Deltabar S PMD70/75, FMD76/77/78: Technical Information
Deltabar S PMD70/75, FMD76/77/78: Technical Information
 TI382P/00/EN/06.09
No. 71095431
                                                                                                                                                                                  Deltabar S
Table of contents
2                                                                                                                                                                       Endress+Hauser
                                                                                                                                                                                      Deltabar S
Human interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Operating elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .           53
Local operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .        55
Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .           55
Hard- und Software for on-site and remote operation . . . . . . . . .                            56
Ordering information . . . . . . . . . . . . . . . . . . . . . . . . 67
PMD70 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .      67
PMD70 (continued) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .            68
PMD70 (continued) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .            69
PMD75 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .      70
PMD75 (continued) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .            71
PMD75 (continued) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .            72
FMD76 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .      73
FMD76 (continued) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .            74
FMD76 (continued) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .            75
FMD77 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .      76
FMD77 (continued) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .            77
FMD77 (continued) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .            78
FMD78 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .      79
FMD78 (continued) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .            80
3                                                                                                                                                                            Endress+Hauser
                                                                                                                                                                                                 Deltabar S
+ –
+ –
                    With ceramic process                    With metallic process                   With ceramic process                  With metallic process                 With metallic process
                    isolating diaphragms                    isolating diaphragms                    isolating diaphragms                  isolating diaphragms                  isolating diaphragms and
                                                                                                                                          and diaphragm seal                    capillary diaphragm
                                                                                                                                          mounted on one side                   seals
Field of            – Flow                                  – Flow                                  – Level                               – Level                               – Level
application         – Level                                 – Level                                                                                                             – Differential pressure
                    – Differential pressure                 – Differential pressure
Process             – 1/4 – 18 NPT                          – 1/4 – 18 NPT                          Low-pressure side (–):                Low-pressure side (–):                – Wide range of
connections         – RC 1/4                                – RC 1/4                                – 1/4 – 18 NPT                        – 1/4 – 18 NPT                          diaphragm seals
                                                                                                    – RC 1/4                              – RC 1/4
                                                                                                    High-pressure side (+):               High-pressure side (+):
                                                                                                    – DN 80 – DN 100                      – DN 50 – DN 100
                                                                                                    – ANSI 3" – 4"                        – ANSI 2" – 4"
                                                                                                    – JIS 80A – 100A                      – JIS 80A – 100A
Measuring ranges    from –25...+25 mbar                     from –10...+10 mbar                     from –100...+100 mbar                 from –100...+100 mbar                 from –100...+100 mbar
                    to –3...+3 bar                          to –40...+40 bar                        to –3...+3 bar                        to –16 bar...+16 bar                  to –40...+40 bar
Overload 1          on one side: max. 100 bar on one side: max. 420 bar on one side: max. 100 bar on one side: max. 160 bar on one side: max. 160 bar
                    on both sides: max. 150 bar on both sides: max. 630 bar                                                 on both sides: max. 240 bar
Process           –20...+85°C                               –40...+120°C                            –20...+85°C                           up to + 400°C (+752°F)                up to +400°C (+752°F)
temperature range (–4...+185°F)                             (–40...+248°F)                          (–4...+185°F)
Ambient           –20...+85°C                               –40...+85°C                             –20...+85°C                           –40...+85°C                           –40...+85°C
temperature range (–4...+185°F)                             (–40...+185°F) 2                        (–4...+185°F)                         (–40...+185°F) 2                      (–40...+185°F) 2
Ambient                                                                                       –40 to +60°C (–40 to +140°F)
temperature range
separate housing
Reference           – Up to 0.075% of the set span                                                                                       – Up to 0.075 % of the set span
Accuracy            – PLATINUM version: up to 0.05% of the set span
Supply voltage      – For non-hazardous areas: 10.5...45 V DC
                    – Ex ia: 10.5...30 V DC
Output              4...20 mA with superimposed HART protocol, PROFIBUS PA or FOUNDATION Fieldbus
Options             – High-pressure version up to pstat 700 bar
                    – PMD75, FMD77, FMD78: Gold-Rhodium-coated process isolating diaphragm, NACE-compliant materials
                    – Separate housing
Specialities        – Metal-free measurement – pstat up to 420 bar                                  – Abrasion-resistant and              – For high media                      – Wide range of
(options)             with PVDF flange        – Process isolating                                     corrosion-resistant                   temperatures                          diaphragm seals
                    – Available with Deltatop   diaphragm: tantalum                                 – No diaphragm-seal
                      as flow compact device – Available with Deltatop                                temperature effects
                                                as flow compact device                              – Metal-free
                                                                                                      measurement possible
                                                                                                      with ECTFE-coated
                                                                                                      process connection
                    – HistoROM®/M-DAT memory module
1)    dependent on the lowest-rated element, with regard to pressure, of the selected components
2)    lower temperature on request
4                                                                                                                                                                                      Endress+Hauser
Deltabar S
Measuring principle
Ceramic process isolating diaphragms used for                         Metallic process isolating diaphragms used for PMD75, FMD77 and FMD78
PMD70 and FMD76
                                                                                                          ➀
                                     ➀                                                                    ➁
                                     ➁                                                                                                                                         ➆
                                                                                                                                                   ➅
    p1                                                p2
                                                                                                          ➂                                                                    ➇
                                     ➂                                     p1                             ➃             p2                    p1                                            p2
                                                                                                                                                                               ➈
                                     ➃                                                                    ➄
                                     ➄
                                       P01-xMD7xxxx-03-xx-xx-xx-000                                       P01-xMD7xxxx-03-xx-xx-xx-002                                       P01-xMD7xxxx-03-xx-xx-xx-003
Ceramic measuring cell PMD70 and FMD76                                Metal measuring cell 10 mbar and 30 mbar                           Metal measuring cell as of 100 mbar
1        Meter body                                                   1     Sensing element                                              6     Sensing element
2        Process isolating diaphragm                                  2     Silicon diaphragm                                            7     Overload diaphragm
3        Electrodes                                                   3     Process isolating diaphragm                                  8     Filling oil
4        Glass frit fixes the process isolating                       4     Filling oil                                                  9     Process isolating diaphragm
         diaphragm onto the meter body                                5     Overload diaphragm
5        Temperature sensor
                                                Metallic process isolating diaphragms used for PMD75, FMD77 and FMD78
                                                The process isolating diaphragms (3/9) are deflected on both sides by the acting pressures. A filling oil (4/8)
                                                transfers the pressure to a resistance circuit bridge (semi-conductor technology). The differential-pressure-
                                                dependent change of the bridge output voltage is measured and further processed.
                                                Advantages:
                                                • Standard operating pressures: 160 bar and 420 bar
                                                • High long-term stability
                                                • Very high single-sided overload resistance
                                                • Second process barrier (Secondary Containment) for enhanced integrity
Endress+Hauser                                                                                                                                                                                       5
                                                                                                                            Deltabar S
                                            +         –                                           +         –
                              Q ~ Δp         p1      p2                               Q ~ Δp        p1     p2
Q Q
P01-PMD7xxxx-15-xx-xx-xx-000
                   Flow measurement with Deltabar S and primary element, left: Orifice plate and right: Pitot tube
                   Q      Flow
                   p     Differential pressure, p = p1 – p2
                   Your benefits
                   • Choice of four flow modes of operation: volume flow, norm volume flow (European norm conditions),
                     standard volume flow (American standard conditions) and mass flow.
                   • Choice of diverse flow units with automatic unit conversion.
                   • A customised unit can be specified
                   • Low flow cut off: when activated, this function suppresses small flows which can lead to large fluctuations
                     in the measured value.
                   • Contains two totalizers as standard. One totalizer can be reset to zero.
                   • The totalizing mode and unit can be individually set for each totalizer. This allows independent daily and
                     annual quantity totalizing.
                   • With the product family Deltatop, Endress+Hauser is offering a universal and reliable solutions for flow
                     measurement:
                     – Deltatop, the compact, ready-to-use flow measuring unit including differential pressure transmitter
                        Deltabar S
                   Note!
                   For more information about flow measurement with the Deltabar S differential pressure transmitter
                   • Deltabar S with orifice plate (TI422P, Deltatop DO6x)
                   • Deltabar S with pitot tube (TI425P, Deltatop DP6x)
6                                                                                                                    Endress+Hauser
Deltabar S
                                                                             –                                    –        Δp
                                                                                                                        h= ρ g
                                                      ➀
                                                                                                     h
                                                                                                                                  ➂
                                                                                        ➁                         +
                                                                             +
                                                                                                                       P01xMD7xxxx-15-xx-xx-xx-000
                            Your benefits
                            •   Choice of three level operating modes
                            •   Volume and mass measurements in any tank shapes by means of a freely programmable characteristic curve
                            •   Choice of diverse level units with automatic unit conversion
                            •   A customised unit can be specified
                            •   Has a wide range of uses, e.g.
                                – for level measurement in tanks with superimposed pressure
                                – in the event of foam formation
                                – in tanks with agitators of screen fittings
                                – in the event of liquid gases
                                – for standard level measurement
Endress+Hauser                                                                                                                                 7
                                                                                                                                                     Deltabar S
                                          Input
Measured variable                         Differential pressure, from which flow (volume or mass current) and level (level, volume or mass) are derived
Measuring range PMD75, FMD77, FMD78 (with metallic process isolating diaphragms)
Nominal              Measurement         Smallest          MWP 1                   Overload 2              Min. operating pressure 3          Versions in the
value                   limit            calibratable                                                                                          order code 4
                                         span 5
Nominal              Measurement         Smallest          MWP 1                   Overload 2              Min. operating pressure 3 Versions in the
value                   limit            calibratable                                                                                order code 4
                                         span 5
1)        The MWP (maximum working pressure; MWP = PN) for the measuring device depends on the weakest element of the components selected with regard to
          pressure, i.e. the process connection ( ä 30 ff) has to be taken into consideration in addition to the measuring cell ( see table above). Also observe
          pressure-temperature dependency. For the appropriate standards and further information, see  ä 29, "Pressure specifications" section.
2)        The maximum pressure for the measuring device is dependent on the lowest-rated element, with regard to pressure, of the selected components. See also
           ä 29, section "Pressure specifications".
3)        The minimum operating pressure indicated in the table applies to silicone oil under reference operating conditions.
          Min. operating pressure at 85°C (185°F) for silicone oil: 10 mbarabs.
          FMD77 and FMD78: Min. operating pressure: 50 mbarabs; observe also the pressure and temperature application limits of the selected filling oil on  ä 59.
          For vacuum applications, please observe the installation instructions on  ä 65 ff.
4)        Versions in the order code  ä 71 ff, feature 40 "Nominal range; PN"
5)        Turn down > 100:1 on request
6)        PN 160 versions with stainless steel A2 screws, PN 420 versions with stainless steel A4 M12 screws
          PN 420 versions for PMD75 only.
7)        PMD75 only
8)        minimum span that can be calibrated for PMD75: 1 mbar; minimum span that can be calibrated for FMD77 and FMD78: 5 mbar
9)        For PMD75 with CRN-approved process connections, the MWP is 315 bar.
10)       "–" side: 100 bar
8                                                                                                                                              Endress+Hauser
Deltabar S
Explanation of terms Explanation of the terms: Turn down (TD), set span and zero based span
                       Case 1:
                       • Lower range value Upper range value                                       ➀=➁
                       Example:                                                    LRL                LRV     URV                URL
                       • Lower range value (LRV) = 0 mbar
                       • Upper range value (URV) = 100 mbar                   –500 mbar                 0     100          +500 mbar
                       • Nominal value (URL) = 500 mbar
                       Turn down:
                       • TD = URL /URV:1
                                                                                                                                  ➂
                       set span:                                                                                  ➃
                       • URV – LRV = 100 mbar
                          This span is based on the zero point.                                         ➄
                                                                                                                      P01-xxxxxxxx-05-xx-xx-xx-001
                       Case 2:
                       • Lower range value Upper range value                               ➀=➁
                       Example:                                                    LRL     LRV          URV                      URL
                       • Lower range value (LRV) = –300 mbar
                       • Upper range value (URV) = 0 bar
                       • Nominal value (URL) = 500 mbar                       –500 mbar                  0                 +500 mbar
Endress+Hauser                                                                                                                                 9
                                                                                                                                      Deltabar S
                        Output
Output signal           • 4...20 mA with superimposed digital communication protocol HART 5.0, 2-wire
                        • Digital communication signal PROFIBUS PA (Profile 3.0)
                          – signal coding: Manchester Bus Powered (MBP); Manchester II
                          – data transmission rate: 31.25 KBit/s, voltage mode
                        • Digital communication signal FOUNDATION Fieldbus
                          – signal coding: Manchester Bus Powered (MBP); Manchester II
                          – data transmission rate: 31.25 KBit/s, voltage mode
                                       847                                                          804
                                                                  ➃                                                              ➃
                                       413
                                                         ➂                                          369
                                                                                                                     ➂
                                          10.5      20       30         40 45 U                        11.5     20        30         40 45 U
                                                                              [V]                                                          [V]
                                                           U – 10.5 V                                                   U – 11.5 V
                                                 RLmax ≤                                                      RLmax ≤
                                                            23 mA                                                        23 mA
                                                                                                                                P01-xMD7xxxx-05-xx-xx-xx-000
                        Load diagram, observe the position of the jumper and the explosion protection ( ä 17, section "Taking 4...20 mA test
                        signal".)
