UltraRAE 3000

User’s Guide

Ex-Ox-Tox Gasdetectie
Westerdreef 5V
Rev. C 2152 CS Nieuw-Vennep
Telefoon: 0252 620885
August 2010 E-mail: info@exoxtox.nl
P/N 059-4023-000 Website: www.exoxtox.nl
Downloaded from www.Manualslib.com manuals search engine
FCC Information

Contains FCC ID: PI4411B

The enclosed device complies with part 15 of the FCC rules.


Operation is subject to the following conditions: (1) This device may
not cause harmful interference, and (2) This device must accept any
interference received, including interference that may cause undesired
operation.

© Copyright 2010 RAE Systems, Inc.

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

Contents
1 Standard Contents...................................................................9
2 General Information................................................................9
3 Physical Description .............................................................11
4 Specifications........................................................................11
5 Replacing Alkaline Batteries ................................................14
6 Charging A Lithium-Ion Battery ..........................................16
6.1 Charging A Spare Battery ............................................18
6.2 Low Voltage Warning..................................................18
6.3 Clock Battery ...............................................................19
6.4 Data Protection While Power Is Off ............................19
7 User Interface .......................................................................20
7.1 Display .........................................................................22
8 Operating The Instrument.....................................................23
8.1 Turning The Instrument On (Simple) ..........................23
8.2 Turning The Instrument On (Power On Zero) ............24
8.3 Turning The Instrument Off.........................................24
8.4 Operating The Built-In Flashlight................................24
8.5 Pump Status .................................................................25
8.6 Calibration Status.........................................................26
9 Operating Modes ..................................................................27
10 Compound-Specific Operation .............................................31
10.1 Compound-Specific Measurement...............................31
10.2 Measurement Phases ....................................................31
10.3 Performing A Measurement.........................................31
10.4 Separation Tube Preparation........................................32
10.5 Inserting The Separation Tube.....................................34
10.6 Measuring ....................................................................35
11 VOC Operation.....................................................................38
11.1 Basic User Level/Hygiene Mode (Default Settings)....38
12 Alarm Signals .......................................................................41
12.1 Alarm Signal Summary................................................42
12.2 Preset Alarm Limits & Calibration ..............................43
12.3 Testing The Alarm .......................................................43
13 Integrated Sampling Pump ...................................................43
14 Backlight...............................................................................44
15 Datalogging ..........................................................................44
15.1 Datalogging Event .......................................................44

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

15.2 Datalogging Sample.....................................................44


15.3 Auto/Manual/Snapshot Datalogging............................45
16 Accessories ...........................................................................45
17 Standard Kit & Accessories..................................................46
17.1 AC Adapter (Battery Charger).....................................46
17.2 External Filter ..............................................................47
18 Optional Accessories ............................................................48
18.1 Calibration Adapter......................................................48
18.2 Calibration Regulator...................................................48
18.3 Organic Vapor Zeroing Kit ..........................................48
19 Standard Two-Point Calibration (Zero & Span)...................49
19.1 Entering Calibration.....................................................50
19.2 Zero (Fresh Air) Calibration ........................................51
19.3 Span Calibration...........................................................52
19.4 Exiting Two-Point Calibration In Basic User Level ....55
20 Three-Point Calibration ........................................................56
20.1 Span 2 Calibration........................................................58
20.2 Exiting Three-Point Calibration...................................59
21 Programming Mode ..............................................................60
21.1 Entering Programming Mode.......................................60
22 Programming Mode Menus ..................................................61
22.1 Exiting Programming Mode.........................................63
22.2 Navigating Programming Mode Menus.......................63
22.3 Calibration....................................................................64
22.3.1 Zero Calibration...................................................................64
22.3.2 Span Calibration...................................................................64
22.4 Measurement................................................................65
22.4.1 Meas. Gas.............................................................................65
22.4.2 Meas. Unit ............................................................................66
22.4.3 Tube Selection......................................................................67
22.5 Alarm Setting ...............................................................68
22.5.1 High Alarm ..........................................................................69
22.5.2 Low Alarm ...........................................................................69
22.5.3 STEL Alarm.........................................................................70
22.5.4 TWA Alarm .........................................................................71
22.5.5 Alarm Mode.........................................................................72
22.5.6 Buzzer & Light ....................................................................73
22.6 Datalog.........................................................................73
22.6.1 Clear Datalog .......................................................................74

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

22.6.2 Interval..................................................................................74
22.6.3 Data Selection ......................................................................75
22.6.4 Datalog Type........................................................................76
22.6.5 Manual Datalog....................................................................76
22.6.6 Snapshot Datalog.................................................................78
22.7 Monitor Setup ..............................................................78
22.7.1 Radio Power.........................................................................78
22.7.2 Op Mode...............................................................................79
22.7.3 Site ID...................................................................................79
22.7.4 User ID .................................................................................80
22.7.5 User Mode............................................................................81
22.7.6 Date.......................................................................................81
22.7.7 Time......................................................................................82
22.7.8 Duty Cycle ...........................................................................82
22.7.9 Temperature Unit.................................................................83
22.7.10 Pump Speed .......................................................................83
22.7.11 Language............................................................................84
22.7.12 Real Time Protocol............................................................84
22.7.13 Power On Zero...................................................................85
22.7.14 Unit ID................................................................................85
22.7.15 LCD Contrast.....................................................................86
22.7.16 Lamp ID .............................................................................86
23 Humidity Compensation.......................................................87
24 Hygiene Mode ......................................................................87
24.1 Basic User Level & Hygiene Mode .............................88
24.2 Entering Search Mode From Hygiene Mode ...............90
24.3 Optional Graphic Screen In Search Mode ...................91
25 Advanced User Level (Hygiene Mode Or Search Mode).....92
25.1 Advanced User Level & Hygiene Mode......................92
25.2 Basic User Level & Search Mode................................94
25.3 Advanced User Level & Search Mode ........................96
25.4 Diagnostic Mode..........................................................97
25.4.1 Entering Diagnostic Mode ..................................................97
25.4.2 Adjusting The Pump Stall Threshold.................................98
25.4.3 Pump High ...........................................................................98
25.4.4 Pump Low............................................................................98
25.4.5 Testing The Humidity Sensor.............................................99
25.4.6 Exiting Diagnostic Mode ................................................. 100
26 Transferring Data To & From A Computer........................102

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

26.1 Downloading The Datalog To A PC..........................102


26.2 Uploading Firmware To The instrument From A PC 103
27 Maintenance........................................................................104
27.1 Battery Charging & Replacement ..............................104
27.1.1 Replacing The Li-ion Battery .......................................... 105
27.1.2 Replacing The Alkaline Battery Adapter........................ 105
27.2 PID Sensor & Lamp Cleaning/Replacement...................107
27.3 Cleaning The PID Sensor...........................................108
27.3.1 Cleaning The Lamp Housing Or Changing The Lamp . 108
27.3.2 Determining The Lamp Type .......................................... 109
27.3.3 Sampling Pump................................................................. 110
27.3.4 Testing The T.H.P. Sensor ............................................... 110
27.3.5 Cleaning The Instrument.................................................. 110
27.3.6 Ordering Replacement Parts ............................................ 110
27.4 Special Servicing Note...............................................111
28 Troubleshooting..................................................................112
29 Technical Support...............................................................113
30 RAE Systems Contacts .......................................................114
31 Controlled Part Of Manual ..................................................117

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

Read Before Operating


This manual must be carefully read by all individuals who have or
will have the responsibility of using, maintaining, or servicing this
product. The product will perform as designed only if it is used,
maintained, and serviced in accordance with the manufacturer’s
instructions. The user should understand how to set the correct
parameters and interpret the obtained results.

CAUTION!
To reduce the risk of electric shock, turn the power off before
removing the instrument cover. Disconnect the battery before
removing sensor module for service. Never operate the instrument
when the cover is removed. Remove instrument cover and sensor
module only in an area known to be non-hazardous.

ATEX WARNING!
To reduce the risk of electrostatic ignition, do not use the instrument
without the rubber boot in place.

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

Special Notes
When the instrument is taken out of the transport case and
turned on for the first time, there may be some residual organic
or inorganic vapor trapped inside the detector chamber. The
initial PID sensor reading may indicate a few ppm. Enter an
area known to be free of any organic vapor and turn on the
instrument. After running for several minutes, the residual
vapor in the detector chamber will be cleared and the reading
should return to zero.

The battery of the instrument discharges slowly even if it is


turned off. If the instrument has not been charged for 5 to 7
days, the battery voltage will be low. Therefore, it is a good
practice to always charge the instrument before using it. It is
also recommended to fully charge the instrument for at least
10 hours before first use. Refer to this User Guide’s section on
battery charging for more information on battery charging and
replacement.

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

WARNINGS
STATIC HAZARD: Clean only with damp cloth.

For safety reasons, this equipment must be operated and serviced by


qualified personnel only. Read and understand the instruction manual
completely before operating or servicing. Use only RAE Systems
battery packs, part numbers 059-3051-000, 059-3052-000, and 059-
3054-000. This instrument has not been tested in an explosive gas/air
atmosphere having an oxygen concentration greater than 21%.
Substitution of components may impair intrinsic safety. Recharge
batteries only in non-hazardous locations.

Do not mix old and new batteries or batteries from different


manufacturers.

The calibration of all newly purchased RAE Systems instruments


should be tested by exposing the sensor(s) to known concentration
calibration gas before the instrument is put into service.

For maximum safety, the accuracy of the instrument should be


checked by exposing it to a known concentration calibration gas
before each day’s use.

Do not use USB/PC communication in hazardous locations.

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

AVERTISSEMENT
DANGER
RISQUE D'ORIGINE ELECTROSTATIQUE: Nettoyer
uniquement avec un chiffon humide.
Pour des raisons de sécurité, cet équipement doit être utilisé,
entretenu et réparé uniquement par un personnel qualifié. Étudier le
manuel d’instructions en entier avant d’utiliser, d’entretenir ou de
réparer l’équipement.
Utiliser seulement l'ensemble de batterie RAE Systems, référence
059-3051-000 ou 059-3052-000 ou 059-3054-000. Cet instrument n’a
pas été testé dans une atmosphère de gaz/air explosive ayant une
concentration d’oxygène plus élevée que 21%. La substitution de
composants peut compromettre la sécurité intrinsèque. Ne charger les
batteries que dans des emplacements désignés non-dangereux.
Ne pas mélanger les anciennes et les nouvelles batteries, ou bien
encore les batteries de différents fabricants.
L’étalonnage de tout instrument de RAE Systems doit être testé en
exposant l’instrument à une concentration de gaz connue avant de
mettre en service l’instrument pour la première fois.

Pour une sécurité maximale, la sensibilité de l’instrument doit être


vérifiée en exposant l’instrument à une concentration de gaz connue
avant chaque utilisation journalière.

Ne pas utiliser de connexion USB/PC en zone dangereuse.

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

1 Standard Contents
Instrument
Calibration Kit
Charging Cradle
AC/DC Adapter
Alkaline Battery Adapter
Data Cable
CD-ROM With User’s Guide, Quick Start Guide, and related materials

2 General Information
The UltraRAE 3000 is a hand-held, programmable compound specific
PID monitor designed to provide instantaneous exposure monitoring
of a specific organic gas. It monitors a specific gas by utilizing a gas
separation tube and the photoionization detector (PID) with a 9.8 eV
gas discharge lamp. It also can be used to measure total volatile
organic compound (VOC) as a broadband monitor by utilizing the
PID with a 9.8 eV, 10.6 eV, or 11.7 eV lamp.
Features include:

Lightweight and Compact


• Compact, lightweight, rugged design
• Built-in sample draw pump
Dependable and Accurate
• Up to 16 hours of continuous monitoring with rechargeable
battery pack
• Designed to continuously monitor VOC vapor at parts-per-
million (ppm) levels
User-friendly
• Preset alarm thresholds for STEL, TWA, low- and high-level
peak values.
• Audio buzzer and flashing LED display are activated when the
limits are exceeded.
Datalogging Capabilities
• 260,000-point datalogging storage capacity for data download to PC

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

The UltraRAE 3000 consists of a PID with associated microcomputer


and electronic circuit. The unit is housed in a rugged case with a
backlit LCD and 3 keys to provide easy user interface. It also has a
built-in flashlight for operational ease in dark locations.

