Electronic
Pressure Measurement
Differential Pressure Transmitter
Model DPT-10
WIKA Data Sheet PE 86.21
Applications
Process engineering
Pharmaceutical
Food and beverage industries
Special features
High measuring accuracy
Scaleable measuring ranges
Various hazardous area approvals
Seven different housing variants
Configuration via DTM (Device Type Manager) according
to FDT (Field Device Tool) - concept (e.g. PACTware)
Bildunterschrift
Differential Pressure Transmitter Model DPT-10
Description
With its 4 ... 20 mA, 4 ... 20 mA HART®, PROFIBUS® PA There are seven different housing variants available, and
or FOUNDATION Fieldbus™ output signals, combined with thus, it is possible to select a variant suited to every operating
either intrinsically-safe or flameproof ignition protection environment. The housing itself can be rotated through 330°,
(in accordance with ATEX), the DPT-10 is ideally suited to and is available in plastic, aluminium and stainless steel.
applications requiring these features. The electronics of all An electropolished stainless steel housing (316 L) is avail-
of these transmitters, even for the flameproof variant, are able to meet the high demands of the food industry and
intrinsically safe. Thus it is possible to make adjustments on pharmaceutical industry.
the instrument in EX areas while the instrument is live.
Easy configuration and operation
Versatile in application Service and configuration at the instrument is carried out
The DPT-10 is suitable for many industrial measuring require- using the optional display and operation module, which can
ments, such as flow measurement using differential pressure be fitted in four positions. The operation menu has a simple
transducers, level measurement or filter and pump monitor- and self-explanatory structure and has seven selectable
ing. With diaphragm seals fitted, the DPT-10 is also suitable languages as standard. Alternatively, the operating param-
for harsh process conditions. As a result of the available eters can be set using the PACTware™ free and non-propri-
measuring ranges of 0 ... 10 mbar up to 0 ... 40 bar and a etary configuration software. An instrument-specific DTM
static pressure limitation of up to 420 bar, the instrument can enables easy integration into a corresponding Distributed
be used in almost any application. The internal digital signal Control Systems.
processing, combined with proven sensors, guarantees high
accuracy and the best long-term stability.
WIKA Data Sheet PE 86.21 ∙ 04/2010 Page 1 of 8
Data sheets showing similar products and accessories:
Process Pressure Transmitter; Model IPT-10, standard version; see Data Sheet PE 86.11
Process Pressure Transmitter; Model IPT-11, flush diaphragm; see Data Sheet PE 86.11
Specifications Model DPT-10
Measuring range 1) bar 0.01 0.03 0.1 0.5 3 16 40 2)
Max. static working pressure bar 160 160 160 160 {420} 160 {420} 160 {420} 160 {420}
Smallest possible span mbar 0.25 0.3 1 5 30 160 400
Lowest static pressure 3) mbar abs 0.1
Overload on one side bar 160 160 {420}
Overload on both sides bar 240 240 {630}
Accuracy
Measuring ranges bar < 0.5 ≥ 0.5
Long-term stability % URL/Year ± 0.18 ± 0.05
Reference accuracy 4) % of the span Measuring ranges 10, 30 mbar: up to TD 15:1 ± 0.075 %
from TD 1:1 ± 0.15 % x TD TD from 15:1 ± (0.0015 x TD + 0.053) %
100 mbar:
up to TD 4:1 ± 0.075 %
from TD 4:1 ± (0.012 x TD + 0.027) %
Total Performance 5) 6) ± 0.15 % ± 0.15 %
Influence of the system pressure 6)
Zero point % URL ± 0.35 / 70 bar ± 0.075 / 70 bar
Measuring range 10 mbar: 0.15 / 7 bar
Span % URL ± 0.14 / 70 bar ± 0.14 / 70 bar
Measuring range 10 mbar: 0.035 / 7 bar
Influence of the media and ambient temperature 6)
-10 ... +60°C 10 mbar and 30 mbar 0.5 bar, 3 bar, 40 bar ± (0.08 x TD + 0.05) %
± (0.31 x TD + 0.06) % 16 bar ± (0.1 x TD + 0.1) %
100 mbar ± (0.18 x TD + 0.06) %
-40 … -10 / +60 … +85 °C 10 mbar and 30 mbar 0.5 bar, 3 bar ± (0.12 x TD + 0.1) %
± (0.45 x TD + 0.1) % 16 bar ± (0.15 x TD + 0.2) %
100 mbar ± (0.3 x TD + 0.15) % 40 bar ± (0.37 x TD + 0.1) %
Mounting position influence mbar ≤4
Permitted temperature ranges
Ambient 7) °C -40 … +80 (without Display) -20 … +70 (with Display)
Transport/Storage -40 … +80
Process limits dependent upon the °C FKM/NBR: -20 … +85
sealing material 7) PTFE, copper: -40 … +85
FKM, free of oil and grease: -10 … +85
for oxygen applications Copper, PTFE: -20 … +60; FKM: -10 … +60
With differential pressure lines longer than 100 mm: -40 … +120 (-10 … +120 with
Temperature limits °C
lateral flanges C22.8)
Material
Wetted parts Process connection C22.8, {316L, C276}
Diaphragm: 316L, C276, {Tantalum, C276 Gold-Rhodium coated, Monel400®}
Seals: FKM/FPM, NBR, copper, {PTFE}
