Application: Precise Time-Scaling of Gas Chromatographic Methods Using Method Translation and Retention Time Locking
Application: Precise Time-Scaling of Gas Chromatographic Methods Using Method Translation and Retention Time Locking
Application
Gas Chromatography
May 1998
2
• Split/splitless inlet Results and Discussion Figure 1 shows the method transla-
tion software. The original method
• Automatic liquid sampler
Locking GC-MS with Other GC conditions for the GC-AED pesticide
The GC-AED system also included an Detectors method are entered in the column
Agilent G2350A atomic emission labeled “Original Method.” The
detector with GC-AED ChemStation When using selective GC detectors in column dimensions, carrier gas type,
software (rev B.00.00) for Microsoft® conjunction with GC-MS, one prob- inlet pressure, outlet pressure, ambi-
Windows NT®. lem that is encountered is knowing ent pressure, and oven temperature
the relationship between retention program are entered here. Note that
The GC-micro-ECD system was con- times on the selective detector and the inlet pressure is in psi (gauge),
trolled by Agilent GC ChemStation that of the GC-MS. In GC-MS, the while the outlet pressure and ambient
software (rev A.05.04). Both the outlet pressure of the column is pressure are psi (absolute). The origi-
GC-AED and the GC-micro-ECD vacuum, while with most other GC nal method here is being used on a
ChemStations contained RTL soft- detectors, the outlet pressure of the GC-AED system, so the outlet pres-
ware for GC ChemStation (G2080AA) column is at or near atmospheric sure is entered as atmospheric pres-
and the Retention Time Locking Pes- pressure. This difference in outlet sure plus 1.5 psi, the operating
ticide Library for GC ChemStation pressures results in large differences pressure of the GC-AED.
(G2081AA). in retention time between GC with
MS detection and GC with other The “Criterion” parameter is set to
The GC-MS system (G1723A) used detectors. Comparison of GC-FID, a “None,” which allows the user to
consisted of an 6890 Series GC general detector, with GC-MS is rea- select a specific value of “speed gain”
equipped with an Agilent 5973 mass sonably straightforward, because the by adjusting the value of hold-up time
selective detector (MSD). The total ion chromatogram (TIC) of the for the translated method (see
process for retention time locking the GC-MS system has similar response figure 1). In the column labeled
GC-MS system is described in to the FID. Retention times on the “Translated Method,” the parameters
reference 2. GC-MS system corresponding to of column dimensions, carrier gas
those on the GC-FID can be deter- type, outlet pressure, and ambient
All systems except the micro-ECD
mined by looking for similar patterns pressure for the GC-MS method are
instrument used 30 m ´ 0.25 mm id ´
of response. With selective detectors, entered. Note that the inlet pressure
0.25 mm HP-5MS columns (part no.
this is much more difficult because and oven program are not entered;
19091S-433). The Agilent micro-ECD
the response patterns from selective they are calculated by the program.
instrument used 10 m ´ 0.1 mm id ´
detectors usually do not resemble the To set the speed gain to a desired
0.1 mm HP-5 column (part no.
TIC. For this reason, matching the value, take the calculated value of
19091J-141).
retention times of selective detectors hold-up time in the first column
RTL measurements were made with a precisely with the GC-MS system sim- (0.996060 minute) and divide it by the
solution of dichlorvos, methyl chlor- plifies data analysis greatly. scale factor. Because in this case the
pyrifos, and mirex, each at 10-ppm desired scale factor (“speed gain”) is
In this first example of scaling the 1, the same hold-up time for both the
concentration in acetone. All injec-
RTL Pesticide Library, the method GC-AED and the GC-MS methods is
tions were 1-mL splitless, except for
will be scaled from the GC-AED required. Clicking the radio button
the micro-ECD experiments, which
method to the GC-MS method. In this next to the hold-up time in the “Trans-
were 1-mL split 100:1. In all methods,
case, the desired scale factor is lated Method” column will do this
inlets were operated at 250 °C and
exactly 1, that is, the GC-MS retention automatically.
detectors at 300 °C.
times are desired to be exactly the
Method translation requires inlets to same as those of the GC-AED. The The method translation indicates that
be run in constant pressure mode to first step is to use the method transla- to obtain the same retention times on
obtain precise scaling of retention tion software to determine the GC the GC-MS system as on the GC-AED,
times. Thus, all methods discussed in conditions to use for GC-MS. use all the same method parameters
the note were run in this mode.
