4112 Dew Point Transmitter
4112 Dew Point Transmitter
Operations Manual
                     70-82-25-08
                          Rev. 2
                           2/99
        Copyright, Notices, and Trademarks
              Printed in U.S.A. – © Copyright 1999 by Honeywell Inc.
                                 Revision 2 – 2/99
                                     Honeywell
                         Industrial Automation and Control
                                Automation College
                              2820 West Kelton Lane
                                 Phoenix, AZ 85023
                                   (602) 313-5669
Abstract
This manual provides technical description and installation and maintenance instructions for the
Honeywell 4112 Moisture Instrument.
Included are all circuit adjustment instructions required to verify proper electrical functions of the
instrument as well as system calibration for ambient temperature and moisture.
Revision Notes
The following list provides notes concerning all revisions of this document.
          0                9/30/96          This document is the initial release of the Honeywell version of the L&N
                                            manual p/n 278159 Rev. A.
          1                 10/97           Additions were made in Chapter 3 for Ferrite Bead installation and in
                                            Chapter 12 for CE approval information.
          2                  2/99           Additions were made for the cooler option and the S200 and S201 dip
                                            switches.
References
Non-Honeywell Documents
The following list identifies select non-Honeywell documents that may be sources of reference for the
material discussed in this publication.
Contacts
The following list identifies important contacts within Honeywell.
1.         INTRODUCTION....................................................................................................1
     1.1    Overview.........................................................................................................................................1
           Transmitter Functions........................................................................................................................1
     1.2     Filters ..............................................................................................................................................1
3.         INSTALLATION .....................................................................................................9
     3.1     Storage and Shipment .....................................................................................................................9
     3.2     Unpacking .......................................................................................................................................9
     3.3     Rough Functional Check.................................................................................................................9
     3.4     Dew Point and Wet Bulb Considerations .......................................................................................9
     3.5     Mechanical Considerations...........................................................................................................10
     3.6     Maintaining CE Mark Approval Procedure..................................................................................10
4.         WIRING................................................................................................................13
     4.1     Introduction...................................................................................................................................13
7. TROUBLESHOOTING.........................................................................................31
1. Introduction
1.1 Overview
Transmitter Functions
       The Honeywell 4112 unit measures and digitally displays Ambient Temperature and a moisture
       measurement. The moisture measurement can be in units of Dew Point Temperature, Relative Humidity,
       Humidity Ratio or Wet Bulb temperature, depending on a customer’s requirements. In addition this
       transmitter unit outputs two 4 mA to 20 mA current loops, proportional to Ambient Temperature and
       moisture measurement. Figure 1-1 shows an exploded view of the 4112 assembly.
       The unit is packaged in an explosion proof housing with a window to permit viewing the digital displays.
       Connection for Transmitter operation is available by unscrewing (counterclockwise) the windowed housing
       cover and exposing a barrier strip with screw terminals on an inside cover.
       A label adjacent to the barrier strip screw terminals identifies connections for 120 VAC or low voltage DC
       operation of the unit. Wiring should be routed through conduit inlets on the housing to the screw terminals
       exposed by removal of the windowed explosion proof cap.
       A green grounding screw is mounted to the housing adjacent to the internal cap for system grounding.
       A metal swing away plate is affixed to the side of the unit below the explosion proof cap. This plate may be
       loosened and swung away to expose adjustment controls for field calibration. These controls may be
       accessed without violating the explosion proof integrity of the housing.
       The sensor may be inserted into the application via 1/2” NPT male threads at the root of the sensor probe
       just under the base of the housing.
       The probe stainless steel shaft is 3/8” diameter and may optionally be inserted into the application via a 3/8”
       compression fitting.
1.2 Filters
       Included with each dew point transmitter are two threaded filters and a fluted shield .
       One of the filters is mounted on the end of the sensing probe. The other filter and the shield are packed in a
       small plastic bag. Following is a description of these three parts.:
       condensing or wet conditions. It will help keep liquid moisture off the sensor itself, allowing it to more
       quickly respond when conditions return to a non-saturated state.
Window
Window
                                                          Digital Display
          Interface Cable Cap
                                                                Screw Terminals for
                                                                Temperature Sensor
              Signal Conditioning
        and     Transmitter Ass’y
                                                                  Co-Axial Connector
                                                                  for Moisture Sensor
Neoprene O-Ring
                                                               1 1/4-16 UN-2A
                                                               to Have 8 Threads
                                                               Minimum Engagement
         1/2-14 NPT Conduit                                    with Housing
                                                             Material: 304 or 316 S.S.
                         1/2-14 NPT
                                                              Material: 300 Series S.S.
Sensor
                                                                                 a/n 23314
                                                   3/8"
INSTALLATION
           Transmitter Housing:         NEMA 4X, IP65 cast low-copper aluminum fully anodized with aliphatic
                                        urethane finish. Weight approximately 4.4 lb. (2.0 kg).
                   Sensor Housing:      Stainless steel probe; fluted shield; porous stainless steel filters.
                        Mounting:       Can be mounted on pipe or wall
    Sensor Operating Environment:       -40oC to 185oC (-58oF to 365oF); to 150°C (300°F) for –110 option
Transmitter Operating Environment:      -40oC to 80oC (-40oF to 176oF); Display:-30oC (-22oF)
              Storage Environment:      -55oC to 85oC (-67oF to 185oF)
              Power Requirements:       115 VAC ± 10%, 60 Hz with Rlmax 500 Ohm or 11Vdc + (Rload x 0.02 A)
                                        min. to 45 Vdc max. for dual 2-wire operation
                                        Operation requires three wires, even if using only one signal
  Intrinsic Safety Entity Parameters:   V max = 30 volts; I max = 360mA; Ci = 0.1 µF; Li = mH
                  Signal Connection:    Screw terminals
                   Sensor Mounting:     1/2” NPT; ¾” NPT (-110 option)
                           Pressure:    0 to 350 psi dew point at pressure; 0 to 2000 psi (-110 option)
SERVICEABILITY
              Periodic Maintenance:     Verify calibration periodically. Wash sensor as needed.