Internal transmission fluid 8) Silicone oil (halocarbon oil for oxygen applications)
Case Plastic (PBT; polyester), {aluminium}, {stainless steel 316L}
Weight kg approx. 4.2 … 4.5 depending on the process connection and case version
Electrical data
Power supply UB V DC non-hazardous: 14 … 36 Ex ia: 14 … 30; Ex d: 20 … 36;
{FOUNDATION fieldbus™ und PROFIBUS® PA Ex ia: 9 ... 24 Ex d 12 … 32}
Output signal 4 … 20 mA, 2-wire {4 ... 20 mA, 2-wire with a superimposed communication signal
HART®}, {FOUNDATION Fieldbus™}, {PROFIBUS® PA}
Dead time ms 100
Time constant (63 %) ms 180 (measuring ranges 10, 30 mbar: 250)
Damping s 0 … 999, adjustable
Permissible max. load RA in Ω RA = (UB – UBmin) / 0.023 A
Page 2 of 8 WIKA Data Sheet PE 86.21 ∙ 04/2010
Explosion protection
Ex protection 9) ATEX Category II 1G, II 1/2G, II 2G Ex ia IIC T6…T1
II 1/2 G, II 2 G Ex d ia IIC T6…T1
Ambient conditions
CE-mark EMC 2004/108/EC interference emission and interference immunity per EN 61 326-1
industrial application
Interference emission Limit Class A and B, 94/9/EC EN 50 014 (general part),
EN 50 020 (intrinsic safe), EN 50 284 (zone 0), {EN 50 281-1 (dust Ex)}
Shock resistance g 100 per IEC 60 068-2-27 (Shock mechanical)
Vibration resistance 10) g 4 (5 … 100 Hz) (Vibration under resonance)
Electrical ingress protection Overvoltage Category III Protection Class II
IP66/67 (standard housing)
{} Items in curved brackets are optional extras for additional price.
1) Other measuring ranges can be set via the respective Turndown.
2) Measuring range 40 bar, "-" side overload range unilateral up to 100 bar.
3) Valid with reference conditions according to IEC 60 770.
4) Includes non-linearity after limit point setting, hysteresis and non-repeatability in accordance with IEC 60 770.
5) Includes non-linearity, hysteresis, non-repeatability, thermal changes, zero point and static pressure influence (Pstat= 70 bar) in the temperature range -10 … +60 °C.
6) Values are not valid for tantalum diaphragm.
7) Restrictions of process temperature for oxygen -20 … +60 °C / the lowest temperature for PN 420: -10 °C.
8) Operating pressure with halocarbon oil - higher than 1 bar abs.
9) The operating conditions and safety-relevant data in the approval documents must be read.
10) Tested in accordance with directives, GL characteristic line 2 (not for double chamber cases in stainless steel).
URL = basic measuring range
TD = Turndown
Electrical connection
Non-hazardous area Hazardous area
Legend:
}
power supply ― supply minus
2-wire connection
load + supply plus
Electrical connection
Non-hazardous area Hazardous area
PLC
Segment coupler
Power hub
WIKA Data Sheet PE 86.21 ∙ 04/2010 Page 3 of 8
Process connection
Oval flange, connection 1/4 -18 NPT or RC 1/4 with vent at rear
Measuring cells 10 and 30 mbar
Measuring cells ≥ 100mbar
Connection Mounting Material Equipment
¼-18 NPT IEC 61518 7/16-20 UNF Stainless steel C 22.8 2 vent valves 1)
¼-18 NPT IEC 61518 7/16-20 UNF AISI 316L 2 vent valves 1)
¼-18 NPT IEC 61518 7/16-20 UNF C276 Without valves / plugs
RC ¼ 7/16-20 UNF AISI 316L 2 vent valves 1)
¼-18 NPT IEC 61518 PN 160: M10; PN 420: M12 Stainless steel C22.8 2 vent valves 1)
¼-18 NPT IEC 61518 PN 160: M10; PN 420: M12 AISI 316L 2 vent valves 1)
¼-18 NPT IEC 61518 PN 160: M10; PN 420: M12 C276 Without valves / plugs
1) Material: AISI316L /1.4404
Oval flange, connection 1/4-18 NPT or RC 1/4, with lateral venting
Measuring cells 10 and 30 mbar
Measuring cells ≥ 100 mbar
Connection Mounting Material Equipment
¼-18 NPT IEC 61518 7/16-20 UNF Stainless steel C 22.8 2 vent valves, 4 plug screws 1)
¼-18 NPT IEC 61518 7/16-20 UNF AISI 316L 2 vent valves, 4 plug screws 1)
¼-18 NPT IEC 61518 7/16-20 UNF C276 Without valves / plugs
RC ¼ 7/16-20 UNF AISI 316L 2 vent valves, 4 plug screws 1)
1) Material: AISI316L /1.4404
Page 4 of 8 WIKA Data Sheet PE 86.21 ∙ 04/2010
Housing variants
Single chamber housing, plastic Single chamber housing, aluminium Single chamber housing, cast stainless
steel
Single chamber housing, stainless
steel, deep-drawn
Dual chamber housing, plastic Dual chamber housing, aluminium Dual chamber housing, cast stainless
steel
WIKA Data Sheet PE 86.21 ∙ 04/2010 Page 5 of 8
Accessories
Three-way valve manifolds Five-way valve manifolds
With upstream pressure equalising valves it is possible to These pressure equalising valves with integrated shut-off
avoid unilateral overpressure loadings during both start-up valves, and also purge and ventilating valves, in addition
and operation as well as enabling zero point checks during enable the venting of the pressure gauge on one or both
operation. Furthermore, they enable the isolation of the sides and enable the supply line to be purged.
