High Voltage Cables


AX

Replacements of Parts
High Voltage Cables

Pirelli and Claymount

07462570
05759787
05759555
07759728
05760090
05759001
05658385
04819855
03848905
03828451
03828394
03827370
03827362
03774119
01170427
01167050

© Siemens AG 2005
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.

Print No.: AX98-000.841.01.01.02 English


Replaces: RA98-031.031.01.07 Doc. Gen. Date: 09.05
2 Revision / Disclaimer
1Revision / Disclaimer

Document revision level


The document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.

Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.

High Voltage Cables AX98-000.841.01.01.02 Page 2 of 38 Siemens AG


09.05 CS PS 211 Medical Solutions
Table of Contents 3
1- 0Table of Contents

1 _______ General Information______________________________________________ 5

General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Establishing the HV Plug-in Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
General Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

2 _______ Claymount High Voltage Cables with Euroflex Hose ___________________ 7

High Voltage Cables, Model G5474 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7


Replacement: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Recognition Features:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Items Included in Shipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Connecting the High Voltage Connectors to the X-ray Tube Unit. . . . . . . . . . . . . . . . 9
Connecting the High Voltage Connectors to the High Voltage Transformer. . . . . . . 11

3 _______ Claymount High Voltage Cable, Locaflex, 3-pole _____________________ 12

Claymount High Voltage Cable, Model X2194 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12


Replacement: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Recognition Features:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Items included in the Shipment: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Connecting the High Voltage Connectors to the X-ray Tube Unit. . . . . . . . . . . . . . . 14
Connecting the High Voltage Connectors to the High Voltage Transformer. . . . . . . 16

4 _______ Claymount High Voltage Cable, Superflex, 3-pole ____________________ 17

High Voltage Cables, Model X2195 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17


Replacement: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Recognition Features:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Items included in the Shipment: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Connecting the High Voltage Connectors to the X-ray Tube Unit. . . . . . . . . . . . . . . 19
Connecting the High Voltage Connectors to the High Voltage Transformer. . . . . . . 21

5 _______ Claymount High Voltage Cable, Superflex, 4-pole ____________________ 22

High Voltage Cables, Model X2199 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22


Replacement: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Recognition Features:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Items included in the Shipment: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Connecting the High Voltage Connectors to the X-ray Tube Unit. . . . . . . . . . . . . . . 24
Connecting the High Voltage Connectors to the High Voltage Transformer. . . . . . . 26

6 _______ Pirelli High Voltage Cables with Euroflex Hose ______________________ 27

High Voltage Cables, Model G141E / G145E. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27


Replacement: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Recognition Features:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Items Included in Shipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Connecting the High Voltage Connectors to the X-ray Tube Unit. . . . . . . . . . . . . . . 29
Connecting the High Voltage Connectors to the High Voltage Transformer. . . . . . . 30

7 _______ Pirelli High Voltage Cables _______________________________________ 32

Siemens AG AX98-000.841.01.01.02 Page 3 of 38 High Voltage Cables


Medical Solutions 09.05 CS PS 211
4 Table of Contents

High Voltage Cables, Model RH098 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32


Replacement: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Recognition Features: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Items Included in Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Connecting the High Voltage Connectors to the X-ray Tube Unit . . . . . . . . . . . . . . . 35
Connecting the High Voltage Connectors to the High Voltage Transformer . . . . . . . 36

8 _______ Changes to Previous Version _____________________________________ 37

High Voltage Cables AX98-000.841.01.01.02 Page 4 of 38 Siemens AG


09.05 CS PS 211 Medical Solutions
General Information 5
2-
General Information
1General Information

Establishing the HV Plug-in Connections 0

To ensure on-going, problem-free function, the following points must be observed:


• Use only the materials provided and ensure that parts are not damaged
• Pay attention to the cleanliness of the connectors, sockets and insulations parts, aids
and tools.
• Do not touch the insulation parts with your bare hand; hand perspiration causes spikes.
• For cleaning, use only clean, soft and lint-free cloths.
• Remove stubborn soiling with undiluted ethyl alcohol and wipe the connectors so
they are dry.
• Examine the HV connectors and sockets for cracks and traces of overheating. Use only
undamaged insulation parts.
• Do not mix silicon oil and transformer oil.
• When exposed to the silicon oil, the silicon disk and the sealer ring swell slightly; when
this happens, this ensures that no moisture can get into the connection. If the plug-in
connector is removed after longer operation (more than 8 days), the silicon disk and the
sealer ring must be replaced.

NOTE If too much force is used, depending on the tolerances, it is still


possible to insert the connector into the socket turned 120° de-
spite the slot and nib.

Siemens AG AX98-000.841.01.01.02 Page 5 of 38 High Voltage Cables


Medical Solutions 09.05 CS PS 211
6 General Information

Safety Information 0

General Safety Information 0

WARNING Safety Information!


¹ When performing the work steps and test, the prod-
uct-specific safety information contained in the docu-
ments as well as the general safety information for
medical products must be observed.

