3001 Manual
3001 Manual
Operator’s Manual
3001 Operator’s Manual
1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.1 Customer Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.2 Standard Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.3 Options and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.4 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.5 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
2. Calibrator Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
2.1 Front Panel Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
2.2 Primary Input/Output Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
2.3 Primary Input/Output Display and Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
2.4 Isolated Input Display, Controls, and Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
2.5 Rear Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
2.6 Display Layouts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
2.7 Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
3. Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
4. Primary Inputs and Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
4.1 DC Voltage Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
4.2 DC Current Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
4.3 Resistance Temperature Detector (RTD) and Ohms Measure . . . . . . . . . . . . . . . . . . . . .16
4.4 Resistance Temperature Detector (RTD) and Ohms Source . . . . . . . . . . . . . . . . . . . . . . .18
4.5 Resistance Temperature Detector (RTD) with Custom Coefficients . . . . . . . . . . . . . . . . .19
4.6 Standard Platinum Resistance Thermometer (SPRT) Coefficients . . . . . . . . . . . . . . . . . . .20
4.7 Thermocouple (T/C) Measure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
4.8 Thermocouple (TC) Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
4.9 Pressure Measure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
5. Isolated Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
5.1 Voltage Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
5.2 Current Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
5.3 Pressure Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
6. Output Setpoints . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
7. Application Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
7.1 P/I Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
7.2 I/P Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
7.3 V/I Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
7.4 RTD Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
7.5 RTD Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
7.6 Thermocouple Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
7.7 Thermocouple Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
7.8 RTD Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
7.9 Precision Current Trip . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34
7.10 I/I Isolator/Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
7.12 Precision Temperature Measurement with IBP-2 Probe . . . . . . . . . . . . . . . . . . . . . . . . .36
8. LCD and Remote Interface Setup Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
9. Remote Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
9.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
9.2 Setting up the RS-232 Port for Remote Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38
9.3 Setting up the IEEE-488 Port for Remote Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38
9.4 Changing Between Local and Remote Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
9.5 IEEE-488 Interface Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40
9.6 Using Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41
9.7 Checking 3001 Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
10. Remote Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50
10.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50
10.2 Command Summary by Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50
10.3 Error Code Listing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52
10.4 Remote Command Listing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54
11. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77
11.1 Cleaning the Calibrator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77
11.2 Replacing a Line Fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77
11.3 Changing the Line Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .78
12. Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .79
12.1 General Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .79
12.2 DC Voltage Specifications, Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .80
12.3 DC Voltage Specifications, Isolated Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .80
12.4 DC Current Specifications, Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .80
12.5 DC Current Specifications, Isolated Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .81
12.6 Resistance Specifications, Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .81
12.7 Resistance Specifications, Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .81
12.8 Thermocouple Specification, Output and Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .82
12.9 RTD and Thermistor Specification, Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .84
12.10 RTD and Thermistor Specification, Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .86
12.11 Pressure Measurement Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .87
13. Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .88
1. Introduction
The Martel series 3001 calibrator is an accurate full-featured temperature, pressure and
DC calibrator intended for R & D, manufacturing and calibration lab applications. The
unit's simple design and ease of operation allow users to quickly familiarize themselves
with its operations and features. Time saving functions like the ability to save, recall and
automatically cycle through setpoints for each output range, the ability to enter user
definable RTD curves, and a complete remote interface are several key features offered
by the 3001.
1.4 Unpacking
Upon receipt of the shipment, inspect the container and equipment for any signs of
damage. Take particular note of any evidence of rough handling in transit. Immediately
report any damage to the shipping agent.
NOTE: The carrier will not honor any claims unless all shipping material is saved for their
examination.
After examining and removing the contents, save the packing material and carton in the
event that re-shipment is necessary.
Remove the Packing List and verify that all of the listed equipment has been received. If
there are any questions about the shipment, please call Martel Electronics at 1-800-821-
0023.
Check to see if your basic calibrator package is complete. It should include:
• 3001 Calibrator
• Instruction Manual
1
• AC Line Cord
• Thermocouple Shorting Jumper
• NIST Certificate
Symbols Used
The following table lists the International Electrical Symbols. Some or all of these symbols
may be used on the instrument or in this manual.
Symbol Description
AC (Alternating Current)
AC-DC
Battery
DC
Double Insulated
Electric Shock
Fuse
PE Ground
Off
On
2
The following definitions apply to the terms "Warning" and "Caution".
• " Warning" identifies conditions and actions that may pose hazards to the user.
• " Caution" identifies conditions and actions that may damage the instrument being
used.
Use the calibrator only as specified in this manual, otherwise personal injury and/or
damage to the calibrator may occur.
Warning
To avoid possible electric shock or personal injury:
• Do not apply more than the rated voltage between the terminals, or between any
terminal and chassis ground. See specifications for supported ranges.
• Follow all equipment safety procedures.
• Always use the power cord and connector suitable for the voltage and outlet of the
location in which you are working.
• Do not use the calibrator if it is damaged. Before you use the calibrator, inspect the
case. Look for cracks or missing plastic. Pay particular attention to the insulation
surrounding the connectors.
• Inspect the test leads for damaged insulation or exposed metal. Check test lead
continuity. Replace damaged test leads before you use the calibrator.
• Never operate the calibrator with the cover removed or the case open, and never
remove the cover or open the case without first disconnecting the power source and all
test leads.
• Select the proper function and range for your measurement.
• Never touch the probe to a voltage source when the test leads are plugged into the
current terminals.
• When using the probes, keep your fingers away from the probe contacts. Keep your
fingers behind the finger guards on the probes.
• Connect the common test lead before you connect the live test lead. When you
disconnect test leads, disconnect the live test lead first.
• Do not use the calibrator if it operates abnormally. Protection may be impaired. When
in doubt, have the calibrator serviced.
• Do not operate the calibrator around explosive gas, vapor, or dust.
• When using a pressure module, make sure the process pressure line is shut off and
depressurized before you connect it, or disconnect, it from the pressure module.
• Disconnect test leads before changing to another measure or source function.
• Have the calibrator serviced only by qualified personnel, and use only specified
replacement parts.
• Use only the replacement fuse(s) specified in this manual.
• To avoid a violent release of pressure in a pressurized system, shut off the valve and
slowly bleed off the pressure before you attach the pressure module to the pressure
line.
3
Caution
To avoid possible damage to the calibrator or to the equipment under test:
• Use the proper terminals, function, and range for your measurement or sourcing
application.
• To avoid mechanically damaging the pressure module, never apply more than 10 ft-lb
of torque between the pressure module fittings, or between the fittings and the body of
the module.
• To avoid damaging the pressure module from overpressure, never apply pressure
above the rated maximum printed on the module.
• To avoid damaging the pressure module from corrosion, use it only with specified
materials. Refer to the pressure module documentation for material compatibility.
2. Calibrator Description
4
2.2 Primary Input/Output Terminals
Figure 2 describes the primary input/output terminals in detail. The display and controls
for these terminals are described in the next section.
Note 1: These terminal binding posts are made of a special copper alloy to reduce thermal
EMF's. They support the use of either discreet wires or standard banana plugs, and the
HI/LO pairs are spaced for standard dual banana plugs.
Note 2: Caution. Do not exceed a maximum of 100 volts to chassis ground.
Note 3: Caution. Do not exceed a maximum of 20 volts to chassis ground.
5
2.3 Primary Input/Output Display and Controls
Figure 3 describes the primary input/output display and controls in detail.
LOCAL Press to regain local control of the 3001 after the remote
command REMOTE has been received; in this case all
keys except this one are ignored. When the remote com-
mand LOCKOUT has been received, all keys are ignored
including this one and the remote command LOCAL must
be received to regain local control.
7
Clears a partial keypad entry and reverts the calibrator out-
put or parameter to its last known value.
8
Item Name Description
1 Display A 2 line, 16 character, display providing all visual user
feedback for the isolated input operations. See section 2.6
for layout details, and section 2.7 for possible error mes-
sages.
2 Function keys
Note 1: These terminal binding posts are made of a special copper alloy to reduce thermal
EMF's. They support the use of either discreet wires or standard banana plugs, and the
HI/LO pairs are spaced for standard dual banana plugs.
Note 2: Caution. Do not exceed a maximum of 100 volts to chassis ground.
9
2.5 Rear Panel
Figure 5 describes the rear panel layout.
Warning To avoid shock hazard, connect the factory supplied 3 conductor power
cord to a properly grounded power outlet. Do not use a 2 conductor adapter or extension
cord as this will break the protective ground.
Use the chassis ground terminal for a protective ground wire if there is any question about
the grounding through the 3 conductor power cord.
5 Standard IEC AC power inlet for 120/240 VAC.
6 Main power on/off switch.
7 Power line voltage selector and fuse compartment. See section 11 for instructions on
changing the line voltage selector and changing fuses.