                        1       Jumper for 4...20 mA test signal inserted in "Non-test" position
                        2       Jumper for 4...20 mA test signal inserted in "Test" position
                        3       Supply voltage 10.5 (11.5)...30 V DC for 1/2 D, 1 GD, 1/2 GD, FM IS, CSA IS, IECEx ia, NEPSI Ex ia
                        4       Supply voltage 10.5 (11.5)...45 V DC for device for non-hazardous areas, 1/2 D, 1/3 D, 2 G Ex d, 3 G Ex nA, FM
                                XP, FM DIP, FM NI, CSA XP, CSA Dust-Ex, NEPSI Ex d
                        RLmax   Maximum load resistance
                        U       Supply voltage
                        Note!
                        When operating via a handheld terminal or via PC with an operating program, a minimum communication
                        resistance of 250 must exist within the loop.
10                                                                                                                             Endress+Hauser
Deltabar S
                                                            I
                                                      100 %
                                                       90 %
63 %
t1 t2 t
P01-xxxxxxxx-05-xx-xx-xx-007
                        Reading cycle
                        • HART commands: on average 3 to 4 per second on average.
                          The Deltabar S commands the BURST MODE function for cyclic value transmission via the HART
                          communication protocol.
                        Response time
                         250 ms
Endress+Hauser                                                                                                                                  11
                                                                                                                            Deltabar S
                      Response time
                      • cyclic: approx. 10 ms per request
                      • acyclic: < 50 ms
                      All values are typical values.
                      Reading cycle
                      • cyclic: up to 5/s, dependent on the number and type of function blocks used in a closed-control loop
                      • acyclic: 10/s
                      Response time
                      • cyclic: < 80 ms
                      • acyclic: < 40 ms
                      All values are typical values.
12                                                                                                                  Endress+Hauser
Deltabar S
DD Revision 01 (hex)
Number VCRs 44
Permanent Entries 44
Client VCRs 0
                         Server VCRs                                5
                         Source VCRs                                8
Sink VCRs 0
                         Subscriber VCRs                            12
                         Publisher VCRs                             19
Link Settings
Slot time 4
Transducer Blocks
                         TRD1 Block          contains all parameters related to the measurement   • Pressure, Flow or Level (Channel 1)
                                                                                                  • Process temperatur (Channel 2)
                         Diagnsotic Block contains diagnostiv information                         Error code via DI channels (channel 0 to
                                                                                                  6)
Display Block contains parameters to configure the local display no output values
Endress+Hauser                                                                                                                         13
                                                                                                                  Deltabar S
Function Blocks
     Resource Block     The Resource Block contains all the data that uni-                                   enhanced
                        quely identifies the field device. It is an electronic
                        version of a nameplate of the device.
     Analog Input       The AI block takes the manufacturer's input data,                        45 ms       enhanced
     Block 1            selected by channel number, and makes it avai-
     Analog Input       lable to other function blocks at its output.
     Block 2            Enhancement: digital outputs for process alarms,
                        fail safe mode
     Digital Input      This block contains the discrete data of the diag-                       40 ms       standard
     Block              nose block (selectable via a channel number 0 to
                        16) and provides them for the blocks at the out-
                        put.
     Digital Output     This block converts the discrete input and thus                          60 ms       standard
     Block              initiates an action (selectable via a channel num-
                        ber) in the dp flow block or in the service block.
                        Channel 1 resets the counter for max. pressure
                        transgressions..
     Signal Characte- The signal characterizer block has two sections,                           30 ms       standard
     rizer Block      each with an output that is a non-linear function
                      of the respective input. The non-linear function is
                      determined by a single look-up table with 21 arbi-
                      trary x-y pairs.
     Analog Alarm       This block contains all process alarm conditions                         35 ms       standard
     Block              (working like a comparator) and represents them
                        at the output.
14                                                                                                        Endress+Hauser
Deltabar S
                        Power supply
Electrical connection   Note!
                        • When using the measuring device in hazardous areas, installation must comply with the corresponding
                          national standards and regulations and the Safety Instructions or Installation or Control Drawings.  ä 82,
                          section "Safety Instructions" and "Installation/Control Drawings".
                        • Devices with integrated overvoltage protection must be earthed.  ä 28.
                        • Protective circuits against reverse polarity, HF influences and overvoltage peaks are installed.
4...20 mA HART
                             ➀                                                                                        ➅    10.5 V DC
                                                                                                                      ➆    11.5 V DC
                                                                                              4…20 mA
                                         Test
Test
                             ➂                                        ➄
                                                                                                                                                 ➇
                                                                                                          4... 20mA     Test
                                                                      ➃
                                                                                                                               P01-xMx7xxxx-04-xx-xx-xx-001
                        PROFIBUS PA
                        The digital communication signal is transmitted to the bus via a 2-wire connection. The bus also provides the
                        auxiliary energy. For further information on the network structure and grounding and for further bus system
                        components such as bus cables, see the relevant documentation, e.g. Operating Instructions BA034S
                        "Guidelines for planning and commissioning PROFIBUS DP/PA" and the PNO Guideline.
                        Cable specifications:
                        • Use a twisted, shielded two-wire cable, preferably cable type A
                        Note!
                        For further information on the cable specifications, see Operating Instructions BA034S
                        Guidelines for planning and commissioning PROFIBUS DP/PA", PNO Guideline 2.092 "
                        PROFIBUS PA User and Installation Guideline" and IEC 61158-2 (MBP).
Endress+Hauser                                                                                                                                       15
                                                                                                                  Deltabar S
     FOUNDATION Fieldbus
     The digital communication signal is transmitted to the bus via a 2-wire connection. The bus also provides the
     auxiliary energy. For further information on the network structure and grounding and for further bus system
     components such as bus cables, see the relevant documentation, e.g. Operating Instructions BA013S
     "FOUNDATION Fieldbus Overview" and the FOUNDATION Fieldbus Guideline.
     Cable specifications:
     • Use a twisted, shielded two-wire cable, preferably cable type A
     Note!
     For further information on the cable specifications, see Operating Instructions BA013S "FOUNDATION
     Fieldbus Overview", FOUNDATION Fieldbus Guideline and IEC 61158-2 (MBP).
                                                                                        +
                                                                                            7
                  + –                                                      –    8                   6
                                                                                            1
                                                                                2                   5
                                                                                    3           4
                                  Han7D
                  + –
                                                                                                            P01-xMD7xxxx-04-xx-xx-xx-000
                  + –
                                                                       –                                +
M12
                  + –
                                                                                                            P01-xMD7xxxx-04-xx-xx-xx-008
     Endress+Hauser offers for devices with M12 plug the following accessories:
     Plug-in jack M 12x1, straight
     • Material: Body PA; coupling nut CuZn, nickel-plated
     • Degree of protection (fully locked): IP67
     • Order number: 52006263
     Plug-in jack M 12x1, elbowed
     • Material: Body PBT/PA; coupling nut GD-Zn, nickel-plated
     • Degree of protection (fully locked): IP67
     • Order number: 51006327
     Cable 4x0.34 mm2 with M12 socket, elbowed, screw plug, 5 m length
     • Material: Body PUR; coupling nut CuSn/Ni; cable PVC
     • Degree of protection (fully locked): IP67
     • Order number: 52010285
16                                                                                                      Endress+Hauser
Deltabar S
                                                                                      –
                                   + –
7/8” +
                                  +–
                                                                                                                          P01-xxx7xxxx-04-xx-xx-xx-003
Kabel gland
                 Terminals
                 for wire cross-sections of 0.5 to 2.5 mm2
Endress+Hauser                                                                                                                                  17
                                                                                                                               Deltabar S
                            4...20 mA HART
                            • Version for non-hazardous areas, jumper for 4...20 mA test signal in "Test" position (delivery status):
                              11.5...45 V DC
                            • Version for non-hazardous areas, jumper for 4...20 mA test signal in "Non-test" position: 10.5...45 V DC
                            PROFIBUS PA
                            • Version for non-hazardous areas: 9...32 V DC
                            FOUNDATION Fieldbus
                            • Version for non-hazardous areas: 9...32 V DC
Current consumption         • PROFIBUS PA: 13 mA ± 1 mA, switch-on current corresponds to IEC 61158-2, Clause 21
                            • FOUNDATION Fieldbus: 15 mA ± 1 mA, switch-on current corresponds to IEC 61158-2, Clause 21
Residual ripple             Without influence on 4...20 mA signal up to 5 % residual ripple within the permitted voltage range [according
                            to HART hardware specification HCF_SPEC-54 (DIN IEC 60381-1)]
18                                                                                                                      Endress+Hauser
Deltabar S
10 ±0.150 -- --
                                100                                          ±0.180                             --                   --
                                500                                          ±0.025                          ±0.050               ±0.075
                                Note!
                                Position-dependent zero shift can be corrected.  ä 23, section "General installation instructions" and
                                 ä 62 ff, section "Response time".
Endress+Hauser                                                                                                                                    19
                                                                                                                                 Deltabar S
                            PMD75
                            Measuring cell                            % of the set span
                                                     • TD 1:1               =   ±0.15
                            10 mbar, 30 mbar
                                                     • TD > 1:1             =   ±0.15 x TD
                            Platinum version:
                                                     • TD 1:1               =   ±0.05
                            00 mbar
                            FMD77, FMD78
                            Measuring                           FMD77                                               FMD78
                            cell
                                                            % of the set span (influence of the diaphragm seal included)
Total performance – PMD75   The "Total performance" specification comprises the non-linearity including hysteresis, non-reproducibility, the
                            thermal change of the zero point as well as the influence of the line pressure (pst = 70 bar).
                            Measuring cell           AISI 316L/1.4435, Alloy, Gold-Rhodium            Tantal process isolating diaphragm
                                                      or Monel process isolating diaphragm
Total Error The total error comprises the long-term stability and the total performance:
                            Measuring cell           AISI 316L/1.4435, Alloy, Gold-Rhodium            Tantal process isolating diaphragm
                                                      or Monel process isolating diaphragm
% of URL/year
                            10 mbar, 30 mbar,
                                                                        0.33                                        0.48
                            100 mbar
20                                                                                                                           Endress+Hauser
Deltabar S
Influence of the operating   Measuring     AISI 316L/1.4435, Alloy, Gold-Rhodium                  Tantal process isolating diaphragm
pressure on zero point and   cell          coated or Monel process isolating diaphragm
span – PMD75, FMD77,
FMD78                                      Influence of the            Influence of the           Influence of the         Influence of the
                                           operating pressure on       operating pressure on      operating pressure on    operating pressure on
                                           the zero point              the span                   the zero point           the span
                             10 mbar       0.15 % of URL/7 bar        0.035 % of URL/7 bar      0.28 % of URL/7 bar     0.28 % of URL/7 bar
                             30 mbar       0.35 % of URL/70 bar       0.14 % of URL/70 bar      0.70 % of URL/70 bar    0.70 % of URL/70 bar
100 mbar 0.15 % of URL/70 bar 0.14 % of URL/70 bar 0.42 % of URL/70 bar 0.42 % of URL/70 bar
500 mbar 0.075 % of URL/70 bar 0.14 % of URL/70 bar 0.14 % of URL/70 bar 0.14 % of URL/70 bar
3 bar 0.075 % of URL/70 bar 0.14 % of URL/70 bar 0.14 % of URL/70 bar 0.14 % of URL/70 bar
16 bar 0.075 % of URL/70 bar 0.14 % of URL/70 bar 0.14 % of URL/70 bar 0.14 % of URL/70 bar
40 bar 0.075 % of URL/70 bar 0.14 % of URL/70 bar 0.14 % of URL/70 bar 0.14 % of URL/70 bar
                             Note!
                             The influence of the operating pressure on the zero point can be calibrated out.
Endress+Hauser                                                                                                                                     21
                                                                                                                                       Deltabar S
                                                                                     • TD 1:1                =   ±0.15
                               25 mbar
                                                                                     • TD > 1:1              =   ±0.15 x TD
Total performance –            The "Total performance" specification comprises the non-linearity including hysteresis, non-reproducibility, the
PMD70, FMD76                   thermal change of the zero point as well as the influence of the line pressure (pst = 70 bar).
Total Error The total error comprises the long-term stability and the total performance:
Influence of the operating     Measuring cell      Influence of the operating pressure on the     Influence of the operating pressure on the
pressure on zero point and                         zero point                                     span
span – PMD70, FMD76
                               25 mbar             0.7 % of URL/7 bar                            0.14 % of URL/7 bar
                               Note!
                               The influence of the operating pressure on the zero point can be calibrated out.
Thermal change of the zero     Measuring cell                     –10...+60 °C                                –20...–10 °C, +60...+85 °C
output and the output span –                                     (+14...+140°F)                              (–4...+14°F, +140...+185°F)
PMD70, FMD76
                                                                                         % of the set span
22                                                                                                                             Endress+Hauser
Deltabar S
                        Level measurement
                        • PMD70, PMD75, FMD76 and FMD77 are best suited to level measurement in open tanks. All Deltabar S
                          devices are suitable for level measurement in closed tanks.
                        Measuring arrangement level measurement in closed tanks and closed tanks with superimposed vapour
                        • PMD70, PMD75: Mount device below the lower measuring connection. Always connect the negative side
                          above the maximum level.
                        • FMD76, FMD77: Mount device direct on the tank. Always connect the negative side above the maximum
                          level.
                        • In the case of level measurement in closed tanks with superimposed vapour, a condensate trap ensures
                          pressure which remains constant on the minus side.
                        Pressure measurement
                        •   The PMD70 and PMD75 are best suited to differential pressure measurement.
                        •   Measuring arrangement for gases: Mount device above the measuring point.
                        •   Measuring arrangement for liquids and steams: Mount device below tapping point.