High Voltage Cables AX98-000.841.01.01.02 Page 6 of 38 Siemens AG


09.05 CS PS 211 Medical Solutions
Claymount High Voltage Cables with Euroflex Hose 7
3-
High Voltage Cables, Model G5474
2Claymount High Voltage Cables with Euroflex Hose

Part No.: Designation Use


100 93 759 HV Cable, 3-pole, AXIOM Artis MP / dMP
20m
100 93 762 HV Cable, 4-pole, AXIOM Artis MP / dMP
20m
100 93 760 HV Cable, 3-pole, AXIOM Artis FA / FC / dFA / dFC / dFCM / BA
28m (Plane A) / BC (Plane A) / dBA (Plane A) / dBC
(Plane A)
100 93 763 HV Cable, 4-pole, AXIOM Artis FA / FC / dFA / dFC / dFCM / BA
28m (Plane A) / BC (Plane A) / dBA (Plane A) / dBC
(Plane A)
100 93 761 HV Cable, 3-pole, AXIOM Artis TA / TC / dTA / dTC / BA (Plane B) /
30m BC (Plane B) / dBA (Plane B) / dBC (Plane B)
100 93 764 HV Cable, 4-pole, AXIOM Artis TA / TC / dTA / dTC / BA (Plane B) /
30m BC (Plane B) / dBA (Plane B) / dBC (Plane B)

Replacement: 0

n.a.

Recognition Features: 0

• High voltage cable color = cream white


• High voltage cables identified with the LOCAFLEX printing.
• Cable diameter of the 3- and 4-pole high voltage cables = 16 mm
• High Voltage Cables covered with the black Euroflex hose

Siemens AG AX98-000.841.01.01.02 Page 7 of 38 High Voltage Cables


Medical Solutions 09.05 CS PS 211
8 Claymount High Voltage Cables with Euroflex Hose

Items Included in Shipment 0

Fig. 2: Claymount accessory pack,


Fig. 1: Claymount accessory pack, 3-pole 4-pole
Pos. 1 Mounting ring
Pos. 1 Mounting ring
Pos. 2 Contact ring
Pos. 2 Contact ring
Pos. 3 Sealer ring
Pos. 3 Sealer ring
Pos. 4 Silicon disk, 3-pole
Pos. 4 Silicon disk, 4-pole
Pos. 5 Silicon oil
Pos. 5 Silicon oil

Fig. 4: Claymount angled cuff, single


parts
Pos. 1 Angled cuff half cowl
Pos. 2 Snap ring
Fig. 3: Claymount angled cuff, complete

NOTE When ordering a 3-pole high voltage cable, the 3-pole accessory
pack, an angled cuff as well as a 20 m Euroflex hose are included
in the shipment.
When ordering a 4-pole high voltage cable, the 4-pole accessory
pack, an angled cuff as well as a 20 m Euroflex hose are included
in the shipment.

High Voltage Cables AX98-000.841.01.01.02 Page 8 of 38 Siemens AG


09.05 CS PS 211 Medical Solutions
Claymount High Voltage Cables with Euroflex Hose 9

Connecting the High Voltage Connectors to the X-ray Tube Unit 0

Fig. 5: Claymount high voltage connectors


Pos. 1 High voltage connectors
Pos. 2 Sealer ring
Pos. 3 Contact ring
Pos. 4 Mounting ring
Pos. 5 Contact pin (replaceable
Pos. 6 Contact flange (part of the high voltage connector)
Pos. 7 Silicon disk
Pos. 8 Safety screw for mounting ring
Pos. 9 Guide nipple on high voltage connector

Fig. 7: Claymount silicon disk, Part No.:


Fig. 6: Claymount silicon disk, Part No.: 10093773
10093772

Siemens AG AX98-000.841.01.01.02 Page 9 of 38 High Voltage Cables


Medical Solutions 09.05 CS PS 211
10 Claymount High Voltage Cables with Euroflex Hose

Fig. 9: Claymount angled cuff, installed


Pos. 1 Angled cuff half cowls
Pos. 2 Mounting ring
Pos. 3 Snap ring

Fig. 8: Claymount angled cuff, installation

• Turn out the safety screw (8/Fig. 5 / p. 9) from the mounting ring until it is no longer pro-
jecting out on the inside of the mounting ring.
• Unscrew the parts of the contact ring (3/Fig. 5 / p. 9).
• Slide the mounting ring (4/Fig. 5 / p. 9) over the high voltage connector.
• Place the halves of the contact ring over the contact flange (6/Fig. 5 / p. 9) and thread
them together.

NOTE If the angled cuff (Fig. 9 / p. 10) is installed, assembly of the con-
tact ring (3/Fig. 5 / p. 9) is skipped.
• Remove the angled cuff by removing the snap ring
(2/Fig. 4 / p. 8).
• Place the two angled cuff half cowls on the contact flange of the
high voltage connector (Fig. 8 / p. 10).
• Secure the two angled cuff half cowls with the snap ring
(3/Fig. 9 / p. 10).

• Slide the sealer ring (2/Fig. 5 / p. 9) over the high voltage connector until it is up against
the contact ring.
• Evenly apply silicon oil using only the tip of the tube on the front of the connector. Then
place on the silicon disk (Fig. 6 / p. 9) and moisten it with silicon oil.

High Voltage Cables AX98-000.841.01.01.02 Page 10 of 38 Siemens AG


09.05 CS PS 211 Medical Solutions
Claymount High Voltage Cables with Euroflex Hose 11

NOTE For AXIOM Artis systems with a:


• MEGALIX CAT 125/35/80-121GW, Part No.: 5764506
• OPTITOP 150/40/80HC-100-4, Part No.: 4802216
there must be a short circuit jumper (Fig. 10 / p. 11) on the cathode
side high voltage cable between the grid (G) and cathode (0) con-
nection pins.

Fig. 10: Short-circuit jumper for 2-focus tube


• Line up the guide nib on the high voltage connector with the slot on the socket and care-
fully insert the connector into the socket. When this is done, the air must be able to es-
cape the socket.
• Hand-tighten the mounting ring on the cable sleeve.
• After about one hour, check the tightness of the mounting ring.
• Secure the mounting ring with the safety screw.

