Ultrarae 3000 User'S Guide: Rev. C August 2010 P/N 059-4023-000
Ultrarae 3000 User'S Guide: Rev. C August 2010 P/N 059-4023-000
User’s Guide
Ex-Ox-Tox Gasdetectie
Westerdreef 5V
Rev. C 2152 CS Nieuw-Vennep
Telefoon: 0252 620885
August 2010 E-mail: info@exoxtox.nl
P/N 059-4023-000 Website: www.exoxtox.nl
Downloaded from www.Manualslib.com manuals search engine
FCC Information
Contents
1 Standard Contents...................................................................9
2 General Information................................................................9
3 Physical Description .............................................................11
4 Specifications........................................................................11
5 Replacing Alkaline Batteries ................................................14
6 Charging A Lithium-Ion Battery ..........................................16
6.1 Charging A Spare Battery ............................................18
6.2 Low Voltage Warning..................................................18
6.3 Clock Battery ...............................................................19
6.4 Data Protection While Power Is Off ............................19
7 User Interface .......................................................................20
7.1 Display .........................................................................22
8 Operating The Instrument.....................................................23
8.1 Turning The Instrument On (Simple) ..........................23
8.2 Turning The Instrument On (Power On Zero) ............24
8.3 Turning The Instrument Off.........................................24
8.4 Operating The Built-In Flashlight................................24
8.5 Pump Status .................................................................25
8.6 Calibration Status.........................................................26
9 Operating Modes ..................................................................27
10 Compound-Specific Operation .............................................31
10.1 Compound-Specific Measurement...............................31
10.2 Measurement Phases ....................................................31
10.3 Performing A Measurement.........................................31
10.4 Separation Tube Preparation........................................32
10.5 Inserting The Separation Tube.....................................34
10.6 Measuring ....................................................................35
11 VOC Operation.....................................................................38
11.1 Basic User Level/Hygiene Mode (Default Settings)....38
12 Alarm Signals .......................................................................41
12.1 Alarm Signal Summary................................................42
12.2 Preset Alarm Limits & Calibration ..............................43
12.3 Testing The Alarm .......................................................43
13 Integrated Sampling Pump ...................................................43
14 Backlight...............................................................................44
15 Datalogging ..........................................................................44
15.1 Datalogging Event .......................................................44
22.6.2 Interval..................................................................................74
22.6.3 Data Selection ......................................................................75
22.6.4 Datalog Type........................................................................76
22.6.5 Manual Datalog....................................................................76
22.6.6 Snapshot Datalog.................................................................78
22.7 Monitor Setup ..............................................................78
22.7.1 Radio Power.........................................................................78
22.7.2 Op Mode...............................................................................79
22.7.3 Site ID...................................................................................79
22.7.4 User ID .................................................................................80
22.7.5 User Mode............................................................................81
22.7.6 Date.......................................................................................81
22.7.7 Time......................................................................................82
22.7.8 Duty Cycle ...........................................................................82
22.7.9 Temperature Unit.................................................................83
22.7.10 Pump Speed .......................................................................83
22.7.11 Language............................................................................84
22.7.12 Real Time Protocol............................................................84
22.7.13 Power On Zero...................................................................85
22.7.14 Unit ID................................................................................85
22.7.15 LCD Contrast.....................................................................86
22.7.16 Lamp ID .............................................................................86
23 Humidity Compensation.......................................................87
24 Hygiene Mode ......................................................................87
24.1 Basic User Level & Hygiene Mode .............................88
24.2 Entering Search Mode From Hygiene Mode ...............90
24.3 Optional Graphic Screen In Search Mode ...................91
25 Advanced User Level (Hygiene Mode Or Search Mode).....92
25.1 Advanced User Level & Hygiene Mode......................92
25.2 Basic User Level & Search Mode................................94
25.3 Advanced User Level & Search Mode ........................96
25.4 Diagnostic Mode..........................................................97
25.4.1 Entering Diagnostic Mode ..................................................97
25.4.2 Adjusting The Pump Stall Threshold.................................98
25.4.3 Pump High ...........................................................................98
25.4.4 Pump Low............................................................................98
25.4.5 Testing The Humidity Sensor.............................................99
25.4.6 Exiting Diagnostic Mode ................................................. 100
26 Transferring Data To & From A Computer........................102
CAUTION!
To reduce the risk of electric shock, turn the power off before
removing the instrument cover. Disconnect the battery before
removing sensor module for service. Never operate the instrument
when the cover is removed. Remove instrument cover and sensor
module only in an area known to be non-hazardous.
ATEX WARNING!
To reduce the risk of electrostatic ignition, do not use the instrument
without the rubber boot in place.
Special Notes
When the instrument is taken out of the transport case and
turned on for the first time, there may be some residual organic
or inorganic vapor trapped inside the detector chamber. The
initial PID sensor reading may indicate a few ppm. Enter an
area known to be free of any organic vapor and turn on the
instrument. After running for several minutes, the residual
vapor in the detector chamber will be cleared and the reading
should return to zero.
