Manual - VT 650
Manual - VT 650
Users Manual
FBC - 0099
October 2017, Rev. 1
©2017 Fluke Corporation. All rights reserved. All product names are trademarks of their respective companies.
                                                     Warranty and Product Support
Fluke Biomedical warrants this instrument against defects in materials and workmanship for one year from the date of original purchase OR two
years if at the end of your first year you send the instrument to a Fluke Biomedical service center for calibration. You will be charged our
customary fee for such calibration. During the warranty period, we will repair or at our option replace, at no charge, a product that proves to be
defective, provided you return the product, shipping prepaid, to Fluke Biomedical. This warranty covers the original purchaser only and is not
transferable. The warranty does not apply if the product has been damaged by accident or misuse or has been serviced or modified by anyone
other than an authorized Fluke Biomedical service facility. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE,
ARE EXPRESSED OR IMPLIED. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL
DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY.
This warranty covers only serialized products and their accessory items that bear a distinct serial number tag. Recalibration of instruments is
not covered under the warranty.
This warranty gives you specific legal rights and you may also have other rights that vary in different jurisdictions. Since some jurisdictions do
not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to
you. If any provision of this warranty is held invalid or unenforceable by a court or other decision-maker of competent jurisdiction, such holding
will not affect the validity or enforceability of any other provision.
7/07
Notices
All Rights Reserved
          Copyright 2017, Fluke Biomedical. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval
         system, or translated into any language without the written permission of Fluke Biomedical.
Copyright Release
         Fluke Biomedical agrees to a limited copyright release that allows you to reproduce manuals and other printed materials for use in
         service training programs and other technical publications. If you would like other reproductions or distributions, submit a written
         request to Fluke Biomedical.
Technical Support
         For application support or answers to technical questions, either email techservices@flukebiomedical.com or call 1-800- 850-4608 or
         1-440-248-9300 (Europe +31-40-2965314).
Claims
         Our routine method of shipment is via common carrier, FOB origin. Upon delivery, if physical damage is found, retain all packing
         materials in their original condition and contact the carrier immediately to file a claim. If the instrument is delivered in good physical
         condition but does not operate within specifications, or if there are any other problems not caused by shipping damage, please contact
         Fluke Biomedical or your local sales representative.
Certification
        This instrument was thoroughly tested and inspected. It was found to meet Fluke Biomedical’s manufacturing specifications when it
        was shipped from the factory. Calibration measurements are traceable to the National Institute of Standards and Technology (NIST).
        Devices for which there are no NIST calibration standards are measured against in-house performance standards using accepted test
        procedures.
WARNING
        Unauthorized user modifications or application beyond the published specifications may result in electrical shock hazards or improper
        operation. Fluke Biomedical will not be responsible for any injuries sustained due to unauthorized equipment modifications.
Manufacturing Location
        The VT650/VT900 Gas Flow Analyzer is manufactured at Fluke Biomedical, 6920 Seaway Blvd., Everett, WA, U.S.A.
                                                                             Table of Contents
Title Page
   Introduction .................................................................................................... 1
   Key Features .................................................................................................. 1
   Safety Information .......................................................................................... 2
   Unpacking and Inspection .............................................................................. 5
   Accessories .................................................................................................... 6
   The Analyzer .................................................................................................. 8
        Power On the Analyzer ......................................................................... 10
        Analyzer Connections ........................................................................... 12
             Airway Flow (Inlet and Exhaust) .................................................... 12
             Ultra-Low Flow + and - (VT900) .................................................... 13
             High Pressure ............................................................................... 13
             Low Pressure (+ and -) .................................................................. 14
             Ultra-Low Pressure (VT900 only) .................................................. 14
   Test Setup .................................................................................................... 15
        Bi-directional Flow Mode ....................................................................... 15
                                            i
VT650/VT900
Users Manual
                                                       ii
                                                                                                Contents (continued)
                                        iii
VT650/VT900
Users Manual
               iv
Introduction                                         Key Features
The VT650/VT900 Gas Flow Analyzer (the               • Full range, bi-directional air flow and volume
Analyzer or Product), is a general purpose gas         channel
flow analyzer with special features for testing      • Ultra-low flow and pressure ranges (VT900)
mechanical patient ventilators. The Analyzer
measures bi-directional air flow, high and           • High pressure, vacuum and differential low
differential low pressure, barometric pressure,        pressure
Oxygen concentration and airway pressure,            • Airway pressure, Oxygen concentration,
airway temperature and airway humidity. The            temperature, and humidity
VT900 also measures ultra-low flow (±750 ml/
                                                     • Barometric pressure
min) and ultra-low pressure (0 mbar to 10 mbar).