3
except inlet pressure. Instead of using
27.6 psi as is used on the GC-AED,
method translation calculates that
17.93 psi on the GC-MS system will
result in matching retention times. As
mentioned above, this inlet pressure
is calculated on the assumed dimen-
sions of the column in the GC-MS
system. To get the retention times to
match precisely, RTL3 is used.
4
• The inlet pressure calculated in Column 4 of table 2 shows the locking The RTL Pesticide Library contains
the “Translated Method” column pressures for the same set of runs but the retention times of the 567 pesti-
will now change to a new value, determined using the GC-MS RTL cal- cides measured with GC-FID. The
corresponding to the pressure ibration points calculated using values measured with the FID would
that would be obtained if the method translation. The calculated be the same observed with any detec-
calibration run were made on a data provide locking pressures that tor that is operated at or near atmos-
GC-MS system. This pressure is agree well with those based on mea- pheric pressure. Because retention
used with the retention time sured data. The range in locking pres- time matching is critical in this appli-
obtained for the corresponding sures pressure is only from 17.72 to cation, the retention times for all the
GC-AED calibration run as a cali-
17.75 psi. This range of 0.03 psi corre- compounds in the table were also
bration point for the GC-MS
sponds to only about a 0.006-minute measured on the GC-MS system after
method.
range in the retention time of methyl scaling as described here. Figure 3 is
When all five points have been chlorpyrifos. a plot of the difference between the
calculated in this way, they are retention times measured on the
Figure 2 shows the locked GC-FID and the GC-MS systems. The
entered into the RTL calibration
chromatograms from a three- plot shows the retention times match
dialog box for the GC-MS method and
component mixture run on GC-AED well within ± 0.1 minute out to 30
saved with the method. Table 1 lists
and GC-MS systems. As can be seen, minutes. A few compounds at the end
the original RTL calibration pressures
the retention times are well matched deviate outside this window, with one
and times with the calculated pres-
between the two methods. compound 0.2-minute different. The
sures and times for the GC-MS
method.
5
deviation is clearly largest in the Gaining Speed in the Same the matrix. This approach can save a
isothermal hold region, which starts Instrument Setup significant amount of analysis time.
at 31.87 minutes. This effect is seen
with GC-MS, but not with scaling to In the analysis of pesticide residues in In this example of scaling the RTL
other atmospheric pressure detectors. food, there are usually only a few Pesticide Library, the method will be
While the cause is not yet clearly compounds encountered in any one increased in speed at the expense of
understood, it appears related to the sample. Because the screening chromatographic resolution. The first
vacuum outlet pressure of the GC-MS method uses selective detectors, it consideration is by what factor to
column. Although this level of match- makes sense to consider trading increase the speed. The method trans-
ing is very good, the table includes speed for chromatographic resolu- lation software is useful for determin-
both the GC-FID and GC-MS retention tion. Selective detectors respond to ing this. A candidate speed gain, in
times so that smaller time windows only those compounds containing a this example threefold, is entered into
can be used in searching unknowns. specific heteroatom(s), and the chro- the method translation software. The
matography only needs to resolve resulting inlet pressure and oven tem-
Locking GC-AED with Other GC those compounds from each other, perature ramp rates are then
Detectors not from every other compound in inspected to see if the instrument on
6
which the new method will be run is GC-AED method. This is done by the Table 4 compares the locking pres-
compatible with those parameters. same process as shown in the GC-MS sures determined with measured and
scaling above. In this case, when one with calculated RTL calibration
Figure 4 shows the method transla- of the original method RTL calibra- points. As in the above GC-MS exam-
tion software with the data entered tion pressures is entered, the result- ple, the range of the locking pressures
for a speed gain of 3. Note that ing holdup time must be divided by 3 from the calculated data is only
columns for “Original Method” and and entered for the holdup time in the 0.11 psi (87.88 to 87.99), which
“Translated Method” are set up as in “Translated Method” column. This corresponds to ~ 0.003 minute.
