X-Ray Diagnostic System

AX

Installation Instructions
Titel der
High voltage
Unterlage
cable hier
RH 098-5DF6
eintragen 071

© Siem ens A G 1 996


The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.

Register 3 English
Print No.: RX0-000.031.01.03.02 Doc. Gen. Date: 06.97
Replaces: RX0-000.031.01.02.02
0-2 Revision

Chapter Page Revision


0 all 02
1 all 02
2 all 01
0 all 03
1 all 03
2 all 02

X-Ray Diagnostic System Register 3 RX0-000.031.01 Page 2 of 4 Siemens AG


Rev. 03 06.97 TD AX 1 Medical Engineering
Contents 0-3

Page
1 _______High-voltage cable RH 098-5DF6 071 ______________________________ 1 - 1

Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 - 1
Validity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Insulating material permissible . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Making HV plug connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Inserting the HV plug into the tube assembly . . . . . . . . . . . . . . . . . . . . . .1-5
Insertion with corona disks . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Inserting the HV plug onto the HV generator . . . . . . . . . . . . . . . . . . . . . .1-6
Insertion with transformer oil. . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6

2 _______Changes to previous version _____________________________________ 2 - 1

Siemens AG Register 3 RX0-000.031.01 Page 3 of 4 X-Ray Diagnostic System


Medical Engineering Rev. 03 06.97 TD AX 1
0-4 Contents

This page intentionally left blank.

X-Ray Diagnostic System Register 3 RX0-000.031.01 Page 4 of 4 Siemens AG


Rev. 03 06.97 TD AX 1 Medical Engineering
High-voltage cable RH 098-5DF6 071 1 1-1

Prerequisites 1

Validity 1

These instructions are valid only for 3-core high-voltage cables RH 098 with the high-volt-
age conductor "SIEMENS 5DF6 071" in connection with the Siemens 0 receptacles.
Valid part numbers:
44 66 389 RH 098 4m
44 66 397 RH 098 6m
44 66 405 RH 098 8m
44 66 413 RH 098 10 m
44 66 421 RH 098 12 m
44 66 439 RH 098 14 m
44 66 447 RH 098 16 m
44 66 454 RH 098 18 m
44 66 462 RH 098 20 m
44 66 470 RH 098 22 m
44 66 488 RH 098 24 m
44 66 496 RH 098 26 m
44 66 371 RV 098 .. m-goods

Max. operating voltage = 75 kV DC


For the assembly is needed in addition:
1 allen-key 2 mm

Abbreviations 1

HV = high-voltage

Insulating material permissible 1

Corona disk, 3 pole 17 86 508 R4176


Seal 17 86 995 R4176
Silicone oil (AK 350) 17 87 035 R4176
Transformer oil accord.
to RA0-000.076.18.01..
e.g. Transformer oil 1l 44 10 023 RV090
Transformer oil 20 l 16 97 432 R1151
Cooling oil 20 l 73 84 860 R1151

The permissible oils are mineral oils.


They are rated as water polluting and shall be disposed
by a qualified waste disposal enterprise !

Siemens AG Register 3 RX0-000.031.01 Page 1 of 6 X-Ray Diagnostic System


Medical Engineering Rev. 03 06.97 TD AX 1
1-2 High-voltage cable RH 098-5DF6 071
Identification 1

Accessories pack

Fig. 1

11 Union
2 setscrews (M4 x 6)
12 2 half-shells
with screws
13 Gasket 17 86 995 R4176
14 Corona disk 17 86 508 R4176
15 Silicone oil (AK 350) 17 87 035 R4176

X-Ray Diagnostic System Register 3 RX0-000.031.01 Page 2 of 6 Siemens AG


Rev. 03 06.97 TD AX 1 Medical Engineering
High-voltage cable RH 098-5DF6 071 1-3
HV cable

Fig. 2

21 HV conductor
22 HV plug
23 Kink protection

HV plug

Fig. 3

31 Insulating piece
32 Nose
33 Plug bottom
34 Contact pins
35 Sealing surface
36 Plug flange

Siemens AG Register 3 RX0-000.031.01 Page 3 of 6 X-Ray Diagnostic System


Medical Engineering Rev. 03 06.97 TD AX 1
1-4 High-voltage cable RH 098-5DF6 071

Making HV plug connections 1

The following items have to be observed in order to


guarantee permanent faultless function:
• Observe the sequence of the work steps.
• Use only specified materials and undamaged parts.
• Make sure that the plugs, the receptacles and also the insulating materials,
auxiliary equipment and tools are clean.
• Do not touch the insulating pieces with your bare hands.
Sweat on the hands can cause spark-overs.
• For cleaning use only a clean and soft cloth free of fluff.
• Remove heavy dirt with undenatured ethyl alcohol and dry the plug connections.
• Check the HV plugs and receptacles for cracks and scorching.
Use intact insulating pieces only.
• Do not mix the silicone oil and the transformer oil.
• Use the "Corona disk" only if the margin of 3 mm on the bottom
of the socket (sealing surface) is not damaged.
A plug with a damaged sealing surface must only be used on the HV generator with
transformer oil.
• The silicone oil allows the corona disk and the sealing ring to swell up,
thus guaranteeing that the plug connection is protected against humidity.
If the plug connections are removed again after a longer period of operation
(after 8 days), the corona disk and the silicone sealing ring have to be renewed.
• In the case of high-voltage generators with receptacles in a vertical operating position,
use only transformer oil without corona disk and sealing ring.
• Do not use transformer oil for X-ray tube assemblies.
• When inserting the plugs, the slot and the nose have to be located opposite one another.