Warning To prevent electrical shock, only remove the line voltage selector and
fuse holder when the power cord is removed.
10
2.6 Display Layouts
11
4 Output state for RTD outputs:
Stby Standby, terminals inactive
Opr Operating, terminals are active with output per the displayed value
Blank for RTD inputs
Cold junction selection for thermocouple inputs and outputs:
XCJC External cold junction compensation; the 3001 automatic cold junction
compensation is turned off, i.e. 0 mV is always 0°C
Blank Internal cold junction compensation; the 3001 automatically measures
the ambient temperature at the thermocouple terminals and compen-
sates the measurement, i.e. 0 mV is ambient temperature
5 Input or output value
6 Units
12
Item Description
1 Selected range and input mode:·
• 10V RANGE or 100V RANGE for DC voltage·
• 50mA RANGE for basic DC current·
• 24mA LPWR for loop powered 2 wire transmitter·
• 24mA LPWR HART for loop powered 2 wire transmitter with 250 ohm resistor in circuit
2 Input value
3 Units
13
3. Getting Started
After unpacking the 3001 and becoming familiar with the layout and general operation of
the unit as described in the previous section, it is ready to set up for operation.
The following steps should be followed to set up the 3001 for operation:
a) Before attaching the power cord to the rear connector, check that the line voltage
selector is set appropriately for your location. The 3001 is shipped from the factory
with the line voltage set for the country of purchase. To verify the line voltage setting,
check the indicator on the power line voltage selector and fuse compartment cover;
see figure 5 in section 2.5 for the location.
Confirm that the setting is correct according to the following guidelines:
Line Voltage (50/60Hz) Selector Position
90 to 135 VAC 120 VAC position
220 to 250 VAC 240 VAC position
If the setting is not correct, follow the instructions in section 11.3 to change it.
b) Once the voltage selection has been made, making sure that the power switch is off,
connect the AC power cord to the 3001; see figure 5 in section 2.5 for the location.
c) Turn on the 3001 using the rear panel mounted rocker switch. The 3001 should
power up within a few seconds, briefly displaying the model number and firmware
version in the primary display before reverting to the normal input/output display.
NOTE: If a proper power up display does not occur within 30 seconds, turn the power
off, wait a few seconds, and repower the unit. If the problem persists, report the problem
to Martel immediately.
Warm up time is twice the time since last warmed up, to a maximum of 30 minutes. For
good stability it is best to leave the 3001 on all the time.
b) Press the key to select DC voltage and current mode, if not already selected.
If DC current mode is displayed, press the key again to cycle to DC voltage mode.
c) Connect the unit under test to the voltage output terminals of the 3001 as shown in
figure 10.
14
Figure 10 - DC Voltage Output Connection
d) Use the numeric keypad to enter the desired output value and press the key.
Alternatively, use the or cursor key to select a digit to modify, followed by the
or cursor key to ramp the digit up or down. This method offers a simple
solution when small changes to an output value are required, or if specific decades need
to be incremented or decremented.
e) When DC voltage mode is first selected, the 3001 is placed in the standby (Stby)
mode which puts the positive (+) output jack into a high impedance state
(>100k ohm) for safety. To place the output into the active state, press the
key which toggles between the standby and operate modes.
The standby mode is also activated in the following situations:
• If a fault occurs during operation, such as an overload or short circuit condition.
• As a safety feature for all new outputs over 30 VDC. Refer to the product specification
section of this manual for maximum drive currents.
• Warning. Scrolling the output when the output value is already over 30V will not place
the 3001 in standby mode for each new value.
• Warning. Automatic setpoints over 30V will not place the 3001 in standby mode for
each new value.
f) The 3001 can be locked to a specific voltage range by entering a value in that range
and then selecting the secondary RNG LOCK function by pressing the and
keys.
b) Press the key to select DC voltage and current mode, if not already selected.
If DC voltage mode is displayed, press the key again to cycle to DC current mode.
15
c) Connect the unit under test to the current output terminals of the 3001 as shown in
figure 11.
d) Use the numeric keypad to enter the desired output value and press the key.
Alternatively, use the or cursor key to select a digit to modify, followed by the
or cursor key to ramp the digit up or down. This method offers a simple
solution when small changes to an output value are required, or if specific decades
need to be incremented or decremented.
e) When DC current mode is first selected, the 3001 is placed in the standby (Stby)
mode which puts the positive (+) output jack into a high impedance state
(>100k ohm) for safety. To place the output into the active state, press the
key which toggles between the standby and operate modes.
The standby mode is also activated in the following situations:
• No connection made to the output terminals.
• The voltage compliance for a given output current is exceeded. The 3001 has a
typical voltage compliance of 10V so that 4-20mA application loads of up to 500
ohms can be driven. At maximum current of 100mA, the maximum load is 100
ohms.
b) Press the key to select thermocouple and RTD/Ω mode, if not already
selected. If thermocouple mode is displayed, press the key again to cycle to RTD/Ω
mode.
d) Press the key to select the desired RTD type, the custom curve, the SPRT, or
the desired ohms range. The set up and use of custom RTD coefficients is described
in section 4.5. The set up and use of SPRT coefficients is described in section 4.6.
e) Connect the unit under test to the 4 wire RTD/Ω input terminals of the 3001 as shown
in figure 12.
f) Press the and keys to toggle the RTD display between °F and °C.
g) For best accuracy, it is advisable to zero the RTD resistance circuit(s) daily, or if the
3001 is being used outside of the ambient temperature range of 18 to 28 °C. The
maximum offset from unit calibration that can be zeroed out is ±0.1 ohm for the high
range and ±0.01 ohm for the low range.
To zero a RTD resistance circuit:
• Select the RTD measure function as described above, and press the key
until the low or high ohms range is selected.
• Short the RTD/Ω terminals with the test leads normally used for RTD/Ω
measurements.
• Allow at least 3 minutes for the test leads and terminals to stabilize to the same
temperature.
17
4.4 Resistance Temperature Detector (RTD) and Ohms Source
The 3001 can source all common RTD types and 5 custom RTD curves in °F or °C, plus
basic resistance from 5 to 4000 ohms.
The following common RTD types are supported:
Pt 385 100Ω, 200Ω, 500Ω, 1000Ω
Pt 3926 100Ω
Pt 3916 (JIS) 100Ω
Ni120 120Ω
Cu 427 (Minco) 10Ω
YSI 400
a) Disconnect any test leads from external devices.
b) Press the key to select thermocouple and RTD/Ω mode, if not already
selected. If thermocouple mode is displayed, press the key again to cycle to RTD/Ω
mode.
c) If input mode is displayed, select output mode by pressing the and keys.
d) Press the key to select the desired RTD curve or ohms range. The set up
and use of custom RTD coefficients is described in section 4.5.
e) Connect the unit under test to the RTD/Ω output terminals of the 3001 as shown in
figure 13.
f) Press the and keys to toggle the RTD display between °F and °C.
g) Use the numeric keypad to enter the desired output value and press the key.
Alternatively, use the or cursor key to select a digit to modify, followed by the
or cursor key to ramp the digit up or down. This method offers a simple
solution when small changes to an output value are required, or if specific decades
need to be incremented or decremented.
18
h) When RTD/Ω mode is first selected, the 3001 is placed in the standby (Stby) mode
which puts the positive (+) output jack into a high impedance state (>100k ohm) for
safety. To place the output into the active state, press the key which toggles
between the standby and operate modes.
c) Press the key to display the curve selection prompt "RTD CUSTOM (1-
5)".
d) Press the numeric key corresponding to the custom RTD curve to be entered.
RTD curve, and press the key. To enter a coefficient that includes an
exponent, enter the mantissa, press the and keys to select the EXP
k) To abort the curve entry without saving any changes, press the key.
To use a custom RTD curve:
a) Select RTD measure or source mode as described in the preceding sections.
c) Press the key to display the curve selection prompt "RTD CUSTOM (1-
5)".
19
d) Press the numeric key corresponding to the custom RTD curve to be used.
f) To use a different custom RTD curve, press the key twice to display the
USR_DEF selection prompt.
The USR_DEF function of the 3001 uses the Calendar-Van Dusen equation for sourcing
and measuring custom RTD's. The C coefficient is only used for the subrange -260 to 0
degrees Celsius. Only the A and B coefficients are needed for the subrange 0 to 630
degrees. The R0 value is the resistance of the probe at 0 degrees Celsius.
All 5 of the custom RTD curves are set to PT385 at the factory, as shown in Table 2.
Table 3 shows the coefficients for RTD types PT391 and PT392. The C coefficient is only
used for temperatures below 0 degrees Celsius.