                        •   For differential pressure measurement in vapour, mount the condensate traps at the same level as the same
                            the tapping point and at the same distance from Deltabar S.
Endress+Hauser                                                                                                                       23
                                                                                                                                       Deltabar S
Heat insulation – FMD77   The FMD77 must only be insulated up to a certain height. The maximum permitted insulation height is labelled
                          on the devices and applies to an insulation material with a heat conductivity
                           0.04 W/(m x K) and to the maximum permitted ambient and process temperature ( see table below). The
                          data were determined under the most critical application "quiescent air".
                                                            TA                                          λ ≤ 0.04 W
                                                                                                                m•K
                                                                                                               TP
                                                                                                                                P01-FMD77xxxx-11-xx-xx-xx-000
FMD77
Wall- and pipe-mounting   Endress+Hauser offers a mounting bracket for installing the device on pipes or walls.  ä 67 ff, feature 110,
                          "Additional options 2".
                          Note!
                          If a valve block is used, its dimensions should also be taken into consideration.
                                                                                                     37.5
                                                                  ➀
                                                     135
                                                                        74
                                                                 12                                             74
                                                                   4                                                       6
                                                                                                                      10
P01-xMD7xxxx-11-xx-xx-xx-008
24                                                                                                                             Endress+Hauser
Deltabar S
"Separate housing" version   With the "separate housing" version, you are able to mount the housing with the electronics insert at a distance
                             from the measuring point. This facilitates trouble-free measurement:
                             • Under particularly difficult measuring conditions (at installation locations that are cramped or difficult to
                                access)
                             • If rapid cleaning of the measuring point is required
                             • If the measuring point is exposed to vibrations.
                             You can choose between different cable versions:
                             • PE (2 m, 5 m and 10 m)
                             • FEP (5 m).
                              ä 68 ff, Feature 110, "Additional options 2", Version "G".
                              For the dimensions, see  ä 50.
                                             FEP cable:
                                             IP 69K
                                             IP 66/68 NEMA 4/6P                                        ➃           IP xx
                                                                                                                   (see chapter
                                             PE cable:                                                             “Ordering information”)
                                             IP 66/68 NEMA 4/6P
                                                                                           ➂
                                                                                  r ³ 120 mm
                                                        ➀
                                                                                ➁
P01-xMD7xxxx-11-xx-xx-en-010
                             In the case of the "separate housing" version, the sensor is delivered with the process connection and cable ready
                             mounted. The housing and a mounting bracket are enclosed as separate units. The cable is provided with a socket at both
                             ends. These sockets are simply connected to the housing and the sensor.
                             1      Process connection with sensor
                             2      Cable, both ends are fitted with a socket
                             3      Mounting bracket provided, suitable for pipe and wall mounting
                             4      Housing with electronic insert
                             Degree of protection for the process connection with sensor with the use of
                             • FEP cable:
                               – IP 69K
                               – IP 66 NEMA 4/6P
                               – IP 68 (1.83 mH2O for 24 h) NEMA 4/6P
                             • PE cable:
                               – IP 66 NEMA 4/6P
                               – IP 68 (1.83 mH2O for 24 h) NEMA 4/6P
                             Technical data of the PE and FEP cable:
                             • Minimum bending radius: 120 mm (4.72 inch)
                             • Cable extraction force: max. 450 N
                             • Resistance to UV light
                             Use in hazardous area:
                             • Intrinsically safe installations (Ex ia/IS)
                             • FM/CSA IS: for Div.1 installatin only
Endress+Hauser                                                                                                                                           25
                                                                                                                                        Deltabar S
Turn the housing              The housing can be rotated up to 380° by loosening the Allen screw.
                              Your benefits
                              • Simple mounting by optimally aligning the housing
                              • Good, accessible device operation
                              • Optimum readability of the on-site display (optional).
max. 380°
P01-xMD7xxxx-17-xx-xx-xx-001
Oxygen applications           Oxygen and other gases can react explosively to oils, grease and plastics, such that, among other things, the
                              following precautions must be taken:
                              – All components of the system, such as measuring devices, must be cleaned in accordance with the BAM
                                 (DIN 19247) requirements.
                              – Dependent on the materials used, a certain maximum temperature and a maximum pressure for oxygen
                                 applications must not be exceeded.
                              The devices suitable for gaseous oxygen applications are listed in the following table with the specification pmax.
                              Order code for devices cleaned for         pmax for oxygen applications              Tmax for oxygen
                              oxygen applications                                                                  applications
Ultra pure gas applications   Endress+Hauser also offers devices for special applications, such as ultra pure gas, cleaned from oil and grease.
                              No special restrictions regarding the process conditions apply to these devices.
                               ä 67 ff, PMD70 and PMD75: feature 80 "Seal", FMD76 and FMD77: feature 70 "Process connection low-
                              pressure side, material, seal".
26                                                                                                                           Endress+Hauser
Deltabar S
Process isolating diaphragms    With regard to materials in which hydrogen build-up takes place, hydrogen atoms can diffuse through the
for materials with hydrogen     metalic process isolating diaphragms. This can result in incorrect measurement results.
build-up (Gold-Rhodium          Endress+Hauser offers process isolating diaphragms with Gold-Rhodium coating for this application.
coating)
                                 ä 71 ff "Ordering information PMD75",  ä 70 "Ordering information FMD77" or  ä 76 "Ordering
                                information FMD78", feature 60 "Process isolating diaphragm material".
Ambient temperature range       • PMD75, FMD77, FMD78: –40...+85°C (–40...+185°F), devices for lower temperatures on request
                                • PMD70, FMD76: –20...+85°C (–4...+185°F)
                                • On-site display: –20...+70°C (–4...+158°F)
                                  Extended temperature application range with restrictions in optical properties such as display speed and
                                  contrast: –40...+85°C (–40...+185°F)
                                • Separate housing: –40 to +60°C (–40 to +140°F)
                                For devices for use in hazardous areas, see Safety instructions, Installation or Control Drawing ( ä 82,
                                sections "Safety Instructions" and "Installation/Control Drawings").
                                The device can be used in this temperature range. The values of the specification, such as thermal change, may
                                be exceeded. See also DIN 16086.
Climate class                   Class 4K4H (air temperature: –20...55°C/–4...+131F, relative humidity: 4...100%) fulfilled as per
                                DIN EN 60721-3-4 (condensation possible)
Electromagnetic compatibility   • Electromagnetic compatibility to EN 61326 and NAMUR recommendation EMC (NE21). For details refer
                                  to the declaration of conformity.
                                • With enhanced immunity against electromagnetic fields as per EN 61000-4-3:
                                  30 V/m with closed cover (for devices with T14 or T15 housing)
                                • Maximum deviation: < 0.5% of span 1
                                • All EMC measurements were performed with a turn down (TD) = 2:1.
1) Larger deviations possible with PMD70 with 25 mbar or 100 mbar sensor
Endress+Hauser                                                                                                                                   27
                                                                                                                                 Deltabar S
Versions for feature 70 in the order code Seal Temperature operating range
28                                                                                                                        Endress+Hauser
Deltabar S
                          Versions for feature 70 in the         Seal on the LP side (–)                      Process temperature range 1
                          order code
Pressure specifications   • The maximum pressure for the measuring device is dependent on the lowest-rated element with regard to
                            pressure, see the following sections for this:
                            –  ä 8 ff, section "Measuring range"
                            – chapter "Mechanical construction".
                            The MWP (maximum working pressure) is specified on the nameplate. This value refers to a reference
                            temperature of 20°C (68°F) or 100°F for ANSI flanges and may be applied to the device for an unlimited
                            time. Observe pressure-temperature dependency.
                          • The pressure values permitted at higher temperatures can be found in the following standards:
                            – EN 1092-1: 2001 Tab. 18 1
                            – ASME B 16.5a – 1998 Tab. 2-2.2 F316
                            – ASME B 16.5a – 1998 Tab. 2.3.8 N10276
                            – JIS B 2220
                          • For PMD70 and PMD75, the MWP applies for the temperature ranges specified in the "Ambient temperature
                            range" ( ä 27) and "Process temperature limits" ( ä 28) sections.
                          • The test pressure corresponds to the over pressure limit of the measuring instrument (Over pressure limits
                            OPL = 1.5 x MWP) and may fit only temporally limited, so that no permanent damage develops.
                          • The Pressure Equipment Directive (EC Directive 97/23/EC) uses the abbreviation "PS". The abbreviation
                            "PS" corresponds to the MWP (maximum working pressure) of the measuring device.
                          • In the case of sensor range and process connections where the OPL (Over Pressure Limit) of the pressure
                            connection is smaller than the nominal value of the sensor, the device is set at the factory, at the very
                            maximum, to the OPL value of the process connection. If you want to use the entire sensor range, select a
                            process connection with a higher OPL value (1.5 x PN; PN = MWP).
                          • In oxygen applications, the values for "pmax and Tmax for oxygen applications" as per  ä 26, "Oxygen
                            applications" may not be exceeded.
                          1)     With regard to their stability-temperature property, the materials 1.4435 and 1.4404 are grouped together under 13EO
                                 in EN 1092-1 Tab. 18. The chemical composition of the two materials can be identical.
Endress+Hauser                                                                                                                                    29
                                                                                                                                     Deltabar S
                               Mechanical construction
Housing dimensions T14,
optional display on the side                      152
                                                   111                                                111
                                                                                    FIELD TERMINALS
                                                                  H
                                                                                                                               P01-xMD7xxxx-06-00-xx-xx-000
P01-xMD7xxxx-06-00-xx-xx-001
P01-xMD7xxxx-06-00-xx-xx-002
30                                                                                                                            Endress+Hauser
Deltabar S
+ – – +
                                                                        7/16-20 UNF                        +
                                                                        M10
                                          41.3
                                          108
                                      H
                                                                        1/4-18 NPT
                                                                        RC1/4
                                                         54                                            96
                                                         84
P01-PMD70xxx-06-09-xx-xx-000
                                 B          1/4-18 NPT IEC 61518     7/16-20 UNF       Steel C 22.8            2 vent valves (AISI 316L/1.4404)     4.0 kg
                                                                                                   2
                                                                                                               included
                                 D          1/4-18 NPT IEC 61518     7/16-20 UNF       AISI 316L                                                    4.0 kg
                                                                                                       3
                                 F          1/4-18 NPT IEC 61518     7/16-20 UNF       Alloy C276              Vent valves (Alloy C276/2.4819), 4.2 kg
                                                                                                               see  ä 68 feature 110
                                                                                                               "Additional options 2".
                                 U          RC 1/4                   7/16-20 UNF       AISI 316L 2             2 vent valves (AISI 316L/1.4404)     4.0 kg
                                                                                                               included
                                 1          1/4-18 NPT IEC 61518     PN 160: M10       Steel C 22.8                                                 4.0 kg
                                 2          1/4-18 NPT IEC 61518     PN 160: M10       AISI 316L   2                                                4.0 kg
                                                                                                       3
                                 3          1/4-18 NPT IEC 61518     PN 160: M10       Alloy C276              Vent valves (Alloy C276/2.4819), 4.2 kg
                                                                                                               see  ä 68, feature 110
                                                                                                               "Additional options 2".
Endress+Hauser                                                                                                                                                     31
                                                                                                                                         Deltabar S
                                             H
                                                  108
                                                        41.3
                                                                                                                             1/4-18 NPT
                                                                   54                                          96
                                                                   84
P01-PMD70xxx-06-09-xx-xx-001
Process connection PMD70, version G, PVDF inlay, PN = 10 bar, process temperature T = –10...+60°C (14...+140°F)
Device height H
32                                                                                                                             Endress+Hauser
Deltabar S
                                                                          7/16-20 UNF
                                                                          M 10
                                       H
                                                 41.3        + –                                                                            – +
                                           85
                                                                            1/4-18 NPT
                                                                            RC1/4
                                                              54                                              106
                                                              85
                                                             100
                                                                           7/16-20 UNF
                                                                           M10 (M12)
                                       H
                                                   41.3
                                           72
                                                             +–             1/4-18 NPT                                                     –+
                                                                            RC1/4
                                                               54                                             100
                                                               87
                                                               98
                                                                                                                                            P01-PMD75xxx-06-09-xx-xx-005
                                  B             1/4-18 NPT IEC 61518    7/16-20 UNF       Steel C 22.8          2 vent valves (AISI 316L/1.4404)     4.2 kg
                                                                                                      2
                                                                                                                included
                                  D             1/4-18 NPT IEC 61518    7/16-20 UNF       AISI 316L                                                  4.2 kg
                                  F             1/4-18 NPT IEC 61518    7/16-20 UNF       Alloy C276      3     Vent valves (Alloy C276/2.4819), 4.5 kg
                                                                                                                see  ä 72, feature 110
                                                                                                                "Additional options 2".
                                  U             RC 1/4                  7/16-20 UNF       AISI 316L 2           2 vent valves (AISI 316L/1.4404)     4.2kg
                                                                                                                included
                                  1             1/4-18 NPT IEC 61518    – PN 160: M10     Steel C 22.8                                               4.2 kg
                                                                        – PN 420: M12
                                  3             1/4-18 NPT IEC 61518    – PN 160: M10     Alloy C276 3          Vent valves (Alloy C276/2.4819), 4.5 kg
                                                                        – PN 420: M12                           see  ä 72 ff, feature 110
                                                                                                                "Additional options 2".
                                  1)       Weight of process connections without vent valves with 10 mbar or 30 mbar sensors, process connections without
                                           vent valves with sensors  100 mbar weght approx. 800 g less. Housing weight see  ä 51.