                       Adjustments:     Vernier, non-interacting, multi-turn potentiometers for zero and span control
                                        of Dry Bulb temperature. Momentary rotary action switches for Dew Point
                                        Calibration. Adjustments are on the outside of the transmitter housing under
                                        a swing-away plate.
            Sensor Break Indication:    Dry Bulb Sensor: upscale both output signals
                                        Dew Point Sensor: downscale on dew point signal only
       Over Temperature Indication:     Dew Point output at 100oC when Dry Bulb > 192oC
                   Sensor Cleaning:     Wash with detergent solution. Rinse with de-ionized water. Camel hair
                                        (organic) brush may be used during cleaning when required.
TABLE IV – OPTIONS*
None                                                                                           000           •
Pipe/Wall Bracket Kit                                                                          100           •
High Pressure Probe: up to 2000 psi                                                            110           c
FM Intrinsically Safe (24 Volt only)                                                           209           e
CE Compliant (24 Volt DC Power)                                                                210           •
Stainless Steel Tag, up to 25 characters each of 3 lines                                       208           •
RESTRICTIONS
       Restriction Letter                   Available Only With                      Not Available With
                                     Table              Selection            Table              Selection
                c                       I                  04, 12              IV               100, 104
                                       II                   D3
                d                     IV                    110
                e                                                              IV                    210
                f                                                              IV               209, 210
*See 4113 Accessories for coolers, insertion removal device, etc.
                                                           6.0
                                                                                                                                                           -10
                                                                             FP
                                                                                       H                                                                      0°C
                                                                            °C
                                                                                                                           -75
                                                           4.0                                                                °C
                                                                                              FP
                                                                         -50
                                                                                             °C
                                                           3.0
- 25
                                                                                                                                                           DP
                                                                                                              DP
                                                                             H
                                                                         3% R
                                                                                                                                    DP
                                                           2.0                        -50
                                                                                                                                                                                   DP
                                                                                                                                                      °C
                                                                                                        0°C
                                                                                                                                 °C
                                                                                         °C
                                                                                                                                                    50
                                                                                                                                                                          °C
                                                                                                                               25
                                                                                                                                                                        75               DP
                                                                                                                                                                                     0°C
                                                           1 .5                                                                                                                    10
                                                                                        RH
                                                           1 .0                   10%
                                                           0.8
                                                                         -25°C
                                                           0.6                          0°C
                                                           0.5
                                                                  0              25        50 75 100 125                                                               150                  175
                                                                                        Dry Bulb Ambient - °C                                                                                          a/n23315
                                                                                                                                                                                           FP
                                                                                                                                                                                        F)
                                                                                                                                                                                      4°
                                                           0.9                                                                                                                C
                                                                                                                                                                                   (1
                                                                                                                                                                            0°
                                                                                                                                                                       -1                         P
                                                                                                                                                                                             )F
                                                           0.8                                                                                                                         3   °F
                                                                                                                                                          it)                       (2
                                                                                                                                                    Lim                        °C
                                                                                                                                                                            -5               DP
                                                                                                                                               mA                                        )
                                                           0.7                                                                              (4                                       2°F
                                                                                                                                                                                   (3
                                                                                                                                                                               C
                                                                                                                                                                            0°       DP
                                                                                                                                                                                 °F)
                                                                                                                                                                              (50
                                                           0.6                                                                                                            10°
                                                                                                                                                                                   DP
                                                                                                                                                                            (68°F)
                                                                                                                                                                       20°C
                                                 0.55
                                                                     0                   10                                    20                               30                            40
                                                                                  Dry Bulb Temperature, °C                                                                                             a/n 23326
                                                                                                          )   FP
                                                                 4                                    6°F                      FP                      P                           FP
                                                                                                  (-7                     F)                         )F                       F)
                                                                                             C                          8°                        0°F                       2°
                                                                                           0°                    (- 5                       (-4                      (- 2
                                                                                        -6                  °C                         °C                 0     °C
                                                                                                     -5
                                                                                                        0                       -4
                                                                                                                                   0
                                                                                                                                                       -3                          FP
                                                                 2                                                                                                       ° F)
                                                                                                                                                                      (-4          FP
                                                                                                                                                           -2   0°C          14°
                                                                                                                                                                                F)
                                                                                                                                                                                        P
                                                                                                                                                                        0°C(      2°F) D
                                                             1                                                                                                       -1
                                                                                                                                                                           0°C
                                                                                                                                                                               (3
                                                                                                                                                                                     DP
                                                                                                                                                                       20°C (68°F)
                                                           0.5
                                                                          -40         -20      0       20                                                                            40
                                                                                  Dry Bulb Temperature - °C                                                                                           a/n 23327
3. Installation
3.2 Unpacking
Step Action
         1      If there are visible signs of damage, do not remove the transmitter from its packing material. Notify
                the carrier and Honeywell immediately.
         2      If there is no visible damage, compare the contents with the packing list. Notify the carrier and
                Honeywell immediately if there is a or shortage.
         4      Carefully remove the Transmitter from its packing case and then remove any shipping ties on
                packing material. Follow the instructions on any attached tags and then remove such tags.
         1      Connect the “H2O”, “TEMP”, and “COM” screw         The display should indicate ambient temperature and
                terminals on the internal cap together. Apply      dew point temperature.
                120 VAC 60 Hz to the L1 and L2 screw
                terminals.