20
b) Press the key until the SPRT type is selected.
custom SPRT, and press the key. To enter a coefficient that includes an
exponent, enter the mantissa, press the and keys to select the EXP
j) To abort the SPRT entry without saving any changes, press the key.
e) To use a different custom SPRT, press the key twice to display the SPRT
selection prompt.
b) Press the key to select thermocouple and RTD/O mode, if not already
21
selected. If RTD/O mode is displayed, press the key again to cycle to thermocouple
mode.
d) Press the key to select the desired thermocouple type or the millivolt range.
e) Connect the unit under test to the thermocouple terminals of the 3001 using a
standard T/C miniplug as shown in figure 14. One pin is wider than the other; do not
attempt to force the plug in the wrong polarization. The T/C wire used for the
connection must match the thermocouple type selected for proper cold junction
compensation. If the 3001 and the T/C miniplug are at different temperatures, wait at
least 3 minutes for the miniplug and terminals to stabilize to the same temperature.
f) Press the and keys to toggle the thermocouple display between °F and
°C.
g) Press the and keys to toggle the cold junction compensation between
the internal temperature sensor and an external reference.
h) For best accuracy, it is advisable to zero the T/C millivolt circuit daily, or if the 3001 is
being used outside of the ambient temperature range of 18 to 28 °C. The maximum
offset from unit calibration that can be zeroed out is ±1 mV.
To zero the T/C millivolt circuit:
• Select the thermocouple measure function as described above, and press the
22
4.8 Thermocouple (TC) Source
The 3001 can source all common thermocouple types in °F or °C, plus basic millivolts
from -10.0 to 75.0 mV.
The following common thermocouple types are supported:
B, C, E, J, K, L, N, R, S, T, U, XK, BP
a) Disconnect any test leads from external devices.
b) Press the key to select thermocouple and RTD/Ω mode, if not already
selected. If RTD/Ω mode is displayed, press the key again to cycle to thermocouple
mode.
d) Press the key to select the desired thermocouple type or the millivolt range.
e) Connect the unit under test to the thermocouple terminals of the 3001 using a
standard T/C miniplug as shown in figure 15. One pin is wider than the other; do not
attempt to force the plug in the wrong polarization. The T/C wire used for the
connection must match the thermocouple type selected for proper cold junction
compensation. If the 3001 and the T/C miniplug are at different temperatures, wait at
least 3 minutes for the miniplug and terminals to stabilize to the same temperature.
f) Press the and keys to toggle the thermocouple display between °F and
°C.
g) Press the and keys to toggle the cold junction compensation between
the internal temperature sensor and an external reference.
h) Use the numeric keypad to enter the desired output value and press the key.
Alternatively, use the or cursor key to select a digit to modify, followed by the
23
or cursor key to ramp the digit up or down. This method offers a simple
solution when small changes to an output value are required, or if specific decades
need to be incremented or decremented.
i) For best accuracy, it is advisable to zero the T/C millivolt circuit daily, or if the 3001 is
being used outside of the ambient temperature range of 18 to 28 °C. This procedure
is described in section 4.7 on thermocouple measurements.
b) Press the key. The 3001 automatically senses which pressure module is
attached and sets its range accordingly.
c) Press the key to select the desired pressure units for display.
d) Before attaching the module to the pressure source, zero the module as described in
the instruction sheet that came with the module. Procedures vary, but all end with
b) Press the key to select isolated DC voltage and current input mode, if not
already selected. If the desired DC voltage mode is not displayed, press the key
again to cycle to the desired DC voltage mode.
c) Connect the unit under test to the isolated voltage/current input terminals of the 3001
as shown in figure 17.
b) Press the key to select isolated DC voltage and current input mode, if not
already selected. If the DC current mode is not displayed, press the key again to
cycle to it.
c) Connect the unit under test to the isolated voltage/current input terminals of the 3001
as shown in figure 18.
25
Figure 18 - Isolated DC Current Input Connection
d) If the UUT is a 2 wire loop powered transmitter that is disconnected from the wiring,
press the key to activate the 3001 internal 24V supply in series with the
current measuring circuit. The top line changes to 24mA LPWR to indicate that the
supply is activated. Press the key again to deactivate the supply, and the top line
reverts to 50mA RANGE.
e) If a 250 ohm resistor is required during a HART calibration procedure, press the
key to switch in the 3001 internal 250 ohm resistor. The word HART is
appended to the top line of the display to indicate that the resistor is switched in.
Press the key again to switch out the resistor, and the display reverts to its previous
state. This resistor lowers the maximum load driving capability from 1000 ohms at 20
mA to 750 ohms at 20 mA.
c) If necessary, press the key again to cycle through the pressure units until the
desired one is displayed.
d) Before attaching the module to the pressure source, zero the module as described in
the instruction sheet that came with the module. Procedures vary, but all end with
6. Output Setpoints
Nine preset output setpoints may be stored and recalled for each of the following output
modes:
• Voltage
• Current
• each thermocouple type, including millivolts
• each RTD type, including each of the five custom curves.
They may be recalled on an individual basis, or as an automatic up and down cycle with
a configurable dwell time between each setpoint. The automatic cycle feature always
starts at setpoint number 1, stepping up to a user specified ending setpoint number, then
back down in reverse order, and then repeats.
To set a setpoint:
a) Select the output mode.
b) Enter the output value for the setpoint.
27
c) At the setpoint number selection prompt "RECALL SPT#", press the numeric
key, 1 to 9, corresponding to the setpoint to be recalled.
To start an automatic setpoint cycle:
a) Select the output mode.
e) The key can be used at any time during the cycle without stopping it.
f) Press any other key to terminate the cycle.
7. Application Notes
28
1. Disconnect any test leads from external devices.
2. Select pressure input on the primary display as described in section 4.9.
3. Select current input on the isolated display as described in section 5.2. Select the
isolated loop power option. If a HART communicator is to be used for set up of the
transmitter, select the HART option.
4. Connect the transmitter as shown in figure 20.
5. Test and calibrate the transmitter per the manufacturer's instructions.
29
7.3 V/I Transmitter
30
1. Disconnect any test leads from external devices.
2. Select RTD input on the primary display as described in section 4.3. Select the RTD
type which corresponds to the RTD being tested.
3. Connect the RTD as shown in figure 23.
4. Test the RTD per the manufacturer's instructions.
31
7.6 Thermocouple Test
32
1. Disconnect any test leads from external devices.
2. Select thermocouple output on the primary display as described in section 4.8.
Select the thermocouple type which corresponds to the transmitter being tested.
3. Select current input on the isolated display as described in section 5.2. Select the
isolated loop power option. If a HART communicator is to be used for set up of the
transmitter, select the HART option.
4. Connect the transmitter as shown in figure 26.
5. Test and calibrate the transmitter per the manufacturer's instructions.
33
7.9 Precision Current Trip
34
7.10 I/I Isolator/Transmitter
35
7.12 Precision Temperature Measurement with IBP-2 Probe
36
8. LCD and Remote Interface Setup Procedures
These procedures are accessed in sequence as follows:
b) At the "LCD CONTRAST" prompt, press the or key to adjust the LCD
c) At the "LCD BACKLIGHT" prompt, press the or key to adjust the LCD
d) At the "Remote Interface" prompt, press the or key to select the RS-232
interface or the GPIB interface. When the desired interface is selected, press
the key.
e) If the GPIB interface was selected, the "GPIB Address" prompt appears. Press
the or key to ramp through the possible addresses, 0 to 30. When the
9. Remote Interface
9.1 Introduction
The 3001 can be controlled remotely from a personal computer (PC) using either a RS-
232 serial connection or an IEEE-488 parallel connection (also called a General Purpose
Interface Bus, or GPIB, connection). In either case, individual commands can be typed
into a terminal emulator program suitable for the connection type, or the calibrator can be
controlled by an automated PC program using the 3001 command set. You can write
your own automated PC program, or it may be possible to purchase a suitable third party
program and configure it for the 3001. Compatible software for IEEE-488 operation,
MET/CAL, may be purchased from Fluke.
The RS-232 connection allows one 3001 to be connected to one PC. The
communications speed is slower than IEEE-488, but no extra equipment is required other
than a low cost null modem cable.
The IEEE-488 connection allows up to 15 different calibrators, PC's, and items of test
equipment to be connected together in a bus arrangement. The communications speed
is much faster than RS-232, but it requires the purchase and installation of special PC
interface card(s) and connecting cable(s).
This section describes the set up of the two types of connections, and the general
operation of the command set. Section 10 describes the individual commands in detail.
37
9.2 Setting up the RS-232 Port for Remote Control
The 3001 is fully programmable over a standard RS-232 link with a PC. The RS-232
cable length for the port should not exceed 15 meters (50 feet), although longer cable
lengths are permitted if the load capacitance measured at a connection point (including
signal terminator) does not exceed 2500 pF.
