of screen).
△T: Horizontal space between Cursor A and Cursor B ( Time value between
two cursors).
△V: Vertical space between Cursor A and Cursor B (Voltage value between two
cursors).
                                    69
                            Picture 2-52
 Auto mode
 Table 2-34 Auto mode function menu:
This mode will take effect with automatic measurements. The instruments will
3. Press the “ MEASURE” button to enter “Auto cursor measure mode menu”
Picture 2-53
                                70
2.11.3 Auto Measurement
                      Auto Measurement
                           button
                                 Picture 2-54
When you take automatic measurements, the oscilloscope does all the
calculating for you. Because the measurements use the waveform record points, they
There are three auto measurement types: Voltage Measure, Time Measure and Delay
                                          71
                            Picture 2-55
Table 2-36 Auto measure function 2-Voltage measure menu:
                               72
 Option     Setting                      Instruction
 Source     CH1、CH2、MATH、                Select input signal source for
            REFA、REFB                    Delay measure.
 Type       Phase、FRR、FRF、FFR、FFF、       Press the “Type” button or turn the
            LRR、LRF、LFR、LFF              “Universal” knob to select Delay
                                         measure parameter.
                    、       、          、
                                         Display the corresponding icon
                    、       、          、
                                         and measure value of your
                                         selected Delay measure parameter.
                    、       、
 Return                                    Return to the first page of auto
                                           measurement menu.
                             73
                              Defined as (Vmax-Vhig)/Vamp after the
     ROVShoot                  waveform rising.
                              Defined as (Vmin-Vlow)/Vamp after the
     FOVShoot                  waveform falling.
                              Defined as (Vmin-Vlow)/Vamp before the
     RPREshoot                 waveform rising.
                              Defined as (Vmax-Vhig)/Vamp before the
     FPREshoot                 waveform falling.
                             Rise Time Measures the time between 10% and
     Rise Time               90% of the first rising edge of the waveform.
                             Fall Time Measures the time between 90% and
     Fall Time               10% of the first falling edge of the waveform.
                             The duration of a burst.
     BWid                    Measured over the entire waveform.
                             + Width Measures the time between the first rising
     + Wid                   edge and the next falling edge at the waveform
                             50% level.
                             -Width Measures the time between the first falling
     - Wid                   edge and the next rising edge at the waveform
                             50% level.
                             Measures the first cycle waveform.
     + Duty                  Positive Duty Cycle is the ratio between positive
                             pulse width and period.
                             Measures the first cycle waveform.
     -Duty                   Negative Duty Cycle is the ratio between negative
                             pulse width and period.
                             The amount one waveform leads or lags another in
     Phase                   time. Expressed in degrees, where 360 degrees
                             comprise one waveform cycle.
                             The time between the first rising edge of source 1
      FRR                    and the first rising edge of source 2.
                             The time between the first rising edge of source 1
      FRF                    and the first falling edge of source 2.
                             The time between the first falling edge of source 1
       FFR                   and the first rising edge of source 2.
                             The time between the first falling edge of source 1
       FFF                   and the first falling edge of source 2.
                             The time between the first rising edge of source 1
      LRR                    and the last rising edge of source 2.
                             The time between the first rising edge of source 1
      LRF                    and the last falling edge of source 2.
                             The time between the first falling edge of source 1
      LFR                    and the last rising edge of source 2.
                             The time between the first falling edge of source 1
       LFF                   and the last falling edge of source 2.
                              74
 menu”.
 3). Select measure type. If you press the “Voltage” option button,
 “Voltage measurement” menu will display on the screen.
 4). Press the “Source” option button to select “CH1” or “CH2”
 according to input signal channel.
 5). Press the “Type” option button to select the parameter type that
 you want to measure. The corresponding icon and value will display
 below the measure parameter.
                              Picture 2-56
 6). Press the “Return” option button to return the home page of “Auto
 Measurement” menu. The selected parameter and the corresponding
 value will display on the top first position of the home page.
 You can display the other parameters and its’ value on the
 corresponding position using the same way, and the screen can
 display five parameters at one time.
If you want to measure time parameters using all measure function,
 please follow next steps:
 1). Press the “MEASURE” button to enter the “Auto Measure menu”.
 2). Press the top option button to enter the second page of “Auto
 Measure menu”.