         2      Hold the sensor tip close to your open mouth       Observe that the temperature display of ambient and
                and breathe on it.                                 dew point temperature responds.
        The highest temperature at which water can exist as a liquid at one atmosphere of pressure is 100oC, hence
        the maximum dew point range of the 4112 is 100oC or 212oF.
        There is no theoretical practical upper limit of temperature at which water vapor can exist as a gas.
        The real practical upper limit of gas ambient temperature for in situ dew point measurement is the 4112
        sensor technology. This is limited to 185oC.
        The 4112 calibrated upper limit of ambient temperature, sometimes referred to as dry bulb temperature, is
        204.7oC.
        Figure 2-3 illustrates the temperature range for which dew point measurements may be taken by the 4112.
        When choosing a site for sensor installation within a system consideration should be given to select a point
        where the gas will be the coolest without the formation of condensation.
        The reason for the above consideration is that accuracy is dependent on the difference of temperature
        between the ambient or dry bulb and dew point temperature (see accuracy chart Figure 2-1). This
        temperature difference is referred to as temperature depression.
     ATTENTION
     This conduit connection can not exceed 30”.
      Step                                                      Action
       1       These installation notes apply to CE marked direct current loop powered units only (option code –
               210). The ferrite bead is not required for CE marked AC line powered units (option –212). The ferrite
               bead is not required for any units not bearing the CE mark.
       2       Install the 4112 per the installation instructions. Before connecting the field wires to the terminal
               block, slip the supplied ferrite bead over all wires. The ferrite bead is installed inside the threaded
               explosion-proof housing of the 4112 unit.
       3       Push the ferrite bead over the field wires as far as possible to the lower part of the enclosure.
       4       Secure a plastic cable tie around the field wires just above the ferrite bead to retain the ferrite bead (
               the exact location of the ferrite bead is not critical to its function.).
       5       Connect the field wires to the terminal block per the installation instructions.
                                           4.4
          4.6" Dia                     (111.8mm)
                                                  7.3
       (116.8mm)                              (185.4mm)
ZSD
              1/2" NPT
                                                                         Optional Pipe/Wall Mount
                                       12"                                     Hardware Kit
                                    (305mm)
                                      STD
                                                              4.4
                             4.6" Dia                       (111.8m
                                                                         7.3
                          (116.8mm)                                  (185.4mm)
ZSD
3/4" NPT
                                                          12"
                                                       (305mm)
                                                         STD
                   Figure 3-2-b 4112 Transmitter with High Pressure Probe (option –110)
                                         Mounting Configuration
4. WIRING
4.1 Introduction
          The 4112 has two transmitter outputs and an internal 120 VAC 60 Hz power supply to power these two
          transmitters.
          The two transmitters are internally connected together at their (+) positive terminals and when powered by
          120 VAC 60 Hz can drive a maximum wire and load resistance of 500 ohms.
          Should 120 VAC 60 Hz operation not be desired the 4112 may be powered by an external low voltage DC
          power supply.
       ATTENTION
       The intrinsically safe model is DC powered only.
          The above values are derived from the relationship 11 Vdc + (Rload x 0.02A) = Vdc loop min..
          11 Vdc is required to keep the transmitter in regulation while the load(resistor + lead wires) creates the
          balance of the voltage drop incurred at the 20 mA level.
       ATTENTION
       Determine load resistor values and mounting location per the requirements of your readout device(s).
       REFERENCE
       Figure 1-1, screw terminal label on inner cap.
       CAUTION
       Both outputs must be connected to the negative side of the power supply even if no 4-20 mA input device is
       used for one or both outputs.
              3         Connect one end of the Dew Point load                   Maximum load resistance for ac powered
                        resistor to the H2O Terminal, and wire to the           transmitters is 500 ohms, 250 ohms per channel.
                        Positive Terminal of the readout device.
              5         Connect one end of the Dry Bulb load resistor           If Dry Bulb read out is not used, use a jumper
                        to the TEMP Terminal, and wire to the Positive          between TEMP and COMM in place of the
                        Terminal of its readout device.                         resistor and disregard step 6 below. See Figure 4-
                                                                                2.
              4112
             Terminal
              Block
     L                                         Hot (Black)
         1
                                           120 VAC Power
     L
                                               Neutral (White)                                                     4 - 20 mA
         2                                                                                                       Input Device
                                                                                                           _
     Com                                                                                                       Loop Com
      (-)
      H2 O
                                                                                                AI 1
                                                                                                       {   +   Dew Point
                                                                                                               4 - 20 mA
                                                                                                               Input
                                                                                                           +
                                                                                                               Temperature
 Temp                                                                                                          4 - 20 mA
                                                                                                AI 2
                                                                                                    {      _
                                                                                                               Input
     Gnd                                                                                                                   Jumper
                                    Water Pipe
                                                                         Load Resistors
                                                                           if Required
                                                                                                                                a/n 23317
Note: Always consult the manual of these devices for correct jumper configuraiton and wiring connections.
                   4112
                  Terminal
                   Block
        L                                       Hot (Black)
            1
                                            120 VAC Power
        L
                                                Neutral (White)                                                     4 - 20 mA
            2                                                                                                     Input Device
                                                                                                            _
       Com                                                                                                      Loop Com
        H2 O
            (-)
                                                                                                 AI 1
                                                                                                        {   + Dew Point
                                                                                                              4 - 20 mA
                                                                                                              Input
                             Jumper
                                                                                                            +
                                                                            Load Resistors                      Temperature
       Temp                                                                   if Required                       4 - 20 mA
                                                                                                    {
                                                                                                 AI 2       _
                                                                                                                Input
       Gnd
                                      Water Pipe
Note: Always consult the manual of these devices for correct jumper configuraiton and wiring connections.