The serial communications parameters in the 3001 are fixed at the following values:
• 9600 baud
• 8 data bits
• 1 stop bit
• no parity
• Xon/Xoff
• EOL (end-of-line) character CR (Carriage Return)
A typical RS-232 connection is shown in Figure 31. Note the use of a null modem cable
for the connection. See section 2.5 for the location of the RS-232 port on the rear panel
of the 3001.
39
d) Remote with Lockout State
When the 3001 is placed in lockout, either via a RS-232 LOCKOUT command,
or via the IEEE-488 message LLO (Local Lockout), the 3001 front panel
controls are totally locked out. The left end of the top line of the display
changes to: rem.
To return the 3001 to the local with lockout state, send the RS-232 LOCAL
command or the IEEE-488 GTL (Go To Local) message.
Table 4 summarizes the possible operating state transitions. For more information on
IEEE-488 GPIB messages, see section 9.5.
41
messages occur as state transitions of dedicated control lines. The rest of the
interface messages are sent over the data lines with the ATN signal true. All
device-dependent and common commands are sent over the data lines with
the ATN signal false.
An important thing to note about interface messages is that, unlike device-
dependent and common commands, interface messages are not sent literally
(in a direct way). They are converted to parallel signal levels on the bus data
and control lines.
IEEE-488 standards define interface messages which are handled automatically
in most cases.
e) Compound Commands
A compound command is two or more commands placed on a single
command line separated from each other with semicolons. For example,
consider the following two individual commands:
OUT 1 V
OPER
These could be combined into the compound command:
OUT 1 V; OPER
These commands instruct the 3001 to source 1 V DC, and then go into operate
mode.
f) Overlapped Commands
Commands that begin execution but require slightly more time than the normal
communication command/response interval to complete are called overlapped
commands. This is because they can be overlapped by receipt the next
command before they have been completed.
The detailed command descriptions in section 10 show a check mark T beside
Overlapped for overlapped commands.
Use the command *WAI to wait until the overlapped command has completed
execution before executing the next command. For example:
OUT 1 V; *WAI
You can also use the status commands *OPC and *OPC? to detect the
completion of overlapped commands.
g) Sequential Commands
Commands that execute immediately are called sequential commands.
The detailed command descriptions in section 10 show a check mark T beside
Sequential for sequential commands.
The majority of commands are sequential.
h) Commands for RS-232 Only
Several commands are used only for RS?232 serial operation because these
functions must be implemented as IEEE uniline (single control line) bus
management messages per the IEEE Standards. For example, the command
REMOTE could be sent as data over the IEEE-488 interface to place the 3001
into remote operating mode, but it is not because the IEEE Standards call for
42
the remote function to be sent to the device as the uniline message REN. This
is also true for several other commands and functions, as shown in table 6
below with their RS-232 equivalents.
For these commands, the detailed command descriptions in section 10 show a
check mark X beside RS-232, but no check mark beside IEEE-488.
Table 6 - Commands for RS-232 Only
IEEE-488 Message RS-232 Equivalent
GTL LOCAL command
GTR REMOTE command
LLO LOCKOUT command
b) General Rules
The general rules for parameter usage are as follows:
• Numeric parameters may have up 15 significant digits and their value can be
in the range +/-1.0E+/-20.
• Including too many or too few parameters causes a command error.
• Null parameters cause an error, e.g., the adjacent commas in OUT 1V,
,;OPER.
• Expressions, for example 4+2*13, are not allowed as parameters.
c) Extra Space or Tab Characters
In the command descriptions, parameters are shown separated by spaces. One
space after a command is required (unless no parameters are required). All other
spaces are optional. Spaces are inserted for clarity in the manual and may be left in
or omitted as desired. You can insert extra spaces or tabs between parameters as
desired. Extra spaces within a parameter are generally not allowed, except between
a number and its associated multiplier or unit.
d) Terminators
Table 8 summarizes the terminator characters for both the IEEE-488 and the RS-232
remote interfaces.
44
Table 8 - Terminator Characters
Terminator Function ASCII Character Control Language
Command Command
Number Program Terminator Terminator
Carriage Return (CR) 13 Chr(13) <Cntl> M \n
Line Feed (LF) 10 Chr(10) <Cntl> J \r
Backspace (BS) 8 Chr(8) <Cntl> H \b
Form Feed (FF) 12 Chr(12) <Cntl> L \f
Examples:
RS-232 Mode, terminal: OUT 1 V <Enter>
RS-232 Mode, program: Comm1.Output = "OUT 1 V" + Chr(10)
IEEE-488 Mode: OUT 1 V
IEEE-488 interface:
The 3001 sends the ASCII character Carriage Return with the EOI control line
held high as the terminator for response messages. The 3001 recognizes the
following as terminators when encountered in incoming data:
• ASCII CR character
• Any ASCII character sent with the EOI control line asserted
RS-232 interface:
The 3001 sends a Carriage Return (CR) character as the terminator for response
messages. The 3001 recognizes the following as terminators when encountered in
incoming data:
• ASCII CR character
• ASCII LF character
e) Incoming Character Processing
The 3001 processes all incoming data as follows:
• The most significant data bit (DIO8) is ignored
• All data is taken as 7-bit ASCII
• Lower case or upper case characters are accepted, with lower case converted to
upper case before processing
f) Response Message Syntax
In the detailed command descriptions in section 10, the responses from the 3001 are
described wherever appropriate. In order to know what type of data to read in, refer
to the beginning of the response description for the command.
Each status register and queue has a summary bit in the Serial Poll Status Byte. Enable
registers are used to mask various bits in the status registers and to generate summary
bits in the Serial Poll Status Byte. For IEEE-488 interface operation, the Service Request
Enable Register is used to assert the SRQ control line on detection of any status
condition or conditions that the programmer chooses. For RS-232 interface operation,
the SRQSTR string is sent over the serial interface when the SRQ line is set.
1) Serial Poll Status Byte (STB)
46
The Calibrator sends the serial poll status byte (STB) when it responds to a serial poll.
This byte is cleared (set to 0) when the power is turned on. The STB byte structure is
shown in Figure 34. Refer to the *STB? command in section 10 for more information on
reading this register.
7 6 5 4 3 2 1 0
0 RQS ESB MAV EAV 0 0 0
MSS
RQS Requesting service. The RQS bit is set to 1 whenever bits ESB, MAV, EAV, or ISCB
change from 0 to 1 and are enabled (1) in the SRE. When RQS is 1, the 3001 asserts
the SRQ control line on the IEEE-488 interface. You can do a serial poll to read this
bit to see if the 3001 is the source of an SRQ.
MSS Master summary status. Set to 1 whenever bits ESB, MAV, EAV, or ISCB are 1 and
enabled (1) in the SRE. This bit can be read using the *STB? Remote command in
place of doing a serial poll.
ESB Set to 1 when one or more ESR bits are 1.
MAV Message available. The MAV bit is set to 1 whenever data is available in the 3001's
IEEE-488 interface output buffer.
EAV Error available. An error has occurred and an error is available to be read from the
error queue by using the FAULT? query.
Figure 34 - Serial Poll Status Byte (STB) and Service Request Enable (SRE) Registers
2) Service Request (SRQ) Line
IEEE-488 Service Request (SRQ) is an IEEE-488.1 bus control line that the 3001
asserts to notify the controller that it requires some type of service. Many instruments
can be on the bus, but they all share a single SRQ line. To determine which
instrument set SRQ, the Controller normally does a serial poll of each instrument. The
calibrator asserts SRQ whenever the RQS bit in its Serial Poll Status Byte is 1. This bit
informs the controller that the 3001 was the source of the SRQ.
The 3001 clears SRQ and RQS whenever the controller/host performs a serial poll,
sends *CLS, or whenever the MSS bit is cleared. The MSS bit is cleared only when
ESB and MAV are 0, or they are disabled by their associated enable bits in the SRE
register being set to 0.
3) Service Request Enable Register (SRE)
The Service Request Enable Register (SRE) enables or masks the bits of the Serial
Poll Status Byte. The SRE is cleared at power up. Refer to Figure 34 for the bit
functions.
4) Programming the STB and SRE
By resetting (to 0) the bits in the SRE, you can mask (disable) associated bits in the
serial poll status byte. Bits set to 1 enable the associated bit in the serial poll status
byte.
5) Event Status Register (ESR)
The Event Status Register is a two-byte register in which the higher eight bits are
always 0, and the lower eight bits represent various conditions of the 3001. The ESR
is cleared (set to 0) when the power is turned on, and every time it is read.
47
Many of the remote commands require parameters. Improper use of parameters
causes command errors to occur. When a command error occurs, bit CME (5) in the
Event Status Register (ESR) goes to 1 (if enabled in ESE register), and the error is
logged in the error queue.