The Analyzer can be controlled externally using      • External trigger input (VT900)
USB commands or automated with available             • Rechargeable Li-ion battery with up to
software. The Analyzer operates on a                   8 hours battery life
rechargeable Li-Ion battery or external power
supply for stationary or portable use. All figures   • USB port
show the VT900 unless otherwise noted.               • Customizable profiles that can be saved
                                                     • Numerical and graphical screens with real
                                                       time data
                                                     • Automation software available
                                                     • On-board memory
                                                                                                    1
VT650/VT900
Users Manual
A Warning identifies conditions and actions that   • Do not use the Product if it is
pose hazards to the user; a Caution identifies       damaged.
conditions and actions that may damage the         • The battery door must be closed and
Product or the equipment under test.                 locked before you operate the
                 XW Warning                          Product.
                                                   • Recharge the battery when the low
    To prevent possible electrical shock, fire,
                                                     battery indicator shows to prevent
    or personal injury:
                                                     incorrect measurements. Low
    • Read all safety information before             battery will also corrupt the memory
      you use the Product.                           card.
    • Use the Product only as specified, or        • Remove all probes, test leads, and
      the protection supplied by the                 accessories before the battery door
      Product can be compromised.                    is opened.
    • Carefully read all instructions.             • Remove all probes, test leads, and
    • Do not use the Product around                  accessories that are not necessary
      explosive gas, vapor, or in damp or            for the measurement.
      wet environments.                            • Use only specified replacement
    • Use this Product indoors only.                 parts.
    • Do not use the Product if it operates        • Have an approved technician repair
      incorrectly.                                   the Product.
2
                                                                      Gas Flow Analyzer
                                                                       Safety Information
                                                                                       3
VT650/VT900
Users Manual
Symbols used on the Analyzer and in this manual are explained in Table 1.
                                             Table 1. Symbols
              Conforms to the Appliance Efficiency Regulation (California Code of Regulations, Title 20,
               Sections 1601 through 1608), for small battery charging systems.
               This product complies with the WEEE Directive marking requirements. The affixed label indicates
               that you must not discard this electrical/electronic product in domestic household waste. Product
      ~        Category: With reference to the equipment types in the WEEE Directive Annex I, this product is
               classed as category 9 "Monitoring and Control Instrumentation" product. Do not dispose of this
               product as unsorted municipal waste.
4
                                                                                 Gas Flow Analyzer
                                                                            Unpacking and Inspection
                                                                                                  5
VT650/VT900
Users Manual
Accessories
Table 2 is a list of the standard accessories provided with the Analyzer.
6
                                                                                  Gas Flow Analyzer
                                                                                         Accessories
                                                                                                  7
VT650/VT900
Users Manual
The Analyzer
Figure 1 shows the top of the Analyzer.