10

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

3 Physical Description
The main components of the portable compound-specific and
VOC monitoring instrument include:
• Three operation/programming keys for normal operation or
programming
• LCD display with back light for direct readout and calculated
measurements
• Built-in flashlight for illuminating testing points in dark
environments
• Buzzer and red LEDs for alarm signaling whenever exposures
exceed preset limits
• Charge contacts for plugging directly to its charging station
• Easy-to-use separation tube holder
• USB communication port for PC interface
• Protective rubber cover

4 Specifications
Size: 10" L x 3" W x 2.5" H
(25.5 cm x 7.6 cm x 6.4 cm)
Weight: 26 oz (738 g) with battery pack
Detector: Photoionization sensor with 9.8, 10.6, or
11.7 eV UV lamp
Battery: A 4.2V rechargeable Lithium-Ion battery
pack (snap in, field replaceable, at non-
hazardous location only)
Alkaline battery holder (for 4 AA batteries)
Battery Charging: Less than 8 hours to full charge
Operating Hours: Up to16 hours continuous operation
Display: Large dot matrix screen with backlight

11

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

Measurement range & resolution

Lamp Range Resolution


10.6 eV 50 ppb to 10,000 ppm 50 ppb
(VOC mode)
9.8 eV 50 ppb to 200 ppm, for 50 ppb
benzene and butadiene;
50 ppb to 5,000 ppm
(VOC mode)
11.7 eV 50 ppb to 2,000 ppm 50 ppb
(VOC mode)

Response time (T90): 2 seconds


Accuracy 3% at calibration point
(Isobutylene):
PID Detector: Easy access to lamp and sensor for cleaning
and replacement
Correction Factors: Over 200 VOC gases built in (based on RAE
Systems Technical Note TN-106)
Calibration: Two-point field calibration of zero and
standard reference gases
Calibration Reference: Store up to 8 sets of calibration data, alarm
limits and span values
Inlet Probe: Flexible 5" tubing (a short tube is also
available)
Separation tube housing with permanent
VOC tube
Radio module: Bluetooth (2.4GHz)
Keypad: 1 operation key and 2 programming keys; 1
flashlight switch
Direct Readout: Instantaneous, average, STEL, TWA and
peak value, and battery voltage

12

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

Intrinsic Safety: US and Canada: Class I, Division 1, Groups


A, B, C, D
Europe: ATEX (0575 Ex II 2G Ex ia
IIC/IIB T4 Gb)
KEMA 07 ATEX 0127
Complies with EN60079-0:2009,
EN60079-11:2007
IECEx CSA 10.0005 Ex ia IIC/IIB T4 Gb
Complies with IEC 60079-0:2007,
IEC 60079-11:2006
(IIC: 059-3051-000 Li-ion bat pack
or 059-3054-000 NiMH bat pack;
IIB: 059-3052-000 alkaline bat pack)
EM Interference: Highly resistant to EMI/RFI.
Compliant with EMC R&TTE (RF Modules)
Alarm Setting: Separate alarm limit settings for Low, High,
STEL and TWA alarm
Operating Mode: Hygiene or Search mode
Alarm: Buzzer 95dB at 12″ (30cm) and flashing red
LEDs to indicate exceeded preset limits, low
battery voltage, or sensor failure
Alarm Type: Latching or automatic reset
Real-time Clock: Automatic date and time stamps on
datalogged information
Datalogging: 260,000 points with time stamp, serial
number, user ID, site ID, etc.
Communication: Upload data to PC and download instrument
setup from PC via USB on charging station.
Sampling Pump: Internally integrated. Flow rate: 450 to 550
cc/min.
Temperature: -20º C to 50º C (-4º to 122º F)
Humidity: 0% to 95% relative humidity (non-
condensing)
Housing (including Polycarbonate, splashproof and dustproof
rubber boot): Battery can be changed without removing
rubber boot.

13

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

5 Replacing Alkaline Batteries


An alkaline battery adapter is supplied with each instrument. The
adapter (part number 059-3052-000) accepts four AA alkaline
batteries (use only Duracell MN1500) and provides approximately 12
hours of operation. (An optional rechargeable lithium-ion battery
pack, part number 059-3051-000, is also available.)
To install the adapter in the instrument:
1. Remove the alkaline battery adapter from the instrument by
sliding the tab and tilting out the adapter.

2. Replace the batteries (follow the procedure below).


3. Tilt the alkaline battery adapter and put it into the instrument.
4. Slide the tab back into place to secure the battery adapter.

To insert batteries into the adapter:


1. Remove the three hex-socket screws to open the compartment
in the adapter.

14

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

2. Insert four fresh AA batteries as indicated by the polarity (+/-)


markings.

3. Replace the cover. Replace the three screws.

IMPORTANT!
Alkaline batteries cannot be recharged. The instrument’s internal
circuit detects alkaline batteries and will not allow recharging. If you
place the instrument in its Travel Charger or Charger Stand, the
alkaline battery will not be recharged. The internal charging circuit is
designed to prevent damage to alkaline batteries and the charging
circuit when alkaline batteries are installed inside the instrument. If
you try to charge an alkaline batteries installed in the instrument, the
Charging Cradle or Travel Charger’s charging LED does not glow,
indicating that it will not charge the alkaline batteries.

Note: When replacing alkaline batteries, dispose of old ones properly.

15

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

6 Charging A Lithium-Ion Battery


Always fully charge the battery before using the instrument. The
instrument’s Li-ion battery is charged by attaching the instrument to
the Travel Charger (or by placing the instrument in the optional
Charger Stand). Contacts on the bottom of the instrument meet the
Travel Charger’s (or Charger Stand’s) contacts, transferring power
without other connections.

Travel
Charger

Charger Stand

Note: Before connecting the charger to the instrument, visually


inspect the contacts to make sure they are clean. If they are not, wipe
them with a soft cloth. Do not use solvents or cleaners.

Follow this procedure to charge the instrument:


1. Plug the AC/DC adapter’s barrel connector into the
instrument’s Charger Stand or Travel Charger.

2. Plug the AC/DC adapter into the wall outlet.


3. Connect the AC/DC adapter to the Travel Charger (or
Charger Stand).

16

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

4. Place the instrument into the Travel Charger or Charger Stand.


The LED in the Travel Charger (or Charger Stand) should
glow.

The instrument begins charging automatically. (If the optional


Charger Stand is used, the “Primary” LED blinks green to
indicate charging.) During charging, the diagonal lines in the
battery icon on the instrument’s display are animated and you see the
message “Charging...”

Note: If the Li-ion battery has been discharged below a certain


threshold, the “Charging...” message does not display immediately.
The charging LED blinks to indicate that it is charging, and after it
has been charging for a while, the “Charging...” message appears.

When the instrument’s battery is fully charged, the battery icon is no


longer animated and shows a full battery. The message “Fully
charged!” is shown. (If the Charger Stand or Travel Charger is used,
its LED glows continuously green.)

Note: If you see the “Battery Charging Error” icon (a battery


outline with an exclamation mark inside), check that the
instrument or rechargeable battery has been properly set into
the Travel Charger (or Charger Stand). If you still receive the
message, check the Troubleshooting section of this guide.

Note: If the instrument or battery has been charging for more than 10
hours and you see the “Battery Charging Error” icon and a message
that says, “Charging Too Long,” this indicates that the battery is not
reaching a full charge. Try changing the battery and make sure the
contacts on the instrument are meeting the Travel Charger’s (or
Charger Stand’s) contacts. If the message is still shown, consult your
distributor or RAE Systems Technical Services.

17

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

6.1 Charging A Spare Rechargeable Battery


(Optional Charger Stand Only)

A rechargeable Li-ion battery can be charged when it is not inside the


monitor. The Charger Stand is designed to accommodate both types
of charging. Contacts on the bottom of the battery meet the contacts
on the Charger Stand, transferring power without other connections,
and a spring-loaded capture holds the battery in place during
charging.

1. Plug the AC/DC adapter into the Charger Stand.


2. Place the battery into the Charger Stand, with the gold-plated
contacts on top of the six matching charging pins.
3. Plug the AC/DC adapter into the wall outlet.

The battery begins charging automatically. During charging, the


Secondary LED in the Charger Stand blinks green. When charging is
complete, it glows steady green.

Release the battery from the Charger Stand by pulling it back toward
the rear of the Charger Stand and tilting it out of its slot.

Note: If you need to replace the Li-ion battery pack, replacements are
available from RAE Systems. The part number is 059-3051-000.

WARNING!
To reduce the risk of ignition of hazardous atmospheres, recharge
and replace batteries only in areas known to be non-hazardous.

6.2 Low Voltage Warning


When the battery’s charge falls below a preset voltage, the
instrument warns you by beeping once and flashing once every
minute, and the “empty battery” icon blinks on and off once
per second. You should turn off the instrument within 10
minutes and either recharge the battery by placing the

18

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

instrument in its cradle, or replace the battery with a fresh one with a
full charge.

6.3 Clock Battery


An internal clock battery is mounted on one of the instrument’s
printed circuit boards. This long-life battery keeps settings in memory
from being lost whenever the Li-ion battery or alkaline batteries are
removed. This backup battery should last approximately five years,
and must be replaced by an authorized RAE Systems service
technician. It is not user-replaceable.

6.4 Data Protection While Power Is Off


When the instrument is turned off, all the current real-time data
including last measured values are erased. However, the datalog data
is preserved in non-volatile memory. Even if the battery is
disconnected, the datalog data will not be lost.

19

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

7 User Interface
The instrument’s user interface consists of the display, LEDs, an
alarm transducer, and four keys. The keys are:

Y/+
MODE
N/-
Flashlight on/off

The LCD display provides visual feedback that includes the reading,
time, battery condition, and other functions.

LEDs and flashlight

Display

Y/+ key
N/- key

MODE key
Flashlight
on/off key

In addition to their labeled functions, the keys labeled Y/+, MODE,


and N/- act as “soft keys” that control different parameters and make
different selections within the instrument’s menus. From menu to

20

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

menu, each key controls a different parameter or makes a different


selection.

Three panes along the bottom of the display are “mapped” to the
keys. These change as menus change, but at all times the left pane
corresponds to the [Y/+] key, the center pane corresponds to the
[MODE] key, and the right pane corresponds to the [N/-] key. Here
are three examples of different menus with the relationships of the
keys clearly shown:

21

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

7.1 Display

The display shows the following information:

Gas info Tells the Correction Factor and type of


calibration gas
Reading Concentration of gas as measured by the
instrument
Calibration needed Indicates that calibration should be
performed
Radio power Indicates whether radio connection is on or
off
Radio signal Indicates signal strength in 5-bar bargraph
Battery Indicates battery level in 3 bars
Pump Indicates that pump is working
Datalog Indicates whether datalog is on or off
Y/+ Y/+ key’s function for this screen
MODE MODE key’s function for this screen
N/- N/- key’s function for this screen

22

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

8 Operating The Instrument


The instrument is designed as a broadband VOC gas monitor and
datalogger for work in hazardous environments. It gives real-time
measurements and activates alarm signals whenever the exposure
exceeds preset limits. Prior to factory shipment, the instrument is
preset with default alarm limits and the sensor is pre-calibrated with
standard calibration gas. However, you should test the instrument and
verify the calibration before the first use. After the instrument is fully
charged and calibrated, it is ready for immediate operation.

8.1 Turning The Instrument On (Simple)

1. With the instrument turned off, press and hold [MODE].


2. When the display turns on, release the [MODE] key.

The RAE Systems logo should appear first. (If the logo does not
appear, there is likely a problem and you should contact your
distributor or RAE Systems Technical Support.) The instrument is
now operating and performs self tests. If any tests (including sensor
and memory tests) fail, refer to the Troubleshooting section of this
guide.

Once the startup procedure is complete, the instrument shows a


numerical reading screen with icons. This indicates that the
instrument is fully functional and ready to use.

23

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

8.2 Turning The Instrument On


(Power On Zero)
If your UltraRAE 3000 has been configured to perform a zero (fresh
air) calibration upon startup, called Power On Zero, then the startup
routine is interrupted so that you can perform a fresh air calibration.
(See page 85 for details on turning this feature on or off.)

If you do not want to perform a zero calibration, press [MODE] to


bypass it. If you start a zero calibration and want to abort it, press
[N/-], and the calibration stops and the main display is shown.

8.3 Turning The Instrument Off


1. Press and hold the Mode key for 3 seconds. A 5-second
countdown to shutoff begins.
2. When you see “Unit off...” release your finger from the
[MODE] key. The instrument is now off.

Note: You must hold your finger on the key for the entire shutoff
process. If you remove your finger from the key during the
countdown, the shutoff operation is canceled and the instrument
continues normal operation.

8.4 Operating The Built-In Flashlight


The instrument has a built-in flashlight that helps you point the probe
in dark places. Press the flashlight key to turn it on. Press it again to
turn it off.

Note: Using the flashlight for extended periods shortens the battery’s
operating time before it needs recharging.

24

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

8.5 Pump Status


IMPORTANT!
During operation, make sure the probe inlet and the gas outlet are free
of obstructions. Obstructions can cause premature wear on the pump,
false readings, or pump stalling. During normal operation, the pump
icon alternately shows inflow and outflow as shown here:

During duty cycling (PID lamp cleaning), the display shows these
icons in alternation:

If there is a pump failure or obstruction that disrupts the pump, the


alarm sounds and you see this icon blinking on and off:

If you see this blinking icon, consult the Troubleshooting section of


this guide.

25

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

8.6 Calibration Status


The instrument displays this icon if it requires calibration:

Calibration is required (and indicated by this icon) if:

• The lamp type has been changed (for example, from 10.6 eV
to 9.8 eV).
• The sensor module has been replaced.
• It has been 30 days or more since the instrument was last
calibrated.
• If you have changed the calibration gas type without
recalibrating the instrument.

26

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

9 Operating Modes
The UltraRAE is actually two monitors in one:

• Compound-specific monitor
• VOC monitor

As a compound-specific monitor, it takes timed measurements and


uses a separation tube in conjunction with software that enables the
UltraRAE 3000 to give specific readings on one particular type of
compound, such as benzene or butadiene.