            5         Connect one end of the Dry Bulb load resistor         If Dry Bulb read out is not used, use a jumper
                      to the TEMP Terminal, and wire to the Positive        between TEMP and power supply Negative in
                      Terminal of its readout device.                       place of the resistor and disregard step 6 below.
                                                                            See Figure 4-4.
                                           Current Loop
                                           Power Supply
     H20
                                                                                         AI 1
                                                                                                {     Dew Point
                                                                                                    + 4 - 20 mA
                                                                                                      Input
                                                                                                    +
                                                                                                        Temperature
Temp                                                                                                    4 - 20 mA
                                                                                         AI 2   {   _
                                                                                                        Input
Note: Always consult the manual of these devices for correct jumper configuraiton and wiring connections.
   H2O
                                                                                          AI 1
                                                                                                 {   +   Dew Point
                                                                                                         4 - 20 mA
                                                                                                         Input
Temp
                                                                          Load Resistor
                                                                           If Required
  Gnd
                              Water Pipe
Note: Always consult the manual of these devices for correct jumper configuraiton and wiring connections.
          When the monitoring device supplies power for the 4112 Transmitter, both loops must be activated. This is
          necessary for the proper operation of the instrument. Wiring diagrams covering monitoring devices with
          internal power supplies are shown in Figures 4-5, 4-6 and 4-7.
                                                                                                      4 - 20 mA
                                                                                                     Input Device
              4112
            Terminal                                                                             + 24VDC
             Block                                                                                 Internal
                                                                                                 - Supply
                                                                                             -
                                                                                                      Loop Com
     Com
      (+)                                                                                                           Jumper
     H2 O
                                                                                    AI 1
                                                                                        {    +
                                                                                                      Dew Point
                                                                                                      4 - 20 mA
                                                                                                      Input
                                                                                             +
                                                                                                      Temperature
     Temp                                                                                             4 - 20 mA
             Note: Always consult the manual of these devices for correct jumper configuration and wiring connections.
                   Two inputs are needed for proper operation. Both outputs must be connected to the negative side of
                   the power supply, whether one or both outputs are used. If 2 inputs are not available call the
                   Technical Assistance Center (T.A.C.) @ 1-800-343-0228 for assistance..
                           Figure 4-5 DC Wiring Diagram with Internal Supply for Dew Point
                                 and Temperature 4-20 mA Current Loops Monitored
                                                                                                     4 - 20 mA
                                                                                                    Input Device
               4112
             Terminal                                                                           +  24VDC
              Block                                                                                Internal
                                                                                                 - Supply
                                                                                                                   Jumper
   Com                                                                                                -
    (+)
   H 2O
                                                                                     AI 1
                                                                                         {            +
                                                                                            Dew Point
                                                                                            4 - 20 mA
   Temp                                                                                     Input
                                                     Load Resistor
       Gnd                        Water Pipe          If Required
              Note: Always consult the manual of these devices for correct jumper configuration and wiring connections.
                    Two inputs are needed for proper operation. Both outputs must be connected to the negative side of
                    the power supply, whether one or both outputs are used. If 2 inputs are not available call the
                    Technical Assistance Center (T.A.C.) @ 1-800-343-0228 for assistance..
                        Figure 4-6 DC Wiring Diagram with Internal Supply for Dew Point only
                                         4-20 mA Current Loops Monitored
      Com                                                                                 -            +                 V-
                                                                                                       -
       (+)                                                                     AI 1
                                                                                          +                    V+
      H2 O
                                                                                                            Excitation
                                                                                                             Voltage
                                                                                AI 2
     Temp                                                                                 -             +
                                                                                                        -
              Note: Always consult the manual of these devices for correct jumper configuration and wiring connections.
                    Two inputs are needed for proper operation. Both outputs must be connected to the negative side of
                    the power supply, whether one or both outputs are used. If 2 inputs are not available call the
                    Technical Assistance Center (T.A.C.) @ 1-800-343-0228 for assistance.
                           Figure 4-7 DC Wiring Diagram with Internal Supply for Dew Point
                                 and Temperature 4-20 mA Current Loops Monitored
5. System Operation
5.1 Introduction
               Figure 5-1 Simplified System Block Diagram shows transmitter connections labeled L1, L2, COM, H2O
               and Temp.
                                                                    EO +5.0
                                                                               +
                                          DC to DC
                                          Convertor
                                                                      5.0 VDC                      RTD
                                                                               -
                                                  +                                                                           COMM
                                      +                                                                                                               Digital Display
                                                                                   Temp    CCT
                                                                                   Zero    Span                                                           204.7°C
                                                                                                                                                         -104.7°C
                                                                                                                                 Temp.                     D.P.
                                  24 VDC                                                                                         A to D                                            RH OSC
                                                               I2                                                                                                                   Freq
                                               6.9                                                                      Zero              Span
  F101                               CR       VDC                                                                      RV301              RV302
          T 100            C101                                                                                                                                                         RV300              Moisture
                                     100
                                                                                                                                                                                                           Sensor
                                                         VR
                                     -
                   BR100                                 402
                                                                                                                                  D.P.
                           F100                                                                                                  D to A                                                            S-201
                                                  -
                                                        EO -6.9
                                                                                                                                                             MICROPROCESSOR
                                                                                                                                                                                             Display °C/°F
                                                                                                                       Zero               Span
                                     4 mA Temp.                                                                       RV200               RV201                                              Cal Protect
                                                        I1                I3              4 mA D.P.       I4
                                                                                                                                                                                             Scale H.B./D.P.
                                                                                                                                                                                             Not Used
                                         RV400
                                                                                           RV202
                                                                                                                                                                                                          Ext.
                                                                                                                                                                                                         Micro
                                                                                                                                                                                                       Interface
                                                                                                                                                                                                         Conn.