the previous example with two excep- will force the “speed gain” back to 3.
tions. Because the scaling is from The resulting inlet pressure is then Figure 5 compares the chro-
GC-AED to GC-AED, the outlet pres- paired with the retention time of the matograms of the RTL locking mix-
sure in both columns is entered as corresponding original GC-AED cali- ture from both the original and the 3´
16.2 psi. The second and most signifi- bration run, but divided by 3 as a cali- scaled methods. Note that while the
cant difference is the holdup time. bration point for the new method. chromatographic resolution is
The desired “speed gain” is 3. reduced, the speed is increased by a
Table 3 shows the RTL calibration factor of 3.
To set the speed gain, the calculated points from the original GC-AED
value of hold-up time in the first method and calculated points for the Figure 6 shows a plot of the differ-
column (0.996060 minute) is divided threefold speed gain (3´) method. ence between the RTL Pesticide
by exactly 3. This value Library retention times, divided by 3,
(0.33202 minute) is entered for the When the calibration data is entered and those of the 3´ method. The data
hold-up time in the second column. into the RTL calibration dialog box, were taken with a 36-component
This will force the speed gain to the target time for methyl chlorpyri- subset of the library. The plot shows
exactly 3. fos is entered as 5.532 minutes, which the retention times match well within
is 16.596 minutes divided by 3. ± 0.05 minute for all compounds, even
The inlet pressure and oven tempera-
ture ramp for the new threefold speed
method are now calculated. The cal-
culated inlet pressure is 87.862 psi,
which is compatible with the EPC
module on the current system (maxi-
mum 100 psi). Note that the helium
source supplying the GC must be
capable of reaching 100 psi of helium.
An optional 150-psi EPC module is
available for the HP 6890 GC to pro-
vide additional inlet pressure, if
necessary.
The next step is to calculate the RTL Figure 4. Method translation software showing scaling RTL Pesticide method scaled to
calibration points from the original threefold faster method.
7
those in the 3.3-minute hold time at Table 3. RTL Calibration Points from Original GC-AED Method and
the end of the run. Calculated Points for Threefold Speed Gain (3´) Method
8
Figure 7 shows the method transla- 0.2
tion from the GC-AED method to the
0.1-mm id column with a scale factor 0.15
of 3. A speed gain of 3 was again
chosen based on oven and inlet limi- 0.1
tations as described above. The same
scaling process as used above is 0.05
Difference (min)
followed.
0
The RTL calibration points for 0 2 4 6 8 10 12 14
the new 3´ 0.1-mm micro-ECD -0.05
method were both calculated with
-0.1
method translation and measured.
Table 5 shows the calculated values.
-0.15
When the locking pressures from the
measured and calculated values were -0.2
9
Table 7 shows a comparison of lock- Table 5. RTL Calibration Points from Original GC-AED Method and
ing pressures calculated using mea- Calculated Points for 3´ 0.1-mm id Micro-ECD Method
sured and predicted 3´ 0.1-mm id Assuming 10-m Column Length
micro-ECD calibration data. The
GC-AED RTL Calibration 3x Micro-ECD RTL Calibration
range of locking pressures from the
Calculated Calculated
measured data (66.03 to 65.93) only Pressure Ret Time Pressure Ret Time
corresponds to a spread in retention (psi) (min) (psi) (min)
times of about 0.004 minute. How- 33.1 15.346 71.03 5.115
ever, with the data calculated based 30.4 15.919 64.90 5.306
on a 10-m assumed length, the spread 27.6 16.578 58.51 5.526
(66.38 to 63.18) is much larger and 24.8 17.338 52.11 5.779
would correspond to a time range of 22.1 18.242 45.91 6.081
0.14 minute. The locking pressures
calculated using the 10.5622 value are
much more consistent with the mea-