As a VOC monitor, the UltraRAE 3000 operates in different modes.


In some cases, you can change modes using a password and using the
instrument’s navigation. In other cases, you must use ProRAE Studio
software.

The following two sections cover operation in the two modes.

• Compound Specific, 31.


• VOC, page 38.

The diagram on the next page shows the basic flow of the UltraRAE
3000’s functions. The area with the gray field is the compound-
specific (tube) mode, while the rest shows VOC mode. Navigate
through the steps by using the [Y/+] and [N/-] keys as shown in the
diagram.

Note: If you use a password to access Programming Mode (see page


60), then the navigation changes slightly, entering part of Advanced
Hygiene Mode’s settings, as shown on page 29.

27

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

Compound-Specific (Tube)

Note: Dashed line indicates automatic progression.

28

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

29

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

The default setting for your instrument is:

User Mode: Basic


Operation Mode: Hygiene

This is outlined in detail on page 88.

The other options, covered later in this guide, are:

User Mode: Advanced (page 92)


Operation Mode: Hygiene

User Mode: Advanced (page 96)


Operation Mode: Search

Using ProRAE Studio allows access to other options. In addition,


Diagnostic Mode (page 97) is available for service technicians.

30

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

10 Compound-Specific Operation

10.1 Compound-Specific Measurement


The UltraRAE 3000 can perform compound-specific measurement in
addition to general VOC measurement. This requires using a RAE-
Sep separation tube (butadiene or benzene) and having the UltraRAE
3000 in Tube Mode, operating with a 9.8eV lamp.

10.2 Measurement Phases


To perform a compound-specific measurement, follow this order:

1. UltraRAE 3000 is ready for sampling


2. Prepare the separation tube
3. Insert the separation tube
4. Start measurement
5. UltraRAE 3000 displays and logs measurement
6. Remove the separation tube

10.3 Performing A Measurement


Before performing a compound-specific measurement for Benzene or
Butadiene using a RAE-Sep separation tube, make sure the UltraRAE
3000 is in Tube Mode and that the appropriate tube type is selected.
The UltraRAE 3000 only acts as a compound-specific measurement
device when it is equipped with a 9.8eV lamp. The UltraRAE 3000 is
designed to auto-sense the lamp type. It can also be manually set to
default to a 9.8eV lamp type.

Make sure the UltraRAE 3000 is set to operate with your selected tube:

1. Enter Programming Mode.


2. Select Measurement.
3. Select Tube Selection.
4. Make a choice of Benzene or Butadiene.
5. Save your choice.

31

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

To begin measuring, turn on the UltraRAE 3000. This screen is


shown, which includes the CF (correction factor) and measurement
gas type for calibration reference:

Press [N/-] to advance. You will see this screen:

Do not begin sampling yet!

Before you start sampling, you must insert a RAE-Sep separation tube
into the inlet/holder. Follow the Separation Tube Preparation and Placing
A Tube Into The UltraRAE 3000 instructions before pressing any
buttons on the UltraRAE 3000. Once the tube is in place, then proceed to
measuring.

IMPORTANT!
Once a tube’s ends are broken off, the material inside is exposed.
Therefore, use the tube for sampling as soon as possible.

10.4 Separation Tube Preparation

CAUTION!
Wear hand and eye protection when breaking tube tips. Use caution
in handling tubes with broken ends. Keep away from children. RAE-
Sep tubes should be disposed of according to local regulations. See
footnotes of data sheets for disposal information.

32

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

1. Open a package of RAE-Sep separation tubes and remove one.


2. Place the tip in the package’s tube tip breaker (the small hole on
the front) and snap off the tip.
3. Turn the tube around and snap off the other end.

CAUTION!
Handle tubes with care. Tube ends are sharp after ends are broken off.

33

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

10.5 Inserting The Separation Tube

1. Unscrew the front of the sampling probe from the base.


2. Slip the tube into the rubber holder in the front portion. Make
sure the arrow on the side of the tube points toward the
instrument.
3. Insert the other end of the tube into the middle of the base while
turning the front portion to tighten it onto the base’s threads.

IMPORTANT!
Do not overtighten any portion of the sampling assembly.

Note: When the UltraRAE 3000 is used for VOC monitoring, no tube
is inserted.

IMPORTANT!
To ensure that there are no leaks, periodically test the seals:

With the UltraRAE 3000 running, place your finger over the end of
the inlet probe. The alarm should sound and the pump-stall icon
should flash on the display. This indicates that all seals are good. Stop
the alarm by pressing [Y/+]. If the pump does not alarm or show the
stalled-pump icon, then check that all inlet parts are tight and inspect
the O-ring for damage (replace it if necessary).

34

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

10.6 Measuring
Once the tube is in place, begin measuring by pressing [Y/+].

The display shows a countdown (60 seconds is shown here, but sampling
time depends on the type of separation tube selected and the temperature):

Note: You can abort the sampling by pressing [N/-] at any time.
Once the countdown is complete, the reading is shown:

Press [Y/+] to continue sampling with the tube for 15 minutes to


establish a STEL reading, or press [N/-] to return to the main menu.

WARNING!
At least 1/4 of the tube should still be yellow-orange at the bottom. If
not, the STEL value is not valid. Abort the measurement and change
the tube. Then do a snapshot test instead of a STEL test. Note: If the
STEL is exceeded, the UltraRAE 3000 goes into alarm.

35

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

If you press [N/-] to return to the main menu, which shows the tube
type instead of the CF (correction factor):

Press [N/-] to advance to this screen:

If you press [Y/+], you are asked, “Clear peak value! Are You Sure?”
to confirm:

Press [Y/+] to clear the Peak value and exit to VOC operation.

If you press [N/-], this display is shown:

36

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

Remove the tube and put the inlet back together. Then press [N/-].
This display is shown:

After a few seconds, the UltraRAE 3000 enters VOC mode and
shows this display:

You can step through the rest of the steps by pressing [N/-] repeatedly
until you reach the main menu again.

37

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

11 VOC Operation
11.1 Basic User Level/Hygiene Mode
(Default Settings)

The instrument is programmed to operate in Basic User Level/Hygiene Mode


as its default. This gives you the most commonly needed features while
requiring the fewest parameter adjustments.

Pressing [N/-] steps you from one screen to the next, and eventually return to
the main display. If you do not press a key within 60 seconds after entering a
display, the instrument reverts to its main display.

Note: While viewing any of these screens, you can shut off your instrument by
pressing [MODE].

38

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

Note: Dashed line indicates automatic progression.

39

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

After the instrument is turned on, it runs through the start-up menu. If
the UltraRAE 3000 is set for “Power On Zero,” then the message
“Apply zero gas…” is displayed.

At this point, you can perform a zero air (fresh air) calibration. If the
ambient air is clean, you can use that. Otherwise, use a cylinder of
zero air. Refer to Zero Calibration on page 51 for a more detailed
description of zero calibration.
Start zero calibration by pressing Start [Y/+]. You see the message
“Zeroing…” followed by a 30-second countdown.
Note: You can press [MODE] to quit, bypassing the zero air
calibration.
When zero calibration is complete, you see the message:
Zeroing is done!
Reading = 0.0 ppm
The instrument is now sampling and collecting data.

Note: At the Average & Peak, Date & Time & Temperature, Calibration
Gas & Measurement Gas & Correction Factor, and PC Communications
screens, the instrument automatically goes to the main display after 60
seconds if you do not push a key to make a selection.

40

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

12 Alarm Signals
During each measurement period, the gas concentration is compared
with the programmed alarm limits (gas concentration alarm limit
settings). If the concentration exceeds any of the preset limits, the
loud buzzer and red flashing LED are activated immediately to warn
you of the alarm condition.
In addition, the instrument alarms if one of the following conditions
occurs: battery voltage falls below a preset voltage level, failure of
the UV lamp, or pump stall.

41

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

12.1 Alarm Signal Summary


Message Condition Alarm Signal
HIGH Gas exceeds “High 3 beeps/flashes per second*
Alarm” limit
OVR Gas exceeds 3 beeps/flashes per second*
measurement range
MAX Gas exceeds electronics’ 3 beeps/flashes per second*
maximum range
LOW Gas exceeds “Low 2 beeps/flashes per second*
Alarm” limit
TWA Gas exceeds “TWA” 1 Beep/flash per second*
limit
STEL Gas exceeds “STEL” 1 Beep/flash per second*
limit
Pump Pump failure 3 beeps/flashes per second
icon
flashes
Lamp PID lamp failure 3 beeps/flashes per second
plus “Lamp” message on
display
Battery Low battery 1 flash, 1 beep per minute
icon plus battery icon flashes on
flashes display
CAL Calibration failed, or 1 beep/flash per second
needs calibration
NEG Gas reading measures 1 beep/flash per second
less than number stored in
calibration
* Hygiene mode only. In Search mode, the number of beeps per
second (1 to 7) depends upon the concentration of the sampled gas.
Faster rates indicate higher concentrations.

42

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

12.2 Preset Alarm Limits & Calibration


The instrument is factory calibrated with standard calibration gas, and
is programmed with default alarm limits.

Cal Gas Cal unit Low High TWA STEL


Span
Isobutylene 100 ppm 50 100 10 25
Benzene 5 ppm 2 5 0.5 2.5
Butadiene 10 ppm 5 10 2 5

12.3 Testing The Alarm


You can test the alarm whenever the main (Reading) display is
shown. Press [Y/+], and the audible and visible alarms are tested.

13 Integrated Sampling Pump


The instrument includes an integrated sampling pump. This
diaphragm-type pump that provides a 450 to 550 cc per minute flow
rate. Connecting a Teflon or metal tubing with 1/8" inside diameter to
the gas inlet port of the instrument, this pump can pull in air samples
from 200' (61 m) away horizontally, or 90' (27.5 m) vertically, at
about 3' (0.9 m) per second flow speed.
Note: In Search Mode, the pump turns on when a sample
measurement is started, and turns off when the sample is manually
stopped.
If liquid or other objects are pulled into the inlet port filter, the
instrument detects the obstruction and immediately shuts down the
pump. The alarm is activated and a flashing pump icon is displayed.
You should acknowledge the pump shutoff condition by clearing the
obstruction and pressing the [Y/+] key while in the main reading
display to restart the pump.

43

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

14 Backlight
The LCD display is equipped with an LED backlight to assist in
reading the display under poor lighting conditions.

15 Datalogging
During datalogging, the instrument displays a disk icon to indicate
that datalogging is enabled. The instrument stores the measured gas
concentration at the end of every sample period (when data logging is
enabled). In addition, the following information is stored: user ID,
site ID, serial number, last calibration date, and alarm limits. All data
are retained (even after the unit is turned off) in non-volatile memory
so that it can be down- loaded at a later time to a PC.
15.1 Datalogging Event
When Datalogging is enabled, measurement readings are being saved.
These data are stored in “groups” or “events.” A new event is created
and stored each time the instrument is turned on and is set to
automatic datalogging, or a configuration parameter is changed, or
datalogging is interrupted. The maximum time for one event is 24
hours or 28,800 points. If an event exceeds 24 hours, a new event is
automatically created. Information, such as start time, user ID, site
ID, gas name, serial number, last calibration date, and alarm limits are
recorded.
15.2 Datalogging Sample
After an event is recorded, the unit records a shorter form of the data.
When transferred to a PC running ProRAE Studio, this data is
arranged with a sample number, time, date, gas concentration, and
other related information.

44

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

15.3 Auto/Manual/Snapshot Datalogging


The instrument has three datalog types:

Auto Default mode. Collects datalog information when the


instrument is sampling.
Manual Datalogging occurs only when the instrument’s
datalogging is manually started (see page 63 for
details).
Snapshot Datalogs only during snapshot (single-event capture,
initiated by pressing [MODE]) sampling. See page 65
for details.

Note: You can only choose one datalog type to be active at a time.

16 Accessories
The following accessories are included with the instrument:
• AC Adapter (Battery Charger)
• Travel Charger
• Alkaline battery adapter
• External Filter
• Organic Vapor Zeroing kit

Hard-case kits also include these accessories:


• Calibration gas, if specified
• Calibration adapter
• Calibration regulator and flow controller
• Charging Cradle (instead of Travel Charger)

45

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

17 Standard Kit & Accessories


17.1 AC Adapter (Battery Charger)

WARNING
To reduce the risk of ignition of hazardous atmospheres, recharge
battery only in area known to be non-hazardous. Remove and
replace battery only in area known to be non-hazardous.
Ne charger les batteries que dans emplacements designés non-
dangereuses.
A battery charging circuit is built into the instrument cradle. It only needs a
regular AC to 12 VDC adapter (wall-mount transformer, part number 500-
0114-000) to charge the instrument.
To charge the battery inside the instrument:
1. Power off the instrument.
2. Connect the AC adapter to the DC jack on the instrument’s cradle.
If the instrument is off, it automatically turns on.
3. While charging, the display message shows “Charging.” The
Primary LED on the cradle flashes green when charging.
4. When the battery is fully charged, the LED changes to glowing
green continuously, and the message “Fully charged” appears on
the display. If there is a charging error, the LED glows red
continuously.
A completely discharged instrument can be charged to full capacity
within 8 hours. Batteries drain slowly even if an instrument is off.
Therefore, if the instrument has been in storage or has not been
charged for several days or longer, check the charge before using it.
The factory-supplied battery is designed to last for 16 hours of normal
operation (no alarm), for a new battery under the optimum
circumstances. As the battery becomes older or is subject to adverse
conditions (such as cold ambient temperature), its capacity will be
significantly reduced.