 3). Press “All Mea” option button to enter the “All Measure menu”.
 4). Press “Source” option button to select input signal channel.
 5). Press the “Time” option button to select “On”. Now all of time
 parameters value will display on the screen at the same time.(See
 Picture 2-57)
                             75
Picture 2-57
    76
2.12 Storage System
     Showing as follow picture, The SAVE/RECALL is the Storage System Functional
Button.
Picture 2-58
ADS1000 series can save 2 groups reference waveforms, 20 groups setups and 20
interface in the front panel of the ADS1000 series oscilloscope and you can save setup
data, waveform data, waveform interface image , CSV file to a USB flash drive
furthest at a time. Setup data and waveform data can be recalled to the current
oscilloscope or the other same model oscilloscopes. Picture data and CSV file can’t be
recalled to the oscilloscope, but it can be opened on the correlative computer software.
                                           77
SAVE/RECALL SETUP
Save/recall interface
Picture 2-61
Picture 2-62
Picture 2-63
                                        78
and CSV stores to USB flash drive all have this function.Users can modify the name
When you recall the setup, the oscilloscope will be in the mode from which the setup
was saved.
The oscilloscope saves the current setup if you wait three seconds after the last
change before you power off the oscilloscope. The oscilloscope recalls this setup the
Picture 2-64
Save setups to the oscilloscope’s internal memorizer or recall saved setups from
For example: Save setup that the oscilloscope displays waveform as the form of
                                        79
2). Press the “Type” option button to select “setups”.
If the display type has been changed, you want to recall this setup.
4) Press the “Setup” option button or turn the “Universal” knob to select “No.1”.
                                     80
                                             Picture 2-65
For example: Save setup that the oscilloscope displays waveform as the form
2) Insert USB flash drive to USB host port of the oscilloscope and wait that the
4) Press the “Save” option button then you’ll go into the Save/Recall interface.
10) Press “the next page ” option button and press the “Rename” option button to
11) Press the “Confirm”option button and you save the set to the USB drive.
Recall setup data from USB flash drive, follow these steps:
3) Insert USB flash drive to USB host port of the oscilloscope and wait that the
5) Press the “Save” option button then you’ll go into the Save/Recall interface .
6) Choose the file you want then press the “Load” option button (about five seconds,
there is a prompting message that “Read data success”), now setup data have been
■ RECALL FACTORY
                                     81
     You can use this option to recall the factory setup.
Picture 2-66
SAVE/RECALL WAVEFORM
display any waveform that you want to save. The oscilloscopes can store
                                             82
                                      Picture 2-67
steps:
1). Input a sine signal to channel 1 and press the “Auto” button.
5). Press the “waveform” option button or turn the “Universal” knob to select
“No.1”.
6). Turn the “Volts/div” knob or “S/div” knob to adjust the waveform that you want
to need.
Recall the saved waveform from the oscilloscope’s internal memorizer, follow
these steps:
4). Press the “waveform” option button or turn the “Universal” knob to select
“No.1”.
                                      83
■ Save waveforms to USB flash drive
Picture 2-68
                                    84
6). Choose the file you want then press the “Load” option button (about five seconds,
there is a prompting message that “Recall data success” displaying on the screen),
now waveform data have been recalled from USB flash drive.
■   Save Picture
    Waveform interface image can be saved to USB flash drive, but they can’t be
recalled. You can open them on correlative computer software.
Table 2-46 Picture storage function menu:
Picture 2-69
Save waveform interface image to USB flash drive, follow these steps:
4). Insert USB flash drive to USB host port of the oscilloscope and wait that the
                                     85
          oscilloscope has initialized USB flash drive (about ten seconds).
5). Press the “Print Key” option button to select “Save Picture”.
6). Press the “Save” option button then go into the Save/Recall interface”.
7). Press the “Save” buton then you’ll go into the Save/Recall interface
8). Create the picture name then press the “Confirm” button(about five seconds, there
is a prompting message that “Save data success” displaying on the screen), now
Reference waveforms are not adjustable, but the oscilloscope does display the
Save/Recall CSV
                                            86
                                      Picture 2-70
3). Insert USB flash drive to USB host port of the oscilloscope and wait that the
4). Press the “Data Depth” option button to select “Displayed” or “Maximum”.