Connecting the High Voltage Connectors to the High Voltage Transformer 0

• Turn out the safety screw from the mounting ring until it is no longer projecting out on
the inside of the mounting ring.
• Unscrew the parts of the contact ring (3/Fig. 5 / p. 9).
• Slide the mounting ring (4/Fig. 5 / p. 9) over the high voltage connector.
• Place the halves of the contact ring over the contact flange (6/Fig. 5 / p. 9) and thread
them together.
• Installation of the High Voltage Connector, depending on the Generator:
- With POLYDOROS IS A(F) open:
Check the oil level in the socket; if needed, add oil. Minimum oil level without the
connector inserted = 5 mm or 5 ml. The high voltage connector is inserted without a
silicon disk.
- With POLYDOROS A100:
Evenly apply silicon oil using only the tip of the tube on the front of the connector.
Then place on the silicon disk and moisten it with silicon oil.
• Line up the guide nib on the high voltage connector with the slot on the socket and care-
fully insert the connector into the socket.
• Hand-tighten the mounting ring.
• Secure the mounting ring with the safety screw.

Siemens AG AX98-000.841.01.01.02 Page 11 of 38 High Voltage Cables


Medical Solutions 09.05 CS PS 211
12 Claymount High Voltage Cable, Locaflex, 3-pole
4-
Claymount High Voltage Cable, Model X2194
3Claymount High Voltage Cable, Locaflex, 3-pole

Part No.: Designation Use


30 65 500 HV Cable, 3-pole, With the following generators:
4m • POLYDOROS 80A
30 65 518 HV Cable, 3-pole, • POLYDOROS 100A
6m
• POLYDOROS IS-A on workstation 1 with 3-pole
30 65 526 HV Cable, 3-pole, cathode-side high voltage cables
8m • POLYDOROS IS-AF on workstation 1 with
30 65 534 HV Cable, 3-pole, 3-pole cathode-side high voltage cables
10m
30 65 542 HV Cable, 3-pole,
12m
30 65 559 HV Cable, 3-pole,
14m
30 65 567 HV Cable, 3-pole,
16m
30 65 575 HV Cable, 3-pole,
18m
30 65 583 HV Cable, 3-pole,
20m
30 65 609 HV Cable, 3-pole,
24m
30 65 625 HV Cable, 3-pole,
28m

NOTE If a 4-pole high voltage cable, model X2195 is used for connection
of grid-controlled or 3-focus X-ray tubes (4-pole cathode-side
socket) with the POLYDOROS IS-A(F), the 3-pole Superflex high
voltage cable, model X2199 must be used on the anode side.

Replacement: 0

Pirelli High Voltage Cables (Pirelli High Voltage Cables / p. 32)

NOTE Anode-side and cathode-side high voltage cables must always be


from the same manufacturer. In other words, if a Claymount high
voltage cable is replaced with a Pirelli high voltage cable, both
high voltage cables have to be replaced.

High Voltage Cables AX98-000.841.01.01.02 Page 12 of 38 Siemens AG


09.05 CS PS 211 Medical Solutions
Claymount High Voltage Cable, Locaflex, 3-pole 13

Recognition Features: 0

• High voltage cable color = cream white


• High voltage cables identified with the LOCAFLEX printing.
• Cable diameter of the high voltage cable = 16 mm

Items included in the Shipment: 0

When ordering a 3-pole high voltage cable, the 3-pole accessory pack is included.

Fig. 11: Claymount accessory pack, 3-pole


Pos. 1 Mounting ring
Pos. 2 Contact ring
Pos. 3 Sealer ring
Pos. 4 Silicon disk, 3-pole
Pos. 5 Silicon oil

Siemens AG AX98-000.841.01.01.02 Page 13 of 38 High Voltage Cables


Medical Solutions 09.05 CS PS 211
14 Claymount High Voltage Cable, Locaflex, 3-pole

Connecting the High Voltage Connectors to the X-ray Tube Unit 0

Fig. 12: Claymount high voltage connectors


Pos. 1 High voltage connectors
Pos. 2 Sealer ring
Pos. 3 Contact ring
Pos. 4 Mounting ring
Pos. 5 Contact pin (replaceable
Pos. 6 Contact flange (part of the high voltage connector)
Pos. 7 Silicon disk
Pos. 8 Safety screw for mounting ring
Pos. 9 Guide nipple on high voltage connector

High Voltage Cables AX98-000.841.01.01.02 Page 14 of 38 Siemens AG


09.05 CS PS 211 Medical Solutions
Claymount High Voltage Cable, Locaflex, 3-pole 15

Fig. 14: Claymount silicon disk, Part No.:


Fig. 13: Claymount contact pin, old 3065872

Fig. 16: Claymount silicon disk, Part No.:


Fig. 15: Claymount contact pin, new 10093772

• Turn out the safety screw (8/Fig. 12 / p. 14) from the mounting ring until it is no longer
projecting out on the inside of the mounting ring.
• Unscrew the parts of the contact ring (3/Fig. 12 / p. 14).
• Slide the mounting ring (4/Fig. 12 / p. 14) over the high voltage connector.
• Place the halves of the contact ring over the contact flange (6/Fig. 12 / p. 14) and
thread them together.
• Slide the sealer ring (2/Fig. 12 / p. 14) over the high voltage connector until it is up
against the contact ring.
• Evenly apply silicon oil using only the tip of the tube on the front of the connector. Then
place on the silicon disk and moisten it with silicon oil.