46

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

17.2 External Filter


The external filter is made of PTFE (Teflon®) membrane with a 0.45
micron pore size to prevent dust or other particles from being sucked
into the sensor manifold, which would cause extensive damage to the
instrument. It prolongs the operating life of the sensor. To install the
external filter, simply connect it to the instrument’s inlet probe.

47

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

18 Optional Accessories
18.1 Calibration Adapter
The calibration adapter for the instrument is a simple 6-inch Tygon
tubing with a metal adapter on one end. During calibration, simply
insert the metal adapter into the regular gas inlet probe of the
instrument and the tubing to the gas regulator on the gas bottle.
18.2 Calibration Regulator
The Calibration Regulator is used in the calibration process. It
regulates the gas flow rate from the Span gas cylinder into the gas
inlet of the instrument during calibration process. The maximum flow
rate allowed by the flow controller is about 0.5L/min (500 cc per
min.). Alternatively, a demand-flow regulator or a Tedlar gas bag
may be used to match the pump flow precisely.
18.3 Organic Vapor Zeroing Kit
The Organic Vapor Zeroing Kit is used for filtering organic air
contaminants that may affect the zero calibration reading. To use the
Organic Vapor Zeroing Kit, simply connect the filter to the inlet port
of the instrument.

48

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

19 Standard Two-Point Calibration


(Zero & Span)
The following diagram shows the instrument’s calibrations in
Basic/Hygiene
Hygiene mode.

Note: Dashed line indicates automatic progression.

49

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

19.1 Entering Calibration

1. Press and hold [MODE] and [N/-] until you see the Password
screen.

2. In Basic User Level, you do not need a password to perform


calibrations. Instead of inputting a password, enter calibration
by pressing [MODE].

Note: If you inadvertently press [Y/+] and change any of the


numbers, simply press [MODE] and you will be directed to
the calibration menu.

The Calibration screen is now visible with Zero Calibration


highlighted.

These are your options:

• Press [Y/+] to select the highlighted calibration (Zero Calib


or Span Calib).
• Press [MODE] to exit calibration and return to the main
display and resume measurement.
• Press [N/-] to toggle the highlighted calibration type.

50

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

19.2 Zero (Fresh Air) Calibration

This procedure determines the zero point of the sensor calibration


curve. To perform a fresh air calibration, use the calibration adapter to
connect the instrument to a “fresh” air source such as from a cylinder
or Tedlar bag (optional accessory). The “fresh” air is clean, dry air
without organic impurities and an oxygen value of 20.9%. If such an
air cylinder is not available, any clean ambient air without detectable
contaminants or a charcoal filter can be used.
At the Zero Calibration menu, you can proceed to perform a Zero
calibration or bypass Zero calibration and perform a Span calibration.
You may also go back to the initial Calibration menu if you want to
exit calibration.

• Press [Y/+] to start calibration.


• Press [MODE] to quit and return to the main calibration
display.

If you have pressed [Y/+] to enter Zero calibration, then you will see
this message:

1. Turn on your Zero calibration gas.


2. Press [Y/+] to start calibration.

Note: At this point, you may press [MODE] if you decide


that you do not want to initiate calibration. This will take you
directly to the Calibration menu, highlighted for Span
calibration.

51

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

3. Zero calibration starts a 30-second countdown and displays


this message:

Zeroing...

During the zeroing process, the instrument performs the Zero


calibration automatically and does not require any action on your part.

Note: To abort the zeroing process at any time and proceed to Span
calibration, press [N/-] at any time while zeroing is being performed.
You will see a confirmation message that says “Zero aborted!” and
then the Span calibration menu appears.

When Zero calibration is complete, you see this message:

Zeroing is done!
Reading = 0.0 ppm

The instrument will then show the Calibration menu on its display,
with Span Calib highlighted.

19.3 Span Calibration


This procedure determines the second point of the sensor calibration
curve for the sensor. A cylinder of standard reference gas (span gas)
fitted with a 500 cc/min. flow-limiting regulator or a flow-matching
regulator is the simplest way to perform this procedure. Choose the
500 cc/min. regulator only if the flow rate matches or slightly exceeds
the flow rate of the instrument pump. Alternatively, the span gas can
first be filled into a Tedlar bag or delivered through a demand-flow
regulator. Connect the calibration adapter to the inlet port of the
instrument, and connect the tubing to the regulator or Tedlar bag.
Another alternative is to use a regulator with >500 cc/min flow but
allow the excess flow to escape through a T or an open tube. In the
latter method, the span gas flows out through an open tube slightly
wider than the probe, and the probe is inserted into the calibration
tube.

52

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

At the Span Calibration menu, you perform a Span calibration. You


may also go back to the Zero calibration menu or to the initial
Calibration menu if you want to exit calibration.

• Press [Y/+] to enter Span calibration.


• Press [N/-] to skip Span calibration and return to Zero
calibration.
• Press [MODE] to exit Span calibration and return to the top
calibration menu.

If you have pressed [Y/+] to enter Span calibration, then you will see
the name of your Span gas (the default is isobutylene) and the span
value in parts per million (ppm).

IMPORTANT!
If you are using the UltraRAE 3000 to test for benzene, it is
recommended that you calibrate with 5 ppm benzene calibration gas
from RAE Systems.

You will also see this message that prompts you:

1. Turn on your span calibration gas.


2. Press [Y/+] to initiate calibration.

Note: You may press [MODE] if you decide that you do not
want to initiate calibration. This will abort the span
calibration and take you directly to the Calibration menu for
Zero calibration.

3. Span calibration starts and displays this message:

Calibrating...

53

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

During the Span calibration process, there is a 30-second countdown


and the instrument performs the Span calibration automatically. It
requires no actions on your part.

Note: If you want to abort the Span calibration process, press [N/-] at
any time during the process. You will see a confirmation message that
says “Span is aborted!” and then the Zero calibration menu appears.
You can then proceed to perform a Zero calibration, perform a Span
calibration, or exit to the topmost Calibration menu.

When Span calibration is complete, you see a message similar to this


(the value is an example only):

Span 1 is done!
Reading = 100.0 ppm

The instrument then exits Span calibration and shows the Zero
calibration menu on its display.

Note: The reading should be very close to the span gas value.

54

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

19.4 Exiting Two-Point Calibration In Basic


User Level
When you are done performing calibrations, press [MODE], which
corresponds with “Back” on the display. You will see the following
message:

Updating settings…

The instrument updates its settings and then returns to the main
display. It begins or resumes monitoring.

55

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

20 Three-Point Calibration
For enhanced accuracy, it is possible to perform a second Span
calibration in addition to the Zero and Span calibrations outlined in
the previous section. Your instrument first must be set to allow this
third calibration. This requires using ProRAE Studio software and a
PC, as well as a higher concentration of calibration gas.

Note: Once the third calibration is set, you do not need to use
ProRAE Studio to allow future 3-point calibrations. Also, you can
only disable 3-point calibration capability by using ProRAE Studio
again.

Perform the Zero and Span calibrations. After the first Span
calibration (Span 1) is completed, the display a second Span
calibration (Span 2) can be performed. The process is identical to the
first calibration. As in the Span 1 calibration, you may exit and return
to the Zero calibration screen if you choose not to perform this
calibration or to abort it.

56

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

Note: Dashed line indicates automatic progression.

57

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

20.1 Span 2 Calibration


A cylinder of standard reference gas (span gas) fitted with a 500
cc/min. flow-limiting regulator or a flow-matching regulator is the
simplest way to perform this procedure.

Note: This gas should be of a higher concentration than the gas used
for Span 1 calibration.

Choose the 500 cc/min. regulator only if the flow rate matches or
slightly exceeds the flow rate of the instrument pump. Alternatively,
the span gas can first be filled into a Tedlar bag or delivered through a
demand-flow regulator. Connect the calibration adapter to the inlet
port of the instrument, and connect the tubing to the regulator or
Tedlar bag.
Another alternative is to use a regulator with >500 cc/min flow but
allow the excess flow to escape through a T or an open tube. In the
latter method, the span gas flows out through an open tube slightly
wider than the probe, and the probe is inserted into the calibration
tube.
At the Span Calibration menu, you perform a Span calibration. You
may also go back to the Zero calibration menu or to the initial
Calibration menu if you want to exit calibration.

• Press [Y/+] to enter Span 2 calibration.


• Press [N/-] to skip Span calibration and return to Zero
calibration.
• Press [MODE] to exit Span calibration and return to the top
calibration menu.

If you have pressed [Y/+] to enter Span calibration, then you will see
the name of your Span gas (the default is isobutylene) and the span
value in parts per million (ppm). You will also see this message that
prompts you:

Please apply gas...

4. Turn on your span calibration gas.

58

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

5. Press [Y/+] to initiate calibration.

Note: You may press [MODE] if you decide that you do not
want to initiate calibration. This will take you directly to the
Calibration menu for Zero calibration.

6. Span calibration starts a 30-second countdown and displays


this message:

Calibrating...

During the Span calibration process, the instrument performs the


Span calibration automatically and does not require any action on
your part.

Note: If you want to abort the Span calibration process, press [N/-] at
any time during the process. You will see a confirmation message that
says “Span is aborted!” and then the Zero calibration menu will
appear. You can then proceed to perform a Zero calibration, perform
a Span calibration, or exit to the topmost Calibration menu.

When Span calibration is complete, you will see a message similar to


this (the value shown here is for example only):

Span 2 is done!
Reading = 1000 ppm

The instrument then exits Span calibration and shows the Zero
calibration menu on its display.

Note: The reading should be very close to the span gas value.

20.2 Exiting Three-Point Calibration


When you are done performing calibrations, press [MODE], which
corresponds with “Back” on the display. You will see the following
message:

Updating settings…

59

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

The instrument updates its settings and then returns to the main
display. It begins or resumes monitoring.

21 Programming Mode

Programming Mode can be entered from either Hygiene Mode or


Search Mode. If the current user mode is Basic, you must provide a 4-
digit password to enter.

21.1 Entering Programming Mode

1. Press and hold [MODE] and [N/-] until you see the Password
screen.

2. Input the 4-digit password:

• Increase the number from 0 through 9 by pressing [Y/+].


• Step from digit to digit using [N/-].
• Press [MODE] when you are done.

If you make a mistake, you can cycle through the digits by pressing
[N/-] and then using [Y/+] to change the number in each position.

Note: The default password is 0000.

60

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

When you have successfully entered Programming Mode, you see


this screen:

Note: The password can only be changed by connecting the


instrument to a PC running ProRAE Studio software. Follow the
instructions in ProRAE Studio to change it.The Calibration label is
shown and its icon is highlighted, but you can press [N/-] to step from
one programming menu to the next, with the name of the menu shown
at the top of the display and the corresponding icon highlighted. As
you repeatedly press [N/-], the selection moves from left to right, and
you see these screens:

Note: When you reach Monitor Setup and press [N/-], the menu
cycles back to Calibration.

22 Programming Mode Menus


The Programming Mode allows anyone with the password to change
the instrument’s settings, calibrate the instrument, modify the sensor
configuration, enter user information, etc. Programming Mode has
five menus. Each menu includes several sub-menus to perform
additional programming functions.

61

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

This table shows the menus and sub-menus:

Calibration Measurement Alarm Datalog Monitor


Setting Setup
Zero Meas. Gas High Clear Radio
Calibration Alarm Datalog Power
Span Meas. Unit Low Interval Op Mode
Calibration Alarm
Tube Selection STEL Data Site ID
Alarm Selection
TWA Datalog User ID
Alarm Type
Alarm User Mode
Mode
Buzzer Date
& Light
Time
Pump Duty
Cycle
Pump Speed
Temperature
Unit
Language
Real Time
Protocol
Power On
Zero
Unit ID
LCD
Contrast
Lamp ID

62

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

Once you enter Programming Mode, the LCD displays the first menu,
Calibration. Each subsequent menu is accessed by pressing [N/-]
repeatedly until the desired menu is displayed. To enter a sub-menu
of a menu, press [Y/+].

22.1 Exiting Programming Mode


To exit Programming Mode and return to normal operation, press
[MODE] once at any of the programming menu displays. You will
see “Updating Settings…” as changes are registered and the mode
changes.
22.2 Navigating Programming Mode Menus
Navigating through the Programming Mode menus is easy and
consistent, using a single interface format of “Select,” “Back” and
“Next” at the top level. The three control buttons correspond to these
choices as shown:

Note: Pressing [MODE] in the Programming Mode’s top level causes


the instrument to exit Programming Mode and return to monitoring.

The three keys perform the following functions in Programming Mode:

Key Function in Programming Mode


[MODE]: Exit menu when pressed momentarily or exit
data entry mode
[Y/+]: Increase alphanumerical value for data entry or
confirm (yes) for a question
[N/-]: Provides a “no” response to a question

63

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

22.3 Calibration

Two types of calibration are available: Zero (fresh air) and Span.