5). Press the “Para Save” option button to select “On” or “Off”.
6). Press the “Save” option button then you’ll go into the Save/Recall interface
7). Create a file name then press the “Confirm” button (About five seconds, there is a
prompting message that “Save data success” displaying on the screen), now CSV file
                                     87
2.13 Utility System
Picture 2-66
                                         80
                            Picture 2-67
                              81
                        Picture 2-68
Picture 2-69
                              82
          Table 2-50 Functional Menu of the Utility System 4:
Picture 2-69-1
Selecting “System Status” from the Utility Menu to display a series of information
Picture 2-70
                                              83
          Table 2-52 System status instruction:
            Option              Introduction
            Starup Times        List the open times.
            Software version    list the software version.
            Hardware Version    List the hardware version.
            Product      type   List the product type.
            Serial No.          List the product serial number.
2.13.2 Language
Picture 2-71
2.13.3 Print
ADS1000 Serials support PictBridge compatible printers. You can connect the back
USB port of the oscilloscope to the USB port of the PictBridge compatible printer
                                          84
   Table 2-53 Print setup option function menu 1:
Picture 2-72
                                     85
 Image Size       Default, 2.5x3.25 in, L (3.5x5 in), 4x6     Displays Settings
                  in, 2L (5x7 in), 8x10 in, 4L (7 x 10 in),   available on your
                  E, Card, Hagaki card, 6 x 8 cm, 7x10        PictBridge
                  cm, 9x13 cm, 10x15 cm, 13x18 cm,            compatible
                  15x21 cm, 18x24 cm, A4, Letter              printer.
 Paper Type       Default, plain, Photo, Fast Photo
 Print Quality    Default, Normal, Draft, Fine
 ID Print         Default, On, Off
 Next Page        Page 2/2                                    Press this button
                                                              to enter first page
                                                              of print setup
                                                              menu.
   Note: 1.The Printer may override your selection for best fit.
   2. If your selection is not supported by the printer, the oscilloscope uses the
   Default setting.
   3. The oscilloscope was designed to print to any PictBridge compatible
   printer. Refer to product documentation for your printer to determine if the
   printer is PictBridge compatible.
Picture 2-73
1) Insert one end of a USB cable into the USB Device port on the oscilloscope.
2) Insert the other end of the cable into the PictBridge port on a PictBridge
compatible printer.
                                       86
       2. Print a Screen image
1) Power on the oscilloscope and the printer (The printer recognizes the oscilloscope
4) Press the “Print Setup” option button to display “Print setup” menu.
5) Set up the print setup option according to your need. The oscilloscope queries the
printer, and only displays options and values that the printer supports.
If you are not sure which setting to choose, select “Default” for each option.
The oscilloscope takes a few seconds to capture the screen image. The settings of
your printer and print speed determine how long it takes to print the data. Additional
        Note:
        1. Before connecting the oscilloscope to the printer, you must be sure that
        “ Back USB” option having been setted to “Printer”.
        2. If you want to use oscilloscope connecting to your Computer, you must be
        sure that “ Back USB” option having been setted to “Computer”
        3. You can use the oscilloscope while the printer prints.
        4. If printing fails, check that the USB cable is connected to the PictBridge
        port on the printer. If the connection have no problem, but the printe failed
        again, you can turn on and off the printer or oscilloscope again, before you
        turn on and off the printer or oscilloscope again , you must first disconnect
        the connection between USB cable and USB Host port of              the Printer or
        USB Deviece port of the oscilloscope, then try again.
Then press the “Utility” button to choose the “Do self cal” to show the self
                                            87
       prompting on the screen.
Picture 2-74
Table2-55
               Option                 Introduction
               Screen Test            Run Screen Test Program.
               Keyboard Test          Run Keyboard Test Program.
               LED Test               Run LED Test Program.
Operation Steps:
1. Screen Test:
Select “Screen Test” to enter the screen test interface. The clew words “Press
                                             88
                                     Picture 2-75
2. Keyboard Test
Select “keyboard Test” to enter the key test interface, the on-screen lathy rectangle
shapes represent the front panel keys. The shapes with two arrows beside them
represent the front panel knobs. The squares represent the knob presses for scale
knobs. Test all keys and knobs and you should also verify that all the backlit buttons
illuminate correctly.