                                                                                                                                                                                      E ² Memory
                                                                                                      Set Zero                RH     +    DP                                           Look Up
                                                      AC Powered                                                 Function/Mode                                                           Ta ble
                                                      Connections                                                  Switches
                                                                                                                                               Down                           Up
 DC Powered                                              See                                                        S-200
 Connections                                          Figure 4-1                                                                                      DP Calibration
     See
  Figure 4-2
a/n 23319
5.4 A to D Conversion
       Once the current output is calibrated a voltage in the temperature circuit proportional to output current
       calibration is routed to 12 bit A to D converter circuitry that digitizes the temperature signal to the
       microprocessor.
       This digitized value is displayed on a microprocessor controlled LCD display.
       Calibration of the A to D converter is accomplished by RV301 and RV302, zero and span controls.
       As moisture levels change and the capacitance of the moisture sensor changes, the RH oscillator changes
       frequency.
       This frequency change is monitored by the microprocessors timer input and provides digitized moisture
       information to the microprocessor.
6.1 Introduction
        Two single – pole, single – throw DIP switch packs are mounted on the microprocessor board and
        interfaced to the microprocessor. S-200 consists of two switches and S-201 consists of four switches.
        The location of the S-200 and S-201 on the dew point assembly is shown in Figure 6-3
S-200
        DIP switch S-200 contains two switches and is located on the microprocessor board perpendicular to the
        display board and directly behind the upper display (dry bulb temperature). S-200 has the letters D.P.
        inscribed in white letters directly beside it (see Figure 6-3). Figure 6-1a illustrates the correct method for
        configuring the switches of DIP switch S-200 in the off position. Each individual switch is OFF when it is
        pointing up toward the white dot and the word OFF located on the DIP switch block. Figure 6-2b illustrates
        the correct method for configuring the switches of DIP switch S-200 in the ON position. The switch is ON
        when it is pointing down toward the switch numbers inscribed on the DIP switch block.
                                    O FF                                            O FF
             A.                                                 B.
             OFF                                                ON
                                           2                                             2
                                     1                                               1
                                    OFF 2 sec, back     Switch S-200-1 OFF only when performing a zero %RH
                                        to ON           calibration. When switch 1 of S200 is turned OFF and then
                                                        back ON, the transmitter assumes a zero %RH standard is
                                                        being applied to the sensor and stores this data in memory.
                                                        Note: Switch 3 of DIP switch S-201 must be in the ON
                                                        position to allow the transmitter to store the zero calibration
                                                        data. If this switch is in the OFF position when activating
                                                        switch 1 of S-200, the lower display will quickly 0.0% RH for
                                                        approximately one second, and then continue displaying the
                                                        value observed before configuring switch 1 of S-200.
S-201
        DIP switch S-201 contains 4 switches and is labeled S201and is located on the reverse side of the
        microprocessor board than S200. See Figure 6-3. Figure 6-2 illustrates the correct method for configuring
        the switches of DIP switch S-201 on and off. Figure 6-2a illustrates the correct method for configuring the
        switches of DIP switch S-201 in the off position. Each individual switch is OFF when it is pointing up
        toward the white dot and the word OFF located on the DIP switch block. Figure 6-2b illustrates the correct
        method for configuring the switches of DIP switch S-201 in the on position. The switch ins ON when it is
        pointing down toward the switch number inscribed on the DIP switch block.
                             O FF
          A.                                                       B.                      O FF
OFF ON
                                            4                                                           4
                                        3                                                         2 3
                                    2
                               1                                                            1
                        D1, D2, D3, W1, R1        Switch 1 of S-201 comes configured in the OFF position from
                                                  the factory for models D1, D2, D3, W1 & R1. The position of
                                                  switch 1 for these models has no effect on the transmitter
                                                  reading/output.
       1    OFF             H1, H2, H3            When in the OFF position and switch 2 of S-201 is in the ON
                                                  position, the transmitter moisture display and output scaling
                                                  units are lb/lb. When switch 2 of S-201 is in the OFF position,
                                                  switch 1 of S-201 has no effect on the transmitter
                                                  reading/output.
            OFF      D1, D2, D3, H1, H2, H3, W1   When switch 2 of S-201 is in the OFF position, the moisture
                                                  display and output scaling are proportional to Dew Point
                                                  temperature.
             ON          D2, D3, H1, H2, H3       When switch 2 of S-201 is in the ON position and switch 1 of S-
                                                  201 is in the ON position, the transmitter moisture display and
                                                  output scaling units are grains/lb. When switch 2 of S-201 is in
                                                  the ON position and switch 1 of S-201 is in the OFF position,
                                                  the transmitter moisture display and output scaling units are
                                                  lb/lb.
       4     ON                 ALL               Dry Bulb and Dew Point/Wet Bulb Displays are in degrees
                                                  Fahrenheit on the transmitter.
            OFF                                   Dry Bulb and Dew Point/Wet Bulb Displays are in degrees
                                                  Celsius on the transmitter.
                                 S201        4 Display On = °F
                          S200                        Off = °C
           1 Zero = Cal                        3 Cal Prot. On = Fact
            Off = Oper                                     Off = Usr
                                                                                                            RTD Temperature
                                              2 Scale On = W.B.
                                                                                                            Sensor Locations
                                                      Off = D.P.
        Span RV201                             1 Not Used
        Zero RV200
                                             2 Display On = D.P.                                            RV400 4mA Temp
                                                      Off = %RH
     DP Loop 4mA
                                                                       Dew Point Up/Down Span
           RV202
                                                                       Counts Pushbuttons
                                                                                                           RTD Card
       External Micro
                                                                             Span Temp 20mA
Interface Connector
                                                                          Mother Card
                                                                                                                      a/n 23320
           Table 6-3 can be used to convert the 4112 transmitter from one model type to another. For example, if we
           take a D1 model type transmitter, the table shows that it can be converted to an R1 or H1 unit by
           appropriately configuring DIP switches S-200 and S-201.