Select Zero or Span Calibration by pressing [N/+]. Once your choice


is highlighted, press [Y/+].

22.3.1 Zero Calibration

The procedure for performing a zero calibration is covered on page


49.

22.3.2 Span Calibration

The procedure for performing a basic span calibration is covered on


page 49.

64

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

22.4 Measurement

The sub-menus for Measurement are Measurement Gas and


Measurement Unit.

22.4.1 Meas. Gas

Measurement gases are organized in four lists:

• My List is a customized list of gases that you create. It contains a


maximum of 10 gases and can only be built in ProRAE Studio on
a PC and transferred to the instrument. Note: The first gas in the
list is always isobutylene (it cannot be removed from the list).
• Last Ten is a list of the last ten gases used by your instrument.
The list is built automatically and is only updated if the gas
selected from Custom Gases or Library is not already in the Last
Ten. This ensures that there is no repetition.
• Gas Library is a library that consists of many of the gases found
in RAE Systems’ Technical Note TN-106 (available online at
www.raesystems.com).
• Custom Gases are gases with user-modified parameters. Using
ProRAE Studio, all parameters defining a gas can be modified,
including the name, span value(s), correction factor, and default
alarm limits.

1. Scroll through each list by pressing [N/-].


2. Press [Y/+] to select one (My List, Last Ten, Gas Library, or
Custom Gases).

65

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

3. Once you are in one of the categories, press [N/-] to scroll


through its list of options and [Y/+] to select one. (If you
press [MODE], you exit to the next submenu.)
4. Press [Y/+] to save your choice or [N/-] to undo your
selection.

Leave the sub-menu and return to the Programming Mode menus by


pressing [MODE].

22.4.2 Meas. Unit

Standard available measurement units include:

Abbreviation Unit UltraRAE 3000


ppm parts per million Yes
ppb parts per billion
mg/m3 milligrams per cubic meter Yes
ug/m3 micrograms per cubic meter

• Scroll through the list by pressing [N/-].


• Select by pressing [Y/+].
• Save your selection by pressing [Y/+] or undo your selection by
pressing [N/-].

Leave the sub-menu and return to the Programming Mode menus by


pressing [MODE].

66

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

22.4.3 Tube Selection


When operating the UltraRAE 3000 in Compound Specific mode, the
internal computer works most effectively when it is told which type
of separation tube is being used.

1. Scroll through the menu by pressing [N/-].


2. Press [Y/+] to make a selection.
3. Press [MODE] when you are done.
4. Press [Y/+] to save your choice or [N/-] to undo your selection.

67

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

22.5 Alarm Setting


During each measurement period, the gas concentration is compared
with the programmed alarm limits (gas concentration alarm limit
settings: Low, High, TWA and STEL). If the concentration exceeds
any of the preset limits, the loud buzzer and red flashing LED are
activated immediately to warn of the alarm condition.

An alarm signal summary is shown on page 42.

In this menu, you can change the High and Low alarm limits, the
STEL limit, and the TWA. Press [Y/+] to to enter the Alarm Setting
menu. Note: All settings are shown in ppb (parts per billion), or
µg/m3 (micrograms per cubic meter), depending on your setting.

1. Scroll through the Alarm Limit sub-menu using the [N/-] key
until the display shows the desired limit to be changed (High
Alarm, Low Alarm, STEL Alarm, and TWA Alarm)
2. Press [Y/+] to select one of the alarm types. The display
shows a flashing cursor on the left-most digit of the
previously stored alarm limit.
3. Press [Y/+] to increase each digit’s value.
4. Press [N/-] to advance to the next digit.
5. Again, use [Y/+] to increase the number.
Repeat this process until all numbers are entered.
Press [MODE] when you are done.

• Press [Y/+] to save the changes.


• Press [N/-] to undo the changes and revert to the previous
settings.
When all alarm types have been changed or bypassed, press [MODE]
to exit to the Programming Menu.

68

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

22.5.1 High Alarm

You can change the High Alarm limit value. The value is typically set
by the instrument to match the value for the current calibration gas. It
is expressed in parts per billion (ppb). Note: The default value
depends on the measurement gas.

To change the High Alarm value:

1. Press [Y/+] to increase each digit’s value.


2. Press [N/-] to advance to the next digit.
3. Again, use [Y/+] to increase the number.

Repeat this process until all numbers are entered.


When you have completed your selections, press [MODE]. You will
see two choices: Save and Undo. You have the opportunity to register
the new settings or to change your mind and revert to your previous
settings.

Press [Y/+] to save the changes.


Press [N/-] to undo the changes and revert to the previous settings.

22.5.2 Low Alarm

You can change the Low Alarm limit value. The value is typically set
by the instrument to match the value for the current calibration gas. It
is expressed in parts per billion (ppb). Note: The default value
depends on the measurement gas.

To change the Low Alarm value:

1. Press [Y/+] to increase each digit’s value.


2. Press [N/-] to advance to the next digit.
3. Again, use [Y/+] to increase the number.

Repeat this process until all numbers are entered.

69

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

When you have completed your selections, press [MODE]. You will
see two choices: Save and Undo. You have the opportunity to register
the new settings or to change your mind and revert to your previous
settings.

• Press [Y/+] to save the changes.


• Press [N/-] to undo the changes and revert to the previous
settings.

22.5.3 STEL Alarm

You can change the STEL Alarm limit value. The value is typically
set by the instrument to match the value for the calibration gas. It is
expressed in parts per billion (ppb). Note: The default value depends
on the measurement gas.

To change the STEL Alarm value:

1. Press [Y/+] to increase each digit’s value.


2. Press [N/-] to advance to the next digit.
3. Again, use [Y/+] to increase the number.
Repeat this process until all numbers are entered.
When you have completed your selections, press [MODE]. You will
see two choices: Save and Undo. You have the opportunity to register
the new settings or to change your mind and revert to your previous
settings.

• Press [Y/+] to save the changes.


• Press [N/-] to undo the changes and revert to the previous
settings.

70

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

22.5.4 TWA Alarm

You can change the TWA (time-weighted average) Alarm limit value.
The value is typically set by the instrument to match the value for the
calibration gas. It is expressed in parts per billion (ppb). Note: The
default value depends on the measurement gas.

To change the TWA Alarm value:

1. Press [Y/+] to increase each digit’s value.


2. Press [N/-] to advance to the next digit.
3. Again, use [Y/+] to increase the number.
Repeat this process until all numbers are entered.
When you have completed your selections, press [MODE]. You will
see two choices:

• Save
• Undo

You have the opportunity to register the new settings or to change


your mind and revert to your previous settings.

• Press [Y/+] to save the changes.


• Press [N/-] to undo the changes and revert to the previous
settings.

71

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

22.5.5 Alarm Mode

There are two selectable alarm modes:


Latched When the alarm is triggered, you can
manually stop the alarm.
The latched setting only controls alarms
for High Alarm, Low Alarm, STEL
Alarm, and TWA alarm.

Note: To clear an alarm when the


instrument is set to “Latched,” press
[Y/+] when the main (Reading) display is
shown.
Automatic Reset When the alarm condition is no longer
present, the alarm stops and resets itself.
1. Press [N/-] to step from one alarm mode to the other.
2. Press [Y/+] to select an alarm mode.
When you have completed your selections, press [MODE].

You will see two choices: Save and Undo. You have the
opportunity to register the new settings or to change your mind
and revert to your previous settings.

• Press [Y/+] to save the changes.


• Press [N/-] to undo the changes and revert to the previous
settings.

72

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

22.5.6 Buzzer & Light

The buzzer and light alarms can be programmed to be on or off


individually or in combination. Your choices are:

• Both on
• Light only
• Buzzer only
• Both off

1. Press [N/-] to step from one option to the next.


2. Press [Y/+] to make your selection (the dark circle in the “radio
button” indicates your selection).
3. When you have completed your selections, press [MODE].

You will see two choices: Save and Undo. You have the
opportunity to register the new settings or to change your mind
and revert to your previous settings.

• Press [Y/+] to save the changes.


• Press [N/-] to undo the changes and revert to the previous
settings.

22.6 Datalog
The instrument calculates and stores the concentration and ID of each
sample taken. In the datalog sub-menu, a user can perform the tasks
and functions shown below.

73

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

1. Scroll through the Datalog sub-menu using the [N/-] key until the
display shows the desired parameter to be changed:

Clear Datalog
Interval
Data Selection
Datalog Type

2. Press [Y/+] to make your selection. Exit by pressing [MODE] for


Back.

22.6.1 Clear Datalog

This erases all the data stored in the datalog.

Note: Once the datalog is cleared, the data cannot be recovered.

Press [Y/+] to clear the datalog. The display asks, “Are you sure?”

• Press [Y/+] if you want to clear the datalog. When it has been
cleared, the display shows “Datalog Cleared!”
• Press [N/-] if you do not want to clear the datalog.

The display changes, and you are taken to the next sub-menu,
Interval.

22.6.2 Interval

Intervals are shown in seconds. The default value is 60 seconds. The


maximum interval is 3600 seconds.
1. Press [Y/+] to increase each digit’s value.
2. Press [N/-] to advance to the next digit.
3. Again, use [Y/+] to increase the number.

74

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

Repeat this process until all numbers are entered.


When you have completed your selections, press [MODE].

You will see two choices: Save and Undo. You have the opportunity
to register the new settings or to change your mind and revert to your
previous settings.

• Press [Y/+] to save the changes.


• Press [N/-] to undo the changes and revert to the previous
settings.

22.6.3 Data Selection

Data Selection allows you to select which types of data are stored and
made available when you offload your datalog to a computer via
ProRAE Studio software.

You can choose any or all of three types of data (you must choose at
least one):

• Average
• Maximum
• Minimum

1. Press [N/-] to step from one option to the next. The highlighter
indicates your choice.
2. Press [Y/+] to toggle your selection on or off (the check box
indicates “on” with an “X”).
3. When you have completed your selections, press [MODE].

You will see two choices: Save and Undo. You have the opportunity
to register the new settings or to change your mind and revert to your
previous settings.

• Press [Y/+] to save the changes.


• Press [N/-] to undo the changes and revert to the previous
settings.

75

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

22.6.4 Datalog Type

The instrument has three datalog types:

Auto Default mode. Collects datalog information when the


instrument is sampling.
Manual Datalogging occurs only when the instrument’s
datalogging is manually started (see below for
details).
Snapshot Datalogs only during single-event capture sampling.
Note: You can only choose one datalog type to be active at a time.

1. Press [N/-] to step from one option to the next.


2. Press [Y/+] to make your selection (the dark circle in the “radio
button” indicates “on”).
3. When you have completed your selection, press [MODE].

You will see two choices: Save and Undo. You have the opportunity
to register the new settings or to change your mind and revert to your
previous settings.

• Press [Y/+] to save the changes.


Press [N/-] to undo the changes and revert to the previous settings.

22.6.5 Manual Datalog

When the instrument is set to Manual Datalog, you turn datalogging


on and off by stepping through the displays from the Main Display,
and then pressing the keys to select datalog on/off functions.

• When you reach the screen that says “Start Datalog?” press
[Y/+] to start it. You see “Datalog Started,” confirming that
datalogging is now on.
When you reach the screen that says “Stop Datalog?” press [Y/+] to
stop it. You see “Datalog Stopped,” confirming that datalogging is
now off.

76

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

Note: Dashed line indicates automatic progression.

77

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

22.6.6 Snapshot Datalog

When the instrument is in Snapshot datalogging mode, it captures a


single “snapshot” of the data at the moment of your choosing.
Whenever the instrument is on and it is set to Snapshot, all you have
to do is press [MODE] each time you want to capture a snapshot of
the data at that instant.

When you send the data to a computer using ProRAE Studio, the data
snapshots are uniquely identified by time and other parameters.

22.7 Monitor Setup

Many settings can be accessed in this menu, including setting the date
and time and adjusting the pump’s on/off duty cycle.

22.7.1 Radio Power

The radio connection can be turned on or off. (The default value is


off.)

1. Press [N/-] to step from one option to the next (on or off).
2. Press [Y/+] to make your selection (the dark circle in the “radio
button” indicates that the option is selected).
3. When you have completed your selection, press [MODE].

• Press [Y/+] to accept the new radio setting (on or off).


• Press [N/-] to discard the change and move to the next sub-
menu.

78

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

22.7.2 Op Mode

Under Monitor Setup is “Op Mode.”

Press [Y/+] to select.

You see two options (one is highlighted):

Hygiene
Search

The current mode is indicated by a dark circle within the circle in


front of either Hygiene or Search.

1. Select Hygiene or Search by pressing [N/-]. The highlighting


changes from one to the other each time you press [N/-].
2. Press [Y/+] to select that mode for the instrument.
3. Press [MODE] when you want to register your selection to
place the instrument in the selected mode.
4. Press [Y/+] to commit the change and exit to the Monitor
Setup screen, or press [N/-] to Undo (exit to the Monitor
Setup screen without changing the Mode).

22.7.3 Site ID

Enter an 8-digit alphanumeric/character Site ID in the programming


mode. This Site ID is included in the datalog report.