Note:
● When you operate, the screen would display the white (Color LCD) and black
● The tested button or knobs corresponding area would display the green (Color
● At the bottom of the screen display “Press‘ RUN/STOP’ Key Three Times to
exit” information prompt to show that press “RUN/STOP” three times for quite the
test .
                                     89
                                      Picture 2-76
3. LED test
● Select “LED Test” to enter the lighten interface, the on-screen lathy rectangle
shapes represent the front panel keys; the lathy rectangle shapes represent the front
panel knobs. The squares represent the knob presses for scale knobs. The clew words
“Press ‘SINGLE’ Key to continue , Press ‘RUN/STOP’ Key to exit” is displayed, you
could press the “Single” continuously for test, when buttons are lighted ,the
LCD ).
Picture 2-77
The software of ADS1000 series oscilloscope can be updated directly via USB Flash drive,
                                              90
     1. Invert USB Flash Drive with firmware procedure to USB Host interface on the front
3. Press “Next Page” option button to enter the third page of “Utility menu”.
5. Press the “SINGLE” button to start updating the software according to prompt on the
screen.
Turn off the oscilloscope and turn on it again, the software has been updated. The
2.13.7 Pass/Fail
The Pass/Fail function can monitor changes of signals and output pass or fail
signals by judging the input signal is within the predefined mask or not.
                                           91
                        Picture 2-78
Picture 2-79
                                92
  ydiv                      <0.04div-4.00div>
Picture 2-80
Picture 2-81
                                 93
   RUN Pass/Fail test, please follow next page:
3) Press the “Next Page Page2/3” option button to enter the third page of “Utility
menu”
4) Press the “Pass/Fail” option button to enter the “Pass/Fail” function menu.
7) Press the “Next Page Page1/2”option button to enter the second page of the
“Pass/Fail menu”.
8) Press the “Mask Setting” option button to enter the first page of “Mask menu”.
9) Press the “X Mask” button, turn the “Universal” knob to adjust the horizontal
clearance.
10) Press the “Y Mask” button, turn the “Universal” knob to adjust the vertical
clearance.
11) Press the “Create Mask” button to create mask and you also can enter the next
12) Enter the second page of “Pass/Fail function menu”, press the “Output” option
13) Enter the first page of “Pass/Fail function menu”, press the “Operate” option
                                      94
                                   Picture 2-82
       Waveform recorder can record input waveform from CH1 and CH2, with a
     maximum record length of 2500 frames. This record behavior can also be activated by
     the pass/fail test output, which makes this function especially useful to capture
     abnormal signals in long term without keeping an eye watching it.
   Waveform recorder: Record the waveforms with specified interval.
       Table 2-60 Waveform recorder function menu:
                                          95
                            Picture 2-89
Play Back: Play back current record waveforms or saved record waveforms.
                                    96
                                           Picture 2-90
Picture 2-91
                                    97
         6. Select the “Start Frame” option, turn the “Universal” knob to adjust the start
             frame of play back waveform.
         7. Select the “End Frame” option, turn the “Universal” knob to adjust the end frame
             of play back waveform.
         8. Press the “Next Page 2/2” option button to return the first page of “Play Back
             menu”.
         9. Press the “   ” option button in “Operate” option to play back waveforms.
2.13.9      Recorder
   The waveform recorder is a kind of seamless and no-gap real time recording of waveform,
means oscilloscope can save and replay waveform every time it captured. It is similar to
waveform recording instrument, and the biggest recording size of its internal memory is 6M
                                            98
    Previous           Recall the waveform again
                              99
                                                      capturing time of each frame
                                         100
2.14 Online Help Function
multi-language help information, and you can recall them to help you operate the
Press the “HELP” button to enter the help status, then press every button to
recall the corresponding help information. Note:If you want to see help information
of SINGLE button and RUN/STOP button, you need to see them when you first enter
the help status because you press the SINGLE button can enter next page of help
information and Press the RUN/STOP button can return to the first page of help
Submenus of every main menu all have themselves’ help information. Note: If
you need to see help information of next page submenus, you need to first press the
“HELP” button to exit the help status and switch to next page menu then press the
“HELP” knob to enter the help status again and press the submenu option buttons to
Picture 2-86
                                          101
Chapter3 Application Examples
     This section presents a series of application examples. These simple examples
highlight the features of the oscilloscope and give you ideas for using it to solve your
                                          102
3.1Taking Simple Measurements
Observe a unknown signal in a circuit, display the signal quickly and measure the
1. Press the CH1 button, set the Probe option attenuation to 10X and set the switch to
The oscilloscope sets vertical, horizontal, and trigger controls automatically. If you
● Press the “Time” option button to enter the “Time Measure menu”.
