Digital Radiography & Image Production Review
Digital Radiography & Image Production Review
Image Production
             Presented by:
John Fleming, M.Ed., RT(R)(MR)(CT)
        St. Petersburg College
       Office: (727) 341-3758
   E-mail: flemingj@spcollege.edu
          Lesson Objectives:
   ARRT Content Specifications
   Image Production
      Image Acquisition & Technical Evaluation
      Equipment Operation & Quality Assurance
   Midterm & Final Exam
                        https://www.arrt.org/Handbooks
                         2017
January-December 2017
ARRT Content Specifications:
         Erased Cassette
CR vs. Background Radiation:
          One Week
CR vs. Scatter Radiation:
CR Collimation Considerations:
   If the collimated borders are not recognized,
    the image data analysis will include all of the
    data outside the collimation.
   The final image may not have the correct level
    of brightness and contrast.
   Tightly collimated images should be placed in
    the center of the IR.
Exposure Indicators vs. mAs (x):
                       Symbol   ½(x)   Perfect (x)    2(x)
Fuji/Konica Minolta/
                        S#      400       200         100
  Siemans/Philips
   S# 397            S# 201           S# 98
 Which of these images would be most ideal?
Direct Radiography
        Unit
    Direct Radiography (DR):
   DR employs a flat panel image receptor (IR)
    that is about the size of a CR cassette.
   Two Types of Flat Panel IRs:
      Indirect Capture
      Direct Capture
Flat Panel IR:
Flat Panel IR:
Indirect Capture Flat Panel IR:
                                      Image Forming Beam
                                          Surface Reflector
                                      Scintillation Detector Array that
                                      contains CsI or GOS
Digital Frequency
                                               Raw Data
     Raw Image
Histograms: Luminance Values
                   Bone Area
                                               Direct Exposure Area
                    Digital Frequency
                                            Bone Area
                                                         Soft Tissue
                                                             Area
Raw Data
Raw Data
         Look-up-table Algorithm
      Look-up-tables:
                              2048
                              0
  0          1024         2048
Pixel Values from the Rescaled Image
           Look-up-tables:
Rescaled                      No Change from
 Image                        Original Image
                 AP Hip
              Look-up-Table
                Algorithm
                              0
  0          1024         2048
Pixel Values from the Rescaled Image
                         Look-up-tables:
                                                      Processed Pixels
Original Pixels
                   45                          11
                               AP Hip
                  600       Look-up-Table      725
                              Algorithm
                  1420                        2011
                               AP Hip
                            Look-up-Table
                              Algorithm
             Histogram          LUT
Image Quality vs. CR Image Processing:
  kVp:       55          65         75
  mAs:       5           2.3        1.2
  S#:        111         109        106
  Contrast   2.68        2.68       2.68
 Note how image quality remained fairly constant.
  Digital Optimum kVp Values:
Chest (Grid)            110 to 130
Chest (Non Grid)        80 to 90
                                   * Be sure to follow
Abdomen                 80 to 85
                                     hospital protocol.
Abdomen (Iodine)        76 to 80 *
Extremities (Non Grid) 65 to 75 **
Extremities (Grid)      85 to 95   ** If possible, avoid
AP Spines               85 to 95      using less than 1
Lateral Spines          85 to 100     mAs during any
Ribs                    80 to 90          exposure.
Skull                   80 to 90
Barium Studies          110 to 120
    Modified from Barry Burns
          ALARA: Dose Creep
   Konica S# Range: 100 to 400
      kVp         mAs         S#     Tube Output
       80         75         107       321 mR
       80         38         211       164 mR
       80         20         406       87 mR
     *92          10         409       60 mR
      Total Reduction in Tube Output: 535%
      Total Reduction in Patient Dose???
      *Using the 15% Rule will decrease tube output by
      approximately one third.
 Konica S# Range 100 to 400
kVp    80           80          80          92
mAs    75           38          20          10
S#     107          211         406         409
mR     321          164         87          60
 Note how image quality remained fairly constant.