NOTE Two types of contact pins are used in the Claymount high voltage
cables:
• With the cone-shaped, bronze-colored contact pin
(Fig. 13 / p. 15), the silicon disk with Part No.: 3065872
(Fig. 14 / p. 15) is used
• With the straight, silver-colored contact pin (Fig. 15 / p. 15), the
silicon disk with Part No.: 10093772 (Fig. 16 / p. 15) is used

Siemens AG AX98-000.841.01.01.02 Page 15 of 38 High Voltage Cables


Medical Solutions 09.05 CS PS 211
16 Claymount High Voltage Cable, Locaflex, 3-pole

• Line up the guide nib on the high voltage connector with the slot on the socket and care-
fully insert the connector into the socket. When this is done, the air must be able to es-
cape the socket.
• Hand-tighten the mounting ring on the cable sleeve.
• After about one hour, check the tightness of the mounting ring.
• Secure the mounting ring with the safety screw.

Connecting the High Voltage Connectors to the High Voltage Transformer 0

• Turn out the safety screw from the mounting ring until it is no longer projecting out on
the inside of the mounting ring.
• Unscrew the parts of the contact ring (3/Fig. 12 / p. 14).
• Slide the mounting ring (4/Fig. 12 / p. 14) over the high voltage connector.
• Place the halves of the contact ring over the contact flange (6/Fig. 12 / p. 14) and
thread them together.
• Check the oil level in the socket; if needed, add oil. Minimum oil level without the con-
nector inserted = 5 mm or 5 ml. The high voltage connector is inserted without a silicon
disk.
• Line up the guide nib on the high voltage connector with the slot on the socket and care-
fully insert the connector into the socket.
• Hand-tighten the mounting ring.
• Secure the mounting ring with the safety screw.

High Voltage Cables AX98-000.841.01.01.02 Page 16 of 38 Siemens AG


09.05 CS PS 211 Medical Solutions
Claymount High Voltage Cable, Superflex, 3-pole 17
5-
High Voltage Cables, Model X2195
4Claymount High Voltage Cable, Superflex, 3-pole

Part No.: Designation Use


30 65 807 HV Cable, 3-pole, With the following generators:
24m • POLYDOROS IS-C
30 65 815 HV Cable, 3-pole, • POLYDOROS IS-CF
26m
• POLYDOROS IS-A on workstation 2 with 4-pole
30 65 823 HV Cable, 3-pole, cathode-side high voltage cables
28m • POLYDOROS IS-AF on workstation 2 with 4-pole
cathode-side high voltage cables

NOTE If a 4-pole high voltage cable, model X2195 is used for connection
of grid-controlled or 3-focus X-ray tubes (4-pole cathode-side
socket) with the POLYDOROS IS-A(F) / IS-C(F), the 3-pole Super-
flex high voltage cable, model X2199 must be used on the anode
side.

Replacement: 0

Pirelli High Voltage Cables (Pirelli High Voltage Cables / p. 32)

NOTE Anode-side and cathode-side high voltage cables must always be


from the same manufacturer. In other words, if a Claymount high
voltage cable is replaced with a Pirelli high voltage cable, both
high voltage cables have to be replaced.

Recognition Features: 0

• High voltage cable color = cream white


• Cable diameter of the high voltage cable = 16 mm

Items included in the Shipment: 0

When ordering a 3-pole high voltage cable, the 3-pole accessory pack is included.

Siemens AG AX98-000.841.01.01.02 Page 17 of 38 High Voltage Cables


Medical Solutions 09.05 CS PS 211
18 Claymount High Voltage Cable, Superflex, 3-pole

Fig. 17: Claymount accessory pack, 3-pole


Pos. 1 Mounting ring
Pos. 2 Contact ring
Pos. 3 Sealer ring
Pos. 4 Silicon disk, 3-pole
Pos. 5 Silicon oil

High Voltage Cables AX98-000.841.01.01.02 Page 18 of 38 Siemens AG


09.05 CS PS 211 Medical Solutions
Claymount High Voltage Cable, Superflex, 3-pole 19

Connecting the High Voltage Connectors to the X-ray Tube Unit 0

Fig. 18: Claymount high voltage connectors


Pos. 1 High voltage connectors
Pos. 2 Sealer ring
Pos. 3 Contact ring
Pos. 4 Mounting ring
Pos. 5 Contact pin (replaceable
Pos. 6 Contact flange (part of the high voltage connector)
Pos. 7 Silicon disk
Pos. 8 Safety screw for mounting ring
Pos. 9 Guide nipple on high voltage connector

Siemens AG AX98-000.841.01.01.02 Page 19 of 38 High Voltage Cables


Medical Solutions 09.05 CS PS 211
20 Claymount High Voltage Cable, Superflex, 3-pole

Fig. 20: Claymount silicon disk, Part No.:


Fig. 19: Claymount contact pin, old 3065872

Fig. 22: Claymount silicon disk, Part No.:


Fig. 21: Claymount contact pin, new 10093772

• Turn out the safety screw (8/Fig. 18 / p. 19) from the mounting ring until it is no longer
projecting out on the inside of the mounting ring.
• Unscrew the parts of the contact ring (3/Fig. 18 / p. 19).
• Slide the mounting ring (4/Fig. 18 / p. 19) over the high voltage connector.
• Place the halves of the contact ring over the contact flange (6/Fig. 18 / p. 19) and
thread them together.
• Slide the sealer ring (2/Fig. 18 / p. 19) over the high voltage connector until it is up
against the contact ring.
• Evenly apply silicon oil using only the tip of the tube on the front of the connector. Then
place on the silicon disk and moisten it with silicon oil.