The corresponding icon and measure value will display on the third option area.
                                          100
⑵ Measure signal peak-to-peak amplitude
● Press the “Voltage” option button to enter the “Voltage measure menu”.
The corresponding icon and measure value will display on the third option area.
                                    101
3.2 Taking Cursor Measurements
You can use the cursors to quickly take time and voltage measurements on a
waveform.
To measure the ring frequency at the rising edge of a signal, follow these steps:
5). Press the “CurA”button and turn the universal knob to place the cursor A on the
6). Press the “Cur B” button and turn the universal knob to place the cursor B on the
You can see the delta time and frequency (the measured ring frequency) on the top
Picture 3-1
                                          102
3.2.2 Measuring Ring Amplitude
5). Press the “CurA” option button and turn the universal knob to place the cursor A
6). Press the “CurB” option button and turn the universal knob to place the cursor B
You can see the following measurements on the top right of the screen:
Picture 3-2
                                          103
3.3 Catch the single signal
To capture a single event, you need to gather some pre-test knowledge of the signal in
order to correctly set up the trigger level and slope .if you are not sure the instance of
the signal, you can observe it through auto or normal trigger mode to ensure the
The following steps show you how to use the oscilloscope to capture a single event.
1. Set the Probe option attenuation to 10X and set the switch to 10X on the
probe.
(2) Under this menu, set the trigger type to “edge”, Edge type is “Rising”,
(3) Adjust the horizontal time base and vertical scale to the satisfied range
When the trigger conditions satisfied, data appears on the display representing;
the data points that the oscilloscope obtained with one acquisition and press the
RUN/STOP button again rearms the trigger circuit and erases the display.
                                            104
3.4 Analyze the signal detail
      You have a noisy signal displayed on the oscilloscope and you need to know more
about it. You suspect that the signal contains much more detail than you can now see
in the display.
The signal appears noisy and you suspect that noise is causing problems in your
2)Press the “Acquisition” option button or turn the “Universal” knob to select “Peak
Detect”.
3) If necessary, press the “DISPLAY” button to see “Display menu”. Turn the
“Universal” knob to adjust waveform intensity and grid brightness to see the noise
more easily.
4). Peak detect emphasizes noise spikes and glitches in your signal, specially when
2)Press the “Acquisition” option button or turn the “Universal” knob to select
“Average”.
3)Press the “Averages” option button to see the effects of varying the number of
Averaging reduces random noise and makes it easier to see detail in a signal.
                                            105
3.5 Triggering on a Video Signal
Observe the video circuit in a piece of medical equipment and Use the video trigger
4). Press the “Sync” option button and select “Odd Field” or “Even Field”.
6). Turn the horizontal “S/Div” knob to see a complete field across the screen.
7). Turn the vertical “Volts/Div” knob to ensure that the entire video signal is visible
on the screen.
3). Press the “Sync” option button and select “Line Num” and turn the universal
5). Turn the “S/Div” knob to see a complete video line across the screen.
6). Turn the “Volts/Div” knob to ensure that the entire video signal is visible on
the screen.
                                             106
3.6. Application of X-Y function
To view the input and output of the circuit in an XY display, follow these steps:
1)Press the “CH 1” MENU button and set the Probe option attenuation to 10X.
2) Press the “CH 2” MENU button and set the Probe option attenuation to 10X.
4). Connect the channel 1 probe to the input of the network, and connect the channel
6). Turn the “Volts/div” knobs to display approximately the same amplitude signals
on each channel.
The oscilloscope displays a Lissajous pattern representing the input and output
9).Turn the VOLTS/DIV and VERTICAL POSITION knobs to optimize the display.