                       Table 6-3 Switch Selectability of Output Units for the 4112 Transmitter
Model Type D1 D2 D3 D4 H1 H2 H3 R1 W1
D1 X — — — X — — X —
D2 — X — — — X — *X —
D3 — — X — — — X *X —
D4 — — — X — — — *X —
H1 X — — — X — — X —
H2 — X — — — X — *X —
H3 — — X — — — X *X —
R1 X — — — X — — X —
W1 X — — — — — — X X
           NOTE: *X indicates a non-temperature compensated relative humidity output signal which is provided
           only for purposes of calibration against a relative humidity reference at room temperature. RH is always 0-
           100% over 4-20 mA. Dry bulb temperature ranges always remain fixed.
       ATTENTION
       •    The RH output is temperature compensated only if the instrument is configured with D1, W1, H1 or R1
            output options.
• With other ranges, RH output reads low above room temperature, and high below room temperature.
                                           7. Troubleshooting
                                                Table 7-1 Troubleshooting
                    Symptom                           Probable Cause                            Action
        Transmitter does not function.        Incorrect connections, reversed     Review Figure 4-1and
        Display is blank.                     polarity
                                                                                  Figure 4-3. Try reversing the
                                                                                  polarity
        Humidity output is locked at 20ma,    Water vapor has condensed on        Withdraw sensor from process. Dry
        dew point reading tracks at or very   the sensor and or the filter        the filter *. Use the sintered
        close to dry bulb temperature                                             stainless steel filter with
        reading. Display reads HHH.H.                                             hydrophobic coating
        Mechanical damage of the              Visible scratches, abrasion, etc.   Use screens or baffles to deflect
        sensor surface                        high flow rates, greater than 30    high speed airborne particulate,
                                              m/sec (100 fps)                     relocate transmitter.
                                                                                  Damaged sensors must be replaced
                                                                                  and the transmitter recalibrated.
        Humidity output is locked             Break in the electrical             Return transmitter to Honeywell for
        permanently, typically at 4ma.        connection between the sensor       service.
                                              and the transmitter electronics:
        Dew point reading does not agree      Reference dew point value is        Convert each reading to the
        with the reading from another         referenced to atmospheric           equivalent humidity ratio (mixing
        reference instrument.                 pressure, while the transmitter’s   ratio) value and then compare the
                                              sensor is reading the gas at        humidity ratio values **
                                              elevated pressure
        * After an hour or so, depending on the amount of condensation, the unit output should be between 4 and 20ma.
       One way to speed the drying of the sensor and filter is to spray them with an isopropyl alcohol solution from a
       simple plastic pump spray bottle.
     ** The partial vapor pressure of a compressed gas mixture is higher than the partial vapor pressure of the same
     gas mixture at atmospheric pressure. Because dew point is directly related to vapor pressure, the dew point of a
     compressed gas will also be higher than the dew point of that same gas at atmospheric pressure. One way to
     compare two dew point readings of the same gas but at different pressures is to convert each reading to the
     equivalent humidity ratio (mixing ratio) value and then compare the humidity ratio values. Given total pressure
     and dew point, the humidity ratio can be calculated. The humidity ratio is commonly expressed as pounds of
     H2O per pound of dry air, grains per pound, kilogram per kilogram, etc.
Sensor Cleaning
     The surface of the sensor may be coated with a film or with a heavy layer of contaminants. This is usually,
     though not always, evident by visual inspection of the sensor surface. The sensor surface is green in color and
     normally appears shiny, sometimes iridescent depending on lighting. A good procedure for cleaning the sensor
     is as follows: Obtain three simple pump spray bottles, Fill the first with a solution of distilled water and mild
     detergent like Ivory soap. Fill the second with distilled water, and the third with a solution of water and
     isopropyl alcohol. Spray the sensor directly with the soap solution, and follow that with distilled water spray to
     rinse it off. Finally a spray of alcohol will help to dry out the sensor quickly. As an alternative to the three
     bottles, the alcohol alone will often be adequate to clear the sensor of surface contaminants. Avoid touching the
     sensor surface directly as it is possible to scratch the sensor’s top protective layer. If necessary, a soft camel hair
     brush may be used to help remove surface dirt.
     For convenience, the sensor may be removed from the end of the probe before cleaning. The sensor connects to
     the probe via four gold-plated pins. When reinstalling the sensor assembly, be sure to match up the notch on the
     sensor assembly housing with the machined flat at the end of the probe.
       REFERENCE
       (Table 6-1 , 6-2 ), Fig. 1-1 , Fig. 6-3, Fig 8-3, and Fig. 9-1
        1       Obtain a screwdriver, suitable power supply,    The Calibration Kit provides the necessary
                Calibration Kit or alternate 0% and 75.3% RH    standards. Near 0% RH can also be obtained
                standards.                                      from bottled Nitrogen gas. A 75.3% RH standard
                                                                may be made by mixing a saturated solution of
                                                                sodium chloride in a suitable container.
        3       Remove inner cover from the transmitter.        Loosen mounting screws(3) and one ground
                                                                screw.
4 Power the transmitter Make sure terminals are covered with black shield
        5       Set S-200-2, labeled DP-2 to OFF; make sure     Bottom display of transmitter will display %RH
                S-201-3 is ON
        7       Insert the probe into the canister with the     Let stand for 30 to 60 minutes. RH reading should
                attached compression fitting, and fasten the    be 0 to 0.1 % after time has elapsed. If so, go to
                fitting around the probe end.                   step 9. If higher, go to step 8.