sured values. The range in retention Table 6. RTL Calibration Points from Original GC-AED Method and
Calculated Points for 3´ 0.1-mm id Micro-ECD Method
times would be ~ 0.03 minute if all the
Assuming 10.5622-m Column Length
calculated points are used, and if the
first value in column 5 is ignored, the GC-AED RTL Calibration 3x Micro-ECD RTL Calibration
range drops to ~ 0.005 minute. Calculated Calculated
Pressure Ret Time Pressure Ret Time
(psi) (min) (psi) (min)
The fact that the agreement in locking
33.1 15.346 80.03 5.115
pressures is much improved by using
30.4 15.919 73.13 5.306
10.56 m instead of 10 m suggests that
27.6 16.578 65.95 5.526
length is probably the largest contrib-
24.8 17.338 58.74 5.779
utor to the discrepancy. These results
22.1 18.242 51.75 6.081
should reinforce the recommendation
that if a method is to be used exten-
sively, it is prudent to obtain mea-
sured RTL calibration data. It should Table 7. Comparison of Locking Pressures Calculated Using Measured and
be noted, however, that even with the Predicted 3´ 0.1-mm id Micro-ECD Calibration Data
RTL calibration from the 10-m 3x Micro-ECD Locking Runs Locking Pressures
assumed length, the worst conse- Measured 3x Micro-ECD RTL Using Measured Using 10-m Calculated Using 10.56-m Calculated
quence would be that the RT locking Cal Points RTL Cal Points RTL Cal Points RTL Cal Points
step would need to be repeated an Pressure Ret Time Pressure Pressure Pressure
extra time to get a more precise (psi) (min) (psi) (psi) (psi)
match. 48.81 6.323 65.95 66.38 65.30
52.66 6.041 66.03 65.77 65.85
Figure 8 compares the chromato- 58.51 5.797 65.95 65.12 65.96
grams of the RTL locking mixture 64.36 5.585 65.93 64.36 65.95
from both the original and the 70.22 5.396 66.00 63.18 65.90
3 ´ 0.1-mm id micro-ECD methods.
10
Note that while the most of the References
chromatographic resolution is pre- 1. P. L. Wylie and B. D. Quimby, “A
served, the speed is increased by a Method Used to Screen for 567
factor of 3. Pesticides and Suspected
Endocrine Disrupters,”
After being locked, the three peaks in Hewlett-Packard Company, Appli-
the 3´ micro-ECD method had reten- cation Note 228-402, Publication
tion times of 1.924, 5.533, and 9.963 5967-5860E, April 1998.
minutes, respectively. These values
2. M. Klee and V. Giarrocco, “Pre-
are very close to the RTL Pesticide
dictable Translation of Capillary
Library retention times for the three
GC Methods for Fast GC,”
compounds divided by 3: 1.932, 5.532,
Hewlett-Packard Company, Appli-
and 9.949. The fact that the largest
cation Note 228-373, Publication
difference between the scaled table
5965-7673E, March 1997.
and the 3´ micro-ECD method is only
0.014 minute again demonstrates the 3. V. Giarrocco, B. D. Quimby, and
precision of retention time matching M. S. Klee,“Retention Time Lock-
achievable with the scaling technique ing: Concepts and Applications,”
described here. Hewlett-Packard Company, Appli-
cation Note 228-392, Publication
5966-2469E, December 1997.
Conclusions
4. Capillary Column Method Transla-
Using method translation combined tor, user contributed software,
with retention time locking provides a free download from:
means of extending the usefulness of www.hp.com/go/mts.
existing capillary GC methods. The
ability to precisely scale a method to
meet the needs of different samples
and instrument types greatly reduces
the effort required to re-use methods,
thus saving time and money.
3
1 GC-AED (1x)
2
0 5 10 15 20 25 30 35 40 min
GC-micro-ECD (3x)
0 2 4 6 8 10 12 min
Figure 8. Chlorine chromatogram from 1´ GC-AED method (top) and 3´ micro-ECD method
(bottom) of three-component locking mixture. Peak identifications: 1. dichlorvos, 2.
methyl chlorpyrifos, 3. mirex.
11
Agilent shall not be liable for errors contained herein or for
incidental or consequential damages in connection with the
furnishing, performance, or use of this material.
Copyright© 2000
Agilent Technologies, Inc.