                                                                                                                        Touchsceen LCD
        1
                                                                                                                        Flow exhaust
    2                                                                                                            3
                                                                                                                     D   Low pressure + and -
                   FLOW
                                                                                                        FLOW
                                                                                                                     E   High pressure and vacuum
                                                                                                        INLET
                  EXHAUST                                                                              ± 300
                                                                                                           lpm
8 7 6 5 4
8
                                                                                       Gas Flow Analyzer
                                                                                             The Analyzer
1 1
6 5 4 3
                                                                                                       9
VT650/VT900
Users Manual
5 4
10
                                                                             Gas Flow Analyzer
                                                                                   The Analyzer
2 3
1 5
12
                                                                                  Gas Flow Analyzer
                                                                                        The Analyzer
                                                                                                    13
VT650/VT900
Users Manual
14
                                                                                                                                                                        Gas Flow Analyzer
                                                                                                                                                                                Test Setup
                                                                                                                                                                            Ventilator
        Test Lung                      FLOW
                                      EXHAUST
                                                                                                                               FLOW
                                                                                                                               INLET
                                                                                                                              ± 300 lpm
Expiratory Hose
                                                                                                                                                                                         15
VT650/VT900
Users Manual
Exhaust Inlet
                                                                                                                                         FLOW
                                                 FLOW                                                                                    INLET
                                                EXHAUST                                                                                 ± 300 lpm
                                                                                                                                                            Ventilator
         Test Lung
                                                           750 ml /min                       0 TO 10 mbar   0.8 TO 10 bar    160 mbar
                                                            ULTRA                            ULTRA            HIGH            LOW
                                                          LOW FLOW                        LOW PRESSURE      PRESSURE        PRESSURE
Expiratory Hose
16
                                                                                                                                                           Gas Flow Analyzer
                                                                                                                                                                   Test Setup
Inspiratory Hose
Exhaust Inlet
                                                                                                                                                           Ventilator
         Test Lung
                                                                        VT900   GAS FLOW ANALYZER
                                                                                                                                        FLOW
                                                FLOW                                                                                    INLET
                                               EXHAUST                                                                                 ± 300 lpm
Expiratory Hose
                                                                                                                                                                          17
VT650/VT900
Users Manual
18
                                                                                  Gas Flow Analyzer
                                                                                         Operations
                                                                                                19
VT650/VT900
Users Manual
20
                                                                                   Gas Flow Analyzer
                                                                                          Operations
                                                                                                 21
VT650/VT900
Users Manual
22
                                                                    Gas Flow Analyzer
                                                                           Operations
        1
                                                            2
         2
                                                                3
                                                                                  23
VT650/VT900
Users Manual
24
                                                                                       Gas Flow Analyzer
                                                                                              Operations
          Setup                                            Description
 Gas                       Select the type of gas. See Table 5.
                           Select the type of Gas correction (temperature and pressure) to different
 Correction Mode
                           gas standards. See Table 6.
                                                                                                       25
VT650/VT900
Users Manual
          Setup                                 Description
                  Sets the options for Breath Detection. Options are:
                  • Mode - Sets the breath detection mode:
                      Bidirectional
                      Uni-directional Inspiratory
                      Uni-directional Expiratory
                      Off - disables breath triggering
                  • Trigger Source - Sets how the breath detection is triggered:
                      Flow - Set the Threshold
 Breath Detect        Pressure - Requires back pressure, such as a test lung
                      External - enables external TTL triggering input on the Analyzer
                         (VT900)
                  • Patient - Sets the type of patient:
                      Adult
                      Pediatric
                  • Threshold - Enter the numeric value for the detection limit.
                      + (Inspiratory threshold)
                      - (Expiratory threshold)
26
                                                                              Gas Flow Analyzer
                                                                                     Operations
         Setup                                  Description
                 Displays basic information about
                 the Analyzer including:              • Date of calibration
Instrument       • Model number                       • Battery charge level
Information
                 • Serial number                      • Memory remaining
                 • Firmware version
                 Options are:
Auto Dim         • On - The LCD dims after 5 minutes of no use. Tap to restore brightness.
                 • Off - The LCD does not auto dim.
                 Options are:
                 • Off
Beeper           • Soft
                 • Medium
                 • Loud
                                                                                            27
VT650/VT900
Users Manual
          Setup                                   Description
                  Options are:
                  • Set Time Format - Select 12 hour with am/pm or 24 hour.
                  • Set Time - Tap the arrows to set the time. If in 12-hour format, select am
 Time and Date      or pm.
                  • Set Date Format - Select MM/DD/YYYY, DD/MM/YYYY, or 
                    YYYY/MM/DD.
                  • Set Date - Tap the arrows to set the date.
                  Options are:
                  • Low
 Brightness
                  • Medium
                  • High
 Language         Select the language.
28
                                                                                                                         Gas Flow Analyzer
                                                                                                                                Operations
                                                                                                                                              29
VT650/VT900
Users Manual
30
                                                                                 Gas Flow Analyzer
                                                                                        Operations
You can change the units of measure for all        Unit                   Description
readings.