6) Event Status Enable (ESE) Register
A mask register called the Event Status Enable register (ESE) allows the controller to
enable or mask (disable) each bit in the ESR. When a bit in the ESE is 1, the
corresponding bit in the ESR is enabled. When any enabled bit in the ESR is 1, the
ESB bit in the Serial Poll Status Byte also goes to 1. The ESR bit stays 1 until the
controller reads the ESR, does a device clear, a selected device clear, or sends the
reset or *CLS command to the 3001. The ESE is cleared (set to 0) when the power
is turned on.
7) Bit Assignments for the ESR and ESE
The bits in the Event Status Register (ESR) and Event Status Enable register (ESE)
are assigned as shown in Figure 35.
15 14 13 12 11 10 9 8
0 0 0 0 0 0 0 0
7 6 5 4 3 2 1 0
PON 0 CME EXE DDE QYE 0 OPC
PON Power on. This bit is set to 1 if line power has been turned off and on since the last
time the ESR was read.
CME Command error. The IEEE-488 interface of the 3001 encountered an incorrectly
formed command and placed an error code in the error queue. The command
FAULT? can be used to fetch error codes from the error queue, which is described
in more detail below.
EXE Execution error. An error occurred when the 3001 tried to execute the last com-
mand and an error code was placed into the error queue. This could be caused,
for example, by a parameter being out of range. The command FAULT? can be
used to fetch error codes from the error queue, which is described in more detail
below.
DDE Device-dependent error. An error related to a device-dependent command has
occurred.
QYE Query error. The 3001 was addressed to talk when no response data was available
or appropriate, or when the controller failed to retrieve data from the output queue.
OPC Operation complete. All commands previous to reception of a *OPC command
have been executed, and the interface is ready to accept another message.
Figure 35 - Event Status Register (ESR) and Event Status Enable (ESE) Registers
8) Programming the ESR and ESE
To read the contents of the ESR, send the remote command, *ESR?. The ESR is
cleared (set to 0) every time it is read. To read the contents of the ESE, send the
remote command, *ESE?. The ESE is not cleared when it is read. When you read
either register, the 3001 responds by sending a decimal number that when converted
48
to binary represents bits 0 through 15.
9) Output Queue
The output queue is loaded whenever a query is processed, and holds up to 250
characters. The controller reads it with a statement such as a BASIC INPUT
statement, removing what it reads from the queue. If the queue is empty, the 3001
does not respond to the INPUT statement from the controller. The Message Available
(MAV) bit in the Serial Poll Status Byte is 1 if there is something in the output queue,
and 0 if the output queue is empty.
10) Error Queue
When a command error, execution error, or device-dependent error occurs, its error
code is placed in the error queue where it can be read by the FAULT? command.
Reading the first error with the FAULT? command removes that error from the queue.
A response of 0 means the error queue is empty. The error queue is cleared when
you turn off the power, and when you use the *CLS (Clear Status) common
command. The error queue contains up to 15 entries. If more than 15 errors occur,
only the first 15 errors are kept in the queue. A 16th entry in the queue is always an
"error queue overflow" error, and all later errors are discarded until the queue is at
least partially read. The first errors are kept, because if many errors occur before the
user can acknowledge and read them, the earliest errors are the most likely to point
to the problem. The later errors are usually repetitions or consequences of the
original problem.
11) Input Buffer Operation
As the 3001 receives each data byte from the controller, it places the byte in a portion
of memory called the input buffer. The input buffer holds up to 250 data bytes and
operates in a first in, first out fashion.
IEEE-488 interface:
The 3001 treats the IEEE-488 EOI control line as a separate data byte and inserts it
into the input buffer if it is encountered as part of a message terminator. Input buffer
operation is transparent to the program running on the controller. If the controller
sends commands faster than the 3001 can process them, the input buffer fills to
capacity. When the input buffer is full, the 3001 holds off the IEEE-488 bus with the
NRFD (Not Ready For Data) handshake line. When the 3001 has processed a data
byte from the full input buffer, it then completes the handshake, allowing the
controller to send another data byte. The calibrator clears the input buffer on power-
up and on receiving the DCL (Device Clear) or SDC (Selected Device Clear) message
from the controller.
RS-232 interface:
The 3001 uses the RS-232-C Xon/Xoff protocol to control buffer overflow. The 3001
sends a Xoff (Ctrl S) character when the input buffer becomes 80% full, and sends a
Xon (Ctrl Q) character when it has processed enough of the input buffer so that it is
less than 40% full.
49
10. Remote Commands
10.1 Introduction
Remote commands duplicate actions that can be initiated from the front panel in local
operating mode. Following the summary table is a complete alphabetical listing of all
commands complete with protocol details. Separate headings in the alphabetical listing
provide the parameters and responses, plus an example for each command. For
information on using the commands, see section 9.
Common Commands
Command Description
*CLS Clear status. Clears the ESR, the error queue, and the RQS bit in the status
byte. This command terminates any pending operation complete com-
mands, *OPC or *OPC?.
*ESE Loads a byte into the Event Status Enable register.
*ESE? Returns the contents of the Event Status Enable register.
*ESR? Returns the contents of the Event Status register and clears the register.
*IDN? Identification query. Returns the manufacturer, model number, serial number,
and firmware revision level of the Calibrator.
*OPC Sets bit 0 (OPC for Operation Complete) in the Event Status Register to 1
when all pending device operations are complete.
*OPC? Returns a 1 after all pending operations are complete. This command caus-
es program execution to pause until all operations are complete. See also
the *WAI command.
*OPT? Returns a list of the installed hardware and software options.
*RST Resets the state of the instrument to the power-up state. This command
holds off execution of subsequent commands until it is complete.
*SRE Loads a byte into the Service Request Enable register (SRE).
*SRE? Returns the byte from the Service Request Enable register.
*STB? Returns the status byte.
*TST? Runs a series of self-tests and returns a "0" for pass or a "1" for fail. If any
faults are detected, they are logged into the fault queue where they can be
read by the FAULT? query.
*WAI Prevents further remote commands from being executed until all previous
remote commands have been completed.
50
External Connection Commands
Command Description
FUNC? Returns the present output, measurement, or calibration function selected on
the isolated and primary displays, in that order.
HART? Returns the HART resistor setting for the isolated milliamp range, ON or OFF.
HART_OFF Turns off the HART resistor on the isolated milliamp range.
HART_ON Turns on the HART resistor on the isolated milliamp range.
ISO_PRES_UNIT Sets the isolated display pressure units.
ISO_PRES_UNIT? Returns the isolated display pressure units.
LOOP_POWER? Returns the 24V loop power setting for the isolated milliamp range, ON or
OFF.
LOOP_POWER _OFF Turns off the 24V loop power on the isolated milliamp range.
LOOP_POWER _ON Turns on the 24V loop power on the isolated milliamp range.
PRES_UNIT Sets the primary display pressure units.
PRES_UNIT? Returns the primary display pressure units.
RTD_TYPE Sets the Resistance Temperature Detector (RTD) type.
RTD_TYPE? Returns the Resistance Temperature Detector (RTD) type.
TC_REF Selects the internal temperature sensor or an external reference value for
cold junction compensation of thermocouple (TC) source and measurement.
TC_REF? Returns the source of the temperature being used for cold junction compen-
sation of thermocouple (TC) source and measurement.
TC_TYPE Sets the thermocouple (TC) type.
TC_TYPE? Returns the thermocouple (TC) type.
TSENS_TYPE Sets temperature mode, RTD or TC.
TSENS_TYPE? Returns the temperature mode.
Output Commands
Command Description
OPER Activates the 3001 output if it is in standby mode.
OPER? Returns the operate/standby mode setting.
OUT Sets the output of the 3001.
OUT? Returns the present output value of the 3001.
RANGE? Returns the present output range, for voltage and current only.
RANGELCK Locks the present output range, for voltage only.
RANGELCK? Returns the RANGELOCK state, for voltage only.
STBY Deactivates the 3001 output if it is in operate mode.
51
Measurement Commands
Command Description
ISO_MEAS Sets the isolated input measurement type.
PRES? Queries the attached pressure module for its manufacturer and serial num-
ber.
PRES_MEAS Changes the operating mode of the primary display to pressure measure-
ment.
RTD_MEAS Changes the operating mode to RTD measurement.
TC_MEAS Changes the operating mode to thermocouple measurement.
VAL? Returns the last values for the isolated and primary measurements, in that
order.
ZERO_MEAS Zeros the pressure module, TC mV, or RTD ohms.
ZERO_MEAS? Returns the zero offset for the pressure module, TC mV, or RTD ohms.
Status Commands
Command Description
FAULT? Returns the most recent error code in the 3001 error queue, and then
removes that error code from the queue.