                                  2)       AISI 316L/1.4435 or 1.4404
                                  3)       Alloy C276/2.4819
Endress+Hauser                                                                                                                                                    33
                                                                                                                                                Deltabar S
Process connections PMD75         Oval flange, connection 1/4-18 NPT or RC 1/4, with side vent
with metallic process isolating
diaphragms (continued)
                                                 41.3
                                                             + –                                                                           – +
                                           85
                                                                                                                        42
                                                                           1/4-18 NPT
                                                                           RC1/4
                                                              54                                     106
                                                              85
                                                             100
                                                                                                                                          P01-PMD75xxx-06-09-xx-xx-004
                                                                           7/16-20 UNF
                                                                                                                     1/4-18 NPT
                                       H
41.3
                                                                                                                        28
                                            72
                                                              +–            1/4-18 NPT                                                        –+
                                                                            RC1/4
                                                               54                                        100
                                                               87
                                                               98
                                                                                                                                          P01-PMD75xxx-06-09-xx-xx-003
                                  C             1/4-18 NPT IEC 61518   7/16-20 UNF        Steel C 22.8         4 locking screws and                4.2 kg
                                                                                                               2 vent valves included
                                  E             1/4-18 NPT IEC 61518   7/16-20 UNF        AISI 316L 2                                              4.2 kg
                                                                                                               (AISI 316L/1.4404)
                                  H             1/4-18 NPT IEC 61518   7/16-20 UNF        Alloy C276 3         Vent valves (Alloy C276/2.4819), 4.5 kg
                                                                                                               see  ä 72, feature 110
                                                                                                               "Additional options 2".
                                  1)       Weight of process connections without vent valves with 10 mbar or 30 mbar sensors, process connections without
                                           vent valves with sensors  100 mbar weght approx. 800 g less. Housing weight see  ä 51
                                  2)       PN 160 bar measuring cells: AISI 316L/1.4435, PN 420 bar measuring cells: AISI 316L/1.4435 or 1.4404
                                  3)       Alloy C276/2.4819
34                                                                                                                                      Endress+Hauser
Deltabar S
Process connections PMD75         Oval flange, prepared for diaphragm seal mount
with metallic process isolating
diaphragms (continued)
                                                                                    7/16-20 UNF
                                       H
                                                                                                                                               ➁
                                                                               ➀
                                                                                                                                               ➂
                                                                                                                      41.3
                                             72
                                                               +–                            M6             –+
                                                               54                                           54
                                                               87
                                                                 98
P01-PMD75xxx-06-09-xx-xx-002
                                  Left: Process connection PMD75, version W, prepared for diaphragm seal mount
                                  Right: Position of the copper ring seal
                                  H        Device height see the following section "Device height H"
                                  1        Diaphragm seal attachment
                                  2        Copper ring seal
                                  3        Process isolating diaphragm
Device height H
1) Values for devices with 10 mbar and 30 mbar measuring cell in brackets
Endress+Hauser                                                                                                                                     35
                                                                                                                                                       Deltabar S
41
41.3
7/16-20 UNF
                                                                                                                                   H
                                                                                                                               h
                                                                           g2
                                                                                                                 1/4-18 NPT
                                                                   2
ø46
                                                                                                                          b
                                                                                                    g
                                                                                                    k
                                                                                                    D
P01-FMD76xxx-06-09-xx-xx-000
           Flange 1                                                                                          Boltholes
Version    Material         Nominal     Shape    2    Nominal       Diameter      Thickness Raised           Quantity     Diameter      Hole            Flange
                            diameter                  pressure                              face                                        circle          weight 3
D b g g2 k
1)   The roughness of the surface in contact with the medium, including the sealing surface of the flanges (all standards), made of Hastelloy C, Monel or
     Tantalum is Ra 0.8 μm. Lower surface roughness on request.
2)   Designation as per DIN 2527 in brackets
3)   Housing weight see  ä 51
4)   ECTFE coating on AISI 316L/1.4435,
     When operating in hazardous area, avoid electrostatic charge of the plastic surfaces.
5)   Alloy C276/2.4819
36                                                                                                                                             Endress+Hauser
Deltabar S
Process connection FMD76                 ANSI flanges, connection dimensions as per ANSI B 16.5, raised face RF and
with ceramic process isolating           JIS flanges, connection dimensions as per JIS B 2220 BL, raised face RF
diaphragms (continued)
                                                                                                                           41
                                                                                                                         1.61 in
                                                                                                       41.3
                                                                                                      1.63 in
7/16-20 UNF
                                                                                                                                                         H
                                                                                                                                                    h
                                                                                         g2
                                                                                                                                   1/4-18 NPT
ø46
                                                                                                                                                b
                                                                                                             ø1.81 in
                                                                               2 (0.078 in)
                                                                                                                     g
                                                                                                                    k
                                                               ANSI
                                                                               JIS
                                                                                                                    D
                                                                                                                                                                       P01-FMD76xxx-06-09-xx-xx-001
                                         Process connection FMD76, high-pressure side: ANSI or JIS flange (see table below),
                                         low-pressure side: connection 1/4-18 NPT
                                         H      Device height see  ä 38, section "Device height H, devices with flange"
                                         h      Height of the device without flange thickness b
              Flange 1                                                                                                         Boltholes
Version       Material           Nominal       Class/          Diameter                        Thickness      Raised face      Quantity    Diameter      Hole circle         Flange
                                 diameter      Nominal                                                                                                                       weight 2
                                               pressure
D b g g2 k
[in] / [mm] [in] / [mm] [in] / [mm] [in] / [mm] [in] / [mm] [kg]
ANSI flanges
P AISI 316/316L 3 3 in 150 lb./sq.in 7.5 / 190.5 0.94 / 23.9 5 / 127 4 0.75 / 19.1 6 / 152.4 4.9
R ECTFE 4 4.9
T AISI 316/316L  4 in 150 lb./sq.in 9 / 228.5 0.94 / 23.9 6.19 / 157.2 8 0.75 / 19.1 7.5 / 190.5 7.1
U ECTFE  7.1
V Alloy C276 8
W AISI 316/316L  4 in 300 lb./sq.in 10 / 254 1.25 / 31.8 6.19 / 157.2 8 0.88 / 22.4 7.88 / 200.2 11.7
JIS flanges
1 AISI 316L 80 A 10 K 7.32 / 185 0.71 / 18 5 / 127 8 0.75 / 19.1 5.9 / 150 3.3
4 AISI 316L 100 A 10 K 8.27 / 210 0.71 / 18 5.95 / 151 8 0.75 / 19.1 6.89 / 175 4.4
1)    The roughness of the surface in contact with the medium, including the sealing surface of the flanges (all standards), made of Hastelloy C, Monel or
      Tantalum is Ra 0.8 μm. Lower surface roughness on request.
2)    Housing weight see  ä 51
3)    Combination of AISI 316 for required pressure resistance and AISI 316L for required chemical resistance (dual rated)
4)    ECTFE coating on AISI 316L.
      When operating in hazardous area, avoid electrostatic charge of the plastic surfaces.
Endress+Hauser                                                                                                                                                                               37
                                                                                                                                                  Deltabar S
T14 housing, optional display on the side 175 mm + flange thickness b (see tables)
T15 housing without display, flat cover 181 mm + flange thickness b (see tables)
T15 housing with display, high cover 193.5 mm + flange thickness b (see tables)
T17 housing, optional display on the side 191 mm + flange thickness b (see tables)
                                                                                41.3
                                                            +–                                                                            –+
                                                                                  1/4-18 NPT
                                           150
                                                             103.7
                                                                                                                                            P01-FMD77xxx-06-xx-xx-xx-001
38                                                                                                                                    Endress+Hauser
Deltabar S
                                                                                                            H
                                                                                                       h
                                                                                              g2
                                                                                                   b
                                                                          dM
                                                1.6 mm
                                                0.06 in
                                                                                                                                                            L
                                                                          g
                                                                          k                                                            d3
                                                                          D                                                             g
P01-FMD77xxx-06-09-xx-xx-002
                                        Process connection FMD77, high-pressure side EN/DIN flange with and without extended diaphragm seal, material
                                        AISI 316L
                                        H       Device height see  ä 41, section "Device height H"
                                        h       Height of the device without flange thickness b
Ver- No-          Nominal     Shape Dia-          Thick-   Raised     Exten-    Exten-    Quan- Dia-            Hole     max.   TK           TK           Flange
                              2
sion minal        pressure          meter         ness     face       sion      sion      tity  meter           circle   Dia-   Am-          Pro-         weight 3
     dia-                                                             length    dia-                                     phragm bient        cess
     meter                                                                      meter                                    dia-
                                                                                                                         meter
D b g L d3 g2 k dM
                                                                      100                                                                                 6.7
                                                                      200                                                                                 7.8
1)    The roughness of the surface in contact with the medium, including the sealing surface of the flanges (all standards), made of Hastelloy C, Monel or
      Tantalum is Ra 0.8 μm. Lower surface roughness on request.
2)    Designation as per DIN 2527 in brackets
3)    Housing weight see  ä 51
Endress+Hauser                                                                                                                                                         39
                                                                                                                                                         Deltabar S
Process connections FMD77                ANSI flanges, connection dimensions as per B 16.5, raised face RF
with metallic process isolating
diaphragms, high-pressure
side (continued)
                                                                                                              H
                                                                                                       h
                                                                                              g2
                                                                                                   b
                                                 1.6 mm                     dM
                                                 0.06 in
                                                                                                                                                              L
                                                                            g
                                                                            k                                                           d3
                                                                            D                                                            g
P01-FMD77xxx-06-09-xx-xx-000
                                         Process connection FMD77, high-pressure side ANSI flange with and without extended diaphragm seal, material
                                         AISI 316/316L
                                         H       Device height see  ä 41, section "Device height H"
                                         h       Height of the device without flange thickness b
D b g L d3 g2 k dM
                   [lb./sq.in]   [in]        [in]          [in]     [in]         [in]                  [in]       [in]     [in]      [mbar/10 K]              [kg]
                                 [mm]        [mm]          [mm]     [mm]         [mm]                  [mm]       [mm]     [mm]
                                                                    4                                                                                         6.6
                                                                    101.6
                                                                    6                                                                                         7.1
                                                                    152.4
                                                                    8                                                                                         7.7
                                                                    203.8
1)     The roughness of the surface in contact with the medium, including the sealing surface of the flanges (all standards), made of Hastelloy C, Monel or
       Tantalum is Ra 0.8 μm. Lower surface roughness on request.
2)     Housing weight see  ä 51
40                                                                                                                                              Endress+Hauser
Deltabar S
Process connections FMD77                 JIS flanges, connection dimensions as per JIS B 2220 BL , raised face RF
with metallic process isolating
diaphragms, high-pressure
side (continued)
                                                                                                                                   H
                                                                                                                               h
                                                                        0.0787"
                                                                                                                     g2
2 mm
                                                                                                                          b
                                                                                                 dM
                                                                                                    g
                                                                                                    k
                                                                                                 D
                                                                                                                                                      P01-FMD77xxx-06-09-xx-xx-001
                                          Process connection FMD77, high-pressure side, JIS flange, material AISI 316L/1.4435
                                          H     Device height see the following section "Device height H"
                                          h     Height of the device without flange thickness b
Ver-    Nominal      Nominal       Diameter Thick-          Raised       Quan-        Diameter Hole          max. Dia-        TK            TK                 Flange
sion    diameter     pressure               ness            face         tity                  circle        phragm           Ambient       Process            weight 2
                                                                                                             diameter
D b g g2 k dM
1)     The roughness of the surface in contact with the medium, including the sealing surface of the flanges (all standards), made of Hastelloy C, Monel or
       Tantalum is Ra 0.8 μm. Lower surface roughness on request.
2)     Housing weight see  ä 51
Device height H
T14 housing, optional display on the side 325 mm + flange thickness b (see tables)
T15 housing without display, flat cover 331 mm + flange thickness b (see tables)
T15 housing with display, high cover 343.5 mm + flange thickness b (see tables)
T17 housing, optional display on the side 341 mm + flange thickness b (see tables)
Endress+Hauser                                                                                                                                                              41
                                                                                                    Deltabar S
P01-FMD78xxx-06-xx-xx-xx-000
Device height H
42                                                                                           Endress+Hauser
Deltabar S
                                                                                                                              b
                                                                                                        dM
P01-FMD78xxx-06-09-xx-xx-000
Version          Nominal              Nominal             max. Diameter       Thickness           max. Diaphragm TK                       Weight of two
                 diameter             pressure 1                                                  diameter       Process                  diaphragm seals
D b dM
Version          Nominal              Nominal             max. Diameter       Thickness           max. Diaphragm TK                       Weight of two
                 diameter             pressure 1                                                  diameter       Process                  diaphragm seals
D b dM
1)    The specified nominal pressure applies to the diaphragm seal. The maximum pressure for the measuring device is dependent on the lowest-rated element,
      with regard to pressure, of the selected components. See also  ä 29, section "Pressure specifications".
Endress+Hauser                                                                                                                                                     43
                                                                                                                                                  Deltabar S
Process connection FMD78          Threaded ISO 228 G 1/2 B and ANSI 1/2 MNPT, separator with PTFE seal
with metallic process isolating
diaphragm (continued)
                                                                   100                                                    100
90
                                                                                                           89
                                                                                               ➀                                                     ➀
                                                                                 SW22                                                  SW22
                                                                                 22 AF                                                 22 AF
                                                                             6                                                         1/2 NPT
                                                                   G½
P01-FMD78xxx-06-09-xx-xx-012
                                  Process connection FMD78, left: with threaded connection ISO 228 G 1/2 B, right: with threaded connection
                                  ANSI 1/2 MNPT
                                  1       PTFE seal as standard max. 260 °C/500 °F (higher temperatures on request)
[mbar/10 K] [kg]
                                                                                                   dM
                                                                                                   C7
P01-FMD78xxx-06-09-xx-xx-005
                                  Process connection FMD78, Material: AISI 316L, surface roughness of the wetted surfaces  0.8 m as standard. Lower
                                  surface roughness on request.