           Spatial Resolution:
   Controlled by the following:
     Size of the Matrix
     Pixel Size
     Field of View (FOV)
     Detector Element Size for DR
     Sampling Frequency for CR
           Spatial Resolution:
   Matrix
     Made of columns and rows of cells (pixels).
            Spatial Resolution:
   Pixel (Picture Element)
      They are generally square.
   Voxel (Volume Element)
           Pixel              Voxel
            Spatial Resolution:
   Pixel Pitch
       Refers to the distance between each pixel.
    Pixel Size vs. Matrix Size:
     Large           Medium            Small
24 x 30
           35 x 43
   Note how the pixel pitch has changed.
Which FOV has the highest spatial resolution?
     Spatial Resolution vs. FOV:
   Some manufactures will decrease the pixel
    pitch as the FOV is reduced.
   The same number of columns and rows are
    packed into a smaller FOV.
   The original matrix is maintained within the
    smaller FOV and results in an increase in
    spatial resolution.
   Spatial Resolution vs. FOV:
24 x 30
             35 x 43
Since the pixel pitch remains the same, the spatial
            resolution is unchanged.
     Spatial Resolution vs. FOV:
   Other manufactures maintain the same pixel
    pitch as the FOV is reduced.
   The net result is a decrease in the size of the
    matrix while the pixel size remains the same.
   Bottom line: there is no change is spatial
    resolution.
CR Spatial Resolution:
         A thinner diameter laser
         will create a higher
         sampling frequency.
          A higher sampling
          frequency will decrease
          pixel pitch and result in
          higher spatial resolution.
CR Spatial Resolution:
    CR Spatial Resolution:
                F/S Latitude
                                  Saturation
 Density
Brightness
Quantum Noise
                 Exposure in mR
        Quantum Noise:
1024
                     0
0     1024       2048
    Inversion:
                    2048
1024
                     0
0     1024       2048
Inversion:
Inversion:
                Equalization:
   This is also referred to as Dynamic Range
    Compression (DRC).
   An algorithm is employed to reduce file size by
    subtracting the extreme light and dark areas of
    the image.
   The remaining dynamic range will be modified
    to improve the overall contrast of the image.
            Equalization:
Dynamic Range
Compression (cropping)
Shape Size
             Abdomen
         *                        **
   *               IR        **
10 mAs at 50 kVp        20 mAs at 50 kVp
                  15% Rule:
   To maintain the RE, increase the kVp by 15% and
    reduce the mAs by 50%.
        Using the 15% Rule in this manner will reduce
        tube output by approximately one third.
   How will reducing the kVp by 15% and doubling
    the mAs affect the RE?
        How about the patient dose?
                  15% Rule:
   The 15% Rule vs. RE and tube output:
       kVp      mAs       Tube Output        RE
       115       5           40 mR         30 mR
       100       10          60 mR         30 mR
       85        20          80 mR         30 mR
   What might be another clinical application for this
    rule?
                 15% Rule:
   How would you change the following technique if
    you wanted to use kVp to help reduce motion but
    maintain the original RE?
          80 kVp 100 mA 200 ms (0.2 s)
              80 kVp x 1.15 = 92 kVp
               200 ms x 0.5 = 100 ms
    New Technique: 92 kVp 100 mA 100 ms (0.1 s)
       The Anode Heel Effect:
   X-rays are emitted isotropically from the anode.
   The intensity of the beam decreases as you travel
    closer to the anode side of the tube.
The Anode Heel Effect:
                               Cathode
Stator
                                 (-)
                    Target
Anode (+)
         Filament
                             Focusing Cup
  The Anode Heel Effect:
Stator         Anode
                (+)              Focusing Cup
                                  69.5
                                  69       90
                 X-ray Photon Energy (keV)
The X-ray Emission Spectrum:
                           With 2.5 mm of
                         Filtration At 90 kVp
  Quantity or
  Intensity of
     X-ray
    Photons
                      30          69.5
                                  69       90
                 X-ray Photon Energy (keV)
Added Filtration vs. the X-ray Emission Spectrum:
                   What  happened
                   What happened     to the beam intensity?