53
10.4 Remote Command Listing
The following is an alphabetical list of all 3001 remote commands and queries, including
the common commands and the device-dependent commands. Each command title
includes a checkbox that indicates the remote interface applicability, IEEE-488 and/or RS-
232, and the command group, Sequential or Overlapped; see section 9.6.1 for a
description of these terms.
54
*ESR? X IEEE-488 X RS-232 X Sequential Overlapped
Event Status Register query. This command returns the contents of the Event Status
Register (ESR) and clears the register. See the Event Status Register (ESR) description in
section 9.7.
Parameter: <None>
Response: <value>
where <value> is the decimal equivalent of the ESR byte, 0 to 255
Example: *ESR?
61
This example returns decimal 61 (binary 00111101) which indicates
that bits 5 (CME), 4 (EXE), 3 (DDE), 2 (QYE) and 0 (OPC) are enabled.
55
FUNC? X IEEE-488 X RS-232 X Sequential Overlapped
This command returns the present output, measurement, or calibration function for the
primary and isolated displays.
Parameter: <None>
Response: <isolated>,<primary>
where <isolated> is one of the following:
DC10V measure DC voltage, 10V range
DC100V measure DC voltage, 100V range
DCI measure DC current
PRESSURE measure pressure
and where <primary> is one of the following:
DCV source DC voltage
DCI source DC current
RTD_OUT source RTD/Ohms
RTD_IN measure RTD/Ohms
TC_OUT source thermocouple
TC_IN measure thermocouple
PRESSURE measure pressure
Example: FUNC?
DC10V,PRESSURE
This example indicates that the isolated display is selected to the 10V
range and the primary display is selected to pressure.
56
HART_OFF X IEEE-488 X RS-232 X Sequential Overlapped
This command disables the isolated DC current input HART resistor.
Parameter: <None>
Response: <None>
Example: HART_OFF
This example disables the isolated DC current input HART resistor.
57
ISO_MEAS X IEEE-488 X RS-232 Sequential X Overlapped
This command sets the isolated measurement type.
Parameter: <value>
where <value> is one of the following:
DC10V measure DC voltage, 10V range
DC100V measure DC voltage, 100V range
DCI measure DC current
PRESSURE measure pressure
Response: <None>
Example: ISO_MEAS DCI
This example sets the isolated measurement to DC current.
58
ISO_PRES_UNIT? X IEEE-488 X RS-232 X Sequential Overlapped
This command returns the isolated pressure unit.
Parameter: <None>
Response: <value>
where <value> is one of the following:
PSI pounds per square inch
INH2O4C inches of water at 4 °C
INH2O20C inches of water at 20 °C
INH2O60F inches of water at 60 °F
CMH2O4C centimeters of water at 4 °C
CMH2O20C centimeters of water at 20 °C
MMH2O4C millimeters of water at 4 °C
MMH2O20C millimeters of water at 20 °C
BAR bars
MBAR millibars
KPA kilopascals
MPA megapascals
INHG inches of mercury at 0 °C
MMHG millimeters of mercury at 0 °C
KG/CM2 kilograms per square centimeter
Example: ISO_PRES_UNIT?
BAR
This example indicates that the isolated pressure unit is bars.
59
LOCKOUT IEEE-488 X RS-232 X Sequential Overlapped
This command puts the 3001 into the lockout state when in remote control (see the
REMOTE command). In this state, no local operation is allowed at the front panel,
including the LOCAL key. To clear the lockout condition, use the LOCAL command. This
command duplicates the IEEE-488 LLO (Local Lockout) message.
Parameter: <None>
Response: <None>
Example: LOCKOUT
This example puts the instrument into the lockout state. No front panel
controls can be used.
60
*OPC X IEEE-488 X RS-232 X Sequential Overlapped
Operations Complete command. This command sets bit 0 (OPC) of the Event Status
Register to 1 when all pending device operations are complete. Also see the *ESR?
command.
Parameter: <None>
Response: <None>
Example: *OPC
This example sets bit 0 of the Event Status Register to 1 when all
pending device operations are done.
61
OPER? X IEEE-488 X RS-232 X Sequential Overlapped
This command returns the present operate/standby mode setting.
Parameter: <None>
Response: <value>
where <value> is "1" for operate mode and "0" for standby mode
Example: OPER?
1
This example indicates that the 3001 is in operate mode.
62
Examples: Explanation
OUT 15.2 V Change to DC volts, output 15.2 V
OUT 1.2 mA Change to DC current, output 1.2 mA, note units prefix
multiplier m
OUT 5 Ohm Change to Resistance, output 5 O
OUT 100 CEL Change to temperature in °C, output 100 °C
OUT 3 No change to output mode, output a value of 3 in the
present units
Examples: Explanation
OUT?
1.88300E-02,A Present output is 18.83 mA
OUT?
1.23000E+00,V Present output is 1.23 V
OUT?
4.00000E+03,OHM Present output is 4.0 KO
OUT?
1.0430E+02,CEL Present output is 104.3 °C
63
PRES? X IEEE-488 X RS-232 X Sequential Overlapped
This command queries the attached pressure module for its manufacturer, serial number,
and firmware version.
Parameter: <None>
Response: <text string>
where <text string> contains the following three fields separated by
commas:
1. Manufacturer
2. Serial number
3. Firmware revision level (always 0)
Example: PRES?
MARTEL,610070,0
This example indicates that the manufacturer is Martel, the serial
number is 610070, and the firmware version is 0.
64
PRES_UNIT X IEEE-488 X RS-232 Sequential X Overlapped
This command sets the primary display pressure units.
Parameter: <value>
where <value> is one of the following:
PSI pounds per square inch
INH2O4C inches of water at 4 °C
INH2O20C inches of water at 20 °C
INH2O60F inches of water at 60 °F
CMH2O4C centimeters of water at 4 °C
CMH2O20C centimeters of water at 20 °C
MMH2O4C millimeters of water at 4 °C
MMH2O20C millimeters of water at 20 °C
BAR bars
MBAR millibars
KPA kilopascals
MPA megapascals
INHG inches of mercury at 0 °C
MMHG millimeters of mercury at 0 °C
KG/CM2 kilograms per square centimeter
Response: <None>
Example: PRES_UNIT BAR
This example sets the primary display pressure units to bars.
66
RANGELCK? X IEEE-488 X RS-232 X Sequential Overlapped
This command returns the DC voltage range lock status.
Parameter: <None>
Response: <value>
where <value> is one of the following:
ON DC voltage range lock is on
OFF DC voltage range lock is off
Example: RANGELCK?
OFF
This example indicates that the range lock is off.
69
*SRE X IEEE-488 X RS-232 X Sequential Overlapped
Service Request Enable command. This command loads a byte into the Service Request
Enable (SRE) register. See the Service Request Enable Register (SRE) description in
section 9.7. Since bit 6 is not used (decimal value 64), the maximum entry is 255 - 64 =
191.
Parameter: <value>
where <value> is the decimal equivalent of the SRE byte, 0 to 191
Response: <None>
Example: *SRE 48
This example enables bits 4 (MAV) and 5 (ESB).
70
STBY X IEEE-488 X RS-232 Sequential X Overlapped
This command places the 3001 in standby mode, deactivating the output at front panel
terminals. This command acts the same as pressing the front panel key when in
operate mode.
Parameter: <None>
Response: <None>
Example: STBY
This example disconnects the selected output from the 3001 front
panel terminals. It also indicates Stby on the display.
71
TC_REF? X IEEE-488 X RS-232 X Sequential Overlapped
This command returns the source of the temperature being used for cold junction
compensation of thermocouple source and measurement.
Parameter: <None>
Response: <value>
where <value> is one of the following:
INT internal temperature sensor in use
EXT external reference value in use
Example: TC_REF?
INT
This example indicates that the internal sensor is in use.
72
TC_TYPE? X IEEE-488 X RS-232 X Sequential Overlapped
This command returns the Thermocouple (TC) sensor type being used for TC
temperature source and measurement.
Parameter: <None>
Response: <value>
where <value> is one of the following:
B B-type thermocouple
C C-type thermocouple
E E-type thermocouple
J J-type thermocouple
K K-type thermocouple
L L-type thermocouple
N N-type thermocouple
R R-type thermocouple
S S-type thermocouple
T T-type thermocouple
U U-type thermocouple
XK XK-type thermocouple
BP BP-type thermocouple
Z 1 mV/°C
Example: TC_TYPE?
K
This example indicates that the thermocouple sensor type is K-type.
73
TSENS_TYPE? X IEEE-488 X RS-232 X Sequential Overlapped
This command returns the present temperature mode, thermocouple (TC) or Resistance
Temperature Detector (RTD).
Parameter: <None>
Response: <value>
where <value> is one of the following:
TC Thermocouple
RTD Resistance Temperature Detector
Example: TSENS_TYPE?
TC
This example indicates that the present temperature mode is
thermocouple.