WARNINGS
STATIC HAZARD: Clean only with damp cloth.
AVERTISSEMENT
DANGER
RISQUE D'ORIGINE ELECTROSTATIQUE: Nettoyer
uniquement avec un chiffon humide.
Pour des raisons de sécurité, cet équipement doit être utilisé,
entretenu et réparé uniquement par un personnel qualifié. Étudier le
manuel d’instructions en entier avant d’utiliser, d’entretenir ou de
réparer l’équipement.
Utiliser seulement l'ensemble de batterie RAE Systems, référence
059-3051-000 ou 059-3052-000 ou 059-3054-000. Cet instrument n’a
pas été testé dans une atmosphère de gaz/air explosive ayant une
concentration d’oxygène plus élevée que 21%. La substitution de
composants peut compromettre la sécurité intrinsèque. Ne charger les
batteries que dans des emplacements désignés non-dangereux.
Ne pas mélanger les anciennes et les nouvelles batteries, ou bien
encore les batteries de différents fabricants.
L’étalonnage de tout instrument de RAE Systems doit être testé en
exposant l’instrument à une concentration de gaz connue avant de
mettre en service l’instrument pour la première fois.
1 Standard Contents
Instrument
Calibration Kit
Charging Cradle
AC/DC Adapter
Alkaline Battery Adapter
Data Cable
CD-ROM With User’s Guide, Quick Start Guide, and related materials
2 General Information
The UltraRAE 3000 is a hand-held, programmable compound specific
PID monitor designed to provide instantaneous exposure monitoring
of a specific organic gas. It monitors a specific gas by utilizing a gas
separation tube and the photoionization detector (PID) with a 9.8 eV
gas discharge lamp. It also can be used to measure total volatile
organic compound (VOC) as a broadband monitor by utilizing the
PID with a 9.8 eV, 10.6 eV, or 11.7 eV lamp.
Features include:
10
3 Physical Description
The main components of the portable compound-specific and
VOC monitoring instrument include:
• Three operation/programming keys for normal operation or
programming
• LCD display with back light for direct readout and calculated
measurements
• Built-in flashlight for illuminating testing points in dark
environments
• Buzzer and red LEDs for alarm signaling whenever exposures
exceed preset limits
• Charge contacts for plugging directly to its charging station
• Easy-to-use separation tube holder
• USB communication port for PC interface
• Protective rubber cover
4 Specifications
Size: 10" L x 3" W x 2.5" H
(25.5 cm x 7.6 cm x 6.4 cm)
Weight: 26 oz (738 g) with battery pack
Detector: Photoionization sensor with 9.8, 10.6, or
11.7 eV UV lamp
Battery: A 4.2V rechargeable Lithium-Ion battery
pack (snap in, field replaceable, at non-
hazardous location only)
Alkaline battery holder (for 4 AA batteries)
Battery Charging: Less than 8 hours to full charge
Operating Hours: Up to16 hours continuous operation
Display: Large dot matrix screen with backlight
11
12
13
14
IMPORTANT!
Alkaline batteries cannot be recharged. The instrument’s internal
circuit detects alkaline batteries and will not allow recharging. If you
place the instrument in its Travel Charger or Charger Stand, the
alkaline battery will not be recharged. The internal charging circuit is
designed to prevent damage to alkaline batteries and the charging
circuit when alkaline batteries are installed inside the instrument. If
you try to charge an alkaline batteries installed in the instrument, the
Charging Cradle or Travel Charger’s charging LED does not glow,
indicating that it will not charge the alkaline batteries.
15
Travel
Charger
Charger Stand
16
Note: If the instrument or battery has been charging for more than 10
hours and you see the “Battery Charging Error” icon and a message
that says, “Charging Too Long,” this indicates that the battery is not
reaching a full charge. Try changing the battery and make sure the
contacts on the instrument are meeting the Travel Charger’s (or
Charger Stand’s) contacts. If the message is still shown, consult your
distributor or RAE Systems Technical Services.
17
Release the battery from the Charger Stand by pulling it back toward
the rear of the Charger Stand and tilting it out of its slot.
Note: If you need to replace the Li-ion battery pack, replacements are
available from RAE Systems. The part number is 059-3051-000.
WARNING!
To reduce the risk of ignition of hazardous atmospheres, recharge
and replace batteries only in areas known to be non-hazardous.
18
instrument in its cradle, or replace the battery with a fresh one with a
full charge.
19
7 User Interface
The instrument’s user interface consists of the display, LEDs, an
alarm transducer, and four keys. The keys are:
Y/+
MODE
N/-
Flashlight on/off
The LCD display provides visual feedback that includes the reading,
time, battery condition, and other functions.