Version     Nominal          Nominal                Nominal             Diameter           max.           Height        TK Process         Weight of two
            diameter         diameter               diameter                               Diaphragm                                       diaphragm
            ISO 2852         DIN 32676                                                     diameter                                        seals
C7 dM h
TD DN 51 DN 50 2 64 48 30 +1.00 2.2
44                                                                                                                                       Endress+Hauser
Deltabar S
                                                                                                                            D
                                                                                                                           d2
                                                                                                         L
P01-FMD78xxx-06-09-xx-xx-001
                                           Process connection FMD78, Material: AISI 316L, surface roughness of the wetted surfaces  0.8 m as standard. Lower
                                           surface roughness on request.
Version         Nominal            Nominal         Diameter       Diameter         Diameter    Height          Face-to-face      TK Process      Weight of two
                diameter           diameter                                                                    length                            diaphragm
                ISO 2852                                                                                                                         seals
                                                   D              d1               d2          h               L
                                   [in]            [mm]           [mm]             [mm]        [mm]            [mm]              [mbar/10 K]     [kg]
1) Including 3.1 and pressure test as per Pressure Equipment Directive, category II
                                                                                                    dM max. = 50
                                                                                                       ø68
                                                                                                                                                 P01-FMD78xxx-06-09-xx-xx-006
                                           Process connection FMD78, surface roughness of the wetted surfaces  0.8 m as standard. Lower surface roughness on
                                           request.
[mbar/10 K] [kg]
Endress+Hauser                                                                                                                                                         45
                                                                                                                                    Deltabar S
41
                                                             23.6
                                                                            dM max. =50               4 x ø11.5
                                                                                ø65–1.2
                                                                                  ø84
                                                                                 ø105
                                                                                                                               P01-FM78xxx-06-09-xx-xx-002
                           Process connection FMD78, surface roughness of the wetted surfaces  0.8 m as standard. Lower surface roughness on
                           request.
[mbar/10 K] [kg]
53.5
ø91.45
ø100
ø132.7
P01-FMD78xxx-06-09-xx-xx-011
                           Process connection FMD78, surface roughness of the wetted surfaces  0.8 m as standard. Lower surface roughness on
                           request.
[mbar/10 K] [kg]
46                                                                                                                           Endress+Hauser
Deltabar S
Process connection FMD78            Taper adapter with coupling nut, DIN 11851 (dairy fitting)
with metallic measuring
diaphragms (continued)
                                                                                                                             40
                                                                                                                     f
                                                                                                                   m
                                                                                                                   k
                                                                                             dM
                                                                                              D
                                                                                              G
                                                                                                                                      P01-FMD78xxx-06-09-xx-xx-007
                                    Process connection FMD78, material AISI 316L, surface roughness of the wetted surfaces  0.8 m as standard. Lower
                                    surface roughness on request.
Ver-      Nominal    Nominal       Diameter    Adapter        Thread            Height        Height        max.        TK Process       Weight of
sion      diameter   pressure                  height                                                       Diaphragm                    two
                                                                                                            diameter                     diaphragm
                                                                                                                                         seals
D f G k m dM
                                                                                             dM
                                                                                             d1
                                                                                             G
                                                                                                                                      P01-FMD78xxx-06-09-xx-xx-008
                                    Process connection FMD78, material AISI 316L, surface roughness of the wetted surfaces  0.8 m as standard. Lower
                                    surface roughness on request.
Version        Nominal          Nominal            Diameter       Height            Thread             max.             TK Process    Weight of two
               diameter         pressure                                                               Diaphragm                      diaphragm
                                                                                                       diameter                       seals
d1 h G dM
Endress+Hauser                                                                                                                                              47
                                                                                                                                                        Deltabar S
Process connection FMD78                  EN/DIN flanges, connection dimensions as per EN 1092-1/DIN 2527
with metallic measuring                   JIS flanges, connection dimensions as per JIS B 2220 BL
diaphragms (continued)
g2
                                                                                                                             b
                                                                                                       dM
                                                                             2
                                                                                                       g
                                                                                                       k
                                                                                                       D
P01-FMD78xxx-06-09-xx-xx-009
Ver-    Nominal Nominal            Shape 2    Dia-        Thick-      Raised      Quan-        Dia-        Hole     max.            TK              Weight of
sion    diameter pressure                     meter       ness        face        tity         meter       circle   Diaphragm       Process         two
                                                                                                                    diameter                        diaphragm
                                                                                                                                                    seals
D b g g2 k dM
1)     The roughness of the surface in contact with the medium, including the sealing surface of the flanges (all standards), made of Hastelloy C, Monel or
       Tantalum is Ra 0.8 μm. Lower surface roughness on request.
2)     Designation as per DIN 2527 in brackets
Ver-    Nominal         Nominal    Diameter       Thickness Raised           Quantity     Diameter     Hole         max.            TK               Weight of
sion    diameter        pressure                            face                                       circle       Diaphragm       Process          two
                                                                                                                    diameter                         diaphragm
                                                                                                                                                     seals
D b g g2 k dM
1)     The roughness of the surface in contact with the medium, including the sealing surface of the flanges (all standards), made of Hastelloy C, Monel or
       Tantalum is Ra 0.8 μm. Lower surface roughness on request.
48                                                                                                                                              Endress+Hauser
Deltabar S
Process connection FMD78                 ANSI flanges, connection dimensions as per ANSI B 16.5, raised face RF
with metallic measuring
diaphragms (continued)
                                                           g2
                                                                                                 b
                                                                             dM
                                                 1.6 mm
                                                 0.06 in
                                                                                                                                                                   L
                                                                                g
k d3
D g
P01-FMD78xxx-06-09-xx-xx-010
Process connection FMD78, ANSI flange with and without extended diaphragm seal, material AISI 316/AISI 316L
Ver-      Nominal Class          Dia-        Thick-             Raised   Exten-     Exten-   Quan-       Dia-      Hole       max. Dia- TK                  Weight of
sion      dia-                   meter       ness               face     sion       sion     tity        meter     circle     phragm    Process             two dia-
          meter                                                          length     dia-                                      diameter                      phragm
                                                                                    meter                                                                   seals
                                 D           b                  g        L          d3                   g2        k          dM
                      [lb/sq.in] [in]        [in]               [in]     [in]       [in]                 [in]      [in]       [in]         [mbar/           [kg]
                                 [mm]        [mm]               [mm]     [mm]       [mm]                 [mm]      [mm]       [mm]         10 K]
                                                                         6                                                                                  14.2
                                                                         152.4
                                                                         8                                                                                  15.4
                                                                         203.6
AH        4           150        9           0.94               6.19     –          –        8           0.75      7.5        3.50         +0.25            14.4
                                 228.6       23.9               157.2                                    19.1      190.5      89
                                                                         6                                                                                  22.3
                                                                         152.4
                                                                         8                                                                                  24.8
                                                                         203.6
1)     The roughness of the surface in contact with the medium, including the sealing surface of the flanges (all standards), made of Hastelloy C, Monel or
       Tantalum is Ra 0.8 μm. Lower surface roughness on request.
Endress+Hauser                                                                                                                                                            49
                                                                                                                                         Deltabar S
                                                                                131
                                                                                45.5
                                                           ø51.7
                                                                                                                                    P01-xxxxxxxx-06-xx-xx-xx-000
Dimensions T14 housing, optional display on the side. Housing weight see  ä 51.
                                                 101               76                                           118
                                                                              147
                                                                              45.5
                                                         ø51.7
                                                                                                                                    P01-xxxxxxxx-06-xx-xx-xx-001
Dimensions T17 housing, optional display on the side. Housing weight see  ä 51.
                                      T14                                              T17
                                                                                                               45
                                                                   45
ø54.1
                                                  ➀                                             ➀                           ➀
                                                                                                                                                       81
+– +– +– +– +–
P01-xxxxxxxx-06-xx-xx-xx-002
                             Reduction of the mounting height of the process connection, for application of the separate housing.
                             1     Process connection adapter.
                             If the separate housing is used, the mounting height of the process connection is reduced by approx. 45 mm
                             as compared to the dimensions of the standard version.
                             The minimum bending radius (r) for the cable is 120 mm (4.7").
50                                                                                                                               Endress+Hauser
Deltabar S
Weight Housing
                                     Process connections
                                     •   Process connections PMD70 with ceramic process isolating diaphragms:  ä 31 ff
                                     •   Process connections PMD75 with metallic process isolating diaphragms:  ä 33 ff
                                     •   Process connection FMD76 with ceramic process isolating diaphragms:  ä 36 ff
                                     •   Process connections FMD77 with metallic process isolating diaphragms, low-pressure side:  ä 38 ff
                                     •   Process connection FMD78 with metallic process isolating diaphragm:  ä 43 ff
                                     Process connections
                                     • "Clamp connections" and "Hygienic connections" (see also Chapter "Ordering information"): AISI 316L/
                                       1.4435
                                     • "Threaded connection" and "DIN/EN flanges" (see also Chapter "Ordering information"): stainless steel AISI
                                       316L with the material number 1.4435 or 1.4404
                                     • With regard to their stability-temperatur property, the materials 1.4435 and 1.4404 are grouped together
                                       under 13E0 in EN 1092-1 Tab.18. The chemical composition of the two materials can be identical.
Endress+Hauser                                                                                                                                   51
                                                                                                          Deltabar S
     !   Note!
         Process wetted device components are listed in the "Mechanical construction" ( ä 30) and "Ordering
         information" ( ä 67) sections.
         Miscellaneous:
         • Measuring cell PMD70/FMD76, filling oil
           – 25 mbar and 100 mbar measuring cell: Silicone oil
           – 500 mbar and 3000 mbar measuring cell: Mineral oil
           – for oxygen and ultra pure gas applications: Inert oil (Voltalef 1A)
         • Measuring cell PMD75/FMD77 and FMD78, filling oil (Silicone oil)
           – for oxygen and ultra pure gas applications: Inert oil (Halocarbon 6.3)
         • Process isolating diaphragm material:
           – PMD70/FMD76: Al2O3 (Aluminium-oxide-ceramic)
           – PMD75, FMD77, FMD78:
             – AISI 316L (1.4435)
             – Alloy C276 (2.4819)
             – Monel 400 (2.4360)
             – Tantal
             – AISI 316L (1.4435) with Gold-Rhodium coating (FMD77/FMD78 only)
             – AISI 316L (1.4435) with 0.09 mm PTFE foil (FMD77/FMD78 only)
         • Process isolating diaphragm PMD70/FMD76: Al2O3 (Aluminium-oxide-ceramic)
         • Mounting accessories: mounting bracket with screws AISI 304 (1.4301)
         • Capillary: AISI 316 Ti (1.4571)
         • Protective hose for capillary: AISI 304 (1.4301)
         • External earth terminal: AISI 304 (1.4301)
         • Screws and nuts for side flanges:
           – PMD70: hex.-headed bolt DIN 931-M10x50-A2-70, hex.-headed nut: DIN 934-M10-A4-70
           – PMD75 PN 160: hex.-headed bolt DIN 931-M12x90-A2-70, hex.-headed nut DIN 934-M12-A2-70
           – PMD75 PN 420: hex.-headed bolt ISO 4014-M12x90-A4, hex.-headed nut ISO 4032-M12-A4-bs
         For process connections, seals and filling oils see ordering information,  ä 67 ff.
52                                                                                                  Endress+Hauser
Deltabar S
                     Human interface
Operating elements   On-site display (optional)
                     A 4-line liquid crystal display (LCD) is used for display and operation. The on-site display shows measured
                     values, dialog text as well as fault and notice messages in plain text, thereby supporting the user in every stage
                     of operation. The liquid crystal display of the device can be turned in 90° steps.
                     Depending on the installation position of the device, this makes it easy to operate the device and read the
                     measured value.
                     Functions:
                     • 8-digit measured value display including sign and decimal point, bar graph for 4 to 20 mA HART as current
                       display or for PROFIBUS PA as graphical display of the scaled value of the AI Block
                     • Simple and complete menu guidance thanks to seperation of the parameters into three levels
                     • Ech parameter is given as 3-digit ID number for easy navigation
                     • Option for configuring the display according to individual requirements and desires, such as language,
                       alternating display, display of other measured values such as sensor temperature, contrast setting
                     • Comprehensive diagnostic functions (fault and warning message, peak-hold indicators, etc.)
                     • Rapid and safe commissioning with the Quick Setup menus
                                                                 Header line
                                                                 Main line
                                                                                                                                    Unit
                                                                 Information
                                                                 line
                               –    +    E                                     Symbol
                                                                                                   Bargraph
                                                                             Editing modes
                             Operating keys
                                                                 Selection
                                                                 options
                                                                 Value that
                                                                 can be edited
Endress+Hauser                                                                                                                                53
                                                                                                                                                                                                                                                                                                Deltabar S
4...20 mA HART
                                                                                                                                                                                                                PROFIBUS PA/
                                                                                                                                                                                                                FOUNDATION Fieldbus
                                                                                                                                                                                                                                          0%
                                                                                                                                                                                                                                                       Zero
P01-PMx7xxxx-19-xx-xx-xx-038
                                                                                  The operating keys located externally on the device work on the Hall sensor principle. As a result, no additional
                                                                                  openings are required in the device. This guarantees:
                                                                                  • Complete protection against environmental influences such as moisture and contamination
                                                                                  • Simple operation without any tools
                                                                                  • No wear.