11).Press the “Intensity” or “Brightness ” option buttons and turn the “Universal”
12). Apply the Ellipse method to observe the phase difference between the two
channels
                                            107
                                    The signal must be mediacy
                                    horizontal
A B
                                 Picture 3-3
sinθ=A/B or C/D    where θ=phase shift (in degrees )between the two signals From
If the main axis of the ellipse is at Ⅰ and Ⅲ quadrant , θ must be in the range of (0~
π/2) or (3π/2~2π).if the main axis is at Ⅰ and Ⅲ quadrant, θ must be in the range of
(π/2~π) or (π~3π/2).
                                      108
3.7 Analyzing a Differential Communication Signal
You are having intermittent problems with a serial data communication link, and you
suspect poor signal quality. Set up the oscilloscope to show you a snapshot of the
serial data stream so you can verify the signal levels and transition times.
Because this is a differential signal, you use the math function of the oscilloscope to
To activate the differential signals connected to channel 1 and channel 2, follow these
steps:
1) .Press the CH 1 button and set the Probe option attenuation to 10X.
2). Press the CH 2 button and set the Probe option attenuation to 10X.
7). Press the “CH1-CH2” option button to display a new waveform that is the
8). You can adjust the vertical scale and position of the Math waveform. To do so,
                                           109
Chapter 4 prompting messages and troubleshooting
4.1 Prompting messages:
■ Trig level at limit: Mention you that the trigger Level is at a limit when you turn
■ Horizon position at limit: Mention you that the horizontal position is at a limit
■ Volts/Div at limit: Mention you that the vertical voltage have already touched the
■ Volts position at limit: The system would display this information when the
■ Sec/Div at limit: Prompts that the Volts/Div is at full range while turning the
■ Holdoff time at limit!: Turn the “Universal” knob when holdoff time have been to
max or min value, now the system will clew this information.
■ Functions isn’t useable: Under several special modes, the some functions could
not be running.
■ No signal!: The system would clew this information when the signal could not
■ Adjust at limit: You could adjust the pulse width by the universal knob till the
■ Location Empty: If you have no stored waveforms or setups on some location, the
screen will display this information when you pressing the “Recall” button on this
location.
■ USB Flash Drive Plug In! : This information will appear when you invert the
                                          110
 USB Flash Drive to the USB Host port.
■ USB Flash Drive Pull Out! : This information will appear when you pull out the
■ Store Data Success!: Save setup data, waveform data or picture data to the internal
■ Ready Data Success!: Read setup data or waveform data from the internal of the
■ Please set Back USB to printer!: Press the “S/div” knob will appear this
information on the screen when the “Print Key” option is set to “Print Picture” and
■ USB Flash Drive isn’t connected!: When the “Save To” option is set to “File” or
the “Print Key” option is set to “Save Picture” in “Save/Recall” menu , Press the
“Save” option button or the “S/div” knob before inverting the USB Flash Drive to
the USB Host port will appear this information on the screen.
■ Record Wave Success!:This message will appear when you finish recording
waveforms.
                                   111
4.2 Troubleshooting
following steps:
⑷ If the oscilloscope is still not used after the checking, please connect with
ATTEN company.
2. If there is no signal wave in the screen after gathering the signal, please do as
following steps:
(2)Check the signal cable connecting with the BNC connector or not.
(3)Check the probe whether connect with the goods tested or not.
3. The value of the tested voltage is 10 times higher/lower than the real one ,
Check the attenuation quotiety whether match the probe attenuation proportion
or not.
(1) Check the signal source on the trigger interface whether or not match the signal
channel.
(2) Check the trigger mode: normal signal should use the “edge” trigger mode. The
video signal should use the “Video” Trigger mode. The signal would display
(3) Attempt to change the “ coupling” into “HF Reject” or “LF Reject” display, so
that the High/low frequency noise disturb the trigger should be filtrated
                                         112
    Check the trigger mode on the trigger interface whether or not in the “normal” or
“single”, and check the trigger level is whether or not over the wave range. If yes,
please put the trigger level to the middle position or set the trigger mode to the
“Auto” position. In another hand, you could choose the “Auto” button to set up
automatically.
6. After the Acquisition is set to Averages or Display Persistence time is set too
(1)This phenomena is normal. The time base maybe is too slow .you should turn
display.