NOTE Two types of contact pins are used in the Claymount high voltage
cables:
• With the cone-shaped, bronze-colored contact pin
(Fig. 19 / p. 20), the silicon disk with Part No.: 3065872
(Fig. 20 / p. 20) is used
• With the straight, silver-colored contact pin (Fig. 21 / p. 20), the
silicon disk with Part No.: 10093772 (Fig. 22 / p. 20) is used

High Voltage Cables AX98-000.841.01.01.02 Page 20 of 38 Siemens AG


09.05 CS PS 211 Medical Solutions
Claymount High Voltage Cable, Superflex, 3-pole 21

• Line up the guide nib on the high voltage connector with the slot on the socket and care-
fully insert the connector into the socket. When this is done, the air must be able to es-
cape the socket.
• Hand-tighten the mounting ring on the cable sleeve.
• After about one hour, check the tightness of the mounting ring.
• Secure the mounting ring with the safety screw.

Connecting the High Voltage Connectors to the High Voltage Transformer 0

• Turn out the safety screw from the mounting ring until it is no longer projecting out on
the inside of the mounting ring.
• Unscrew the parts of the contact ring (3/Fig. 18 / p. 19).
• Slide the mounting ring (4/Fig. 18 / p. 19) over the high voltage connector.
• Place the halves of the contact ring over the contact flange (6/Fig. 18 / p. 19) and
thread them together.
• Check the oil level in the socket; if needed, add oil. Minimum oil level without the con-
nector inserted = 5 mm or 5 ml. The high voltage connector is inserted without a silicon
disk.
• Line up the guide nib on the high voltage connector with the slot on the socket and care-
fully insert the connector into the socket.
• Hand-tighten the mounting ring.
• Secure the mounting ring with the safety screw.

Siemens AG AX98-000.841.01.01.02 Page 21 of 38 High Voltage Cables


Medical Solutions 09.05 CS PS 211
22 Claymount High Voltage Cable, Superflex, 4-pole
6-
High Voltage Cables, Model X2199
5Claymount High Voltage Cable, Superflex, 4-pole

Part No.: Designation Use


30 65 658 HV Cable, 4-pole, With the following generators:
6m • POLYDOROS IS-C
30 65 666 HV Cable, 4-pole, • POLYDOROS IS-CF
8m
• POLYDOROS IS-A on workstation 2 with 4-pole
30 65 674 HV Cable, 4-pole, cathode-side high voltage cables
10m • POLYDOROS IS-AF on workstation 2 with 4-pole
30 65 682 HV Cable, 4-pole, cathode-side high voltage cables
12m • POLYDOROS 80A with 3-focus X-ray tubes
30 65 690 HV Cable, 4-pole, • POLYDOROS 100A with 3-focus X-ray tubes
14m
30 65 708 HV Cable, 4-pole,
16m
30 65 724 HV Cable, 4-pole,
20m
30 65 740 HV Cable, 4-pole,
24m
30 65 757 HV Cable, 4-pole,
26m
30 65 765 HV Cable, 4-pole,
28m

NOTE If a 4-pole high voltage cable, model X2195 is used for connection
of grid-controlled or 3-focus X-ray tubes (4-pole cathode-side
socket) with the POLYDOROS IS-A(F) / IS-C(F), the 3-pole Super-
flex high voltage cable, model X2199 must be used on the anode
side.

Replacement: 0

Pirelli High Voltage Cables (Pirelli High Voltage Cables / p. 32)

NOTE Anode-side and cathode-side high voltage cables must always be


from the same manufacturer. In other words, if a Claymount high
voltage cable is replaced with a Pirelli high voltage cable, both
high voltage cables have to be replaced.

High Voltage Cables AX98-000.841.01.01.02 Page 22 of 38 Siemens AG


09.05 CS PS 211 Medical Solutions
Claymount High Voltage Cable, Superflex, 4-pole 23

Recognition Features: 0

• High voltage cable color = cream white


• Cable diameter of the high voltage cable = 19mm

Items included in the Shipment: 0

When ordering a 4-pole high voltage cable, the 4-pole accessory pack is included.

Fig. 23: Claymount accessory pack, 4-pole


Pos. 1 Mounting ring
Pos. 2 Contact ring
Pos. 3 Sealer ring
Pos. 4 Silicon disk, 4-pole
Pos. 5 Silicon oil

Siemens AG AX98-000.841.01.01.02 Page 23 of 38 High Voltage Cables


Medical Solutions 09.05 CS PS 211
24 Claymount High Voltage Cable, Superflex, 4-pole

Connecting the High Voltage Connectors to the X-ray Tube Unit 0

Fig. 24: Claymount high voltage connectors


Pos. 1 High voltage connectors
Pos. 2 Sealer ring
Pos. 3 Contact ring
Pos. 4 Mounting ring
Pos. 5 Contact pin (replaceable
Pos. 6 Contact flange (part of the high voltage connector)
Pos. 7 Silicon disk
Pos. 8 Safety screw for mounting ring
Pos. 9 Guide nipple on high voltage connector

High Voltage Cables AX98-000.841.01.01.02 Page 24 of 38 Siemens AG


09.05 CS PS 211 Medical Solutions
Claymount High Voltage Cable, Superflex, 4-pole 25

Fig. 26: Claymount silicon disk, Part No.:


Fig. 25: Claymount contact pin, old 4775065

Fig. 28: Claymount silicon disk, Part No.:


Fig. 27: Claymount contact pin, new 10093773

• Turn out the safety screw (8/Fig. 24 / p. 24) from the mounting ring until it is no longer
projecting out on the inside of the mounting ring.
• Unscrew the parts of the contact ring (3/Fig. 24 / p. 24).
• Slide the mounting ring (4/Fig. 24 / p. 24) over the high voltage connector.
• Place the halves of the contact ring over the contact flange (6/Fig. 24 / p. 24) and
thread them together.
• Slide the sealer ring (2/Fig. 24 / p. 24) over the high voltage connector until it is up
against the contact ring.
• Evenly apply silicon oil using only the tip of the tube on the front of the connector. Then
place on the silicon disk and moisten it with silicon oil.

NOTE Two types of contact pins are used in the Claymount high voltage
cables:
• With the cone-shaped, bronze-colored contact pin
(Fig. 25 / p. 25), the silicon disk with Part No.: 3065872
(Fig. 20 / p. 20) is used
• With the straight, silver-colored contact pin (Fig. 27 / p. 25), the
silicon disk with Part No.: 10093772 (Fig. 22 / p. 20) is used

Siemens AG AX98-000.841.01.01.02 Page 25 of 38 High Voltage Cables


Medical Solutions 09.05 CS PS 211
26 Claymount High Voltage Cable, Superflex, 4-pole

• Line up the guide nib on the high voltage connector with the slot on the socket and care-
fully insert the connector into the socket. When this is done, the air must be able to es-
cape the socket.
• Hand-tighten the mounting ring on the cable sleeve.
• After about one hour, check the tightness of the mounting ring.
• Secure the mounting ring with the safety screw.

Connecting the High Voltage Connectors to the High Voltage Transformer 0

• Turn out the safety screw from the mounting ring until it is no longer projecting out on
the inside of the mounting ring.
• Unscrew the parts of the contact ring (3/Fig. 24 / p. 24).
• Slide the mounting ring (4/Fig. 24 / p. 24) over the high voltage connector.
• Place the halves of the contact ring over the contact flange (6/Fig. 24 / p. 24) and
thread them together.
• Check the oil level in the socket; if needed, add oil. Minimum oil level without the con-
nector inserted = 5 mm or 5 ml. The high voltage connector is inserted without a silicon
disk.
• Line up the guide nib on the high voltage connector with the slot on the socket and care-
fully insert the connector into the socket.
• Hand-tighten the mounting ring.
• Secure the mounting ring with the safety screw.

High Voltage Cables AX98-000.841.01.01.02 Page 26 of 38 Siemens AG


09.05 CS PS 211 Medical Solutions
Pirelli High Voltage Cables with Euroflex Hose 27
7-
High Voltage Cables, Model G141E / G145E
6Pirelli High Voltage Cables with Euroflex Hose

Part No.: Designation Use


65 89 506 HV Cable, 3-pole, AXIOM Artis MP / dMP
20m
65 89 514 HV Cable, 4-pole, AXIOM Artis MP / dMP
20m
65 55 978 HV Cable, 3-pole, AXIOM Artis FA / FC / dFA / dFC / dFCM / BA
28m (Plane A) / BC (Plane A) / dBA (Plane A) / dBC
(Plane A)
65 55 986 HV Cable, 4-pole, AXIOM Artis FA / FC / dFA / dFC / dFCM / BA
28m (Plane A) / BC (Plane A) / dBA (Plane A) / dBC
(Plane A)
65 89 639 HV Cable, 3-pole, AXIOM Artis TA / TC / dTA / dTC / BA (Plane B) /
30m BC (Plane B) / dBA (Plane B) / dBC (Plane B)
65 89 647 HV Cable, 4-pole, AXIOM Artis TA / TC / dTA / dTC / BA (Plane B) /
30m BC (Plane B) / dBA (Plane B) / dBC (Plane B)

Replacement: 0

Claymount High Voltage Cables with Euroflex Hose (Claymount High Voltage Cables with
Euroflex Hose / p. 7)

NOTE Mixed use of anode-side and cathode-side high voltage cables is


permitted. In other words, if a Pirelli high voltage cable is replaced
with a Claymount high voltage cable, both high voltage cables do
not have to be replaced.

Recognition Features: 0

• Color of the high voltage cable connectors = blue or black


• High voltage cables identified with the SIEMENS printing.
• Cable diameter of the 3- and 4-pole high voltage cables = 16 mm
• High Voltage Cables covered with the black Euroflex hose

Siemens AG AX98-000.841.01.01.02 Page 27 of 38 High Voltage Cables


Medical Solutions 09.05 CS PS 211
28 Pirelli High Voltage Cables with Euroflex Hose

Items Included in Shipment 0

Fig. 29: Pirelli accessory pack, 3-pole Fig. 30: Pirelli accessory pack,
Pos. 1 Mounting ring 4-pole
Pos. 2 Set screws Pos. 1 Mounting ring
Pos. 3 Contact ring Pos. 2 Set screws
Pos. 4 Sealer ring Pos. 3 Contact ring
Pos. 5 Silicon disk, 3-pole Pos. 4 Sealer ring
Pos. 6 Silicon oil Pos. 5 Silicon disk, 4-pole
Pos. 6 Silicon oil

NOTE When ordering a 3-pole high voltage cable, two 3-pole accessory
packs are included.
When ordering a 4-pole high voltage cable, two 4-pole accessory
packs are included.