                                 to the
                   average energy of the
                             How about
                   primary beam?           the RE?
    Quantity or
    Intensity of                     2.5 mm
       X-ray
      Photons
                                    3.5 mm
                         30          69.5
                                     69       90
                    X-ray Photon Energy (keV)
Collimation vs. RE:
Patient
                            IR
         Image
   Tight Collimation
Collimation vs. RE:
Patient
                                IR
      Radiographic Image
  *   No Collimation       *
Collimation vs. RE:
Patient
                                IR
      Radiographic Image
  *   No Collimation       *
Collimation vs. RE: Magnified
                IR
Radiographic Image
       *                    *
           No Collimation
Collimation vs. RE: Magnified
                    IR
Radiographic Image
         *
  Bottom Line:
                                *
                 Lack of collimation adds
    unwanted scatter (noise) to the IR.
Collimation vs. RE: Konica CR
 24 x 30 cm Collimation          No Collimation
S = 142 S = 103
                          new SID
       The Direct Square Law:
   This formula is employed to maintain beam
    intensity at a new distance
                   I1    D12
                   I2    D22
                            old SID
                   Example #1:
   If your original beam intensity was 5.0 mR at a 30” SID,
    what would be your new beam intensity if the SID was
    changed to 60”?
                              •
               new mR = old mR      old SID   2
new SID
60”
                      new mR = 1.25 mR
                 Example #2:
   Your original SID was 40” and 75 mAs produced an
    acceptable RE. What new mAs would you use at 60” if
    you wanted to maintain the original RE?
                            •
             new mAs = old mAs new SID      2
old SID
40”
                       Patient
           *   *
       *
                          Grid
                   *
                           IR
           Grid Construction:
   Grid Frequency
   Grid Ratio (GR)
      The height of the lead strips is divided by the
      distance between them.
      GR = h/D
      With all other grid construction factors
      constant, the higher the GR, the greater the
      scatter clean-up.
      Higher GRs also require more accuracy in their
      use and result in a higher patient dose.
Grid Ratio Calculation:
              H
              H == 24
                   4.0mm
                       mm
                   GRGR= =h/D
                            H/D
                   GRGR= =4.0/0.5
                            24/4
                   GRGR= =8:1
                            6:1
DD==0.5
     4 mm
        mm         6060Lines/Inch
                        Lines/Inch
              H
              H ==24
                   4.0mm
                       mm
                   GR
                   GR == 4.0/0.25
                         24/1
                   GR
                   GR == 16:1
                         12:1
                   120
                   120 Lines/Inch
                       Lines/Inch
 DD = 2 mm
  = 0.25 mm
           Grid Construction:
   Grid Radius or Focal Range
       Parallel vs. Focused Grids
  Parallel Linear Grid:
72”
SID
                    IR
  Focused Linear Grid:
72”
SID
                   IR
        Grid Cap:
Front               Back
Grid Cap: Magnified
Front
        Grid Cap: Magnified
           Ratio: 8:1
           Focus: 40 to 72
           LPI: 132
The LPI range for digital IRs is between103 and 200
           with 140 being the average.
Grid Cut-Off:
    Patient Factors to Consider:
   Additive Diseases
      Require a 50% increase in mAs or a 7.5%
      increase in kVp to maintain RE
Additive Disease: Ascites
    Normal          Ascites
    Patient Factors to Consider:
   Destructive Diseases
      Require a 30% decrease in mAs or a 5%
      decrease in kVp to maintain RE
Destructive Disease: Osteoporosis
       Normal          Osteoporosis
Patient Factors to Consider: Casts
   Fiberglass: increase mAs by 30% or increase
    kVp by 5%
Patient Factors to Consider: Casts
                Contrast Agents
    Patient Factors to Consider:
   Body Habitus
      Hypersthenic
      Sthenic
      Hyposthenic
      Asthenic
The Contrast Unit
             Subject Contrast:
   The magnitude of the signal difference in the
    remnant beam as a result of the different
    absorption characteristics of the tissues and
    structures making up that part.