(2) Maybe the display Type is set to “Vectors”, You could set it to Dots mode to
                                    113
Chapter 5 Service and Support
defects in materials and workmanship for a period of three (3) years from the date of
defective within the respective period, ATTEN will provide repair or replacement as
To arrange for service or obtain a copy of the complete warranty statement, please
makes no warranty of any kind, express or implied, including without limitation the
                                          114
5.2 Contract us
518055,P.R.China
Fax:86-755-6161 8298
E-mail:atten@atten.com.cn
http//www.atten.com.cn
                                      115
  Appendix A: Specifications
All specification apply to 10X probe and All ADS1000 Digital Storage Oscilloscopes.
To verify that the oscilloscope meets specifications, the oscilloscope must first meet
● The oscilloscope must have been operating continuously for twenty minutes within
● You must perform the Do Self Cal operation, accessible through the Utility menu,
  Specifications
                 Inputs
                 Input Coupling         AC, DC,GND
                 Input Impedance        1MΩ+/-2% || 17F±3pF
                 Maximum input          400V(DC+AC PK-PK,1MΩ input impedance),CAT I,CAT
                 voltage                II
                 Probe Attenuator       1X、10X
                 Probe Attenuator       1X、10X、100X、1000X
                  Factors Set
Horizontal System
Real Time Sampling       ADS1000C+ : Single Channel 500MSa/s
Rate                     ADS1000CA/CM : Single Channel 1GSa/s
Equivalent Sampling      50GSa/s
Rate                     (ADS1022C+ equivalent Sampling 10GSa/S;
                                117
                           ADS1042C+/CM equivalent Sampling 25GSa/S )
Measure        Display    MAIN, WINDOW, WINDOW ZOOM, ROLL, X-Y
 Modes
Horizontal nicety error   ±0.01%
Horizontal        Scan    2.5nS/DIV - 50S/DIV(ADS1152CA/CM)
Range                     2.5nS/DIV - 50S/DIV(ADS1102C+/CA/CM)
                          5nS/DIV - 50S/DIV(ADS1062C+/CA/CM)
                          10nS/DIV - 50S/DIV(ADS1042C+/CM)
                          25nS/DIV - 50S/DIV( ADS1022C+)
                          Roll: 100mS/DIV ~50S/DIV (1-2-5 sequence)
Delay range               Pretrigger: 12 DIV, Delay trigger: 130 DIV
 Trigger System
 Trigger Types             Edge, Pulse Width, Video,Slope, Alternative
 Trigger Modes                  Auto, normal, Single
 Trigger Coupling               AC, DC, LF rej, HF rej
 Trigger Level Range       CH1,CH2: ±6divisions from center of screen
                                EXT: ±1.2V
                                EXT/5: ±6V
 Trigger Displacement           Pretrigger : 6DIV, Delay Trigger: 6DIV
 Holdoff range                  100ns – 1.5s
 Edge Trigger                   Edge type: Rising, Falling, Rising and Falling
 Pulse Width Trigger            Trigger Modes: (>,<, =)positive Pulse Width,
                           (>, <, =)Negative Pulse Width
                                Pulse Width Range: 20ns – 10s
 Video Trigger                  Support signal Formats: PAL/SECAM, NTSC
                                Trigger condition : odd field, even field, all lines,
                           line Num
 Slope Trigger                  (>,<, =) Positive slope, (>,<, =) Negative
                           slope
                                Time: 20ns-10s
 Alternative Trigger            CH1 trigger type: Edge, Pulse, Video, Slope
                                CH2 trigger type: Edge, Pulse, Video, Slope
 X-Y Mode
 X-pole Input / Y-Pole     Channel 1    (CH1) / Channel 2 (CH2)
 Input
 Phase Error               ±3 degrees
 Sample Frequency          XY mode has a breakthrough that trad
                           oscilloscopes restrict sampling rate at 1MSa/s and
                           support 5KSa/s~200MSa/s (1-2-5 sequence)
 Measure System
 Auto Measure              Vpp, Vmax, Vmin, Vamp, Vtop, Vbase, Vavg,
 (32 Types)                Mean,Crms, Vrms, ROVShoot, FOVShoot,
                           RPREShoot, FPREShoot, Rise time, Fall time, Freq,
                           Period,+ Wid,-Wid, +Dut, -Dut, BWid, Phase,
                           FRR, FRF, FFR, FFF, LRR, LRF, LFR, LFF
 Cursor Measure            Manual mode, Track mode and Auto mode
                                118
Control Panel Function
Auto Set               Auto adjusting the Vertical, Horizontal system and
                       Trigger Position
Save/Recall            Support 2 Group referenced Waveforms,20 Group
                       setups,20 Group captured Waveforms internal
                       Storage/Recall function and USB flash driver storage
                       function.