Display
Y/+ key
N/- key
MODE key
Flashlight
on/off key
20
Three panes along the bottom of the display are “mapped” to the
keys. These change as menus change, but at all times the left pane
corresponds to the [Y/+] key, the center pane corresponds to the
[MODE] key, and the right pane corresponds to the [N/-] key. Here
are three examples of different menus with the relationships of the
keys clearly shown:
21
7.1 Display
22
The RAE Systems logo should appear first. (If the logo does not
appear, there is likely a problem and you should contact your
distributor or RAE Systems Technical Support.) The instrument is
now operating and performs self tests. If any tests (including sensor
and memory tests) fail, refer to the Troubleshooting section of this
guide.
23
Note: You must hold your finger on the key for the entire shutoff
process. If you remove your finger from the key during the
countdown, the shutoff operation is canceled and the instrument
continues normal operation.
Note: Using the flashlight for extended periods shortens the battery’s
operating time before it needs recharging.
24
During duty cycling (PID lamp cleaning), the display shows these
icons in alternation:
25
• The lamp type has been changed (for example, from 10.6 eV
to 9.8 eV).
• The sensor module has been replaced.
• It has been 30 days or more since the instrument was last
calibrated.
• If you have changed the calibration gas type without
recalibrating the instrument.
26
9 Operating Modes
The UltraRAE is actually two monitors in one:
• Compound-specific monitor
• VOC monitor
The diagram on the next page shows the basic flow of the UltraRAE
3000’s functions. The area with the gray field is the compound-
specific (tube) mode, while the rest shows VOC mode. Navigate
through the steps by using the [Y/+] and [N/-] keys as shown in the
diagram.
27
Compound-Specific (Tube)
28
29
30
10 Compound-Specific Operation
Make sure the UltraRAE 3000 is set to operate with your selected tube:
31
Before you start sampling, you must insert a RAE-Sep separation tube
into the inlet/holder. Follow the Separation Tube Preparation and Placing
A Tube Into The UltraRAE 3000 instructions before pressing any
buttons on the UltraRAE 3000. Once the tube is in place, then proceed to
measuring.
IMPORTANT!
Once a tube’s ends are broken off, the material inside is exposed.
Therefore, use the tube for sampling as soon as possible.
CAUTION!
Wear hand and eye protection when breaking tube tips. Use caution
in handling tubes with broken ends. Keep away from children. RAE-
Sep tubes should be disposed of according to local regulations. See
footnotes of data sheets for disposal information.
32
CAUTION!
Handle tubes with care. Tube ends are sharp after ends are broken off.
33
IMPORTANT!
Do not overtighten any portion of the sampling assembly.
Note: When the UltraRAE 3000 is used for VOC monitoring, no tube
is inserted.
IMPORTANT!
To ensure that there are no leaks, periodically test the seals:
With the UltraRAE 3000 running, place your finger over the end of
the inlet probe. The alarm should sound and the pump-stall icon
should flash on the display. This indicates that all seals are good. Stop
the alarm by pressing [Y/+]. If the pump does not alarm or show the
stalled-pump icon, then check that all inlet parts are tight and inspect
the O-ring for damage (replace it if necessary).
34
10.6 Measuring
Once the tube is in place, begin measuring by pressing [Y/+].
The display shows a countdown (60 seconds is shown here, but sampling
time depends on the type of separation tube selected and the temperature):
Note: You can abort the sampling by pressing [N/-] at any time.
Once the countdown is complete, the reading is shown:
WARNING!
At least 1/4 of the tube should still be yellow-orange at the bottom. If
not, the STEL value is not valid. Abort the measurement and change
the tube. Then do a snapshot test instead of a STEL test. Note: If the
STEL is exceeded, the UltraRAE 3000 goes into alarm.
35
If you press [N/-] to return to the main menu, which shows the tube
type instead of the CF (correction factor):
If you press [Y/+], you are asked, “Clear peak value! Are You Sure?”
to confirm:
Press [Y/+] to clear the Peak value and exit to VOC operation.
36
Remove the tube and put the inlet back together. Then press [N/-].
This display is shown:
After a few seconds, the UltraRAE 3000 enters VOC mode and
shows this display:
You can step through the rest of the steps by pressing [N/-] repeatedly
until you reach the main menu again.
37
11 VOC Operation
11.1 Basic User Level/Hygiene Mode
(Default Settings)
Pressing [N/-] steps you from one screen to the next, and eventually return to
the main display. If you do not press a key within 60 seconds after entering a
display, the instrument reverts to its main display.
Note: While viewing any of these screens, you can shut off your instrument by
pressing [MODE].
38
39
After the instrument is turned on, it runs through the start-up menu. If
the UltraRAE 3000 is set for “Power On Zero,” then the message
“Apply zero gas…” is displayed.
At this point, you can perform a zero air (fresh air) calibration. If the
ambient air is clean, you can use that. Otherwise, use a cylinder of
zero air. Refer to Zero Calibration on page 51 for a more detailed
description of zero calibration.
Start zero calibration by pressing Start [Y/+]. You see the message
“Zeroing…” followed by a 30-second countdown.