High Voltage Cables AX98-000.841.01.01.02 Page 28 of 38 Siemens AG


09.05 CS PS 211 Medical Solutions
Pirelli High Voltage Cables with Euroflex Hose 29

Connecting the High Voltage Connectors to the X-ray Tube Unit 0

Fig. 31: Pirelli high voltage connectors, installation


Pos. 1 Mounting ring
Pos. 2 Contact ring
Pos. 3 Sealer ring
Pos. 4 Silicon disk
Pos. 5 Guide nipple on high voltage connector

Fig. 33: Pirelli silicon disk, Part No.:


1786516
Fig. 32: Pirelli silicon disk, Part No.: 1786508

• Turn out the two safety screws (2/Fig. 29 / p. 28) from the mounting ring until they are
no longer projecting out on the inside of the mounting ring.
• Slide the mounting ring (1/Fig. 31 / p. 29) over the connector, threads facing the con-
nector.
• Place the contact ring half cowls (2/Fig. 31 / p. 29) over the flange with the slot close to
stop.
• Evenly tighten the screws on the contact ring half cowls.

Siemens AG AX98-000.841.01.01.02 Page 29 of 38 High Voltage Cables


Medical Solutions 09.05 CS PS 211
30 Pirelli High Voltage Cables with Euroflex Hose

• Pull the sealer ring (3/Fig. 31 / p. 29) over the connector up to the flange.
• Evenly apply silicon oil using only the tip of the tube on the front of the connector. Then
place on the silicon disk (4/Fig. 31 / p. 29) and moisten it with silicon oil.

NOTE For AXIOM Artis systems with a:


• MEGALIX CAT 125/35/80-121GW, Part No.: 5764506
• OPTITOP 150/40/80HC-100-4, Part No.: 4802216
there must be a short circuit jumper (Fig. 34 / p. 30) on the cathode
side high voltage cable between the grid (G) and cathode (0) con-
nection pins.

Fig. 34: Short-circuit jumper for 2-focus tube


• Line up the guide nib on the high voltage connector (5/Fig. 31 / p. 29) with the slot on
the socket and carefully insert the connector into the socket.
• Hand-tighten the mounting ring; when doing this, hold the high voltage cable so it does
not twist.
• Tighten both set screws on the mounting ring until there is slight resistance (securing
against turning).

Connecting the High Voltage Connectors to the High Voltage Transformer 0

• Turn out the two safety screws (2/Fig. 29 / p. 28) from the mounting ring until they are
no longer projecting out on the inside of the mounting ring.
• Slide the mounting ring (1/Fig. 31 / p. 29) over the connector, threads facing the con-
nector.
• Place the contact ring half cowls (2/Fig. 31 / p. 29) over the flange with the slot close to
stop.
• Evenly tighten the screws on the contact ring half cowls.
• Installation of the High Voltage Connector, depending on the Generator:
- With POLYDOROS A100:
Evenly apply silicon oil using only the tip of the tube on the front of the connector.
Then place on the silicon disk and moisten it with silicon oil.
- With POLYDOROS IS A(F) open:
Check the oil level in the socket; if needed, add oil. Minimum oil level without the
connector inserted = 5 mm or 5 ml. The high voltage connector is inserted without a
silicon disk.

High Voltage Cables AX98-000.841.01.01.02 Page 30 of 38 Siemens AG


09.05 CS PS 211 Medical Solutions
Pirelli High Voltage Cables with Euroflex Hose 31

• Line up the guide nib on the high voltage connector (5/Fig. 31 / p. 29) with the slot on
the socket and carefully insert the connector into the socket.
• Hand-tighten the mounting ring; when doing this, hold the high voltage cable so it does
not twist.
• Tighten both set screws on the mounting ring until there is slight resistance (securing
against turning).

Siemens AG AX98-000.841.01.01.02 Page 31 of 38 High Voltage Cables


Medical Solutions 09.05 CS PS 211
32 Pirelli High Voltage Cables
8-
High Voltage Cables, Model RH098
7Pirelli High Voltage Cables

Part No.: Designation Use


44 66 389 HV Cable, 3-pole, With the following generators:
4m • POLYDOROS 80A
44 66 397 HV Cable, 3-pole, • POLYDOROS 100A
6m
• POLYDOROS IS-A
44 66 405 HV Cable, 3-pole, • POLYDOROS IS-AF
8m
• POLYDOROS IS-C
44 66 413 HV Cable, 3-pole,
• POLYDOROS IS-CF
10m
• POLYDOROS SX50/80
44 66 421 HV Cable, 3-pole,
12m • POLYDOROS SX65/80
• POLYDOROS LX30/50
44 66 439 HV Cable, 3-pole,
14m • POLYDOROS LX30/50 Lite
• POLYDOROS LX 80
44 66 447 HV Cable, 3-pole,
16m • POLYDOROS IT
44 66 454 HV Cable, 3-pole,
18m
44 66 462 HV Cable, 3-pole,
20m
44 66 470 HV Cable, 3-pole,
22m
44 66 488 HV Cable, 3-pole,
24m
44 66 496 HV Cable, 3-pole,
26m
58 90 194 HV Cable, 3-pole,
28m
58 90 186 HV Cable, 3-pole,
30m

High Voltage Cables AX98-000.841.01.01.02 Page 32 of 38 Siemens AG


09.05 CS PS 211 Medical Solutions
Pirelli High Voltage Cables 33

Part No.: Designation Use


58 90 160 HV Cable, 4-pole, With the following generators:
16m • POLYDOROS IS-C and IS-CF
58 90 152 HV Cable, 4-pole, • POLYDOROS IS-A and IS-AF on workstation 2 with
20m 4-pole cathode-side high voltage cables
58 90 145 HV Cable, 4-pole, • POLYDOROS 80A and 100A with 3-focus X-ray
24m tubes

58 90 137 HV Cable, 4-pole,


26m
58 90 129 HV Cable, 4-pole,
28m
58 90 103 HV Cable, 4-pole,
30m

NOTE When used with the POLYDOROS IS-A(F) / IS-C(F), the D45 board
(Part No.: 2797962) must have Rev. 02 or higher. This is necessary
to ensure adaptation of the capacitance to the high voltage cable
Model RH098.