74
expressed in scientific notation.
and where <primary-units> is one of the following:
FAR °F
CEL °C
OHM Ohms
V DC volts (thermocouple millivolts)
OVER measurement is over or under range
NONE primary display is presently set to a source mode
or one of the pressure units listed with the PRES_UNIT?
command
Example: VAL?
2.137000E+00,V,5.763300E+01,CEL
This example indicates that the isolated measurement is 2.137 volts
and that the primary measurement is 57.633 °C.
76
11. Maintenance
Warning
To avoid personal injury and/or damage to the Calibrator, use only the specified
replacement parts and do not allow water into the case.
Caution
To avoid damaging the case, do not use solvents or abrasive cleaners.
Clean the calibrator and pressure modules with a soft cloth dampened with water, or mild
soap and water.
Warning
To avoid electrical shock hazard disconnect line power before opening the case or
fuse compartment.
The line power fuses and line voltage selector are located in the compartment above the
power switch on the right rear of the Calibrator. See the rear panel layout in section 2.5.
Table 10 shows the correct replacement fuse for each line voltage setting.
Table 10 - Replacement Fuses
Fuse Description Line Voltage Setting
0.25 A/250 V fast 120 V (90 V to 132 V)
0.125 A/250 V fast 240 V (198 V to 264 V)
77
11.3 Changing the Line Voltage
The calibrator arrives from the factory configured for the line voltage appropriate for the
country of purchase, or as specified when it is ordered. To verify the line voltage setting,
check the line voltage indicator on the power line fuse compartment cover.
Confirm that the line voltage selection is set for 120 V for line voltages between 90 V and
132 V, or that the selector is set to 240 V for line voltages between 198 V and 264 V.
Warning
To avoid electrical shock hazard disconnect line power before opening the case or
fuse compartment.
To change the line voltage:
1. Disconnect line power.
2. Using the blade of a suitable flat screwdriver, pry up the tab at the base of the
line fuse compartment by inserting the blade in the center slot under the tab.
The compartment cover will pop part way out.
3. Remove the compartment cover.
4. Remove the line voltage selector assembly by gripping the line voltage
indicator tab with pliers and pulling it straight out of the compartment.
5. Rotate the line voltage selector assembly to the desired voltage and reinsert.
6. Verify you are using the appropriate fuse for the selected line voltage (see table
10 above), and reinstall the fuse compartment by pushing it in until the tab
locks in place.
78
12. Specifications
79
12.2 DC Voltage Specifications, Output
Absolute Uncertainty,
tcal ±5 °C
± (ppm of output +µV) Stability
24 hours, ±1 °C Maximum
Ranges1 90 days 1 year ± (ppm of output +µV) Resolution Burden2
0 to 100.000 mV 25 3 30 3 5 ppm +2 1 µV 10 mA
0 to 1.00000 V 25 10 30 10 4 ppm + 10 10 µV 10 mA
0 to 10.0000 V 25 100 30 100 4 ppm + 100 100 µV 10 mA
0 to 100.000 V 25 1 mV 30 1 mV 5 ppm + 1 mV 1 mV 1 mA
TC Output and Input
-10 to 75.000 mV 25 3 mV 30 3 µV 5 ppm + 2 µV 1 µV 10 Ω
1. All outputs are positive only.
2. Remote sensing is not provided. Output resistance is < 1Ω
Noise
Bandwidth 0.1 to 10 Hz p-p Bandwidth
Ranges ± (ppm of output +µV) 10 Hz to 10 kHz rms µV
0 to 100.000 mV 1 µV 6 µV
0 to 1.00000 V 10 µV 60 µV
0 to 10.0000 V 100 µV 600 µV
0 to 100.000 V 10 ppm+1 mV 20 mV
80
Noise
Bandwidth Bandwidth
Ranges 0.1 to 10 Hz p-p 10 Hz to 10 kHz rms µV
0 to 100.000 mA 2000 nA 20 µA
81
12.8 Thermocouple Specification, Output and Input
Absolute Uncertainty,
tcal ±5 °C, ±(°C)1
Range (° C) Output/Input
TC Type Minimum Maximum 90 days 1 year
B 600 °C 800 °C 0.42 °C 0.46 °C
800 °C 1000 °C 0.39 °C 0.39 °C
1000 °C 1550 °C 0.40 °C 0.40 °C
1550 °C 1820 °C 0.44 °C 0.45 °C
C 0 °C 150 °C 0.25 °C 0.30 °C
150 °C 650 °C 0.21 °C 0.26 °C
650 °C 1000 °C 0.23 °C 0.31 °C
1000 °C 1800 °C 0.38 °C 0.50 °C
1800 °C 2316 °C 0.63 °C 0.84 °C
E -250 °C -100 °C 0.38 °C 0.50 °C
-100 °C -25 °C 0.16 °C 0.18 °C
-25 °C 350 °C 0.14 °C 0.15 °C
350 °C 650 °C 0.14 °C 0.16 °C
650 °C 1000 °C 0.16 °C 0.21 °C
J -210 °C -100 °C 0.20 °C 0.27 °C
-100 °C -30 °C 0.18 °C 0.20 °C
-30 °C 150 °C 0.14 °C 0.16 °C
150 °C 760 °C 0.14 °C 0.17 °C
760 °C 1200 °C 0.18 °C 0.23 °C
K -200 °C -100 °C 0.25 °C 0.33 °C
-100 °C -25 °C 0.19 °C 0.22 °C
-25 °C 120 °C 0.14 °C 0.16 °C
120 °C 1000 °C 0.19 °C 0.26 °C
1000 °C 1372 °C 0.30 °C 0.40 °C
L -200 °C -100 °C 0.37 °C 0.37 °C
-100 °C 800 °C 0.26 °C 0.26 °C
800 °C 900 °C 0.17 °C 0.17 °C
N -200 °C -100 °C 0.33 °C 0.40 °C
-100 °C -25 °C 0.20 °C 0.24 °C
-25 °C 120 °C 0.16 °C 0.19 °C
120 °C 410 °C 0.14 °C 0.18 °C
410 °C 1300 °C 0.21 °C 0.27 °C
1. Does not include thermocouple wire error.
82
Thermocouple Specification, Output and Input (continued)