Note: You can press [MODE] to quit, bypassing the zero air
calibration.
When zero calibration is complete, you see the message:
Zeroing is done!
Reading = 0.0 ppm
The instrument is now sampling and collecting data.
Note: At the Average & Peak, Date & Time & Temperature, Calibration
Gas & Measurement Gas & Correction Factor, and PC Communications
screens, the instrument automatically goes to the main display after 60
seconds if you do not push a key to make a selection.
40
12 Alarm Signals
During each measurement period, the gas concentration is compared
with the programmed alarm limits (gas concentration alarm limit
settings). If the concentration exceeds any of the preset limits, the
loud buzzer and red flashing LED are activated immediately to warn
you of the alarm condition.
In addition, the instrument alarms if one of the following conditions
occurs: battery voltage falls below a preset voltage level, failure of
the UV lamp, or pump stall.
41
42
43
14 Backlight
The LCD display is equipped with an LED backlight to assist in
reading the display under poor lighting conditions.
15 Datalogging
During datalogging, the instrument displays a disk icon to indicate
that datalogging is enabled. The instrument stores the measured gas
concentration at the end of every sample period (when data logging is
enabled). In addition, the following information is stored: user ID,
site ID, serial number, last calibration date, and alarm limits. All data
are retained (even after the unit is turned off) in non-volatile memory
so that it can be down- loaded at a later time to a PC.
15.1 Datalogging Event
When Datalogging is enabled, measurement readings are being saved.
These data are stored in “groups” or “events.” A new event is created
and stored each time the instrument is turned on and is set to
automatic datalogging, or a configuration parameter is changed, or
datalogging is interrupted. The maximum time for one event is 24
hours or 28,800 points. If an event exceeds 24 hours, a new event is
automatically created. Information, such as start time, user ID, site
ID, gas name, serial number, last calibration date, and alarm limits are
recorded.
15.2 Datalogging Sample
After an event is recorded, the unit records a shorter form of the data.
When transferred to a PC running ProRAE Studio, this data is
arranged with a sample number, time, date, gas concentration, and
other related information.
44
Note: You can only choose one datalog type to be active at a time.
16 Accessories
The following accessories are included with the instrument:
• AC Adapter (Battery Charger)
• Travel Charger
• Alkaline battery adapter
• External Filter
• Organic Vapor Zeroing kit
45
WARNING
To reduce the risk of ignition of hazardous atmospheres, recharge
battery only in area known to be non-hazardous. Remove and
replace battery only in area known to be non-hazardous.
Ne charger les batteries que dans emplacements designés non-
dangereuses.
A battery charging circuit is built into the instrument cradle. It only needs a
regular AC to 12 VDC adapter (wall-mount transformer, part number 500-
0114-000) to charge the instrument.
To charge the battery inside the instrument:
1. Power off the instrument.
2. Connect the AC adapter to the DC jack on the instrument’s cradle.
If the instrument is off, it automatically turns on.
3. While charging, the display message shows “Charging.” The
Primary LED on the cradle flashes green when charging.
4. When the battery is fully charged, the LED changes to glowing
green continuously, and the message “Fully charged” appears on
the display. If there is a charging error, the LED glows red
continuously.
A completely discharged instrument can be charged to full capacity
within 8 hours. Batteries drain slowly even if an instrument is off.
Therefore, if the instrument has been in storage or has not been
charged for several days or longer, check the charge before using it.
The factory-supplied battery is designed to last for 16 hours of normal
operation (no alarm), for a new battery under the optimum
circumstances. As the battery becomes older or is subject to adverse
conditions (such as cold ambient temperature), its capacity will be
significantly reduced.
46
47
18 Optional Accessories
18.1 Calibration Adapter
The calibration adapter for the instrument is a simple 6-inch Tygon
tubing with a metal adapter on one end. During calibration, simply
insert the metal adapter into the regular gas inlet probe of the
instrument and the tubing to the gas regulator on the gas bottle.
18.2 Calibration Regulator
The Calibration Regulator is used in the calibration process. It
regulates the gas flow rate from the Span gas cylinder into the gas
inlet of the instrument during calibration process. The maximum flow
rate allowed by the flow controller is about 0.5L/min (500 cc per
min.). Alternatively, a demand-flow regulator or a Tedlar gas bag
may be used to match the pump flow precisely.
18.3 Organic Vapor Zeroing Kit
The Organic Vapor Zeroing Kit is used for filtering organic air
contaminants that may affect the zero calibration reading. To use the
Organic Vapor Zeroing Kit, simply connect the filter to the inlet port
of the instrument.
48
49
1. Press and hold [MODE] and [N/-] until you see the Password
screen.
50
If you have pressed [Y/+] to enter Zero calibration, then you will see
this message:
51
Zeroing...
Note: To abort the zeroing process at any time and proceed to Span
calibration, press [N/-] at any time while zeroing is being performed.
You will see a confirmation message that says “Zero aborted!” and
then the Span calibration menu appears.