process lines without interference to the running process.
Diaphragm seals Differential pressure transducer
Model DPT-10 differential pressure transmitters can be Differential pressure transducers for flow measurement are
adapted to the harshest conditions in the process industry by available as accessories. Depending on the application, the
using diaphragm or diaphragm in-line seals. Thus, the transmit- differential pressure transducers are designed as simple
ters can be used at extreme temperatures, and with aggressi- push plugs, measuring flange or complete measuring path.
ve, corrosive, heterogeneous, abrasive, highly viscous or toxic
media. As a result of the wide variety of aseptic connections
(such as clamp, threaded pipe or DIN 11 864 aseptic connec-
tions) measuring assemblies meet the high demands of sterile
process engineering.
Page 6 of 8 WIKA Data Sheet PE 86.21 ∙ 04/2010
Display and operating module
Menu languages:
German
English
French
Display and menu item number
Spanish
Polish
LC-Display
Italian
Dutch
Operating keys
DTM user interface
For HART output signals, Profibus-PA and FF, a DTM is
available in accordance with the FDT-standard. The DTM
provides a self-explanatory and clear user interface for all
setup and control processes of the transmitter. For testing
purposes, it is also possible to simulate process values
and archive the parameter data.
Recording of the measured values is available for
diagnostic purposes.
Tank volume calculation
The additional tank volume calculation of the DTM
function can be used to reproduce any optional tank
geometry. The corresponding linearisation table is
generated automatically. The linearisation table can be
transferred directly to the transmitter.
WIKA Data Sheet PE 86.21 ∙ 04/2010 Page 7 of 8
Accessories
Model Special features Order No.
DIH52-F and DIH62-F Indication module DIH52-F and DIH62-F, 5-digit display, 20 segment bargraph, without On request
separate power supply, with additional HART®-functionality. Automatic adjustment of
measuring range and span. Local-master functionality: Setting the measuring range and
unit of the connected transmitters using HART®-standard commands. explosion protection
according to ATEX.
Model 010031 HART®-Modem for USB-interface, specifically designed for use with modern notebooks 11025166
Model 010001 HART®-Modem for RS232-interface 7957522
Model 010041 HART®-Modem for Bluetooth-interface [EEx ia] IIC 11364254
FC475HR1EKL9 HART® protocol, Li-Ion battery, power supply 100 ... 240 V On request
coloured indication with background lighting, Bluetooth and infrared interface ATEX, FM,
CSA and IECEx(i) (including FISCO if available).
FC475HR1EKLU HART® protocol, NIMH battery, power supply AC 90 ... 240 V, with EASY UPGRADE, ATEX On request
II 2G (1GD) EEx ia IIC T4
MFC4150 HART® protocol, universal power supply, cable set with 250 Ω resistance, with DOF up- 11405333
grade, with Ex-protection
DTM Collection
incl. PACTware, contains DTMs for WIKA field devices 12513636
Display and operation module, housing cover in aluminium, with window 12298884
Display and operation module, housing cover in cast stainless steel, with window 12298906
Display and operation module, housing cover in plastic, with window 13315277
Display and operation module, housing cover in stainless steel, electropolished, with window 13315269
Three-way valve
On request
CrNi-steel, PN 420, shape A, Nace compliant, 3.1 material certificate
Five-way valve
On request
CrNi-steel, PN 420, shape A, Nace compliant, 3.1 material certificate
Oval flange 1/4 NPT Cr-Ni-steel
On request
CrNi-steel, PN 420, shape A, Nace compliant, 3.1 material certificate
Oval flange ½ NPT Cr-Ni-steel
On request
CrNi-steel, PN 420, shape A, Nace compliant, 3.1 material certificate
Mounting bracket for wall or pipe mounting, stainless steel 11553945
The specifications given in this document represent the state of engineering at the time of publishing.
We reserve the right to make modifications to the specifications and materials.
Page 8 of 8 WIKA Data Sheet PE 86.21 ∙ 04/2010
04/2010 GB
WIKA Alexander Wiegand SE & Co. KG
Alexander-Wiegand-Straße 30
63911 Klingenberg/Germany
Tel. (+49) 9372/132-0
Fax (+49) 9372/132-406
E-mail info@wika.de