          ➃➄                                                                                             ➅➆                                                                                                               ➄➅
           t        ➅                                                                                        t
                                                                                                                                                                                                                               Sim.
    on
                            ➀                                                                      on
                                                                                                                          ➀              ➁                          ➂                                               on
                                                                                                                                                                                                                                          ➀                                 ➁
                                                                                       ➁                                                                                                   ➃                                                                                                                   ➂
    off                                                                                            off                              0%                                                                              off    1   2
                                                                                                                                                                                                                                                                                                                                ➃
                                                                                                                                                                                                                                                                   0%
                                                                                                                                Sensor
                                                                                                                                               Zero
                                                                                                                                                                        Display                             ➄                                                   Sensor
                                                                                                                                                                                                                                                                             Zero
                                                                                                                                                                                                                                                                                            Display
                                                                                                                                                                                                                                          Simulation
➂ on on ON CK S D A 0 8
                                                                                                                                                                                                                                                          on
                                                                                                                                                                                                                                                                                             HW
                1   2                                                                                             1   2   off     off      1    2   3   4   5   6   7    8
                                                                                                                                                                                                   PC
                                                                                                                                                                                                                                      1   2               off
                                                                                  PC
                                                                                                                                                                                                                                                                                                                      PC
                                                                                                                                         1 2 3 4 5 6 7 8
                                                                                                                                             Address
                                                                                                                                                                                                Histo ROM
Histo ROM
                                                                                                                                                                        SW
                                                                                                                                                                        HW
                            HART
                                              R
                                                                    250002271-–
                                                                                                                                                                                                                                                                                 TM
                                                                                                                                                                                  250002272-–
250002273-–
                            HW-Version:                    1                                                                                                                                                                                                             FOUNDATION
                            SW-Version:                    2                                                                             HW-Version:                1                                                                                                    HW-Version:    1
                                                                                                                                         SW-Version:                2                                                                                                    SW-Version:    2
54                                                                                                                                                                                                                                                                                Endress+Hauser
Deltabar S
Remote operation Depending on the position of the write protection switch at the device, all software parameters are accessible.
                   HART
                   Remote operation via:
                   • Handheld terminal Field Communicator 375 (see Chapter "Hard- und Software for on-site and remote
                     operation"  ä 56)
                   • FieldCare (see Chapter "Hard- und Software for on-site and remote operation"  ä 56 ff) mit
                     – Commubox FXA191 (see Chapter "Hard- und Software for on-site and remote operation"  ä 56 ff)
                     – Commubox FXA195 (see Chapter "Hard- und Software for on-site and remote operation"  ä 56 ff)
                   • Field Xpert:
                     Field Xpert is an industrial PDA with integrated 3.5" touchscreen from Endress+Hauser based on Windows
                     Mobile. It communicates via wireless with the optional VIATOR Bluetooth modem connected to a HART
                     device point-to-point or wireless via WiFi and Endress+Hauser’s Fieldgate FXA520. Field Xpert also works
                     as a stand-alone device for asset management applications. For details refer to BA060S/00/en.
                   PROFIBUS PA
                   Remote operation via:
                   • FieldCare (see Chapter "Hard- und Software for on-site and remote operation"  ä 56 ff)
                     – Profiboard: For the Connection of a Personal Computer to PROFIBUS
                     – Proficard: For the Connection of a Laptop to PROFIBUS
                   FOUNDATION Fieldbus
                   Remote operation via:
                   • Handheld terminal Field Communicator 375 (see Chapter "Hard- und Software for on-site and remote
                     operation"  ä 56 ff)
                   • Use an FF-configuration program for e.g. NI-FBUS configurator, to
                     – connect devices with "FOUNDATION Fieldbus signal" into an FF-network
                     – set FF-specific parameter
                     Operation with NI-FBUS Configurator:
                     The NI-FBUS Configurator is an easy-to-use graphical environment for creating linkages, loops and a
                     schedule based on the fieldbus concept.
                     You can use the NI-FBUS Configurator to configure a fieldbus network as follows:
                     – Set block and device tags
                     – Set device addresses
                     – Create and edit function block control strategies (function block applications)
                     – Configure vendor -defined function and transducer blocks
                     – Create and edit schedules
                     – Read and write to function block control strategies (function block applications)
                     – Invoke Device Description (DD) methods
Endress+Hauser                                                                                                                 55
                                                                                                                                      Deltabar S
                                   –   Display DD menus
                                   –   Download a configuration
                                   –   Verify a configuration and compare it to a saved configuration
                                   –   Monitor a downloaded configuration
                                 Commubox FXA195
                                 For intrinsically safe communication with FieldCare via the USB interface. For details refer to TI404F/00/en.
                                 Commubox FXA291
                                 The Commubox FXA291 connects Endress+Hauser field instruent with CDI interface (=Endress+Hauser
                                 Common Data Interface) to the USB interface of a personal computer or a notebook. For details refer to
                                 TI405C/07/en.
                                 Note!
                                 For the following Endress+Hauser instruments you need the "ToF Adapter FXA291" as an additional accessory:
                                 • Cerabar S PMC71, PMP7x
                                 • Deltabar S PMD7x, FMD7x
                                 • Deltapilot S FMB70
56                                                                                                                           Endress+Hauser
Deltabar S
                 HistoROM®/M-DAT (optional)
                 HistoROM®/M-DAT is a memory module, which is attached to the electronic insert. The HistoROM®/M-DAT
                 can be retrofitted at any stage (Order number: 52027785).
                 Your benefits
                 • Quick and safe commissioning of the same measuring points by copying the configuration data of one
                   transmitter to another transmitter
                 • Reliable process monitoring thanks to cyclical recording of pressure and sensor temperature measured values
                 • Simple dagnosis by recording diverse events such as alarms, configuration changes, counters for measuring
                   range undershoot and overshoot for pressure and temperature as well as user limit overshoot and undershoot
                   for pressure and temperature etc.
                 • Analysis and graphic evaluation of the events and process parameters via software (contained in scope of
                   supply).
                 HistoROM®/M-DAT can be ordered via feature 100 "Additional options 1" or feature 110 "Additional options
                 2" or as spare parts.  ä 70 ff. A CD with Endress+Hauser operating program is also included in the scope of
                 delivery.
                 You can copy data from one transmitter to another transmitter when operating a FOUNDATION Fieldbus
                 device via an FF configuration program. You need the Endress+Hauser FieldCare operating program and the
                 Commubox FXA291 service interface and the ToF Adapter FXA291 to be able to access the data and events
                 saved in the HistoROM®/M-DAT.
                 FieldCare
                 FieldCare is an Endress+Hauser asset management tool based on FDT technology. With FieldCare, you can
                 configure all Endress+Hauser devices as wella s devices from other manufacturers that support the FDT
                 standard.
                 FieldCare supports the following functions:
                 • Configuration of transmitter in offline and online operation
                 • Loading and saving device data (upload/download)
                 • HistoROM®/M-DAT analysis
                 • Documentation of the measuring point
                 Connection options:
                 • HART via Commubox FXA191 and the RS232C serial interface of a computer
                 • HART via Commubox FXA195 and the USB port on a computer
                 • PROFIBUS PA via segment coupler and PROFIBUS interface card
                 • FOUNDATION Fieldbus via Commubox FXA193 and the RS232C serial interface of a computer
                 • Service interface with adapter Commubox FXA291 and ToF Adapter FXA291 (USB).
                 For further information see È www.endress.com
Endress+Hauser                                                                                                             57
                                                                                                                                Deltabar S
Applications                Diaphragm seal systems should be used if the process media and the device should be separated. Diaphragm
                            seal systems offer clear advantages in the following instances:
                            • In the case of high process temperatures ( ä 28, section "Process temperature limits".)
                            • In the case of process media that crystallise
                            • In the case of corrosive or highly various process media or process media with solids content
                            • In the case of heterogeneous and fibrous process media
                            • If good and rapid measuring point cleaning is necessary
                            • If the measuring point is exposed to vibrations
                            • For mounting locations that are difficult to access
Design and operation mode   Diaphragm seals are separating equipment between the measuring system and the process medium.
                            A diaphragm seal system consists of:
                            • A diaphragm seal in a one-sided system, e.g. FMD77 or two diaphragm seals, in a two-sided system, e.g.
                              FMD78
                            • One capillary tube or two capillary tubes
                            • Fill fluid and
                            • A differential pressure transmitter.
                            The process pressure acts via the process isolating diaphragm of the diaphragm seal on the liquid-filled system,
                            which transfers the process pressure via the capillary tube onto the sensor of the differential pressure
                            transmitter.
                            Endress+Hauser delivers all diaphragm seal systems as welded versions. The system is hermetically sealed,
                            which ensures the highest reliability.
                            Note!
                            The correlations between the individual diaphragm seal components are presented in the following section.
                            For further information and comprehensive diaphragm seal system designs, please contact your local
                            Endress+Hauser Sales Center.
                            Diaphragm seal
                            The diaphragm seal determines the application range of the system by
                            • the process isolating diaphragm diameter
                            • the process isolating diaphragm stiffness and material
                            • the design (oil volume).
                            Diaphragm diameter
                            The larger the diaphragm diameter (less stiffness), the smaller the temperature effect on the measurement
                            result.
                            Note: To keep the temperature effect in practice-oriented limits, you should select diaphragm seals with a
                            nominal diameter of DN 80, in as far as the process connection allows for it.
                            Diaphragm stiffness
                            The stiffness is dependent on the process isolating diaphragm diameter, the material, any available coating and
                            on the diaphragm thickness and shape. The diaphragm thickness and the shape are defined constructively. The
                            stiffness of a process isolating diaphragm of the diaphragm seal influences the temperature operating range and
                            the measuring error caused by temperature effects.
                            Capillary
                            Diaphragm seals are used with the following capillary internal diameters as standard:
                            • DN 50: 1 mm
                            • > DN 50: 2 mm
                            The capillary tube influences the TK zero point, the ambient temperature operating range and the response time
                            of a diaphragm seal system as a result of its length and internal diameter.
                             ä 60 ff, sections "Influence of the temperature on the zero point", "Ambient temperature range" and
                            "Response time".
58                                                                                                                       Endress+Hauser
Deltabar S
                                        Observe the installation instructions regarding capillary tubes. See  ä 64 ff, section "Installation
                                        instructions".
                                        Filling oil
                                        When selecting the filling oil, fluid and ambient temperature as well as the operating pressure are of crucial
                                        importance. Observe the temperatures and pressures during commissioning and cleaning. A further selection
                                        criterion is the compatibility of the filling oil with the requirements of the process medium. For this reason,
                                        only filling oils that are harmless to health are used in the food industry, such as vegetable oil or silicone oil. 
                                        See also the following section "Diaphragm seal filling oils".
                                        The filling oil used influences the TK zero point and the temperature operating range of a diaphragm seal system
                                        and the response time.  ä 60 ff, sections "Influence of the temperature on the zero point" and "Response
                                        time".
FMD77: F        Inert oil          –40...+80°C              –40...+175°C            1.87      27           0.000876      0.91          Oil for ultra pure gas
FMD78: D, 4                        (–40...+176°F)           (–40...+347°F)                                                             and oxygen
                                                                                                                                       applications
FMD77: D        Vegetable oil      –10...+120°C             –10...+200°C            0.94      9.5          0.00101       1.05          suitable for foods
FMD78: B, 2                        (+14...+248°F)           (+14...+392°F)                                                             FDA 21 CFR 172.856
FMD77: L        Low                –70...+80°C2             –70...+180°C2           0.92      4.4          0.00108       1.12          low temperatures
FMD78: E, 5     temperature oil    (–94...176°F)            (–94...+356°F)
Endress+Hauser                                                                                                                                              59
                                                                                                                                           Deltabar S
Influence of the temperature   A temperature change results in a volume change of the filling oil. The volume change is dependent on the
on the zero point              coefficient of thermal expansion of the filling oil and on the volume of the filling oil at calibration temperature
                               (constant in the range: +21 to +33°C (+69.8 to 91.4°F)).  ä 59, section "Diaphragm seal filling oils".
                               For example, the filling oil expands in the event of a temperature increase. The additional volume presses
                               against the process isolating diaphragm of the diaphragm seal. The stiffer a diaphragm is, the greater its return
                               force, which counteracts a volume change and acts on the measuring cell together with the operating pressure,
                               thus shifting the zero point. For the "TK Process", see  ä 43 ff, section "Process connections FMD78".
                               The following diagrams display the temperature coefficient "TK Ambient" dependent on the capillary length.
                               The following application is displayed: capillary temperature and transmitter temperature (ambient
                               temperature) change, the process temperature corresponds to the calibration temperature.
                               The temperature coefficients obtained from the diagrams apply to silicone oil and the process isolating
                               diaphragm material AISI 316L/1.4435. For other filling oils, these temperature coefficients must be multiplied
                               by the TK correction factor of the corresponding filling oil. For the TK correction factors, see  ä 59, section
                               "Diaphragm seal filling oils".