Zeroing is done!
Reading = 0.0 ppm
The instrument will then show the Calibration menu on its display,
with Span Calib highlighted.
52
If you have pressed [Y/+] to enter Span calibration, then you will see
the name of your Span gas (the default is isobutylene) and the span
value in parts per million (ppm).
IMPORTANT!
If you are using the UltraRAE 3000 to test for benzene, it is
recommended that you calibrate with 5 ppm benzene calibration gas
from RAE Systems.
Note: You may press [MODE] if you decide that you do not
want to initiate calibration. This will abort the span
calibration and take you directly to the Calibration menu for
Zero calibration.
Calibrating...
53
Note: If you want to abort the Span calibration process, press [N/-] at
any time during the process. You will see a confirmation message that
says “Span is aborted!” and then the Zero calibration menu appears.
You can then proceed to perform a Zero calibration, perform a Span
calibration, or exit to the topmost Calibration menu.
Span 1 is done!
Reading = 100.0 ppm
The instrument then exits Span calibration and shows the Zero
calibration menu on its display.
Note: The reading should be very close to the span gas value.
54
Updating settings…
The instrument updates its settings and then returns to the main
display. It begins or resumes monitoring.
55
20 Three-Point Calibration
For enhanced accuracy, it is possible to perform a second Span
calibration in addition to the Zero and Span calibrations outlined in
the previous section. Your instrument first must be set to allow this
third calibration. This requires using ProRAE Studio software and a
PC, as well as a higher concentration of calibration gas.
Note: Once the third calibration is set, you do not need to use
ProRAE Studio to allow future 3-point calibrations. Also, you can
only disable 3-point calibration capability by using ProRAE Studio
again.
Perform the Zero and Span calibrations. After the first Span
calibration (Span 1) is completed, the display a second Span
calibration (Span 2) can be performed. The process is identical to the
first calibration. As in the Span 1 calibration, you may exit and return
to the Zero calibration screen if you choose not to perform this
calibration or to abort it.
56
57
Note: This gas should be of a higher concentration than the gas used
for Span 1 calibration.
Choose the 500 cc/min. regulator only if the flow rate matches or
slightly exceeds the flow rate of the instrument pump. Alternatively,
the span gas can first be filled into a Tedlar bag or delivered through a
demand-flow regulator. Connect the calibration adapter to the inlet
port of the instrument, and connect the tubing to the regulator or
Tedlar bag.
Another alternative is to use a regulator with >500 cc/min flow but
allow the excess flow to escape through a T or an open tube. In the
latter method, the span gas flows out through an open tube slightly
wider than the probe, and the probe is inserted into the calibration
tube.
At the Span Calibration menu, you perform a Span calibration. You
may also go back to the Zero calibration menu or to the initial
Calibration menu if you want to exit calibration.
If you have pressed [Y/+] to enter Span calibration, then you will see
the name of your Span gas (the default is isobutylene) and the span
value in parts per million (ppm). You will also see this message that
prompts you:
58
Note: You may press [MODE] if you decide that you do not
want to initiate calibration. This will take you directly to the
Calibration menu for Zero calibration.
Calibrating...
Note: If you want to abort the Span calibration process, press [N/-] at
any time during the process. You will see a confirmation message that
says “Span is aborted!” and then the Zero calibration menu will
appear. You can then proceed to perform a Zero calibration, perform
a Span calibration, or exit to the topmost Calibration menu.
Span 2 is done!
Reading = 1000 ppm
The instrument then exits Span calibration and shows the Zero
calibration menu on its display.
Note: The reading should be very close to the span gas value.
Updating settings…
59
The instrument updates its settings and then returns to the main
display. It begins or resumes monitoring.
21 Programming Mode
1. Press and hold [MODE] and [N/-] until you see the Password
screen.
If you make a mistake, you can cycle through the digits by pressing
[N/-] and then using [Y/+] to change the number in each position.
60
Note: When you reach Monitor Setup and press [N/-], the menu
cycles back to Calibration.
61
62
Once you enter Programming Mode, the LCD displays the first menu,
Calibration. Each subsequent menu is accessed by pressing [N/-]
repeatedly until the desired menu is displayed. To enter a sub-menu
of a menu, press [Y/+].
63
22.3 Calibration
Two types of calibration are available: Zero (fresh air) and Span.
64
22.4 Measurement
65
66
67
In this menu, you can change the High and Low alarm limits, the
STEL limit, and the TWA. Press [Y/+] to to enter the Alarm Setting
menu. Note: All settings are shown in ppb (parts per billion), or
µg/m3 (micrograms per cubic meter), depending on your setting.
1. Scroll through the Alarm Limit sub-menu using the [N/-] key
until the display shows the desired limit to be changed (High
Alarm, Low Alarm, STEL Alarm, and TWA Alarm)
2. Press [Y/+] to select one of the alarm types. The display
shows a flashing cursor on the left-most digit of the
previously stored alarm limit.