   The difference in the thickness and atomic
    numbers of the structures that comprise the body
    part of interest.
   kVp is the primary controlling factor for subject
    contrast.
        This is the basis for optimum kVp.
Image (Radiographic) Contrast:
   The visible difference between any two selected
    brightness levels within a displayed image.
   It is determined by algorithms that are applied to
    raw data collected during image capture.
Image (Radiographic) Contrast:
   Grayscale
       The number of brightness levels or shades of
       gray visible on an image.
       Linked to the bit depth of the system which a
       reference to the total number of shades of gray
       available.
       How would noise affect grayscale?
Image (Radiographic) Contrast:
   Short Scale or High Contrast
       Major differences between shades of gray.
   Long Scale or Low Contrast
       Slight differences between shades of gray.
   kVp dictates the starting point for image contrast.
       All other factors that impact image contrast
       will cause it to increase or decrease from here.
Image (Radiographic) Contrast:
                         Note the
                       relationship
                      between kVp
                       and subject
                         contrast.
Patient
                          IR
           Image
      Tight Collimation
Collimation vs. Contrast:
Patient
                                 IR
        Radiographic Image
     * No Collimation *
Collimation vs. Contrast:
Patient
                                 IR
        Radiographic Image
     * No Collimation *
Collimation vs. Contrast:
             IR
Radiographic Image
    *                    *
        No Collimation
Collimation vs. Contrast:
                  IR
Radiographic Image
       *
Bottom Line:
                              *
               Lack of collimation adds
          No Collimation
  unwanted scatter (noise) to the IR.
Collimation vs. Contrast: Konica CR
 24 x 30 cm Collimation          No Collimation
S = 142 S = 103
                        30          69.5
                                    69       90
                   X-ray Photon Energy (keV)
Filtration vs. Contrast: Konica CR
 2.5 mm   3.5 mm   4.5 mm   5.5 mm
 Filtration vs. Contrast: Konica CR
                   2.5 mm     5.5 mm
                             Patient
         What happens    *
         to contrast?
     Patient                  4” OID
 *
        1” OID
10 mAs at 75 kVp
                   IR 20 mAs at 75 kVp
OID vs. Contrast: Konica CR
   4” OID                     10” OID
          *
                              8:1 GR
                      *
                                 IR
Grids vs. Contrast: Konica CR
    6:1 Grid Ratio           15:1 Grid Radio
S = 259 S = 257
Voluntary      Involuntary
           Geometric Factors:
1.   Focal Spot (FS)
2.   Object-to-Image-receptor Distance (OID)
3.   Source-to-Image-receptor Distance (SID)
4.   Source-to-Object Distance (SOD)
           Geometric Factors:
   Focal Spot
      Primary controlling factor for SR.
      Only variable that exclusively affects SR.
      Generally, techniques > 200 mA will require a
      large focal spot.
      Penumbra vs. Umbra
--
     Focal Spot Size vs. SR:
     0.6 mm FS             2.0 mm FS
Patient
                   IR
--
     FS Size vs. SR: Konica CR
        Small FS      Large FS
           Geometric Factors:
   OID
      Of the geometric factors, this is the most
      critical factor to consider.
      What is the overall MOST critical factor to
      consider regarding SR?
--
                 OID vs. SR:
     2.0 mm FS                2.0 mm FS
                    Patient    4” OID
     1” OID
                      IR
          OID vs. SR: Konica CR
      1” OID           4” OID              8” OID
--
40” SID
Patient
                    IR
       SID vs. SR: Konica CR
   40” SID            60” SID         72” SID
                    x   40
                      =
                   20   36
                  36x = 800