WARNING When using force, it is possible that, due to the tolerances, the
plug - turned by 120°- can be inserted into the receptacle despite
the slot and the nose.

When installing the high-voltage cables pay attention


to the minimal bending radius !

At a single permanent installation r ≥ 24 mm


At a fixed installation r ≥ 100 mm
At a free installation r ≥ 165 mm

X-Ray Diagnostic System Register 3 RX0-000.031.01 Page 4 of 6 Siemens AG


Rev. 03 06.97 TD AX 1 Medical Engineering
High-voltage cable RH 098-5DF6 071 1-5

Inserting the HV plug into the tub assembly 1

Fig. 4

Insertion with corona disks 1

• Push the union (Fig. 4/11) over the plug, the thread points to the plug.
• Attach the half-shells over the flange (Fig. 3/36)
with the slot close to the limit stop.
Observe the installation direction, no height adjustment possible!
• Fasten the screw on the half-shells even at both sides (max. 1Nm).
The shielding braid is connected to the ground via the tightened clamp,
important for functioning!
• Clean the insulating piece (Fig. 3/31) on the plug, do not touch with your bare hands.
• Pull the gasket (Fig. 4/13) over the plug up to the flange.
• Moisten the corona disk (Fig. 4/14) generously with silicone oil (Fig. 1/15)
on both sides and push it over the plug contacts.
• Insert the HV plug - pay attention to the nose (Fig. 3/32) and the slot -
into the receptacle until the contact pins are engaged.
• Tighten the union manually by holding the HV cable so that it does not get twisted.
• Tighten both setscrews with the allen key
until slight counter-pressure (max. 1Nm) occurs.
Locking element!

The upper edge of the setscrew may be turned only max. 2mm
from the surface of the union (Fig. 4/11).
If no counter-pressure occurs, turn back the setscrew
until the upper edge is flush with surface of the union.

Siemens AG Register 3 RX0-000.031.01 Page 5 of 6 X-Ray Diagnostic System


Medical Engineering Rev. 03 06.97 TD AX 1
1-6 High-voltage cable RH 098-5DF6 071

Inserting the HV plug onto the HV generator 1

Insertion with transformer oil 1

In the case of HV generators with HV receptacles in vertical position,


the HV plugs are inserted with transformer oil without corona disk and sealing.
If the HV receptacles are not in a vertical or variable position,
the HV plugs are inserted the same way as into the tube assembly, see page 1-6.
• Check the oil level and, if necessary, fill it up.
Minimum oil level without inserted plug = 5 mm or 5 ml.
• Push the union (Fig. 4/11) over the plug, the thread points to the plug.
• Attach the half-shells over the flange (Fig. 3/36)
with the slot close to the limit stop.
Observe the installation direction, no height adjustment possible!
• Fasten the screw on the half-shells even at both sides (max. 1Nm).
The shielding braid is connected to the ground via the tightened clamp,
important for functioning!
• Clean the insulating piece (Fig. 3/31) on the plug, do not touch with your bare hands.
• Insert the HV plug - pay attention to the nose (Fig. 3/32) and the slot -
into the respectable until the contact pins are engaged.
• When the plug is inserted, the oil level must reach at least up to half the height
of the insulating piece and max. up to the upper edge of the plug flange.
• Tighten the union manually by holding the HV cable so that it does not get twisted.
• Tighten both setscrews with the allen key
until slight counter-pressure (max. 1Nm) occurs.
Locking element!

The upper edge of the setscrew may be turned only max. 2mm
from the surface of the union (Fig. 4/11).
If no counter-pressure occurs, turn back the setscrew
until the upper edge is flush with surface of the union.

X-Ray Diagnostic System Register 3 RX0-000.031.01 Page 6 of 6 Siemens AG


Rev. 03 06.97 TD AX 1 Medical Engineering
Changes to previous version 2 2-1

Chapter 0: Revision page and Table of Contents have been updated.

Chapter 1: Page 4: Bending radius changed; wrong: ≤ ; o.k.: ≥

Chapter 2: "’Changes to previous version" has been added.

Siemens AG Register 3 RX0-000.031.01 Page 1 of 2 X-Ray Diagnostic System


Medical Engineering Rev. 02 06.97 TD AX 1
2-2 Changes to previous version

TD AX 4 / Link
TD AX 1 / Kern R.

X-Ray Diagnostic System Register 3 RX0-000.031.01 Page 2 of 2 Siemens AG


Rev. 02 06.97 TD AX 1 Medical Engineering