3. Press [Y/+] to increase each digit’s value.
4. Press [N/-] to advance to the next digit.
5. Again, use [Y/+] to increase the number.
Repeat this process until all numbers are entered.
Press [MODE] when you are done.
68
You can change the High Alarm limit value. The value is typically set
by the instrument to match the value for the current calibration gas. It
is expressed in parts per billion (ppb). Note: The default value
depends on the measurement gas.
You can change the Low Alarm limit value. The value is typically set
by the instrument to match the value for the current calibration gas. It
is expressed in parts per billion (ppb). Note: The default value
depends on the measurement gas.
69
When you have completed your selections, press [MODE]. You will
see two choices: Save and Undo. You have the opportunity to register
the new settings or to change your mind and revert to your previous
settings.
You can change the STEL Alarm limit value. The value is typically
set by the instrument to match the value for the calibration gas. It is
expressed in parts per billion (ppb). Note: The default value depends
on the measurement gas.
70
You can change the TWA (time-weighted average) Alarm limit value.
The value is typically set by the instrument to match the value for the
calibration gas. It is expressed in parts per billion (ppb). Note: The
default value depends on the measurement gas.
• Save
• Undo
71
You will see two choices: Save and Undo. You have the
opportunity to register the new settings or to change your mind
and revert to your previous settings.
72
• Both on
• Light only
• Buzzer only
• Both off
You will see two choices: Save and Undo. You have the
opportunity to register the new settings or to change your mind
and revert to your previous settings.
22.6 Datalog
The instrument calculates and stores the concentration and ID of each
sample taken. In the datalog sub-menu, a user can perform the tasks
and functions shown below.
73
1. Scroll through the Datalog sub-menu using the [N/-] key until the
display shows the desired parameter to be changed:
Clear Datalog
Interval
Data Selection
Datalog Type
Press [Y/+] to clear the datalog. The display asks, “Are you sure?”
• Press [Y/+] if you want to clear the datalog. When it has been
cleared, the display shows “Datalog Cleared!”
• Press [N/-] if you do not want to clear the datalog.
The display changes, and you are taken to the next sub-menu,
Interval.
22.6.2 Interval
74
You will see two choices: Save and Undo. You have the opportunity
to register the new settings or to change your mind and revert to your
previous settings.
Data Selection allows you to select which types of data are stored and
made available when you offload your datalog to a computer via
ProRAE Studio software.
You can choose any or all of three types of data (you must choose at
least one):
• Average
• Maximum
• Minimum
1. Press [N/-] to step from one option to the next. The highlighter
indicates your choice.
2. Press [Y/+] to toggle your selection on or off (the check box
indicates “on” with an “X”).
3. When you have completed your selections, press [MODE].
You will see two choices: Save and Undo. You have the opportunity
to register the new settings or to change your mind and revert to your
previous settings.
75
You will see two choices: Save and Undo. You have the opportunity
to register the new settings or to change your mind and revert to your
previous settings.
• When you reach the screen that says “Start Datalog?” press
[Y/+] to start it. You see “Datalog Started,” confirming that
datalogging is now on.
When you reach the screen that says “Stop Datalog?” press [Y/+] to
stop it. You see “Datalog Stopped,” confirming that datalogging is
now off.
76
77
When you send the data to a computer using ProRAE Studio, the data
snapshots are uniquely identified by time and other parameters.
Many settings can be accessed in this menu, including setting the date
and time and adjusting the pump’s on/off duty cycle.
1. Press [N/-] to step from one option to the next (on or off).
2. Press [Y/+] to make your selection (the dark circle in the “radio
button” indicates that the option is selected).
3. When you have completed your selection, press [MODE].
78
22.7.2 Op Mode
Hygiene
Search
22.7.3 Site ID
1. Press [Y/+] and the display shows the current site ID.
Example: “RAE00001.” Note that the left-most digit flashes
to indicate it is the selected one.
2. Press [Y/+] to step through all 26 letters (A to Z) and 10
numerals (0 to 9).
Note: The last four digits must be numerals.
3. Press [N/-] to advance to the next digit. The next digit to the
right flashes.
79
Repeat this process until all eight digits of the new site ID are
entered.
Press [MODE] to exit.
If there is any change to the existing site ID, the display shows
“Save?” Press [Y/+] to accept the new site ID. Press [N/-] to discard
the change and move to the next sub-menu.
22.7.4 User ID
1. Press [Y/+] and the display shows the current User ID.
Example: “RAE00001.” Note that the left-most digit flashes
to indicate it is the selected one.
2. Press [Y/+] to step through all 26 letters (A to Z) and 10
numerals (0 to 9).
3. Press [N/-] to advance to the next digit. The next digit to the
right flashes.
Repeat this process until all eight digits of the new User ID
are entered.
Press [MODE] to exit.
If there is any change to the existing User ID, the display shows
“Save” Press [Y/+] to accept the new site ID. Press [N/-] to discard
(undo) the change and move to the next sub-menu.