                                    TK Ambient      12
                                    [mbar/10 K]                                                                                     7
                                                                                                                                        6
                                                    10
                                                                                                                                       5
                                                     6
                                                                                                                                            4
                                                                                                                                3
                                                     4
                                                                                                                                            2
                                                     2
                                             mbar
                                       1.4                                                                                                 1
                                             10 K
                                                     0
                                                         0    1        2       3       4        5        6       7        8       9        10
                                                                                                                              Capillary length
                                                                                                                                    [m]
                                                                                                                                    P01-FMD78xxx-05-xx-xx-en-004
                               Example for:
                               – Diaphragm seal versions "B5, EN/DIN Flange DN 80 PN 10-40 B1, AISI 316L"
                               – Capillary length: 5 m
                               – Ambient temperature, capillary/transmitter: 45°C
                               – Filling oil: silicone oil
                               1.     Select characteristic curve type for the diaphragm seal versions "B5" in accordance with the following
                                      table.
                                      Result: characteristic curve type 2
                               2.     Obtain value for TK Ambient from the diagram.
                                      Result: 1.4 mbar/10 K
                               3.     TAmbient – TCalibration = 45°C – 25°C = 20°C 1.4 mbar/10 K x 20 K = 2.8 mbar
                               Result: In this application, the zero point is shifted by 2.8 mbar.
                               Note!
                               • The influence of temperature on the zero point can be corrected with position calibration.
                               • The temperature influence can be minimised by using a filling oil with a smaller coefficient of thermal
                                 expansion, shorter capillaries, diaphragm seals with larger diaphragm diameter or by using a smaller capillary
                                 internal diameter.
60                                                                                                                             Endress+Hauser
Deltabar S
                     TK Ambient      90
                     [mbar/10 K]
                                     80
                                     70
                                                                                                           11
                                     60
50
                                     40
                                                                                                                 10
                                     30
                                                                                                                      9
                                     20
                                                                                                                8
                                     10
                                        0
                                            0      1        2         3        4         5        6         7         8         9            10
                                                                                                                           Capillary length
                                                                                                                                 [m]
                                                                                                                               P01-FMD78xxx-05-xx-xx-en-005
Endress+Hauser                                                                                                                                       61
                                                                                                                                          Deltabar S
Ambient temperature range   The operating temperature range of a diaphragm seal system depends on Fill fluid, "Capillary length and
                            internal diameter, Process temperature and Diaphragm seal oil volume.
                            The range of application can be extended by using a fill fluid with a smaller expansion coefficient and a shorter
                            capillary. The permitted operating temperature ranges in relation to the capillary length can be calculated
                            online at "Applicator Sizing Diaphragm Seal":
                            http://www.endress.com/applicator -> Applicator Sizing Diaphragm Seal -> Horncurve
P01-xxxxxxxx-20-xx-xx-en-001
                            Note!
                            • Endress+Hauser recommends you use a low temperature oil for applications that require short response
                              times or are close to the lower temperature limit (see "diaphragm seal fill fluid").
                            • Please contact your Endress+Hauser sales office for further information, comprehensive diaphragm seal
                              system designs and measuring technology solutions that are close to the application limits.
Response time               The viscosity of the filling oil, the capillary length and the capillary internal diameter influence the frictional
                            resistance. The greater the frictional resistance, the longer the response time.
                            Furthermore, the volume change of the measuring cell influences the response time. The lower the
                            volume change of the measuring cell is, the less filling oil has to be shifted in the diaphragm seal system.
                            The following diagram shows typical response times (T90) for the various filling oils dependent on the
                            measuring cell and the capillary internal diameter. The values given are in seconds per metre of capillary length
                            and must be multiplied by the actual length of the capillary. The response time of the transmitter must also be
                            taken into consideration.
62                                                                                                                               Endress+Hauser
Deltabar S
                                                     0
                                                              DN 50      DN 80        DN 50    DN 80         Nominal diameter
                                                             (1 mm)     (2 mm)       (1 mm)   (2 mm)         Capillary internal diameter
                                                                   500 mbar               3 bar              Measuring cell
                                                                                                                                              P01-FMD78xxx-05-xx-xx-en-000
                                                                          100 %
                                                                                        ➀
                                                                           90 %
T90 t
P01-xxxxxxxx-05-xx-xx-xx-006
Larger capillary internal diameter The temperature effect increases with increasing diameter.
Shorter capillaries –
                 Filling oil with lower viscosity                                                 – Observe compatibility of the filling oil with the process
                                                                                                    fluid.
                                                                                                  – Observe the filling oil operating limits.
Endress+Hauser                                                                                                                                                       63
                                                                                                                                              Deltabar S
Selecting the measuring cell (observe the hydrostatic pressure of the filling fluid column in the capillaries!)
                                                           pi
                                                                       –
ΔH = 1 m Hv = 1.8 m
                                                                min.   +
                                                         Hu = 0.2 m
                                                                              H1 = 0.3 m
                                                                                                      + –
Pressure on the negative side of the differential pressure transmitter (p–) when the tank is empty (min. level)
                                                  kg                          kg
                               = 1.8 m • 0.96        • 9.81 m                           m
                                                            s + 0.3 m • 0.96 dm3 • 9.81 s + pi
                                                 dm3
                               = 197.77 mbar + p i
Pressure on the positive side of the differential pressure transmitter (p+) when the tank is empty (min. level)
                                              kg                          kg
                               = 0.2 m • 1       • 9.81 m                           m
                                                        s + 0.3 m • 0.96 dm3 • 9.81 s + pi
                                             dm3
                               = 47.87 mbar + pi
64                                                                                                                                   Endress+Hauser
Deltabar S
                 Result:
                 If the tank were full, a differential pressure of –51.80 mbar would be present at the differential pressure transmitter. When
                 the tank is empty, a differential pressure of –149.90 mbar is present. Therefore, a 500 mbar measuring cell is required for
                 this application.
                 Installation instructions
                 In order to obtain more precise measurement results and to avoid a defect in the device, mount the capillaries
                 as follows:
                 • vibration-free (in order to avoid additional pressure fluctuations)
                 • not in the vicinity of heating or cooling lines
                 • insulate if the ambient temperature is below ore above the reference temperature
                 • with a bending radius of 100 mm.
                 Vacuum applications
                 For applications under vacuum, Endress+Hauser recommends mounting the pressure transmitter underneath
                 the lower diaphragm seal. A vacuum load of the diaphragm seal caused by the presence of filling oil in the
                 capillaries is hereby prevented.
                 When the pressure transmitter is mounted above the lower diaphragm seal, the maximum height difference
                 H1 in accordance with the following illustration on the left must not be exceeded. The maximum height
                 difference is dependent on the density of the filling oil and the smallest ever pressure that is permitted to occur
                 at the diaphragm seal on the positive side (empty tank), see the following illustration, on the right.
                                                                                                                     12.0
                                                                                                                                                                  Low temperature oil
                                                                                                                     10.0
                                                                                          Height difference H1 [m]
                                                                                                                                                                Vegetable oil
                                                                                                                     8.0
                                 –                                                                                                               Silicone oil
                                                                                                                     6.0
                                                                                                                                                                  High temperature
                                         –           +                                                                                                            oil
                                              + –
                                                                                                                     4.0
                                                                 H1                                                                                                    Inert oil
                                                                                                                     2.0
                                 +
                                                                                                                     0.0
                                                                                                                           50 100 200   300   400   500    600      700     800     900 1000
                                                                                                                                        Pressure, diaphragm seal positive side [mbarabs]
P01-FMD7xxxx-11-xx-xx-xx-001 P01-FMD7xxxx-05-xx-xx-en-002
                 Installation above the lower diaphragm seal                       Diagram of maximum installation height above the lower diaphragm
                                                                                   seal for vacuum applications dependent on the pressure at the
                                                                                   diaphragm seal on the positive side
Endress+Hauser                                                                                                                                                                                    65
                                                                                                                                   Deltabar S
Ex approvals                   •   ATEX
                               •   FM
                               •   CSA
                               •   NEPSI
                               •   IECEx
                               •   GOST on request
                               •   also combinations of different approvals
                               All explosion protection data are given in separate documentation which is available upon request. The Ex
                               documentation is supplied as standard with all devices approved for use in explosion hazardous areas.
                                ä 82 ff, sections "Safety Instructions" and "Installation/Control Drawings".
Functional Safety SIL / IEC    The Deltabar S with 4 to 20 mA output signal have been developed to IEC 61508 standard. These devices can
61508 Declaration of confor-   be used for flow, level and differential pressure monitoring up to SIL 3.
mity (optional)                È For a detailed description of the safety functions with Deltabar S, settings and characteristic quantities for
                               functional safety, please refer tot the "Manual for Functional Safety- Deltabar S" SD189.
                               È For devices with SIL / IEC 61508 declaration of conformity see  ä 67 ff, Feature 100 "Additional option
                               1" and Feature 110 "Additional option 2" version E "SIL / IEC 61508, declaration of Conformity".
Overspill protection WHG (German Water Resources Act). See "Ordering information"  ä 67 (see also ZE259P/00/de).
CRN approvals                  Some device versions have CRN approval. For a CRN-approved device, a CRN-approved process connection
                               ( ä 68, feature 70 "Process connection") has to be ordered with a CSA approval ( ä 67, feature 10
                               "Approval"). These devices are fitted with a separate plate bearing the registration number 0F10524.5C.
Pressure Equipment Directive   The devices PMD70, PMD75, FMD76, FMD77 und FMD78 corresponds to Article 3 (3) of the EC directive
(PED)                          97/23/EC (Pressure Equipment Directive) and has been designed and manufactured according to good
                               engineering practice.
                               Additionally applies:
                               – FMD78 with pipe diaphragm seal 1.5"/PN40:
                                 Suitable for stable gases in group 1, category II
                               – PMD75, PN 420
                                 Suitable for stable gases in group 1, category I
66                                                                                                                          Endress+Hauser
Deltabar S
                                                Ordering information
PMD70                                           This overview does not mark options which are mutually exclusive.
                                                10        Approval:
                                                          A   For non-hazardous areas
                                                          1   ATEX II 1/2 G Ex ia IIC T6
                                                          6   ATEX II 1/2 G Ex ia IIC T6, overspill protection WHG
                                                          2   ATEX II 1/2 D
                                                          4   ATEX II 1/3 D
                                 m                        8   ATEX II 1 GD Ex ia IIC T6
                                                          3   ATEX II 1/2 GD Ex ia IIC T6
                                                          7   ATEX II 3 G Ex nA II T6
                                                          S   FM IS, Class I, II, III Division 1, Groups A – G; NI Class I Division 2, Groups A – D; AEx ia
             –   +   E                                    Q   FM DIP, Class II, III Division 1, Groups E – G
                                                          R   FM NI, Class I, Division 2, Groups A – D
                                                          U   CSA IS, Class I, II, III Division 1, Groups A – G; Class I Division 2, Groups A – D, Ex ia
                                                          W   CSA Class II, III Division 1, Groups E – G (Dust-Ex)
                                                          E   Combi-certification
                                                              ATEX II Ex ia + FM IS + CSA IS
                                                              ATEX II 1/2G Ex ia IIC T6 +
                                                              FM/CSA IS Class I, II, III Division 1 Group A - G
                                                          H   NEPSI Ex ia IIC T6
                                                          I   IECEx Zone 0/1 Ex ia IIC T6
                                                20            Output; Operation:
                     n                                        A   4...20 mA HART, SIL operation outside, LCD (see Fig. ➀, n)
                                                                  4...20 mA HART, SIL operation inside, LCD (see Fig. ➀, o)
                                       ➂                      B
                                                              C   4...20 mA HART, SIL operation inside (see Fig. o)
                                                              M   PROFIBUS PA, operation outside, LCD (see Fig. ➀, n)
                         1   2
                                                              N   PROFIBUS PA, operation inside, LCD (see Fig. ➀, o)
                                           PC
Endress+Hauser                                                                                                                                                     67
                                                                                                                   Deltabar S
                    80            Seal:
                                  A    FKM Viton
                                  B    EPDM
                                  D    Kalrez
                                  E    Chemraz
                                  1    FKM Viton, cleaned from oil and greace
                                  2    FKM Viton, cleaned for oxygen service
                                       Note application limits pressure/temp.
68                                                                                                        Endress+Hauser
Deltabar S
Endress+Hauser                                           69
                                                                                                                                                                    Deltabar S
PMD75 This overview does not mark options which are mutually exclusive.