1. Press [Y/+] and the display shows the current site ID.
Example: “RAE00001.” Note that the left-most digit flashes
to indicate it is the selected one.
2. Press [Y/+] to step through all 26 letters (A to Z) and 10
numerals (0 to 9).
Note: The last four digits must be numerals.
3. Press [N/-] to advance to the next digit. The next digit to the
right flashes.

79

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

Repeat this process until all eight digits of the new site ID are
entered.
Press [MODE] to exit.
If there is any change to the existing site ID, the display shows
“Save?” Press [Y/+] to accept the new site ID. Press [N/-] to discard
the change and move to the next sub-menu.

22.7.4 User ID

Enter an 8-digit alphanumeric User ID in the programming mode.


This User ID is included in the datalog report.

1. Press [Y/+] and the display shows the current User ID.
Example: “RAE00001.” Note that the left-most digit flashes
to indicate it is the selected one.
2. Press [Y/+] to step through all 26 letters (A to Z) and 10
numerals (0 to 9).
3. Press [N/-] to advance to the next digit. The next digit to the
right flashes.
Repeat this process until all eight digits of the new User ID
are entered.
Press [MODE] to exit.
If there is any change to the existing User ID, the display shows
“Save” Press [Y/+] to accept the new site ID. Press [N/-] to discard
(undo) the change and move to the next sub-menu.

80

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

22.7.5 User Mode

The instrument has two user modes:


Basic Basic users can only see and use a basic set of
functions.
Advanced Advanced users can see all screens and perform all
available functions.
Note: The default value for User Mode is Basic.
To change the User Mode:

1. Press [N/-] to step from one option to the next. The highlighting
changes each time you press [N/-].
2. Press [Y/+] to make your selection (the dark circle in the “radio
button” indicates “on”).
3. When you have completed your selection, press [MODE].
4. Press [Y/+] to accept the new User Mode. Press [N/-] to discard
the change and move to the next sub-menu.

22.7.6 Date

The Date is expressed as Month/Day/Year, with two digits for each.

1. Press [Y/+] and the display shows the current date. Note that
the left-most digit flashes to indicate it is selected.
2. Press [Y/+] to step through all 10 numerals (0 to 9).
3. Press [N/-] to advance to the next digit. The next digit to the
right flashes.
Repeat this process until all six digits of the new date are
entered.
Press [MODE] to exit.
• Press [Y/+] to save the new date.
• Press [N/-] to undo the change and move to the next sub-
menu.

81

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

22.7.7 Time

The Time is expressed as Hours/Minutes/Seconds, with two digits for


each. The time is in 24-hour (military) format.

1. Press [Y/+] and the display shows the current time. Note that
the left-most digit flashes to indicate it is selected.
2. Press [Y/+] to step through all 10 numerals (0 to 9).
3. Press [N/-] to advance to the next digit. The next digit to the
right flashes.
Repeat this process until all six digits of the new time are
entered.
Press [MODE] to exit.
• Press [Y/+] to save the new date.
• Press [N/-] to undo the change and move to the next sub-
menu.

22.7.8 Duty Cycle

The pump’s duty cycle is the ratio of its on time to off time. The duty
cycle ranges from 50% to 100% (always on; this is the default value),
and the period is 10 seconds. Therefore, a duty cycle of 60% means that
the pump is on for 6 seconds and off for four seconds. Duty cycling is
employed by the instrument to clean the PID. A lower duty cycle has a
greater effect on keeping the PID clean than a higher duty cycle.

Important! Pump duty cycling is interrupted when the instrument


senses a gas. The pump’s duty cycle is disabled when the
measurement is greater than the 2ppm threshold and is re-enabled
when the reading falls below 90% of the threshold (1.8 ppm).

1. Press [Y/+] to increase the value.


2. When you have completed your selection, press [MODE].
• Press [Y/+] to save the new duty cycle value.
• Press [N/-] to undo the change and move to the next sub-
menu.

82

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

22.7.9 Temperature Unit

The temperature display can be switched between Fahrenheit and


Celsius units.

1. Press [N/-] to step from one option to the next.


2. Press [Y/+] to make your selection (the dark circle in the “radio
button” indicates “on”).
3. When you have completed your selection, press [MODE].
• Press [Y/+] to save the new temperature unit.
• Press [N/-] to undo the change and move to the next sub-
menu.

22.7.10 Pump Speed

The pump can operate at two speeds, high and low. Running at low
speed is quieter and conserves a small amount of power. There is
almost no difference in sampling accuracy.

1. Press [N/-] to step from one option to the next.


2. Press [Y/+] to make your selection (the dark circle in the “radio
button” indicates “on”).
3. When you have completed your selection, press [MODE].
• Press [Y/+] to save the new temperature unit.
• Press [N/-] to undo the change and move to the next sub-
menu.

83

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

22.7.11 Language

English is the default language, but other languages can be selected


for the instrument.

1. Press [N/-] to step from one option to the next.


2. Press [Y/+] to make your selection (the dark circle in the “radio
button” indicates “on”).
3. When you have completed your selection, press [MODE].
• Press [Y/+] to save your new language choice.
• Press [N/-] to undo it and return to the previous language
selection.

22.7.12 Real Time Protocol

Real Time Protocol is the setting for data transmission.

The choices are:

P2M (cable) Point to multipoint. Data is transferred from the


instrument to multiple locations using a wired
connection. Default data rate: 19200 bps.
P2P (cable) Point to point (default). Data is transferred only
between the instrument and one other location,
such as a computer. Default data rate: 9600 bps.
P2M (wireless) Point to multipoint, wireless. Data is transferred
wirelessly and can be received by multiple
receivers. Use this setting with a RAELink3.

1. Press [N/-] to step from one option to the next.


2. Press [Y/+] to make your selection (the dark circle in the “radio
button” indicates “on”).
3. When you have completed your selection, press [MODE].
• Press [Y/+] to save the new real-time communications protocol.
• Press [N/-] to undo the change and move to the next sub-menu.

84

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

22.7.13 Power On Zero

When Power On Zero is on, the instrument is ready to perform a zero


calibration when it is turned on.

1. Press [N/-] to step from one option to the next.


2. Press [Y/+] to make your selection (the dark circle in the “radio
button” indicates your selection).
3. When you have completed your selection, press [MODE].
• Press [Y/+] to save the change.
• Press [N/-] to discard the change and move to the next sub-
menu.

22.7.14 Unit ID

This three-digit number keeps data separated by instrument when


more than one instrument is used in a network. If multiple sensing
units are attempting to communicate with the same Host, then the
units must all have a different Unit ID.

1. Press [Y/+] to step through all 10 numerals (0 to 9). If you pass


the numeral you want, keep pressing [Y/+]. After it counts up to
9, it starts counting up from 0 again.
2. Press [N/-] to advance to the next digit. The next digit to the right
flashes.
Repeat this process until all three digits of the Unit ID are entered.

3. Press [MODE] when you are done.


• Press [Y/+] to save the change.
• Press [N/-] to discard the change and move to the next sub-
menu.

Note: If you are using an UltraRAE 3000 with a RAELink3, this unit
ID is displayed by the RAELink3.

85

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

22.7.15 LCD Contrast

The display’s contrast can be increased or decreased from its default


setting. You may not need to ever change the default setting, but
sometimes you can optimize the display to suit extreme temperature
and ambient brightness/darkness conditions.

• The minimum value is 20.


• The maximum value is 60.

1. Press [Y/+] to increase the value or [N/-] to decrease the value.


2. Press [MODE] to save your selection.
• Press [Y/+] to save your new contrast value.
• Press [N/-] to undo it and return to the previous value.

22.7.16 Lamp ID
The UltraRAE can automatically identify the type of lamp, or you can
select a lamp type manually.

1. Scroll through the menu by pressing [N/-].


2. Press [Y/+] to make a selection.
3. Press [MODE] when you are done.
4. Press [Y/+] to save your choice or [N/-] to undo your selection.

86

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

23 Humidity Compensation
The UltraRAE 3000 has a humidity sensor and humidity
compensation circuitry. By default it is on, but it can be turned off or
on by using ProRAE Studio software. RAE Systems recommends
testing the humidity sensor once a year. See page 99 for the simple
procedure.

24 Hygiene Mode
The instrument usually operates in Hygiene Mode, which provides
basic functionality. However, it is possible to operate it in a second
mode called Search Mode. Here are the primary differences:

Hygiene Mode: Automatic measurements, continuously


running and datalogging, and calculates
additional exposure values.
Search Mode: Manual start/stop of measurements and display
of certain exposure values.

87

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

24.1 Basic User Level & Hygiene Mode

The default setting is navigated in the following way:

Note: Dashed line indicates automatic progression.

88

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

Pressing [N/-] steps you from screen to screen. Options include


clearing the Peak value and turning on the instrument’s PC
Communications for data transfer to a PC.

89

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

24.2 Entering Search Mode From Hygiene


Mode

In order to change the instrument’s operational mode from Hygiene


Mode to Search Mode, you must enter the password-protected
Programming Mode:

1. Hold [MODE] and [N/-] until you see the password screen.
2. Use [Y/+] to increment to the number you want for the first
digit. (If you pass by the desired number, press [Y/+] until it
cycles through to 0 again. Then press [Y/+] until you reach
the desired number.)
3. Press [N/-] to advance to the next digit.
4. Again press [Y/+] to increment the number.
5. Press [N/-] to advance to the next digit.

Continue the process until all four numbers of the password have
been input. Then press [MODE] to proceed.

The screen changes to icons with the label “Calibration.”

1. Press [N/-] to advance to “Monitor Setup.”


2. Press [Y/+] to select Monitor Setup.

Under Monitor Setup, you will see “Op Mode.”

Press [Y/+] to select.

You will see:

Hygiene
Search

The current mode is indicated by a dark circle within the circle in


front of either Hygiene or Search.

1. Select Hygiene or Search by pressing [N/-].


2. Press [Y/+] to place the instrument into the selected mode.

90

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

3. Press [MODE] when you want to register your selection to


place the instrument in the selected mode.
4. Press [Y/+] to commit the change and exit to the Monitor
Setup screen, or press [N/-] to Undo (exit to the Monitor
Setup screen without changing the Mode).

24.3 Optional Graphic Screen In Search Mode


Using ProRAE Studio, you can set your instrument to show a graphic
display instead of a numeric display of ongoing data. Consult your
ProRAE Studio disc for information.

During sampling, the display’s readings are shown numerically, plus


the graph tracks the highest readings over time. The numeric reading
alternates between the value and the measurement units, as well:

91

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

25 Advanced User Level (Hygiene


Mode Or Search Mode)
The User Mode called Advanced User Level allows a greater number
of parameters to be changed than Basic User Level. It can be used
with either of the Operation Modes, Hygiene Mode or Search Mode.

25.1 Advanced User Level & Hygiene Mode

With the instrument in Operation Mode: Hygiene Mode, enter User


Mode: Advanced User Level (refer to the section called Monitor
Mode for instructions).

Once you are in Advanced User Level and Hygiene Mode together,
you can change the calibration reference and measurement gas, in
addition to performing normal monitoring functions.

Pressing [N/-] progresses through the screens, while pressing [Y/+]


selects options. Pressing [MODE] makes menu choices when it is
shown for “Done” or “Back.” Pressing and holding [Mode] whenever
the circle with a vertical line in the middle is shown activates the
countdown to shutoff.

92

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

Note: Dashed line indicates automatic progression.

93

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

25.2 Basic User Level & Search Mode


With the instrument in Operation Mode: Search Mode, enter User
Mode and select Basic User Level (refer to the section called User
Mode for instructions).

When the instrument is in Search Mode, it only samples when you


activate sampling. When you see the display that says, “Ready…Start
sampling?” press [Y/+] to start. The pump turns on and the instrument
begins collecting data. To stop sampling, press [N/-] while the main
display is showing. You will see a new screen that says, “Stop
sampling?” Press [Y/+] to stop sampling. Press [N/-] if you want
sampling to continue.

94

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

Note: Dashed line indicates automatic progression.

95

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

25.3 Advanced User Level & Search Mode

With the instrument in Operation Mode: Search Mode, enter User


Mode and select Advanced User Level (refer to the section called
Monitor Mode for instructions). Operation is similar to Basic User
Level & Sampling Mode, but now allows you to change calibration
and measurement reference gases. Refer to the section on
measurement gases on page 65 for more details.

Note: Dashed line indicates


automatic progression.

96

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

25.4 Diagnostic Mode

IMPORTANT! Diagnostic Mode is designed for servicing and


manufacturing, and therefore is not intended for everyday use, even
by advanced users. It provides raw data from sensors and about
settings, but only allows adjustment of pump stall parameters, which
should only be changed by qualified personnel.

Note: If the instrument is turned on in Diagnostic Mode and you


switch to User Mode, datalog data remains in raw count form. To
change to standard readings, you must restart the instrument.

25.4.1 Entering Diagnostic Mode


Note: To enter Diagnostic Mode, you must begin with the instrument
turned off.

Press and hold [Y/+] and [MODE] until the instrument starts.