Replacement: 0

n.a.

Recognition Features: 0

• Color of the high voltage cable connectors = blue or black


• High voltage cables identified with the SIEMENS printing.
• Cable diameter of the 3- and 4-pole high voltage cables = 16 mm

Siemens AG AX98-000.841.01.01.02 Page 33 of 38 High Voltage Cables


Medical Solutions 09.05 CS PS 211
34 Pirelli High Voltage Cables

Items Included in Shipment 0

Fig. 35: Pirelli accessory pack, 3-pole Fig. 36: Pirelli accessory pack,
Pos. 1 Mounting ring 4-pole
Pos. 2 Set screws Pos. 1 Mounting ring
Pos. 3 Contact ring Pos. 2 Set screws
Pos. 4 Sealer ring Pos. 3 Contact ring
Pos. 5 Silicon disk, 3-pole Pos. 4 Sealer ring
Pos. 6 Silicon oil Pos. 5 Silicon disk, 4-pole
Pos. 6 Silicon oil

NOTE When ordering a 3-pole high voltage cable, two 3-pole accessory
packs are included.
When ordering a 4-pole high voltage cable, two 4-pole accessory
packs are included.

High Voltage Cables AX98-000.841.01.01.02 Page 34 of 38 Siemens AG


09.05 CS PS 211 Medical Solutions
Pirelli High Voltage Cables 35

Connecting the High Voltage Connectors to the X-ray Tube Unit 0

Fig. 37: Pirelli high voltage connectors, installation


Pos. 1 Mounting ring
Pos. 2 Contact ring
Pos. 3 Sealer ring
Pos. 4 Silicon disk
Pos. 5 Guide nipple on high voltage connector

Fig. 39: Pirelli silicon disk, Part No.:


1786516
Fig. 38: Pirelli silicon disk, Part No.: 1786508

• Turn out the two safety screws (2/Fig. 35 / p. 34) from the mounting ring until they are
no longer projecting out on the inside of the mounting ring.
• Slide the mounting ring (1/Fig. 37 / p. 35) over the connector, threads facing the con-
nector.
• Place the contact ring half cowls (2/Fig. 37 / p. 35) over the flange with the slot close to
stop.
• Evenly tighten the screws on the contact ring half cowls.

Siemens AG AX98-000.841.01.01.02 Page 35 of 38 High Voltage Cables


Medical Solutions 09.05 CS PS 211
36 Pirelli High Voltage Cables

• Pull the sealer ring (3/Fig. 37 / p. 35) over the connector up to the flange.
• Evenly apply silicon oil using only the tip of the tube on the front of the connector. Then
place on the silicon disk (4/Fig. 37 / p. 35) and moisten it with silicon oil.
• Line up the guide nib on the high voltage connector (5/Fig. 37 / p. 35) with the slot on
the socket and carefully insert the connector into the socket.
• Hand-tighten the mounting ring; when doing this, hold the high voltage cable so it does
not twist.
• Tighten both set screws on the mounting ring until there is slight resistance (securing
against turning).

Connecting the High Voltage Connectors to the High Voltage Transformer 0

• Turn out the two safety screws (2/Fig. 35 / p. 34) from the mounting ring until they are
no longer projecting out on the inside of the mounting ring.
• Slide the mounting ring (1/Fig. 37 / p. 35) over the connector, threads facing the con-
nector.
• Place the contact ring half cowls (2/Fig. 37 / p. 35) over the flange with the slot close to
stop.
• Evenly tighten the screws on the contact ring half cowls.
• Installation of the High Voltage Connector, depending on the Generator:
- With POLYDOROS IT:
Evenly apply silicon oil using only the tip of the tube on the front of the connector.
Then place on the silicon disk and moisten it with silicon oil.
- With all other generators:
Check the oil level in the socket; if needed, add oil. Minimum oil level without the
connector inserted = 5 mm or 5 ml. The high voltage connector is inserted without a
silicon disk.
• Line up the guide nib on the high voltage connector (5/Fig. 37 / p. 35) with the slot on
the socket and carefully insert the connector into the socket.
• Hand-tighten the mounting ring; when doing this, hold the high voltage cable so it does
not twist.
• Tighten both set screws on the mounting ring until there is slight resistance (securing
against turning).

High Voltage Cables AX98-000.841.01.01.02 Page 36 of 38 Siemens AG


09.05 CS PS 211 Medical Solutions
Changes to Previous Version 37
9-
8Changes to Previous Version

None; initial publication.

Siemens AG AX98-000.841.01.01.02 Page 37 of 38 High Voltage Cables


Medical Solutions 09.05 CS PS 211
38 Changes to Previous Version

High Voltage Cables AX98-000.841.01.01.02 Page 38 of 38 Siemens AG


09.05 CS PS 211 Medical Solutions