               [1]
     NOTE:           ----ADS1000C+/CA   not support Long Memory
                              119
Generic Specification
          Display System
          Display Mode            Color TFT 5.7in.(145mm)diagonal Liquid Crystal
                                  Display
          Resolution              320 horizontal by 234 vertical pixels
          Display Color           64K color
          Display      Contrast   150:1
          (Typical state)
          Backlight Intensity     300nit
          (Typical state)
          Wave display range      8 x 12 div
          Wave Display Mode       Point, Vector
          Persist                 Off, 1 sec, 2 sec, 5 sec, Infinite
          Menu Display            2 sec, 5 sec, 10 sec, 20 sec, Infinite
          Skin                    Classical, Modern, Tradition, Succinct
          waveform                Sin(x)/x、Linear
          interpolation
          Color model             Normal , Invert
          Language                Simplified Chinese, Traditional Chinese,English,
                                  Arabic, French, German, Russian,Spanish,Portuguese
                                  Japanese,Korean
          Environmenta
          Temperature             Operating:10℃ to +40℃
                                  Nonoperating: -20℃ to +60℃
          Humidity                Operating: 95%RH, 40℃, 24 hours
                                  Nonoperating: 90%RH, 65℃, 24 hours
          Height                  Operating: 3000m
                                  Nonoperating: 15,266m
          pollution               It Usually only have dry insulative pollution
                                  It must be prospective that it is instant electric because
                                  of icing once in a way.
          Power Supply
          Input Voltage           100-240 VAC, CAT II, Auto selection
          Frequency Scope         45Hz to 440Hz
          Power                   50VA Max
         Mechanical
         Dimension length                        305mm
                    Width                        133mm
                    Height                       154mm
         Weight     2.3 kg
                                       120
Appendix B: Default setup
                     Gird
                     Menu Display            2 sec
    REF              Type                    waveform
                     Source                  CH1
                     REFA                    off
    TRIGGER (edge)   Type                    edge
                                       122
                                                      Appendix D:default setup
                Source                  CH1
                Slope                   Rising
                Mode                    Auto
                Coupling                DC
                LEVEL                   0.00V
TRIGGER         Type                    pulse
(pulse)         Source                  CH1
                When                    =
                Set Pulse Width         1.00ms
                Mode                    Auto
                Coupling                DC
TRIGGER         Type                    Video
(Video)         Source                  CH1
                Polarity                Normal
                Sync                    All Lines
                Standard                NTSC
                Type                    Slope
TRIGGER         Source                  CH1
(Slope)
                Mode                    Auto
TRIGGER         Type                    Alternative
(Alternative)   Source                  CH1
                                  123
Appendix C: Daily Maintain and Cleaning
Daily Maintain
DO not store or leave the instrument in where the LCD display will be exposed to
Cleaning
      If this instrument requires cleaning, disconnect it from all power sources and clean it
with a mild detergent and water. Make sure the instrument is completely dry before
1). Remove loose dust on the outside of the instrument and probes with a lint-free
cloth. Use care to avoid scratching the clear plastic display filter.
2). Use a soft cloth dampened with water to clean the instrument. Use an aqueous
Note:
To avoid damage to the surface of the instrument or probes, do not use any abrasive
                                            124
                                                                                                                                       目录
Index
                                                                       125
Troubleshooting ...................................... 112                Voltage Cursor ...........................................60
Universal knob .......................................... 18              Vpp ............................................................67
UTILITY................................................... 80             Wave Cancel ..............................................20
Utility System ........................................... 80             Window .....................................................38
Vertical System ......................................... 19              Window Zone ............................................39
Vmax......................................................... 67          X-Y Format ...............................................57
Vmin ......................................................... 67         YT..............................................................55
Volt/div...................................................... 28
126