The instrument goes through a brief startup, and then displays raw
data for the PID sensor. These numbers are raw sensor readings
without calibration. The instrument is now in Diagnostic Mode.

Note: In Diagnostic Mode, the pump and lamp are normally on.

You can enter Programming Mode and calibrate the instrument as


usual by pressing both [MODE] and [N/-] for three seconds.

You can enter Monitoring Mode by pressing [MODE] and [Y/+]


together for three seconds.

Once the instrument is started up in Diagnostic Mode, you can switch


between Diagnostic Mode and Monitoring Mode by pressing and
holding [MODE] and [Y/+] simultaneously for two seconds.

In Diagnostic mode, you can step through parameter screens by


pressing [MODE].

97

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

25.4.2 Adjusting The Pump Stall Threshold


If the gas inlet is blocked but the pump does not shut down, or the
pump shuts down too easily with a slight blockage, the pump stall
threshold value may be set too high or too low.

Use the following steps to adjust the pump stall threshold:

25.4.3 Pump High


In Diagnostic Mode, press the [MODE] key until “Pump High” is
displayed. The display shows the maximum, minimum, and stall
values for the pump at its high speed. Write down the “Max” reading.

Block the gas inlet and watch the pump current reading (labeled “I”)
increase. Write down its blocked reading. Note: If the pump current
reading does not increase significantly (less than 10 counts), then
there may be a leak in the gas inlet or the pump is weak or defective.

Add the two readings you wrote down. This is the average of the
maximum block count and the maximum idle count. Divide that
number by 2. Use the [Y/+] or [N/-] key to increase or decrease the
stall value to equal that number.

Press the [MODE] key to exit this display.

25.4.4 Pump Low


In Diagnostic Mode, press the [MODE] key until “Pump Low” is
displayed. The display shows the maximum, minimum, and stall
values for the pump at its low speed. Write down the “Max” reading.

Block the gas inlet and watch the pump current reading (labeled “I”)
increase. Write down its blocked reading. Note: If the pump current
reading does not increase significantly (less than 10 counts), then
there may be a leak in the gas inlet or the pump is weak or defective.

Add the two readings you wrote down. This is the average of the
maximum block count and the maximum idle count. Divide that

98

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

number by 2. Use the [Y/+] or [N/-] key to increase or decrease the


stall value to equal that number.

Press the [MODE] key to exit this display.

25.4.5 Testing The Humidity Sensor

1. Press [MODE] to step through the diagnostic screens until you


reach a screen that says “T.H.P” (for “temperature, humidity,
pressure”) at the top.

There are three numbers for the humidity reading (“H”). The first
number is the current humidity reading from the sensor. The
second is the reference number for 0% humidity, and the third
number is the reference for 100% humidity.

2. Fill a cup with warm water (>25° C/77° F).


3. Put a filter on the UltraRAE 3000’s inlet probe.
4. Place the inlet probe over the warm water.

99

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

5. Cover the cup with your hand.

25° C / 77° F

6. Wait a few seconds and check the high-humidity reading.


7. The humidity reading should be within ±10% of the 100%
humidity reading. If it is not, then the THP Sensor (part number
023-3011-000-FRU) should be replaced.
8. Check the low-humidity reading by connecting the inlet probe to
a tank of zero gas (air at 0% humidity).
9. Turn on the zero gas and wait a few seconds for the sensor
reading to stabilize. It should read within 10% of the low-
reference number. If it does not, replace the T.H.P. sensor.
10. Once you have finished testing the humidity sensor, exit
Diagnostic Mode by shutting down the UltraRAE 3000 (hold
[MODE] through the countdown, and then release when it is off).

25.4.6 Exiting Diagnostic Mode


You can exit Diagnostic Mode and go directly to Programming Mode
or Monitor Mode as outlined above, or you can exit Diagnostic Mode
completely.

To exit Diagnostic Mode so that it cannot be re-entered without a


restart:

100

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

Shut down the instrument. When it is off, restart it by holding the


[MODE] key. Diagnostic Mode cannot be entered until the instrument
is restarted as outlined in “Entering Diagnostic Mode.”

101

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

26 Transferring Data To & From


Computer
Once you have connected your instrument cradle to the PC, you can
can transfer data, including a download of the datalog to the computer
and updates of firmware to the instrument (should this ever be
necessary).

26.1 Downloading The Datalog To A PC

1. Connect the data cable to the PC and the cradle.


2. Place the instrument into its cradle. The charging LED should
be illuminated.
3. Start ProRAE Studio on your PC.
4. From ProRAE Studio, select “Operation” and select Setup
Connection.
5. Select the COM port to establish a communication link
between the PC and the instrument.
6. To receive the datalog in the PC, select “Downlog Datalog.”
7. When you see “Unit Information,” click OK.

During the data transfer, the display shows a progress bar.

When the transfer is done, you will see a screen with the datalog
information. You can now export this datalog for other use or
printing.

102

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

26.2 Uploading Firmware To The instrument


From A PC
Uploading new firmware to your instrument requires connecting the
instrument and PC. Follow these steps to make the connection:

1. Connect the data cable to the PC and the cradle.


2. Place the instrument into its cradle. The charging LED should
be illuminated.
3. Start RAEProgrammer 7000 on your PC.
4. From RAEProgrammer 7000, select “Operation” and select
Setup Connection.
5. Select the COM port to establish a communication link
between the PC and the instrument.
6. Select Operation Æ Download Firmware.

Once communication is established, follow the instructions that


accompany RAEProgrammer 7000 and the firmware to upload the
new firmware to your instrument.

Note: Check for the latest updates to ProRAEProgrammer 7000 at


www.raesystems.com.

103

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

27 Maintenance
The major maintenance items of the instrument are:
• Battery pack
• Sensor module
• PID lamp
• Sampling pump
• Inlet connectors and filters
In addition, you can test the humidity sensor (this should be done
annually, in order to ensure the most accurate operation).
Note: Maintenance should be performed by qualified personnel
only.
NOTE: The printed circuit board of the instrument is connected
to the battery pack even if the power is turned off. Therefore, it is
very important to disconnect the battery pack before servicing or
replacing any components inside the instrument. Severe damage
to the printed circuit board or battery may occur if the battery
pack is not disconnected before servicing the unit.

27.1 Battery Charging & Replacement


When the display shows a flashing empty battery icon, the battery
requires recharging. It is recommended to recharge the instrument
upon returning from fieldwork. A fully charged battery runs a
instrument for 16 hours continuously. The charging time is less than 8
hours for a fully discharged battery. The battery may be replaced in
the field (in areas known to be non-hazardous), if required.
WARNING!
To reduce the risk of ignition of hazardous atmospheres, recharge
battery only in area known to be non-hazardous. Remove and
replace battery only in areas known to be non-hazardous.

104

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

27.1.1 Replacing The Li-ion Battery


1. Turn off the instrument.
2. Located on the rear of the instrument is a battery tab. Slide it down to
unlock the battery.

3. Remove the battery pack from the battery compartment by tilting


it out.

4. Replace a fully charged spare battery pack inside the battery


compartment. Make sure the battery pack is oriented properly
inside the compartment.
5. Slide the capture tab back up to its locked position.

27.1.2 Replacing The Alkaline Battery Adapter


An alkaline battery adapter is supplied with each instrument. The
adapter (part number 059-3052-000) accepts four AA alkaline
batteries (use only Duracell MN1500) and provides approximately 12
hours of operation. The adapter is intended to be used in emergency
situations when there is no time to charge the Li-ion battery pack.
To insert batteries into the adapter:
1. Remove the three hex-socket screws to open the
compartment.
2. Insert four fresh AA batteries as indicated by the polarity (+/-)
markings.
3. Replace the cover. Replace the three screws.

105

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

To install the adapter in the instrument:


1. Remove the Li-ion battery pack from the battery
compartment by sliding the tab and tilting out the battery.
2. Replace it with the alkaline battery adapter
3. Slide the tab back into place to secure the battery adapter.

IMPORTANT!
Alkaline batteries cannot be recharged. The instrument’s internal
circuit detects alkaline batteries and will not allow recharging. If you
place the instrument in its cradle, the alkaline battery will not be
recharged. The internal charging circuit is designed to prevent
damage to alkaline batteries and the charging circuit when alkaline
batteries are installed inside the instrument.

Note: When replacing alkaline batteries, dispose of old ones properly.

WARNING!
To reduce the risk of ignition of hazardous atmospheres, recharge the
battery only in areas known to be non-hazardous. Remove and replace
the battery only in areas known to be non-hazardous.

Note: The internal charging circuit is designed to prevent charging to


alkaline batteries.

106

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

27.2 PID Sensor & Lamp Cleaning/Replacement


The sensor module is made of several components and is attached to
the lamp-housing unit as shown below.
Inlet Probe Assembly
PN 023-3012-000-FRU

O-ring, 35mm x 2mm


Tube Adapter
Assembly
Porous Metal Filter PN 059-3015-000-FRU
Lamp 1/2" (9.8eV)
PN 050-0020-000
Rubber Adapter
Seal Nut
Sensor Detector
PN 023-3010-001
Sensor Body
Assembly Stainless Steel Washer
PN 023-3006- O-ring, 36.5mm x 2.65mm
000-FRU
Sensor Module Assembly
PN 023-3005-600-FRU
THP (temperature, humidity, pressure) Sensor Module
PN 023-3011-000-FRU

Sensor Components
Note: The cleaning procedure is not normally needed. Clean the PID
sensor module, the lamp and the lamp housing only if:
1. The reading is inaccurate even after calibration.
2. The reading is very sensitive to air moisture.
3. A liquid has been sucked into the unit and damaged the unit.
Use of the external filter helps to prevent contamination of the sensor.
To access the sensor components and lamp, gently unscrew the lamp-
housing cap, remove the sensor adapter with the gas inlet probe and
the metal filter all together. Then hold the PID sensor and pull it
straight out. A slight, gentle rocking motion helps release the sensor.
Note: The 10.6eV lamp requires a Teflon O-ring. The 9.8eV and
11.7eV lamps do not require the O-ring.

107

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

27.3 Cleaning The PID Sensor


Place the entire PID sensor module into GC grade methanol. It is
highly recommended that an ultrasound bath to be used to clean the
sensor for at least 15 minutes. Then dry the sensor thoroughly. Never
touch the electrodes of the sensor by hand.
Also use a methanol-soaked cotton swab to wipe off the lamp housing
where it contacts the sensor when the sensor is installed.
Turn over the sensor so that the pins point up and the sensor cavity is
visible. Examine the sensor electrodes for any corrosion, damage, or
bending out of alignment. The metal sensor electrode “fingers”
should be flat and straight. If necessary, carefully bend the sensor
fingers to ensure that they do not touch the Teflon portions and that
they are parallel to each other. Make sure that the nuts on the sensor
pins are snug but not overtight. If the sensor is corroded or otherwise
damaged, it should be replaced.

27.3.1 Cleaning The Lamp Housing Or Changing The


Lamp
If the lamp does not turn on, the instrument will display an error
message to indicate replacement of the lamp may be required.
1. If the lamp is operational, clean the lamp window surface and the
lamp housing by wiping it with GC grade methanol using a cotton
swab using moderate pressure. After cleaning, hold the lamp up
to the light at an angle to detect any remaining film. Repeat the
process until the lamp window is clean. Never use water solutions
to clean the lamp. Dry the lamp and the lamp housing thoroughly
after cleaning.
CAUTION: Never touch the window surface with the fingers
or anything else that may leave a film. Never use acetone or
aqueous solutions.
2. If the lamp does not turn on, remove the lamp from the lamp
housing. Place the lamp O-ring onto the new lamp. Insert the new
lamp, avoiding contact with the flat window surface.
3. Reinstall the PID sensor module.
4. Tighten the Lamp Housing Cap.

108

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

27.3.2 Determining The Lamp Type


The monitor can accommodate three lamp values: 10.6eV (standard),
9.8eV, and 11.7eV. The monitor automatically reads a marking on the
side of the lamp to set the proper Correction Factor. There are two
ways to determine the lamp type:

Remove the lamp and look for markings (bars) on the side:

• No bars: 10.6eV
• 1 bar: 11.7eV
• 2 bars: 9.8eV

Also, when the monitor is running, the lamp type is shown along with
the calibration and measurement gas and Correction Factor:

Note: This screen can be accessed from the reading screen by


pressing [N/-] four times.

109

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

27.3.3 Sampling Pump


When approaching the end of the specified lifetime of the pump, it
will consume higher amount of energy and reduce its sample draw
capability significantly. When this occurs, it is necessary to replace or
rebuild the pump. When checking the pump flow, make sure that the
inlet connector is tight and the inlet tubing is in good condition.
Connect a flow meter to the gas inlet probe. The flow rate should be
above 450 cc/min when there is no air leakage.
If the pump is not working properly, refer the instrument to qualified
service personnel for further testing and, if necessary, pump repair or
replacement.

27.3.4 Testing The T.H.P. Sensor


It is recommended that you periodically test the humidity sensitivity
of the T.H.P. (temperature/humidity/pressure) sensor. See page 99 for
a simple procedure.

27.3.5 Cleaning The Instrument


Occasional cleaning with a soft cloth is recommended. Do not use
detergents or chemicals.