Absolute Uncertainty,
tcal ±5 °C, ±(°C)1
Range (° C) Output/Input
TC Type Minimum Maximum 90 days 1 year
R 0 °C 250 °C 0.58 °C 0.58 °C
250 °C 400 °C 0.34 °C 0.35 °C
400 °C 1000 °C 0.31 °C 0.33 °C
1000 °C 1750 °C 0.30 °C 0.40 °C
S 0 °C 250 °C 0.56 °C 0.56 °C
250 °C 1000 °C 0.36 °C 0.36 °C
1000 °C 1400 °C 0.30 °C 0.37 °C
1400 °C 1750 °C 0.35 °C 0.46 °C
T -250 °C -150 °C 0.51 °C 0.63 °C
-150 °C 0 °C 0.18 °C 0.24 °C
0 °C 120 °C 0.13 °C 0.16 °C
120 °C 400 °C 0.12 °C 0.14 °C
U -200 °C 0 °C 0.56 °C 0.56 °C
0 °C 600 °C 0.27 °C 0.27 °C
XK -200 °C -100 °C 0.22 °C 0.22 °C
-100 °C 300 °C 0.12 °C 0.13 °C
300 °C 800 °C 0.19 °C 0.20 °C
BP 0 °C 200 °C 0.42 °C 0.42 °C
200 °C 600 °C 0.32 °C 0.32 °C
600 °C 800 °C 0.39 °C 0.40 °C
800 °C 1600 °C 0.45 °C 0.46 °C
1600 °C 2000 °C 0.57 °C 0.58 °C
2000 °C 2500 °C 0.67 °C 0.80 °C
1. Does not include thermocouple wire error.
83
12.9 RTD and Thermistor Specification, Output
Absolute Uncertainty,
tcal ±5 °C, ±(°C)1
Range (° C) Output/Input
RTD Type Minimum Maximum 90 days 1 year
Pt 385, 100 Ω -200 °C -80 °C 0.03 °C 0.04 °C
-80 °C 0 °C 0.04 °C 0.05 °C
0 °C 100 °C 0.04 °C 0.05 °C
100 °C 300 °C 0.03 °C 0.04 °C
300 °C 400 °C 0.04 °C 0.04 °C
400 °C 630 °C 0.04 °C 0.05 °C
630 °C 800 °C 0.04 °C 0.05 °C
Pt 3926, 100 Ω -200 °C -80 °C 0.03 °C 0.04 °C
-80 °C 0 °C 0.03 °C 0.04 °C
0 °C 100 °C 0.03 °C 0.04 °C
100 °C 300 °C 0.03 °C 0.04 °C
300 °C 400 °C 0.03 °C 0.04 °C
400 °C 630 °C 0.04 °C 0.05 °C
Pt 3916, 100 Ω -200 °C -190 °C 0.03 °C 0.03 °C
-190 °C -80 °C 0.03 °C 0.04 °C
-80 °C 0 °C 0.03 °C 0.04 °C
0 °C 100 °C 0.03 °C 0.04 °C
100 °C 260 °C 0.03 °C 0.04 °C
260 °C 300 °C 0.03 °C 0.04 °C
300 °C 400 °C 0.03 °C 0.04 °C
400 °C 600 °C 0.04 °C 0.05 °C
600 °C 630 °C 0.04 °C 0.05 °C
Pt 385, 200 Ω -200 °C -80 °C 0.31 °C 0.38 °C
-80 °C 0 °C 0.32 °C 0.38 °C
0 °C 100 °C 0.33 °C 0.39 °C
100 °C 260 °C 0.33 °C 0.39 °C
260 °C 300 °C 0.36 °C 0.43 °C
300 °C 400 °C 0.36 °C 0.43 °C
400 °C 600 °C 0.42 °C 0.50 °C
600 °C 630 °C 0.42 °C 0.50 °C
Pt 385, 500 Ω -200 °C -80 °C 0.13 °C 0.15 °C
-80 °C 0 °C 0.13 °C 0.15 °C
0 °C 100 °C 0.13 °C 0.16 °C
100 °C 260 °C 0.14 °C 0.17 °C
260 °C 300 °C 0.14 °C 0.17 °C
300 °C 400 °C 0.15 °C 0.18 °C
400 °C 600 °C 0.16 °C 0.19 °C
600 °C 630 °C 0.16 °C 0.19 °C
1. 2-wire output
84
RTD and Thermistor Specification, Output (continued)
Absolute Uncertainty,
tcal ±5 °C, ±(°C)1
Range (° C) Output/Input
RTD Type Minimum Maximum 90 days 1 year
Pt 385, 1000 Ω -200 °C -80 °C 0.06 °C 0.07 °C
-80 °C 0 °C 0.06 °C 0.08 °C
0 °C 100 °C 0.07 °C 0.08 °C
100 °C 260 °C 0.07 °C 0.08 °C
260 °C 300 °C 0.07 °C 0.09 °C
300 °C 400 °C 0.07 °C 0.09 °C
400 °C 600 °C 0.08 °C 0.09 °C
600 °C 630 °C 0.08 °C 0.09 °C
Ni 120, 120 Ω -80 °C 0 °C 0.02 °C 0.02 °C
0 °C 100 °C 0.02 °C 0.02 °C
100 °C 260 °C 0.01 °C 0.02 °C
Cu 427, 10 Ω2 -100 °C 260 °C 0.30 °C 0.38 °C
YSI 400 15 °C 50 °C 0.005 °C 0.007 °C
1. 2-wire output
2. Based on MINCO Application Aid No. 18.
85
12.10 RTD and Thermistor Specification, Input
Absolute Uncertainty,
tcal ±5 °C, ±(°C)1
Range (° C) Output/Input
RTD Type Minimum Maximum 90 days 1 year
Pt 385, 100 Ω -200 °C -80 °C 0.011 °C 0.012 °C
-80 °C 0 °C 0.018 °C 0.020 °C
0 °C 100 °C 0.018 °C 0.020 °C
100 °C 300 °C 0.027 °C 0.030 °C
300 °C 400 °C 0.031 °C 0.035 °C
400 °C 630 °C 0.042 °C 0.047 °C
630 °C 800 °C 0.050 °C 0.057 °C
Pt 3926, 100 Ω -200 °C -80 °C 0.011 °C 0.011 °C
-80 °C 0 °C 0.014 °C 0.015 °C
0 °C 100 °C 0.018 °C 0.019 °C
100 °C 300 °C 0.026 °C 0.029 °C
300 °C 400 °C 0.031 °C 0.034 °C
400 °C 630 °C 0.041 °C 0.046 °C
Pt 3916, 100 Ω -200 °C -190 °C 0.006 °C 0.006 °C
-190 °C -80 °C 0.011 °C 0.012 °C
-80 °C 0 °C 0.014 °C 0.015 °C
0 °C 100 °C 0.018 °C 0.019 °C
100 °C 260 °C 0.025 °C 0.028 °C
260 °C 300 °C 0.026 °C 0.029 °C
300 °C 400 °C 0.031 °C 0.034 °C
400 °C 600 °C 0.040 °C 0.045 °C
600 °C 630 °C 0.042 °C 0.047 °C
Pt 385, 200 Ω -200 °C -80 °C 0.008 °C 0.009 °C
-80 °C 0 °C 0.012 °C 0.013 °C
0 °C 100 °C 0.015 °C 0.017 °C
100 °C 260 °C 0.020 °C 0.022 °C
260 °C 300 °C 0.050 °C 0.053 °C
300 °C 400 °C 0.053 °C 0.057 °C
400 °C 600 °C 0.070 °C 0.075 °C
600 °C 630 °C 0.071 °C 0.076 °C
Pt 385, 500 Ω -200 °C -80 °C 0.007 °C 0.008 °C
-80 °C 0 °C 0.019 °C 0.020 °C
0 °C 100 °C 0.023 °C 0.025 °C
100 °C 260 °C 0.030 °C 0.033 °C
260 °C 300 °C 0.032 °C 0.035 °C
300 °C 400 °C 0.037 °C 0.041 °C
400 °C 600 °C 0.047 °C 0.052 °C
600 °C 630 °C 0.048 °C 0.053 °C
1. 4-wire mode. Uncertainties listed do not include probe uncertainties.
86
RTD and Thermistor Specification, Input (continued)
Absolute Uncertainty,
tcal ±5 °C, ±(°C)1
Range (° C) Output/Input
RTD Type Minimum Maximum 90 days 1 year
Pt 385, 1000 Ω -200 °C -80 °C 0.011 °C 0.012 °C
-80 °C 0 °C 0.014 °C 0.015 °C
0 °C 100 °C 0.019 °C 0.020 °C
100 °C 260 °C 0.025 °C 0.028 °C
260 °C 300 °C 0.027 °C 0.030 °C
300 °C 400 °C 0.030 °C 0.034 °C
400 °C 600 °C 0.041 °C 0.045 °C
600 °C 630 °C 0.042 °C 0.047 °C
Ni 120, 120 Ω -80 °C 0 °C 0.009 °C 0.010 °C
0 °C 100 °C 0.010 °C 0.011 °C
100 °C 260 °C 0.011 °C 0.012 °C
Cu 427, 10 Ω2 -100 °C 260 °C 0.067 °C 0.069 °C
YSI 400 15 °C 50 °C 0.005 °C 0.007 °C
SPRT -200 °C 660 °C 0.05 °C 0.06 °C
1. 4-wire mode. Uncertainties listed do not include probe uncertainties
2. Based on MINCO Application Aid No. 18.
87
Range Accuracy and Resolution Units
Determined by the Determined by the
pressure module pressure module PSI(pounds per square inch)
in H2O 4ºC
(inches of water at 4 degrees Celsius)
in H2O 20°C
(inches of water at 20 degrees Celsius)
in H2O 60°F
(inches of water at 60 degrees Fahrenheit)
cm H2O 4ºC
(centimeters of water at 4 degrees Celsius)
cm H2O 20°C
(centimeters of water at 20 degrees Celsius)
mm H2O 4ºC
(millimeters of water at 4 degrees Celsius)
mm H2O 20°C
(millimeters of water at 20 degrees Celsius)
BAR
(bars)
mBAR
(millibars)
kPa
(kilopascals)
MPa
(megapascals)
in HG 0ºC
(inches of mercury at 0 degrees Celsius)
mm HG 0ºC
(millimeters of mercury at 0 degrees Celsius)
Kg/cm2
(kilograms per square centimeter)
13. Warranty
Martel Electronics Corporation warrants all products against material defects and
workmanship for a period of twelve (12) months after the date of shipment. Problems or
defects that arise from misuse or abuse of the instrument are not covered. If any product
is to be returned, a "Return Material Authorization" number must be obtained from our
Customer Service Department. This number must be indicated on the return package as
notice to our Receiving Department to accept the shipment. Any package not so marked
will not be accepted and will be returned to the shipper. Martel will not be responsible for
damage as a result of poor return packaging. Out of warranty repairs and recalibration
will be subject to specific charges. Under no circumstances will Martel Electronics be
liable for any device or circumstance beyond the value of the product.
88
89
www.martelcorp.com
e-mail: sales@martelcorp.com
Fax: (603) 434-1653 Tel: (603) 434-1433
800-821-0023
Martel Electronics
PO Box 770
Londonderry, NH 03053