        8       Activate DIP switch S-200-1,labeled DP-1 to     S-200 is on the top edge of the microprocessor
                OFF position for approximately 2 seconds then   board in Figure 6-3. The display should be
                return to ON position.                          reading 0.0% RH.
Note: Factory calibration values can be restored by setting DIP switch S-201-3 to OFF, while power is applied to
transmitter.
         2     Add distilled water (about 10 ml) to form a           Slush/slurry appearance, salt crystals should be
               saturated solution.                                   visible in solution Avoid wetting the plastic filter in
                                                                     the bottle.
         3     Insert the probe via the plastic filter fitting.      Let stand for approximately two hours. The lower
               Insert the bottle into a temperature stable           transmitter display should read 75.3% RH @
               environment, see Figure 9-1 Calibration Setup         75°F/24°C after the two hours.
         5     Remove the Honeywell nameplate from the               Moisture display increments up or down. CAL is
               face of the transmitter.                              displayed until actuator is released. After
                                                                     adjustment, CCC.C appears in the display.
               Use a flathead screwdriver to rotate slot D
               clockwise to increase, counter clockwise to
               decrease display of %RH until lower displays
               read 75.3% RH.
         6     Momentarily press and release again.                  If CCC.C does not clear, the switch button is
                                                                     stuck, or there is an internal fault. Have the
                                                                     transmitter serviced.
         7     Set S-200-2 back to the ON position. The              Do not turn OFF S-201-3 or the original factory
               calibration is complete.                              span and zero calibration will be restored.
                                           N.C.           L1
                                                                             120 VAC          R1              R2
                                           N.C.           L2                 60 Hz             DVM              DVM
                                         Comm             Comm
                                           H2 O           H2 O
Temp Temp
INC
Z S D
a/n 23321
       CAUTION
       •        Do not attempt this procedure unless temperature bath can be measured and controlled to 0.01deg c via
                NIST traceable standards.
• Do not remove the P.C. assemblies from the lower housing as this affects calibration
       REFERENCE
       (Table 6-1, 2 ), Fig. 1-1 , Fig. 6-1, fig 8-1, and Fig. 9-1
            1         Obtain a power supply, resistance decade          Connections are shown in Fig. 8-1, electrical test
                      box(0-5000 ohms, settable to .01 ohm), load       set up
                      resistors and DVM. Also tip cover( Honeywell
                      part number 225-0158)
5 Connect resistance box to RTD block Set for zero ohms(or short box terminals)
            7         adjust RV-400 trimpot on the RTD Board            Temperature Channel output of 3.98 to 4.00
                                                                        milliamps cutoff
            8         adjust RV-301on the A to D Board                  Temperature Display reading of 000.0. Dew Point
                                                                        display should now read -104.7.
            9         adjust RV-200 trimpot on the Microprocessor       0.000 millivolts output +/- 50 microvolts at pins 1
                      Board                                             and 2 of U-200 IC
            10        adjust RV-202 trimpot on the Microprocessor       Dew Point Channel output of 3.98 to 4.00
                      Board                                             milliamps cutoff.
       12     adjust the Zero trimpot on the Mother Board     for slightly over 4 milliamps Temperature Channel
                                                              output and then slowly back to 4.00 milliamps
13 adjust RV-301 trimpot on the A/D board Temperature Display reading of 000.0.
       15     adjust the SPAN trimpot on the Mother Board     20.00 milliamps output on the Temperature
                                                              Channel
       16     adjust trimpot RV-302 located on the A/D        Temperature Display to read 204.7.
              Board
       18     adjust RV-201 trimpot on the Microprocessor     for Dew Point channel 20.00 milliamps output
              Board
22 Place protective tip cover on end of probe prevents water/oil Sensor contamination
       24     adjust the Zero trimpot located on the Mother   for slightly over 4 milliamps Temperature Channel
              Board                                           output then slowly back to 4.00 milliamps output.
       26     adjust the Span trimpot located on the Mother   for the Temperature Channel milliamps output by
              Board                                           using the following transfer equation:
                                                                                               o
                                                              milliamps output = (( Bath Temp C / 204.7 ) x 16
                                                              mA )) + 4 mA.
       30   using screw driver slot switch activator D (Fig.   read 75.3 % RH.
            8-1), increment or decrement the Display
                                               N.C.             L1
                                                                                          120 VAC
                                               N.C.             L2                        60 Hz
                                            Comm                Comm
                                              H 2O              H 2O
                                             Temp               Temp
Jumper
INC
                                                      Z S D
             Ambient Temp. Zero
              (±Multiturn Pot.)                                                      Dew Point
                                                                                   Cal. Actuator
                                                                                (±10° multiturnOnly)
                Ambient Temp. Span
                  (±Multiturn Pot.)
Any Container
                Fine Gravel
                Thermal Mass
                (I.E. Aquarium Gravel)
3"
                                                                            Solution with
                                                                            Sodium Chloride
                                                       2" min               Crystals in a
                                                                            Saturated Solution
a/n 23322
10.1 Introduction
          The Dew Point Transmitter has a maximum sensor operating temperature of 185°C. The optional Cooler
          allows the safe operation of the sensor at process temperatures of up to 540°C (1000°F).
          The Cooler surrounds the Dew Point probe with a jacket of flowing external air. Customer supplied air is
          directed through the inlet port on the Cooler. The air is vented through a muffler on the outlet port. A safe
          operating temperature can be maintained by controlling the air pressure.
          Table 10-1 Cooler Performance, shows the cooling capability of the cooler. The data was obtained by
          inserting the Dew Point Transmitter and Cooler into an oven in still air. The table lists sensor temperatures
          for various oven temperatures and inlet pressures. For example, with oven temperature at 427°C and inlet
          pressure at 20 psig, the dew point sensor temperature is 109°C, well below the 185°C maximum sensor
          rating.