Visually inspect the contacts at the base of the instrument, on the


battery, and on the charging cradle to make sure they are clean. If
they are not, wipe them with a soft, dry cloth. Never use solvents or
cleaners.

27.3.6 Ordering Replacement Parts


If you need replacement parts, contact your local RAE Systems
distributor. A list is available online:
http://www.raesystems.com
In the U.S., you can order sensors, replacement batteries, and other
accessories online at:
http://istore.raesystems.com/

110

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

27.4 Special Servicing Note


If the instrument needs to be serviced, contact either:
1. The RAE Systems distributor from whom the instrument was
purchased; they will return the instrument on your behalf.

or

2. The RAE Systems Technical Service Department. Before


returning the instrument for service or repair, obtain a Returned
Material Authorization (RMA) number for proper tracking of
your equipment. This number needs to be on all documentation
and posted on the outside of the box in which the instrument is
returned for service or upgrade. Packages without RMA Numbers
will be refused at the factory.

111

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

28 Troubleshooting
Problem Possible Reasons & Solutions
Cannot turn on power Reasons: Discharged battery.
after charging the Defective battery.
battery
Solutions: Charge or replace battery.
Lost password Solutions: Call Technical Support at
+1 408-752-0723 or toll-
free at
+1 888-723-4800
Reading abnormally Reasons: Dirty filter.
High Dirty sensor module.
Excessive moisture and
water condensation.
Incorrect calibration.

Solutions: Replace filter.


Blow-dry the sensor
module.
Calibrate the unit.
Reading abnormally Reasons: Dirty filter.
Low Dirty sensor module.
Weak or dirty lamp.
Incorrect calibration.

Solutions: Replace filter.


Remove Calibration
Adapter.
Calibrate the unit.
Check for air leakage.
Buzzer Reasons: Bad buzzer.
Inoperative
Solutions: Check that buzzer is not
turned off.
Call authorized service
center.

112

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

Inlet flow too low Reasons: Pump diaphragm damaged


or has debris.
Flow path leaks.

Solutions: Check flow path for leaks;


sensor module O-ring, tube
connectors, Teflon tube
compression fitting.
Call Technical Support at
+1 408-752-0723
or toll-free at
+1 888-723-4800
“Lamp” message Reasons: Lamp drive circuit.
during operation Weak or defective PID
lamp, defective.

Solutions: Turn the unit off and back


on.
Replace UV lamp

29 Technical Support
To contact RAE Systems Technical Support Team:

Monday through Friday, 7:00AM to 5:00PM Pacific (US) Time


Phone (toll-free): +1 888-723-4800
Phone: +1 408-952-8461
Email: tech@raesystems.com

Life-critical after-hours support is available:

+1 408-952-8200 select option 8

113

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

30 RAE Systems Contacts


RAE Systems
World Headquarters
3775 N. First St.
San Jose, CA 95134-1708 USA
Phone: +1 408.952.8200
Fax: +1 408.952.8480

E-mail: customerserv@raesystems.com
Web Site: www.raesystems.com

RAE Systems Technical Support


Monday through Friday, 7:00AM to 5:00PM Pacific Time
Phone: +1.408.952.8461
Email: tech@raesystems.com

Life-critical after-hours support is available:


+1.408.952.8200 select option 8

RAE Systems Europe ApS


Kirstinehøj 23 A
DK-2770 Kastrup
Denmark
Phone: +45 86 52 51 55
Fax: +45 86 52 51 77
orders@raeeurope.com
sales@raeeurope.com
service@raesystems.com
Web: www.raesystems.dk

114

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

RAE Systems UK Ltd


D5 Culham Innovation Centre
Culham Science Centre
Abingdon, Oxon OX14 3DB
United Kingdom
Phone: +44 1865408368
Fax: +44 1235531119
Mobile: +44 7841362693
Email: raeuk@raeeurope.com

RAE Systems France


336, rue de la fée des eaux
69390 Vernaison
France
Phone: +33 4 78 46 16 65
Fax: +33 4 78 46 25 98
Email: info-france@raeeurope.com
Web: www.raesystems.fr

RAE BeNeLux BV
Hoofdweg 34C
2908 LC Capelle a/d IJssel
The Netherlands
Phone: +31 10 4426149
Fax: +31 10 4426148
Email: info@rae.nl
Web: www.rae.nl

RAE Systems Spain, s.l.


Av. Remolar, 31
08820 El Prat de Llobregat
Spain
Phone: +34 933 788 352
Fax: +34 933 788 353
Mobile: +34 687 491 106
Email: mdelgado@raespain.com
Web: www.raespain.com

115

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

RAE Middle East


Lot 7, Ground Floor, Office 19
Jebel Ali Free Zone
Dubai
United Arab Emirates
Phone: +971 4 887 5562
Fax: +971 4 887 5563
Email: mesales@raesystems.com

RAE Systems (Hong Kong) Ltd.


Room 8, 6/F, Hong Leong Plaza
33 Lok Yip Road
Fanling, N.T, Hong Kong
Phone: +852.2669.0828
Fax: +852.2669.0803
Email: hksales@raesystems.com

RAE Systems Japan


403 Plaza Ochanomizu Bldg. 2-1
Surugadai Kanda Chiyoda-Ku
Tokyo, Japan
Phone: 81-3-5283-3268
Fax: 81-3-5283-3275
Email: jpsales@raesystems.com

RAE Systems Korea


#1010, DaeMyungAnsVill First,
Sang-Dong 412-2, Wonmi-Gu, Bucheon,
Kyungki-Do, Korea
Phone: 82-32-328-7123
Fax: 82-32-328-7127
Email: krsales@raesystems.com

116

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

Controlled Part of Manual


Intrinsic Safety:
US and Canada: Class I, Division 1, Groups A,B,C,D T4

Europe: ATEX (0575 Ex II 2G Ex ia IIC/IIB T4 Gb)


KEMA 07 ATEX 0127
Complies with EN60079-0:2009, EN60079-11:2007
IECEx CSA 10.0005 Ex ia IIC/IIB T4 Gb
Complies with IEC 60079-0:2007, IEC 60079-11:2006
Temperature: -20º C to 50º C (-4º to 122º F)
Humidity: 0% to 95% relative humidity (non-
condensing)

Basic Operation

Turning The Instrument On


1. With the instrument turned off, press and hold [MODE].
2. When the display turns on, release the [MODE] key.

The instrument is now operating and performs self tests. Once the self
tests are complete, the display shows a graph or numerical gas
reading. This indicates that the instrument is fully functional and
ready to use.

Turning The Instrument Off


1. Press and hold the Mode key for 3 seconds. A 5-second
countdown to shutoff begins.
2. When you see “Unit off...” release your finger from the [MODE]
key. The instrument is now off.

Note: You must hold your finger on the key for the entire shutoff
process. If you remove your finger from the key during the

117

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

countdown, the shutoff operation is canceled and the instrument


continues normal operation.

118

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

Alarm Signals
During each measurement period, the gas concentration is compared
with the programmed alarm limits (gas concentration alarm limit
settings). If the concentration exceeds any of the preset limits, the
loud buzzer and red flashing LED are activated immediately to warn
you of the alarm condition.
In addition, the instrument alarms if one of the following conditions
occurs: battery voltage falls below a preset voltage level, failure of
the UV lamp, pump stall, or when the datalog memory is full.

Alarm Signal Summary


Message Condition Alarm Signal
HIGH Gas exceeds “High 3 beeps/flashes per second*
Alarm” limit
OVR Gas exceeds 3 beeps/flashes per second*
measurement range
MAX Gas exceeds electronics’ 3 beeps/flashes per second*
maximum range
LOW Gas exceeds “Low 2 beeps/flashes per second*
Alarm” limit
TWA Gas exceeds “TWA” 1 Beep/flash per second*
limit
STEL Gas exceeds “STEL” 1 Beep/flash per second*
limit
Pump Pump failure 3 beeps/flashes per second
icon
flashes
Lamp PID lamp failure 3 beeps/flashes per second
plus “Lamp” message on
display

119

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

Battery Low battery 1 flash, 1 beep per minute


icon plus battery icon flashes on
flashes display
CAL Calibration failed, or 1 beep/flash per second
needs calibration
NEG Gas reading measures 1 beep/flash per second
less than number stored in
calibration

Preset Alarm Limits & Calibration


The instrument is factory calibrated with standard calibration gas, and
is programmed with default alarm limits.

Cal Gas Cal unit Low High TWA STEL


Span
(Isobutylene)
ppbRAE 10 ppm 10 25 10 25
3000
MiniRAE 100 ppm 50 100 10 25
3000
MiniRAE 100 ppm 50 100 10 25
Lite
UltraRAE 100 ppm 50 100 10 25
3000

120

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

Charging The Battery


Always fully charge the battery before using the instrument. The
instrument’s Li-ion/NiMH battery is charged by placing the
instrument in its cradle. Contacts on the bottom of the instrument
meet the cradle’s contacts, transferring power without other
connections.

Note: Before setting the instrument into its charging cradle, visually
inspect the contacts to make sure they are clean. If they are not, wipe
them with a soft cloth. Do not use solvents or cleaners.

Follow this procedure to charge the instrument:

1. Plug the AC/DC adapter’s barrel connector into the


instrument’s cradle.

2. Plug the AC/DC adapter into the wall outlet.


3. Place the instrument into the cradle, press down, and lean it
back. It locks in place and the LED in the cradle glows.

Note: To release the instrument, press down and tilt the top out
of the cradle and lift up.

The instrument begins charging automatically. The LED on the front


of the cradle marked “Primary” blinks during charging. During
charging, the diagonal lines in the battery icon on the instrument’s
display are animated and you see the message “Charging...”

When the instrument’s battery is fully charged, the battery icon is no


longer animated and shows a full battery. The message “Fully
charged!” is shown and the Primary LED on the cradle glows
continuously green.

121

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

Note: A spare Li-ion battery (059-3051-000) or NiMH (059-3054-


000) can be charged by placing it directly in the charging port on the
back of the cradle. It can be charged at the same time as the
instrument. Press the battery in place, sliding it slightly toward the
front of the cradle. This locks it in the cradle. To release the battery,
slide it forward again and tilt it up.

Note: An Alkaline Battery Adapter (part number 059-3052-000),


which uses four AA alkaline batteries (Duracell MN1500), may be
substituted for the Li-Ion battery.

WARNING!
To reduce the risk of ignition of hazardous atmospheres, recharge
and replace batteries only in areas known to be non-hazardous.
Remove and replace batteries only in areas known to be non-
hazardous.

Low Voltage Warning


When the battery’s charge falls below a preset voltage, the instrument
warns you by beeping once and flashing once every minute, and the
battery icon blinks once per second. You should turn off the
instrument within 10 minutes and either recharge the battery by
placing the instrument in its cradle, or replace the battery with a fresh
one with a full charge.

Clock Battery
An internal clock battery is mounted on one of the instrument’s
printed circuit boards. This long-life battery keeps settings in memory
from being lost whenever the Li-ion, NiMH, or alkaline batteries are
removed. This backup battery should last approximately five years,
and must be replaced by an authorized RAE Systems service
technician. It is not user-replaceable.

122

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

WARNING
To reduce the risk of ignition of hazardous atmospheres, recharge
battery only in area known to be non-hazardous. Remove and
replace battery only in an area known to be non-hazardous.

Replacing Rechargeable Li-Ion or NiMH


Battery
Caution: Turn off the instrument before removing or replacing the
battery.

Alkaline Battery Adapter


An alkaline battery adapter is supplied with each instrument. The
adapter (part number 059-3052-000) accepts four AA alkaline
batteries (use only Duracell MN1500).

Do not mix old and new batteries or different type batteries.

123

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

Troubleshooting
Problem Possible Reasons & Solutions
Cannot turn on power Reasons: Discharged battery.
after charging the Defective battery.
battery
Solutions: Charge or replace battery.
Lost password Solutions: Call Technical Support at
+1 408-752-0723 or toll-
free at
+1 888-723-4800
Reading abnormally Reasons: Dirty filter.
High Dirty sensor module.
Excessive moisture and
water condensation.
Incorrect calibration.

Solutions: Replace filter.


Blow-dry the sensor
module.
Calibrate the unit.
Reading abnormally Reasons: Dirty filter.
Low Dirty sensor module.
Weak or dirty lamp.
Incorrect calibration.

Solutions: Replace filter.


Remove Calibration
Adapter.
Calibrate the unit.
Check for air leakage.
Buzzer Reasons: Bad buzzer.
Inoperative
Solutions: Check that buzzer is not
turned off.
Call authorized service
center.

124

Downloaded from www.Manualslib.com manuals search engine


UltraRAE 3000 User’s Guide

Inlet flow too low Reasons: Pump diaphragm damaged


or has debris.
Flow path leaks.

Solutions: Check flow path for leaks;


sensor module O-ring, tube
connectors, Teflon tube
compression fitting.
Call Technical Support at
+1 408-752-0723
or toll-free at
+1 888-723-4800
“Lamp” message Reasons: Lamp drive circuit.
during operation Weak or defective PID
lamp, defective.

Solutions: Turn the unit off and back


on.
Replace UV lamp

125

Downloaded from www.Manualslib.com manuals search engine