80
1. Press [N/-] to step from one option to the next. The highlighting
changes each time you press [N/-].
2. Press [Y/+] to make your selection (the dark circle in the “radio
button” indicates “on”).
3. When you have completed your selection, press [MODE].
4. Press [Y/+] to accept the new User Mode. Press [N/-] to discard
the change and move to the next sub-menu.
22.7.6 Date
1. Press [Y/+] and the display shows the current date. Note that
the left-most digit flashes to indicate it is selected.
2. Press [Y/+] to step through all 10 numerals (0 to 9).
3. Press [N/-] to advance to the next digit. The next digit to the
right flashes.
Repeat this process until all six digits of the new date are
entered.
Press [MODE] to exit.
• Press [Y/+] to save the new date.
• Press [N/-] to undo the change and move to the next sub-
menu.
81
22.7.7 Time
1. Press [Y/+] and the display shows the current time. Note that
the left-most digit flashes to indicate it is selected.
2. Press [Y/+] to step through all 10 numerals (0 to 9).
3. Press [N/-] to advance to the next digit. The next digit to the
right flashes.
Repeat this process until all six digits of the new time are
entered.
Press [MODE] to exit.
• Press [Y/+] to save the new date.
• Press [N/-] to undo the change and move to the next sub-
menu.
The pump’s duty cycle is the ratio of its on time to off time. The duty
cycle ranges from 50% to 100% (always on; this is the default value),
and the period is 10 seconds. Therefore, a duty cycle of 60% means that
the pump is on for 6 seconds and off for four seconds. Duty cycling is
employed by the instrument to clean the PID. A lower duty cycle has a
greater effect on keeping the PID clean than a higher duty cycle.
82
The pump can operate at two speeds, high and low. Running at low
speed is quieter and conserves a small amount of power. There is
almost no difference in sampling accuracy.
83
22.7.11 Language
84
22.7.14 Unit ID
Note: If you are using an UltraRAE 3000 with a RAELink3, this unit
ID is displayed by the RAELink3.
85
22.7.16 Lamp ID
The UltraRAE can automatically identify the type of lamp, or you can
select a lamp type manually.
86
23 Humidity Compensation
The UltraRAE 3000 has a humidity sensor and humidity
compensation circuitry. By default it is on, but it can be turned off or
on by using ProRAE Studio software. RAE Systems recommends
testing the humidity sensor once a year. See page 99 for the simple
procedure.
24 Hygiene Mode
The instrument usually operates in Hygiene Mode, which provides
basic functionality. However, it is possible to operate it in a second
mode called Search Mode. Here are the primary differences:
87
88
89
1. Hold [MODE] and [N/-] until you see the password screen.
2. Use [Y/+] to increment to the number you want for the first
digit. (If you pass by the desired number, press [Y/+] until it
cycles through to 0 again. Then press [Y/+] until you reach
the desired number.)
3. Press [N/-] to advance to the next digit.
4. Again press [Y/+] to increment the number.
5. Press [N/-] to advance to the next digit.
Continue the process until all four numbers of the password have
been input. Then press [MODE] to proceed.
Hygiene
Search
90
91
Once you are in Advanced User Level and Hygiene Mode together,
you can change the calibration reference and measurement gas, in
addition to performing normal monitoring functions.
92
93
94
95
96
Press and hold [Y/+] and [MODE] until the instrument starts.
The instrument goes through a brief startup, and then displays raw
data for the PID sensor. These numbers are raw sensor readings
without calibration. The instrument is now in Diagnostic Mode.
Note: In Diagnostic Mode, the pump and lamp are normally on.
97
Block the gas inlet and watch the pump current reading (labeled “I”)
increase. Write down its blocked reading. Note: If the pump current
reading does not increase significantly (less than 10 counts), then
there may be a leak in the gas inlet or the pump is weak or defective.
Add the two readings you wrote down. This is the average of the
maximum block count and the maximum idle count. Divide that
number by 2. Use the [Y/+] or [N/-] key to increase or decrease the
stall value to equal that number.
Block the gas inlet and watch the pump current reading (labeled “I”)
increase. Write down its blocked reading. Note: If the pump current
reading does not increase significantly (less than 10 counts), then
there may be a leak in the gas inlet or the pump is weak or defective.
Add the two readings you wrote down. This is the average of the
maximum block count and the maximum idle count. Divide that
98
There are three numbers for the humidity reading (“H”). The first
number is the current humidity reading from the sensor. The
second is the reference number for 0% humidity, and the third
number is the reference for 100% humidity.
99
25° C / 77° F
100
101
When the transfer is done, you will see a screen with the datalog
information. You can now export this datalog for other use or
printing.
102
103
27 Maintenance
The major maintenance items of the instrument are:
• Battery pack
• Sensor module
• PID lamp
• Sampling pump
• Inlet connectors and filters
In addition, you can test the humidity sensor (this should be done
annually, in order to ensure the most accurate operation).