50 ----- ----- 89 62 49 37
40 23 3 102 74 53 39
30 16 2.3 120 88 60 43
20 10 2 152 109 75 50
       CAUTION
       Maximum sensor temperature is 185°c. exceeding this temp. will damage the sensor and void the warranty.
       REFERENCE
       Data Developed using a 12 in. probe and cooler with 5 (0.187 dia.) vent holes. Actual sensor temperatures
       depend upon actual process ambient (moving vs static air) and inlet process air temperature.
.700
                                                                                                         90 o Street Elbow
               1.25"                                                                                     304 SST
               DIA.
10.3 Installation
     CAUTION
     Installation must not permit exceeding the 185°C sensor limit or the 80°C max electronics operating limit. To
     protect the sensor, the air supply must be reliable and over -temp and loss of air alarms are recommended.
Step Action
        1      Obtain a pressure regulator and gauge capable of delivering 5 scfm @ 50 psig., suitable tubing to
               connect the air supply to the ¼ “ NPT inlet port, and a means to measure the cavity temperature.
2 Remove the Cooler from the transmitter. Loosely plug the ½” NPT mounting port temporarily.
4 Install the pressure regulator and gauge from the air supply to the Cooler inlet.
        6      When the process temperature is at its hottest, allow 30 minutes for the Cooler cavity to reach its
               maximum.
7 Measure the cavity temperature. Make sure it does not exceed 185°C.
        8      For best performance, adjust the pressure to make the cavity temperature about 25 deg. above the
               normal process dew point. See
11.1 Introduction
       The Honeywell model 4112 Dew Point Transmitter is approved by Factory Mutual:
       Case: NEMA 4X, approved units have FM labeling.
       Explosion proof: Class I, Division 1, Groups B, C, and D; dust-ignition proof for Class II, Division 1,
       Groups E, F, and G indoor and outdoor locations.
       Intrinsically Safe: Class I, Division 1, Groups A, B, C, and D; dust-ignition proof for Class II, Division 1,
       Groups E, F, and G indoor and outdoor locations in accordance with Drawing No. A-21008, Rev 4-7-94
                                    3 Positive
 2 Loop                              Barriers
 24V (40 mA MIN)                  (Notes 3 & 4)
 DC Supply
               0.1A                                        8-40 mA
               Fuse
                                                                        COM
                                                                                             Notes:
            Power Supply Ground                                                              1. Vmax must be greater than Vt of the FMRC approved
                                                           4-20 mA +                            barrier combination.
                                                                        H2 O
                                                                                                Imax must be greater than It of the FMRC approved
      Alternate Load Resistor
      Location (Note 2)                                                                         barrier combination.
                                                           4-20 mA +                            Ci plus interconnecting wiring must be less than C0 of the
                                                                          TEMP                  FMRC approved barrier combination.
                                                                                             2. Preferred location of 2500 load resistor is for high
                                                                    CAT 4112                    impedence receiver input circuit for less than 250k Ohm
              Signal Common                                         Dew Point/                  input impedance move load resistor to receiver terminals.
                                                                   Temperature                  This adds barrier resistance to transmitter load.
 1-5V Receiver                                                      Transmitter              3. Install in accordance with manufacturers instructions and
 Input Terminals                                                     (Note 1)                   NEC Article 504 and ANSI/ISA RP12.6.
 Dew Point                                                                                   4. Typical connections are shown to barrier unit terminals.
      Alternate Load Resistor                                                                   Maximum fault voltage in any circuit connected to barrier
      Location (Note 2)                                         Preferred Load                  terminals must not exceed 250V RMS or 250VDC.
                                                                Resistor Location            5. A Approved transmitters:
                                                                (Note 2)                          4112 02-D1.d2-C-100 207
                                                                                                           to D3H1-f-104 208
               Signal Common                                                                              60 H3W1               209
                                                               1/4 X 1 3/4 Copper Bus Bar       B See sheet 2 of this drawing for Stahl barrier system
 1-5V Receiver                                                                                      and sheet 3 for MIL barrier system.
                                                               for Mounting Barriers. Must   6. Use only NRTL listed dust tight fitting or threaded conduit
 Input Terminals                                               Have 2 Earth Ground
 Temperature                                                   Connections. Bus Bar to          for installation in Class II, Division I, Groups E, F, and G.
                                                               Ground Resistance 1 Ohm
                                                               Maximum.
                                                                                                          Intrinsic Safety Entity Paramters:
            NONHAZARDOUS LOCATION                                                                         Vmax = 30 volts
                                                             HAZARDOUS LOCATION
                                                             Class I, II, III                             Imax = 360 mA
                                                             Division 1                                   Ci = 0.1uF
                                                             Groups A, B, C, D, E, F, G                   Li =      mH
OPTION I
I S Ground
 HAZARDOUS LOCATION
                                              NONHAZARDOUS LOCATION
 Class I, II, III
 Div 1
 Groups C, D, E, F, G
12. CE Approvals
Introduction
       The following statements covers the subject of CE approvals for the 4112 Dew Point Transmitter.
       CE conformity: This product is in conformance with the protection requirements of the following
       European Council Directive: 89/336/EEC, the EMC directive. Conformance of this product with any
       other “CE Mark” Directive(s) shall not be assumed.
                                                       Attention
       The emission limits of EN 50081-2 are designed to provide reasonable protection against harmful
       interference when this equipment is operated in an industrial environment. Operation of this equipment in a
       residential area may cause harmful interference. This equipment generates, uses, and can radiate radio
       frequency energy and may cause interference to radio and television reception when the equipment is used
       closer than 30 meters to the antenna(e). In special cases, when highly susceptible apparatus is used in close
       proximity, the user may have to employ additional mitigating measures to further reduce the
       electromagnetic emissions of this equipment.