                                                                     Volume
To set the units:
                                              l             Liters
1. Select Menus > Units.
                                              ml            Milliliters
2. Select the type of reading.
                                              cf            Cubic feet
3. Select the unit of measure. See Table 7.
                                                                 Temperature
4. Select OK.
        Table 7. Available Units of Measure   °C            Celsius
                       Flow
 lpm          Liters per minute
 lps          Liters per second
 mlpm         Milliliters per minute
 mlps         Milliliters per second
 cfm          Cubic feet per minute
                                                                                                 31
VT650/VT900
Users Manual
32
                                                                                 Gas Flow Analyzer
                                                                                        Operations
Trend Tests show how a particular breath            The test stops when the number of breaths
parameter performs over time. To do a trend         is reached or you select End.
test:                                            The High Frequency Test measures high-
1. Select Menus > Special Tests > Trend          frequency flow and airway pressure.
   Test.                                         1. Select Menus > Special Tests > High
2. Use the arrows to select the parameter to        Frequency Test
   measure.                                      2. Apply the high-frequency flow or pressure.
3. Use the keypad to enter the % limit for       Calibrate Oxygen
   maximum allowed deviation.
                                                 Calibrate the oxygen sensor at the beginning of
4. Tap Start.
                                                 each day that oxygen will be measured, and
The Stacked Volume Test measures the volume      after replacing the sensor.
of multiple breaths. Breath volume determines
                                                 To calibrate the sensor:
the accuracy of the ventilator in accumulating
volume.                                          1. Select Menus > Calibrate Oxygen.
1. Select Menus > Special Tests > Stacked        2. Follow the on-screen instructions. Make sure
   Volume Test.                                     to complete both steps.
2. Use the key pad to enter the number of           If the calibration fails, replace the Oxygen
   breaths.                                         sensor. If the problem persists, contact Fluke
                                                    Biomedical for repair.
3. Select Start.
                                                                                               33
VT650/VT900
Users Manual
34
                                                                                      Gas Flow Analyzer
                                                                      Maintenance, Service and Calibration
                                                                                                      35
VT650/VT900
Users Manual
36
                                                                              Gas Flow Analyzer
                                                              Maintenance, Service and Calibration
Battery Replacement
To check the battery charge level, remove the
                                                  1
battery door and push the TEST button on the
battery. The battery displays the percentage of                             2
charge.
Use only the replacement battery listed in
Table 8.
                                                                 3
To replace the battery (See Figure 9):
1. Loosen the screws on the battery door. (The
   screws do not come out.)
2. Remove the battery door.
3. Use the strap on the battery to lift out the
   battery.
4. Insert the new battery and replace the
   battery door.
                                                                                              37
VT650/VT900
Users Manual
38
                                                                                                Gas Flow Analyzer
                                                                                                     Specifications
                                                                                                                   39
VT650/VT900
Users Manual
40
                                                                                                                                Gas Flow Analyzer
                                                                                                                                     Specifications
                                                                                                                                   3.0 % of rdg or
        Table 9. Airway Flow Range and Accuracy                               Nitrous Oxide (N2O)              ±0.400 slpm
                                                                                                                                   0.02 slpm, typical
             Gas                       Range            Specification          Note: Ultra-low flow that is over range can be detected
                                                                               as unstable. If this occurs, reduce flow.
  Main Airway Flow
                                                                            Flow Temperature
  Air, Nitrogen (N2),                                  1.7 % of rdg or      Coefficient ...............................Add 0.2 % of reading per °C
                                   ±300 slpm                                                                           between 10 °C and 18 °C
  Oxygen (O2)                                          0.04 slpm
                                                                                                                       Add 0.2 % of reading per °C
  Argon, Heliox,                                       3.0 % of rdg or                                                 between 28 °C and 40 °C
                                   ±300 slpm
  O2 bal He, O2 bal N2                                 0.08 slpm, typical
                                                                                                                                                     41
VT650/VT900
Users Manual
Accuracy ............................ ±1 % (VT900), ±2 % (VT650) General .............................. IEC 61010-1: Pollution Degree 2
42
                                                                        Gas Flow Analyzer
                                                                             Specifications
                                                                                       43
VT650/VT900
Users Manual
44
                                                                                                   Gas Flow Analyzer
                                                                                                        Specifications
                                                                                                                   45
VT650/VT900
Users Manual
46