Note: Maintenance should be performed by qualified personnel
only.
NOTE: The printed circuit board of the instrument is connected
to the battery pack even if the power is turned off. Therefore, it is
very important to disconnect the battery pack before servicing or
replacing any components inside the instrument. Severe damage
to the printed circuit board or battery may occur if the battery
pack is not disconnected before servicing the unit.
104
105
IMPORTANT!
Alkaline batteries cannot be recharged. The instrument’s internal
circuit detects alkaline batteries and will not allow recharging. If you
place the instrument in its cradle, the alkaline battery will not be
recharged. The internal charging circuit is designed to prevent
damage to alkaline batteries and the charging circuit when alkaline
batteries are installed inside the instrument.
WARNING!
To reduce the risk of ignition of hazardous atmospheres, recharge the
battery only in areas known to be non-hazardous. Remove and replace
the battery only in areas known to be non-hazardous.
106
Sensor Components
Note: The cleaning procedure is not normally needed. Clean the PID
sensor module, the lamp and the lamp housing only if:
1. The reading is inaccurate even after calibration.
2. The reading is very sensitive to air moisture.
3. A liquid has been sucked into the unit and damaged the unit.
Use of the external filter helps to prevent contamination of the sensor.
To access the sensor components and lamp, gently unscrew the lamp-
housing cap, remove the sensor adapter with the gas inlet probe and
the metal filter all together. Then hold the PID sensor and pull it
straight out. A slight, gentle rocking motion helps release the sensor.
Note: The 10.6eV lamp requires a Teflon O-ring. The 9.8eV and
11.7eV lamps do not require the O-ring.
107
108
Remove the lamp and look for markings (bars) on the side:
• No bars: 10.6eV
• 1 bar: 11.7eV
• 2 bars: 9.8eV
Also, when the monitor is running, the lamp type is shown along with
the calibration and measurement gas and Correction Factor:
109
110
or
111
28 Troubleshooting
Problem Possible Reasons & Solutions
Cannot turn on power Reasons: Discharged battery.
after charging the Defective battery.
battery
Solutions: Charge or replace battery.
Lost password Solutions: Call Technical Support at
+1 408-752-0723 or toll-
free at
+1 888-723-4800
Reading abnormally Reasons: Dirty filter.
High Dirty sensor module.
Excessive moisture and
water condensation.
Incorrect calibration.
112
29 Technical Support
To contact RAE Systems Technical Support Team:
113
E-mail: customerserv@raesystems.com
Web Site: www.raesystems.com
114
RAE BeNeLux BV
Hoofdweg 34C
2908 LC Capelle a/d IJssel
The Netherlands
Phone: +31 10 4426149
Fax: +31 10 4426148
Email: info@rae.nl
Web: www.rae.nl
115
116
Basic Operation
The instrument is now operating and performs self tests. Once the self
tests are complete, the display shows a graph or numerical gas
reading. This indicates that the instrument is fully functional and
ready to use.
Note: You must hold your finger on the key for the entire shutoff
process. If you remove your finger from the key during the
117
118
Alarm Signals
During each measurement period, the gas concentration is compared
with the programmed alarm limits (gas concentration alarm limit
settings). If the concentration exceeds any of the preset limits, the
loud buzzer and red flashing LED are activated immediately to warn
you of the alarm condition.
In addition, the instrument alarms if one of the following conditions
occurs: battery voltage falls below a preset voltage level, failure of
the UV lamp, pump stall, or when the datalog memory is full.
119
120
Note: Before setting the instrument into its charging cradle, visually
inspect the contacts to make sure they are clean. If they are not, wipe
them with a soft cloth. Do not use solvents or cleaners.
Note: To release the instrument, press down and tilt the top out
of the cradle and lift up.
121
WARNING!
To reduce the risk of ignition of hazardous atmospheres, recharge
and replace batteries only in areas known to be non-hazardous.
Remove and replace batteries only in areas known to be non-
hazardous.
Clock Battery
An internal clock battery is mounted on one of the instrument’s
printed circuit boards. This long-life battery keeps settings in memory
from being lost whenever the Li-ion, NiMH, or alkaline batteries are
removed. This backup battery should last approximately five years,
and must be replaced by an authorized RAE Systems service
technician. It is not user-replaceable.
122
WARNING
To reduce the risk of ignition of hazardous atmospheres, recharge
battery only in area known to be non-hazardous. Remove and
replace battery only in an area known to be non-hazardous.
123
Troubleshooting
Problem Possible Reasons & Solutions
Cannot turn on power Reasons: Discharged battery.
after charging the Defective battery.
battery
Solutions: Charge or replace battery.
Lost password Solutions: Call Technical Support at
+1 408-752-0723 or toll-
free at
+1 888-723-4800
Reading abnormally Reasons: Dirty filter.
High Dirty sensor module.
Excessive moisture and
water condensation.
Incorrect calibration.
124
125
Rev. C
August 2010
P/N 059-4023-000