                                                 10        Approval:
                                                           A   For non-hazardous areas
                                                           1   ATEX II 1/2 G Ex ia IIC T6
                                                           6   ATEX II 1/2 G Ex ia IIC T6, overspill protection WHG
                                                           2   ATEX II 1/2 D
                                                           4   ATEX II 1/3 D
                                                           8   ATEX II 1 GD Ex ia IIC T6
                                                           3   ATEX II 1/2 GD Ex ia IIC T6
                                                           5   ATEX II 2 G Ex d IIC T6
                                                           7   ATEX II 3 G Ex nA II T6
                                                           S   FM IS, Class I, II, III Division 1, Groups A – G; NI Class I Division 2, Groups A – D; AEx ia
                                                           T   FM XP, Class I Division 1, Groups A – D; AEx d
                                                           Q   FM DIP, Class II, III Division 1, Groups E – G
                                                           R   FM NI, Class I, Division 2, Groups A – D
                                                           U   CSA IS, Class I, II, III Division 1, Groups A – G; Class I Division 2, Groups A – D, Ex ia
                              m                            V   CSA XP, Class I Division 1, Groups B – D; Ex d
                                                           W   CSA Class II, III Division 1, Groups E – G (Dust-Ex)
                                                           G   NEPSI Exd IIC T6
                                                           H   NEPSI Ex ia IIC T6
         –    +   E
                                                           I   IECEx Zone 0/1 Ex ia IIC T6
                                                           B   Combined certificates: ATEX II 1/2 G Ex ia IIC T6 + II G Ex d IIC T6
                                                           C   Combined certificates: FM IS and XP Class I Division 1, Groups A – D
                                                           D   Combined certificates: CSA IS and XP Class I Division 1, Groups A – D
                                                           E   Combined certificates: FM/CSA IS and XP Class I Division 1, Groups A – D
                                                           F   Combined certificates: ATEX II Ex ia / Ex d + FM/CSA IS + XP; ATEX II 1/2G Ex ia IIC T6+; ATEX II 2G Ex d IIC T6+;
                                                               FM/CSA IS + XP Cl.I Div.1 Gr.A-D
                                                 20            Output; Operation:
                  n                                            A   4...20 mA HART, SIL operation outside, LCD (see Fig. ➀, n)
                                                                   4...20 mA HART, SIL operation inside, LCD (see Fig. ➀, o)
                                        ➂                      B
                                                               C   4...20 mA HART, SIL operation inside (see Fig. o)
                                                               M   PROFIBUS PA, operation outside, LCD (see Fig. ➀, n)
                      1   2
                                                               N   PROFIBUS PA, operation inside, LCD (see Fig. ➀, o)
                                            PC
             T17                                                   N    Aluminium T15 housing, optional display on the top, IP 66/67/NEMA 4X/ 6P, 7/8" FF plug
                                                                   P    Aluminium T15 housing, optional display on the top, IP 65/NEMA 4X, Han7D plug, 90°
                                                                   1    AISI 316L T14 housing, optional display on the side, IP 66/67/NEMA 4X/ 6P, Gland M 20x1.5
                                                                   2    AISI 316L T14 housing, optional display on the side, IP 66/67/NEMA 4X/ 6P, Thread G 1/2
                                                                   3    AISI 316L T14 housing, optional display on the side, IP 66/67/NEMA 4X/ 6P, Thread 1/2 NPT
                                                                   4    AISI 316L T14 housing, optional display on the side, IP 66/67/NEMA 4X/ 6P, M 12x1 PA plug
                                                                   5    AISI 316L T14 housing, optional display on the side, IP 66/67/NEMA 4X/ 6P, 7/8" FF plug
                                                                   6    AISI 316L T14 housing, optional display on the side, IP 65/NEMA 4X, Han7D plug, 90°
                                                                   R    T17 316L Hygiene IP66/68 NEMA6P; M20 gland, T17 = side cover
                                                                   S    T17 316L Hygiene IP66/68 NEMA6P; G1/2 thread, T17 = side cover
                                                                   T    T17 316L Hygiene IP66/68 NEMA6P; NPT1/2 thread, T17 = side cover
                                                                   U    T17 316L Hygiene IP66/68 NEMA6P; M12 plug, T17 = side cover
                                                                   V    T17 316L Hygiene IP66/68 NEMA6P; 7/8" plug, T17 = side cover
                                                                   Z    Housing: see additional specifications
70                                                                                                                                                          Endress+Hauser
Deltabar S
                    50        Calibration; Unit:
                              1   Nominal range; mbar/bar
                              2   Nominal range; kPa/MPa
                              3   Nominal range; mmH2O/mH2O
                              4   Nominal range; inH2O/ftH2O
                              6   Nominal range; psi
                              8   Adjusted for Deltatop; see additional specification
                              B   Customised; see additional specification
                              C   Factory certificate 5-point; see additional specification
                              D   DKD certificate; see additional specification
                              K   Platinum; see additional specification
                              L   Platinum and factory certificate 5-point; see additional specification
                              M   Platinum and DKD certificate; see additional specification
                    80                     Seal:
                                           A    FKM Viton
                                           C    PTFE
                                           F    NBR
                                           K    Copper seal ring, cleaned for oxygen service
                                           1    FKM Viton, cleaned from oil and greace
                                           2    FKM Viton, cleaned for oxygen service
                                                Note application limits pressure/temp.
                                           3    PTFE, cleaned for oxygen service
                                           H    Copper seal ring
Endress+Hauser                                                                                                    71
                                                                                                    Deltabar S
                    995              Identification:
                                     Z1    Measuring point (TAG)
                                     Z2    Bus address
72                                                                                         Endress+Hauser
Deltabar S
FMD76 This overview does not mark options which are mutually exclusive.
                                                 10        Approval:
                                                           A   For non-hazardous areas
                                                           1   ATEX II 1/2 G Ex ia IIC T6
                                                           6   ATEX II 1/2 G Ex ia IIC T6, overspill protection WHG
                                  m                        2   ATEX II 1/2 D Ex ia IIC T6
                                                           8   ATEX II 1 GD Ex ia IIC T6
                                                           3   ATEX II 1/2 GD Ex ia IIC T6
                                                           7   ATEX II 3 G Ex nA II T6
             –   +   E                                     S   FM IS, Class I, II, III Division 1, Groups A – G; NI Class I Division 2, Groups A – D; AEx ia
                                                           R   FM NI, Class I, Division 2, Groups A – D
                                                           U   CSA IS, Class I, II, III Division 1, Groups A – G; Class I Division 2, Groups A – D, Ex ia
                                                           E   Combi-certification
                                                               ATEX II Ex ia + FM IS + CSA IS
                                                               ATEX II 1/2G Ex ia IIC T6 +
                                                               FM/CSA IS Class I, II, III Division 1 Group A - G
                                                           H   NEPSI Ex ia IIC T6
                                                           I   IECEx Zone 0/1 Ex ia IIC T6
                                                 20            Output; Operation:
                     n                                         A   4...20 mA HART, SIL operation outside, LCD (see Fig. ➀, n)
                                                                   4...20 mA HART, SIL operation inside, LCD (see Fig. ➀, o)
                                        ➂                      B
                                                               C   4...20 mA HART, SIL operation inside (see Fig. o)
                                                               M   PROFIBUS PA, operation outside, LCD (see Fig. ➀, n)
                          1   2
                                                               N   PROFIBUS PA, operation inside, LCD (see Fig. ➀, o)
                                            PC
Endress+Hauser                                                                                                                                                      73
                                                                                                          Deltabar S
74                                                                                                    Endress+Hauser
Deltabar S
                    995             Identification:
                                    Z1     Measuring point (TAG)
                                    Z2     Bus address
Endress+Hauser                                                                                                   75
                                                                                                                                                             Deltabar S
FMD77 This overview does not mark options which are mutually exclusive.
                                          10        Approval:
                                                    A   For non-hazardous areas
                                                    1   ATEX II 1/2 G Ex ia IIC T6
                                                    6   ATEX II 1/2 G Ex ia IIC T6, overspill protection WHG
                           m                        2   ATEX II 1/2 D
                                                    4   ATEX II 1/3 D
                                                    8   ATEX II 1 GD Ex ia IIC T6
                                                    3   ATEX II 1/2 GD Ex ia IIC T6
           –
                                                    5   ATEX II 2 G Ex d IIC T6
               +       E
                                                    7   ATEX II 3 G Ex nA II T6
                                                    S   FM IS, Class I, II, III Division 1, Groups A – G; NI Class I Division 2, Groups A – D; AEx ia
                                                    T   FM XP, Class I Division 1, Groups A – D; AEx ia
                                                    Q   FM DIP, Class II, III Division 1, Groups E – G
                                                    R   FM NI, Class I, Division 2, Groups A – D
                                                    U   CSA IS, Class I, II, III Division 1, Groups A – G; Class I Division 2, Groups A – D, Ex ia
                                                    V   CSA XP, Class I Division 1, Groups B – D; Ex d
                                                    W   CSA Class II, III Division 1, Groups E – G (Dust-Ex)
                                                    G   NEPSI Ex d IIC T6
                                                    H   NEPSI Ex ia IIC T6
                                                    I   IECEx Zone 0/1 Ex ia IIC T6
                                                    B   Combined certificates: ATEX II 1/2 G Ex ia IIC T6 + II G Ex d IIC T6
                                                    C   Combined certificates: FM IS and XP Class I Division 1, Groups A – D
                                                    D   Combined certificates: CSA IS and XP Class I Division 1, Groups A – D
                                                    E   Combined certificates: FM/CSA IS and XP Class I Division 1, Groups A – D
                                                    F   Combined certificates: ATEX II Ex ia / Ex d + FM/CSA IS + XP; ATEX II 1/2G Ex ia IIC T6+; ATEX II 2G Ex d IIC T6+;
                                                        FM/CSA IS + XP Cl.I Div.1 Gr.A-D
                                          20            Output; Operation:
               n                                        A   4...20 mA HART, SIL operation outside, LCD (see Fig. ➀, n)
                                                            4...20 mA HART, SIL operation inside, LCD (see Fig. ➀, o)
                               ➂                        B
                                                        C   4...20 mA HART, SIL operation inside (see Fig. o)
                                                        M   PROFIBUS PA, operation outside, LCD (see Fig. ➀, n)
                   1   2
                                                        N   PROFIBUS PA, operation inside, LCD (see Fig. ➀, o)
                                     PC
76                                                                                                                                                   Endress+Hauser
Deltabar S
                    50        Calibration, Units:
                              1   Calibration: nominal range, mbar/bar
                              2   Calibration: nominal range, kPa/MPa
                              3   Calibration: nominal range, mmH2O/mH2O
                              4   Calibration: nominal range, inH2O/ftH2O
                              6   Calibration: nominal range, psi
                              B   Custom calibration: see additional specification
                              C   Factory calibration: see additional specification, Factory calibration certificate, 5-point
                              D   DKD calibration: see additional specification
Endress+Hauser                                                                                                                  77
                                                                                                    Deltabar S
                    995                   Identification:
                                          Z1 Measuring point (TAG)
                                          Z2 Bus address
78                                                                                         Endress+Hauser
Deltabar S
FMD78 This overview does not mark options which are mutually exclusive.
                                                 10        Approval:
                                                           A For non-hazardous areas
                                                           1 ATEX II 1/2 G Ex ia IIC T6
                                                           6 ATEX II 1/2 G Ex ia IIC T6, overspill protection WHG
                                                           2 ATEX II 1/2 D
                                  m                        4 ATEX II 1/3 D
                                                           8 ATEX II 1 GD Ex ia IIC T6
                                                           3 ATEX II 1/2 GD Ex ia IIC T6
                                                           5 ATEX II 2 G Ex d IIC T6
              –   +   E
                                                           7 ATEX II 3 G Ex nA II T6
                                                           S FM IS, Class I, II, III Division 1, Groups A – G; NI Class I Division 2, Groups A – D; AEx ia
                                                           T FM XP, Class I Division 1, Groups A – D; AEx ia
                                                           Q FM DIP, Class II, III Division 1, Groups E – G
                                                           R FM NI, Class I, Division 2, Groups A – D
                                                           U CSA IS, Class I, II, III Division 1, Groups A – G; Class I Division 2, Groups A – D, Ex ia
                                                           V CSA XP, Class I Division 1, Groups B – D; Ex d
                                                           W CSA Class II, III Division 1, Groups E – G (Dust-Ex)
                                                           G NEPSI Ex d IIC T6
                                                           H NEPSI Ex ia IIC T6
                                                           I  IECEx Zone 0/1 Ex ia IIC T6
                                                           B Combined certificates: ATEX II 1/2 G Ex ia IIC T6 + II G Ex d IIC T6
                                                           C Combined certificates: FM IS and XP Class I Division 1, Groups A – D
                                                           D Combined certificates: CSA IS and XP Class I Division 1, Groups A – D
                                                           E Combined certificates: FM/CSA IS and XP Class I Division 1, Groups A – D
                                                           F Combined certificates: ATEX II Ex ia / Ex d + FM/CSA IS + XP; ATEX II 1/2G Ex ia IIC T6+; ATEX II 2G Ex d IIC T6+;
                                                              FM/CSA IS + XP Cl.I Div.1 Gr.A-D
                                                 20            Output; Operation:
                      n                                        A 4...20 mA HART, SIL operation outside, LCD (see Fig. ➀, n)
                                                               B 4...20 mA HART, SIL operation inside, LCD (see Fig. ➀, o)
                                        ➂                      C 4...20 mA HART, SIL operation inside (see Fig. o)
                                                               M PROFIBUS PA, operation outside, LCD (see Fig. ➀, n)
                          1   2
                                                               N PROFIBUS PA, operation inside, LCD (see Fig. ➀, o)
                                            PC
Endress+Hauser                                                                                                                                                               79
                                                                                                                           Deltabar S
                    50        Calibration, Units:
                              1   Calibration: nominal range, mbar/bar
                              2   Calibration: nominal range, kPa/MPa
                              3   Calibration: nominal range, mmH2O/mH2O
                              4   Calibration: nominal range, inH2O/ftH2O
                              6   Calibration: nominal range, psi
                              B   Custom calibration: see additional specification
                              C   Factory calibration: see additional specification, Factory calibration certificate, 5-point
                              D   DKD calibration: see additional specification, DKD-Certificate
80                                                                                                                Endress+Hauser
Deltabar S
                    995                        Identification:
                                               Z1     Measuring point (TAG)
                                               Z2     Bus address
Endress+Hauser                                                                                                     81
                                                                                                                                Deltabar S
                               Additional documentation
Field of Activities            • Pressure measurement: Powerful instruments for process pressure, differential pressure, level and flow:
                                 FA004P/00/en
82                                                                                                                       Endress+Hauser
Deltabar S
Endress+Hauser                                                                                                                    83
                                                                                                                Deltabar S
Instruments International
Endress+Hauser
Instruments International AG
Kaegenstrasse 2
4153 Reinach
Switzerland
TI382P/00/EN/06.09
No. 71095431